Easy Electrolytes

Andry Nofrizal,ST
Surakarta,3 Desember 2015

PT SABA INDOMEDIKA

Berdiri pada January 2005


DISTRIBUTOR SYSMEX
DISTRIBUTOR DFI (URINE CHEMISTRY)
DISTRIBUTOR HANTER IT (PNEUMATIC TUBE SYSTEM)
Distributor MEDICA
Distributor ABON
OFFICES : 1 HEAD OFFICE
3 BRANCH OFFICES
11 REPRENSENTATIVE OFFICES

AREA COERAGE

Whole Blood, Serum, Plasma


Na+ Detection Range: 100-200 mmol/L
+ Detection Range: 1.0-10 mmol/L
K
- Detection Range: 50-150 mmol/L
Cl
or
+ Detection Range: 0.20-3.50 mmol/L
Li

Urine
Na+ Detection Range: 10-300mmol/L
+ Detection Range:2.0-200 mmol/L
K
- Detection Range: 15-400 mmol/L
Cl

Specifications Easy Electrolytes


Sample: Whole Blood, Serum, Plasma or Dilute Urine.
Sample Size: 55 uL Sample Container Mode; 50 uL Capillary
Mode; 300 uL Urine Mode.
Method: Direct measurement by Ion Selective Electrode (ISE).
Analysis Time: 35 sec. (Serum) / 60 sec. (Urine).
Data Storage: up to 64 Patient results.
QC - Normal, Low, and High.
Patient ID : 9 Digits.

Electrode
Na+
K+

Cl- / Li+
Spacer
Electrode

REF

Calibration
B
A
Drift
Slope for all channels
Drift

B
A

Calibration
145

135

Calibrant A

130
125

Calibrant B

Concentration (mmol/l)

16
0

15
0

14
0

13
0

12
0

11
0

10
0

90

80

120

70

milliVolts

140

Calibration
145

135
130
125

Concentration (mmol/l)

16
0

15
0

14
0

13
0

12
0

11
0

10
0

90

80

120

70

milliVolts

140

Calculating Patient Results


145

Patient Sample 136.5 mV

135
130

14
0

13
0

12
0

11
0

10
0

90

80

120

Concentration (mmol/l)

16
0

Patient Result
138.5 mmol/L

15
0

125

70

milliVolts

140

Flow Path

Electrode Measurement Diagram

Analyzer Installation

Power Up
- Check Power Voltage
Setting (Set 220 V,
fuse 0.4 A)
- Install Printer Paper
When the Easy Electrolytes analyzer is powered
up, the display will prompt you to set the
correct date and time.

The Easy Electrolytes analyzer has a 24 hour internal clock.

Do not install reagent module or electrodes at


this time !!

Install Electrode

- Prior to installing the electrodes, select


under REPLACE COMPONENTS
SECOND MENU

Install Pump Tubing

Make sure the two middle collars are pushed back


completely on the pump shelf.

Install Reagent Module

When correctly installed, the message


RGNT MODULE IN is printed.

Operation and
Maintenance

Calibrate

Analyze Sample

Analyze Sample

Syringe samples

Capillary samples

Analyze QC

QC Setting
- Home Menu -> Second Menu
Setup Menu Reference Limits
(QC Level1 / QC Level2 / QC Level3)

Levey-Jennings Chart

Second Menu

Analyze Urine

Standby

Replace Components

Diagnostics

Preventative Maintenance

Daily Run Cleaning Solution 2118


Weekly None
Monthly None
Semi-annually Replace reference
electrode, probe wiper and pump tubing

Daily Cleaner
Perform Daily Cleaner after every 50
samples, or once a day when fewer than 50
samples are analyzed.

As preventative maintenance, a more aggressive cleaning


procedure can be performed to remove accumulated
protein within the sampler and preheater (once every 6
months). To perform this preventative maintenance, refer to the Sample
probe/sensor module obstructions procedure found under FLOW in
Troubleshooting.

Daily Cleaner

RECOMMENDED COMPONENT REPLACEMENT SCHEDULE

Flow Problems

Flow Problems

Sensor Module

Flow Problems

Sensor Module
Valve Module

Flow Problems

Sensor Module
Valve Module
Reagent Module

Flow Problems

Sensor Module
Valve Module
Reagent Module
Pump Tubing