FertiOne Manual (30.11.15)
FertiOne Manual (30.11.15)
FOREIGN LANGUAGES
In the event that you are reading this manual in a language other than the English language, you
acknowledge and agree that the English language version shall prevail in case of inconsistency or
contradiction in interpretation or translation.
2 FERTIONE MANUAL
CONTENTS
Use of symbols
Preface
Aims of this manual
General instructions
Safety instructions
When using acid/chemicals
5
5
5
6
Description
Introduction
Advantages
Basic functions
Operating principle
Modularity
Compatibility
Service
Maintenance
System and setup illustration
Dimensions
Weights
7
7
8
8
8
8
8
8
9
11
11
Installation requirements
Electrical installation
Flow rate stability
Source water
12
12
12
On-site preparations
Hydraulic infrastructure preparation
Electrical preparation
Initial operation
13
17
20
21
Operation
Maintenance
Winterization
Calibration
24
Commissioning
25
Troubleshooting
Symptoms regarding the dosing channel
Switchboard warning
Warranty
26
28
29
List of parts
30
31
32
Main parts
Dosing booster
Dosing channels
Appendices
Appendix 1
Appendix 2
Appendix 3
34
34
35
Installation
Unpacking and placement
Hydraulic installation
Electrical installation
18
18
19
FERTIONE MANUAL 3
USE OF SYMBOLS
The symbols used in this manual refer to the following:
WARNING
The following text contains instructions aimed at preventing bodily injury or direct damage to the
crops, the FertiOne and/or the infrastructure.
CAUTION
The following text contains instructions aimed at preventing unwanted system operation,
installation or conditions that, if not followed, might void the warranty.
ATTENTION
The following text contains instructions aimed at enhancing the effective usage of the
instructions in the manual.
NOTE
The following text contains instructions aimed at emphasizing a certain aspect of the operation
of the system or installation.
ACID HAZARD
The following text contains instructions aimed at preventing bodily injury or direct damage to the
crops, the product and/or the infrastructure in the presence of acid.
ELECTRICAL HAZARD
The following text contains instructions aimed at preventing bodily injury or direct damage to the
FertiOne and/or the infrastructure in the presence of electricity.
SAFETY FOOTWEAR
The following text contains instructions aimed at preventing foot injury.
PROTECTIVE EQUIPMENT
The following text contains instructions aimed at preventing damage to health or bodily
injury in the presence of fertilizers, acid or other chemicals.
EXAMPLE
The following text provides an example to clarify the operation of settings, method of operation
or installation.
The values used in the examples are hypothetical. Do not apply these values to your own
situation.
TIP
The following text provides clarification, tips or useful information.
4 FERTIONE MANUAL
PREFACE
Aims of this manual
The aims of this manual are to provide the technician with detailed instructions for installation of the
FertiOne and to provide the user with instructions for its operation and troubleshooting.
General instructions
Installation must be performed by authorized technicians only.
Consult with your supervisor if problems occur during installation procedure.
Installation should be performed on a hard, leveled floor or on a flat, hard, leveled surface.
Do not apply force or pressure on components during the installation procedure.
Verify that field components work properly.
Make sure fertilizers and acid are on site at the time of installation.
Electricity
Ensure that a suitable electrical power supply is available in the vicinity of the installation for the FertiOne
electrical connection (see Electrical preparation, page 17).
Ensure that an electrical socket is available near the FertiOne, for the purposes installation and service.
Safety instructions
All safety regulations must be applied.
Ensure that the installation is carried out in a manner that prevents leaks from the FertiOne,
the fertilizer/acid tank and line, the peripherals and the accessories (leaks could contaminate the
environment, soil or ambient area).
Electrical installation must be performed by an authorized electrician only.
The electrical installation must comply with the local safety standards and regulations.
Protection provided by the equipment is liable to be impaired if the equipment is used in a manner other
than that specified by the manufacturer.
WARNING
In an agricultural environment - always wear protective footwear.
WARNING
Always use protective equipment, gloves and goggles when handling fertilizers, acid and
other chemicals!
WARNING
Measures must be taken to prevent fertilizer infiltration of the water source, in orderto avoid water
pollution.
CAUTION
When opening or closing any manual valve, always do it gradually, to prevent damage to the
system by water hammer.
NOTE
The maximum sound level produced by the equipment does not exceed 70 dB.
FERTIONE MANUAL 5
PREFACE
When using acid/chemicals
ACID HAZARD
When using acid - always observe the acid manufacturer's safety instructions.
WARNING
Always use protective equipment, gloves and goggles when handling fertilizers, acid and
other chemicals!
Nitric
(HNO3)
Phosphoric
(H 3PO4)
Sulfuric
(H2SO4)
Hydrochloric
(HCl)
Hydrogen
peroxide
(H2O2)
Chlorine (as
hypochloride)
Diaphragm
and O-rings
ATTENTION
When dosing acid, use a dosing channel fitted with the appropriate components according to
the type and concentration of the acid used*:
For pH correction
For maintenance of drippers
EPDM
<3%
<85%
<30%
<10%
<30%
<1%
% is by weight at 21 C (70 F)
*The table indicates the resistance of the dosing channel components to acid,
and is not a recommendation to use the acids mentioned.
WARNING
Exceeding the recommended acid concentrations will damage the dosing channel.
WARNING
Substances such as chemicals for pest/disease control might be corrosive and damage the
FertiOne. When using any substance other than fertilizers or acids that do not exceed the
concentrations in the table above, always observe the manufacturer's instructions for corrosivity.
In case of any doubt, contact your local Netafim representative.
6 FERTIONE MANUAL
DESCRIPTION
Introduction
The FertiOne offers a Nutrigation solution for open field sites fitted with electricity where injection of a
single fertilizer solution is needed.
It is a plug-and-play, fully configurable, simple-to-operate single channel fertilizer/acid dosing unit.
It ensures the highest efficiency with minimum investment - a highly cost-effective solution for Nutrigation.
Built on a sturdy, corrosion-resistant aluminum frame, the FertiOne can accommodate different types
of dosing channels and a variety of single- or three-phase dosing boosters to suit the various needs of
different sites.
The FertiOne is available as a basic manual system or equipped with several control options to meet a
vast range of Nutrigation needs:
The MiniAg controller for simple straightforward operation.
The NMC-Junior controller for more advanced quantitative or proportional control.
Other controllers on the market are optional.
The FertiOne accommodates 3 different types of dosing channels for fertilizer or diluted acid:
Manual dosing channel.
Electric dosing channel.
Electric dosing channel with fertilizer meter.
The FertiOne's electrical switchboard is equipped with a three-way switch:
Auto - for operation by a controller
Off - for safe maintenance
Manual - for manual operation or controller bypass
Capacity range
The FertiOne ensures a satisfactory mixture in a vast range of flow capacities.
It will accommodate a 0.1 Ha (0.25 acre) nursery or a 400 Ha (1000 acre) plantation.
Suitable for main line pressure: 2 - 6 bars (29 - 87 PSI) according to the selected dosing booster.
Suction flow rate range: from 100 to 1000 l/h (from 0.44 to 4.4 GPM).
Advantages
A modular Nutrigation system for soil or substrate applications with minimum investment
Efficient usage of water, fertilizers and energy
Unrivaled range of irrigation water capacities
Designed for any application where quantitative or proportional Nutrigation is required
Highly profitable price/performance ratio
Venturi operating principle - no moving parts
Integrates easily into any existing irrigation system
Precise Nutrigation based on a high-accuracy dosing channel
Quick action dosing valves
Can be operated manually or fully computerized
NMC and other controllers can be assembled on the FertiOne for advanced Nutrigation control
A wide variety of accessories and peripherals can be integrated into the FertiOne to enhance its functions
High-quality components and PVC pipe work
Aluminum, corrosion-resistant frame with adjustable legs
Easy to install and to maintain
Made by Netafim
FERTIONE MANUAL 7
DESCRIPTION
Basic functions
The FertiOne supports the following functions:
Manual or controlled dosing of fertilizers/acid with source water to produce a homogenous nutrient solution
Water pre-treatment
Operating principle
The FertiOne doses the fertilizer or acid solution and injects it into the irrigation water
main line. The suction of the fertilizers and acid in the dosing channel is based on the
Venturi principle. This requires a pressure differentiation - supplied by the dosing booster
in the FertiOne.
The low pressure at the Venturi outlet (around 0 bars/PSI) permits the use of a
high-efficiency Venturi with high suction capacity and low motive flow consumption.
Modularity
The modular FertiOne concept is based on an array of interchangeable components that enables rapid
assembly of a wide range of configurations.
Each FertiOne is delivered according to the precise customers order, either fully factory assembled or
assembled by the local dealer.
The dealer stocks the full assortment of the interchangeable FertiOne components.
This concept enables the dealer to assemble any specific FertiOne according to the customers order,
saving the need to stock a large quantity of fully assembled FertiOne units of various configurations.
The modular FertiOne concept ensures prompt delivery schedules without delays!
Stock selection
When operating with the NMC Junior controller, the stock selection option enables the dosing of multiple
fertilizers through the single dosing channel (in cases where simultaneous dosing is not required). Available
in a wide variety of configurations, from 2 fertilizers to as many fertilizers tanks as required.
For further information, contact Netafim.
Compatibility
The FertiOne can be incorporated in any existing or a planned project; in either case it offers a highly
cost-effective solution for Nutrigation.
ATTENTION
Calculations are either in metric or in US units - consistency in the type of units used is essential.
Service
Servicing the FertiOne is a prompt and simple process. The dealer keeps a small quantity of
interchangeable components on hand, for replacement on site within a few minutes.
Maintenance
To prevent failures and extend the life cycle of the FertiOne, the user must carry out regular maintenance,
such as periodic rinsing of filters. For full maintenance instructions, see Maintenance, page 22.
8 FERTIONE MANUAL
DESCRIPTION
System and setup illustration
Front view
Back view
Schematic diagram
11
15
14
12
10
10
13
2
3
4
5
FERTIONE MANUAL 9
DESCRIPTION
Typical dosing channel description and flow scheme
1 Motive flow = the flow of water required through the
Venturi to enable suction of fertilizer/acid.
2 Suction flow = the flow of fertilizer/acid through the Venturi.
3 Total flow = motive flow + suction flow.
1
D
A
B
C
D
E
A
3
Rotameter
Needle valve
Dosing valve
Check valve
Venturi
Typical setup
7
13
8
9
10
FertiOne
10
14
12
15
11
9 Fertilizer/acid filter
2 Pressure gauge
3 Controller
11 Field valve
4 Dosing booster
12 Water meter
6 Check valve
Color code:
10 FERTIONE MANUAL
DESCRIPTION
Dimensions
Without controller
Configuration
Without controller
67/63/99 cm (26.5/25/39")
60/57.5/83.5 cm (24/23/33")
67/63/99 cm (26.5/25/39")
60/57.5/100 cm (24/23/39.5")
75/85/150 cm (29.5/33.5/59")
Weights
Package dimensions
(W/D/H**)
FERTIONE MANUAL 11
INSTALLATION REQUIREMENTS
Electrical installation
An electrical mains that meets the power consumption requirements of the FertiOne should be installed.
The electrical installation must include a circuit breaker and comply with the local safety standards and
regulations.
Source water
The water entering the FertiOne should be within a temperature range of 10C to 35C
(50F to 95F).
The source water to the FertiOne should be of a satisfactory chemical quality. If water
pre-treatment is required, apply chemical conditioning before the water reaches the FertiOne.
12 FERTIONE MANUAL
ON-SITE PREPARATIONS
Hydraulic infrastructure preparation
In order to enable operation of the FertiOne, main line pressure and infrastructure conditions must be
met.
Required proportions
To enable optimal operation of the FertiOne, piping must be installed while maintaining the following
proportions.
FertiOne
H
d
X4
X5
X3 X2
X1
Required proportions
32 - 40 mm (1.0 - 1.25 in)
10 x D
5xD
Minimum 1 m (3.3 ft)
Less than 3 m (10 ft)
Less than 10 m (33 ft)
Higher than 30 cm (12 in)
FERTIONE MANUAL 13
ON-SITE PREPARATIONS
Pump house (Filter house / Fertilizer house) requirements
CAUTION
The FertiOne should be:
placed in a roofed building
protected from direct sunlight
kept at an ambient temperature between 10C and 40C (50F and 104F)
kept at a maximum relative air humidity of 85%
properly ventilated
protected from dust
protected from splashes or direct spraying with water or chemicals
NOTE
In order to prevent penetration of fertilizer or acid into the soil, it is recommended that the
floor of the pump house have a slope of at least 1% towards a gutter at its lower edge, and an
underground tank at the lower end of the gutter, enabling drainage of any spill or excess.
Inlet
7 cm*
(2.75")
*The height may vary by 1 cm (0.5") according to the adjustment of the legs.
14 FERTIONE MANUAL
53 cm*
(20.9")
44 cm*
(17.3")
Outlet
ON-SITE PREPARATIONS
The water distribution system
For the setup of the water distribution system, the following components should be installed
Components of the water distribution system
Item
10 Manual valve (isolation)
Specifications
To be installed at the inlet and at the outlet of the FertiOne, for use
during system maintenance.
11 Irrigation valve
Controllable.
12 Water meter
Suitable for the required pressure and flow rate according to the
configuration and the selected dosing booster of the FertiOne,
and the field requirements (ensure stable pressure).
15 Main line
Should be installed on the main line, between the main line filters and the
water meter and be able to reduce the main line pressure according to the
field requirements.
Flow-meter output
l/pulse
1
10
100
Flow rate
GPM
Up to 88
88 - 1000
1000-4500
Flow-meter output
US gal/pulse
1
10
100
Air-release valve
ATTENTION
An air-release valve must be installed upstream from
the FertiOne at the highest point of the water
distribution system.
In many cases the highest point of the water
distribution system is at the filtration array.
If there is no air-release valve
installed on the water distribution
system or if the highest point is
at the FertiOne inlet pipe,
install the air-release valve as
shown.
FERTIONE MANUAL 15
ON-SITE PREPARATIONS
The fertilizer/acid tanks and lines
For the setup of the fertilizer/acid tanks and lines, the following components should be installed:
Items of the fertilizer/acid tanks and lines
Component
Specifications
A manual ball valve on the fertilizer/acid line at the stock tank outlet
9 Fertilizer/acid filter
16 FERTIONE MANUAL
ON-SITE PREPARATIONS
Electrical preparation
NOTE
In cases where 3-phase electricity is supplied in addition to a single phase, take note:
The controller is single phase.
The dosing booster can be 1-phase or 3-phase.
In most cases, a 3-phase dosing booster is preferable.
Mains connection
CAUTION
Only qualified electricians are permitted to perform electrical installations!
The following components must be provided in the installation:
A readily accessible circuit breaker, rated according to the FertiOne's total rated power and certified as a
branch circuit overcurrent protector (see Appendix 3 - Dosing booster, page 35), compliant with the
national code and requirements.
Grounding connection: 10 .
For the selection of the wire size - take into account (see Appendix 3 - Dosing booster, page 35):
The FertiOne's total rated power.
Whether the electricity supply is 1-phase or 3-phase.
2.5 mm ( 13 awg)
1.5-2
4 mm2 ( 11 awg)
2-3
6 mm2 ( 9 awg)
1 X 200-250 VAC
3 X 200-250 VAC
3 X 400-480 VAC
1.5-2
2-3
FERTIONE MANUAL 17
INSTALLATION
Unpacking and placement
Check the ShockWatch label attached to the
packaging and ensure the indicator is white.
If the indicator is red - follow to the instructions
on the ShockWatch label.
Place the FertiOne package close to the irrigation
system.
Gently open the packaging.
Remove the screws and bolts that connect the
FertiOne to the wooden pallet.
If there is a controller, remove the plastic cover from it.
Place the FertiOne in its position.
Adjust the legs so that the FertiOne is steady.
Hydraulic installation
WARNING
Always use protective equipment, gloves and goggles when handling fertilizers, acid and
other chemicals!
Diameter
1"
Fittings (interchangeable)
PVC, hose nozzle insert connector (installed)
16 mm
1/2"
3/4"
18 FERTIONE MANUAL
INSTALLATION
Before connecting the flexi-tube, thread a stainless steel band clamp A on each flexi-tube hose.
Heat the tip of each flexi tube hose in boiling water to make it flexible for insertion.
Connect the flexi-tube hose from the fertilizer line to the hose-nozzle-insert connector of the dosing
channel B , ensuring 0.5 meter (1.7 feet) of transparent, chemical-resistant, reinforced PVC water hose
with an internal diameter of 16 mm that will not contract when there is a vacuum in the system.
Slide the stainless steel band clamp onto the flexi-tube connection C and secure tightly.
Flexi-tube specification:
16 mm (") reinforced EVA tubing.
Material: EVA (ethylene vinyl acetate) tube compound specially formulated to meet USDA requirements.
Polyester cord reinforcement with color tracer.
Pressure: 18 to 20 bar (250 to 300 PSI).
Recommended temperature range: -5 oC to +50oC (23oF to +122oF).
TIP
You can rotate the hose connectors of the dosing channel to face the
desired direction according to the location of the stock tanks.
Loosen the Rotameter's lower connector A , rotate the hose connector to
the desired direction B and fasten the Rotameter's lower connector A .
Electrical installation
CAUTION
Only qualified electricians are permitted to perform electrical installations!
Have a qualified electrician connect the FertiOne to the mains, following the instructions on the
switchboard electrical diagram enclosed.
FERTIONE MANUAL 19
INITIAL OPERATION
NOTE
The following procedure describes the operations that should be performed, regardless of the type
of controller in use. For the operation of a specific controller interface, see the Controller Manual.
ATTENTION
Make sure all the wire connectors and terminals are tightened before switching on the power!
OPEN
CLOSE
20 FERTIONE MANUAL
INITIAL OPERATION
Switch the dosing booster ON.
Check that the dosing booster is rotating in the correct direction (see the Dosing Booster Manual).
HIGH VOLTAGE ELECTRICAL HAZARD
Before unplugging the system from the main power source,
switch the unit and the main power source OFF!
FERTIONE MANUAL 21
Maintenance
CAUTION
When opening or closing any manual valve, always do so gradually, to prevent damage to the
system by water hammer.
To prevent failures and extend the life cycle of the FertiOne, be sure to carry out regular maintenance.
Keep the FertiOne dosing unit and its immediate environment clean and dry.
The FertiOne dosing unit and the supply water and irrigation system must be inspected regularly.
Regular inspection
Description
Rinsing of fertilizer filters*
Rinsing of supply water filters*
Water and fertilizer leak inspection
How often
Once a day
Once a day
Once a week
Instructions
Visual inspection
22 FERTIONE MANUAL
FERTIONE MANUAL 23
CALIBRATION
Calculation of dosing ratio
For fine calibration of the FertiOne in order to achieve homogeneous and stable dosing, perform the
following calculation to determine the amount of suction reduction needed to attain the required fertilizer/
acid flow rate.
Metric units
EXAMPLE
m3 /hr
X
l/m3
=
l/hr
X 1.25 =
l/hr
120
X
3
=
360
X 1.25 =
450
m3 /hr
l/m3
l/hr
l/hr
DEFINITION
The quantity of fertilizer/acid (l)
Dosing ratio =
1 m3 irrigation water
US units
EXAMPLE
GPM
X
US gal/1000 US gal
X 0.06 =
GPH
X 1.25 =
GPH
500
X
15
X 0.06 =
45
X 1.25 =
56
GPM
US gal/1000 US gal
GPH
GPH
DEFINITION
The quantity of fertilizer/acid (US gal)
Dosing ratio =
1 THG (1000 US gal) irrigation water
NOTE
The Rotameter's scale is calibrated by the manufacturer for measurement of the flow rate of water
(H2O). Certain inaccuracies may be observed when the flow rates of liquids with different densities,
such as fertilizers and acids, are measured.
24 FERTIONE MANUAL
COMMISSIONING
After completing the calibration process, the FertiOne is ready for commissioning.
The user must be familiar with the FertiOne, the user manual and the accompanying documents and pay
special attention the following:
The importance of following all the safety instructions (page 5).
The warranty and its restrictions (page 29).
Calibration - the user must be aware of the importance of having thorough knowledge of the calibration
process for the operation of the FertiOne (page 24).
Proper dissolving of fertilizers for optimal utilization of the dosing channel capacity
(see Calculation of dosing ratio, page 24).
The importance of regular maintenance and its impact on the warranty (page 22).
The importance of periodic comparison of the hydraulic data with the references in the
Hydraulic conditions checklist (page 34).
Make sure the Hydraulic conditions checklist has been properly filled.
Fill out the Commissioning Report.
Keep both documents for future reference and send a copy of them to cmt.support@netafim.com
This manual is provided with the FertiOne
and can be downloaded at https://www.netafim.com/product/fertione-
FERTIONE MANUAL 25
TROUBLESHOOTING
This chapter is a systematic guide to the actions to be taken in the case of a malfunction of the FertiOne.
ATTENTION
Before proceeding to troubleshoot any malfunction, make sure that:
The controller settings for the dosing channel are correct and match the dosing channel
of the FertiOne (see the Controller Manual).
The controller settings regarding the field valves are correct (see the Controller Manual).
Perform the actions in their order of appearance until the malfunction is fixed.
If you identify faulty parts - consult your local Netafim representative.
CAUTION
Only qualified electricians are permitted to perform electrical installations and repairs!
CAUTION
If isolation valves have been installed on the system, ensure that they are in the closed position
before troubleshooting any hydraulic malfunction.
C
F
H
Dosing channel
26 FERTIONE MANUAL
TROUBLESHOOTING
Symptoms regarding the dosing channel
If one or more of the following symptoms occur, perform the actions listed below:
Rotameter reading
Low fertilizer/acid flow rate
Action
1)
2) Check that the manual valve B of the stock tank is in the OPEN position.
3) Check that the fertilizer/acid filter C is clean - if not, it should be dismantled and thoroughly cleaned.
4)
Check the fertilizer/acid line D (from the stock tank to the dosing channel) for leaks and breaches
and make sure all the connectors are tightened.
5) Make sure the needle valve E of the dosing channel is open according to the reference data in the
FertiOne Hydraulic conditions checklist.
6) Visually check the needle valve E for chemical damage (internal deformation).
If internal deformation is present - replace the needle valve.
7)
8) If the dosing channel is equipped with an electric dosing valve, check that the dosing valve F is
functioning:
With the controller in MANUAL mode, set the dosing valve F to ON (see the Controller Manual).
The LED on the dosing valve should light up.
If it does not - have a qualified electrician check the dosing valve's cable for electrical continuity.
If the cable is in working order - check the controller (see the Controller Manual).
If the controller and the cable are in working order - toggle the dosing valve F to OFF and again to ON
in the controller (see the Controller Manual). A "click" should be heard from the dosing valve with each
toggle - If a "click" is not heard, replace the dosing valve (consult your local Netafim representative).
If a "click" is heard and the dosing valve F still does not open - disconnect the dosing valve from the
dosing channel, set the dosing valve to ON in the controller (see the Controller Manual) and check for
clogging by injecting water at low pressure through the dosing valve.
If there is clogging - thoroughly clean the dosing valve F with running water.
If there is no clogging and the dosing valve F still does not open - replace it
(consult your local Netafim representative).
9) Visually check the non-return valve G for any internal deformation or damage to its flat ring gasket.
If deformation or damage are present - replace the non-return valve (consult your local Netafim
representative).
10) Check the non-return valve G for clogging by injecting water at low pressure through it
according to the direction of flow.
If there is clogging - thoroughly clean the non-return valve G with running water.
11) Disconnect the Venturi H (top, bottom and dosing channel connections) and check it for clogging,
visually and by injecting water at low pressure through it.
FERTIONE MANUAL 27
TROUBLESHOOTING
12) Visually check the Venturi H for chemical damage (internal deformation).
If internal deformation is present - replace the Venturi (consult your local Netafim representative).
If after implementing all the above steps the malfunction is still not fixed - consult your local Netafim
representative.
Switchboard warning
WARNING
Only authorized electricians are permitted to perform electrical repairs!
Electrical repairs must comply with the local safety standards and regulations.
If the following symptom occurs during operation, perform the actions listed below:
Action
1)
28 FERTIONE MANUAL
WARRANTY
Netafim warrants all the components of the FertiOne to be free of defects in material and
workmanship for 1 (one) year from the date of installation, provided the installation is reported to Netafim
within 30 days of installation.
If the installation is not reported or is reported later than 30 days from the date of installation, Netafim
will warrant the FertiOne for a period of 18 months from the date of production, according to its serial
number.
If a defect is discovered during the applicable warranty period, Netafim will repair or replace, at its
discretion, the product or the defective part.
This warranty does not extend to repairs, adjustments or replacements of a FertiOne or any part that
results from misuse, negligence, alteration, force majeure, lightning, power surge, improper installation or
improper maintenance.
If a defect arises in your Netafim product during the warranty period, contact your Netafim supplier.
Limited warranty
This warranty is subject to the conditions in Netafim's official warranty statement.
(For the full text of Netafim's official warranty statement, please contact Netafim).
FERTIONE MANUAL 29
LIST OF PARTS
Main parts
2
5
4
8
10
11
30 FERTIONE MANUAL
LIST OF PARTS
Part
2
3
4
5
7
8
9
10
11
Description
MiniAg controller 110 v
MiniAg controller 230 v
NMC Junior controller 115V
NMC Junior controller 230V
NMC Junior controller 115V double door
NMC Junior controller 230V double door
NMC Junior controller 230V Korean
NMC Junior controller 230V Chinese
Aluminum frame for MiniAg controller
Aluminum frame for NMC Junior controller
Switchbox
Inlet to Venturi 1000 l/h
Pressure gauge - 250 GLZ 8 bar (116 PSI) " BSP
Outlet from dosing booster MTX 5-2
Outlet from dosing booster MTX 5-3
Outlet from dosing booster MTX 5-4
Outlet from dosing booster MTX 5-5
Dosing channel
Venturi 1000 l/h
Dosing booster inlet adaptor 1"
Dosing booster
Basic frame
Cat. No.
73240-041100
73240-041090
74340-004983
74340-004982
74340-005150
74340-005160
74340-013892
74340-023900
45000-040510
45000-008758
Not supplied separately
33254-001040
77540-003350
33254-001000
33254-001010
33254-001020
33254-001030
See pages 32-33
33254-001060
33254-001050
See table below
Not supplied separately
Dosing booster
For 50 Hz installations
Phases Voltage (V)
1
230
230/400
For 60 Hz installations
Dosing
booster
M 5-3T
M 5-4T
M 5-5T
M 5-3T
M 5-4T
M 5-5T
Cat. No.
77800-025530
77800-025510
77800-002618
77800-025535
77800-025540
77800-025545
220-230
220/
380-460
Dosing
booster
M 5-2T6
M 5-3T6
M 5-2T6
M 5-3T6
Cat. No.
77800-025520
77800-023250
77800-025525
77800-025515
FERTIONE MANUAL 31
LIST OF PARTS
Dosing channels
Manual dosing channel
7.09
7.04
7.05
7.06
7.07
7.08
7.04
7.03
7.05
7.06
7.07
32 FERTIONE MANUAL
7.08
LIST OF PARTS
7.10
7.11
7.12
7.02
7.13
7.04
Part
7.01
7.02
7.03
7.04
7.05
7.06
7.07
7.08
7.09
7.10
7.11
7.12
7.13
7.14
7.14
Description
Connection between FIP S22 dosing valve and Venturi with EPDM ISO check valve
Dosing valve FIP S22 24VAC 50/60 Hz VITON
Connection between needle valve and FIP S22 dosing valve
Needle valve
Connection between DFM185 ISO Rotameter and needle valve
Rotameter DFM185 100-1000 l/hr (26.5-265 GPH)
Elbow connection " TO DFM185 ISO Rotameter
Hose nozzle end connector for dosing channel
Manual ball valve and EPDM check valve
Connection between fertilizer meter and Venturi with check valve
Fertilizer meter - PVC PB 3/4", EV 1L
Connection between FIP S22 dosing valve and fertilizer meter
Connection between needle valve and FIP S22 dosing valve with fertilizer meter
PVC threaded reducer M*F
7.08
Cat. No.
33240-004600
77540-008500
33254-001090
76400-011376
33240-004400
77540-007580
33254-001120
33240-005250
33254-001110
33254-001070
70240-005300
33254-001100
33254-001080
77400-010180
FERTIONE MANUAL 33
APPENDIX 1
List of configurator items
See the online configurator at https://cmtconfig.netafim.com.
FertiOne
B Dosing channel
Code
M
E
FM
Description
Manual
Electric s22
Electric s22 w fertilizer meter
Frequency
Code Description
50H 50 Hz
60H 60 Hz
D Voltage
Code
1X220
3X400
3X440
Code
WOC
MG1
MG2
J1
J2
J3
J4
JC2
JK3
Description
50 or 60 Hz
Europe 50 Hz
Mexico 60 Hz
Dosing booster
Code
MTX5-2
MTX5-3
MTX5-4
MTX5-5
Description
Matrix 5-2
Matrix 5-3
Matrix 5-4
Matrix 5-5
Controller
Description
Without controller
MiniAg 110V
MiniAg 220V
NMC Junior 110V
NMC Junior 220V
NMC Junior 110V double door
NMC Junior 220V double door
NMC Junior 220V Chinese
NMC Junior 220V Korean
APPENDIX 2
Hydraulic conditions checklist
Consult the main line water meter and pressure gauge
and the FertiOne pressure gauges and make sure all
the hydraulic conditions are in range according to the
following table.
Serial number
Project name
Technician name
Installation date
Notes
Testing date
34 FERTIONE MANUAL
Dosing channel
Q-suction
l/hr / GPM
APPENDIX 3
Dosing booster
Select the dosing booster according to the available main line pressure.
NOTE
When the FertiOne is operated at a main line pressure higher than the suitable range for the
selected dosing booster, the suction of the Venturi gradually decreases.
WARNING
Never operate the FertiOne at a main line pressure lower than the suitable range for the selected
dosing booster. There is a danger of serious damage to the dosing booster due to cavitation.*
ATTENTION
Make sure that the selected dosing booster fits the electricity voltage, phases and frequency on site.
For 50 Hz installations
Phases
Voltage (V)
230
230/400
Dosing booster
M 5-3T
M 5-4T
M 5-5T
M 5-3T
M 5-4T
M 5-5T
Suitable for
main line
pressure range
Bar
PSI
2.0-3.5 29.0-50.7
3.0-4.0 43.5-58.0
4.0-6.0 50.7-87.0
2.0-3.5 29.0-50.7
3.0-4.0 43.5-58.0
4.0-6.0 50.7-87.0
Current (A)
Total power
consumption
of system
(kW)
4.5
5.7
7.8
2.8/1.6
4.3/2.5
5.6/3.2
0.70
0.95
1.35
0.70
0.95
1.35
Current (A)
Total power
consumption
of system
(kW)
6.3
7.7
3.8/2.2-2.2
5.7/3.3-2.9
0.95
1.35
0.95
1.35
For 60 Hz installations
Phases
Voltage (V)
220-230
220/380-460
Dosing booster
M 5-2T6
M 5-3T6
M 5-2T6
M 5-3T6
Suitable for
main line
pressure range
Bar
PSI
2.0-4.0 29.0-58.0
4.0-6.0 50.7-87.0
2.0-4.0 29.0-58.0
4.0-6.0 50.7-87.0
*Cavitation - The formation of vapor cavities ("bubbles" or "voids") in a liquid. It usually occurs when a
liquid is subjected to rapid changes of pressure that cause the formation of cavities, where the pressure
is relatively low. When subjected to higher pressure, the voids implode and can generate an intense
shockwave, causing significant damage to the pump's impeller and chamber.
FERTIONE MANUAL 35