Professional Documents
Culture Documents
Tsfluxus F601v1-5en Leu
Tsfluxus F601v1-5en Leu
Tsfluxus F601v1-5en Leu
FLUXUS F601
Features
Precise bi-directional and highly dynamic flow measurement with the non-intrusive clamp-on technology
High precision at fast and slow flow rates, high temperature and zero point stability
Portable, easy-to-use flow transmitter with 2 flow channels, multiple inputs/outputs, an integrated data logger
with a serial interface
Water and dust-tight (IP65); resistant against oil, many
liquids and dirt
Li-Ion battery provides up to 14 hours of measurement
operation
Automatic loading of calibration data and transducer
detection for a fast and easy set-up (less than 5 min),
providing precise and long-term stable results
User-friendly design
Transducers available for a wide range of inner pipe diameters ( 6...6500 mm) and fluid temperatures
( -40...+400 C )
Probe for wall thickness measurement available
Robust, water-tight (IP67) transport case with comprehensive accessories
HybridTrek automatically switches between transit time
and NoiseTrek mode of measurement when high particulate flows are encountered
QuickFix for fast mounting of the flow transmitter in difficult conditions
Applications
Designed for the following industries:
Chemical industry
Water and wastewater industry
Oil and gas industry
Cooling systems and air conditioners
Facility management
Aviation industry
FLUXUS F601
Technical Specification
Table of Contents
Function ........................................................................................................................................................... 3
Measurement Principle ..................................................................................................................................... 3
Calculation of Volumetric Flow Rate ................................................................................................................. 3
Number of Sound Paths.................................................................................................................................... 4
Typical Measurement Setup ............................................................................................................................. 5
Flow Transmitter ............................................................................................................................................. 6
Technical Data .................................................................................................................................................. 6
Dimensions ....................................................................................................................................................... 8
Standard Scope of Supply ................................................................................................................................ 9
Connection of Adapters................................................................................................................................... 10
Example for the Equipment of a Transport Case ............................................................................................ 11
Transducers................................................................................................................................................... 12
Transducer Selection ...................................................................................................................................... 12
Transducer Order Code .................................................................................................................................. 13
Technical Data ................................................................................................................................................ 14
Transducer Mounting Fixture ...................................................................................................................... 17
Coupling Materials for Transducers............................................................................................................ 21
Connection Systems..................................................................................................................................... 22
Transducer Cable............................................................................................................................................ 22
Clamp-on Temperature Probe (optional) ................................................................................................... 23
Wall Thickness Measurement (optional)..................................................................................................... 24
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
Technical Specification
FLUXUS F601
Function
Measurement Principle
Transit Time Difference Principle
In order to measure the flow of a medium in a pipe, ultrasonic signals are used, employing the transit time difference principle. Ultrasonic signals are emitted by a transducer installed on the pipe and received by a second transducer. These signals are emitted alternately in the flow direction and against it.
As the medium in which the signals propagate is flowing, the transit time of the ultrasonic signals in the flow
direction is shorter than against the flow direction.
The transit time difference, t, is measured and allows the flowmeter to determine the average flow velocity
along the propagation path of the ultrasonic signals. A flow profile correction is then performed in order to obtain the area averaged flow velocity, which is proportional to the volumetric flow rate.
Two integrated microprocessors control the entire measuring process. This allows the flowmeter to remove
disturbance signals, and to check each received ultrasonic wave for its validity which reduces noise.
HybridTrek
If the gaseous or solid content in the medium increases occasionally during measurement, a measurement
with the transit time difference principle is no longer possible. NoiseTrek mode will then be selected by the
flowmeter. This measurement method allows the flowmeter to achieve a stable measurement even with high
gaseous or solid content
The transmitter can switch automatically between transit time and NoiseTrek mode without any changes to
the measurement setup.
t
Path of the ultrasonic signal
t
1
t
2
V
kRe
A
ka
t
tfl
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
FLUXUS F601
Technical Specification
a - transducer distance
a<0
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
Technical Specification
FLUXUS F601
U L T R A S O N IC
F L O W M E T E R
C H A N N E L A
C H A N N E L B
Q
N E X T
Q -
E N T E R
M U X
1
0
O N
5
Q
6
2
O F F
L IG H T
3 x O F F
B R K
D IS P
B A T T E R Y
Q +
E N T E R
D IS P
M O D E
O N
transmitter
analog/binary outputs
analog inputs
Example of a measurement setup in reflection mode
return line
supply line
transducers
temperature probe
supply temperature
temperature probe
return temperature
U L T R A S O N IC
F L O W M E T E R
C H A N N E L A
C H A N N E L B
Q
N E X T
Q -
E N T E R
M U X
O N
5
Q
6
2
O F F
L IG H T
3 x O F F
B R K
transmitter
D IS P
B A T T E R Y
Q +
E N T E R
D IS P
M O D E
O N
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
FLUXUS F601
Technical Specification
Flow Transmitter
Technical Data
FLUXUS
design
measurement
measurement principle
flow velocity
repeatability
medium
temperature compensation
accuracy1
with standard calibration
with extended calibration
(optional)
with field calibration2
flow transmitter
power supply
battery
power consumption
number of flow measuring
channels
signal attenuation
measuring cycle (1 channel)
response time
housing material
degree of protection
according to IEC/EN 60529
dimensions
weight
fixation
operating temperature
display
menu language
measuring functions
physical quantities
totalizer
calculation functions
diagnostic functions
data logger
loggable values
capacity
F601
portable
for transit time difference principle, reference conditions and v > 0.15 m/s
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
Technical Specification
FLUXUS
communication
interface
serial data kit
software (all Windows
versions)
able
adapter
transport case
dimensions
outputs
number
accessories
range
accuracy
active output
passive output
range
open collector
FLUXUS F601
F601
RS232/USB
- FluxData: download of measurement data, graphical presentation,
conversion to other formats (e.g. for Excel)
- FluxKoef: creating medium data sets
RS232
RS232 - USB
500 x 400 x 190 mm
The outputs are galvanically isolated from the transmitter.
see standard scope of supply on page 9, max. on request
output adapter (if number of outputs > 4)
current output
0/4...20 mA
0.1 % of reading 15 A
Rext < 200
Uext = 4...16 V, depending on Rext
Rext < 500
frequency output
0...5 kHz
24 V/4 mA
binary output
26 V/100 mA
optorelay
binary output as alarm output
- functions
limit, change of flow direction or error
binary output as pulse output
- pulse value
0.01...1000 units
- pulse width
1...1000 ms
inputs
The inputs are galvanically isolated from the transmitter.
number
see standard scope of supply on page 9, max. 4
accessories
input adapter (if number of inputs > 2)
temperature input
type
Pt100/Pt1000
connection
4-wire
range
-150...+560 C
resolution
0.01 K
accuracy
0.01 % of reading 0.03 K
current input
accuracy
0.1 % of reading 10 A
passive input
Ri = 50 , Pi < 0.3 W
- range
-20...+20 mA
voltage input
range
0...1 V
accuracy
0.1 % of reading 1 mV
internal resistance
Ri = 1 M
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
FLUXUS F601
Technical Specification
Dimensions
59
FLUXUS F601
C H A N N E L B
Q
N E X T
O N
Q -
E N T E R
M U X
D IS P
5
Q
M O D E
3 x O F F
B R K
E N T E R
D IS P
O F F
L IG H T
Q +
213
U L T R A S O N IC F L O W M E T E R
C H A N N E L A
B A T T E R Y
O N
226
in mm
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
Technical Specification
FLUXUS F601
F601 Energy
F601 Double Energy
F601 Multifunctional
flow measurement on liquids
2 independent measuring channels
temperature-compensated calculation of mass flow rate
integrated heat flow computer for monitoring of energy flows
simultaneous monitoring of simultaneous monitoring of flow measurement taking
into account other process
energy flow and flow,
2 energy flows,
quantities, e.g. density,
e.g. heating systems
e.g. heating systems,
viscosity
heat exchangers)
application
outputs
passive current output
binary output
inputs
temperature input
passive current input
accessories
transport case
power supply unit,
mains cable
battery
output adapter
input adapter
adapter for voltage and
current inputs
QuickFix pipe
mounting fixture for
transmitter
serial data kit
measuring tape
user manual,
Quick Start Guide
connector board at the
upper side of the
transmitter
2
2
2
2
2
2
4
2
2
-
4
-
2
2
x
x
x
x
x
x
x
x
x
-
x
-
x
2
-
x
x
2
2
x
x
x
x
x
x
x
x
x
x
x
x
C H A
C H B
-
C O M M
P 4
P 3
P 2
D C
P 1
- IN
C H
T 1 /T 3
A
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
T 2 /T 4
C H
C O M M
P 4
P 3
P 2
+
O u tp u t
D C
P 1
- IN
C H
T 1 /T 3
A
T 2 /T 4
C H
C O M M
P 4
P 3
P 2
+
O u tp u t
D C
P 1
- IN
C H
T 1 /T 3
P 3 ...P 8
A
O u tp u t
T 2 /T 4
C H B
C O M M
P 2
D C - IN
P 1
+
+
In p u t
In p u t
O u tp u t
In p u t
FLUXUS F601
Technical Specification
O u tp u t
P 6
P 7
P 8
Connection of Adapters
P 3
P 4
P 5
input adapter
output adapter
transducers
measuring channel A
transducers
measuring channel B
C H
P 3 ...P 8
A
T 1 /T 3
T 2 /T 4
C H
C O M M
P 2
D C
P 1
- IN
+
O u tp u t
outputs
In p u t
inputs
RS232
10
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
Technical Specification
FLUXUS F601
coupling compound
transducers
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
transmitter
11
FLUXUS F601
Technical Specification
Transducers
Transducer Selection
transducer order code
FSK
100
FSM
50
FSQ
10
FSS
6
5
recommended
12
150
10
3600
2000
100
25
10
200
6500
3400
400
70
50
100
500
1000
5000
inner pipe diameter [mm]
possible
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
Technical Specification
FLUXUS F601
9...11
12, 13
no. of character
option
extension cable
description
connection system
5, 6
explosion protection
operating
temperature
transducer
frequency
transducer
1, 2
FS
0.5 MHz
1 MHz
4 MHz
8 MHz
N
example
FS
NN
NL
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
000
13
FLUXUS F601
Technical Specification
Technical Data
Shear Wave Transducers
technical type
order code
transducer frequency
inner pipe diameter d
min. extended
min. recommended
max. recommended
max. extended
pipe wall thickness
min.
max.
material
housing
CLK1NZ7
FSK-NNNNL/LC
0.5
CDM1NZ7
FSM-NNNNL
1
mm
mm
mm
mm
100
200
3600
6500
100
200
3600
6500
50
100
2000
3400
mm
mm
1699
5
1699
9
1699
4
mm
mm
mm
126.5
51
67.5
126.5
51
67.5
60
30
33.5
contact surface
degree of protection
according to
IEC/EN 60529
transducer cable
type
length
dimensions
length l
width b
height h
dimensional drawing
CDK1NZ7
FSK-NNNNL
MHz 0.5
operating temperature
min.
C
max.
C
temperature compensation
14
-40
+130
x
-40
+130
x
-40
+130
x
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
Technical Specification
FLUXUS F601
CDS1NZ7
FSS-NNNNL
8
mm
mm
mm
mm
10
25
150
400
6
10
70
70
mm
mm
1699
3
1699
2
mm
mm
mm
42.5
18
21.5
25
13
17
contact surface
degree of protection
according to
IEC/EN 60529
transducer cable
type
length
dimensions
length l
width b
height h
dimensional drawing
CDQ1NZ7
FSQ-NNNNL
MHz 4
operating temperature
min.
C
max.
C
temperature compensation
-40
+130
x
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
-30
+130
x
15
FLUXUS F601
Technical Specification
CDQ1EZ7
FSQ-ENNNL
4
mm
mm
mm
mm
50
100
2000
3400
10
25
150
400
mm
mm
1699
4
1699
3
mm
mm
mm
60
30
33.5
42.5
18
21.5
h
contact surface
degree of protection
according to
IEC/EN 60529
transducer cable
type
length
dimensions
length l
width b
height h
dimensional drawing
CDM1EZ7
FSM-ENNNL
MHz 1
operating temperature
min.
C
max.
C
temperature compensation
16
-30
+200
x
-30
+200
x
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
Technical Specification
FLUXUS F601
7...9
no. of character
outer pipe
diameter
description
fixation
size
4
measuring mode
transducer
transducer
mounting fixture
1, 2
FS
fastening shoes
VP
portable Variofix
TB
tension belts
WL
all transducers
reflection mode
S
small
medium
C
chains
without fixation
010
10...100 mm
025
10...250 mm
055
10...550 mm
150
50...1500 mm
210
50...2100 mm
055
example
VP
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
17
FLUXUS F601
Technical Specification
transducer frequency: M, Q
material: stainless steel 304
(1.4301), 301 (1.4310), 303
(1.4305)
dimensions:
420 x 48 x 68 mm
chain length: 0.5/1/2 m
outer pipe diameter:
max. 150/310/600 mm
transducer frequency: S
material: stainless steel 304
(1.4301), 301 (1.4310), 303
(1.4305)
dimensions:
210 x 32 x 44 mm
chain length: 0.5 m
outer pipe diameter:
max. 150 mm
18
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
Technical Specification
FLUXUS F601
tension belts TB
transducer frequency: K
material: steel, powder coated
and textile tension belt
length: 5/7 m
operating temperature:
max. 60 C
outer pipe diameter:
max. 1500/2100 mm
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
19
FLUXUS F601
Technical Specification
20
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
Technical Specification
FLUXUS F601
<2h
< 24 h
< 3 months
WaveInjector WI-400
< 280 C
coupling foil
type A
coupling foil
type A
coupling foil
type A
280...400 C
coupling foil
type B
coupling foil
type B
coupling foil
type B
Technical Data
type
order code
990739-1
operating
temperature
C
-30...+130
material
990739-2
-30...+200
silicone paste
coupling compound
type N
coupling compound
type E
coupling compound
type H
coupling foil type A
990739-3
-30...+250
fluoropolymer paste
990739-7
max. 280
plomb
990739-8
> 280...400
silver
-10...+150,
short-time peak
max. 200
fluoroelastomer
remark
coupling foil not to be used for transducer mounting fixture with magnets
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
21
FLUXUS F601
Technical Specification
Connection Systems
connection system NL
x
transmitter
transducer frequency
(3d character of transducer
order code)
N
L
1
cable length m
cable length (option LC) m
G, H, K
x
2
2
M, P
y
3
7
l1
25
25
x
2
-
y
2
-
l1
25
-
x
2
-
y
1
-
l1
25
-
x
1
-
y
1
-
l
20
-
Transducer Cable
Technical Data
type
standard length
max. length
operating temperature
sheath
material
outer diameter
cable jacket
material
outer diameter
thickness
color
shield
22
transducer cable
1699
see table above
m
C
-55...+200
extension cable
2551
5
10
see table above
-25...+80
mm
PTFE
2.9
0.3
brown
x
TPE-O
8
mm
mm
black
x
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
Technical Specification
FLUXUS F601
PT12N
670415-1
PT12F
PT12F
670415-2
670414-2
short response time
Pt100
Pt100 matched according
to EN 1434-1
4-wire
-50...+250
(0.15 C + 2 . 10-3 . T [C]), class A
0.1 K, (3K < T < 6 K),
more corresponding to
EN 1434-1
8
PEEK, stainless steel 304 (1.4301), copper
IP66
Pt100
connection
measuring range
accuracy T
accuracy T
response time
housing
degree of protection
according to
IEC/EN 60529
weight (without
connector)
fixation
accessories
kg
0.25
0.5
0.32
clamp-on
plastic protection plate,
insulation foam
15
15
20
14
30
27
B
mm
mm
mm
A
extension cable
0.64
clamp-on
-
dimensions
length l
width b
height h
dimensional drawing
PT12N
670414-1
Connection
Temperature Probe
red/blue
red
white/blue
white
Connector
pin
1
2
3, 4, 5
6
7
8
extension cable
blue
gray
2
4
1
red
white
5
8
3
7
Cable
type
standard length
max. length
cable jacket
m
m
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
extension cable
LIYCY 8 x 0.14 mm2 gray
5/10/25
200
PVC
23
FLUXUS F601
Technical Specification
Technical Data
technical type
DWQ1xZ7
DWP1EZ7
reverse polarity protected
1...200
0.01
1 % 0.01 mm
measuring range1
mm
resolution
mm
accuracy
operating temperature C
-20...+60
cable length
1.5
-20...+200,
short-time peak max. 540
1.2
The measuring range depends on the attenuation of the ultrasonic signal in the
pipe. For strongly attenuating plastics (e.g. PFA, PTFE, PP) the measuring range
is smaller.
DWQ1xZ7
DWP1EZ7
24
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09
FLUXUS F601
Technical Specification
FLEXIM GmbH
internet: www.flexim.com
Wolfener Str. 36
e-mail: info@flexim.com
12681 Berlin
26
Germany
TSFLUXUS_F601V1-5EN_Leu, 2012-01-09