Professional Documents
Culture Documents
Mitutoyo Projector
Mitutoyo Projector
Mitutoyo Projector
Profile Projector
PJ/PV/PH Series
Catalog No.E14005(2)
PJ-H30 Series
Projector PJ-203
(Produced in 1958)
Each Mitutoyo profile projector is a measuring machine that performs measurement, inspection and observation efficiently by
projecting an image of a test workpiece on the stage onto a viewing screen under accurate magnification.
The inherently non-contact measurement method of profile projectors makes this type of instrument highly suitable for measuring
small parts that are unmeasurable with general-purpose contact instruments or easily deformed plastic parts, and can also be used
to observe the surface profiles of workpieces or inspect minute assemblies with surface illumination. Additionally, a wide selection of
accessories allows advanced measurement and inspection of various workpieces.
PJ Series
Screen diameter
300mm
PV
PJ-A3000
P4 - 5
PJ-H30
P6 - 7
PV-5110
Series
P8 - 9
Screen diameter
500mm
Clock components, electronic parts, precious metal parts, precision parts, etc.
PH Series
Screen diameter
350mm
PH-3515F,
PH-A14
P10 - 11
Stages
P12 - 13
Accessories
P14 - 18
Optical terms
basic knowledge
P19
3
PJ-A3000
Dimensions
Unitmm
742
374.9
400
1093
PJ-A3010F-200
854
Focal
point
Technical Data
Projected image:
Protractor screen
Effective diameter:
Screen material:
Screen rotation:
Angle reading:
Cross hairs:
Projection lens:
Inverted
315mm (12.4 )
Fine-ground glass
360
, The counter displays up to 370
.
Digital counter (LED), Resolution: 1 or 0.01(switchable)
Range: 370
, ABS/INC mode switching, Zero Set
90Solid lines
10X (Standard accessory) Optional: 20X, 50X, 100X
External half-reflecting mirror for surface illumination
(only for 10x, 20x).
Bayonet mount
330
408
A
360
A: 95mm (3.74
67mm (2.64
86mm (3.39
233mm (9.17
Lens mount:
Magnification accuracy
Contour illumination: 0.1% or less
Surface illumination: 0.15% or less
Maximum workpiece height: Refer to the projection lenses (H) right.
Contour illumination
Light source:
Halogen bulb (24V, 150W)
Optical system:
Telecentric
Functions:
2-step (High/Low) brightness switch,
Heat-absorbing filter, Cooling fan
Surface illumination
Light source:
Halogen bulb (24V, 150W)
Optical system:
Vertical illumination with adjustable condenser lens
Functions:
Heat-absorbing filter, Cooling fan
Resolution for X/Y counter: 0.001mm or .0001 /0.001mm
(.00005 /0.001mm: digital head)
Power supply:
100 - 240V AC, 50/60Hz, power cord (2m)
Mass:
105kg - 140kg
Power consumption:
Approx. 400W
*For the stage specification, refer to the page 12.
706
28
) ----------- PJ-A3005D-50
) ----------- PJ-A3010F-100
) ----------- PJ-A3005F-150
) ----------- PJ-A3010F-200
Halfreflecting
mirror F
for 20X
10X Lens
50X Lens
20X Lens
Set
100X Lens
Set
Working
W distance
Max. workpiece
H height
Unit: mm
View field
W
H -50 models*
-100 models
-150 models
200 models
D -50 models*
-100 models
-150 models
200 models
10X
31.5
66 (20)
123.5
91
103.5
92.5
224 (198)
182
207 (198)
185
Magnification
20X
50X
15.7
6.3
32.5 (2)
12.6
123.5
123.5
91
91
103.5
103.5
92.5
92.5
87 (61)
27
87 (61)
27
87 (61)
27
87 (61)
27
100X
3.1
5
123.5
91
103.5
92.5
10
10
10
10
System Diagram
Standard scale 50mm / 2
No.172-116 / No.172-117
Green filter
No.172-160-2
Green filter
No.172-160-3
Overlay chart set
No.12AAM027
(12 pcs.)
Rotary table
No.176-106
PJ-A3005D-50
Machine stand
No.172-269
External dimensions
500W830D655Hmm
PJ-A3005F-150
PJ-A3010F-100
PJ-A3010F-200
Stage adapter C
No.176-317
Note: If an optional unit is installed on the stage, the H (Max. workpiece height) length is reduced by the optional unit height.
PJ-A3000 Series
main unit
Lamp replacement
Adjustable stand
No.172-270
515530: High-luminance
halogen bulb of 24V/150W
512305 *3: Long-life halogen
bulb of 24V/150W
(Long-life specification, Rating
approx. 500 hours)
Detector attachmentA
No.12AAE671
OPTOEYE 200
No.332-151
QM-Data200
(Arm type)
No.264-156*1
PJ-H30
Design patent pending in Japan
Powerful PJ-series machine with the ultimate bright and crisp projection image.
Equipped with a high-rigidity main unit and linear scales, this series achieves
high-accuracy measurement.
A total of 8 models are available including one equipped with the long-stroke
stage of 300 x 179mm.
Provided with quick-release wheels for smooth and rapid operation of the stage.
Standard-equipped turret changes the projection lens smoothly and efficiently.
Dimensions
Unitmm
796
Focal point
185~235
155
0~105
PJ-H30A2010B
Adjustment range
480585
995
480
382
Projected image:
Protractor screen
Effective diameter:
Screen material:
Screen rotation:
Angle Reading:
Cross hairs:
Projection lens:
137
Technical Data
Erect
100
306mm (12 )
Fine-ground glass
360
, The counter displays up to 370
.
Digital counter (LED)
Resolution: 1 or 0.01(switchable)
Range: 370
ABS/INC mode switching, Zero Set
Solid lines
10X (Standard accessory)
Optional: 5X, 20X, 50X, 100X
Half-reflecting mirror for surface illumination
Parfocal lens
3-mount turret, Bayonet mount
374
404
283.5
270
351
100
50
745
786
365
16
Lens mount:
Magnification accuracy
Contour illumination: 0.1% or less
Surface illumination: 0.15% or less
Maximum workpiece height: 105mm (when rotary table is not equipped).
Contour illumination
Light source:
Halogen bulb (24V, 150W)
Optical system:
Zoom Telecentric
Functions:
Non-stepped brightness adjustment,
Heat-absorbing filter, Cooling fan
Surface illumination
Light source:
Halogen bulb (24V, 150W)
Optical system:
Vertical / oblique illumination with an adjustable
condenser lens
Functions:
Non-stepped brightness adjustment,
Heat-absorbing filter, Cooling fan
Focusing:
Projection screen head driving
PJ-H30A(manual), PJ-H30D(power drive)
Resolution for X/Y counter: 0.001mm or .0001 /0.001mm
Power supply:
ON/OFF switch, 100 - 240V AC, GND terminal, 50/60Hz,
power cord (2m)
Mass:
176kg - 212kg
Power consumption:
Approx. 420W
*For the stage specification, refer to the page 12.
5X Lens
10X Lens
Max. workpiece
diameter whose peripheral
line can be focused on
the screen center
20X Lens
50X Lens
100X Lens
W Working distance
D
Max.
H workpiece
height
Unit: mm
View field
H
W
D
5X
61.2
105
66
148
10X
30.6
105
70.5
197
Magnification
20X
15.3
105
56.5
137
50X
6.12
105
50
114
100X
3.06
105
50
114
System Diagram
Standard scale 50mm / 2
No.172-116 / No.172-117
Projection lens 5X
No.172-271
Green filter
No.12AAG981
Adjustable stand
No.172-270
Lens cleaning set
No.12AAA165
Size 1010
Size 2010
Stage adapter C
No.176-317
Stage adapter
No.176-304
Size 2017/3017
Note: If an optional unit is installed on the stage, the H (Max. workpiece height) length is reduced by the optional unit height.
Lamp replacement
515530 *3: High-luminance
Detector attachmentA
No.12AAE671
OPTOEYE 200
No.332-151
Foot switch
No.12AAA846
RS-232C cablecross
No.12AAA807
Note) The connectors at
both ends are D-sub
9-pin female.
Adjustable stand
No.172-270
PV-5110
Dimensions
Unitmm
300
68
150
Projected image:
Protractor screen
Effective diameter:
Screen material:
Screen rotation:
Angle reading:
Cross hairs:
0 Line (Index):
Projection lens:
960
Technical Data
1294
1627
PV-5110
Inverted
508mm (20 )
Fine-ground glass
360
, The counter displays up to 370
.
Digital counter (LED)
Resolution: 1 or 0.01(switchable)
Range: 370
ABS/INC mode switching, Zero Set
90Solid lines
Built-in, With a LED back light
10X (Standard accessory)
Optional: 5X, 20X, 50X, 100X
Insert type mount
470
1073
Lens mount:
Magnification accuracy
Contour illumination: 0.1% or less
Surface illumination: 0.15% or less
Maximum workpiece height: Refer to the projection lenses (H) right.
Contour illumination
Light source:
Halogen bulb (24V, 150W)
Optical system:
Zoom Telecentric
Functions:
2-step (High/Low) brightness switch,
Heat-absorbing filter, Cooling fan
Surface illumination
Light source:
Halogen bulb (24V, 150W)
Optical system:
Vertical illumination
Functions:
Adjustable condenser lens, Oblique
illumination (for 5X, 10X and 20X),
Heat-absorbing filter, Cooling fan
Focusing:
Stage part drive, Manual
Resolution for X/Y counter: 0.001mm or .0001 /0.001mm
(using optional KA counter)
100 - 240V AC, 50/60Hz, power cord (2m)
Power supply:
Mass:
190kg
Power consumption:
Approx. 560W
Note) X and Y counters are not built into the PV-5110 main unit. If a counter display is
required, it is recommended that a QM-Data200 or KA-212 is purchased separately
*For the stage specification, refer to the page 13.
5X, 10X, 20X Oblique surface illumination
535
620
PV-5110
5X
Lens set
10X
Lens set
Max.
workpiece H
height
20X
Lens set
50X
Lens set
100X
Lens set
Working
distance
W
D Max. workpiece
diameter whose periphery
can be focused on
the screen center
Stage glass
Projection lens
Unit: mm
View field
H
W
D
5X
101.6
125
60 (27)
120
10X
50.8
181
60
120
Magnification
20X
25.4
206
60
120
50X
10.16
87
32.4
64.8
100X
5.08
87
22.5
45
System Diagram
Green filter
No.172-160-2
PV-5110
No.304-919
Canopy
No.172-319
External view
1200W1500D2000Hmm
KA-212 (Counter)
No.174-146*1
Stage adapter C
No.176-317
Rotary table with
fine feed wheel
No.172-198
Note: If an optional unit is installed on the stage, the W (Working distance) length is reduced by the optional unit height.
*1 To denote your AC power cable add the following suffixes to the order No. :
A for UL/CSA, D for CEE, DC for CCC, E for BS, K for KC, C and No suffix are required for PSE.
Detector attachmentB
No.12AAE672
OPTOEYE 200
No.332-151
Thermal printer
(with connecting cable)*3
Lamp replacement
512305 *4: Long-life halogen
bulb of 24V/150W
PH-3515F, PH-A14
Dimensions
1160
Unitmm
940
1170
460
PH-3515F
PH-A14
Erect*(PH-3515F), Inverted(PH-A14)
135
Technical Data
Projected image:
Protractor screen
Effective diameter:
Screen material:
Screen rotation:
Angle reading:
(for PH-3515F)
424
460
549
928
45
540
PH-3515F
*Refer to the page 13 for the dimensions of PH-A14.
To mount the optional counter (KA-212) and counter stand, approximately
300mm space is required on the right side of the main unit.
Angle reading:
(for PH-A14)
Cross hairs:
90Solid lines
Projection lens:
10X (Standard accessory)
Optional:
5X (PH-3515 only), 20X, 50X, 100X
Lens mount:
Screw mount
Magnification accuracy
Contour illumination: 0.1% or less of nominal maginification
Surface illumination: 0.15% or less of nominal maginification
Maximum workpiece height: Refer to the projection lenses (L1) right.
Contour illumination
Light source:
Halogen bulb (24V, 150W)
Optical system:
Telecentric
Functions:
2-step (High/Low) brightness switch: PH-3515F,
(Lit together with main power activation: PH-A14)
Heat-absorbing filter, Cooling fan
Surface illumination
Light source:
Parabolic halogen bulb (24V 200W: PH-3515F),
(24V 150W, common to the contour illumination: PH-A14)
Optical system:
Vertical illumination
Functions:
Adjustable condenser lens,
Vertical/ Oblique surface illumination selectable: PH-3515F,
(Lit together with main power activation: PH-A14)
Heat-absorbing filter, Cooling fan
Focusing:
Stage part drive, Manual
Resolution for X/Y counter: 0.001mm or .0001 /0.001mm (using optional KA counter)
Power supply:
100 - 240V AC, 50/60Hz, power cord (2m)
Mass:
150kg (PH-3515F), 140kg (PH-14)
Power consumption:
Approx. 410W (PH-3515F), 200W (PH-A14)
Note1) X and Y counters are not built into the projector main unit. If a counter display is
required, it is recommended that a QM-Data200 or KA-212 is purchased separately.
Note2) The indicated value of a measurement may be slightly smaller than the actual value due
to optical distortion caused by the illumination conditions.
* The projected image of the workpiece is erect but inverted horizontally, which means that
the vertical orientation and displacement direction of the image is the same as on the
workpiece side, but the horizontal orientation and displacement direction are reversed.
*For the stage specification, refer to the page 13.
L1
L2
D
H
Half-reecting mirror F
Max. workpiece width
L
Working distance
W
Max.
workpiece H
height
PH-3515F
View field
L
W
D
H
Unit: mm
5X
70.6
175
160 (64)
152.4
152.4
10X
35.3
235
93 (35)
152.4
152.4
Magnification
20X
17.65
235
40
116
152.4
50X
7.06
80
14.6
30.4
152.4
PH-A14
View field
L
W
D
H
10
Unit: mm
10X
35.6
235
93
130
102
Magnification
20X
50X
17.8
7.12
235
109
40
14.6
116
31.3
102
102
100X
3.56
109
9.5
19.2
102
100X
3.5
109
9.5
19
152.4
System Diagram
Projection lens 5X set
No.172-145 (PH-3515F)
Green filter
No.172-286
PH-3515F, PH-A14
main unit
KA-212 (Counter)
No.174-146*2
No.172-001 No.172-002
*2 To denote your AC power cable add the following suffixes to the order No. :
A for UL/CSA, D for CEE, DC for CCC, E for BS, K for KC, C and No suffix are
required for PSE.
Vertical holder
No.172-132
Rotary vice
No.172-144
Center support
No.172-142
PH-3515F
PH-A14
main unit
QM-Data200
Arm type
No.264-156*3
Lamp replacement
515530 *5: High-luminance
Thermal printer
(with connecting cable)*4
Detector attachmentA
No.12AAE671
Adjustable stand
No.172-270
OPTOEYE 200
No.332-151
11
Stage
PJ-A3000
XY range
PJ-A3000
main unit
Model
Order No.
Measuring unit
Quick-release mechanism
Top surface size
Effective size of stage glass
Stage glass thickness
Stage glass No.
Swivel adjustment range
Maximum loading
5050mm
PJ-A3005D-50
302-704*
Digital micrometer head
152152mm
8282mm
5mm
380405
100100mm
PJ-A3010F-100
302-703*
Digital scale
X and Y axes
250250mm
142142mm
5mm
12BAE041
15050mm
PJ-A3005F-150
302-702*
Digital scale
X and Y axes
280152mm
18584mm
6mm
381349
10kg
10kg
8kg
200100mm
PJ-A3010F-200
302-701*
Digital scale
X and Y axes
380250mm
266170mm
6mm
382762
3
8kg
* To denote your AC power cable add the following suffixes to the order No.: A for UL/CSA, -1D for CEE, -1 DC for CCC, -1E for BS, -1K for KC, C and
No suffix are required for PSE.
PJ-H30
XY range
Model
Protractor screen
Order No.
Protractor screen/ Model
OPTOEYE built-in/
motor-driven focusing Order No.
Measuring unit
Quick-release mechanism
Top surface size
Effective size of stage glass
Stage glass thickness
Stage glass No.
Swivel adjustment range
Maximum loading
Measuring accuracy
100100mm
PJ-H30A1010B
303-712-1*
PJ-H30D1010B
200100mm
PJ-H30A2010B
303-713-1*
PJ-H30D2010B
303-732-1*
200170mm
PJ-H30A2017B
303-714-1*
PJ-H30D2017B
303-733-1*
303-734-1*
High-accuracy digital scale
X and Y axes standard
350280mm
410342mm
250150mm
270240mm
6mm
8mm
382762
12BAD363
300240mm
180150mm
6mm
380412
300170mm
PJ-H30A3017B
303-715-1*
PJ-H30D3017B
303-735-1*
510342mm
370240mm
8mm
12BAD330
3 (right)
5 (left)
10kg
20kg
30.02Lm LMeasured lengthmm The measurement method conforms to JIS B 7184.
* To denote your AC power cable add the following suffixes to the order No.: A for UL/CSA, D for CEE, DC for CCC, E for BS, K for KC, C and No suffix
are required for PSE.
12
PV-5110
XY range
PV-5110 main unit
Order No.
Measuring unit
Quick-release mechanism
Top surface size
Effective size of stage glass
Stage glass thickness
Stage glass No.
Swivel adjustment range
Maximum loading
200100mm
304-919*
Digital scale
X and Y axes
380250mm
266170mm
6mm
382762
3
5kg
* To denote your AC power cable add the following suffixes to the order No.: A for UL/
CSA, D for CEE, DC for CCC, E for BS, K for KC, C and No suffix are required for PSE.
PH-3515F
PH-A14
Picture right is PH-3515F
Dimensions of PH-A14
Unitmm
1158
Model No.
PH-3515F
PH-A14
XY range
254152mm
200100mm
PH main unit Order No.
172-868 *1
172-810*2
Measuring unit
Digital scale
Quick-release mechanism
Only X-axis
Top surface size
450146mm
407152.4mm
Dovetail groove
Two (Pitch = 43mm)
Minimum swivel angle
30
reading
Maximum measuring diameter
340mm
300mm
(horizontally fixed)*3
10
454
612
*1 To denote your AC power cable add the following suffixes to the order No.: A for UL/
CSA, D for CEE, DC for CCC, E for BS, K for KC, C and No suffix are required for PSE.
*2 To denote your AC power cable add the following suffixes to the order No.: -10A for
UL/CSA, -20D for CEE, -20DC for CCC, -20E for BS, -20K for KC.
*3 When using the projection lens 10X (Standard accessory) .
13
158
1240
Accessories (Optional)
High contrast color graphic displays on the large LCD screen with LCD back light.
One-key operation for combined measurements that are often used (circle-circle distance, etc.)
Equipped with the measurement procedure teaching function and measuring position navigation in Repeat mode.
Easy measurement using combination of visual cross-hair alignment and automatic edge detection
(Optoeye positioning function).
The AI measurement function (automatic identification of measuring item) eliminates switching between
the measurement command keys.
The user menu function allows user to store measurement commands or part programs to create his/her own menu.
Tolerance zone judgment of data processing result and statistical processing for each item are possible.
Measurement result output to MS-Excel in spreadsheet (CSV) format*
The measurement procedure and measurement result can be saved, using USB memory.**
Two models are available: a stand-alone type with tilt system and a flexible arm type that can be mounted on a
Profile Projector.
Measurement possible even during printout
Specifications
Code
Operation panel
QM-Data200
Stand-mount type
Arm-mount type
264-155 Note1
264-156 Note1
Japanese/English/German/French/Italian/Spanish/Portuguese/Czech/Chinese/
Display languages
Korean/Turkish/Swedish/Polish/Dutch/Hungarian
(selectable)
Measured value unit
Length: mm Angle: degree
Resolution
0.1m / degree-minute-second (selectable)
Program functions
Part program creation, execution, editing
Number of data, maximum value, minimum value,
Statistical processing
mean value, standard deviation, range, histogram,
statistics on a measuring function basis (by command)
Display system
COLOR TFT LCD (with LED backlight)
Edge Sensor Position Compensation
Supported (Projector)
X,Y,Z: Maximum of three Linear Scales
FS: For connecting to optional foot switch
RS-232C 1: For connecting to external PC
PRINTER: For connecting to optional printer
RS-232C 2: For connecting to counter
Input/Output
USB-FD: For connecting to USB-FDD
of measuring instrument
USB-MEMORY: For connecting to
OPTOEYE: For inputting edge signal
USB memory
from OPTOEYE (OPTOEYE M2)
Measurement result
RS-232C output (CSV format, MUX-10 format)
file output
Power
AC100 - 240V
Maximum power
17W (does not include optional accessories)
consumption
External dimensions
Approximately 260242310
Approximately 318153275mm
(WxDxH)
(including the stand)
(when the arm is in the horizontal posture)
Mass
Approximately 2.9kg
Approximately 2.8kg
PJ-2500/PJ-3000 Series
PJ2500/PJ-3000 Series
PJ-A3000 Series, PJ-H30, PV-5110
Applicable models
PJ-H3000 Series, PV600A
PH-3515F, PH-A14
PH-3515F
Standard accessories
AC adapter, power cable, Easy operation guide
Order No.
Note1: To denote your AC power cable add the following suffixes to the order No.: A for UL/CSA, D for CEE, 00 for CCC
(power cord for CCC and User's Manual set of Simplified Chinese are provided for separately), E for BS, K for KC,
C and No suffix are required for PSE.
Note2: Refer to the QM-Data200 and Vision Unit leaet (E14008) for more details.
Measurement result
14
Rotary tables
120
84
172-198
172-198
176-106
176-106
Rotary table
112mm
Order No.
Product Name
Rotary stage size
Fine adjustment
Effective glass diameter
Minimum angle reading
External dimensions
WxDxH)mm
Mass
PJ-A3000
Applicable PJ-H30
models
PV-5110
176-306
176-305
172-198
176-305
176-306
Rotary table with fine feed wheel Rotary table with fine feed wheel A Rotary table with fine feed wheel B
146mm
240mm
270mm
3
3
3
96mm
182mm
238mm
60mm
6
152x152x21.5
240x172x19.7
280x280x23.7
342x342x23.2
1.7kg
3
2.4kg
3
3
3
5.5kg
6.5kg
120
172-378
176-105
120
78
172-197
176-107
176-107
PJ-A3000 Series
Applicable models
PJ-H30 Series
PV-5110
Maximum width to be clamped
0 - 35mm
Mass
0.4kg
Order No.
Order No.
Applicable models
172-234
172-378
PJ-A3000
Series
PH-3515F
PJ-H30 Series
PH-A14 PV-5110
Maximum workpiece
diameter to be clamped 50mm
Central height from a 38 - 48mm
mounting surface
Mass
1.24kg
Adapter
Order No.
Product Name
External dimensions
WxDxH)mm
Mass
PJ-A3000
Applicable PJ-H30
models
PV-5110
176-304
Stage adapter
176-317
Stage adapter (C)
50x340x15
73x278x17
1.5kg
1.8kg
3
3
3
25mm
38 - 48mm
0.8kg
220
Order No.
176-105
172-197
Maximum
70mm
80mm
workpiece size
to be clamped* 45x140mm 65x140mm
10
10
Inclination
Mass
2.4kg
2.5kg
*The maximum possible size to be measured differs
depending on the projection magnification selected.
The size enclosed in parentheses ( ) indicates that for an
inclination of 10
15
15
Accessories (Optional)
Center support
Rotary vise
172-144
172-143
172-142
172-144
PH-3515F
PH-A14
360
76mm
Order No.
Applicable models
Rotation range
Size between mounting
surface and top surface
Minimum angle reading
Mass
5
2.8kg
Order No.
Applicable models
Maximum
workpiece diameter
to be clamped
Mass
Vertical holder
172-142
PH-3515F
PH-A14
120mm240mm
*
Order No.
Applicable models
Height
Mass
172-132
172-143
PH-3515F
PH-A14
60mm
2.2kg
Order No.
Applicable models
Glass size
Mass
172-132
PH-3515F
PH-A14
1.3kg
3.3kg
Standard scale
Reading scale
Glass scale specially designed for inspecting the magnified image of a standard
scale on the projection screen
172-116
Order No.
Range
Graduation
Accuracy
20C
172-116
50mm
172-330
80mm
0.1mm
3+5L/1000m
LMeasured lengthmm
172-117
2
.01
120+5L
10-6
L=Measured length (inch)
172-118
Order No.
Range
Graduation
Accuracy
20C
Order No.
Range
Graduation
Accuracy
20C
16
172-118
200mm
172-161
172-329
300mm
600mm
0.5mm
15+15L/1000
m
LMeasured lengthmm
172-119
8
172-162
12
.02
10-6
600+15L
LMeasured lengthinch
200
PJ-H30D
PJ-H30D
*Mitutoyo s condition
332-151
OPT-200
Contour/surface*
Non direction
2mm
1mm
1000mm/s
30 - 1500X
20X or higher
1m*
Connected to QM-Data200
Adjustable stand
For QM-Data200 (stand-type
specification), thermal printer,
etc.
Machine stand
* Mitutoyo s condition
Order No.
Platform
position
Platform size
17
172-270
Adjustable to a height of
720 to 1020mm
600x450mm
Accessories (Optional)
Overlay charts
To quickly check an image projected on the screen, an appropriate chart is used. 12 types of overlay charts are available according to the application.
Overlay chart No.11
Product name
Order No.
12AAM587
12AAM588
12AAM589
12AAM590
12AAM591
12AAM592
Specification
Set of 12 charts
(Overlay charts No.11 No.22)
Upper side: radial lines (at intervals of 1)
Lower side: concentric circles (at intervals
of 1mm in radius)
Concentric circles (at intervals of 5mm in
radius) with cross hairs (1mm graduation)
Concentric circles (at intervals of 1mm in
radius) with cross hairs
Horizontal: Parallel lines at intervals of
50mm (50-times enlargement of 1mm)
Vertical: Parallel lines at intervals of 20mm
(20-times enlargement of 1mm)
10mm-interval grids
Cross hairs (0.5mm graduation)
Product name
Overlay chart No.17
Overlay chart No.18
Overlay chart No.19
Overlay chart No.20
Overlay chart No.21
18
Order No.
12AAM593
12AAM594
12AAM595
Specification
1mm-interval grids
1
-interval radial lines
Horizontal: 1mm-interval parallel lines
Concentric circles (at intervals of 1mm in radius)
12AAM596
)
and radial lines (at intervals of 1
Metric screw for 20X lens: P = 0.2 to 2mm
12AAM597 Unified screw: 28 to 12 threads/inch
Whitworth screw: 20 to 10 threads/inch
Metric screw for 100X lens: P = 0.08 to 0.25mm
Involute tooth profile for 20X lens
12AAM598 (reference rack tooth profile)
pressure angle: 0.2 to 1
20
pressure angle: 0.2 to 1
14.5
Projection screen
F
An erect image
An inverted image
An optical system based on the principle that the principal ray is aligned
parallel to the optical axis by placing a lens stop on the focal point on
the image side. Its functional feature is that the image will not vary in
size though the image blurs as the object is shifted along the optical axis.
For measuring projectors and measuring microscopes, an identical effect
is obtained by placing a lamp filament at the focal point of a condenser
lens instead of a lens stop so that the object is illuminated with parallel
beams. (See the figure below.)
Principal ray
Optical axis
Light source
(lamp)
Object surface
Condenser lens
Workpiece
Projection lens
Projection
screen surface
Working distance
F Workpiece
X-axis movement
Y-axis movement
Magnification Accuracy
Projection lens
L
M(%)
X 100
M
scale is always viewed from the same direction, ideally from directly
above, the image will appear to shift against the reading scale
M(%): Magnification accuracy expressed as a percentage
graduations and thus cause a measurement error.
of the nominal lens magnification
L : Length of the projected image of the reference object
measured on the screen
: Length of the reference object
M : Magnification of the projection lens
Type of Illumination
PARALLAX ERROR
19
Mitutoyo Corporation