AMBIENT TEMPERATURE CALIBRAION:

The ambient temperature calibration is to be performed upon initial installation, yearly afer
site installation after any major maintenance that might affect the temperature reading, and
at any time thereafter when the sampler fails a verification check following manufacturers
guidelines.
FILTER TEMPERATURE CALIBRATION
The filter temperature calibration is to be performed upon initial installation, yearly after
site installation, and at any time thereafter when the sampler fails eithera single point or
multi-point temperature verification check. To perform the temperature calibration of filter
follows the manufacturers instructions.
PRESSURE VERIFICATION CHECK
Single-point pressure verification must be performed at least onec every month. The
pressure check is performed following manufacturers instuctions.
PRESSURE CALIBRATION
The pressure calibration is to be ferformed upon initial installation, yearly after site
installaion, and at nay time thereafter when the sampler fails a single- point pressure
verification check. Pressure calibraion shall be performed following manufacturers
instructions.
SINGLE- POINT FLOW VERIFICATION CHECK
A single-piont flow verification check must be performed at least every month. The flow
check is performed following maufacturers instructions.
MULTI-POINT FLOW CALIBRATION PROCEDURE
A multi-point flow calibration must be performed upon initial installation and once per
year thereafter. In addition, the multi-point calibration check indicators that the sampler
flow deviates from the flow tansfter standard by more than 4%. The multi-point
calibration is performed following manufacturers instructions.