Room planning

91

Building vibration specifications


NOTE

2.7

Vibrations at the site have the ability to affect the stability and
homogeneity of the magnetic field. As a result, external
vibrations or shocks affecting the magnet may degrade
image quality. In the three spatial orientations, the building
must not exceed the following vibrational specifications.

General information

NOTE

The mass of the floor plate should be about 600 kg/m2 (corresponding to a thickness of about min. 20cm or 8") to
achieve a good vibration and structure-borne sound isolation.

NOTE

The fringe field of the MR system may make its location critical as specified in the planning documentation. Additional
room shielding can be calculated and recommended by the
ES PM PS planning department in Erlangen.
See also: (Magnetic shielding / p. 83)

Siemens, 2007
For internal use only

M4-060.891.01.06.02
07.11

Page 91 of 320
CS SD MR

MAGNETOM

92

Room planning

Vibration specification diagram

Fig. 56: Permissible acceleration

frequency [Hz]

acceleration [m/s2 rms]


Vibration isolators

MAGNETOM

1.0

0.0040

2.0

0.0020

3.0

0.0013

4.0

0.0013

5.0

0.0013

6.0

0.0013

M4-060.891.01.06.02
07.11

Page 92 of 320
CS SD MR

Siemens, 2007
For internal use only

Room planning
frequency [Hz]

93
acceleration [m/s2 rms]
Vibration isolators

7.0

0.0013

8.0

0.0013

9.0

0.0018

10.0

0.0024

11.0

0.0031

12.0

0.0040

13.0

0.0050

14.0

0.0061

15.0

0.0075

16.0

0.0090

17.0

0.0106

18.0

0.0125

19.0

0.0146

20.0

0.0168

21.0

0.0193

22.0

0.0220

23.0

0.0250

24.0

0.0282

25.0

0.0316

26.0

0.0316

27.0

0.0316

28.0

0.0316

29.0

0.0316

30.0

0.0316

35.0

0.0316

40.0

0.0316

50.0

0.0316

Siemens, 2007
For internal use only

M4-060.891.01.06.02
07.11

Page 93 of 320
CS SD MR

MAGNETOM

94

Room planning
Tab. 18

Building vibration specifications

The acceleration of
0.001 m/s2

is
1/10000
or -80 dB(g)

of the acceleration of
gravity
g=9.81 m/s2

The requirement for accelerating amax is measured as maximum rms value per
frequency component < 0.5 Hz in the Fourier Transformation of the recorded signal (spectrum).
The vibration level of continuous vibrations (caused by air condition generator,
compressor, etc.) at the location of the magnet must not exceed the specified values. For all non-continuous transient vibrations, the figures should be multiplied
by 4 (or 12 dB).
The acceleration of 0.001 m/s2 is about 1/10000 (or about -80 dB) of the acceleration of gravity (g=9.81 m/s2).