Professional Documents
Culture Documents
SPE 59532 Electrical Dissipation Factor: A Novel Way To Detect Changes in Shale Water Content
SPE 59532 Electrical Dissipation Factor: A Novel Way To Detect Changes in Shale Water Content
Electrical Dissipation Factor: A Novel Way To Detect Changes in Shale Water Content
P.S. Adisoemarta, L.R. Heinze, Texas Tech U., and S.L. Morriss, U of Texas at Austin
SPE Members
Abstract
The alteration of shales, caused by adsorption of water while
drilling, is one of many contributors to the wellbore stability
problem that costs the industry in the order of $ 400-500
million annually (Bol et.al., 1992). This alteration of shale
problem has acquired a logging perspective due to the
increasing use of measurements while drilling. The capability
of taking real time and time-lapse measurements while still
drilling introduces the possibility of detecting a swelling
problem while something can still be done about it.
The objective of this study (Adisoemarta, 1999) is to
observe the changes of complex electrical characteristic of
shale as a function of water content. The word complex in
the electrical characteristic term means this study will not only
observe the in-phase electrical response but also the out-ofphase response.
By taking the ratio of the in-phase to the out-of-phase
electrical response, this study found that this ratio, the
dissipation factor, changes linearly as a function of shale
water content. This method can be easily applied to both the
drilling / wellbore stability or formation evaluation areas.
Electromagnetic Theory
Maxwells equations and certain constitutive relationships
describe the macroscopic electrical behavior of conducting
dielectrics subjected to a harmonic sinusoidal field.
All electromagnetic fields are created from distributions of
charges and currents in which the electric field (resulting from
the charge distributions), and the current densities (from the
current distributions) are related through the complex transfer
functions that result from Maxwells equations:
(1)
E B / t = 0
H D / t J = 0
(2)
(3)
(4)
(5)
J = E
(6)
B = H
(7)
where
= dielectric permittivity (farad/m)
= electric conductivity (mho/m or siemens/m)
= magnetic permeability (henry/m).
The equation for total current density, JT, is the result of
solving equations (2), (5), and (6):
J T = H = E + dE / dt
(8)
jt
(9)
where
j = (-1)1/2
= frequency (radians / sec),
(10)
=
j
(11)
=
j
(12)
(13)
where eff = ' + " and eff = ' - "/ are real functions of
frequency that constitute the experimentally measured rock
parameters (Fuller and Ward, 1970).
In the simplest form, equation (13) becomes the familiar
Ohms law that relates electric current to the voltage across
the medium and the conductance of the medium itself.
Ohms law is as follows:
I = GV
(14)
where
I = electric current (amperes)
G = electric conductance (mho or siemens)
V = electric potential (volt).
Conductivity and Susceptivity
The output from the measurement equipments are not directly
comparable to each other, and they have to be converted to
conductivity and susceptivity before they can be compared to
each other. For the impedance analyzer, the outputs are
resistance and reactance. The conversions to resistivity,
reactivity, conductivity and susceptivity are calculated using
the equations below:
r=
Rd 2
t
g=
b=
x=
(15)
SPE 59532
Xd 2
t
r
r + x2
2
x
r + x2
2
(16)
(17)
(18)
where
R = resistance (ohm)
X = reactance (ohm)
r = resistivity (ohmmeter)
x = reactivity (ohmmeter)
t = sample thickness (meter)
d = electrode diameter (meter)
g = conductivity (Siemens/meter)
b = susceptivity (Siemens/meter).
For the network analyzer, the outputs are in real and
imaginary (in-phase and out-of-phase) permittivity, and the
conversions to conductivity and susceptivity are as follows:
= eff
g = 0
(19)
b = 0
= eff
(20)
where
g = conductivity (Siemens/meter)
b = susceptivity (Siemens/meter)
' = relative permittivity, real part (dimensionless)
" = relative permittivity, imaginary part (dimensionless)
0 8.854x10 -12 (Farads/meter).
Dissipation Factor
A dielectric parameter that will be used in this research is the
dissipation factor. The dissipation factor is defined as the ratio
of in-phase to quadrature current components. This term
describes the ratio of electromagnetic energy that is given to a
medium to the amount that is lost as heat
D=
(21)
Electromagnetic Measurement
To cover a wide range of frequency, this study uses the fourelectrode parallel-plate setup for the low frequency
electromagnetic measurements, and the open-ended coaxial
probe method for the high frequency electromagnetic
measurements.
SPE 59532
ELECTRICAL DISSIPATION FACTOR: A NOVEL WAY TO DETECT CHANGES IN SHALE WATER CONTENT
Parallel-Plate Method
The material under test is formed as a thin flat sheet and
sandwiched between two electrodes, thus making a capacitor,
as shown in Figure 1. This method is very dependable and
very easy to implement.
The disadvantages of this method are a frequency limit of
around 15 MHz; fringe effect or the stray current path at the
sample edge; necessity for a smooth, flat with parallel faces
sample; and sample maximum thickness of around 10 mm.
Methods for parallel-plate measurement can be separated
into two general categories, which are two-electrode and fourelectrode measurements. Figure 2 shows these two methods.
The two-electrode measurement is where the setup is exactly
as shown in the Figure 1. The sample is placed between two
current electrodes, and these electrodes are also the voltage
electrodes. This method is prone to errors at the lower end of
the frequency, less than 10 kHz, due to the contact impedance
from the ionic build-up at the sample-electrode interface
(Lewis et al., 1986; Garrouch, 1992). The ionic build-up is the
zone where the frequency changes are slow enough that ions
have sufficient time to pile up at the sample-electrode
interface. As the frequency increases, the possibility of buildup diminishes, and at sufficiently high frequency, the ionic
build-up zone vanishes completely. To reduce the effect of
ionic build-up, Scott et al. (1967) used a combination of
platinized platinum electrodes and blotter pads to increase the
surface area at the sample-electrode interface, hence
minimizing the ionic build-up.
In a four-electrode set-up, separate voltage electrodes are
used at each sample face located away from the current
electrodes, outside the ionic build-up zone. As the voltage
electrodes have a high impedance, they will draw almost no
current, so that essentially no ionic build-up will occur at the
voltage electrodes.
Open-Ended Coaxial Probe
The open-ended coaxial probe is basically a cut-off coaxial
line where the sample touches the face of the probe, as shown
in Figure 3. The face of the sample must be flat and smooth.
The electromagnetic field is reflected off the sample, and the
reflection (S11) is related to the complex permittivity of the
sample.
The advantages of this method are that the frequency range
is relatively broad (100 MHz up to 2.6 GHz), the probe is very
small and portable, the probe can work with both solid and
liquid samples, and there is no need to machine the sample for
a perfect fit, such as for the coaxial-method. The disadvantage
of this method is that because the depth-of-investigation is
very small, the reading is mostly affected by a small portion of
the sample and is thus very prone to heterogeneity of the
sample.
Frequency Coverage
Each shale sample in this study will be measured for its
electromagnetic properties in the range from 5 Hz up to 1.3
GHz (1.3 x 109 Hz) using two different measurement setup,
shown in Figure 4. The low frequency range, 5 Hz up to 13
SPE 59532
SPE 59532
ELECTRICAL DISSIPATION FACTOR: A NOVEL WAY TO DETECT CHANGES IN SHALE WATER CONTENT
Conclusions
Dissipation factor plot of shale shows a distinct maxima
and minima.
The location and magnitude of the dissipation factors
maxima and minima decreases as water loss increases
For the case of insitu shale, this study shows the changes
in dissipation factor for both the drying and swelling of
shale.
Dissipation factor characteristics from MWD tools can be
displayed at the drill floor in real time and can be used to
predict the onset of wellbore instability.
MWD tools need to be modified to work at low
frequencies, and also be of variable frequency
By comparing the dissipation factor characteristic from
different well-logs on the same formation type, the
formation water saturation can be qualitatively
approximated.
Nomenclature
B
= magnetic flux density (webers/meter 2 )
b
= susceptivity (Siemens/meter).
D = electric displacement (coulomb/meter 2 )
D = dissipation factor (dimensionless)
d
= electrode diameter (meter)
E
= electric field intensity (volt/meter)
G = electric conductance (mho or siemens)
g
= conductivity(Siemens/meter)
H = magnetic field intensity (ampere-turn/meter)
I
= electric current (amperes)
J
= electric current density (ampere/meter 2 )
j
= (-1)1/2
q
= charge density (coulomb/meter 3 ).
R
= resistance (ohm)
r
= resistivity (ohmmeter)
t
= sample thickness (meter)
V = electric potential (volt).
X = reactance (ohm)
x
= reactivity (ohmmeter)
= electric conductivity (mho/m or siemens/m)
Crystalline Components
Smectite
Illite
Kaolinite
Chlorite Fe
Calcite
Dolomite
Pyrite
Feldspar Na
Feldspar K
Quartz
Weight % in Crystalline
Portion Wellington Shale
0
12
2
1
14
5
4
5
2
56
Material
under test
Electrodes
Voltage Electrodes
Voltage
Electrodes
Ion Buildup
Ion Buildup
Ion Buildup
Ion Buildup
Current Electrodes
SPE 59532
1.0
DC Conductivity from
Schlumberger Chart
Book
chart book
NaCl = 0.37 S/m
KCl = 0.35 S/m
Current Electrodes
Effective conductivity,
eff (S/m)
0.8
0.6
NaCl from
Schlumberger
chart book
NaCl
KCl
0.4
KCl
KCl from
Schlumberger
chart book
NaCl
0.2
High-Frequency Setup
Parallel-Plate
Method
0.0
10 4
10 5
10 2
10 3
10 4
10 5
10 8
10 9
10 10
10 6
10 7
10 8
EPT
MWD
Dual
Induction
Shallow
Laterolog
Deep
Laterolog
10 1
10 7
Frequency (Hz)
Low-Frequency Setup
10 0
10 6
10 9 10 10
Frequency (Hz)
Alcryn ALC-1170BK
15
Sample thickness: 2 mm for parallel plate
25 mm for coaxial probe
Parallel plate diameter: 5 mm
10
35
30
25
Jordan
Marsland
Adisoemarta
-5
20
10 7
10 8
10 9
10 10
Frequency (Hz)
Figure 5: Electrical properties of Methanol from this study
overlaid with two published data.
40
eff = eff / 0
15
eff / 0
Methanol
10
eff = eff /0
eff
Parallel-plate
Open-ended
-5
10 4
10 5
10 6
10 7
10 8
10 9
10 10
Frequency (Hz)
Figure 7: Alcryn ALC-1130BK as standard material used for
calibrating the equipment before and after every measurement.
SPE 59532
ELECTRICAL DISSIPATION FACTOR: A NOVEL WAY TO DETECT CHANGES IN SHALE WATER CONTENT
Deionized Water
Wellington WN4 Absolute Reading
10 4
3
Dissipation Factor
Parallelplate
10 3
10 2
Probe
0
10
Hasted formula
10
10
25
10
Fre
que
ncy
,
Hasted data
10
20
10
Hz
15
10
10
5
10
10 4
10 5
10 6
10 7
10 8
10 9
2 hrs. in Isopar
equilibration
pct
oss,
ater L
ble W
a
e
v
Mo
5min air
5min air
5min air
2 hrs. in Isopar
equilibration
10 10
Frequency (Hz)
Wellington WN4
dessicator
@0.96
Dissipation Factor
4
10000000
3.5
1000000
EM Measurements
Wellington Shale
Assuming 2.4g of water OIP
Minima
3
100000
Frequency (Hz )
2.5
10000
2
Maxima
1000
1.5
100
67
Maxima
25
10
66.8
15
66.7
10
66.6
5
66.5
66.4
10
20
30
40
50
Air Exposure Time (min)
60
70
0.5
Minima
20
Water in place loss (% )
Sample Weight
66.9
Magnitude
eff = eff / 0
10
0
80
10
15
Water Loss (pct)
20
0
25