OPERATIONMANUAL
Page1
OPERATIONMANUAL
ARCFORPROCEDURES
SURGICALANDVASCULARWHA200
MODS.OPESCOPEACTIVOE
OPESCOPEPLENO
Version1.6inPortuguese
Tousetheproductcorrectly,readthroughthe
Thismanualandkeepitas
Futureconsultations.
TheoperatingmanualwasoriginallywritteninEnglish.
https://translate.googleusercontent.com/translate_f
1/135
11/28/2016
OPERATIONMANUAL
Page2
INSTRUCTIONSFORUSE
BowforSurgicalandVascularProceduresWHA200
MODS.OPESCOPEACTIVOANDOPESCOPEPLENO
SHIMADZUCORPORATION
ImporterandDistributorinBrazil:
ShimadzudoBrasilComrcioLtda.
Av.MarqusdeSoVicente,n1771
SaoPauloSP
Authorization/MS:1036901
Manufacturer:
ShimadzuCorporation
1,NishinokyoKuwabaracho
NakagyoKu
Kyoto6048511Japan
InternationalDistributors:
ShimadzuCorporation
1,NishinokyoKuwabaracho
NakagyoKuKyoto6048511
Japan
ShimadzuCorporation
TokyoOffice,3
KandaNishikicho1chome
Chiyodaku,Tokyo1018448
Japan
ShimadzuMedicalSystems,
USACompany
20101SouthVermontAve.
Torrance,California90502
USA
ShimadzuEuropaGmbH
AlbertHahnStrasse610,
47269
Duisburg,FR
Germany
ShimadzuMedical(INDIA)
PVT.LTD.
1stfloor,1063,Munusamy
Salai
KKNagar,Chennai600078
India
ShimadzuMedicalSystems
(OCEANIA)PTY.LTD
UnitsE,1016SouthStreet
Rydalmere,NSW2116
Australia
Shimadzu(HongKong)LTD.
Suite1028OceanCenter
HarborCity,TsimShaTsui
Kowloon,HongKong
ShimadzuInternational
Trading(Shanghai)CO.LTD
BeijingOffice14F
ChinaLifeTower,No.16
ChaoyangmenwaiStreet
ChaoyangDistrict
Beijing100020PR
China
BeijingShimadzuMedical
EquipmentCO.LTD.
No.207LongShengIndustry
ZoneNo.7RongChangDong
street
BeijingEconomicTechnology,
Developmentarea
Beijing100176PR
China
Producttechnicalname:ARCOSURGICAL
TechnicalDirector:ConstantineLuizDiPIPICREA/SP0682526854
ANVISARegistrationNo:10369010062
https://translate.googleusercontent.com/translate_f
2/135
11/28/2016
OPERATIONMANUAL
Page3
ActiveOpescopeSetting:
XRayTube
XrayControllerUnit
Collimator
XrayGenerator
CarHolderwithCBow
ImageIntensifier
TVsystemwithCCDcamera
LCD/TFTmonitor(s)
Monitorstand(s)
Cassetteholder
FullOpescopeSetup
XRayTube
XrayControllerUnit
Collimator
XrayGenerator
CarHolderwithCBow
ImageIntensifier
TVsystemwithCCDcamera
LCD/TFTmonitors
MonitorStand
Cassetteholder9"
Optional
ClearprotectivecoverforCbow(disposablewith
50units/carton)
Transparentcoverformonotank(disposablewith
100units/carton)
TransparentprotectivecoverforImageEnhancer(disposable
With100units/box)
https://translate.googleusercontent.com/translate_f
3/135
11/28/2016
OPERATIONMANUAL
Page3
Sterilizablecoverforarcc,monotankand
images
Sterilizableprotectioncoverformonotankand
7"pictures
Sterilizableprotectioncoverformonotankand
9"pictures
Sterilizableprotectioncoverformonotankand
6pictures
IndicatorLaser(laserpointer)toII7"
IndicatorLaser(laserpointer)for9II"
IndicatorLaser(laserpointer)for6II"
Dualstagetrigger
FibergrillforII7"
FibergrillforII6"
FibergrillforII9"
ControlknobforIl
Cineloopmemory(ACTIVEonly)
DigitalImagingUnitwith
Digitalsubtraction(DSA)(ACTIVEonly).
MODrecorder(ACTIVEonly)
MODmedia(ACTIVEonly)
Videoprinter(thermalprinter)
DVDR/RWrecorder
LCD/TFTwithtouchscreen(touchscreen)
DICOMStorageKit(includedwithCDROMwithsoftware)
DICOMPrintKit(includingsoftwareCDROM)
DICOMWorklistKit(includedwithCDROMwithsoftwareandPC,
ConsistingofMonitor,CPU,MouseandKeyboard)
Fluoroimagedisplayfunctiononthearcscreen(Only
FULL)
XrayDoseMeterDAP(PLENARYOnly)
ExtensionofEmittingSourceSkinDistance.
https://translate.googleusercontent.com/translate_f
4/135
11/28/2016
OPERATIONMANUAL
Page5
StorethismanualcarefullyalongwiththeOpescope.
Thesymbolsshowninthismanualfollowbelow.
Alertthatindicatesdirectdangerandmaycausedeathorseriousinjuryif
Notbeavoided.
Alertindicatingapotentialhazardwhichcouldcausedeathorserious
Damageifnotavoided.
Hazardwarningthatcancauseslightdamagetothepatientordamagetothe
Equipmentorfireifnotavoided
Alertswithinformationthatcanhelpyouusethesystemcorrectly.
IndicationofReferences
ProvidesinformationtomakeiteasiertouseOpescope
ProvidesinformationforreferencewhenOpescopeisused.
AbbreviationsandSpecialTerms
ReferringtoFig.
https://translate.googleusercontent.com/translate_f
5/135
11/28/2016
OPERATIONMANUAL
Page6
Introduction
WethankyouforchoosingtheArchforSurgicalandVascularProcedures
OPESCOPEACTIVO/OPESCOPEPLENO.
Beforeusingtheproduct,readtheoperationmanualandmakesurethe
Understandingofitscontent.
Afterreadingthismanual,storeitinasafeplace
Accessedeasily.
Copyright
ShimadzuCorporationownsthecopyrightsinthismanualandpartof
Contentorallofitcannotbecopiedorreproducedwithoutdue
AuthorizationfromShimadzuCorporation.
Thecontentsofthismanualcanbemodifiedforimprovementswithoutany
Notice.
ShimadzuCorporationAllrightsreserved.
https://translate.googleusercontent.com/translate_f
6/135
11/28/2016
OPERATIONMANUAL
Page7
OperatingPrecautions
1.Onlyqualifiedpersonnelshouldoperatetheequipment.
2.Wheninstallingtheequipment,payattentiontothefollowingitems:
(2.1)Donotinstalltheequipmentnearwatertapsorsimilar.
(2.2)Installawayfrompotentialsourcesofpressure,temperatureor
Moisture,directsunlight,hydrochloricorsulfurdustorgases.
(2.3)Duringtransportandoperationoftheequipmentavoidvibrations,
Inclinations,shocksandimpacts.
(2.4)Keepequipmentawayfromareaswheregasesorchemicals
Arestored.
(2.5)Useonlytherecommendedpowersource(frequency,currentand
voltage).
(2.6)Checktheconditionsofthebattery'spowersource(powerand
Polarity)beforeoperatingtheequipment.
(2.7)Properlygroundtheequipment.
3.Beforeoperatingtheequipment,payattentiontothefollowingitems:
(3.1)Checktheconditionsofthesockets,polarity,etc.
(3.2)Checkthatgroundingisconnectedcorrectly.
(3.3)Checktheelectricalinstallation.
(3.4)Payattentionwhenusingmorethanoneelectricalequipment,
Maycauseanincorrectdiagnosisorcausedamage.
(3.5)Checktheconditionoftheexternalelectricalcircuit,whichwillbe
Connecteddirectlytothepatient.
(3.6)Checktheconditionofthebatterypowersource
4.Whileoperatingtheequipment,payattentiontothefollowingitems:
(4.1)Donotexceedexposuretimefordiagnosisortherapy.
(4.2)Observethepatientandtheequipmentconstantlyfordetection
Problems.
(4.3)Whenaproblemisdetectedwiththeequipmentorpatient,
https://translate.googleusercontent.com/translate_f
7/135
11/28/2016
OPERATIONMANUAL
Page8
Takethenecessarystepstostoptheequipmentwithout
Hurtthepatient.
(4.4)Donotlettheequipmenttouchthepatient.
5.Afteroperatingtheequipment,payattentiontothefollowingitems:
(5.1)Turnoffthekeysandreturnthedialtoitsoriginalposition.Then,turnoff
Themainkey.
(5.2)Donotpullthepowercordviolentlyfromtheoutlet.
(5.3)Whenstoringtheequipment,payattentiontothefollowing
items:
(I)Keepawayfromwater.
(Ii)Storeawayfrompotentialcausesofproblemssuchas
Abnormalpressure,temperatureandhumidity,airflow,sunlight
Direct,dust,chlorineorsulfurgas.
(Iii)Duringtransportationandstorageofequipment,avoid
Inclinations,vibrationsandimpacts.
(Iv)Storeequipmentawayfromstorageareas.
Chemicalsandgases.
(5.4)Cleanallaccessories,cables,andcontacts,andstoretheminalocation
Appropriate.
(5.5)Keepequipmentcleantoavoidproblemsduringoperation.
Nextuse.
6.Whenequipmentisbroken,donotattempttorepairit.
Properlyidentifyandcontacttheauthorizedservicecenter.
7.Donotmodifyanypartoftheequipment.
8.PreventiveMaintenance
(8.1)Theequipmentanditspartsshallbeperiodicallytested.
(8.2)Iftheequipmentisnotinoperationforanextendedperiod,
Testbeforeusetoensure
Evenifitisworkingproperlyandsafely.
9.Forotheritems,usethemaccordingtotheoperatinginstructions.
ReferringtoFig.
Page9
https://translate.googleusercontent.com/translate_f
8/135
11/28/2016
OPERATIONMANUAL
Usingthesystemsafely...
Onlyqualifiedpersonnelshouldusethesystem.
Theresponsibilitytocontroltheuseandmaintenanceofmedicalequipment
Fallsontheuser(s)ofthesystem.
Onlyqualifiedpersonnelortheirsubordinatesmayusethesystem.
Itisdangeroustocarryoutinternalrepairsorinspections,
Ifrepairs/inspectionsarenecessary.
Donotmodifythesystem.
Ingeneral,allmodificationsarestrictlyprohibitedbytherequirements
Regulatoryframeworks.Pleasecontactyourauthorizedservice
Anymodification.
Performperiodicinspections.
Tomaintainsystemsecurityandperformanceforalongtime,
Performpreventivemaintenance.
Repairsandmaintenanceoftheequipmentcanonlybeperformed
Technicalassistanceorauthorizedtechnicians
ReferringtoFig.
Page10
https://translate.googleusercontent.com/translate_f
9/135
11/28/2016
OPERATIONMANUAL
MainPointstoWatch
Beforeusingtheproduct,readtheoperationmanualandmakesurethe
Understandingofitscontent.
Theuseofthesystemwithoutreadingthismanual,orwitha
Incompleteunderstandingoftheinformationofitscontents,cancauseserious
Damagetotheequipment.
Donotturnonthepowerortouchthebuttonsorothercontrolswithoutfirst
Havereadtheinstructionmanual.
IfyoudonotunderstandanypartoftheInstructionManual,please
Contactyourauthorizedservicecenter.Contactdetailswillbe
Theendofthemanual.
Storeitinasafeplacethatcaneasilybeaccessedwhen
Necessary.
Familiarizeyourselfwiththelocationsofrisksandthenatureoftheserisks.
Makesureeverythinghasbeendonetoavoidsuchrisksandactions.
Correctivemeasurestobetakenintheeventofsomeofthese
Happen.
Onlyqualifiedpersonnelshouldusethesystem.
Theresponsibilitytocontroltheuseandmaintenanceofmedicalequipment
Fallsontheuser(s)ofthesystem.
Onlyqualifiedpersonnelortheirsubordinatesmayusethesystem.
Itisdangeroustocarryoutinternalrepairsorinspections,
Ifrepairs/inspectionsarenecessary.
Donotmodifythesystem.
Ingeneral,allmodificationsarestrictlyprohibitedbytherequirements
Regulatoryframeworks.Pleasecontactyourauthorizedservice
Anymodification.
10
Page11
https://translate.googleusercontent.com/translate_f
10/135
11/28/2016
OPERATIONMANUAL
LocationofRisks
HighVoltageSection
Nevertouchtheinternalpartsoftheequipment,asitcontains
Highvoltagesectors.
Thiscouldresultinseriousaccidentsincludingdeathby
electrocution.
Figure1HighVoltageSection
Inspectthecableusedinthesystem.
Checktheinsulationofthecableoranydamagebeforeusingit.
inthesystem.
Insulationfaultscancauseseriousaccidentsincludingdeath
byelectrocution
11
Page12
https://translate.googleusercontent.com/translate_f
11/135
11/28/2016
OPERATIONMANUAL
MovingSection
Beawareofhazardswhentouchingequipment
Youcaninjureyourarmsorlegsbypositioningthembetween
Carmandtherestofthestructure.
WhenmovingtheCarm,grasponlyattheappropriatepointsand
Operatecarefully.
Donotplaceanyperipheraldevicesoranyother
Intherangeofmovementoftheequipment
Thelaserbeamgeneratedhasawavelengthof630nm
(Nanometers)andtheoutputislessthanmWandconsequentlyis
CharacterizedasaClassIIlaserbytheFDAandasalaser
Class3AbytheIEC.
Figure2RangeofMotion
12
Page13
https://translate.googleusercontent.com/translate_f
12/135
11/28/2016
OPERATIONMANUAL
PreventiveMaintenanceNotes
PerformPeriodicInspections.
Preventivemaintenancewillberequiredtomaintainsafety
Andsystemperformance.
Maintenanceshouldonlybecarriedoutbytechniciansor
Authorizedtechnicalassistance.
Repairsandmaintenanceshouldonlybeperformedby
Techniciansand/orauthorizedtechnicalassistance.
RadiationCare
Beawareofxrayirradiation.
Iftheequipmentisnotusedproperlybytheuser,the
Patientandothersmayreceiveahigherdoseof
Radiationthannecessary.Nooneelse,exceptthepatientandthe
Personsstrictlynecessaryfortheprocedure(staff
Doctor)shouldremainintheexaminationroomduring
Exposureofxrays.Ifforsomereasonanotherperson
Needtostayinsidetheexaminationroom,itshouldtakethe
Measurestoprotectagainstradiation.
Tominimizetheabsorbeddose,focusthe
Asfaraspossible.
13
Page14
Careofthelaserbeam
https://translate.googleusercontent.com/translate_f
13/135
11/28/2016
OPERATIONMANUAL
Thelaserbeamgeneratedhasawavelengthof630nm
(Nanometers)andtheoutputislessthanmWandconsequentlyis
CharacterizedasaClassIIlaserbytheFDAandasalaser
Class3AbytheIEC.
Neverlookdirectlyatthelaserbeam.
Whenlookingatthelaserbeam,theeyescanbeinjured.
Turnontheequipmentonlywhennecessarytoavoid
Unnecessaryexposure.
Donotattemptanyoperation,adjustmentorprocedurethatdoesnot
Describedinthismanual.
Explosioncare
Equipmentnotsuitableforuseinthepresenceofa
Inflammableanestheticmixturewithair,oxygenor
nitrous.
Takecarewhenusingdisinfectantspray.
Donotusespraydisinfectantsthatareflammableastheymay
Causeexplosions.
Vaporfromadisinfectantspraymaytriggeranexplosion
Causingdeaththreatstothesurroundingpeopleanddamageto
system.
14
Page15
FlameCare
https://translate.googleusercontent.com/translate_f
14/135
11/28/2016
OPERATIONMANUAL
Implementmeasurestoeliminateflames.
Usingthesysteminanunsuitableenvironmentmaycause
Fireorexplosion.
Installappropriatefireextinguishersandensurethatthe
Procedurefordealingwithfiresiswidelyknown.
Usefireextinguishersuitableforfire
Chemicalandelectrical.
Usingwaterorotherliquidsinfirefighting
Cancauseseriousdamage.
Cellphonerelatedcare
Keepcellphonesorthelikeoutoftheroom.
exam.
Donotplacedevicesthatemitelectromagneticwaves(eg
Cellphones)withintheexaminationroom,evenif
Off.Suchdevicesmayexceedthewave
Electromagneticfields,andundercertainconditionsmay
Functioningofthesystem.Intheworstcase,itcancauseserious
Damagesorclinicalerrors.
15
Page16
Careaboutfieldcompatibility
electromagnetic
https://translate.googleusercontent.com/translate_f
15/135
11/28/2016
OPERATIONMANUAL
TheWHA200requiresspecialprecautionsregardingEMC
(ElectromagneticCompatibility)andneedstobeinstalledand
UsedinaccordancewiththeEMCinformationprovidedinthis
Operatingmanual.
Checkthattheelectromagneticfieldsarecompatible.
Allperipheraldevicesmustconformtothe
EMCstandardsfortheemissionofelectromagneticwavesand
Sensitivitytoemitelectromagneticwaves.Devicesthat
Donotmeetthesestandardsmaycausethe
Thesystem.Intheworstcase,itcouldcauseseriousdamageor
Errors.
TheWHA200shouldnotbeusedtooclosetoorstackedwith
Otherequipment,butifuseinthiswayisnecessary,the
WHA200shouldbeconstantlymonitoredtoverifythe
Normaloperationintheconfigurationwhichwillbeused.
16
Page17
Cleaninganddisinfectioncare
https://translate.googleusercontent.com/translate_f
16/135
11/28/2016
OPERATIONMANUAL
Cleaninganddisinfectionofthesystem.
Cleaninganddisinfectionareveryimportanttoensurethat
Thesystemwillbecleanandsafetouse.
Disconnectthemainpowercablebeforestartingthe
cleaning.
Itisstrictlynecessarytoswitchtheunitoffbeforestartingthe
Cleaninganddisinfecting.
Otherwise,thereisariskofelectricshock.
Donotusewaterinthesystem.
Cleanthesystembywipingwithaclothmoistenedwithdisinfectant.
Usingwatertocleanthesystemmaycause
Electricalinstallations.
Donotapplydisinfectantorspraydirectlyonthe
equipment.
Ifdisinfectantentersthesystem,itmaycauseproblemsor
Accidents.
17
Page18
Cautiononuseandstorage
https://translate.googleusercontent.com/translate_f
17/135
11/28/2016
OPERATIONMANUAL
Installasatisfactoryground.
Thesystemrequiresgrounding.
Unsatisfactorygroundingmaycauseelectricshock.
Usetheequipmentinanenvironmentthatmeetsthe
Guidelinesaddressedhere.
Usetheequipmentonlyinanareaorroomthatis
Accordancewithgovernmentalandelectricallysafelawsforthis
kindofequipment.
Conditionsforproperuseandstorage.
Thelocationwheretheequipmentwillbeusedandstoredshall
Followingconditions:
Atmosphere:Cannotbeexplosiveorcontaincorrosivegases
Temperature:10Cto40C
Relativehumidity:30%to80%(noncondensing)
Atmosphericpressure:700hPaat1060hPa
Note:theenvironmentshouldnotundergosudden
Humidityortemperature,asthismaycausesystemproblems.
18
Page19
Caringforthesystem
https://translate.googleusercontent.com/translate_f
18/135
11/28/2016
OPERATIONMANUAL
Becarefulofimpactsorsystemcrashes.
Thesystemweighsmorethan200kg.
Impactsandfallswilldamagethesystemandcauseseriousdamageto
Peoplewhoaremanipulatingit.
Whenmovingthesystem,payattentiontothefollowingpoints:
CheckthattheCarmisinthetransportposition.
Whenyouarecrossingrampsorentrances,and
Movethesystemassmoothlyaspossible.Becareful
Particularlyonsteepramps.
Alwaysusethebrakewhenstoppingthesystem.
Onrampswithaslopegreaterthan5,thesystemmayfall.
Donotmoveoroperatetheequipmentinlocationswherethe
Exceed5.
Whenmovingthesysteminplacesthatgeneratestrongvibrations,
Outdoors,avoidrapidprogress,stop
Abrupt,andsteepramps.Donotstoptheequipmentonroutes
Sloping.
Whilemovingthesystem,becareful
obstacles.
19
Page20
WarningLabels
Donotremoveordamagethewarninglabels.
Warninglabelsindicateimportantcarewhenusing
thesystem.
https://translate.googleusercontent.com/translate_f
19/135
11/28/2016
OPERATIONMANUAL
Ifthelabelisunreadablecontacttheauthorizedservice.
Figure3WarningLabels
20
Page21
CAUTION
Fordriving,turntheCArmtothe
Position.
https://translate.googleusercontent.com/translate_f
20/135
11/28/2016
OPERATIONMANUAL
CAUTION
CRUSHING
Donotputyourhandhere.
WARNING
ThisXrayunitcanbedangerous
Forthepatientandoperatorunlessthe
Safeexposurefactors,exposure
Operationandmaintenancescheduling
Areobserved.
WARNING
RISKOFEXPLOSION
Donotuseinthepresenceofanesthetics
Flammable.
CAUTION
RISKOFFALL
Donotusetheequipmentonafloor
Withaslopegreaterthan5degrees.
CAUTION
RISKOFELECTRICSHOCK
Keeptheequipmentdry.
CAUTION
LASERRADIATION
DONOTLOOKFORLASER
1.0mWMax.CW645nm
ClassIILaserProductIEC608251
21
Page22
Warrantyterm
ShimadzudoBrasilComrcioLtdaguaranteesthenormaloperationfor
Allequipmentsuppliedbyit,againstdefectsorfailuresthatunderuse,
Maintenanceandoperation,occurdueto
Ormaterial,foraperiodoftwelve(12)months,countedfromthedateof
Terminationoftheinstallationandconsequentacceptanceofthe
Maynotexceedfifteen(15)months
OftheBillofLadingorBillofSale,
https://translate.googleusercontent.com/translate_f
21/135
11/28/2016
OPERATIONMANUAL
Whicheveroccursfirst.Thebuyershallcommunicateinwritingtothe
Shimadzu,anydefectsthatmayoccur,withtheevidence
Satisfactory,assoonaspossible.
Defectiveparts,whichareundertheWarrantyperiod,willbe
ReplacedorrepairedatShimadzu'sdiscretionandwithoutchargetothepurchaser.
ThecostsrelatedtothestayandtraveloftheEngineersand/or
Shimadzuanditsrepresentatives,whentheequipmentisinstalled
OutsidethecityofSoPaulo,oroutsidethehostcityofoneoftheRepresentatives
Aswellasotherexpensesnecessaryforsuch
Becoveredbythebuyer.Thereplacementofpartswillbedoneinthesame
Purchaseofequipment.Therepairorreplacementofpart
Equipmentdoesnotextendorinterruptthewarrantyperiod.AShimadzu
Reservestherighttorequestthereturnofreplacedpartsandcomponents
UnderWarranty.
ExceptfortheGuarantee,anyandallobjectssubjecttothe
Deterioration,naturalwearorconsumption,suchas:accumulators,batteries
Droughts,rubberorplasticobjects,includingprotective,filmandphoto
Chemicalproducts,incandescentbulbs,cablesingeneral,materials
Processing,etc.
Defectsordamagesresultingfromaccidentsarealsoexcluded.
Suchas:fires,floods,electricpowerlines,shocks
Maintenance,cleaning,improperuseof
Equipmentornegligence.Alsonotincludedinthiswarranty
Anyindemnityforlossofprofit,personalinjuryandotherproperty.
22
Page23
Chapter1
https://translate.googleusercontent.com/translate_f
22/135
11/28/2016
OPERATIONMANUAL
WHA200Overview
OPESCOPEACTIVE/FULL
ThischapterprovidesanoverviewoftheWHA200OPESCOPEACTIVEand
Describestheenvironmentalcharacteristicsandconditionswhichshouldbeused.
23
Page24
1.1Outline
Thisequipmentisamobileimagingsystemforradiographyand
Fluoroscopyforuseduringsurgery,
AswellasinXrayexaminationrooms.
Pleaseattachtheoptional"ExtensionKitforSourceSkinDistance"
Whentheequipmentisusedforanotherapplicationthan
Forsurgerypurposes.
https://translate.googleusercontent.com/translate_f
23/135
11/28/2016
OPERATIONMANUAL
1.1.1AbouttheWHA200Active/FullOpescope
Innormalconfiguration,themonitorstandhasamonitor
(Forlivefluoroscopyonly).Theexamplebelowreferstothe
Configurationwith2monitors,whichoneisoptional.
Figure1.1SystemOverview
24
Page25
1.2Features
Thisunitincorporatesthefollowingsupportfeaturestoenable
Radiographyandsmoothedfluoroscopywhileensuringfunctionalityand
Requiredintheoperatingroomsand
Emergencies.
Highimagequality
ThesystemincorporatesaCCDcameraandaveryhighxraytube
Capacityandhighdissipationforhigh
Withoutexposuretimelimits.
Lowirradiation
https://translate.googleusercontent.com/translate_f
24/135
11/28/2016
OPERATIONMANUAL
Inthefluoroscopypulse,xraysareemittedasapulse.ITIS
Possibletochoosethelowdosemodethatminimizesirradiationand
Highqualitypicturemodeaccordingtoyourneeds.
The"LastImageHold"functionis
Forcaseswherefluoroscopyisnotrequiredforconfirmation.
Eightimagescanbestoredinmemory.
Speed
Thefreedomofmovementfacilitatesthepositioningofthesystemto
Workonanylimborspine,whichhelpsreducetime
Positionandthetimerequiredforsurgery.
Lightweightandcompact
Thesystemwasmadewithfreeandcompactangle.
Cleaning
ThecablesareembeddednexttotheCarm,sotheydonotgointo
Contactwiththepatientorsurgicalinstruments,allowingthesafeuseof
ofthesystem.
LastImageMaintenance
(LIH)
ThelastXrayimageisdisplayedonthemonitor.
25
Page26
1.3EnvironmentalConditions
1.3.1ConditionsofUse
Inorderfortheequipmenttoachievetheintendedperformance,
Itmustbeusedundertheconditionslaiddown.
Environmentalconditions
Usetheequipmentinanenvironmentwiththe
Specifiedbelow.
https://translate.googleusercontent.com/translate_f
25/135
11/28/2016
OPERATIONMANUAL
Evenwhenthefollowingconditionsaremet,
Makesurethattherearenosuddenchangesin
Temperatureorhumidity.
Thisleadstocondensation,whichcancauseproblems.
Condition
Stipulation
Temperature
10Cto40C
Relativehumidity
30%to85%(noncondensing)
Atmosphericpressure
700hPaat1060hPa
Atmosphereconditions
Theatmospheremustbefreeof
Corrosiveandexplosivegases
Powersupply
Thepowersourcemusthavethefollowingcharacteristics.
Phase
singlephase
B.C
Voltage
100,110,120V10%,0.2
Max.
200,220,230,240V10%,
0.8max.
Capacity
3.5kVAor
larger
Frequency
50or60Hz
Selectoneofthevoltagesindicatedaboveatthetimeof
installation.
26.
Page27
Donotusethepowersourcethatdoesnotconformtothe
Rulesabove.
Theuseofanenergysourcethatisnotincompliancewiththe
Requirementsspecifiedabovecancauseserious
Involvingfire,smokegenerationandexplosion.
EarthRequirements
Theearthresistancemustbelessthan100ohm.
https://translate.googleusercontent.com/translate_f
26/135
11/28/2016
OPERATIONMANUAL
Usealeakagecurrentbreakerthat
Harmonicoscillations.
1.3.2Storage
Evenwhenthefollowingconditionsaremet,
Makesurethattherearenosuddenchangesin
Temperatureorhumidity.
Thisleadstocondensation,whichcancauseproblems.
Condition
Stipulation
Temperature
10Cto40C
Relativehumidity
30%to85%(noncondensing)
Atmosphericpressure
700hPaat1060hPa
Atmosphereconditions
Theatmospheremustbefreeof
Corrosiveandexplosivegases
27
Page28
1.4ClassificationofDevices
1.4.1ClassificationaccordingtotheShockProtectionSystem
Electric.
ClassIDevices
ClassIdevicesarethosethatdependexclusivelyon
Basicinsulationforprotectionagainstelectricshock,inorderto
Toensurethatmetalpartsthatmaycomeintocontactwith
Ifthisbasicinsulationfails,donotcauseanelectricshock,
Throughthegroundwireinisprovidedasameasureof
Additionalsecurity.
https://translate.googleusercontent.com/translate_f
27/135
11/28/2016
OPERATIONMANUAL
1.4.2ClassificationaccordingtoDegreeofProtectionagainstShocks
Electric.
TypeBDevices
TypeBdevicesaredevicesthatprovidealevelof
Particularprotectionagainstelectricshock,anditapplies
Especiallyintheallowableandreliableleakagecurrentandin
Protectiveearthconnections.
28
Page29
1.4.3ClassificationaccordingtoDegreeofProtectionagainstEntry
HarmfulofWater.
StandardDevices
Usetheequipmentinaplacewherethereisnodangerof
Waterinlet.
Ifwatergetsinsidetheequipment,
Causeadangerouselectricshock.Donotallowwaterto
Reachthesurfaceoftheequipmentormoveinsideit.
Ifwatergetsintotheequipmentforanyreason,turnitoff
ImmediatelyandcalltheAuthorizedServiceCenter.
https://translate.googleusercontent.com/translate_f
28/135
11/28/2016
OPERATIONMANUAL
1.4.4ClassificationaccordingtoDegreeofProtectionagainstShocks
Electric.
Equipmentnotsuitableforuseinaninpresenceenvironment
Ofaflammableanestheticmixturewithairoranestheticmixture
Flammablewithoxygenornitrousoxide.
Useinanenvironmentcontainingflammableanestheticgas
Maycauseanexplosion.
1.4.5ClassificationaccordingtotheModeofOperation.
Continuousoperationwithintermittentload
(ContinuousoperationdependsonthespecificationoftheXraytube
mounted)
29
Page30
1.5SymbolsandNameplates
Manufacturer
1.5.1Symbols
Thesymbolsusedontheequipmentareindicatedbelow.
Symbol
Location
IdentificationPlate
Affixedtothehousing
Onthe
Force,wherethe
Protectionunit
Earth.
IdentificationPlate
Affixedtothehousing
CautionLabel/
Warning
https://translate.googleusercontent.com/translate_f
Meaning
Alternatingcurrent
B.C
EarthProtection
Classificationof
Safety:
TypeBdevice
Referenceto
Manualof
Instructions
29/135
11/28/2016
OPERATIONMANUAL
driverof
PotentialEqualization
driverof
Equalization
Potential
30
Page31
1.5.2Tags
Manufacturer'sIdentificationLabels
Manufacturer'sIdentificationLabelsareaffixedtotheindicatedlocations
below.
https://translate.googleusercontent.com/translate_f
30/135
11/28/2016
OPERATIONMANUAL
Figure1.2LabelPosition
31
Page32
https://translate.googleusercontent.com/translate_f
31/135
11/28/2016
OPERATIONMANUAL
Figure1.3ManufacturerIdentificationLabels
32
Page33
Others
LabelsCertifiedbytheEuropeanUnionareaffixedtotheindicatedplace
Asnumber1infigure1.2.
TUVApprovedSticker
CEMarking
EuropeanRepresentative
https://translate.googleusercontent.com/translate_f
32/135
11/28/2016
OPERATIONMANUAL
Figure1.4EuropeanUnionCertifiedLabels
33
Page34
1.6OperatorProfile
Item
Age
Sex
Nationality
Education
Knowledge
Language
Experience
https://translate.googleusercontent.com/translate_f
Details
AgethattheindividualcanobtainthelicenseofTechnician
Radiologyoranequivalentlicense.
withoutlimitation
withoutlimitation
Required
Eachoperatorneedstoconduct
Operatetheequipmentbeforeusingit.
Required
Eachoperatorneedstoconduct
Operatetheequipmentbeforeusingit.
withoutlimitation
Required
Eachoperatorneedstoconduct
Operatetheequipmentbeforeusingit.
33/135
11/28/2016
OPERATIONMANUAL
34
Page35
Chapter2
Configurationofthe
System
https://translate.googleusercontent.com/translate_f
34/135
11/28/2016
OPERATIONMANUAL
Thischapterpresentsthesystemconfiguration.
35
Page36
2.1StandardConfiguration
ThetwolargestcomponentsofthesystemaretheCarmsupportcar
Andstandwithmonitors.
2.1.1SupportCarwithArmC
TheCArmSupportCarcanbeoperatedtoreachposition
Moresuitableforfluoroscopyorradiography.
TheCArmSupportCarisconnectedtothemonitorcartusing
thesystem.
https://translate.googleusercontent.com/translate_f
35/135
11/28/2016
OPERATIONMANUAL
Figure2.1CArmCart
36
Page37
(1)IIControlKnob(optional)
Thisisthehandleusedbythesurgeontooperatethesystem.Thehandleisprovided
WithaCarmlockswitch.
(2)Xrayimageintensifierunit
TheabbreviationforimageintensifierisII
Xraysareconvertedtoavisibleimageonascreen
Thisfluorescentimageisconvertedtophotoelectronby
Photocathode,andthephotoelectronsareacceleratedandconvertedbyalens
Electrostatic,directingthemtothefluorescentoutputscreen.
(3)BeamLimitingDevice(Collimator)
ControlstherangeofXrayradiationwhenperformingfluoroscopyor
CaptureanXray.
(4)XrayGenerator
Thisisthexraygeneratingunit.Thexraytubeisoftheanodetype
fixed.
(5)SidePanel
Throughthispanelitispossibletolock/releasetheelectromagneticlocksofthe
ArmC,rotatethecamera/image,inverttheimage(up/downand
Right/left),open/closethecollimatorandactivatethememoryfunction.
https://translate.googleusercontent.com/translate_f
36/135
11/28/2016
OPERATIONMANUAL
Figure2.2SidePanel
(6)ArmC
TheCarmiscounterbalancedandoperatedmanually.O
Iselectromagnetictype,andcanbeswitchedonoroffwitha
37
Page38
Touch.Thearmcaneasilybepositioned.
(7)Mainpanel
Usedtodisplayandadjusttheradiologicaltechniques,writedatain
memory,andselectimagesdisplayedmemory.
Figure2.3aMainPanelACTIVE
https://translate.googleusercontent.com/translate_f
37/135
11/28/2016
OPERATIONMANUAL
Figure2.3bMainPanelFULL
(8)Handlesformaneuver
UsedtomovethestandarmC
38
Page39
(9)Parkingbrake
ThisistheparkingbrakemanualforthestandarmC
Figure2.4Parkingbrake
(10)Connector
Connectthemonitorstandstandout
https://translate.googleusercontent.com/translate_f
38/135
11/28/2016
OPERATIONMANUAL
Figure2.5Connector
39
Page40
(11)Pedal
Selectthefluoromode.
Leftfootswitch:continuousfluoroscopy,pulsedfluoroscopy
Switchrightpedal:Radiography
Figure2.6Pedal
Subtitle
1LeftPedalcontinuousfluoroscopy,pulsedorHighQuality
2RightPedalXray
(12)UnitXrayControl
https://translate.googleusercontent.com/translate_f
39/135
11/28/2016
OPERATIONMANUAL
40
Page41
2.1.2CartMonitor
Thisisthestandofthemonitor,andisusedconnectedtothestandarmC,
andcontrolsthepowersupplytoallWHA200system.
Figure2.7MonitorStand
Onthedisplaymonitor
Dependingonthedisplayyoucanviewconditions
verysmalldotsorirregularities,butdoesnotindicatea
failure.
https://translate.googleusercontent.com/translate_f
40/135
11/28/2016
OPERATIONMANUAL
Somepixelsmayhavedisplaydefects(Itcan
notlightoralwaysstaylit).
Incoldplaces,theimagemayappearwithadashorthe
poorimage,butthisdoesnotindicatefailure,theequipmentwillreturn
tonormaloncethetemperaturerises.
Donotleaveastillimageforalongtime,thiscan
causedamagetothedisplay.
Theimagemayappearindifferentshadesofcolorbetweenthe
MonitorandtheReferenceMonitor.
41
Page42
(1)Monitor
Displaysthefluoroscopicimagesduringfluoroscopyimages
Radiographic,maintaininglastimage(LIH)andrevisedimages.
(2)ReferenceMonitor
Displaysreferenceimages.
Itallowsrecordingpatientinformationandmakevariousadjustmentsto
imaging(fluoroscopy,radiography),postprocessing,etc.
Inadditiontooperationswiththemouseandkeyboard,youcanalsoperform
operationswiththetouchpanel(optional).
(3)MonitorPositionAdjustment
Usedtomoveupordowntoadjustthepositionof
monitor.
(4)Keyboard
Locationofthekeyboard,mouse,etc.
(5)SystemSwitch
Turnonoroffthesystem.
(6)IndicatorIrradiation
TheindicatedisilluminatedduringradiationbyXrays.
(7)ArmSupportAssistant
Fixedthedisplaypositionintheverticaldirection.Toadjustthisposition,
WemustfirstreleasethelockSupportFlyArm.
(8)PowerSwitch
Turnonoroffthesystem.
https://translate.googleusercontent.com/translate_f
41/135
11/28/2016
OPERATIONMANUAL
42
Page43
2.2MainPanel
2.2.1OpescopeActivefunctionofeachkey
Figure2.8MainPanelOpescopeActive
Legendand
1Releasingtheexposurerestrictionofxrayspresstheswitch
exposuretoxraysorthepedaltoreleasetherestrictionofxraysto
fluoroscopyandradiography.
2Exposuretoxray(fluoroscopy)pressthiskeytostart
exposuretofluoroscopy.
3filmradiographymodepressthiskeytomakearecord
photoonthefilm.TheLEDwilllightup.
4Optionaldigitalsubtractionsystem
5FilterSelectionpressthiskeytoselectthefilter.Thefilter
selectedwillchangeeverytimethekeyispressed..
https://translate.googleusercontent.com/translate_f
42/135
11/28/2016
OPERATIONMANUAL
Filter1(fine):LeftLEDlights
Filter2(medium):TherightLEDlights
Filter3(thick):ThetwoLEDswilllight
Noadditionalfilter:BothLEDsareoff
6Optionaldigitalsubtractionsystem
43
Page44
7Optionaldigitalsubtractionsystem
8Optionaldigitalsubtractionsystem
9APRThecorrespondingLEDwilllighttheAPR.
10IBSpressthiskeytoturnon/offtheIBSfunction(setfunction
Automaticbrightness)
11SelectingthesizeofthemagnificationIIpressthiskeyto
selectthesizeoftheIILEDremainslitwhenmagnification
the6inchfieldisselected,andturnsoffwhenthe9field
inchisselected.
12LaserPointerLEDwillremainlitwhilethelaserpointer
Itisbeingused.
13continuousfluoroscopyratedisplay/PulsedHighQuality
14/15SelectionFluoromodetopressthesekeystoselectthe
Fluoromode.TheLEDslightupasfollows:
Highquality(highqualitymode):
"HQ"light.
Lowdoses(lowdosemode)
"LD"willlightup.
Normal(normalmode)
AllLEDswilllightup.
16TemperaturegeneratorxrayIndicatesthetemperatureofthegenerator
Xray.asthetemperaturebecomeshigh,thelampcolorchangesfromgreento
orangeandred.Whentheindicationisorangeorred,itmeans
thatfluoroscopyisdisabledassoonasthetemperatureofthexraygenerator
istall.
17NumberDisplayindicatesthecurrentmemoryimagenumber.if
noimageisdisplayedwillnotshowanynumber.
18MemoryAddsanimageinmemory
19SelectionPresstochangetheimagedisplayedmemory.
https://translate.googleusercontent.com/translate_f
43/135
11/28/2016
OPERATIONMANUAL
20PrinteroutputPrintsimages(onlyeffectivewhena
printerisinstalled)
21Exchange/displaysthetubevoltageIndicatesthetubevoltage.Press[]/[+]
tomanuallychangethetensionadjustment.However,notethatyoucannot
44
Page45
changingthetubevoltagewhentheIBSfunction(Auto)ison
22CurrentDisplayofthexraytubeToedit/displaythecurrent
InFluoromode:
Thetubecurrentisautomaticallydisplayedinaccordancewiththefluoromode.THE
Currentcurrentcannotbechangedmanually.
Intheradiographicfilmmode:
Indicatestheproductofthetubecurrenttime.Press[]/[+]tochangethis
time.
23CumulativefluoroscopyTime/displayradiographytime:
InFluoromode:
Itshowsthecumulativetimefluoroscopy.theminutesandsecondsaredisplayed.
Whenthetimeexceeds99minutesand59seconds,oneminuteisdisplayed.
Inradiographicfilmmode:
Beforeradiography,displaystheadjustedtimeradiographic
AftertheXray,radiographicdisplaysthemeasuredtime
24Displayoffluoroscopywristband,Camera,collimatorangle,
Errordisplay:
Normallydisplaystherangeofpulsedfluoroscopy(fpslights)
Duringcontinuousfluoroscopy,nothingisdisplayed(fpsoff)
Whenthecameraoperationandthecollimator(ie,theswitchesonthepanel
sidebeingoperated),thecurrentangleisdisplayedflashing.
Ifanerroroccurs,anerrorcodeisdisplayed.
25Resetpressthiskeytoresetadisplayederror.Holding
thiskeyforonesecondthefollowinginformationwillbereset:
cumulativefluoroscopytime
Registrationinformationinthekey
MemoryImage
26APRRegistrationpressthiskeytoregistertheAPR.
27FluoroModeDisplaywilllightthewayFluoro
28ErrorDisplaystayonafteranerrorisactive
https://translate.googleusercontent.com/translate_f
44/135
11/28/2016
OPERATIONMANUAL
29Standbydisplaylightwhiletheirradiationofxraysispossible
30DisplayexposureXraysduringirradiationilluminatexray
31AlarmResetTimerfluoroscopyFortimerfluoroscopy
45
Page46
2.2.2OpescopeFullFunctionofeachkey
Figure2.9MainPanelOpescopeFull
Subtitle:
1releasebuttonlockdisplayxrays:pressto
releasethelockexposurexrays.
2Exposurekeyxray:presstostartfluoroscopy.
3Alarmresetbuttonoffluoroscopictimer:fortheaudiblealarm
timerfluoroscopy.
4Imageexchangekey(optional):Thiskeyisusedtochangethepicture
oroperationscreen.
5Automaticadjustmentkeywindow/level:automaticallyadjuststhecontrastand
thebrightnessoftheradiographicimage.
6windowadjustmentbutton:adjuststhecontrastoftheimage.
7leveladjustmentbutton:adjuststhebrightnessoftheimage.
8depthadjustmentbutton:adjuststhedepthoftheimage.
9examcompletionkey:presstoendtheexam.
10MENUbutton:pressthisbuttontoadjustthemenu.
11Multipurposekey:thiskeyisusedinconjunctionwiththedisplayedbutton
https://translate.googleusercontent.com/translate_f
45/135
11/28/2016
OPERATIONMANUAL
onthescreeninimageprocessing.
46
Page47
2.2.2.1FullOpescopeLCDDisplay
Atouchscreenisbuiltintotheunit,allowing
operationdirectlybypressingthebuttonsonthescreen.
Thebuttonsdisplayeddifferaccordingtothescreen.Thebasicscreenappears
below:
1DisplayoffluoroscopymodeLitinfluoroscopymode.
2IBS(imagebrightnessstabilizationfunction)Pressthiskey
toenableIBSfunction.ThekeyislitwhentheIBSfunctionison.
3voltageoftheXraytubeIndicatesthetubevoltage.Theadjustmentvariesfrom40to
110kV.Press[+]/[]tochangethisvaluemanually.
4currentoftheXraytubeInthefluoroscopymode,thetubecurrent
Itisautomaticallyselected.Thisvaluecannotbechangedmanually.
5SystemStatusIndicatorDisplaysthestatusofthesystem.
https://translate.googleusercontent.com/translate_f
46/135
11/28/2016
OPERATIONMANUAL
Whenfluoroscopicradiographyispossible.
47
Page48
WhenthereisthepresenceofXrayradiation.
WhenthetubeXrayisinhightemperatureorhighhumidity.
6FluoroscopyAccumulatedTimeInorderfluoroscopytime
accumulatedfluoroscopyisindicatedhere.Whenthetimeexceeds99minutes59
seconds,thedisplaychangestominuteunitsonly.
7FilterUsedtoselectthefilter.Eachtimethiskeyis
Pressthefilterchanges.Thefiltersequenceis1Filter<Filter2<Filter3.
8MoviemodeRadiographicPressthiskeywhenyouwanttoperform
rayperfilm.Thescreenradiographappears.
9XraytubetemperatureIndicatesthetemperatureoftheXraytube.
whenthetemperatureincreases,thedisplaybarextends.
10DisplaymodeDisplaysthedisplaymodedialog.Thisallowsmenu
selectthetypeofdisplay,thumbnaildisplay,performing
Zoomoperations,reversethepolarityoftheimage,reverseimage
vertically,orremasktheimage.
11SendREFPresstosendtheimagedisplayedonthemonitortothe
referencemonitor.
12SavefluoroscopicimageSavethefluoroscopicimagedisplayedon
monitor(includingloops)ontheharddrive.
13VideoPrintingPresstoprinttheimagedisplayedonthescreen
monitorinanimageprinter.(Onlyappearswhentheprinter
specifiedisconnected.)
14SelectionAPRDisplaystheAPRscreen.
15IntensifiedimagefieldsizeselectionPressto
select16cmor23cmasintensifierfieldsize
Image.
16LoopFluoroscopyUsedtoregisterafluoroscopicimage
inmemory(Maximum100imagesbeforefinalizingfluoroscopy).Stay
litwhenthefunctionison.
17LaserPointeronoroffthefunction.
18ProductDosebyAreaIndicatesthedoseperareadetectedbyasensor.
https://translate.googleusercontent.com/translate_f
47/135
11/28/2016
OPERATIONMANUAL
48
Page49
2.3Sidepanel
2.3.1lockswitcharmC
UsedtolockorreleasethearmC.
Figure2.10lockswitcharmC
1slipmovementofthearmC
LEDon:ThelockarmCisreleased
LEDoff:ThearmCislocked
2RotationalmovementofthearmC
3lateralandposteriorslidingmovementofthearmC
4longitudinalmovementofthearmC
2.3.2ControlKeys
Figure2.11ControlKeys
https://translate.googleusercontent.com/translate_f
48/135
11/28/2016
OPERATIONMANUAL
49
Page50
1/2MemoryKeyusedtostoretheconditionsoffluoroscopyand
thecollimatorsettings
3Open/Closethecollimatoroctagonal
4Open/Closetheparallelcollimator
5Rotatestheparallelcollimator
Figure2.12Collimatorsoctagonalandparallel
6FastFoodholdingthiskeywhilepressingthe
rotationoftheparallelcollimatororcamerarotationspeedinthese
operationsincreases
7Rotationcamerausedtorotatethecamera
8Monitorimagereversalusedtoreversetheimagedisplayedon
themonitor.
https://translate.googleusercontent.com/translate_f
49/135
11/28/2016
OPERATIONMANUAL
50
Page51
Chapter3
Basicoperation
Thischapterexplainsthebasicproceduresforusingthesystem.
51
https://translate.googleusercontent.com/translate_f
50/135
11/28/2016
OPERATIONMANUAL
Page52
3.1BASICOPERATIONS
PoweringOn/Offpower
Toavoidproblemswiththesystemwhileexaminingapatient,
alwaysmakeaninspectiontocheckthateverythingisinaccordancewiththe
specifications.
3.1.1ConnectingtheUnits
1Unrollthesignalcableonthebackofthecarsupport
monitors.
2Connectthesignalcabletothemainunittothispositiontheconnectorand
rotateitclockwiseturnasfig.below.
Figure3.1Connector
3Connectthepowercordatthebackofthecarsupport
monitorsthelocaloutlet,confirmingthattheunitvoltageiscompatiblewith
makingselected.
52
https://translate.googleusercontent.com/translate_f
51/135
11/28/2016
OPERATIONMANUAL
Page53
3.1.2)Connecttheequipment
1Setthepowerswitchonthebackofthemonitorstandto"|"
[switchedon].
Figure3.2On/offswitch
3
2Pressthefrontofthekeymonitorstand.
AllLEDsonthemainpanelandsidepanelwilllightupfor
about1second.
Initializationofthecollimator,camerarotation,etc.,willcontinuefor
approximately30seconds.Duringthistime,thekey"init"
lightsinthepulseratedisplayedonthemainpanel.
Whenyouturnofftheequipmentandneedtoconnectagain,waita
10minuteinterval.
53
https://translate.googleusercontent.com/translate_f
52/135
11/28/2016
OPERATIONMANUAL
Page54
3.2Turningthepoweroff
3.2.1ActiveOpescope
Afteryoufinishusingthemachineperformthefinalinspectionandfilethe
results.
1MakesurethearmCisinthehome.
Theinitialpositionisoneatallscalesrecordzero.
Figure3.3StartPositionArmCOpescopeActive
3
2Pressthekeyto[OFF].
3Settheswitchonthebackofthemonitorstandto[off].
4Disconnectthemainpowercable.
54
Page55
https://translate.googleusercontent.com/translate_f
53/135
11/28/2016
OPERATIONMANUAL
3.2.2FullOpescope
strictlyfollowtheprocedurebelowtoturnoffthe
equipment.
OnlypresstheOFFbuttononthebackofthemonitorstand(s)or
unplugthepowercordafterthecompletionoftheprocess
systemshutdown.Butthesystemandthedataislost.
Also,iftheOFFkeyispressedduringthedatatransfer
imagetoamedia(CDR),theimagedatawillbelost.
1MakesurethearmCisinthehome.
Theinitialpositionisoneatallscalesrecordzero.
Figure3.4StartPositionArmCFullOpescope
2Proceedwiththeshutdownoperationoftheequipmentthroughthemonitor
ofreference.
Pressthefunctionkey
Choosefunction
todisplaythefunctionsonthemonitor.
.
Thepatient'sfilewillbeclosedandaconfirmationdialogbox
askingifyouwanttoexittheapplicationisdisplayed,confirm
55
Page56
https://translate.googleusercontent.com/translate_f
54/135
11/28/2016
OPERATIONMANUAL
clicking"Yes".
3Makesurethegreenlightontheswitchfromthefrontofthecart
monitor(s)off.
4Pressthepowerbutton(OFF)onthebackofthecart
monitor(s)
Figure3.5PowerButtonFullOpescope
5Unplugthepowercord
3.3POSITIONING(ARMCOPERATION)
PlacetheArmCintherequiredpositionfluoroscopy/radiography.
Figure3.6Locationofthepointwherethefingerscangetcaught
56
Page57
https://translate.googleusercontent.com/translate_f
55/135
11/28/2016
OPERATIONMANUAL
Beawareofthedangerofshockorholdanypartofthebodyto
equipment.
WhenoperatingthearmC,carefulnottobumptheequipmentdevices
peripherals.ManipulatethearmCbythehandles,itmayhappenaccidents
howtoholdyourfingerstotheequipment.
3
Donotplaceperipheraldeviceswithintherangeofmovement
system.
Ifperipheraldevicessuchasthemonitorstandisplacedwithinthearea
showninFigure15,thereisadangerofcollisionorcontact.
Figure3.7DriveRange
57
Page58
https://translate.googleusercontent.com/translate_f
56/135
11/28/2016
OPERATIONMANUAL
1PressthearmCofthebeaminthedirectionyouwanttomove.
TheLEDwilllightandthecorrespondingbrakeisreleased.
2Holdingthehandles,movethearmCfortheposition.
3WhenyoufinishpositioningthearmC,engagethelock.
TheLEDwillturnoffandthearmCislocked.
4Makesurethebrakeisapplied.
Evenwhenthebrakeisactivated,youcanmovetheframein
oppositiontothebrakebyapplyingaforcegreaterthan10kgf.Thisisameasure
toenabletheoperationofallpartsofthesystemevenin
criticalsituationasoverhead.
Ifmorethanoneminutepassedwithoutanyoperationintheunlockedstate
(LED),thesystemautomaticallyreturnstothelockedstate(LED
wipedout).
3.3.1VerticalmovementofarmC
Figure3.8VerticalmovementofarmC
58
Page59
https://translate.googleusercontent.com/translate_f
57/135
11/28/2016
OPERATIONMANUAL
3.3.2longitudinalmovementandlateralarmC
Figure3.9longitudinalandlateralmovementsofthearmC
59
Page60
3.3.3axialrotationalmovementofthearmC
https://translate.googleusercontent.com/translate_f
58/135
11/28/2016
OPERATIONMANUAL
Figure3.10axialrotationalmovementofthearmC
3.3.4rotationalmovementofthearmC
Figure3:11rotationalmovementofthearmC
60
Page61
3.3.5scales
Everymovementfeaturesascalethatyoucanseeasaguide
https://translate.googleusercontent.com/translate_f
59/135
11/28/2016
OPERATIONMANUAL
duringthemove.Notethattherangeofcolormatchesthecolorkey
armClock.
ThezerooneachscaleistheinitialpositionofthearmC.
Figure3.12Scale
61
Page62
3.3.6GauntletofIIControl(Optional)
TherearetwokeysonthehandleII
https://translate.googleusercontent.com/translate_f
60/135
11/28/2016
OPERATIONMANUAL
Whilethiskeyispressed,thelockonthereturnslip
sideandarmClongitudinalmovementarereleasedandtheLEDsbelow
Theywillbeilluminated.
Figure3.13lockkeysonthehandleII
62
Page63
3.4SELECTINGAPR
3.4.1OPESCOPEASSETS
https://translate.googleusercontent.com/translate_f
61/135
11/28/2016
OPERATIONMANUAL
1Pressthe[APR]key.TheLEDwilllightup.
Thereare6typesofRPAaccordingtotheobservationposition.Theconditions
greatfluoroscopyandcamerarotationpositionsaresetintheAPR.
TheAPRsaresetinthesystematthefactory,butyoucanchangeto
Accordingtooptimizeitsuse.
Figure3.14APRKey(OpescopeActive)
T
LegendPositionNote:
1HeadHead
2AbdomenAbdomen
3Upperarm,handjoints,fingers,shoulderjoints,etc.upperarm,gaskets
hand,fingers,jointsofshoulders,etc.
4LegsLegs
5/6UsersettingsUsersettings
63
Page64
3.4.2OPESCOPEFULL
AfterregisteringthepatientorpresstheAPRselectionbuttononthe
mainpanel,itwillbedisplayed.
https://translate.googleusercontent.com/translate_f
62/135
11/28/2016
OPERATIONMANUAL
1Check[APR]
TochangetheAPR,presstheselectkey.
TheselectedAPRwillappearhighlightedonthemonitor.
2Press[OK]
Theprimaryscreenisdisplayed.
Figure3.15APRKey(OpescopeFull)
64
Page65
3.4.2.1ChangingtheAPRscreen
Theywillbeshownsixbuttons[APR]simultaneouslyonthescreen
APR.Press[next]togotothenextpage.It'spossible
https://translate.googleusercontent.com/translate_f
63/135
11/28/2016
OPERATIONMANUAL
viewupto12APRs.
Figure3.16APRSelectionScreen
65
Page66
3.4.2.2ChangingthecombinationoffluoroscopyandacquisitionAPRAPR
1Presstheacquisition/fluoroscopyAPR.
Theselectionoptionsaredisplayed.
https://translate.googleusercontent.com/translate_f
64/135
11/28/2016
OPERATIONMANUAL
Figure3.17Screenselectionofacquisition/fluoroscopyAPR
2ChangetheselectionoffluoroscopyAPRandAPRacquisitionandpress[closes].
APRselectionscreenwillberedisplayed.
FluoroscopyAPRandAPRacquisitionwillbeindicatedonthedisplay.
66
Page67
https://translate.googleusercontent.com/translate_f
65/135
11/28/2016
OPERATIONMANUAL
Figure3.18Screentoclosethewindow
3Press[OK]
Theprimaryscreenisdisplayed.
67
Page68
3.5FLUOROSCOPY
3.5.1OPESCOPEASSETS
https://translate.googleusercontent.com/translate_f
66/135
11/28/2016
OPERATIONMANUAL
Figure3.19resetbutton
Whenfluoroscopyoptionequipmentistestedbeforestart
cycletests,youneedtoreset(delete)thefluoroscopytime
totalized.
Pressthe[reset]buttonforabout1second.
1Press[releasingthexrayexposurerestriction].
TheLEDlights[releasingthexrayexposurerestriction].
68
Page69
https://translate.googleusercontent.com/translate_f
67/135
11/28/2016
OPERATIONMANUAL
Figure3.20Freeingexposurerestrictionofxrays
Thexrayradiationisonlypossiblewhenthekey
ordertopreventfaultyoperation,theLEDisoffwhenthe
poweristurnedon.WhiletheLEDisoff,youcannot
initiatingradiation.
ison.In
3
2Checktheconditionoffluoroscopy.Thevariablesareset
automaticallyaccordingtotheselectedAPR,butyoucanchangethe
manuallyadjustments.
Figure3:21Checkingtheconditionsoffluoroscopy
69
Page70
3Pressthepedalorpress[exposurexray(fluoroscopy)].
Xraysareirradiatedwhilethepedalispressedorwhile
keyispressed.
Theradiationindicatorlights.
Thefluoroscopicimageisdisplayedonthemonitor.
https://translate.googleusercontent.com/translate_f
68/135
11/28/2016
OPERATIONMANUAL
Figure3.22Startfluoroscopy(exposurekeyxray)
Whenthesettingsoffluoroscopyconditions,orcollimator,orrotation
camera,areexchangedmanually,theLEDwillflashAPR.
3
70
Page71
changingthe
soFluoro
HQmodetheLEDwilllight.
InLDmode,theLEDwilllight
Inordinarywordstate,pressthebuttonthatislitto
turnofftheLED
HQ:HighQuality(highqualitymode)
Theareaisirradiatedwithahigherdoseofxraystogenerate
Highqualityimagesinpatientsdifficulttopenetratethe
Xray.
LD:LowDose(lowdosemode)
Thelowdosexraysisusedtoreducetheamount
irradiatedxrays.
Normal(normalmode)
Theirradiateddoseisstandard.
Swapping
inbetween
fluoroscopy
pulsedand
Pressthekeys
toswitchbetweenfluoroscopy
continuousorpulsedfluoroscopy.Forpulsedfluoroscopy,the
Pulserateisdisplayedwithanumericalvalue.
Tobecontinued
https://translate.googleusercontent.com/translate_f
69/135
11/28/2016
OPERATIONMANUAL
fluoroscopy
tobecontinued
Press
pulsed
Calling/
disconnectingthe
functionIBS
(Function
stabilization
Shining
image)
TheLEDwillremainlitwhiletheIBSfunction
isoff.
WhenIBSfunctionisconnected,thedoseofxraysis
automaticallyadjustedaccordingtothearea.
Changingthe
tubevoltage
Increasesvoltage
Decreasestension
OnlyvalidwhileIBSisOff
Changebypressingthekeys[+]/[].
Exchanging
filter
Thefilterischangedeverytimethekey
Filter1(fine):LeftLEDlights.
Filter2(medium):TherightLEDlightsup.
Filter3(thick):BothLEDslightup.
Noadditionalfilter:BothLEDsareoff
ispressed.
71
Page72
Selecting
thesizeofthe
magnification
II
Press
toselectthesizeof
magnificationII
Whenthefield16cm(6inches)isused,theLED
remainon.Whenthepitchof23cm(9inches)is
used,theLEDisoff(Validfor9Activemodels"
andFull.
ForcontinuousfluoroscopyinHQmode,thehighdoseisused.Forthisreason,a
warningtoneissoundedcontinuously.
https://translate.googleusercontent.com/translate_f
70/135
11/28/2016
OPERATIONMANUAL
Ifthefluoroscopytimeisexceeded,thealarmwillsound.Thealarmcanbe
stoppedbypressing
Youcancontinuewithfluoroscopyevenwhenthefluoroscopytime
setisexceededandthealarmsounding.However,thelimitforfluoroscopy
Continuousis10minutes.Asthislimitisreached,theexposureofxraysis
automaticallystop.Tocontinueafterthispointfluoroscopy
releasetheleftpedalandpressitagain(orreleasethebutton
[Exposurexray(fluoroscopy)]andpressitagain).
72
Page73
3.5.2OPESCOPEFULL
1Checkthefluoroscopicconditions.
Thefluoroscopicconditionsareautomaticallyadjustedaccordingtothe
APRselection.
Youcanchangetheseconditionsmanually.
https://translate.googleusercontent.com/translate_f
71/135
11/28/2016
OPERATIONMANUAL
Figure3:23Mainscreenoffluoroscopicconditions
2Ifthekeyisoff,pressit.
Tomaketheirradiationbuttonshouldbeon.
Figure3.24exposurelockreleasebuttonxray
73
Page74
3Presstheleftpedal.
Irradiationforfluoroscopystarts.
Duringirradiation,thekeywillremainonthemainpanel.
Whilexraysareirradiated,theirradiationindicatorislitin
Standmonitor.Thefluoroscopicimageisdisplayedonthemonitor.
https://translate.googleusercontent.com/translate_f
72/135
11/28/2016
OPERATIONMANUAL
Figure3.25Exposurebuttonxray(fluoroscopy)
3.5.2.1Changingthefluoroscopicconditions1(fluoroscopyprogram)
1Press[fluoroscopyProgram].
Fluoroscopyprogramselectionscreenisdisplayed.
Figure3.26fluoroscopyProgram
2Selectthefluoroscopyprogram.
Theselectedbuttonishighlightedandthewindowcloses.
74
Page75
Figure3.27fluoroscopyProgram
[Next]fluoroscopyprogramselectionscreenshowsfour
fluoroscopyprogramsonapage.Press[next]todisplaythe
nextpagefluoroscopyprogramselectionscreen.
3.5.2.2Changingconditionsfluoroscopy2(fluoroscopymode)
https://translate.googleusercontent.com/translate_f
73/135
11/28/2016
OPERATIONMANUAL
1Press[fluoroscopymode].
Theselectedbuttonwillappearprominently.
[HQ]HighQuality(highqualitymode)Theareaisirradiatedwithhigher
dosexraystogeneratehighqualityimagesinpatientsdifficult
penetrationofXrays.
[N]:Normal(normalmode)thestandarddoseisirradiated.
[LD]Lowdose(lowdosemode)lowdosexraysisradiated.
Figure3.28fluoroscopymode
75
Page76
3.5.2.3Changingthefluoroscopicconditions3(pulserate)
1Press[pulserate]
Pulserateselectionscreenisdisplayed.
Figure3:29Pulserate
https://translate.googleusercontent.com/translate_f
74/135
11/28/2016
OPERATIONMANUAL
2Selectthefluoroscopicpulserate.
Select[CONTINUOUS]forcontinuousfluoroscopy.Whenselectingany
anotherbutton,pulsedfluoroscopyisperformedandthepulseratewillbe
changesinrelationtotheselectedvalue.
Ifyouprefernottochange,press[closes].
Figure3.30PulseRateSelection
76
Page77
3.6CAMERAROTATION,REVERSEIMAGE,
COLLIMATOROPERATION
Figure3.31Viewofthecontrolbuttons
3.6.1CameraRotationKey
camerarotation/image
Pressingthisbuttonthecamera/imagerotates.Duringfluoroscopy,theimage
displayedtheexposedareaalsorotatesaccordingtotherotationofthecamera.ITIS
possibletoincreasethespeedofrotationofthecameraholdingthekey
fastfoodwhilepressingthecamerakeyrotation.
https://translate.googleusercontent.com/translate_f
75/135
11/28/2016
OPERATIONMANUAL
Whileholdingthecamerarotationkey,therotationangle
thecameraisdisplayedinthemainpanel.(OpescopeActive)
Figure3.32Screenrotationangle(OpescopeActive)
77
Page78
3.6.2ImagebuttonReverse
reverseimage
Pressingthisbuttonfluoroscopicimagereversed.
1.ReverseH(horizontalinversion)
2.InversionV(verticalflip)
https://translate.googleusercontent.com/translate_f
76/135
11/28/2016
OPERATIONMANUAL
78
Page79
3.6.3OperationKeyofCollimator
Pressthesebuttonstoopen/closeandrotatethecollimatorandoctagonal
parallelcollimator.
Whilepressingthebutton,theangleofrotationisindicated
onthescreeninthemainpanel.Theparallelcollimatorallowsrotations
repeated.
Figure3:33Collimatoroctagonalandparallel
1.Open/Closethecollimatoroctagonal
Pressingthisbuttonopensthecollimatoroctagonal.
Pressingthisbuttontheoctagonalcollimatorcloses.
https://translate.googleusercontent.com/translate_f
77/135
11/28/2016
OPERATIONMANUAL
2.Open/Closeparallelcollimator
Pressingthisbuttonopensthecollimatorparallel.
Pressingthisbuttonclosestheparallelcollimator.
3.Rotationoftheparallelcollimator
Pressingthisbuttontheparallelcollimatorrotatesin
schedule.
Pressingthisbuttontheparallelcollimatorrotatescounterclockwise
schedule.
4.Speed
Pressingthiskeytogetherwiththekeyrotation
theparallelcollimatorspeedisincreased
79
Page80
3.7FUNCTIONOFMEMORYFRAME
OPESCOPEASSETS
Youcantemporarilystoreupto8imagescurrentlydisplayed.
Settingonamonitor(standard)
Memoryimagesframearedisplayedonthemonitor.Note,however,
thatduringthefluoroscopicimagefluoroscopyappears,butthen
fluoroscopytheLIHimageisdisplayed.Youcanswitchtodisplaythe
framememory,performingtheoperationsoftheframememory.
Configurationoftwomonitors(usedonlyasthesecond
monitoroptionalisselected)
Theframememoryoftheimageisalwaysdisplayedonthereferencemonitor.
storingimages
Press
[Framememory].
Thecurrentimageisdisplayedonthemonitorisstored,andthenumberof
https://translate.googleusercontent.com/translate_f
78/135
11/28/2016
OPERATIONMANUAL
memoryisdisplayed.
Imageselection
Press
[Selectionframe].
Pressingthisbutton
theframenumberincreases.
Pressingthisbutton
framenumberdecreases.
Theexchangeimagedisplayedwhenthememorynumberischanged.
Memorynumber1:latestimageinmemory
MemoryNumber8:olderpictureinmemory
80
Page81
Whenthenumberofpayrollimagestotheframememoriesexceeds
8,thenewimageisstoredasnumber[1]andtheimage
storedinthenumber[8]iserased.
Whenthesystemisturnedoff,thestoredimageisdeleted.
https://translate.googleusercontent.com/translate_f
79/135
11/28/2016
OPERATIONMANUAL
81
Page82
3.8RADIOGRAPHY(CRRadiographicorMovie)
OPESCOPEASSETS
WhenRadiographicMoviemodeisselected,thecollimatoroctagonal
Itwillautomaticallybefullyopened.Ifnecessaryadjustthe
collimatorparallel,makethesettingonthesidepanel.
Setthetapeandtaketheradiograph.
[FilmradiographymodeorCR].
1Press
2SetthekVandmAsvaluesbypressing[+]/[].
3Releasetherightpedalorpressthehandbutton(optional).
StartpreparationsforXray.
The[ofthescreen]keywilldeletetemporarily.
https://translate.googleusercontent.com/translate_f
80/135
11/28/2016
OPERATIONMANUAL
Whenpreparationsforexposurearecompleted,the
keywillreturn
4Whenusingthepedal,holdthepedal.
Whenusingthehandcontrol(optional),pressthecontrolforthesecond
level.
Thexrayirradiationstarts.
Duringirradiation,theindicatorlightirradiationwillbe
on.
82
Page83
Keepthepedalorhandcontrol(optional)Pressed
FullXrayradiation.Ifthecommandsarereleased,the
radiationXrayswillendprematurely.
Ifthesystemhasthehandcontrol(optional),youcannottake
Xrayswiththerightpedal.
https://translate.googleusercontent.com/translate_f
81/135
11/28/2016
OPERATIONMANUAL
Figure3:34FilmRadiography(OpescopeActive)
83
Page84
DependingontheXrayparameters,onemaybenecessary
waitingtimeafterradiography.Inthiscase,theword"COOL"willbe
displayed,alongwiththewaitingtime.Duringthistimewillnot
possiblexrayfluoroscopybutyes.
https://translate.googleusercontent.com/translate_f
82/135
11/28/2016
OPERATIONMANUAL
Figure3.35DisplayCOOLandDisplayTimeout
84
Page85
OPESCOPEFULL
1Placethesupporttape
2Press[radiographymodefilm].
Thekeywilllight.
Figure3:36radiographymodeinmovie
3Press[+]/[]toadjustthetubevoltage(kV)andtheproductof
timextubecurrent(mAs).
https://translate.googleusercontent.com/translate_f
83/135
11/28/2016
OPERATIONMANUAL
Figure3:37radiographymodeinmovie
85
Page86
4Releasetherightpedal.Alternatively,pressthehandcontrol
(Optional)tothefirstlevel.
Figure3:38RadiographyinFilm
5Continuereleasingtherightpedal,orpressthehandcontrolto
thesecondlevel.
Thexraysareirradiated.Duringirradiation,thecorresponding
indicator(onthemonitorstand)willremainon.
Keepthepedalorhandcontrol(optional)Pressed
FullXrayradiation.Ifthecommandsarereleased,the
radiationXrayswillendprematurely.
6Press[radiographymodefilm]
https://translate.googleusercontent.com/translate_f
84/135
11/28/2016
OPERATIONMANUAL
ThekeygoesoutandtheXrayiscompleted.Theprimaryscreenisre
displayed.
86
Page87
3.9USEFULFUNCTIONS
3.9.1MemoryKey
Storetherotationalpositionofthecamera,collimatorposition,thefactor
loadingandotherinformation.
Thisoperationisnotpossibleduringfluoroscopy.
Informationstoredwiththisfunctioncanbeerased
pressingthebutton.
Thefollowingparameterscanbestored.
Camerarotationalposition
Positionopen/closeandrotationalangleofthecollimators
ImageInversion
FluoroMode
Pulserate
Filter
Switchingon/offIBS(imagebrightnessstabilizationfunction)
Tubevoltage(IBSisconnectedwhenthetubevoltagestarts)
magnificationsizeII
https://translate.googleusercontent.com/translate_f
85/135
11/28/2016
OPERATIONMANUAL
Figure3:39MemoryKey
87
Page88
1Holdthebutton(forabout1second).
Thecurrentconditionsarestored.
WhenthestorageendtheLED.
Ifthestoredconditionschange,theLEDwillflash,indicatinga
incompatibilitybetweenthecurrentconditionsandstoredconditions
therelevantmemorykey.
2Pressthememorykeythatisflashing.
Theconditionsstoredinthememorykeywillbecalled.
Atthesametime,theLEDstopsblinkingandstayson
continuously.
differentusesforthetwomemorykeys
1HoldthememorybuttonAforabout1second.
ThecurrentconditionswillbestoredinmemorykeyA.
WhenthestoreisfulltheLEDwillflash.
2Modifytheconditionsasthecamerarotationalpositionmode
Fluoro.
MemorykeyAflashes.
3PresstheBbuttonbriefly.
ThemodifiedconditionswillbesavedinthememorykeyB.
Whenstorageiscomplete,theLEDwillflash.
TheLEDAoff.
4PressthememorykeyA.
TheAbuttonwilllight,andthesystemwillreturntothesavedconditionsin
https://translate.googleusercontent.com/translate_f
86/135
11/28/2016
OPERATIONMANUAL
A.memorykey
TheLEDBoff.
5PressthememorykeyB.
TheLEDBlightsupandthesystemwilladopttheconditions
savedinthememorykeyB.
TheLEDAoff.
88
Page89
3.9.2ChangingtheAPRadjustment(OpescopeActive)
YoucanchangetheAPRsettings.
1ChangetheconditionsasFluoromodesettingsandconditions
Xray.
Thefollowinginformationmaybelogged.
Camerarotationalposition
Positionopens/closesthecollimatorandrotationangle
ImageInversion
fluoroscopymode
Pulserate
Filter
On/OffIBS(imagebrightnessstabilizationfunction).
Tubevoltage(asIBSisconnected,thetubevoltage
starts)
magnificationsizeII
2Press[APR]andthekeyatthesametime.
TheadjustmentintheAPRwhosebuttonsarepressedwillbewritten.
Whenregistrationiscomplete,analarmwillsound.
https://translate.googleusercontent.com/translate_f
87/135
11/28/2016
OPERATIONMANUAL
Figure3:40RegisteringthesettingsmodifiedinAPR
89
Page90
3.9.3Changingthepresettimerinfluoroscopy
Youcanadjustthetimetothepresettimerinfluoroscopy.
Screenwithpresetvalueofthetimer
Figure3:41Changingthepresettimerfluoroscopy
1InFluoromode,pressthe[changingthetimerpre
adjustedfluoroscopy].
Thevalueofthetimersetwillbedisplayedflashing.
2Eachtimeyoupressthebuttonwhilethedisplayisflashing,the
timervaluewillbeincreasedby30seconds.
Youcanadjustthevalueofthetimerasfollows:
3thecourseof3secondsaftertheoperationinstep2,thescreenreturns
thecumulativedisplaytimefluoroscopy.
https://translate.googleusercontent.com/translate_f
88/135
11/28/2016
OPERATIONMANUAL
90
Page91
3:10DIGITALRADIOGRAPHY
OPESCOPEFULL
Thedigitalradiographyconditionsareautomaticallyadjusted
basedonfluoroscopicconditions.Forthisreason,beforetakingthe
Xray,youmustconnecttheIBSfunctionandperformfluoroscopyto
thefluoroscopicimageisstabilized.
1Checktheradiographicconditions.
Figure3:42ScreenRadiography
https://translate.googleusercontent.com/translate_f
89/135
11/28/2016
OPERATIONMANUAL
Thesefunctionsareonlydisplayedwhenatechniquethatinvolves
serialradiography,suchasDSAorserialisselected.
91
Page92
2Releasetheleftpedal.
Irradiationstartsandtheradiographicimageappears
onthemonitor.
Figure3:43Gettingradiography
https://translate.googleusercontent.com/translate_f
90/135
11/28/2016
OPERATIONMANUAL
92
Page93
Changingtheradiographicprogram(acquisitionprogram)
Changesmadethroughthemainpanelaretemporary.
1Press[selectionacquisitionprogram]
Theacquisitionscreenappears.
Figure3:44Selectionacquisitionprogram
https://translate.googleusercontent.com/translate_f
91/135
11/28/2016
OPERATIONMANUAL
93
Page94
2Selecttheacquisitionprogram.
Thebuttonishighlightedandthewindowcloses.
Figure3:45Selectionacquisitionprogram
[Next]Theselectionwindowacquisitionprogramshows
fouracquisitionprogramsonapage.Press[next]
toviewthenextprogramselectionpage
acquisition.
Ifyoudonotwanttochangetheselectionoftheacquisitionprogram,performthe
operationsdescribedbelowtoreturntothebasicscreenwithoutchangingthe
acquisitionprogram.
Press[Close]
StartAcquisition
https://translate.googleusercontent.com/translate_f
92/135
11/28/2016
OPERATIONMANUAL
94
Page95
Changingtheacquisitionrate
1Press[acquisitionrate]
Thewindowwiththeselectionoftheacquisitionrateisdisplayed.
Figure3:46ChangeofAcquisitionRate1
[Acquisitionrate]isonlydisplayedwhenatechniqueinvolving
continuousradiographyas"serial"or"DSA"isused.
Ifaprogramsequencehasanacquisitionrateexceeding5
fps,themodeishighspeedandtubevoltageandthequalityof
imagecanbechanged.
https://translate.googleusercontent.com/translate_f
93/135
11/28/2016
OPERATIONMANUAL
95
Page96
2Selecttheacquisitionrate
Thebuttonwillbehighlightedandtheselectionwindowoftheacquisitionrate
Itwillbeclosed.
Ifyouwanttokeeptherateofpurchase,pressthe[close]
Figure3:47ChangeofAcquisitionRate2
Figure3:48ChangeofAcquisitionRate3
https://translate.googleusercontent.com/translate_f
94/135
11/28/2016
OPERATIONMANUAL
96
Page97
Changingtheprogramsequence
Press[sequence]intheselectionwindowoftheacquisitionrateandmake
selection.Youwillneedtoadjusttheprogramsequenceonthemonitor.if
thereisnosetsequence,the[detailsdetails]not
appears.
1Press[details]
Theprogramsequenceselectionwindowappears.
Figure3:49acquisitionsequencedetails
2Press[]/[]toselecttheprogramsequence.
Theselectedprogramsequenceisdisplayed.
Figure3:50Selectionofprogramsequence
97
https://translate.googleusercontent.com/translate_f
95/135
11/28/2016
OPERATIONMANUAL
Page98
3Press[closes]
Theprogramsequenceselectionwindowisclosed.
ProgramSequenceDetails
Figure3:51ProgramSequencedetails.
Thefollowingdetailsscreenshowsdetailsoftheprogram
sequenceofselectedprograms.Itdisplays5levelsrecorded
acquisitionrateandtime.
98
https://translate.googleusercontent.com/translate_f
96/135
11/28/2016
OPERATIONMANUAL
Page99
3:11Tweaking
FullOpescope
Youcancalibratethetouchpanel,adjustthesound,setthetimer
Fluoroscopy,checkthesettingsstoredinmemory.
Makesurethedisplayisturnedoffbeforetheoperation.
Thechangesthataremadetotheunitwillremainthesameeffective
themachineisrestarted.
1Pression[Menu]
TheSystemSetupmenuscreenappears.
Figure3:52SystemSetupMenuScreen
99
https://translate.googleusercontent.com/translate_f
97/135
11/28/2016
OPERATIONMANUAL
Page100
Chapter4
Troubleshooting
Thischapterexplainsthecorrectiveactionstobetakeniftheequipment
presentproblem.
100
Page101
https://translate.googleusercontent.com/translate_f
98/135
11/28/2016
OPERATIONMANUAL
4.1Respondingtoproblems
Whenaproblemoccurswiththesystem,followtheinstructions
flowchartfollowingappears:
Ifyouhaveanyquestionsonprocedure,contacttechnicalassistance
authorized.
Ifyounoticeanyproblemswiththesystem,stopuseimmediately.
Repairsandmaintenanceshouldbecarriedoutonlybytechnicaland/or
authorizedservice.
101
Page102
https://translate.googleusercontent.com/translate_f
99/135
11/28/2016
OPERATIONMANUAL
FluoxogramaprocedureConcealer
Occurrenceofproblem
Ifamessage
appears,
seethelistof
errors.
No
Got
solvethe
problem?
Yes
END
Telephoneassistance
authorizedservice.
102
Page103
https://translate.googleusercontent.com/translate_f
100/135
11/28/2016
OPERATIONMANUAL
4.2Listoferrors
Ifanerroroccurs,thekeywilllightupdependingonthenatureof
errorandacodewillbedisplayedontheerrorscreen.
Code
Er1to9
Er1019
Er20to29
Er30to39
Er40to49
Er50to59
Er60to99
HEAT
Message
CPUerror
sourceoferrorofinternalenergy
controlcircuiterror
Errorinthehighvoltagegenerationcontrolsystem
Errorsassociatedwithtemperaturedetectionand
communications
Errorinmechanicalcontrolsystemcollimator
Errorsassociatedwiththesystem
Hightemperaturethroughoutthexraytube
Whilethebuttonisflashing,youcanperformfluoroscopy
radiograph.
Whilethekeyislitcontinuously,youcannot
performfluoroscopyandradiography.
Whenthexraytubeassemblyattainsahightemperature,
HEATcodeappears,andyoucannotperformradiographyand
fluoroscopy.Whenthetemperaturelowers,themessageis
automaticallyerasedandthenyoucanperformfluoroscopy
radiograph.
Figure4.1AssetOpescopeerrorscreen
103
Page104
https://translate.googleusercontent.com/translate_f
101/135
11/28/2016
OPERATIONMANUAL
Figure4.2OpescopeFullErrorScreen
4.3OperatingChange
Intheeventofoperatingchange,turnoffthemachineandenter
contactyourdealer.
ShimadzuofBrazilLtda
Address:MarquesAvenueStVincent,1771SP
Tel.:(11)21341688/Fax:(11)36112209
email:webmaster@shimadzu.com.br
website:www.shimadzu.com.br
104
Page105
https://translate.googleusercontent.com/translate_f
102/135
11/28/2016
OPERATIONMANUAL
Chapter5
MaintenanceandInspection
Thischapterexplainsthedailyandperiodicinspections.Inspectionsshouldbe
heldastomaintainequipmentperformanceandensure
safety.
105
Page106
https://translate.googleusercontent.com/translate_f
103/135
11/28/2016
OPERATIONMANUAL
5.1MaintenanceandInspections
Ourcompletequalitycontrolandinspectionprocedures
ensurethatthesystemistransportedinoptimalconditions.For
maintainoptimalconditions,weneedtoimplementandmaintain
inspection.
Themaintenanceandinspectionsforthesystemareclassifiedas
followbelow:
Typeofmaintenance/
inspection
startingthe
inspections inspections
daily
finalizingthe
inspections
Cleaninganddisinfection
periodicinspections
implementedby
Operatorormanager
equipment
Operatorormanager
equipment
Operatorormanager
equipment
Technicalassistance
authorized
Interval
Daily
Daily
Whennecessary
Every6months
5.2Dailyinspections
Dailychecksincludethebeginningandendofinspections.
Dailyinspectionsarenecessarytomaintainthesystemforalong
period.
Checkallinspectionsandcorrectiveactionshavebeencompletedbefore
tousethesystem.
106
Page107
https://translate.googleusercontent.com/translate_f
104/135
11/28/2016
OPERATIONMANUAL
ReferringtoFig.
5.3Cleaninganddisinfection
Youwillneedtocleananddisinfectthesystem.
Cleaninganddisinfectionareveryimportanttoensurethatthe
systemcanbeusedsafely.Youwillneedtofollowthemethods
indicatedbythemanufacturer
1Makesurethatthesystem'smainswitchisturnedoffanddisconnectthe
maincable.
Unplugthemaincablebeforeyoustartcleaning.
Youmustdisconnectthesystemfrommainsbeforestarting
cleaninganddisinfection.
Turnoffthesystembeforedisconnectingthemaincable.
Otherwisethereisariskofelectricshockandleakagecurrent.
2Performcleaninganddisinfection.
Cleantheoutersurfaceofthesystemwithadampcloth
glutaraldehydeina2%solution.
Donotusewateronthesystem.
Cleanthesystemwithacleanclothsoakedwithdisinfectant.
107
Page108
https://translate.googleusercontent.com/translate_f
105/135
11/28/2016
OPERATIONMANUAL
Donotspraydisinfectantdirectlyonthesystem.
Ifthedisinfectantintothesystemmaycauseproblemsor
accidents.
Donotusethebelowlistedproducts
Disinfectantschlorinebase
Disinfectantsthatcorrodemetals,plastics,rubbers,etc.
Disinfectantsatomized
volatileDisinfectants
Usetheminimumamountofdisinfectant
Overtime,thedisinfectionmaycausedamageto
paintdiscolorationordeteriorationofresinssuchasplasticand
rubbers.
3Aftercompletingthework,checkthefollowingpoints:
Thereshouldbenowaterordisinfectantjoinedthesystem.
Thetoolsusedtocleananddisinfectthesystemshouldbe
cleanafteruse.
108
Page109
https://translate.googleusercontent.com/translate_f
106/135
11/28/2016
OPERATIONMANUAL
Chapter6
optional
Thischapterexplainshowtooperatetheoptional(operation,maintenance,
setting,etc.).
109
Page110
6.1LaserIndicator(Laserpointer)
ThelaserconsistsofaprojectorunitintheXraytube,plate
withcentralidentificationintheimageintensifierandtheON/OFF
https://translate.googleusercontent.com/translate_f
107/135
11/28/2016
OPERATIONMANUAL
ofthelaserbeamonthetopside.
1Press"LaserPointer"onthemainpanelinitiatestheissueof
laserbeam.(Lamplightsuponthedashboard)
Alaserbeamisemittedwhileitappearsa
markingacrossonthemonitorscreenimage.
Thebuttonisilluminatedwhenactive.
Whenyoupressthebuttonagain,thelaserbeamis
interrupted.
2Pressingthesamekeyagainthelaserbeamisturnedoff.
Markingimagecrossdisappearsfromthemonitor.
Thelaserbeamcontinuestobeoutputwhilethebuttonis
on.
Figure6.1LaserPointer(IndicatorLaser)
110
Page111
Neverlookdirectlyintothelaserbeam.
https://translate.googleusercontent.com/translate_f
108/135
11/28/2016
OPERATIONMANUAL
laserradiationinfo:
Power:1.0mWmax.
Powerconsumption:5V200mA
CW:635nm
Class2(IEC8251)
Thisoptionwasdevelopedtomarktheareatobeexposedto
xraysmoreaccurately,soyoucanoptimizetheamountof
XraysirradiatedwhileavoidingdoseofXraysunnecessaryforthepatientand
medicalteam.
AdditionalInformation:
Thevaluesandindicationscanbeobservedinthemachinescreenor
onthemonitors.
111
Page112
6.2CassetteSupport
Youcanmountacassetteholderintheintensifierunit
Image.
https://translate.googleusercontent.com/translate_f
109/135
11/28/2016
OPERATIONMANUAL
Installthecassetteholdercorrectlyforitnot
presentlaterfailures.
Adjustingthecassetteholder
1Unscrewthescrew(1).
2Lettherunner(2)contactastheflangemounted(3),push
thecassetteholderforwardandlockthefixedguide(4)mountedontheflange
(3).
3Tightenthescrew(1)untilthetipcontactface
rearoftheblock(5).
Inthispositionthecassetteholdercanbemountedanditwill
rotatefreelyintheappendixoftheintensifierunit
image.
4Adjustthecassetteholderpositionaccordingtotheposition
Radiographic.
Figure6.2CassetteSupport
112
Page113
6.3HandTrigger
Thison/offcontroltheradiationXrays.
1Pressthehandshutterbuttontothefirstlevel(level
Radiographicpreparation).
Startspreparationforradiography.
https://translate.googleusercontent.com/translate_f
110/135
11/28/2016
OPERATIONMANUAL
TheStandbyindicatorlampequipmentwillturnoff
temporarily.
Whentheradiographicpreparationiscomplete,thelamp
standbyindicatorwilllightupagain.
2Pressthehandcontroltothesecondlevel(levelofpreparation
RadiographicandexposuretoXrays).
TheXrayradiationwillstart.
DuringexposuretoXrays,theradiationmonitorindicator
willlight.
113
Page114
https://translate.googleusercontent.com/translate_f
111/135
11/28/2016
OPERATIONMANUAL
Chapter7
Technicaldata
114
Page115
7.1TechnicalSpecifications
7.1.1)CararmCsupport
Item
FocusII
verticaloffset
horizontalscroll
axialrotationofthearc
Arcrotation
lateralmovement
https://translate.googleusercontent.com/translate_f
OpescopeActive
900mm
450mm
200mm
120th
420
12.5
FullOpescope
900mm
450mm
200mm
120th
420
12.5
112/135
11/28/2016
OPERATIONMANUAL
freeareabetweentube/II
DepthofArc
700mm
780mm(9II")
700mm
780mm(9II")
7.1.2)Unitcontroller/generatorxray
Item
Ratedpower
voltagerange(RAD/Fluoro)
mAsrange(RAD/Fluoro)
Maximumcurrent(RAD)
Maximumcurrent(FLUORO)
pulsedfluoroscopy
fluoroscopycontinues
FluoroscopyHighPulsed
Quality(BOOST)
fluoroscopytime
tobecontinued
Inverterfrequency
OpescopeActive
FullOpescope
2kW
2kW
40to110kVin1kVstep40to110kVin1kVstep
0.5to27insteps200mAs
0.5to27insteps200mAs
20mA
20mA
16mA
16mA
0.5/1/2/3.75/7.5and15fps0.5/1/2/3.75/7.5and15fps
Yes
Yes
Yes
Yes
45minutesormore
45minutesormore
50kHz
50kHz
7.1.3)tubexray
Item
caloriccapacityoftheanode
dissipationcapacity
Callusanode
caloriccapacity
set
SpotSize
AnodeAngle
anodematerial
Typeanode
feeding
inherentfiltration
OpescopeActive
100.000HU
FullOpescope
100.000HU
550W
550W
680.000HU
680.000HU
0.6mm2
8th
Tungsten
Fixed
Highfrequency
2,9mmAl
0.6mm2
8th
Tungsten
Fixed
Highfrequency
2,9mmAl
115
Page116
7.1.4)Collimator
Item
collimatoroctagonal
collimatedparallel
compensatingfilters
OpescopeActive
Motorizedwithbladesof
Lead(Pb)
Motorizedwithbladesof
Copper(Cu)andsystem
rotation.
Motor.Coppermade
(Cu)andthicknesses
0.1mm,0.2mmand0.3mm
FullOpescope
Motorizedwithbladesof
Lead(Pb)
Motorizedwithbladesof
Copper(Cu)andsystem
rotation.
Motor.Coppermade
(Cu)andthicknesses
0.1mm,0.2mmand0.3mm
7.1.5)Intensifierimages(II)
https://translate.googleusercontent.com/translate_f
113/135
11/28/2016
OPERATIONMANUAL
Item
Intensifier6Image"/4"
Intensifier7Image"
Intensifier9Image"/6"
Item
Sizeoffields
centralResolution
maxim
OpescopeActive
Yes
Optional
Optional
II6"/4"
6"/4"(16cm/10cm)
FullOpescope
Yes
II7"
7"(18cm)
54/64
50
II9"/6"
9"/6"(23cm/16cm)
50
7.1.6)ImageProcessingUnit
Item
Cameratype
acquisitionmatrix
Resolution
fluoroscopyacquisitionrate
serialacquisitionrate
Imageprocessing
OpescopeActive
CCD400kP
768x494(EIC)or752x582
(CCIR)
8bits
30frames/second
5frames/second
FullOpescope
CCD1MP
1024x1024
12bits
30frames/second
15frames/sec
windowsettings/level,brightness
andcontrast,gamma,highlight
windowsettings/level,brightnessedges,zoom/pan2xzoom,
andcontrast,gamma,highlight displayingmultipleimages
edges,zoom/pan2xzoom,
anascreenannotation,measurements,
displayingmultipleimages
SubtractionAngiography
anascreenannotation,measurements,
Digital(DSA)inrealtime,
SubtractionAngiography
remascara,displacement
Digital(DSA)inrealtime,
pixel(pixelshift),landmarking
remascara,displacement
(Anatomicalreference)
pixel(pixelshift),landmarking
positive/negativeinversion,
(Anatomicalreference)
right/leftinversion,
positive/negativeinversion,
reversingup/downmode
right/leftinversion,
roadmappingcineandRSM
reversingup/downmode
DSA(angiography
cineandroadmapping.
DigitalSubtractionintime
realtimewithmask
real).
116
Page117
AcquisitionofDSA
Capacityof
imagestorage
DICOM3.0protocol
0.5to5frames/second
0.5to7.5frames/second
Upto30,000imagesonthedisc
Upto16,770imagesonthedisc
hard
hard
Upto100imagesinmemory
Upto68imagesinmemory
RAM
RAM
Upto300images/CDR
DICOM3.0format
DICOMPrint
DICOMStorage(send/receive)
DICOMWorklist
DICOMMediaStorage
7.1.7)Monitor
Item
LCDMonitor18"
LCDmonitor18"touchscreen
https://translate.googleusercontent.com/translate_f
OpescopeActive
Optional
Optional
FullOpescope
Optional
Optional
114/135
11/28/2016
OPERATIONMANUAL
LCDMonitor15"
CRTMonitor17"
LCDMonitor19"
Monochrome
LCDmonitor19"touchscreen
LCDMonitor19"color
Optional
Yes
Optional
Optional
Optional
Yes
Optional
Optional
Optional
Optional
7.1.7)antidifusoraGrade
Item
Gridreason
griddensity
Interespaador
OpescopeActive
10:1or8:1
40lines/cm
Aluminum(Al)orfiber
FullOpescope
10:1or8:1
40lines/cm
Aluminum(Al)orfiber
Note:Thetechnicalinformationsuchascircuitdiagrams,lists
components,descriptions,instructionsforcalibrationandtestingareprovided
bythemanufacturertostaffand/orqualifiedandauthorizedtechnicalassistance
Brazil.
117
Page118
https://translate.googleusercontent.com/translate_f
115/135
11/28/2016
OPERATIONMANUAL
Figure7.1xraytube
Subtitle:
Referenceaxistheaxisofreference
Focalspotfocalpoint
UnitUnit
118
Page119
https://translate.googleusercontent.com/translate_f
116/135
11/28/2016
OPERATIONMANUAL
Figure7.2Heatingandcoolingcurvesofthexraytubeanodeand
Subtitle:
XRaytubeassemblyheatingandcoolingcurveheatingandcoolingcurve
xraytube
AnodeHeatingandCoolingCurveheatingandcoolingcurveoftheanode
119
Page120
https://translate.googleusercontent.com/translate_f
117/135
11/28/2016
OPERATIONMANUAL
Figure7.3Loadevaluationboardsandcharacteristicsofemissionandfilament
Subtitle:
SingleLoadRatingChartsApplicationAssessmentFrameworksinasinglecharge
AndFilamentEmissionCharacteristicsemissioncharacteristicsandfilament
XrayTubecurrentTubecurrentxray
TimeTime
FilamentCurrentFilamentCurrent
FilamentVoltageFilamentvoltage
120
Page121
Monitor
Kind
LCDMonitor18"
LCDmonitor18"touchscreen
LCDMonitor15"
CRTMonitor17"
https://translate.googleusercontent.com/translate_f
Specification
ScreenMatrix:1280x1024
ScreenType:flat(flat)withtechnology
TFT
ScreenSize:18"(diagonal)
ScreenMatrix:1280x1024
ScreenType:flat(flat)withtechnology
TFTandtouch
ScreenSize:18"(diagonal)
ScreenMatrix:1280x1024
ScreenType:flat(flat)withtechnology
TFT
ScreenSize:15"(diagonal)
ScreenMatrix:1280x1024
ScreenType:CRT
ScreenSize:17"(diagonal)
ScreenMatrix:1280x1024
ScreenType:flat(flat)withtechnology
118/135
11/28/2016
OPERATIONMANUAL
LCDMonitor19"Monochrome
LCDmonitor19"touchscreen
LCD19"ColorMonitor
TFT
ScreenSize:19"(diagonal)
ScreenMatrix:1280x1024
ScreenType:flat(flat)withtechnology
TFTandtouch
ScreenSize:19"(diagonal)
ScreenMatrix:1280x1024
ScreenType:flat(flat)withtechnology
TFT
ScreenSize:19"(diagonal)
Videotape
Videosignal:NTSCstandardcolorsignal
Power:AC120V,60Hz,0.6A
Videosignal:
Input:BNC1.0Vpp,75ohms.Sin(4P)Y:1.0Vpp,75ohms.W:
0.286Vpp,75ohms.
Output:BNC1.0Vpp,75ohms.Sout(4P)Y:1.0Vpp,75ohms.W:
0.286Vpp,75ohms.
Audiosignal:
Input:LINEINCH1/CH2(PINRCA)10dBu,47kohms.Microphone
121
Page122
NormalaudioCH1(RCAJackQ)to600ohmmic60dBu,3kohms.
Output:AUDIOMONITOR(RCAPIN)5dBu,47kohm.LINEOUTCH
1/CH2(RCAPIN)10dBu,47kohm.
tapeformat/speed:SVHS/VHS/33.35mm/sec
VideoColorPrinter
Resolution:403DPI
PrintedImagesize:(NTSC)3"x27/8"/5"x35/8"
Printspeed(NTSC)1929seconds
Powersupply:100120volts,1.8A220240volt,1.0A
VideoPrinterBlackandWhite
Resolution:1280DOT
Effectivepixels:508linesx1280dots
PrintSpeed:3.07.0onthe2nd
https://translate.googleusercontent.com/translate_f
119/135
11/28/2016
OPERATIONMANUAL
Powersupply:100120volts,1.8A220240volt,1.0A
optional
TapeSupport
IndicatorLaser
SterilisationCase
HandControl
LCDMonitor
VideoSonyPrinter
DVDrecorderSony
ExtensionDistanceSourceSkin
OptionalOPESCOPEFULL
LCDMonitorTouch
DICOM3.0Kit(Print(IncludesCDROMwithsoftware)/Storage(send/receive)
(IncludesCDROMwithsoftware)andWorklist(IncludesCDROMwithsoftwareandPC,
consistsofMonitor,CPU,mouseandkeyboard))
122
Page123
FluoroimageinSupportCarDisplaywithCArm
Dosemeter(DAP)
OptionalOPESCOPEASSETS
VariableField(9"/6")toII
Fibergrid
HandleControlII(ImageIntensifier)
TypeMemory"LOOP"
LCDMonitor6.4"
DigitalImageProcessingUnitwithSubtractionAngiography
Digital(DSA)
MODRecorder
MODMedia
WeightCarHolderwithArmC:250kg
WeightCartMonitors:145kg
https://translate.googleusercontent.com/translate_f
120/135
11/28/2016
OPERATIONMANUAL
Thecontrollerunitofxrayconstitutesthejunctionhousing(outsideand
xraygenerator)
TheCarchsupport,supportxraytube,xraytube,collimator,
highvoltagegenerator,imageintensifier,powersupplyand
controllerunitxraysareofexclusiveuseoftheequipment.
Thelaserpoint,monitors,standandsupportDICOMkitaremanufacturing
Shimadzu,buttheymaybeusedinotherShimadzuequipment.
ThevideoprinterandvideocassettearenotmanufacturedbyShimadzu,
gettingthepurchasethecustomer'sdiscretion.(Seeminimumspecifications)
graphicillustrations
123
Page124
Figure7.4xraytube
https://translate.googleusercontent.com/translate_f
121/135
11/28/2016
OPERATIONMANUAL
Figure7.5Collimator
124
Page125
Figure7.6highvoltagegeneratorphotos
https://translate.googleusercontent.com/translate_f
122/135
11/28/2016
OPERATIONMANUAL
Figure7.7imageintensifierPhoto
125
Page126
Figure7.8Monitor
https://translate.googleusercontent.com/translate_f
123/135
11/28/2016
OPERATIONMANUAL
Figure7.9Figurepowersupplyboard
Subtitle:
2Baseboard
3Limiter
4Label
5Spacers
7Limiter
8,9,10,11and12Powersupply
13Filter
14Label
126
Page127
Figure7.10Laserpointillustration
https://translate.googleusercontent.com/translate_f
124/135
11/28/2016
OPERATIONMANUAL
Figure7.11StandformonitorPhoto
127
Page128
https://translate.googleusercontent.com/translate_f
125/135
11/28/2016
OPERATIONMANUAL
Figure7.13SupportCarBoomCwith9II"
128
Page129
https://translate.googleusercontent.com/translate_f
126/135
11/28/2016
OPERATIONMANUAL
Figure7.14SupportCarBoomCwith7II"
129
Page130
Figure7.15SupportCarBoomCIIwith6"/4"
https://translate.googleusercontent.com/translate_f
127/135
11/28/2016
OPERATIONMANUAL
130
Page131
7.2ConditionsforInstallationandUse
1.EnvironmentalConditions:
Temperature
10Cto40C
Relativehumidity
30%to85%(noncondensing)
Atmosphericpressure
700hPato1060hPa
Atmosphereconditions
Theatmosphereshouldbegasfree
corrosiveandexplosive
2.SourceofEnergy:
Equipment
SimplephaseAC
Frequency
50or60Hz
Capacity
3.5kVA
VoltageandImpedance
100,110,120V10%0.2Max.
200,220,230,240V10%0.8Max.
Earth
typeD
3.PowerConsumption
Inuse:
35A(100,110,120V),17,5A(200.220,230,240V)
standby:
15A(100,110,120V),10A(200.220,230,240V)
Installation
Installationoftheequipmentshouldbeperformedonlybypersonnel
and/ordealer.
https://translate.googleusercontent.com/translate_f
128/135
11/28/2016
OPERATIONMANUAL
AuthorizedTechnicalAssistance
ShimadzuofBrazilLtda
Address:MarquesAvenueStVincent,1771SP
Tel.:(11)21341688/Fax:(11)36112209
email:webmaster@shimadzu.com.br
website:www.shimadzu.com.br
131
Page132
7.3ConditionsforStorage
Temperature
10Cto40C
Relativehumidity
30%to85%(noncondensing)
Atmosphericpressure
700hPato1060hPa
Atmosphereconditions
Theatmosphereshouldbegasfree
corrosiveandexplosive
7.4DoseInformation
7.4.1LocationofSignificantZonesofOccupation
ArcCposition
TheforwardverticalpositionseeFigure7:16
AnteroposteriorverticalpositionseeFigure7:17
HorizontalpositionseeFigure7:18
SignificantzoneofoccupationandmeasurementpointsseeFigures
7:16,7:17and7:18
Fantom
Externaldimensions:35x35x15(cm)
https://translate.googleusercontent.com/translate_f
129/135
11/28/2016
OPERATIONMANUAL
132
Page133
https://translate.googleusercontent.com/translate_f
130/135
11/28/2016
OPERATIONMANUAL
Figure7.16PositionposteroanteriorVertical(measuredinmm)
133
Page134
https://translate.googleusercontent.com/translate_f
131/135
11/28/2016
OPERATIONMANUAL
Figure7:17PositionAnteroPosteriorVertical(measuredinmm)
134
Page135
https://translate.googleusercontent.com/translate_f
132/135
11/28/2016
OPERATIONMANUAL
Figure7.18Horizontalposition(measuredinmm)
135
Page136
Heightfromtheground
Figure7.19posterioranteriorpositionvertical(medicalcm)
https://translate.googleusercontent.com/translate_f
133/135
11/28/2016
OPERATIONMANUAL
Heightfromtheground
Figure7.20anteroposteriorverticalposition(measuredincm)
136
Page137
Heightfromtheground
Figure7.21Horizontalposition(measuredincm)
RepresentativeandTechnicalAssistanceinBrazil
ShimadzuofBrazilLtda
AvenueMarquesdeSoVicente,1771BarraFunda
SoPauloSPCEP:01139003
Tel.:(11)21341688
Fax:(11)36112209
email:marketing@shimadzu.com.br
https://translate.googleusercontent.com/translate_f
134/135
11/28/2016
OPERATIONMANUAL
______________________
Technicalmanager
ConstantineLuizDiPIPI
CREA/SP0682526854
________________________
Legalresponsible
AkihikoKato
CEO
137