Professional Documents
Culture Documents
Border Separation For Adjacent Orthogonal Fields
Border Separation For Adjacent Orthogonal Fields
BORDER
SEPARATION
FOR ADJACENT
1991 Amencan
0739-021 l/91
Association
of Medical
ORTHOGONAL
$3.00 + .oo
Dosnnetrists
FIELDS
INTRODUCTION
S=d-
iL
SSD
(1)
ANALYSIS
80
Medical Dosimetxy
(4
(b)
Fig. 1. Orthogonal radiation fields incident on a cylindrical
phantom: (a) light fields on the surface and gap width S; (b)
intersection of the fields internally.
nal cord at distance S below the mastoid tip. The superior border of the posterior spinal field is positioned so
as to pass through point A. Distance S is given by
(2)
where T is the lateral head thickness at the mastoid
tip, ;L, is the distance from the center of the field to
the field border closest to point A in the plane perpen-
81
fields used to treat the spinal
NOTATION:
. . This
[recommendation] is based on very precise ahgnment of fields which may not be obtainable in all departments.
Treating without a gap may cause concern to individual radiation
oncologists. Therefore, discretion is allowed to individuals to put in
a gap if they have reservations about this recommendation. The gap
should normally be not greater than 0.5 cm. .
12.26634
A wedge filter may be used to broaden the penumbra around
the posterior border of the large head field so that an appropriate
match is made with the penumbra of the leading edge of the posterior spinal field.
12.26635
The junction line is moved 1 cm cephalad after each 1260
cGy to smooth out any dose inhomogeneity occurring at the junction.
We constructed a polystyrene copy of the Alderson Rando phantom* with sections that are interchangeable with the original phantom. The polystyrene sections each have a &-in hole at the spinal cord
(usually about 4 cm from the back). The spinal cord
holes were filled with polystyrene plugs partially
sliced at .5locations, 5 mm apart, each slice becoming
a shelf, holding a $ in. X $ in X .015 in LiF thermoluminescent dosimeter (TLD) chip snugly. (See Fig. 5.)
See a similar study described in Ref. 13.
We chose to study the rotated, orthogonal setup
shown in Fig. 4 using a cobalt-60 therapy unit with a 2
cm diameter source and a 45 cm source to collimator
distance. Phantom sections numbered 4 through 8
(Rando phantom section numbers) were loaded with
25 TLD chips, spaced 5 mm apart, in a line along the
spinal cord. Sections 4 and 5 were wholly in the cranial fields. By adjusting the upper border of the spinal
field according to the geometrical method described
above, sections 7 and 8 were wholly in the spinal field
and section 6 was in the gap region.
The cranial fields had an SSD of 80 cm, the unblocked field size was 2 1 X 16.5 cm, the effective field
size was 19 X 15 cm (the field borders extended
beyond the head and were partially blocked as shown
in Fig. 4) and the lateral head thickness at the mastoid tip was 13.7 cm.
The lateral head thickness at the mastoid tip was
13.7 cm, the unblocked field size of the cranial fields
was 2 1 x 16.5 cm, the effective field size was 19 X 15
cm (the field borders extended beyond the head and
were partially blocked as shown in Fig. 4) and the
SSD was 80 cm. The calculated gap was 0.7 cm. The
spinal field had an SSD of 160 cm, an open field size
* Alderson Research Laboratory, Stamford, CT.
82
Medical Dosimetry
AND
DISCUSSION
Phantom
ity of the gap. The spinal field falls off because the gap
lies at the end of a long narrow field. The cranial fields
fall off because of the loss of phantom scatter at point
A. Matching the field borders according to the geometrical method results in overlapping the 45% isodose
lines not the 50% isodose lines. Figure 7 shows that a
25% local cold spot results from using the gap calculated according to Eqn. 2.
Figure 8 shows the effect of widening or reducing
the field separation. The curves shown were constructed by adding the cranial dose distribution to the
spinal dose distribution after shifting their spatial relationship by a variable gap error. Widening or reducing
the field separation by l/2 cm increases or decreases
the junction dose as shown. Van Dyke et al. show
the results of measurements,
similar to the ones de-
Holder
Top View
of Phantom
Sections
90
60-
4
Position
6
Along
Spinal
8
Cord
IO
12
(cm)
Fig. 6. Dose distribution at the inferior edge of the cranial field and at the superior edge of the spinal field. Doses
were measured with TLD chips along the spinal cord. Dose is normalized to 100% at the center ofthe brain and at
al.
83
108
calculated
.$
84 80 76 -
Sect/on
, Section
4
Position
6
8
along Spinal Cord (cm)
10
12
of the gap region, but moving the junction would reduce its significance. According to the graph shown in
section 12.266 1 of the CCSG protocol. the hot spot
should only have been about 5%.
Differences in beam characteristics and patient
shape will result in different dose distributions in the
junction region. To be sure of the dose distribution
everywhere in the junction region, isodose distribu-
180
160
120
2
-
100
8
g
80
60
Fig. 8. Dose distribution in the gap region. The curves represent the arithmetic sum of the dose distribution
measured in the cranial field alone and in the spinal field alone, as shown in Fig. 6, after adding or subtracting the
indicated distance from the calculated gap. Dose is normalized to 100% at the center of the brain, and at a depth of
4 cm on the central axis of the spinal field.
Medical Dosimetry
84