Professional Documents
Culture Documents
Hioki PW3360 21.compressed
Hioki PW3360 21.compressed
Hioki PW3360 21.compressed
Now with
QUICK SET
Convenience
Harmonic Measurement Model
PW3360-21
Function
Enhancement
Make proper
connections
simply by
observing
the colors of
the displayed
leads.
PASS
Check wire
connection
status and
judgment
indicators.
FAIL
If FAIL appears, move the cursor
to the indicator and press the [ENTER] key.
Corrective action
tips appear
Select the
current range
Correct
Orientation
FAIL
PASS
Load side
Changed
I3 Clamp
P: 20.6kW
P dem+
(kW)
Maximum
Demand
Values
Automatically refreshes
with latest values
Time
Switched Display
P demkW
Time
Power Sold
Average Value
Minimum Value
Maximum data
Maximum data
Average
data
Graph Display
Measured value
(one selected item)
Automatically refreshes
with latest values
Average
data
Minimum
data
Minimum data
Record 3 data points
per interval
Time
Record to
SD card
Time
* Store up to one years data acquired at one minute intervals. Performance cannot be guaranteed on storage media other than Hioki-specified SD card options.
48 mm
Compact
(1.89")
100 mm
(3.94")
180mm (7.09")
In dim environments
Easy-to-see color LCD
8-Hour
operation
Voltage Line
Power Adapter
Power Source
Obtains power from the measurement lines
50C
-10C
Pulse Input
The pulse input function can be used to record power data and
production volume counts simultaneously. The power data and
pulse volume (production volume) information are useful for unit
cost production management.
Load
Power Source
Load Load
Power Source
Load
Pulse Input
Pulse Output
Use the Pulse Output function to acquire temperature and pulse (electrical energy) data simultaneously with a data
logger. Evaluate the relationship between air conditioner temperature control settings and power consumption.
Load
Power Source
Application Example
Temperature
Measurement
Pulse Output
Options
Load
30mm
(1.18 in)
Load
Power Source
Load
9675
9657-10
Ideal for quick investigation of intermittent leakage by continuous calculation processing every 200 ms.
(Select to save the average, maximum and/or minimum at every interval.)
PW3360-21
(vector display)
Remote
Monitor
Power Source
Load
Power Source
LAN
HUB
Files recorded in the Clamp On Power Logger's internal memory or SD card are accessible via a LAN or
USB connection, and are downloadable using the free PW3360 Setup and Download Software.
Click
Click
USB or LAN
Indication example
PC
Processing
Data saved to an SD card or internal memory can be loaded into a PC for expanded display, aggregation and analysis.
Supported models: PW3360, PW3365, 3169-20
Air Conditioner A
Power Source
Air Conditioner B
Power Source
Air Conditioner C
Power Source
Input specifications
Measurement
line type
Measurement
line Frequency
Number of input
channels
Voltage range
Measurement items
Voltage
Current
50/ 60 Hz
Voltage: 3 channels U1 to U3
Current: 3 channels I1 to I3
600 V AC
Total display area: 5V to 1000 V (less than 5 V displays as 0 V)
When RMS voltage is zero, zero is displayed for all orders of
harmonic voltage.
Effective measurement range: 90 V to 780 V, peak: 1400V
[OVER] indicates over-range warning
Current ranges
Load current
CLAMP ON SENSOR 9694
: 500m/1/5/10/50 A
CLAMP ON SENSOR 9695-02
: 500m/1/5/10/50 A
CLAMP ON SENSOR 9660
: 5/10/50/100 A
CLAMP ON SENSOR 9695-03
: 5/10/50/100 A
CLAMP ON SENSOR 9661
: 5/10/50/100/500 A
CLAMP ON SENSOR 9669
: 100/200/1k A
AC FLEXIBLE CURRENT SENSOR CT9667-01 : 50/100 /500/1k/5k A
AC FLEXIBLE CURRENT SENSOR CT9667-02 : 50/100 /500/1k/5k A
AC FLEXIBLE CURRENT SENSOR CT9667-03 : 50/100 /500/1k/5k A
Leakage current
LEAK CLAMP ON SENSOR 9657-10 : 50m/100m/500m/1/5 A
LEAK CLAMP ON SENSOR 9675
: 50m/100m/500m/1/5 A
Total display range: Within 0.4 to 130% of the range
(zero is suppressed for less than 0.4%)
When RMS current is zero, zero is displayed for all orders of
harmonic current.
Effective measurement range: Within 5 to 110% of the range
peak: 400% of range, however, maximum range is 200%.
[OVER] indicates over-range warning
300.00 W to 9.0000 MW
Power ranges
Depends on voltage/current combination and measured line
type (see Measurement Range Configuration Tables)
Total display range: Within 0 to 130% of the range
(0W display indicates zero rms voltage and/or current)
When RMS voltage and current are zero, zero is displayed
for all orders of harmonic active power and harmonic reactive
power.
Effective measurement area: Within 5 to 110% of the range
Any (0.01 to 9999.99)
VT ratio settings
Selections (1/60/100/200/300/600/700/1000/2000/2500/5000)
Any (0.01 to 9999.99)
CT ratio settings
Selections (1/40/60/80/120/160/200/240/300/400/600/800/1200)
Voltage: Insolated inputs (except between U1, U2, U3 and N)
Input methods
Current: Isolated input using a clamp-on sensor
Voltage input part: 3 M 20% (50/ 60 Hz)
Input resistance
Maximum rated voltage Voltage input section: 1000 VAC, 1400 Vpeak
Current input section: 1.7 VAC, 2.4 Vpeak
between terminals
Voltage input section: 600V Measurement Category III
Maximum rated
300V Measurement Category IV
voltage to earth
Current input section: Depends on clamp sensor in use.
Pulse input
Input specifications No-voltage contact input (counts when shorted terminals open)
Voltage input (Hi: 2 V to 45 V, Lo: 0 V to 0.5 V, counts at Lo to Hi)
Maximum rated input between terminals: 45 V DC
Maximum rated input to ground: not isolated (GND is equipment common)
Measurement range 0 to 9999 (maximum pulse count per save interval)
Filter On (for mechanical contacts) 25 Hz or less, and at least 20
Filter
ms Hi and Lo pulse width
Filter Off (for solid-state contacts) 5 kHz or less, and at least 100
s Hi and Lo pulse width
Displays product of pulse count and scaling factor setting
Scaling
Setting ranges: 0.001 to 1.000, and 1.000 to 100.00
Power
Demand
Harmonic
Measurement screen
List
Voltage RMS value, current RMS value, frequency, total active power,
total reactive power, apparent power, power factor or displacement
power factor, active energy (consumption), elapsed time
Voltage RMS value, voltage fundamental wave value,voltage
waveform peak, voltage fundamental wave phase angle,
current RMS value, current fundamental wave value, current
waveform peak, current fundamental wave phase angle
Per-channel and total active power, apparent power, reactive
power,power factor or displacement power factor
Active energy (consumption, regeneration), reactiv energy (lag,lead),
recording start time, recording stop time, elapsed time, energy cost
Active power demand value (consumption, regeneration),
reactive power demand value (lag, lead),
power factor demand value, or pulse input
Displays the maximum active power demand value and the time
at which it occurred (this information is not saved).
(data from up to 48 intervals is internally stored, then refreshed oldest-first).
Graph (voltage, current and power levels, content percentage and phase angle)
List (voltage, current and power levels, content percentage and phase angle)
Displays voltage and current waveform, voltage and current
RMS values, and frequency.
With a 3P3W3M connection, displays the phase voltage
waveform from the virtual neutral point.
Enlarged view of 4 user-selected parameters
For one selected measurement item (except demand and harmonics),
displays maximum, average and minimum values, with cursor
calculations available (Note: with Trend display, there is no poweroff backup function).
U/I
Power
Integ
Demand
Harmonic
Waveform
Zoom
Trend
SD card Interface Settings data, measurement data, screen data, waveform data
10BASE-T/100BASE-TX IEEE802.3 Compliance
LAN interface
- HTTP server function
- Download settings and data by communication application program
USB Ver 2.0, Windows 8 (32/64bit)/Windows 7 (32/64bit) /
USB interface
Vista (32bit) /XP
- When connected to a computer, the SD Card and internal
memory are recognized as removable storage devices.
- Download settings and data by communication application program
Pulse output
Function
Pulse rate
Pulse width
Output signal
WIRE SPECIFICATIONS
Pulse output
device
Pulse counter
9
Specifications in orange available in Model PW3360-21 only
General Specifications
Display device
Operating
environment
Operating
temperature and
humidity
(no condensation)
Storage
temperature and
humidity
(no condensation)
Dielectric strength
Applicable standards
Power supply
Charge function
Maximum rated
power
Continuous
battery
operation time
Backup battery life
Dimensions
Mass
Accessories
Standard
Window width
Points per window
Analysis orders
THD calculation selection
Analysis items
Measurement
accuracy
Measurement Specifications
Connection
Simultaneous
power/current
measurement modes
Calculation
selection
Measurement
accuracy
Functions Specifications
Trend graph
display function
Summary display
function
Waveform display
Recording Specifications
Harmonic display
Copy function
Print function
Report printing
10
9694
9660
9661
9669
9695-02
Insulated conductor
Not CE marked
Not CE marked
Appearance
Measurable conductor
diameter
Primary current rating
Accuracy
Cord length: 3 m
(9.84ft)
Cord length: 3 m
(9.84ft)
Cord length: 3 m
(9.84ft)
15mm (0.59")
15mm (0.59")
46mm (0.81")
5A AC
0.3% rdg.
0.02% f.s.
Within 2
100A AC
0.3% rdg.
0.02% f.s.
Within 1
500A AC
0.3% rdg.
0.01% f.s.
Within 0.5
Frequency characteristic
Effect of external magnetic field
(with a magnetic field of 400 A/ m AC)
Within 1.5%
CAT III 600Vrms
1000 A continuous
Within 0.5%
CAT III 300Vrms
60 A continuous 130 A continuous
5 21D (0.83") mm
5 21D (0.83") mm
5 42D(1.65") mm
50.5W(2.28") 5 58H(2.28")
5 18.7D(0.74") mm
100 mm
(3.94")
Phase
Frequency characteristic
CT9667-03
Mass
Power supply
180 mm
254 mm
(7.09")
(10.00")
500A AC/5,000A AC
2.0% rdg. 0.3% f.s.
Within 1
Cord length: 3 m
(9.84ft)
40mm (1.57")
Measurable conductor diameter
10A AC*
Primary current rating
Accuracy Amplitude (45 to 66 Hz) 1.0% rdg. 0.05% f.s.
Phase angle (@50 or 60 Hz)
Within 3
Frequency characteristic
Within 5%
40Hz to 5kHz
(deviation from accuracy)
Within 3.0%
CAT III 1000Vrms, CAT IV 600Vrms
10000 A continuous
35W (1.38") 120H (4.74") 34D (1.34") mm
7.4 mm(0.29)
13 mm(0.51)
Insulated conductor
Appearance
Within 3 dB
Dimensions
100A AC
0.3% rdg.
0.02% f.s.
Within 1
Equivalent to 1 A or less
(45 to 66Hz)
50A AC
0.3% rdg.
0.02% f.s.
Within 2
Appearance
(45 to 66Hz)
15mm (0.59")
Within 1.0%
15mm (0.59")
Within 2.0%
Within 0.5%
Effect of conductor position
CAT III 300Vrms
CAT III 600Vrms
Maximum rated voltage to earth CAT III 300Vrms
130 A continuous
550 A continuous
Maximum input (45 to 66Hz) 50 A continuous
46W (1.81") 5 135H (5.31") 46W (1.81") 5 135H (5.31") 77W (3.03") 5 151H (5.94")
Dimensions
55mm (2.17"),
80 (3.15")20 (0.79")mm
1000A AC
1.0% rdg.
0.01% f.s.
Within 1
Within 1.0%
40Hz to 5kHz
(deviation from accuracy)
Mass
9695-03
Insulated conductor
Cord length: 3 m
(9.84ft)
30mm (1.18")
10A AC*
1.0% rdg. 0.005% f.s.
Within 5
Within 5%
7.5 mA max.
7.5 mA max.
Within 0.1%
Insulated conductor
Within 0.1%
Insulated conductor
30 A continuous
10 A continuous
Dimensions
74W(2.91") 145H(5.71")
42D(1.65")
60W(2.36") 112.5H(4.43")
23.6D(0.95")
(45 to 66Hz)
Mass
Notes
PW3360-20 and PW3360-21 with Z4001 2-GB SD card, measuring 3P3W2M wiring
Saved Items: ALL data (Saves all data: average, maximum, and minimum values)
Screen save: OFF Waveform save: OFF
Interval time
1 seconds
2 seconds
5 seconds
10 seconds
15 seconds
Save Time
PW3360-20
PW3360-21
PW3360-21
(Saving of harmonic (Saving of harmonic
data: OFF)
data: ON)
15.9 days
24.7 hours
31.9 days
2.1 days
79.7 days
5.1 days
159 days
10.3 days
242 days
15.4 days
Interval time
30s
1 minutes
2 minutes
5 minutes
More than
10 minites
Save Time
PW3360-20
PW3360-21
PW3360-21
(Saving of harmonic (Saving of harmonic
data: OFF)
data: ON)
1 year
30.8 days
1 year
61.7 days
1 year
123 days
1 year
308 days
1 year
1 year
11
Voltage
Connection
1P2W
1P3W
1P3W1U
600.00 V
3P3W2M
3P3W3M
3P4W
600.00 W
1.2000 kW
6.0000 kW
12.000 kW
60.000 kW
900.00 W
1.8000 kW
9.0000 kW
18.000 kW
90.000 kW
*1. For the 9694 sensor, the range of guaranteed accuracy is from 500 mA to 5 A, and for the 9695-02, from 500 mA to 50 A.
Current
Voltage
Connection
500.00 A
300.00 kW
1P2W
1P3W
1P3W1U
6.0000 kW
12.000 kW 60.000 kW
120.00 kW
600.00 kW
600.00 V
3P3W2M
3P3W3M
9.0000 kW
18.000 kW 90.000 kW
180.00 kW
900.00 kW
3P4W
*2. For the 9660 and 9695-03 sensors, the range of guaranteed accuracy is from 5 A to 100 A, and for the 9661, from 5 A to 500 A.
Current
CLAMP ON SENSOR 9669
Voltage
Connection
1P2W
1P3W
1P3W1U
600.00 V
3P3W2M
3P3W3M
3P4W
100.00 A
60.000 kW
120.00 kW
200.00 A
120.00 kW
240.00 kW
1.0000 kA
600.00 kW
1.2000 MW
180.00 kW
360.00 kW
1.8000 MW
Current
Range
1P3W1U
3P3W2M
3P3W3M
3P4W
Display range
650 A
Effective measurement range
2A
25 A
550 A
130 A
Effective measurement range
5A
110 A
65 A
Effective measurement range
100 A
0.4 A
50 A
0.2 A
2.5 A
0A
1P3W
600.00V
Range
500.00 A
100.00 A
50.000 A
10.000 A
5.0000 A
9660, 9695-03
Range
9669
1.3% rdg. 0.11% f.s.
1.3% rdg. 0.15% f.s.
1.3% rdg. 0.2% f.s.
1.0000 kA
200.00 A
100.00 A
Range
5.0000kA
1.0000kA
500.00A
100.00A
50.000A
9661
0.6% rdg. 0.11% f.s.
0.6% rdg. 0.15% f.s.
0.6% rdg. 0.2% f.s.
0.6% rdg. 0.6% f.s.
0.6% rdg. 1.1% f.s.
-01
CT9667-02
-03 5000A range
2.3% rdg. 0.4% f.s.
2.3% rdg. 1.6% f.s.
2.3% rdg. 3.1% f.s.
-01
CT9667 -02
500A range
-03
10 A
100 A
1000 A
13 A
10 A
0.04 A
0.5 A
5A
0A
100 mA
6.5 A
11 A
5.5 A
0.25 A
1A
10 A
100 A
Input Current
n Measurement accuracy
55 A
1A
0.02 A
500 A
Current
Conditions of guaranteed accuracy After 30 minute warm-up, with 50/60 Hz sine wave input
Apparent power
Reactive power
Accessories
VOLTAGE CORD L9438-53 (1 set) , AC ADAPTER Z1006 (1) , USB cable (1),
instruction manual (1), measurement guide (1), color spiral tubes (1 set): red, yellow, blue/two
each, for color-coding clamp sensors, spiral tubes for grouping clamp sensor cords (5)
AC ADAPTER Z1006
n Options
CLAMP ON ADAPTER
9290-10
Secondary side
100A
When purchasing the 9695-02 and 9695-03, we recommend also purchasing the
separately sold 9219 Connection Cord.
BATTERY SET
PW9003
Rated voltage: 240 V AC
(supplies power from measurement lines) Operating temperature and humidity range: -10 to 50C, 80% RH or less
Z4001
CARRYING CASE
Power Source
9804-01 Red
PW9002
MAGNET ADAPTER
C1005
SF1001
LAN CABLE
9642
9804-02 Black
11mm (0.43 in)
(generally compatible with M6 pan screws)
Magnetic tip for use with the standard
VOLTAGE CORD L9438-53
Dimension:
Approx. 390W (15.4")275H (10.8")110D (4.3") mm
Note: Company names and Product names appearing in this catalog are trademarks or registered trademarks of various companies.
DISTRIBUTED BY
HEADQUARTERS
E-mail: hioki@hioki.in
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 FAX +81-268-28-0568
HIOKI SINGAPORE PTE. LTD.
http://www.hioki.com / E-mail: os-com@hioki.co.jp TEL +65-6634-7677 FAX +65-6634-7477
E-mail: info-sg@hioki.com.sg
PW3360E10-5XB
Printed in Japan