vPad Safety Analyzer

Datrend Systems Inc.


File: 960703846_20170113_170413_P.txt
Date: 2017 Jan 13
Time: 17:04:13
vPad-ES: VPR15100055
Equipment Information...
Control Number: 960703846
Description:
BIOFEEDBACK SYSTEM
Manufacturer:
GYMNA UNIPHY
Model:
MYO 200
Serial Number: 69126
Location:
PHYSIOTHERAPY
Facility:
BEMS
Technician ID: 20045
Tech Time (hrs):
Overall Result: PASS
Signature:______________________________ Date:__________
Detailed Test Data...
AutoSequence Filename:
Title of AutoSequence:
AutoSequence Run Time:

vpad_as_31.asq
IEC60601 Class 1
4 minutes

Elec. Safety Standard:


Ground Resistance Limit:
Chassis Leakage Limits:
Ground Leakage Limits:
Lead Leakage Limit (BF):

IEC60601
Maximum 0.200 ohms
Maximum 100 A NC; 500 A SFC
Maximum 5000 A NC;10000 A SFC
Maximum 100 A NC; 500 A SFC; 5000 A MAP

Line Cord Resistance: 0.184 ohm - PASS


Line Voltage L1-L2: 232.4 V
Line Voltage L1-GND: 232.5 V
Line Voltage L2-GND: 1.6 V
EUT power switch "ON"...
Ground Leakage (POL-R;NEU-O;GND-O): 396 A - PASS
Chassis Leakage (POL-R;NEU-O;GND-N): 0.5 A - PASS
Lead-GND LA/AP1 (POL-R;NEU-O;GND-N): 0.9 A - PASS
Chassis Leakage (POL-R;NEU-N;GND-O): 202 A - PASS
Lead-GND LA/AP1 (POL-R;NEU-N;GND-O): 79.5 A - PASS
Ground Leakage (POL-R;NEU-N;GND-O): 202 A - PASS
Chassis Leakage (POL-R;NEU-N;GND-N): 0.5 A - PASS
Lead-GND LA/AP1 (POL-R;NEU-N;GND-N): 1.2 A - PASS
Iso Test LA/AP1 (POL-R;NEU-N;GND-N): 334 A - PASS
Ground Leakage (POL-N;NEU-O;GND-O): 395 A - PASS
Chassis Leakage (POL-N;NEU-O;GND-N): 0.5 A - PASS
Lead-GND LA/AP1 (POL-N;NEU-O;GND-N): 0.9 A - PASS
Chassis Leakage (POL-N;NEU-N;GND-O): 204 A - PASS
Lead-GND LA/AP1 (POL-N;NEU-N;GND-O): 80.1 A - PASS
Ground Leakage (POL-N;NEU-N;GND-O): 204 A - PASS
Chassis Leakage (POL-N;NEU-N;GND-N): 0.6 A - PASS
Lead-GND LA/AP1 (POL-N;NEU-N;GND-N): 1.2 A - PASS

Iso Test LA/AP1 (POL-N;NEU-N;GND-N): 332 A - PASS


Load Current:
0.100 A
Equipment Power: 23 W
END OF RECORD