It can also be offered with a PCI card that will allow it to be connected to an existing PC.
Features
Field generated in three axes, up to 500T per axis for DC and 100T at 5kHz
Applications
Calibration of three-axis magnetic field sensors
Electrical
Amplifier input 1V/100T differential
Input connector Hirose RM15TRD-10S
Coils output connector Neutrik Speakon 8 poles
Current monitor output 1V/100T differential
Current monitor output connector Hirose RM15TRD-10P
Power requirement 100V230V AC, <10A, 50/60Hz
Maximum power 1kW
Mechanical
Coil dimensions X Coil: nominal diameter 600mm
Y Coil: nominal diameter 554mm
Z Coil: nominal diameter 508mm
Coil assembly dimensions Height: 632mm
Width: 620mm
Depth: 574mm
Coil system volume 0.5m3
Coil homogeneous volume (<0.1% error) 4.5cm3
Coil homogeneous volume (<1% error) 165cm3
Coil gross weight 30kg
Electronics enclosure dimensions 6U 19in rack mount box
Height: 270mm
Width: 510mm
Depth: 600mm
Electronics enclosure gross weight 59kg
Environmental
Operating temperature range +15C to +30C
Storage temperature +10C to +40C
Operating humidity Up to 90% RH non-condensing
www.bartington.com
HCS1 Three-Axis Helmholtz Coil System
Current sense Three axis 10V differential connected to current sense output
on Helmholtz drive amplifier
Input filters Type Butterworth 2-pole 12dB per octave low pass
Frequencies (software selectable): 10kHz,1kHz, 100Hz & 10Hz
Sensor power output Magnetometer 0 to 20V programmable, limited to 100mA
(compatible with single or dual supply devices)
AUX & current 15V limited to 50mA per sensor channel. Power output
sense protected by self resetting semiconductor fuses
DAC output channels Three axes 10V balanced/differential connected to Helmholtz drive amplifier
ESD protection Comprehensive ESD surge protection on input and output channels for ESD and
lightning protection
Electrical
Maximum cable length 20m all inputs and outputs
Power supply voltage 100230 VAC 50/60 Hz
Power supply current 1100 mA (115V) / 640 mA (230V) type
Power input (rear panel) 3-way IEC 320 with integral filter (mains cable provided)
Working signal input voltage Up to 10V differential or single-ended (see above)
Working common mode input voltage Up to 8V: balanced / differential output sensors
Up to 3V: single ended sensors
Absolute maximum input voltage 15V on any input pin
Input connectors (front panel) Hirose RM15TRD-10P
Input mating connector Hirose RM15TPD-10S
Output connectors (front panel) Hirose RM15TRD-10S
Output mating connector Hirose RM15TPD-10P
Output connector (rear panel) 2 x 68-way Harting 6001 068 5232; interfaces with National Instruments 6289 DAQ card
using NI cables SHC68-68EPA
Mechanical
Dimensions 930 x 575 x 360mm (in wheeled flight case)
Weight 46kg (in wheeled flight case)
Environmental
Operating temperature range +10C to +50C
Operating humidity 0 to 50% RH non-condensing
Storage temperature -40C to +85C
PXI Data Acquisition Unit Specifications
PXI Cardframe NI PXI-1042*
Power supply voltage 100 to 240 VAC 50/60 Hz
Power supply current 8A (peak)
Power requirement 500W (max)
Operating temperature 0 to 55C
Storage temperature -20 to 70C
Operating humidity 10 to 90% RH non-condensing
Height 177mm
Width 271mm
Depth 397mm
Weight 8.4kg
Processor NI PXI-8106*
Processor Intel Core 2 Duo processor T7400 (2.16 GHz dual core)
Hard drive 60 GB minimum, internal 9.5 mm Serial ATA 1.0 interface
Chipset Mobile Intel 945GM Express chipset
RAM 2 x SO-DIMM sockets, DDR2 SDRAM, PC2 5400, dual channel
512 MB (1 x 512 DIMM) standard, 4 GB (2 x 2 GB DIMMs) maximum
Ethernet 10/100/1000BaseTX, RJ45 connector
Video Intel GMA 950 graphics media accelerator, DVI-I connector
Serial RS232
Parallel port IEEE 1284 Type C miniature connector
GPIB PCI-GPIB/TNT, micro D25 connector
Hi-speed USB 4
/ continued
www.bartington.com
HCS1 Three-Axis Helmholtz Coil System
* See the National Instruments website at www.ni.com for any changes to these specifications.