Professional Documents
Culture Documents
SPE-184814-MS Long Term Conductivity vs. Point Specific Conductivity
SPE-184814-MS Long Term Conductivity vs. Point Specific Conductivity
This paper was prepared for presentation at the SPE Hydraulic Fracturing Technology Conference and Exhibition held in The Woodlands, Texas, USA, 24-26 January
2017.
This paper was selected for presentation by an SPE program committee following review of information contained in an abstract submitted by the author(s). Contents
of the paper have not been reviewed by the Society of Petroleum Engineers and are subject to correction by the author(s). The material does not necessarily reflect
any position of the Society of Petroleum Engineers, its officers, or members. Electronic reproduction, distribution, or storage of any part of this paper without the written
consent of the Society of Petroleum Engineers is prohibited. Permission to reproduce in print is restricted to an abstract of not more than 300 words; illustrations may
not be copied. The abstract must contain conspicuous acknowledgment of SPE copyright.
Abstract
Long term conductivity of proppants, currently defined by API RP 19D, has been used since the early 1970's
as the standard for evaluating the conductivity and permeability of proppants for use in hydraulic fracturing.
Proppants are typically tested across their useful range in 2,000-psi (137.89-bar) increments (i.e. 2-4-6-8K-
psi) and held 50 hours at each stress level, ramp rates between each stress and the test temperatures are also
defined by API RP 19D. Frac sand is tested at 150F (65.5C) and ceramic proppants at 250F (121.11C).
Proppant suppliers use this, among other testing parameters, to develop data to market their products to the
industry. As with any standard, the test has its limitations. The test is expensive, time consuming and in
many cases does not match the application.
Point specific 50 hour (PS50) conductivity tests follow the procedures outlined in API RP 19D but use
a point specific stress and point specific temperature, modeling a specific application. Although not yet
sanctioned by the API, the test can generate meaningful data in less time, for a specific purpose and save
money. Point specific testing has maximized Research & Development budgets and increased the time to
market for new products. Operators have benefited by tailoring proppants for their specific well conditions.
In this work we plan to compare point specific test data to API RP 19D data over a range of stress and
time as the industry pushes the envelope on the applicable stress for proppants. Aven, Tang and Weaver
brought out that "it is normal to observe a continuous decline in conductivity, although less than 5% on a
daily basis, if the measurement is carried longer than two days. It is rare that the system actually achieves
equilibrium conductivity during that time, but the rate of change of conductivity does slow down1."
Could the point specific test be extended to achieve equilibrium and still be comparable to existing data
from the standardized test? Multiple tests were conducted to compare these two methods of conductivity
tests at normal duration, a point specific 50 hour (PS50) conductivity test and the API RP 19D conductivity
test which the test spends 50 hours at each 2,000 psi (137.89-bar) increment. Results presented in this paper
will help answer these questions and aid engineers make more informed decisions faster and with a lower
economic cost.
Introduction
With the advancement of technology, previously undiscovered and unreachable reservoirs are now within
reach. At the same time technology has allowed the materials being used and placed in wells to advance.
2 SPE-184814-MS
The standards to test these materials, specifically proppant, on the other had has not changed at the same rate
and has left the industry and companies who use them with a standard that is time consuming and costly.
Long term conductivity tests are not only a significant economic commitment to have performed but
can take several weeks to complete. Improvements have been made in equipment such as harder stainless
conductivity cells to test at temperatures and stresses beyond the API standard recommendation. Equipment
such as a custom built hydraulic press capable of reaching and controlling pressure beyond what was
considered standard has now become standard. However, for time consideration in this paper we will
examine the results of stresses up to 12,000-psi. Higher temperature conductivity cells require additional
engineering and customization. Ohio sandstone has to be purchased, cut, and shaped using hydraulic surface
grinders and milling machines and becomes another potential source of error to any conductivity test when
used beyond 12,000-psi.
Each stress in a multi-stack conductivity test is held for 50 hours. Typical proppants in use today are tested
up to and past 12,000-psi and these tests can last up to 300 hours or 13 days. Does the added time at lower
stress levels affect the data at the higher stress levels? The equipment and personnel that are committed
to a single test for the duration of the test are also significant. Stress levels above 8,000-psi (551.57-bar)
increase the potential damage from fines generated by the Ohio sandstone. Downhole formation core is
more representative but expensive to obtain and also see fines damage, spalling and embedment. Is there
a test that can be developed to lessen the economic, time, and man hours commitment that is currently
required for the standard long-term conductivity test? How does the developed PS50 test compare to the
standard API long-term conductivity test?
PS50 is designed to target specific wells or fields or to accelerate the research and development time-
line and extend R&D budgets. PS50 measure conductivity, width and permeability at 2 target stress values,
1000-psi (per ISO 13503-5) for 24 hours then the target stress for 50 continuous hours which is defined as
a long-term conductivity test. Reduced press time means overall reduced cost to obtain the desired data.
similar and demonstrated it was the total stress not how fast it was applied. R.D. Barree8 in SPE 84306-
MS determined that the packing arrangement of the proppant when placed in the conductivity cell was
the predominant factor in measuring the permeability. Variance as higher than 25% is common due to the
loading procedures whether using a pluviator or a technician. Data generated for this paper utilized a single
conductivity press, a single set of test cells and the same technician throughout to minimize any errors.
Table 1
Percent Difference 1 3 4 1 26 0
Figure 1
SPE-184814-MS 5
The long-term conductivity test as shown in Table 2 and Figure 2 tested 30/50 white sand. The percent
difference started out initially low as expected but at 12,000-psi (827.36-bar) significant deviation was
noticeable. The original PS50 value for 12,000-psi (827.36-bar) was 33 mD-ft however after a statistical
analysis of that point it was found to have a Z score that was classified as an outlier. Therefore, the test
was repeated and the results of the repeated test were accepted which was 83 mD-ft. The results in Figure
2 and Table 2 reflect the repeated test.
Table 2
Percent Difference 2 5 18 3 4 22
Figure 2
For the 40/70 mesh tests, the PS50 conductivity value compared to the standard API test remained no
greater than 12% difference throughout the test. See Figure 3 and Table 3.
6 SPE-184814-MS
Figure 3
Table 3
Percent Difference 0 7 5 5 12 12
Lightweight ceramic was also tested for this paper to determine if the type of proppant could be affected
by the PS50 test when comparing to the standard API test method. Figure 4 and Table 4 show that the
PS50 test was never more than 15 % different from the standard API test which falls within our internal
acceptable variation.
SPE-184814-MS 7
Figure 4
Table 4
Percent Difference 1 13 15 4 2 10
Table 5
The table shows that on average the PS50 test is about 5-6% lower than the standard API testing
procedure. Now this is a small sample size so caution needs to be taken when drawing a conclusion from
this limited data set. 3 API standard tests were run and an average was taken from them. One standard API
test was thrown out of the 30/50 due to core complications which led to inability to get stable readings. The
20/40 LWC API test is not an average as time constraints only allowed for one test but the results were in
line with the other sand tests. Also due to time constraints only one set of PS50 tests were run so the data
shown in the graphs is of one data point.
The Z score is a common method used to determine if a value within a data set is an outlier. The Z score is
the difference between the value and the mean divided by the standard deviation. For this paper we decided
to reject data points that had a Z score of less than -3.0 or greater than +3.0. All data points were given a
Z score and can be seen in Table 6.
Table 6
The mean for the complete data set is 0.94 and the standard deviation is 0.16. The only point that that
failed our criteria and was classified as an outlier was the 30/50 sand 12K reading. The Z score of this
data point is 3.86. Because it is classified as an outlier the test was repeated. The result of the repeated test
resulted in a conductivity of 83 mD-ft which had a Z score of 1.9. This Z score resulted in the acceptance
of the test and the new Mean for the updated data is 0.95 with a standard deviation of 0.09.
Conclusions
1. While ISO long term conductivity and PS50 conductivity results fall both above and below each other,
the overall results indicate that the PS50 test is about 5% lower than the standard API test method.
This is based on a limited number of samples so caution should be used when taking these results
into consideration.
2. Comparing long term conductivity to point specific conductivity the percent change from the
beginning of the test and at the end of the test remained similar from beginning to end and well within
the statistical mean.
3. For various proppants and mesh sizes normalization of results indicate that data falls within predicted
results for the various particle distributions in both the API and PS50 conductivity tests.
SPE-184814-MS 9
4. PS50 tests produce valuable and comparable data used in material selection and quality control at
a significant cost savings. The amount of cost savings can vary anywhere from 35% up to 57%
depending on the stress. This allows for R&D budgets to stretch much further than before.
5. PS50 tests are a valuable tool for operators, service companies and proppant manufactures that allow
them to quickly determine the expected performance characteristics of proppants
References
1. Nevil Kunnath Aven (Halliburton) | Tingji Tang (Halliburton) | Jim D Weaver (Halliburton)
SPE-164082-MS, Society of Petroleum Engineers SPE International Symposium on Oilfield
Chemistry, 8-10 April, the Woodlands, Texas, USA.
2. Pearson, C.Mark,(Liberty Resources LLC), Griffin, Larry (Liberty Resources LLC), Chikaloff,
Juliana(Liberty Resources LLC, SPE 168641-MS, MS, Society of Petroleum Engineers SPE
International Symposium on Oilfield Chemistry, Woodlands, Texas, USA Measuring Field
Supplied Proppant Conductivity : Issues Discovered in an Operator.
3. C. E. Cooke, Jr., SPE-AIME,( Esso Production Research Co.),SPE 4117-PA, SPEAIME 47th
Annual Fall Meeting, held in San Antonio. Tex., Oct. 811, 1972, "Conductivity of Fracture
Proppants in Multiple Layers"
4. B.W. McDaniel, Halliburton Services SPE Member, Copyright 1986, SPE 15067-MS, Society
of Petroleum Engineers, presented at the 56th California Regional Meeting of the Society of
Petroleum Engineers held in Oakland, CA, April 2-4, 1986. "Conductivity Testing of Proppants at
High Temperature and Stress"
5. International Standards Organization, ISO 13503-5 "Procedures for Measuring the Long Term
Conductivity of Proppants" ISO 13503-5:2006(E)
6. American Petroleum Institute, API RP 61 "Recommended Practices for Evaluating Short Term
Proppant Pack Conductivity", First Edition, October 1, 1989.
7. API RP 19D "Recommended Practice for Measuring the Long-term Conductivity of Proppants",
First Edition (ISO 13503-5:2006, Identical) (Includes July 2008 Errata)
8. R.D.Barree, Barree and Associates, S.A.Cox,SPE 84306, SPE Annual Technical Conference and
Exhibition, Denver, Co. 5-8 October, 2003, "Realistic Assessment of Proppant Pack Conductivity
for Material Selection"
9. G.Penny, Stimlab, SPE 16900, SPE 62nd Annual Technical Conference and Exhibition, Dallas
Texas Sept. 27-30, 1987, "An Evaluation of the Effects of Enviromental Conditions and
Fracturing Fluids Upon the Long Term Conductivity of Proppants"
10. R.R. McDaniels, BJ Hughes, J.R. Willingham, BJ Hughes, SPE 53rd Annual Technical
Conference and Exhibition, Houston, Texas, Oct.1-3, 1978. "The Effect of Various Proppants and
Proppant Mixtures on Fracture Permeability"