Professional Documents
Culture Documents
Feedback Control of Type 1 Diabetes: Sandip D. Kulkarni
Feedback Control of Type 1 Diabetes: Sandip D. Kulkarni
By Sandip Kulkarni 1
Type 1 Diabetes Mellitus
Hyperglycemia: Kidney failure, heart
disease, blindness, brain damage
180 mg/dl
Target
Blood
Glucose (BG)
range
Overcom-
70 mg/dl pensation
Hypoglycemia: Neurological
60 mg/dl damage, unconsciousness,
seizures, death Severe
Hypoglycemia
By Sandip Kulkarni 2
Glucose Sensing
Self Monitoring of Blood Continuous Glucose
Glucose (SMBG) Monitoring (CGM) Device
By Sandip Kulkarni 6
Linearized discrete-time model
u(k)
insulin infusion GLUCOSE
x(k+1)=Ax(k)+Bu(k)+Gc(k) y(k)=Cx(k) +v(k)
SENSOR
d(k)
meal
States
G glucose
X Insulin sensiKvity
I sc1
Insulin in Intradermal
I sc 2 Compartments
x=
I p Insulin in Plasma
Gsc Infused Glucose
Q
1 Stomach Compartments C = ( 1 0 0 0 0 0 0 0 )
Q2 By Sandip Kulkarni 7
What happens today...
Meals,
Exercise
INSULIN
PUMP
bolus
basal
Open-loop
Subject
therapy
GLUCOSE
SENSOR
By Sandip Kulkarni 8
Control Block Diagram
MPC
bolus
basal + Safety
Mod.
Subject
GLUCOSE
SENSOR
By Sandip Kulkarni 9
18 Hour Closed Loop Control Experiment in Hospital
4 PM: 30 min low- 7 PM: manual 4.4u pre-meal bolus Overnight maintenance of
220 intensity exercise meal underinsulinized by 40% basal rate 2.2
200 2
180 1.8
160 1.6
140 1.4
12:00 14:00 16:00 18:00 20:00 22:00 00:00 02:00 04:00 06:00 08:00
Finger SKcks CGM Insulin injecKon
By Sandip Kulkarni 10