18 SOP Monitoring of Autoclave
18 SOP Monitoring of Autoclave
18 SOP Monitoring of Autoclave
1.0 PURPOSE
2.0 SCOPE
3.0 RESPONSIBILITIES
Microbiologist
Lab Assistant
4.0 PROCEDURE
Theory:
o In a plastic vial, a white paper strip on which spores of Geobacillus
stearothermophilus (ATCC 7953) have inoculated and a glass ampoule in
which nutrient agar and also a pH-color indicator has covered.
Process:
o 03 biological indicators are placed at different places (places are shown in
monitoring checklist) in autoclave, so that to check the proper sterilization
process (01 biological indicator is placed in incubator as a Control). Close the
access door and continue to operate the apparatus.
Written By Reviewed by Approved by
Interpretation of Result:
o The control vial shows growth after 1-2 days.
o If the test vials dont change their color after 2 days, the sterilization was
successful.
o If the test vials change their color, stop the autoclave and repeat the test
procedure, placing several indicator vials throughout the autoclave.
o If the above color change observed again, examine the autoclave and check its
function.
Geobacillus.
Stearothermophilus Violet Yellow-Green
spores
Frequency:
o Autoclaves (Injectable & Microbiology Lab) are monitored after every 15 days
(twice a month).
Note:
o Do not crush the inner ampoule until the vial is at room temperature! Hot glass
ampoules may burst the plastic vial.
o The bio-indicator must not be used for dry heat (risk of glass ampoule and
plastic vial explosion).