Professional Documents
Culture Documents
Polytronirex Manual PDF
Polytronirex Manual PDF
Polytron
Transmitter IR Ex, IR Ex ...
Instructions for Use
00123655_04.eps
Contents
For Your Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
EC type test certificate in accordance with Directive 94/9/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Installing transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Transmitter IR Ex, IR Ex ES, IR Ex HC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Transmitter IR Ex IL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Installation of NPT pipe connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Installation with hard pipe connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Mounting to a wall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Using the maintenance opening for gas inlet in diffusion operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Installing gas path with shut-off valves for calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Mounting IR Ex display unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Mounting IR junction box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Connecting the transmitter to the central controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Connecting transmitter and remote IR Ex display unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Connecting transmitter and IR junction box for hand-held terminal IR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
After connecting electrical installation: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Removing dust filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Fitting splash guard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Connecting tubing for remote calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Initial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Maintenance Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Replacing dust filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Cleaning the optical system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Menu Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Entering password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Entering Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Entering Text . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Entering Decisions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Entering Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Handling Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Maintenance Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Calibrating zero point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Calibrating sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Configuring the calibration signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Checking 4 to 20 mA analog interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Display warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Display errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Display sensor temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Adjust pre-amplification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Setting H2O offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Calibrating analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Display software-specific information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Testing the display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Configuration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Changing gas configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Entering temperature coefficients A0 to A2 and their check sums . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Configuring fault signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Configuring zero clamping (positive values) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Changing password for the maintenance menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Changing password for configuration menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Changing the language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Switching measured value display on and off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Configuring warning signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Configuring a special gas name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Transmitter performance IR Ex and IR Ex IL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Transmitter performance IR Ex ES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Transmitter performance IR Ex FP and IR Ex FP IL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Transmitter performance IR Ex HC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Cross sensitivities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
IR Ex HC cross sensitivities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Measuring, using cross sensitivities (IR Ex HC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
IR Ex ES cross sensitivities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Measuring other hydrocarbons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Polytron IR Ex and IR Ex ES transmitters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Polytron IR Ex IL transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Polytron IR Ex HC transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Polytron IR Ex display unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Polytron IR junction box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Measurement Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Order List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
IECEx - Approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Maintenance
Repair of the transmitter may only be carried out by trained service personnel.
We recommend that a service contract is obtained with DrgerService and that all
repairs are also carried out by them.
Only authentic Drger spare parts may be used for maintenance.
Observe chapter "Maintenance Intervals".
Intended Use
For stationary, continuous monitoring of selectable gas mixtures containing
hydrocarbons by infra-red absorption at wavelength 3.4 m (IR Ex ES: 3.3 m).
With a 4 mA to 20 mA analog output with 3 or 4 wire connection for easy
replacement of catalytic transmitters, for instance.
IR Ex IL and IR Ex FP IL:
Gas-tight connection to the optical system of the transmitter for in-situ
measurement under atmospheric pressure.
IR Ex HC and IR Ex FP:
Not suitable for detecting ethene (ethylene) and methane.
IR Ex FP and IR Ex FP IL transmitters:
These models differ from the IR Ex and IR Ex IL versions by their very quick response
time to propane.
However, these models cannot be congured for methane or ethene, nor is the
% by vol. measuring range congurable.
The same technical data and measuring properties apply to these versions as stated
for IR Ex or IR Ex IL in the propane substance group, unless otherwise stated.
Polytron IR Ex IL volume ow
The volume ow for the Polytron IR Ex IL should be set taking into account the
response times (see Technical Data on page 59).
Undervoltage
The minimum operating voltage for all the transmitters described here is 13 V.
At operating voltages between 10 and 13 V, the sensor behaviour is undened
and may cause false alarms at the central controller. Below 10 V the transmitter
will reliably switch off.
Remark:
The ex-approval is valid for different types and shapes of the IR * (e.g. different
interference lters or cuvettes). The device description is indicating the different
types and shapes by additional abbreviations, separated by a comma,
e.g. "Polytron IR NDF, Ex IL".
The transmitter Polytron IR with its versions NDF and NDH is intended to be used for
the detection of ammable gases and vapours under atmospheric conditions. The
non-intrinsically safe power supply and signal transmission is realized by either one
3-core- or two 2-core-cables, which have to be connected to the terminals K1, K2, K3
and K4 placed in an "e"-terminal box (explosion protection increased safety "e").
The intrinsically safe output circuit of the transmitters Polytron IR NDF or Polytron IR
NDH can be connected by a cable (length max. 50 meters) to either the display box
(Polytron IR Display box) or the hand held terminal (Polytron IR HHT).
The rating of the cable must not exceed the max. inductivity and capacity of
Lo = 100 H resp. Co = 100 nF.
The transmitter Polytron IR, the versions Polytron IR NDF and Polytron IR NDH, the
Polytron IR Display box and the hand held terminal Polytron IR HHT are associated
to the device category II 2G and approved to be used in potentially explosive
atmospheres zone 1and zone 2.
Safety-relevant parameters:
The following max. values for the non-intrinsically safe power supply and signal
transmission are valid for Polytron IR, Polytron IR NDF and Polytron IR NDH
(terminals K1, K2, K3 and K4):
Max. DC-voltage Um 32 V DC
Max. AC-voltage Um 250 V AC
Environmental conditions:
operation: 40 to +65 oC 1)
20 to +55 oC 1) (only applies to transmitters, concerning IECEx approval)
700 hPa to 1300 hPa
(in potentially explosive atmospheres:
800 to 1100 hPa)
0 to 100 % rel. humidity, non condensing 2)
during storage: 40 to +65 oC
700 hPa to 1300 hPa
0 to 100 % rel. humidity, non condensing 2)
IP-rating acc. EN 60 529: IP 65
____________
1)
The given temperatures are min. and max. temperatures concerning the explosion protection approval. Concerning the
measuring performance the temperature range might be smaller, e.g. for the detection of ethene: 25 to +50 oC.
2)
IR Ex HC: 0 to 95 % rel. humidity
Design
The modular design allows several assembly options:
00223655_04.eps
for transmitters which are easily accessible to the operator.
The transmitter is operated directly via a built-in display and keypad.
04723655_04.eps
for transmitters which are easily accessible to the operator.
The transmitter is operated via the hand-held terminal using the plug connection.
IR Ex, IR Ex ES IR Ex IL IR Ex HC
Installing transmitter
Select installation point according to local regulations.
Keep well clear of extraneous factors such as flooding water, oil etc. as well
as any chance of mechanical damage.
Removing cover:
Remove four screws, carefully lift off cover.
Polytron IR Ex
ESC
01023655_04.eps
IR Ex, IR Ex ES IR Ex IL IR Ex HC
3 3
2 3 2
1 1 1
01123655_04.eps
1 Remove the five plug connections (IR Ex HC = three plug connections) on the
housing cover.
The plug connections are keyed or interchangeable, see page 21.
2 Mount transmitter with four screws through the housing or
3 Fix mounting brackets to transmitter and mount transmitter with four screws
through the brackets.
Mount in vertical position (sensor at the bottom).
Leave at least 30 cm of open space below the transmitter to allow access for
calibration purposes.
Do not fit cover for electrical connections yet.
Transmitter IR Ex IL
Make sure that the pressure range of the sensor is always observed (see section
Technical Data, page 56).
Connect gas supply line using two 1/2" NPT female threads, seal with Teflon tape
(connection to commercial systems possible).
Flow monitoring is necessary in flow mode.
00423655_04.eps
Installation for diffusion operation
Use of the maintenance opening, page 15.
00523655_04.eps
Mounting to a wall
Gas supply must be flexible tubing, flexible gas-tight conduit etc.
00623655_04.eps
ESC
00723655_04.eps
Using the maintenance opening for gas inlet in diffusion operation
A sealing plug (maintenance opening) is provided at the rear of the sensor for
maintenance and inspection of the optical system.
Thread size: G 3/4
This maintenance opening can be used for diffusion operation:
Screw gas supply connecting piece with G 3/4 male thread to maintenance
opening.
Make sure that the connecting piece between sensor and the duct to be
monitored is as short as possible.
Seal the two NPT openings with an appropriate sealing plug.
Make sure that the transmitter is protected against mechanical damage.
Observe dead weight of the transmitter.
The response time of the instrument may be considerably increased when the
maintenance opening is used as gas inlet.
00823655_04.eps
Installing gas path with shut-off valves for calibration
especially in case of pipe installation, the gas path should have shut-off valves.
Then the maintenance opening is used as gas inlet during calibration.
00923655_04.eps
01223655_04.eps
2
01323655_04.eps
2
Electrical connections
Routing and connecting the electrical installation must only be done by trained
personnel, observing appropriate regulations
use 2, 3 or multi-core, screened cable (braided screen, cover 80 %,).
When installing the transmitter, proper grounding has to be ensured.
Covers of transmitter (display unit, IR junction box) should not be put back on yet.
On the transmitter:
1 Remove two screws from the terminal cover.
2 Remove terminal cover. 1
Connect according to the applicable procedure below.
01423655_04.eps
When using 4-wire-connection:
Remove blind plug on transmitter.
3 Screw second cable gland (accessory, see Order List) into housing and secure
with nut.
01523655_04.eps
The cable gland is suitable for cable diameters of 6 to 12 mm.
For cable diameters of 13 to 15 mm the cable gland must be replaced by Drger
cable gland (part-no. 83 11 370) without the inner of the two rings.
3-wire connection:
The cable resistance per core for the connection to transmitter terminals K1 and K3 must not exceed:
Rmax = (U0[V] 13) x 2 per core (e.g. 22 per core at 24.0 V)
U0 is the power supply at the central controller.
The cable resistance for the connection to the transmitter terminal K4 must not exceed 450 (individual internal resistance
of the central controller).
Connect screen to terminal E of the transmitter.
Ground the transmitter via external terminal, but do not ground screen at the central controller or do not ground transmitter
via external terminal, but ground screen in the central controller.
IR Ex, IR Ex ES
K1
K3
K4 E
Supply unit
24 V 20 % , 0.3 A
IR Ex IL + 24 V
K1 0V
K3
K4 0V
K1 E
K3
K4 4...20 mA
E
Central controller
IR Ex HC
K1
K3
K4 E
01823655_04_en.eps
4-wire connection
The cable resistance per core for the connection to transmitter terminals K1 and K2 must not exceed:
Rmax = (U0[V] 13) x 2 per core (e.g. 22 per core at 24.0 V)
U0 is the power supply at the central controller.
The cable resistance per core for the connection to the transmitter terminal K3 and K4 must not exceed 350 individual internal
resistance of central controller).
Connect screen to terminal E of the transmitter.
Ground the transmitter via external terminal, but do not ground screen in the central controller or do not ground transmitter
via external terminal, but ground screen in the central controller.
IR Ex, IR Ex ES
K1
K2
K3
K4
E
Supply unit
IR Ex IL 24 V 20 % , 0.3 A
E
K1 + 24 V
K2 0V
K1
K2
K3 K3 0V
K4 K4 4...20 mA
E E
Central controller
IR Ex HC
K1
K2
K3
K4
E
01923655_04_en.eps
1 All three cables must be looped through the ferrite core once, following the same direction.
Connect screen to terminal E of the transmitter and terminal E of the display unit.
K13 E
K11
K12
E K12
K11 K13
K3 E
K2
K1
K1 K3
E 1
02023655_04_de.eps
K2
K13 E
K11
K12
E K11 K13
K1 K2 K3
K12
K1 K3
K2
02123655_04.eps
After connecting electrical installation:
1 Screw terminal cover back on to transmitter, two screws.
Reconnect five plug connections (IR Ex HC = three plug connections) on the
housing cover,
where:
2 Sensor cable (7-pin plug)
1
Not IR Ex HC:
3 Heater cable (2-pin socket). Polarity does not matter.
4 Heater cable (2-pin socket). Polarity does not matter.
These heater cables are interchangeable.
02223655_04.eps
5 Supply cable 2 3 4
6 Cable to infra-red source (3-pin socket). Polarity does not matter . 5 6
Note: if cable 6 (infra-red source) is connected when the power supply is switched
on, the infra-red source may switch off. In this case, disconnect plug 5 (supply cable)
from the printed board for about 10 seconds.
Then reconnect plug 5.
Screw all housing covers back on transmitter, the remote display unit and on to
the IR junction box:
Each housing cover is secured with four screws do not pinch any cables!
Polytron IR Ex
ESC
02323655_04.eps
Accessories
IR Ex, IR Ex ES only
The transmitter comes with a dust lter for protection against dust, splash water and
salt mist.
To ensure optimum sensor functioning under various operating conditions the
following congurations and accessories are recommended:
3
2
02423655_04.eps
Fitting splash guard
Accessory, order separately
02523655_04.eps
Slide splash guard onto sensor until it fits into support plate.
Secure splash guard with 2 screws.
02623655_04.eps
Initial Operation
Switch on power supply.
Note:
As the transmitter briey (approx. 100 ms) delivers current of up to 12 mA to
the central controller when the power supply is switched on, the alarm limits
set at the central controller may be exceeded.
Analog Output:
Fault signal for approx. 30 seconds; after that normal operation (see
Measurement)
Check zero point calibration; calibrate zero point, if necessary see "Operating
Functions for Maintenance and Configuration", page 37.
Check analog output see "Operating Functions for Maintenance and
Configuration", page 41.
Check that displayed value on transmitter (display unit, hand-held terminal) and
central controller match.
If necessary:
Calibrate analog output see "Operating Functions for Maintenance and
Configuration", page 45.
The transmitters are factory calibrated and congured (see label on transmitter).
Changing conguration see "Operating Functions for Maintenance and
Conguration", page 48.
Measurement
The current output of the transmitter during normal operation is between 4 and
20 mA and is proportional to the detected concentration.
Polytron IR Ex transmitter uses different current values to indicate various modes
of operation.:
Meaning: Current at conguration Current at conguration
Fault signal 3 mA Fault signal 1 mA
Fault 3 mA 1 0.2 mA
Warning (maintenance required) 10 s measuring signal and 10 s measuring signal and 3 s 2 0.2 mA
3 s calibration signal or 2 0.2 mA constant
Calibration signal 2 to 5 mA oscillating at 1 Hz 3 0.2 mA
Zero point 4 mA 4 mA
Full scale of measuring range 20 mA 20 mA
Measuring range exceeded >20 mA >20 mA
If values fall below the measuring range, a current with a minimum of 3.5 mA flows
through the output.
Maintenance Intervals
EN 50073 and the national regulations in the country of use must be observed.
Daily
Visual check to determine readiness for operation.
At regular intervals,
as specied by persons responsible for the maintenance of the gas warning system,
recommendation: every six months:
Check zero point and sensitivity calibration, page 37 to page 38.
When using the transmitter for methane measurement in the 0 to 20 % LEL range, if
the average operating temperatures vary greatly (e.g. by more than 20 K seasonally),
it should be recalibrated at the different average operating temperatures in order to
comply with the error limits according to EN 50057.
It is only possible to extend the calibration interval if no blocking of the gas supply to
the optical system due to e.g. dust or oil can happen in this particular application
during the period between two calibrations.
Annually
Inspection by trained personnel.
The inspection intervals in each individual case are subject to technical safety
considerations, engineering processes and the technical requirements for the
equipment.
We recommend a contract with DrgerService which also covers repairs.
Maintenance
Replacing dust lter
IR Ex and IR Ex ES transmitters
When tted, disconnect splash guard:
02823655_04.eps
1 Loosen screw on protective screen slightly.
2 Twist protective screen and pull off to the right
3 Hold dust filter in the hand and pull off to the right, gently twisting back and forth.
Push new dust filter over sensor gently twisting back and forth
the dust filter must not get torn!
Put protective screen back in place and secure with screw.
Fit splash guard back. 1
3
2
02923655_04.eps
In order to suppress a false alarm while cleaning the optical system: Set
calibration signal: menu item CALIBRATION SIGNAL, ON, page 40.
To remove splash guard and dust filter, proceed like previously described
"Replace dust filter", page 28.
03023655_04.eps
Blow off any loose particles with clean, dry air.
Rinse optical system with distilled water if necessary clean with alcohol or
mild detergent when heavily soiled clean with a lint-free soft cloth.
03223655_04.eps
IR Ex IL transmitter
It is absolutely necessary to protect the optical system from particles and aerosols by
preparing the sample gas appropriately.
When there is build-up of deposit on the optical system, the transmitter gives a
warning.
In order to suppress a false alarm while cleaning the optical system: Set
calibration signal: menu item CALIBRATION SIGNAL, ON, page 40.
Open maintenance plug at rear of optical system if necessary, remove
transmitter from installation.
Blow off any loose particles in the optical system with clean, dry air (do not exceed
the maximum pressure of the optical system).
Rinse optical system with distilled water if necessary, clean with alcohol or mild
detergent.
Vent out excess alcohol or cleaning liquid or let them dry.
Check the result of the cleaning process with special mirror, if necessary repeat
cleaning procedure.
Re-install maintenance plug, if necessary replace gasket. Re-install transmitter.
03323655_04.eps
IR Ex, IR Ex ES and IR Ex IL transmitters:
Cancel calibration signal: menu item CALIBRATION SIGNAL, OFF, page 40.
Return to measurement mode, page 34.
Menu Navigation
There are six keys on the transmitter (display unit, hand-held terminal):
four cursor keys [ q ], [ l ], [ k ], [ j ],
one Escape key [ ESC ]
one Enter key [ u ].
IR Ex, IR Ex ES IR Ex IL IR Ex HC
03423655_04.eps
to change between:
menu items / functions
numbers / letters at the position of the cursor
default values
use [ q ], [ l ] keys
to acknowledge:
inputs
messages
use [ u ] key
Menu structure
There are two main submenus in the menu structure:
MAINTENANCE and CONFIGURATION.
NOTE
Since maintenance might be performed by staff other than those who congure the
system, the instrument has a different password to access each group of functions.
However, the password for the submenu conguration also gives the user access to
the submenu maintenance.
Maintenance Menu
This level gives access to a number of functions for regular maintenance of the
transmitter. It is suggested that access be given to service department personnel.
Maintenance Menu, page 37
Conguration Menu
This level allows the user to change sensor parameters and to congure the
transmitter. It is suggested that access be given to authorized personnel in the
measurement and control or health and safety departments. Both the maintenance
and the conguration menus can be accessed with the conguration password.
Conguration Menu, page 48
Menu Structure:
MENU:
MEASUREMENT:
MENU: MENU:
MAINTENANCE: CONFIGURATION:
MAINTENANCE: CONFIGURATION:
page 37 page 48
CALIBR. ZERO GAS CONFIGURATION
1)
MAINTENANCE: CONFIGURATION:
page 38 page 50
CALIBR. SENSITIVITY TEMP-COEFFICIENT
MAINTENANCE: CONFIGURATION:
CALIBRATION-SIGNAL page 40 FAULT SIGNAL page 51
MAINTENANCE: CONFIGURATION:
TEST 4-20 mA page 41 ZERO CLAMPING page 51
MAINTENANCE: CONFIGURATION:
SHOW WARNINGS page 41 PASSWORD MAINTENANCE page 52
MAINTENANCE: CONFIGURATION:
SHOW ERRORS page 42 PASSWORD CONFIG. page 52
MAINTENANCE: CONFIGURATION:
SHOW SENSOR-TEMP. page 44 LANGUAGE page 53
MAINTENANCE: CONFIGURATION:
PREAMPLFIER ADJ. page 44 DISPLAY page 53
MAINTENANCE: CONFIGURATION:
H2O-OFFSET page 45 WARNINGS page 54
MAINTENANCE: CONFIGURATION:
CALIBR. 4-20 mA page 45 SPECIAL GAS NAME page 55
MAINTENANCE:
SOFTWARE-INFO page 46
MAINTENANCE:
DISPLAY TEST page 47
1)
On the IR Ex HC transmitter, this function is carried out solely by DrgerService in the case of sensor replacement.
Entering password
Display during normal measurement, e.g.:
0.0 %PROPANE
LEL
NOTE
The instrument will continue to monitor the concentration and give a 4 to 20 mA
output while navigating through and executing most of the submenu features.
Depending on the hierarchy level of the password, access for the following menus
can be selected, using [ q ] or [ l ] key. MENU:
Selection: MEASUREMENT, MAINTENANCE or CONFIGURATION MAINTENANCE
Press [ q ] or [ l ] key to change the menu.
Press [ u ] key, to select the menu, e.g.: MAINTENANCE:
CALIBR. ZERO
Use [ q ] or [ l ] keys, to navigate within the submenus, e.g.: MAINTENANCE:
SHOW SENSOR-TEMP
Press [ u ] key, to access a function , e.g.: SENSOR-TEMPERATURE:
20.4 DEGREES
The different functions in the maintenance and conguration menus are described in
detail below.
Press [ ESC ] key to return to the next higher menu level, e.g.: MAINTENANCE:
SHOW SENSOR-TEMP
When pressing the [ ESC ] key several times, the instrument will go to the top level MENU:
(measurement menu). MEASUREMENT:
To return to the measurement mode, press the [ u ] key once, and the instrument
will display the currently detected value, e.g.:
0.0 %PROPANE
LEL
Entering Numbers
For some of the functions it is necessary to enter numbers. Whenever such an entry
is required, the end of the first line in the display will show a question mark. At the
same time the second line displays the last entry with a blinking cursor under one of
the numbers, e.g.: CALIBR. GAS CONC. ?
40.0 % LEL
Use the [ k ] or [ j ] keys, to change the position of the cursor. It will skip over any
decimal point.
Use the [ q ] or [ l ] keys, to change the number at the position of the cursor.
Entries which do not make sense, such as .1 , -2-.22 , -.3, 44.-4 will not
accepted.
Keeping a key pressed is regarded as repeated activation of that key.
Press [ u ] key, the number set is accepted and verified whether it is within the
permissable range.
If the number entered is outside the permissible range a message will be shown.
Press [ u ] key, to acknowledge the message the number set previously will be
shown.
Entering Text
For some of the functions it is necessary to enter text. Whenever such an entry is
required, the end of the rst line in the display will show a question mark. At the
same time the second line displays the last entry with a blinking cursor under the last
letter of the text, e.g.: PASSWORD ?
Exception: when entering a password while in measurement no text will be DAVID
displayed. The second line is blank.
Entering Decisions
For some of the functions it is necessary to enter a decision. Whenever such an
entry is required, the end of the first line in the display will show a question mark.
At the same time the second line displays the options. They are separated by / .
The cursor is on the "safe" position, where no changes would be accepted, e.g.: 100 % CORRECT ?
Y/N
Press [ k ] key, to change the position of the cursor.
Press [ u ] key, to accept the decision.
Entering Options
For some of the functions it might be necessary to choose an option. Whenever
such a situation occurs, the rst line in the display will show a question mark. At the
same time, the second line shows the last selection. The cursor is not shown, e.g.: DISPLAY ?
ON
Use the [ q ] or [ l ] key, to select a different option, e.g.: DISPLAY ?
Keeping a key pressed, this is regarded as repeated activation of that key. OFF
Press [ u ] key, to accept the selected option.
Handling Messages
Within some functions, messages are issued. They are identified by an exclamation
point at the end of the text, e.g.: INPUT VALUE
TOO LARGE !
Press [ u ] key, to acknowledge the message.
Maintenance Menu
The maintenance menu contains all functions which are necessary for regular
maintenance of the transmitter.
Zero gas:
Nitrogen or alternatively synthetic air (compressed air) with a hydrocarbon content
under 50 ppm may be used to calibrate the zero point.
When using ambient air to calibrate the zero point, it is important to ensure that
ESC
there are no hydrocarbons in the vicinity of the transmitter.
Note:
As the zero point can be set even with very low calibration gas concentrations,
zero gas should never be confused with calibration gas.
IR Ex, IR Ex ES
2
1 In still air and with a dust filter fitted: apply zero gas to the gas connector into
03523655_04.eps
the sensor at a flow of approx. 3 L/min (not evaluated in the 1st supplement
to EC type test certificate DMT 02 ATEX E 0178X). 1
or
2 Attach calibration adapter 68 08 667 and apply zero gas to the sensor at a
flow of approx. 0.5 L/min.
IR Ex IL
Apply zero gas into the sensor at a flow of approx. 0.5 L/min.
03623655_04.eps
IR Ex HC
With adapter 68 09 240 and calibration adapter 68 06 978
Apply zero gas to the calibration adapter at a flow of approx. 0.5 L/min.
68 09 240
03723655_04.eps
All transmitters
Select function. MAINTENANCE:
CALIBR. ZERO
Press [ u ] key to access the function, display: APPLY NITROGEN
Apply zero gas to sensor. TO SENSOR !
Press [ u ] key to acknowledge the message.
Wait until the displayed value stabilizes (up to 3 minutes), confirm decision VALUE CONSTANT ?
with "yes" (Y) 2.05 Y/N
Press [ u ] key
Press [ u ] key to acknowledge the message. NEW CALIBRATION-
FACTOR STORED !
Tolerances:
IR Ex, IR Ex ES, IR Ex IL
% LEL % by vol. ppm
Methane: 0.2 0.01 ---
Ethene: 0.6 0.01 ---
Propane: 0.1 0.01 10
IR Ex, IR Ex IL, full scale of range 5 to 100 % by vol. methane: 0.1 % by vol
IR Ex HC: 1 % LEL
Calibrating sensitivity
When this function is activated, the 4 to 20 mA output changes from the
concentration to the calibration signal.
Calibration gas:
Use standard calibration gas (such as 40 % LEL methane in air or nitrogen with a full
scale of range of 100 % LEL methane) to calibrate sensitivity. The concentration of
other hydrocarbons in the calibration gas should be less than 50 ppm.
The calibration gas concentration can be between 10 % and 100 % of the full scale
of measurement, but should if possible be between 40 % and 70 % of the full scale
of measurement.
Note:
The calibration gas concentration should not be below 14 % LEL methane for
a full scale of measurement of 20 % LEL methane (IR Ex, IR Ex IL).
Caution:
Never inhale calibration gas. Health hazard.
Refer to appropriate Material Safety Data Sheets for handling calibration gas.
IR Ex, IR Ex ES
1 In still air and with a dust filter fitted: apply calibration gas to the gas connector
into the sensor at a flow of approx. 3 L/min (not evaluated in the 1st supplement 2
to EC type test certificate DMT 02 ATEX E 0178X).
03523655_04.eps
or
2 Attach calibration adapter 68 08 667 and apply calibration gas to the sensor 1
at a flow of approx. 0.5 L/min.
IR Ex IL
Apply calibration gas into the sensor at a flow of approx. 0.5 L/min.
03623655_04.eps
IR Ex HC
With adapter 68 09 240 and calibration adapter 68 06 978, vent calibration
gas safely to the outside via a tube.
Apply calibration gas through the calibration adapter at a flow of approx.
0.5 L/min.
68 09 240
03723655_04.eps
All transmitters
Select function. MAINTENANCE:
CALIBR. SENSITIVITY
If the calibration signal is activated the Z symbol is displayed in the upper right
corner of the display.
Display warnings
This function is used to display any warning in plain text.
If there is a warning message available, the i symbol will be displayed in the
upper right hand corner of the display.
Warning messages offer an opportunity for the operator to perform preventive
maintenance.
Press [ u ] key.
The function terminates.
Display errors
This function is used to identify an error by a code.
If there is a error message available, the F symbol will be displayed in the upper
right hand corner of the display.
Press [ u ] key to display the next message. The function automatically terminates
if there are no more message. ERROR-CODE
1!
Possible error messages:
ERROR-CODE 1
to
ERROR-CODE 17
see Table.
Adjust pre-amplication
This function is used to optimize signal pre-amplication.
This must be followed, for instance, after cleaning the optical system.
IR Ex HC transmitter: DrgerService only.
When this function is activated, the 4 to 20 mA output changes to the maintenance/
calibration signal.
Apply synthetic air or nitrogen, free from hydrocarbon gases or vapours, to the
sensor (see page 37).
Select function. MAINTENANCE:
PREAMPLIFIER ADJ.
Press [ u ] key to access the function
display shows the current pre-amplification factors, e.g.: OLD PREAMPL FACTORS:
REF: 1 SIG: 4
Press [ u ] key, the best pre-amplification is determined (takes about
20 seconds). Display: SEARCHING FOR
PREAMP FACTORS !
Then new pre-amplification factors are displayed, e.g.: NEW PREAMPL FACTORS:
REF: 5 SIG: 5
Press [ u ] key to access the Delay function.
The current delay values are displayed, e.g.: OLD DELAY-VALUES:
REF: 42 SIG: 43
Press [ u ] key to determine and display the new delay values, e.g.: NEW DELAY-VALUES:
REF: 42 SIG: 45
Press [ u ] key to save the new delay values.
Note:
The zero point and sensitivity should be calibrated after adjusting the
pre-amplication.
When using the transmitter in the ethene gas category, the IR sensor may have a
cross-sensitivity to water vapour. This can be corrected by using a wet zero gas and
entering a numerical value (99 to +100). The measured value will then automatically
be corrected by this offset.
Apply wet zero gas with a humidity corresponding to the typical humidity of the
gas to be measured to the sensor and wait for the measured value to stabilize
(up to 3 minutes). Read the average value of the display (H2O offset) in % LEL,
round to an integer and enter the value as the H2O offset.
X2
03823655_04.eps
mV
Press [ u ] key to obtain informations how to start and end the display test: DISPLAY TEST
ON: ENTER OFF: ESC
Press [ u ] key to observe the alternatively activated pixels of the display.
Faulty pixels can by identified.
Display test keeps being activated until [ESC] key is pressed.
Conguration Menu
The conguration menu contains all functions which are necessary to congure
the transmitter to application specic requirements. (Menu Structure, page 32).
When changing the engineering unit between % LEL and a volumetric unit
(% by vol. or ppm) without recalibrating the sensitivity, a conversion factor is used,
see table below:
Gas type Conversion factor 1) Adjustment range
Methane 4.4 % by vol. corresponds to 100 % LEL 4.4 5.0 % by vol.
Ethene 2.3 % by vol. corresponds to 100 % LEL 2.3 2.7 % by vol.
Propane 1.7 % by vol. corresponds to 100 % LEL 1.7 2.1 % by vol.
1) EN 61779-1, addendum A
When this conversion factor cannot be used (e.g. for measuring fuel in the propane
gas category), the sensitivity must be recalibrated after changing the engineering
unit.
Enter conversion factor for the gas to be measured (e.g. 2.3 % by vol. for ethene). 100 % LEL = ?:
2.30
Especially if high gas or vapour concentrations should be monitored, it is
recommended to recalibrate sensitivity after the range has been changed.
When changing the gas type all temperature coefcients and the check sum must
be changed accordingly, otherwise the linearization of the detector signal will not
work properly. Then, zero point and sensitivity must be calibrated.
If the temperature coefcients and the check sums do not correspond to the gas
type, the transmitter will go into error.
When changing the gas type, the following steps must be taken:
1 Switch calibration signal on, page 40
2 Change gas configuration, page 48
3 Change temperature coefficient, page 50
4 Calibrate zero point, page 37
5 Calibrate sensitivity, page 38
6 Switch calibration signal off, page 40.
03923655_04_en.eps
Temperature coefficient A1 Temperature coefficient A2
Fault 1 mA:
A transmitter fault is indicated by a current of 1 0.2 mA.
The calibration signal is a constant current of 3 0.2 mA.
After having congured the fault signal the warning signal is always switched off and
has to be switched on or newly congured if necessary (see page 54).
OFF
No warning signal on the analog output (default setting).
Depending on the conguration of the fault signal the following warning signals
can be selected:
2 mA steady
When a warning condition has been recognized a current of 2 0.2 mA is output.
DEFAULT GAS NAME reactivates the default text (methane, propane or GAS NAME ?
ethene). DEFAULT GAS NAME
A name of up to 11 characters can be input under SPECIAL GAS NAME . GAS NAME ?
SPECIAL GAS NAME
On completion of input, press [ u ] key. SPECIAL GAS NAME ?
BUTADIENE
The message SPECIAL GAS NAME ENABLED! appears. SPECIAL GAS NAME ?
ENABLED !
Press [ u ] key to end the function.
Technical Data
Overview
See page 5 for adjustable full scale of measuring range.
Transmitter type Linearity and temperature compensation Digital display resolution
IR Ex und IR Ex IL congurable: 0.1 % LEL
1 Methane 0.01 % vol. or 10 ppm
IR Ex ES 2 Ethene
3 Propane (also for other gases and vapours,
such as butane and ethane)
IR Ex HC 1 Propane 0.1 % LEL
IR Ex FP, IR Ex FP IL 1 Propane 0.1 % LEL or 10 ppm
Electrical data
Operating voltage: 13 V to 30 V
Power input: 4 W (IR Ex HC: <2 W)
Cable entry: Pg 13.5 cable diameter 6 to 12 mm
Wire cross-section: 0.5 to 1.5 mm2
Weightt
Transmitter: 1.9 kg (IR Ex IL: 3.5 kg)
Display unit: 1.0 kg
Junction box: 0.3 kg
Hand-held terminal: 0.35 kg
1)
factory setting
Environmental parameters
during operation IR Ex, IR Ex ES, IR Ex IL: 25 to 65 oC 1)
IR Ex HC: 25 to 50 oC
700 to 1300 hPa 2)
IR Ex, IR Ex ES, IR Ex IL: 0 to 100 % r.h. (non condensing)
IR Ex HC: 0 to 95 % r.h.
during storage 40 to 65 oC
700 to 1300 hPa
IR Ex, IR Ex ES, IR Ex IL: 0 to 100 % r.h.
IR Ex HC: 0 to 95 % r.h.
Classication: IP 65
Approval:
Transmitter: II 2G EEx me [ib] d IIB + H2 T4
40 oC Ta +65 oC
DMT 02 ATEX E 178 X
CE marking with notied body No.
Year of construction by serial number3)
Ex me [ib] d IIB + H2 T4
20 oC Ta +55 oC
IECEx BVS 05.0007
Ex ib IIC T4
40 oC Ta +65 oC
IECEx BVS 05.0007
Ex ib IIC T4
40 oC Ta +65 oC
IECEx BVS 05.0007
The following measurement data still apply to the measuring ranges and gases listed:
IR Ex: Effect of ow up to 6 m/s no effect
IR Ex: Response time at 25 oC:
Volume ow Time t50 Time t90 Volume ow Time t50 Time t90
0.1 L/min 35 s 97 s 1 L/min 10 s 17 s
0.2 L/min 24 s 68 s 3 L/min 9s 12 s
0.3 L/min 20 s 51 s 5 L/min 8s 11 s
0.4 L/min 16 s 35 s 10 L/min 7s 10 s
0.5 L/min 13 s 26 s
Transmitter performance IR Ex ES
(not evaluated in the 1st supplement to EC type test certicate DMT 02 ATEX E 0178X).
Type of gas Methane Propane Ethene
Full scale of range 100 % LEL 10 % by vol. 100 % LEL 10 % by vol. 100 % LEL 10 % by vol.
Concentration of calibration gas 50 % LEL 10 % by vol. 50 % LEL 10 % by vol. 50 % LEL 10 % by vol.
Values in % LEL % by vol. % LEL % by vol. % LEL % by vol.
Linearity error, max.: 3.0 2.0 3.0
Repeatability (DIN 1319)
zero point: 1.0 0.5 1.0
calibration gas: 1.0 0.2 1.0 0.7 2.0 0.3
Long-term drift, 6 months
zero point errors, max.: 0.5 1.0 1.0
calibration gas errors, max.: 5.0 7.0 8.0 0.5
Effect of temperature (25 to +55 oC)
Errors with respect to 20 oC, max.:
zero point: 1) 0.1 0.1 0.1
calibration gas: *) 0.5 0.1 0.3 0.03 0.1 0.01
Effect of pressure (950 to 1100 hPa)
Errors with respect to 1013 hPa, max.:
zero point: 0.1 0.1 0.1
calibration gas: 5.0 5.0 3.0
Effect of humidity (5 to 95 % r.h., 40 oC),
Errors with respect to 55 % r.h., max.:
zero point: 1.5 2.0 2.0
calibration gas: 5.0 4.0 5.0
Response time t90 at 25 oC:
without dust filter < 3 s in moving air (> 1 m/s), < 5 s in still air
with dust filter: < 15 s in moving air (>1 m/s), <30 s in still air
with dust filter and splash guard: < 30 s in moving air (>1 m/s), <60 s in still air
Effect of ow up to 6 m/s: no effect
1) gures per Kelvin
Transmitter performance IR Ex HC
(not evaluated in the 1st supplement to EC type test certicate
DMT 02 ATEX E 0178X).
Gas type Propane
Full scale of measuring range 100 % LEL
Linearity error, max.: 4.0 % LEL
Repeatability (DIN 1319)
zero point: 0.8 % LEL
50 % LEL: 1.0 % LEL
Long-term drift, 6 months
zero point errors, max.: 1.0 % LEL / month
errors at 50 % LEL, max.: 2.0 % LEL / month
Effect of temperature (25 to +50 oC)
Errors with respect to 20 oC, max.:
zero point: 0.05 % LEL / K
50 % LEL: 0.1 % LEL / K
Effect of pressure, max.
at 950 to 1100 hPa: 0.1 % of measured value / hPa
Effect of humidity, max.
at 0 to 95 % r.h., 40 oC:
zero point: 0.01 % LEL / % r.h.
50 % LEL: 0.02 % LEL / % r.h.
Response time t90 at 25 oC <45 s
Effect of ow up to 6 m/s: no effect
Cross sensitivities
(not evaluated in the 1st supplement to EC type test certicate DMT 02 ATEX E
0178X).
IR Ex HC cross sensitivities
when calibrated with propane
The values shown in the table are approximate values. The table does not claim to be
complete. The sensor may also be sensitive to other gases. Concentration of target-
gas 50 % LEL.
Gas / Vapour Chemical symbol Display in
% LEL propane1)
Acetone (CH3)2CO < 10
Benzene C6H12 <2
Butadiene CH2CHCHCH2 <4
Ethene C2 H4 <4
Ethine C2 H2 <1
Ethyl acetate CH3COOC2H5 < 25
Methane CH4 < 16
Methylethylketone CH3COC2H5 < 25
Toluene C6H5CH3 < 10
Xylene C6H4(CH3)2 < 17
1) Effect on measured values in mixtures with hydrocarbons (e.g. in ambient air) proportional to
displacement (partial pressure measurement).
IR Ex ES cross sensitivities
when calibrated with propane
The values shown in the table are approximate values and may deviate by 30 %.
The table does not claim to be complete. The sensor may also be sensitive to other
gases. Concentration of target gas 20 % LEL.
The conversion factors between % by vol. and LEL used in the tables are listed on
page 48.
Gas / Vapour Chemical symbol Display of Polytron
IR Ex ES in % LEL
Acetone CH3COCH3 12
Benzene C6 H6 11
Ethene C2 H4 8
Ethanol C2H5OH 31
Ethylacetate CH3COOC2H5 25
Hexane C6H14 15
i-Propanol (CH3)2CHOH 23
Methane CH4 19
Methanol CH3OH 47
Methylethylketone CH3COC2H5 17
Pentane C5H12 23
Propane C3 H8 20
Toluene C6H5CH3 15
o-Xylene C6H4(CH3)2 13
Process application
Not IR Ex IL transmitter!
Using the transmitter in a process applications continuous exposure to high
concentrations with certain hydrocarbons: on request.
Dimensions
Polytron IR Ex and IR Ex ES transmitters
145
120 103
106
152
120
82
225
04023655_04.eps
Polytron IR Ex IL transmitter
145
120 112
154
D Polytron IR Ex
120
82
ESC
200
04123655_04.eps
174
Polytron IR Ex HC transmitter
120
106
152
120
82
205
103
04223655_04.eps
145
106
152
120
170
82
04323655_04.eps
120 103
68
108
45
95
75
04423655_04.eps
80 58
Measurement Principle
The Polytron IR Ex, IR Ex ES, IR Ex IL and IR Ex HC transmitters are suitable for the
measurement of the concentration of hydrocarbons in ambient air, by the absorption
of infra-red radiation.
Operating principle
1 Infrared light source
2 Reference detector 1 2
3 Interference filter
4 Measuring detector 3
5 Beam splitter
6 Mirror
7 Sample gas
4
5
6 7 6
04523655_04.eps
IR Ex HC:
The ambient air to be monitored diffuses or is pumped into the optical system. A
replaceable dust lter protects the optical system against dust and dirt. From the
infra-red source broad-banded radiation passes through a window into the optical
system, is reected on the spherical mirror and focused, passes through another
window and hits a beam splitter. The radiation transmitted by the beam splitter
passes through a narrow-band interference lter, (measuring wavelength) into the
measuring detector.
The radiation reected by the beam splitter passes through another interference
lter (reference wavelength) into the reference detector.
Hydrocarbons in the gas mixture in the optical system cause the radiation to be
absorbed at the measuring wave length and, therefore, the intensity is reduced at the
detector. The stability of the zero point is guaranteed by a reference detector which
compensates the effects of build-up of deposit on the mirrors and windows as well
as possible uctuations of intensity in the infra-red light source.
Operating principle
1 Infrared source 4
2 Window
3 Cuvette
4 Mirror
5 Window
6 Beam splitter
7 Measurement filter 3 5
8 Measurement detector
9 Reference filter 9
10 Reference detector 2
10
1 7
8
04623655_04.eps
6
Order List
Name and description Order-Nr.
IR Ex Polytron transmitter 83 16 030
with built-in display and keypad
IR Ex NDF Polytron transmitter 83 16 031
without display and keypad,
with cable inlet for remote display unit
IR Ex NDH Polytron transmitter 83 16 032
without display and keypad,
with plug connector for hand-held terminal
Accessories
Splash guard (IR Ex, IR Ex ES) 68 08 671
Protective screen for sensor (IR Ex, IR Ex ES) 68 08 906
Cable gland for second connecting cable 83 11 370
(for 13 to 15 mm cable diameter)
Cable gland for second connecting cable 83 16 563
(for 6 to 12 mm cable diameter)
Calibration accessories
Calibration adapter (IR Ex, IR Ex ES) 68 08 667
Adapter (stainless steel) IR Ex HC (for calibration) 68 09 240
Calibration adapter (IR Ex HC) 68 06 978
Replacement parts
Dust lter (IR Ex, IR Ex ES) 68 08 516
Filter disc (IR Ex HC) 68 09 241
packing qty. 10
IECEx - Approval
Declaration of Conformity
Index
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22, 72
Adjust pre-amplification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
After connecting electrical installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Air movement/draft . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Density of gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Display errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Display sensor temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Display software-specific information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Display warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Dust filter and splash guard fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Dust filter, removed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Handling Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Infra-red technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Initial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Installation for diffusion operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Installation for flow operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Installation point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Installation with inflexible tube connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Installing electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Installing gas path for calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Installing transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Instrument fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Maintenance Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Maintenance Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Maintenance opening . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Maintenance plug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Measurement Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Measuring other hydrocarbons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Measuring ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Microprocessor technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Models differ from the IR Ex and IR ExIL versions . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Mounting IR Ex display unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Mounting IR junction box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Mounting to a wall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Removing cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Removing dust filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Replacement parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Replacing dust filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Transmitter performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Zero clamping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
90 23 655 - GA 4675.601 en
Drger Safety AG & Co. KGaA
4th edition - May 2006
Subject to alteration