Manual Supplement

Manual Title: CuffLink Operators Supplement Issue: 2


Part Number: 2242915 Issue Date: 10/08
Print Date: December 2007 Page Count: 1
Revision/Date:

This supplement contains information necessary to ensure the accuracy of


the above manual. Enter the corrections in the manual if either one of the
following conditions exist:

1. The revision letter stamped on the indicated PCA is equal to or higher


than that given with each change.
2. No revision letter is indicated at the beginning of the change.

2008 Fluke Corporation. All rights reserved. Printed in U.S.A.


CuffLink Operators Manual Supplement

Change #1
On page 1-16, Table 1-5, change:
From:
Adult cuff mandrel 2392381
To:
Adult cuff mandrel spacer blocks 2392381

Delete the following rows:


Neonate cuff mandrel 2224008

Adult cuff mandrel spacer blocks 2230310


(three required)

Change #2, 48806


On page 1-13, under Target Blood Pressure Selections, under Adult and Neonate,
Change: 400/65 (75)
To: 100/65 (75)

On page 1-14, under Accuracy,


Change: Systolic/diastolic Mean Arterial Pressure (MAP) 1.0 % of Target Value
To: Dynamic NIBP Response Repeatability (Systolic/Diastalic/Mean)..1 % of Target Value