1. Fill a pot with water and adjust the water temperature to lower value of the temperature
transmitter/gauge range by using ice or warm water.
2. Place the certified thermometer (liquid in glass) and temperature transmitter into water.
Wait for both temperatures reading to stabilize.
3. Fill a pot with water and adjust the water temperature to higher value of the temperature
transmitter/gauge range by using warm water/heat.
4. Place the certified thermometer (liquid in glass) and temperature transmitter into hot
water. Wait for both temperatures reading to stabilize.