Download as pdf or txt
Download as pdf or txt
You are on page 1of 144

1

MC5 Training General, Part 1


MC5 Hardware
2

Modules on the Front Panel


The adjacent picture presents three of MC5s
modules:
E le c tr ic a l M o d u le
The Electrical Module (E M o d u le )
(E module). E le c tr ic a l a n d
T e m p e r a tu r e M o d u le
The Electrical and Temperature Module (E T M o d u le )
(ET module). R e fe re n c e
J u n c tio n M o d u le
The Reference Junction Module (R J M o d u le )

(RJ module).

In addition to the modules mentioned above, MC5


may also have Internal Pressure Modules and be
connected to External Pressure Modules.
They are discussed further on in this section.
E, ET and RJ modules
3

E Module
The E module is capable of performing the following 2 -w x m t r

tasks:
+ 2 4 V 1 m e a s / s in k

Current measurement using either MC5s own


24 Volt supply or an external supply. V , ,
H A R T

Voltage measurement with separate terminals for M E A S U R E

measuring low voltages (1 Volt or less).


E
M a x in p u t : L o w V
6 0 V D C / 3 0 V A C

Frequency measurement and pulse counting. E module


Switch state sensing.
The E module is typically used for measuring the
output signal of instruments.

The optional HART communication capability is


presented in a separate section of this MC5 training
material.
4

ET Module
The ET modules measuring capabilities: T / C IN T . R J

Low Voltage and T/C measurement using either T / C , L o w V R , R T D


4 -w m e a s
V , 1,

the internal reference junction module (RJ) or the T / C O R E X T

Low Voltage connectors. W IR E S O N L Y

3 & 4 -w m e a s

Resistance and RTD measurement. S E N S O R M E A S U R E & S IM U L A T E O U T P U T


E T
M a x in p u t :

Generation/simulation capabilities: 6 0 V D C / 3 0 V A C

ET module
T/C simulation using either the internal reference
junction module or the Low Voltage connectors.
Resistance and RTD simulation.
Voltage, Current, Frequency and Pulse
generation.

The ET module is typically used for


measuring/simulating/generating the input signal of a
temperature or electrical instrument.
5

RJ Module
The RJ module is an optional addition to the ET T / C IN T . R J

module.
4 -w m e a s
T / C , L o w V R , R T D V , 1,

It performs reference junction compensation when W


T / C O R E X T
IR E S O N L Y

calibrating thermocouples or instruments connected 3 & 4 -w m e a s

to a thermocouple. S E N S O R M E A S U R E & S IM U L A T E O U T P U T
E T
M a x in p u t :
6 0 V D C / 3 0 V A C

RJ module
6

Keyboard, pt. 1
Keys on the left side of the display
C u rs o r
The cursor keys are used when moving the cursor on K e y s
the screen. They also have several special functions (A rro w K e y s )
in certain situations, e.g. when tuning the contrast of
the display. E n te r K e y

The Enter key finishes the entering of values.


7

Keyboard, pt. 2
Keys above the display O n /O ff K e y

Each key above display have a special task. See the


adjacent picture for descriptions. H e lp K e y S ig n K e y

B a c k lig h t K e y D e c im a l K e y
8

Keyboard, pt. 3
Keys on the right side of the display

These keys have several functions:


Entering numbers when a numeric field is
highlighted.
N u m e r ic
Selecting menu items when a menu is open on K e y s
the right side of the display. (M e n u K e y s )

Keys 0 and 8 are also used for opening a


possible submenu
Key 9 can be used for accepting values (like the
Enter key) in non-numerical fields.
9

Keyboard, pt. 4
Keys below the display

The meaning of the Function Keys vary depending


on the situation. The lower part of the display
indicates what the Function Keys stand for at the
moment. F u n c tio n K e y s
10

Upper Panel Connections


MC5 can include up to three Internal Pressure E x te rn a l P re s -
Modules with different measurement ranges. s u r e M o d u le s

In te r n a l P r e s -
Additionally, a connector for an External Pressure s u r e M o d u le s
Module is available. This further expands the
pressure measuring capabilities of MC5.

Pressure module connections


11

Connections on the Left Side


The uppermost connector is dedicated to the optional
Environmental Temperature Sensor. The sensor
can be used for measuring the environmental
temperature and/or the temperature of the instrument
to be calibrated.
E n v ir o n m e n t
T e m p e ra tu re
The Computer/Printer Interface (COMP/PRT) is S e n s o r In te r fa c e
meant for communicating with a PC with calibration C o m p u te r/
software or the optional printer for printing screen P r in te r In te r fa c e
dumps. A u x ilia r y In s tr u -
m e n t In te r fa c e
The Auxiliary Instrument Interface (AUX) is used
when communicating with external devices, e.g. a
pressure controller or a temperature bath / dry block.
12

Connectable Devices
A list of devices connectable to MC5:

External Pressure Modules (EXT).

Environment Temperature Sensor (ENV).

Battery operated portable printer (COMP/PRT).

Calibration software with suitable communication


drivers (COMP/PRT), e.g CMX Beamex Calibration
Maintenance Management software.

Pressure Controllers (AUX), e.g.:


Beamex POC4, Druck DPI510 and Druck DPI515.

Temperature Baths/Dry Blocks (AUX), e.g.:


Ametek/Jofra, Isotech, TekKnow
and HART Scientific.
13

MC5 Training General, Part 2


MC5 Firmware
14

Menu Structure
The adjacent picture presents the basic structure of S T A R T U P
MC5s menus. P R O C E D U R E
M e a s u re m e n t,
G e n e r a tio n D a ta
After the startup procedure MC5 always ends up in & S im u la tio n B A S IC L o g g in g
T r a n s m itte r M O D E S te p p in g
Basic Mode. There MC5 is like an advanced S im u la tio n a n d R a m p in g

multimeter with, e.g. the possibility to simulate a C A L IB R A T IO N M A IN T E -


transmitter. M O D E N A N C E
In s tr u m e n t C o n fig u r in g
C a lib r a tio n th e C a lib r a to r
Calibration Mode is the place to calibrate In s tr u m e n t S e ttin g
A d ju s tm e n t
instruments. V ie w in g
In s tr u m e n t
D a ta b a s e
T im e a n d D a te
A d ju s tin g
th e R e s u lts M a in te n a n c e th e C a lib r a to r

Maintenance Mode allows to save your own


preferences, like the language used in the user MC5 menu structure
interface.
15

User Interface, pt. 1


Status Bar, Window 1 and Window 2 B a t t e r y 's c h a r g e le v e l S ta tu s B a r
1 3 .0 1 .2 0 0 5 1 5 :3 0
1 P re s s u re A la r m < - ( - - )
The Status Bar displays the charge level of the M e a s u r e m e n t r e a d in g P 1 : IN T 2 C / -1 .0 ... 2 b a r g

battery, the current date and time as well as some A d d itio n a l d a ta r o w s : 0 .4 5 2 1 g a u g e


b a r
S e n s o r S u p p ly E T : V o lta g e G e n e r a tio n 5 .0
other information, e.g. when HART communication E x te rn a l
S p e c ia l M e a s u
D e
re m
v ic e
e n t
P O C 4
M a x im
, b a r
u m V a lu e
0 .5
1 .1 2 6 2
is active. E x tra In fo In te rn a l T e m p e ra tu re [ C ] 2 2 .1 C
2 C u rre n t
E : C u rre n t M e a s u re m e n t
D is p la y a r e a d iv id e d
The two Windows display the main in to W in d o w 1 a n d 2 0 .1 1 2 0
measurement/generation/simulation value using a
large font. Possible additional data is shown using M in im u m V a lu e 0 .0 0 0 0

smaller font. C a lib r a tio n


M o d e M o d e M E N U

The Menu Function Key opens up menus for setting A sample picture of a Basic Mode screen
up the windows and configuring the calibrator.

The Calibration Mode Function Key advances to


Calibration Mode.
16

User Interface, pt. 2


Function Key and Menu Elements M e n u
1 3 .0 1 .2 0 0 5 1 5 :4 4
1 P re s s u re A Q l a u r a m n t i< t y
The adjacent picture displays the Window 1 Setup P 1 : IN T 2 C / -1 .0
P a
h P a
... 2 b a r g [P r e s s u r e ]
In d ic a to r
fo r m u ltip le
menu. Other menus open from the Function Key 0 .4 5 2 1
m b a r
k P a
(g a u g e )
mF u b n a c r t / P o r t
[P 1 : IN T 2 C ]
m e n u p a g e s
E T : V o l t ab g a e r G e n e r a tio n
menu. P O C 4 , b Ma rP a
M a x i m u m l b /v f a t 2 l u e : 1 .1 2 6 2
D i s 0p . l 5a y b M a o r d e
[ E b na r g . U n i t ]
P o p -u p lis t
I n t e r n a l tp e s m i p e r a U n it
g f/c m 2 P re s e n t
Use the numeric keys 1 to 7 to select menu items. 2 C u r r e kn g t f / m 2
E : C u r r e kn gt M f / c e m a s 2 u re m e n t
[b a r]
s e le c tio n

Use keys 0 and 8 to see more menu pages (if 0 .1 1 2


k p /c m 2
a t
H A R T

m m H 2 O P re s s u re

applicable). c m H 2 O
m H 2 O
T y p e
[g a u g e ] In d ic a to r
M in im u m v a lu e : 0 .0 0 0 0 Z e ro fo r m u ltip le
iw c P re s s u re
M o d u le
m e n u p a g e s
To scroll through a pop-up list, use the cursor keys or W in d o w
S e tu p
1 W in d o w 2
S e tu p
O th e rs
M o d e
C lo s e
M E N U

the same numeric key that opened the pop-up menu. F u n c tio n k e y b a r

A sample picture of a Basic Mode menu


17

User Interface, pt. 3


Numeric Fields

To start editing a numeric field, move the highlight R a n g e 0 % 4 .0 0 0 0


1 0 0 % m A
(the rectangle) to the field and press either the 1 6 .0 0 0 0

C/Edit Function Key or the Enter Key. Numeric fields, one of them highlighted.
Enter/edit the value and save by pressing either the
D/OK Function Key or the Enter Key.
R a n g e 0 % 4 .0 0 0 0
To help out editing the numeric value, MC5 provides 1 0 0 % 2 0 _ m A

the following Function Keys:


Numeric field in edit mode.
A/Cancel Cancel all edits
B/Clear Clear the field
C/Delete Delete the character to the left of the
cursor
18

User Interface, pt. 4


Text Fields P o s itio n ID P T 1 0 6 .1

To start editing a text field, do as you did for a Highlighted text field.
numeric field.
1 6 .1 0 .2 0 0 0 1 5 :2 1

The editing is done by selecting characters from the P T 1 0 6 .1 A


D
B
E F
C

opened menu. Remember to use keys 0 and 8 to G H I


see more characters. J K L
M N O

Save your edits by pressing either the D/Accept D e v ic e ID P Q R


S T U
Function Key or the Enter Key. D e v ic e N a m e V W X
Y Z
To help out editing the text field, MC5 provides the E r r o r C a lc . M e th o d
R e je c t if
% o f s p a n
> 0 .5 0
O

following Function Keys: A d ju s t if > 0 .3 0






D o n o t A d ju s t if < 0 .1 0
A d ju s t to
< 0 .1 0
A/Cancel Cancel all edits

C/Delete Delete the character to the left of the C a n c e l D e le te A c c e p t

cursor
Text field in edit mode.
19

User Interface, pt. 5


Drop Down Fields In p u t M e th o d M e a s u re d

MC5 uses drop down fields for presenting a fixed list Highlighted drop down field.
of options. To start editing a drop down field, do as
you did for a numeric field.

Use the up and down cursor keys or keys 0 and In p u t M e th o d M e a s u re d


8 to scroll through the available items. M e a
K e y
s u re d
e d
C o n tr o lle d
Use D/Select Function Key or the Enter Key to
select an item. Drop down field in edit mode.

To help out editing the drop down field, MC5


provides the following Function Keys:

A/Cancel Cancel all edits


B/Top of List Jump back to the beginning of the list
20

User Interface, pt. 6


Selection Lists 1 6 .1 0 .2 0 0 0 1 5 :2 1

P O S IT IO N /D E V IC E ID
MC5 uses selection lists for presenting lists with no 1 0
1 1
1 -X
2 -T
L -0 0 1 .1
T -0 0 3 .1
predetermined amount of options. 1 1
E S
2 -T
w
T -0 0 7 .1

P T 1 0 6 .1

Use the up and down cursor keys or keys 0 and P T


P T
1 1
1 1
2 .1
2 .1
2
5 -1

8 to scroll through the available items. P T


P T
1 1
1 1
2 .1
2 .1
5 -2
6
T I
V V
Use C/Select Function Key or the Enter Key to D e v ic e ID

select an item. T T 1 1 2 .0 9
P o s itio n N a m e
P ro d u c t te m p e ra tu re
C a lib r a te d
2 2 .0 2 .2 0 0 0

B a s ic S e le c t M E N U
M o d e

Selection list.
21

MC5 Training Settings and Maintenance


How to Customize Your MC5
and Set Time and Date
22

Opening the Settings Window

To open the settings window (starting from Basic


Mode), press D/Menu, C/Others and
1/Settings.

C a lib r a tio n C W a il ni b d r o a wt i o 1 n W in d o w 2
M E N U O th e rs M C lE o N s U e
M o d e MS e o t d u e p S e tu p M E N U

S e ttin g s
g a u g e
bM a a r i n t e n a n c e

H A R T
C o m m u n ic a tio n
S e ttin g s

D a ta
L o g g in g
m A
T r a n s m itte r /
S w itc h
S im u la tio n
23

1st Settings Page

Set the Date and Time Format as well as the 2 5 .0 9 .2 0 0 0 8 :5 3


S E T T IN G S
User Interface Language according to local
D a te F o rm a t d d .m m .y y y y
needs. T im e F o r m a t h :m m 2 4 h

L a n g u a g e E n g lis h

Set Auto-off Delays and Sound Volumes A u to - o ff D e la y s ( 0 = n e v e r ) [m in ]


C a lib r a to r
according to your own needs.
0
B a c k lig h t 5

S o u n d V o lu m e s
K e y C lic k L o w
A la r m H ig h
A tte n tio n M e d iu m

Note.
O u ts id e S p a n M e d iu m
E rro r L o w

All MC5s include English among the available N e x t


languages. The available other languages C lo s e
P a g e
E d it

depend on the MC5 at hand.


Settings, 1st page
24

2nd Settings Page

Open the second settings page by pressing the 0 1 .0 2 .2 0 0 5 1 3 :3 6


S E T T IN G S
B/Next Page Function Key.
N e t F re q u e n c y 5 0 H z
B a c k lig h t P o w e r N o rm a l

Set the Net Frequency, Temperature Unit and T e m p e r a tu r e U n it C


the Temperature Scale according to local T e m p e r a tu r e S c a le IT S 9 0

needs. C o p y E d it e d I n s t r u m e n t 'P o s it io n I D '


t o 'D e v ic e I D ' IT S 9 0

Set Backlight Power and ENV Sensor Usage


according to your own needs.

N e x t
C lo s e E d it
P a g e

Note.
All changes in settings become valid Settings, 2nd page
immediately.
25

Opening the Time and Date


Settings Window

To open the settings window (starting from Basic


Mode), press D/Menu, C/Others and
1/Settings.

C a lib r a tio n C W a il ni b d r o a wt i o 1 n W in d o w 2
M E N U O th e rs M C lE o N s U e
M o d e MS e o t d u e p S e tu p M E N U

S e ttin g s T im e /D a te
g a u g e g a u g e
b a r Cb a a r l i b r a t o r
M a in te n a n c e
A d ju s tm e n t
H A R T
C o m m u n ic a tio n
S e ttin g s

D a ta
L o g g in g
m A m A
T r a n s m itte r /
S w itc h
S im u la tio n
26

Setting Time and Date

The date must always be given in dd.mm.yyyy 1 3 .0 6 .2 0 0 2 1 6 :0 2


T IM E /D A T E S E T T IN G S
format no matter what the configured date format
D a te (d d .m m .y y y y ) 1 3 0 6 2 0 0 2
is.
T im e ( h h :m m :s s ) 1 6 0 2 2 2

The time must always be given in the 24 hour


format no matter what the configured time format
is.

The date and time will be updated when you


press the D/Accept Function Key.
C a n c e l E d it A c c e p t

Time/Date Settings Page


27

MC5 Training Basic Mode, Part 1


Setting Up Window 1 and 2
28

Window Setup
To configure a window in Basic Mode, select
D/MENU and then either
A/Window 1 Setup or C a lib r a tio n
M o d e
M E N U

B/Window 2 Setup.
1 4 .0 1 .2 0 0 5 8 :3 6
1 P re s s u re Q u a n tit y
P 1 : IN T 2 C /-1 .0 ... 2 b a r g [P re s s u re ]

0 .7 0 8 3 g a u g e
Fb ua r n c t / P o rt
[P 1 : IN T 2 C ]

D is p la y M o d e
[E n g . U n its ]

U n it
2 C u rre n t [b a r]
E : C u rre n t M e a s u re m e n t
H A R T
0 .4 3 7 4 m A
P re s s u re
T y p e
[g a u g e ]
Z e ro
P re s s u re
M o d u le
C W a il ni b d r o a wt i o 1 n W in d o w 2 O th e rs M C El o N s U e
MS e o t d u e p S e tu p M E N U

Opening the Window Setup menu


29

Selecting the Quantity


The first thing to do is selecting a Quantity from 1 4 .0 1 .2 0 0 5 8 :3 6
1N o n P e r e s s u r e
the opened menu. To do that, select P P 1 r e : s I N s Tu 2r e C / - 1 . 0 ... 2 b a r g
Q u a n tit
[P re s s u
y
re ]

1/Quantity. C u rre n t
V o lta g e
0 .7 0 8 3 g a u g e
Fb ua r n c t / P o rt
[P 1 : IN T 2 C ]
R e s is ta n c e
S w itc h D is p la y M o d e
Note. F re q u e n c y
P u ls e s
[E n g . U n its ]

The available Quantities depend on the installed R T D -te m p e ra t


2 T / C C- t eu m r r ep n e tr a t u
u re U n it
[b a r]
re
modules and what is already reserved for use in E :N C V u Tr r e e m n t p M . e a s u re m e n t
H A R T
the other window. 0 .4 3 7 4 m A
P re s s u re
T y p e
[g a u g e ]
Z e ro
P re s s u re
M o d u le
C W a il ni b d r o a wt i o 1 n W in d o w 2 O th e rs M C lE o N s U e
MS e o t d u e p S e tu p M E N U

Selecting the Quantity


30

Selecting the Function/Port


The other important thing is selecting a 1 4 .0 1 .2 0 0 5 8 :3 6
1 P re s s u re
Function/Port by pressing P 1 : IN T 2 C /-1 .0 ... 2 b a r g
Q u a n tit
[P re s s u
y
re ]

2/Funct/Port. (Just 2/Port in some cases) P 1 : -1 .0 ... 2 b a r g0 .7 0 8 3


P 2 : -0 .4 1 ... 0 .4 1 b a r g
g a u g e
Fb ua r n c t / P o rt
[P 1 : IN T 2 C ]
P B : 0 .8 ... 1 .2 b a r a b s
D is p la y M o d e
Note. [E n g . U n its ]

The available Functions/Ports depend on the 2 C u rre n t


U n it
[b a r]
installed modules and what is already reserved E : C u rre n t M e a s u re m e n t
H A R T
in the other window. 0 .4 3 7 4 m A
P re s s u re
T y p e
[g a u g e ]
Z e ro
P re s s u re
M o d u le
C W a il ni b d r o a wt i o 1 n W in d o w 2 O th e rs M C lE o N s U e
MS e o t d u e p S e tu p M E N U

Selecting the Function/Port.


31

Other Window Settings


The two first items (Quantity and Q u a n tity
[C u rre n t]
Function/Port) are the ones defining what is F u n c t / P o r Qt u a n t i t y

measured/generated/simulated and which [ E T : I ( g e n [) S] w i t c h ]

D i s p l a y M Fo ud ne c t / P o r Qt u a n t i t y
connectors/terminals are in use. [ E n g . U n i t [ s E ] : S w i t c h [ ]F r e q u e n c y ]

U n it T r i g g e r L e F v u e n l c t / P o r Qt u a n tity
[m A ] [1 .5 V ]
All other settings (Display Mode, Unit etc) refine
[ E T : f ( g e n [) P] u l s e s ]
S ta te
H A R T I n d i c a t i o n D i s p l a y M Fo ud ne c t / P o r Qt u a n t i t y
the windows configuration. [ / [ ]E n g . U n i t[ sE ] : P l s ( c o [ u R n T t )D ] - T e m p ]

H A R T U n it D i s p l a y M Fo ud ne c t / P o r Q t u a n t i t y
[k H z ] [ E n g . U n i t [ s E ] T : R T D ( [ m T /e C a - ) T ] e m p ]
The adjacent picture contains samples of menus H A R T T r i g g e r E d D g i se p l a y M F o u d n e c t / P o r t

for certain Quantity and Function/Port S o u n d W a v e fo rm


[ r i s i n g ] [ E n g . U n i t[ sE ] T : T C x ( m e a ) ]

combinations. H A R T U n it D is p la y M o d e
[O ff] [p o s . s q u a re ] [C ] [E n g . U n its ]

C l e a r C o u H n At e R r T U n it
[C ]
S e n s o r T y p e
H A R T
[P t1 0 0 a 3 8 5 ]

S e n s o r T y p e
[K N iC r /N iA l]
R e fe re n c e
J u n c tio n
[P t1 0 0 a 3 8 5 ]
32

MC5 Training Basic Mode, Part 2


Active Terminals/Connections for Different
Quantity and Function/Port Settings
33

Pressure Measurement

P 1 P 2 P 3
Quantity: Pressure
Function/Port: P1 , P2 or P3
(availability of ports and
measurement ranges depend
T / C I N T . R J 2 - w x m t r

4 - w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

on the installed modules) W


T / C
IR
O R
E S O
E X T
N L Y

S E N S O R M E A S U R E & S I M U L A
3

T E
& 4 -w m e a s

O U T P U T
V

M
H
,
A

E A S
,
R

U
T

E
E T E
M a x in p u t : L o w V
6 0 V D C / 3 0 V A C

Quantity: Pressure
Function/Port: EXT
(measurement range depend
on the connected EXT module)
T / C I N T . R J 2 - w x m t r

4 - w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T
W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S I M U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t : L o w V
6 0 V D C / 3 0 V A C
34

Current Measurement

Quantity: Current
Function/Port: E: I(meas) 100 mA T / C IN T . R J 2 -w x m t r

4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T
W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t : L o w V
6 0 V D C / 3 0 V A C

Internal Supply

Notes.
Based on the connections, MC5 automatically T / C IN T . R J 2 -w x m t r

detects whether the internal or an external


4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T

supply is used.
W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

The partly highlighted terminal is the HART


S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t : L o w V
6 0 V D C / 3 0 V A C

communication terminal. More of HART in a


External Supply
separate section.
35

Current Generation

Quantity: Current
Function/Port: ET: I(gen) 25 mA
T / C IN T . R J 2 -w x m t r

4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T
W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t :
L o w V
6 0 V D C / 3 0 V A C

Quantity: Current
Function/Port: E: I(control) 0 to 25 mA
T / C IN T . R J 2 -w x m t r T / C IN T . R J 2 -w x m t r

4 -w m e a s + 2 4 V 1 4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1, m e a s / s in k T / C , L o w V R , R T D V , 1,

T / C O R E X T T / C O R E X T
W IR E S O N L Y W IR E S O N L Y

3 & 4 -w m e a s V , , 3 & 4 -w m e a s V , ,
H A R T H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E E T E
M a x in p u t : M a x in p u t : L o w V
L o w V 6 0 V D C / 3 0 V A C
6 0 V D C / 3 0 V A C

External Supply Internal Supply


Note.
Again, based on the connections, MC5 automatically detects whether the internal or an external supply is used.
36

Voltage Measurement

Quantity: Voltage
Function/Port: ET: LowV(mea) 500 mV
T / C IN T . R J 2 -w x m t r

4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T
W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t :
L o w V
6 0 V D C / 3 0 V A C

Quantity: Voltage
Function/Port: E: LowV(mea) 1 V
T / C IN T . R J 2 -w x m t r

4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T
W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t : L o w V
6 0 V D C / 3 0 V A C

Quantity: Voltage
Function/Port: E: V(meas) 50 V
T / C IN T . R J 2 -w x m t r

4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T
W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t : L o w V
6 0 V D C / 3 0 V A C
37

Voltage Generation

Quantity: Voltage
Function/Port: ET: LowV(sim) 500 mV
T / C IN T . R J 2 -w x m t r

4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T
W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t :
L o w V
6 0 V D C / 3 0 V A C

Quantity: Voltage
Function/Port: ET: V(gen) 12 V
T / C IN T . R J 2 -w x m t r

4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T
W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t : L o w V
6 0 V D C / 3 0 V A C
38

Resistance
Measurement/Simulation

Quantity: Resistance
Function/Port: ET: R(meas) 0 to 4000 ohm
T / C IN T . R J 2 -w x m t r

4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T

The partly highlighted terminals are optional


W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

(used in 3 & 4-wire measurement).


S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t :
L o w V
6 0 V D C / 3 0 V A C

Quantity: Resistance T / C IN T . R J 2 -w x m t r

Function/Port: ET: R(sim) 1 to 4000 ohm T / C , L o w V R , R T D


4 -w m e a s
V , 1,
+ 2 4 V 1 m e a s / s in k

T / C O R E X T
W IR E S O N L Y

Connect three or more cables as shown in the


3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E

lowermost picture.
E T E
M a x in p u t : L o w V
6 0 V D C / 3 0 V A C
39

Switch State Sensing

Quantity: Switch
Function/Port: E: Switch
T / C IN T . R J 2 -w x m t r

4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T

Also check the Sound setting.


W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t :
L o w V
6 0 V D C / 3 0 V A C
40

Frequency
Measurement/Generation

Quantity: Frequency
Function/Port: E: f(meas) 0.0028 to 50 kHz
T / C IN T . R J 2 -w x m t r

4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T

Also check the Unit setting.


W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t :
L o w V
6 0 V D C / 3 0 V A C

Quantity: Frequency T / C IN T . R J 2 -w x m t r

Function/Port: ET: f(gen) 0.00028 to 50 kHz T / C , L o w V R , R T D


4 -w m e a s
V , 1,
+ 2 4 V 1 m e a s / s in k

T / C O R E X T
W IR E S O N L Y

Also check the Waveform setting.


3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t :
L o w V
6 0 V D C / 3 0 V A C

Symmetric square: Positive square: Sine:


V o lta g e , V V o lta g e , V V o lta g e , V
6 6 6
3 3 3
0 0 0
-3 -3 -3
-6 -6 -6
T im e T im e T im e
41

Pulse Counting/Generation

Quantity: Pulses
Function/Port: E: Pls(count)
T / C IN T . R J 2 -w x m t r

0 to 9 999 999 pulses


4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T
W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

Also check the Trigg. Edge setting.


S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t :
L o w V
6 0 V D C / 3 0 V A C

Quantity: Pulses
Function/Port: ET: Pls(gen)
T / C IN T . R J 2 -w x m t r

+ 2 4 V 1

0 to 9 999 999 pulses


4 -w m e a s m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T
W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

Also check the Polarity setting.


S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t : L o w V
6 0 V D C / 3 0 V A C

Symmetric: Positive:
V o lta g e , V V o lta g e , V
6 6
3 3
0 0
-3 -3
-6 -6
T im e T im e
42

RTD Temperature
Measurement/Simulation

Quantity: RTD-Temp.
Function/Port: ET: RTD(mea)
T / C IN T . R J 2 -w x m t r

Pt sensors: -200 to 850C


4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T
W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

Also check the Sensor Type setting.


S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t :
L o w V
6 0 V D C / 3 0 V A C

The partly highlighted terminals are optional


(used in 3 & 4-wire measurement).

Quantity: RTD-Temp. T / C IN T . R J 2 -w x m t r

Function/Port: ET: RTD(sim) T / C , L o w V R , R T D


4 -w m e a s
V , 1,
+ 2 4 V 1 m e a s / s in k

Pt sensors: -200 to 850C T / C O R E X T


W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E

Also check the Sensor Type setting.


M a x in p u t : L o w V
6 0 V D C / 3 0 V A C

When connecting three or more cables, do it as


shown in the lowermost picture.
43

T/C Temperature
Measurement

Quantity: T/C-Temp
Function/Port: ET: TCi(mea) T / C IN T . R J 2 -w x m t r

(when using the


4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T

Internal Reference Junction)


W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t :
L o w V
6 0 V D C / 3 0 V A C

Also check the Sensor Type setting.

Quantity: T/C-Temp T / C IN T . R J 2 -w x m t r

Function/Port: ET: TCx(mea) T / C , L o w V R , R T D


4 -w m e a s
V , 1,
+ 2 4 V 1 m e a s / s in k

(for other Reference Junction T / C O R E X T


W IR E S O N L Y

compensation methods)
3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t : L o w V
6 0 V D C / 3 0 V A C

Also check the Sensor Type and Reference


Junction settings.

Measurement range depend on the T/C type.


44

T/C Temperature
Simulation

Quantity: T/C-Temp
Function/Port: ET: TCi(sim) T / C IN T . R J 2 -w x m t r

(when using the


4 -w m e a s + 2 4 V 1 m e a s / s in k
T / C , L o w V R , R T D V , 1,

T / C O R E X T

Internal Reference Junction)


W IR E S O N L Y

3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t :
L o w V
6 0 V D C / 3 0 V A C

Also check the Sensor Type setting.

Quantity: T/C-Temp T / C IN T . R J 2 -w x m t r

Function/Port: ET: TCx(sim) T / C , L o w V R , R T D


4 -w m e a s
V , 1,
+ 2 4 V 1 m e a s / s in k

(for other Reference Junction W


T / C O R E X T
IR E S O N L Y

compensation methods)
3 & 4 -w m e a s V , ,
H A R T

S E N S O R M E A S U R E & S IM U L A T E O U T P U T M E A S U R E
E T E
M a x in p u t :
L o w V
6 0 V D C / 3 0 V A C

Also check the Sensor Type and Reference


Junction settings.

Simulation range depend on the T/C type.


45

MC5 Training Basic Mode, Part 3


Generation/Simulation Specifics
46

Changing the Value of the


Generated/Simulated Signal

Move the field indicator on the display until it 1 8 .0 1 .2 0 0 5 1 5 :4 5

surrounds the value of generated/simulated 1 T /C -te m p e r a tu r e


E T : T /C - S im , In t R J

signal (use the B/Field Function Key or the 1 2 0 .0 0 _ K N iC r /N iA l


C (IT S 9 0 )
cursor keys). R J In te rn a l 1 .1 0 2 4

Enter the new value using the numeric keys.


You may cancel the editing by pressing the 2 C u rre n t
E : C u rre n t M e a s u re m e n t

A/Cancel Function Key. The C/Delete 0 .4 3 7 4 m A


Function Key removes the rightmost digit.

Accept the new value by pressing either the


Enter key or the D/OK Function Key. C a n c e l C le a r D e le te O K
47

Fine Tuning the


Generated/Simulated Signal

Press either of the left or right cursor keys to 1 8 .0 1 .2 0 0 5 1 5 :4 8

start the Fine Tuning of the selected numeric 1 T /C -te m p e r a tu r e


E T : T /C - S im , In t R J

field. Then one of the digits in the number is 1 2 0 .0 0 K N iC r /N iA l


C (IT S 9 0 )
underlined indicating which digit can be fine R J In te rn a l 1 .1 0 2 4

tuned. If needed, press the left or right arrow


buttons again to select another digit for Fine 2 C u rre n t
Tuning. E : C u rre n t M e a s u re m e n t

0 .4 3 7 4 m A

To change the value of the underlined digit, use


the up and down cursor keys. The changes take
effect immediately.
C le a r E d it C lo s e

To end Fine Tuning, press either the Enter key


or the D/Close Function Key. The C/Edit
Function Key ends Fine Tuning and allows you
to enter a number using the numeric keys.
48

MC5 Training Calibration, Part 1


Adding Instruments and Editing Instrument Data
49

Adding Instruments
Adding instruments to be calibrated can be 0 3 .1 0 .2 0 0 0 0 8 :5 1

downloaded from a calibration software that


communicates with MC5 (e.g. CMX).
P O S IT IO N /D E V IC E ID 2 4 .0 1 .2 0 0 5 9 :2 7
C re a te N e w In s tr u m e n t... C re a te N e w

If MC5 have no instruments in its memory, the


P O S IT IO N /D E V IC E ID
In s tr u m e n t
1 0 1 -X I-0 0 1 .1
1 1 2 -T T -0 0 3 .1
Calibration Mode's startup window contains a 1 1 2
P T 1
-T T -0 0 7 .1
0 1 .6

possibility to create a new instrument (see the D e v ic e ID P T 1 1 2 .1 2


P T 1 1 5 .1 5 -1
P o s itio n N a m e P T 1 1 5 .1 5 -2

uppermost picture). If there are already some C a lib r a te d


P T 1
T T 1
T T 1
1 2 .1 6
1 2 .0 8
1 2 .1 0
instruments, open the menu to find the Create a B a s ic
S e le c t
T T 1
M ED N e Uv ic e I D
1 2 .1 2

New Instrument option.


M o d e
T T 1 1 2 .0 7 D e le te A ll
P o s itio n N a m e In s tr u m e n ts
In ta k e T e m p e ra tu re o n F e e d 1
C a lib r a te d
N o t C a lib r a te d
B a s ic C lo s e
S e le c t M E N U
M o d e M E N U

Adding instruments
50

Instrument
Input and Output

When creating an all new instrument, you will 0 3 .1 0 .2 0 0 0 1 2 :1 4

first have to select the Input Quantity and the


Output Quantity. These selections cannot be
S E L E C T IN S T R U M E N T IN P U T
V o lta g e
L o w V o lta g e
changed afterwards. C u rre n t
R e s is ta n c e
0 3 .1 0 .2 0 0 0 1 2 :1 5

P re s s u re
T e m p e ra tu re S E L E C T IN S T R U M E N T O U T P U T
T C T e m p e ra tu re
V o lta g e
R T D T e m p e ra tu re
L o w V o lta g e
F re q u e n c y
C u rre n t
V a lu e
R e s is ta n c e
P re s s u re
T e m p e ra tu re
T C T e m p e ra tu re
R T D T e m p e ra tu re
F re q u e n c y
B a c k O K
V a lu e
S w itc h

Input Quantity list


B a c k O K

Output Quantity list


51

General Instrument Data

The following five instrument data pages may be 1 9 .1 0 .2 0 0 0 1 0 :0 4


IN S T R U M E N T G E N E R A L D A T A
edited later on too.
P o s itio n ID P T 1 0 1 .6

Either the Position ID or Device ID field need to P o s itio n N a m e F e e d 3 , p re s s u re

be entered. All others are optional. D e v ic e ID

D e v ic e N a m e

E r r o r C a lc . M e th o d % o f s p a n
R e je c t if > 0 .5 0
A d ju s t if > 0 .3 0
D o n o t A d ju s t if < 0 .1 0
A d ju s t to < 0 .1 0
N e x t
S a v e E d it M E N U
p a g e

Instrument General Data page


52

Instrument Input Data

The Input Quantity defines which fields are 2 7 .0 9 .2 0 0 1 1 3 :0 1


IN S T R U M E N T IN P U T
required/visible. Common fields for all Quantities P re s s u re

are: In p u t M e th o d C o n tr o lle d
Input Method, P o rt 1 IN T 1 C
Unit and
Range P re s s u re T y p e G a u g e

U n it b a r

R a n g e 0 % 0 .0 0 0 0 0 0
1 0 0 % 1 .0 0 0 0 0 0

S e n s o r S u p p ly E T :V (g e n )
L e v e l ? ? ? V
N e x t
S a v e E d it M E N U
P a g e

Instrument Input Data page for


Pressure Input Quantity
53

Instrument Output Data

The Output Quantity defines which fields are 0 3 .1 0 .2 0 0 0 1 2 :4 7


IN S T R U M E N T O U T P U T

required/visible. Common fields for all Quantities, C u rre n t

except Switch are:


O u tp u t M e th o d M e a s u re d

P o rt E
Output Method, U n it m A

Unit and R a n g e 0 % 4 .0 0 0 0 0 0
0 3 .1 0 .2 0 0 0 1 3 :0 5
IN S T R U M E N T O U T P U T

Range 1 0 0 % 2 0 .0 0 0 0 0 0 S w itc h : P r e s s u r e [b a r ]

S w itc h T y p e N o r m a lly C lo s e d

N o m in a l P o in ts
A c tu a tin g 0 .7 0 0 0 0 0
T r a n s fe r F u n c tio n L in e a r D e a c tu a tin g 0 .6 0 0 0 0 0
Switches have an altogether different group of S a v e
N e x t
p a g e
E d it M E E r rN o U r C a l c u l a t i o n D i r e c t i o n s

settings as shown in the lower adjacent picture. A c tu a tin g


D e a c tu a tin g
U p /D o w n
U p /D o w n
Instrument Output
S c a n R a n g e
Data page for
Scan Range values define the range where MC5
S ta r t P o in t 0 .4 0 0 0 0 0

Current Output E n d P o in t 0 .9 0 0 0 0 0

searches for the approximate actuation and Quantity S a v e


N e x t
p a g e
E d it M E N U

deactuation point at the beginning of an


Instrument Output Data
automatic calibration of a switch. page for Switches
54

Instrument Calibration
Settings Data

Calibration settings define how MC5 performs 1 9 .1 0 .2 0 0 0 1 0 :0 6


C A L IB R A T IO N S E T T IN G S
the calibration.
C a lib r a tio n M e th o d A u to m a tic

Automatic calibration can only be used when C a lib r a tio n P o in ts 5

MC5 is able to generate or simulate the S e tp o in t D e la y 5 .0 0 s

instruments input signal (or optionally when MC5 M a x im u m A llo w e d C a lib r a tio n P o in t
D e v ia tio n < 0 .0 0 %

is able to communicate with an external device C a lib r a tio n R e p e a ts


capable of generate or simulating the A s F o u n d
A s L e ft
0
1
instruments input signal).
C a l. P e r io d , D a y s < 3 6 0

N e x t
S a v e E d it M E N U
p a g e

Calibration Settings page


55

Instrument Calibration
Instructions Data

In addition to the actual instrument data, there 0 3 .1 0 .2 0 0 0 1 3 :5 2


C A L IB R A T IO N IN S T R U C T IO N S
are three note fields. Starting Guide, Adjusting
S ta r tin g G u id e
Guide and Finishing Guide. R e q u ir e d to o ls : P r e s s u r e p u m p o r
a P r e s s u r e C o n tr o lle r

They can be used to provide the calibrating A d ju s tm e n t G u id e


technician with any kind of instructions that may R e m e m b e r th e " D o n o t A d ju s t if" lim it...

be practical during the calibration procedure. The


note texts may be written in MC5 or they can be F in is h in g G u id e

downloaded from a calibration software. C le a n u p a fte r y o u r w o r k !

N e x t
S a v e E d it M E N U
p a g e

Save the instrument data once all data is Calibration Instructions page
entered.
56

Instrument Window

The Instrument window gives you an overview of 2 5 .0 9 .2 0 0 0 1 0 :2 0


IN S T R U M E N T

the selected instrument. 1 1 2 -T T -0 0 3 .1


In ta k e T e m p e r a tu r e o n F e e d 1

T r a n s fe r F u n c tio n L in e a r
To edit the Instrument data, press the B/Edit C a l. P o in ts 5

Function Key. IN P U T 0 .0 0 ... 1 0 0 .0 0 C


M e th o d S im u la te d
S e n s o r T y p e P t1 0 0 a 3 8 5

O U T P U T 4 .0 0 0 0 ... 2 0 .0 0 0 0 m A
M e th o d M e a s u re d

B a c k E d it C a lib r a te M E N U

Instrument overview window


57

Making a Copy of an
Existing Instrument

There is one more way to create a new 2 4 .0 1 .2 0 0 5 1 1 :1 6


I N S T R U M E N TV i e w
instrument: Select an existing instrument, open 1 1 2 -T T -0 0 3 .1 In s tr u m e n t
D e ta ils
the menu when viewing the instruments In ta k e T e m p e r a tu r e o n F e e d 1
M a k e a C o p y
overview data. There is an option to make a copy T r a n s fe r F u n c tio n L in e a r
V ie w
of the selected instrument. C a l. P o in ts 5 C a lib r a tio n
R e s u lts

IN P U T 0 .0 0 ... 1 0 0 .0 0 C
M e th o d S i m u l a t e Dd e l e t e
S e n s o r T y p e P t 1 0 0 a 3 I8 n 5 s t r u m e n t

O U T P U T 4 .0 0 0 0 ... 2 0 .0 0 0 0 m A
M e th o d M e a s u re d
C o n tr o lle r
S e ttin g s
C lo s e
B a c k E d it C a lib r a te M E N U
M E N U

Making a copy of the selected


instrument
58

Deleting Instruments

To remove a single instrument (and its 1 4 .0 6 .2 0 0 0 8 :0 4


I N S T R U M E N TV i e w

calibration data), select it from the Position ID 1 1 2 -T T -0 0 3 .1 In s tr u m e n t


D e ta ils
In ta k e T e m p e r a tu r e o n F e e d 1

list. When viewing the instrument window, press T r a n s fe r F u n c tio n L in e a r


M a k e a C o p y

V ie w
D/Menu and 5/Delete Instrument (see the C a l. P o in ts 5 C a lib r a tio n
R e s u lts

upper adjacent picture). IN P U T 0 .0 0 ... 1 0 0 .0 0 C


2 5 .0 9 .2 0 0 0 1 0 :1 6
C re a te N e w
P O S IT IO N /D E V IC E ID
M e th o d S i m u l a t e Dd e l e t e In s tr u m e n t
1 0 1 -X I-0 0 1 .1
S e n s o r T y p e P t 1 0 0 a 3 I8 n 5 s t r u m e n 1 t 1
To remove all instruments (and their calibration O U T P U T 4 . 0 0 0 0 . . . 2 0 . 0 0 0 0 m P TA
1 1
2 -T
2 -T
1 0 1 .6
T -0 0 3 .1
T -0 0 7 .1

data), open the Position List and press D/Menu, M e th o d M e a s u re d


C o n t r o l l e rP T
P T 1 1
1 1
2 .1 2
5 .1 5 -1

6/Delete All Instruments (see the lower


S e ttin g s P T 1 1 5 .1 5 -2
P T 1 1 2 .1 6
C lo s e
B a c k E d it C a lib r a te M ET NT U1 1 2 .0 8

adjacent picture).
M E N U
T T 1 1 2 .1 0
T T 1 1 2 .1 2

Deleting an D e v ic e
T T 1 1
ID
2 .0 7 D e le te A ll
instrument P o s itio
In ta k
n N a m e In s tr u m e n ts
e T e m p e ra tu re o n F e e d 1
C a lib r a te d
N o t C a lib r a te d
B a s ic C lo s e
S e le c t M E N U
M o d e M E N U

Deleting all instruments


59

MC5 Training Calibration, Part 2


Calibrating Instruments
60

Selecting the Instrument


to be Calibrated

This section describes the calibration of 2 5 .0 9 .2 0 0 0 1 0 :1 6

instruments that already exist in MC5's memory 1 0 1


P O S IT IO N /D E V IC E ID
-X I-0 0 1 .1

(actually the instrument data is entered into


1 1 2 -T T -0 0 3 .1
1 1 2 -T T -0 0 7 .1
P T 1 0 1 .6

MC5). P T 1
P T 1
P T 1
1 2 .1 2
1 5 .1 5 -1
1 5 .1 5 -2
P T 1 1 2 .1 6
T T 1 1 2 .0 8

To select an existing instrument, go from Basic T T 1


T T 1
1 2 .1 0
1 2 .1 2

Mode to Calibration Mode and select the D e v ic e


T T 1 1
P o s itio
ID
2 .0 7
n N a m e
instrument from the displayed selection list. In ta k
C a lib r a
e T e m p e ra tu re o n F e e d 1
te d
N o t C a lib r a te d
B a s ic
S e le c t M E N U
M o d e

Selecting the instrument to be


calibrated
Note.
Creating new instruments is described in another
part of this training material.
61

Selecting the Instrument


to be Calibrated

If your instrument is calibrated using an external 8 .1 1 .2 0 0 1 1 5 :2 4


I N S T R U M E N TV i e w

device that is communicating with MC5, open the 1 1 2 -T T -0 0 3 .1 In s tr u m e n t


D e ta ils
In ta k e T e m p e r a tu r e o n F e e d 1

instrument overview windows menu (press


D e le te
In s tr u m e n t
T r a n s fe r F u n c tio n L in e a r
V ie w
D/MENU Function Key and 7/Controller C a l. P o in ts 5 C a lib r a tio n
R e s u lts
2 5 .0 9 .2 0 0 0 1 0 :2 0
IN S T R U M E N T

Settings) to select the device and start the IN P U T 0 .0 0 ... 1 0 0 .0 0 C


1 1 2 -T T -0 0 3 .1
In ta k e T e m p e r a tu r e o n F e e d 1

communication. M e th o d
S e n s o r T y p e
C o n tr o lle d
P t1 0 0 a 3 8 5 T r a n s fe r F u n c tio n L in e a r
C a l. P o in ts 5
O U T P U T 4 .0 0 0 0 ... 2 0 .0 0 0 0 m A
To open the calibration windows, select the M e th o d M e a s u re d
C o n tr o lle r
S e ttin g s IN P U T 0 .0 0 ... 1 0 0 .0 0 C
C/Calibrate Function Key. B a c k E d it C a lib r a te
M e th o d
C lo s e
M E N U
M S E e N n Us o r T y p e
S im u la te d
P t1 0 0 a 3 8 5

Controller Settings O U T P U T
M e th o d
4 .0 0 0 0 ... 2 0 .0 0 0 0 m A
M e a s u re d

B a c k E d it C a lib r a te M E N U

Opening calibration
windows.
62

Calibration Windows
and Starting Calibration

The calibration screen is divided into three In p u t In p u t


2 5 .0 9 .2 0 0 0 1 0 :4 6
R T D T e m p e r a tu r e [E T : S im u l]

windows as shown in the adjacent picture. S ig n a l


W in d o w
0 .0 0
P t1 0 0 a 3 8 5
C (IT S 9 0 )

O u tp u t C u rre n t [E : M e a s ]
Check the connections before selecting B/Start O u tp u t
S ig n a l 3 .9 7 6 2 m A

Function Key to actually start the calibration run. W in d o w


E rro r
1 .0 0 -0 .1 5 % o f s p a n
E rro r +

G ra p h
0

-
0 % 5 0 % 1 0 0 %
B a c k S ta rt A d ju s t M E N U

Note.
Calibration windows
Checking connections can easily be done. Just
change the instruments input signal and see
what is shown in the three calibration windows.
63

Calibration Under Way

During calibration, you can pause the calibration 2 5 .0 9 .2 0 0 0 1 0 :5 2


In p u t R T D T e m p e r a tu r e [E T : S im u l]
at any time using the A/Pause Function Key. P t1 0 0 a 3 8 5
7 5 .1 0 C (IT S 9 0 )

See the lower part of the adjacent picture to see O u tp u t C u rre n t [E : M e a s ]

the available options when the calibration is 1 5 .9 8 3 9 m A

paused. E rro r
1 .0 0 -0 .2 0 % o f s p a n
+

The B/Accept Readings Function Key is


available only when doing manual calibration. 0

In automatic calibration it is replaced by B/Force


Accept Function Key. It allows you to accept
-
0 % 5 0 % 1 0 0 %
A c c e p t
points even if the output signal of the instrument
P a u s e
R e a d in g s

is so unstable that MC5 cannot automatically


snatch the input/output values. C o n tin u e
R e je c t S k ip U n d o
C a lib r a tio n P o in t P o in t

Function Keys during calibration


64

End of Calibration

When all calibration points have been accepted 2 5 .0 9 .2 0 0 0 1 0 :5 2


In p u t R T D T e m p e r a tu r e [E T : S im u l]
(or skipped), the available options are: P t1 0 0 a 3 8 5
0 .0 0 C (IT S 9 0 )
Saving the results or
rejecting them. O u tp u t C u rre n t [E : M e a s ]

4 .0 2 9 3 m A

E rro r
1 .0 0 0 .1 8 % o f s p a n
+

-
0 % 5 0 % 1 0 0 %
R e je c t
S a v e
C a lib r a tio n

End of Calibration
65

Saving the Calibration Results

MC5 allows you to enter some environmental 2 5 .0 9 .2 0 0 0 1 0 :5 5


E N V IR O N M E N T A L D A T A
data before saving the results.
D e v ic e T e m p e ra tu r e 2 1 .0 C
E n v ir o n m e n t T e m p . 2 1 .0 C

Some of the temperature data may be pre- In p


O u
u t M
tp u t
o d u le
M o d u
T e
le T
m p
e m
.
p .
3 1 .6
2 9 .6
C
C
entered. This depends on what modules where H u m id ity 3 0 %
used during calibration. C a lib r a te d b y :
A n d y H a n d y m a n

N o te s :
J u s t T e s tin g

S a v e S a v e S a v e
B a c k
A s F o u n d A s L e ft A s B o th

Entering environmental data and


saving the results
66

MC5 Training Calibration, Part 3


Calibration Results
67

Opening the Calibration


Results Window

You can open the calibration result windows from 1 4 .0 6 .2 0 0 0 8 :0 6


I N S T R U M E N TV i e w

two places: 1 1 2 -T T -0 0 3 .1 In s tr u m e n t
D e ta ils
In ta k e T e m p e r a tu r e o n F e e d 1
M a k e a C o p y
T r a n s fe r F u n c tio n L in e a r
When viewing the instrument overview window, C a l. P o in ts 5
V ie w
C a lib r a tio n
R e s u lts

press D/Menu and 3/View Calibration Results IN P U T 0 . 0 0 . . . 1 0 0 2 . 9 0 . 00 5 . C2 0 0 2 1 6 : 0 1


(see upper adjacent picture). M e th o d S i m u lI an tp e u D d t e l e t e R T D T e m p e r a t u rS e t a [ Er t T
P H t 1A 0 R 0 T
: S im u l]
a 3 8 5
P t 1 0 0 a 3 I8 n 5 s t r
0 .0 0
S e n s o r T y p e u m e n t
A C d ( Ij Tu s S t 9 m 0 e) n t
P re s c a n
O u tp u t
When viewing the calibration windows, and you
4 .0 0 0 0 ... 2 0 .0 0 0 0 m A [O n ]
M e th o d M e a s O u r u e t dp u t C u rre n t [E : M e a s ]
C o n tr o lle r
3 .9 7 6 2
haven't started a calibration run, press D/Menu B a c k E d it C a lib r a te
S e ttin g s
C lo s e
M E N U
m A
V ie w

and 2/View Calibration Results (see lower E rro r


1 .0 0
+
M E N U C a lib r a tio n
- 0 . 1 5 %R e o s f u s l t p s a n

adjacent picture).
0

Z e ro
- P re s s u re
0 % 5 0 % M o d u le 1 0 0 %
C lo s e
B a c k S ta rt A d ju s t M E N U
M E N U

Opening the Calibration Results window


68

Calibration Result Table

The results are shown starting from the table 0 3 .1 0 .2 0 0 0 1 5 :2 2


2 5 .0 9 .2 0 0 0 1 0 :5 2 - A s L e ft - P a s s e d
with numeric results. In p u t O u tp u t E rro r

[C ] [m A ] [% ]

The table may include more rows than what can 0 .0 0


2 5 .0 0
3 .9 9 3 0
7 .9 8 9 0
-0
-0
.0
.0
4 4
6 9
be displayed. Use the vertical cursor keys to 5 0 .0 0
7 5 .0 0
1 1 .9 8 3 4
1 5 .9 7 7 7
-0
-0
.1
.1
0 4
3 9
scroll through all rows. 1 0 0 .0 0 1 9 .9 7 2 9 -0 .1 7 0
7 5 .0 0 1 5 .9 8 7 5 -0 .0 7 8
5 0 .0 0 1 1 .9 9 2 5 -0 .0 4 7

Use the B/Next Page, to change from one result 2 5 .0 0


0 .0 0
7 .9 9 7 9
4 .0 0 3 2
-0
0
.0
.0
1 3
2 0

page to another.
1 2

N e x t
B a c k M E N U
P a g e

Calibration result table


69

Calibration Result Graph

The graph page shows a larger graph than what 2 5 .0 1 .2 0 0 5 1 0 :3 8


O u tp u t E rro r [% o f s p a n ]
was seen during calibration. 0 .4 0
+

-
0 % 5 0 % 1 0 0 %
M a x im u m E rro r -0 .1 7 0
M a x im u m H y s te r e s is 0 .0 6 4
M a x im u m U n lin e a r ity 0 .0 4 4
S p a n E rro r -0 .1 2 6
Z e ro E rro r -0 .0 4 4
N e x t
B a c k M E N U
P a g e

Calibration result graph


70

Environmental Data

The environmental data page is the same as 2 5 .0 1 .2 0 0 5 8 :4 8


E N V IR O N M E N T A L D A T A
shown when saving calibration results, but now
the data cannot be edited. D e
E n
v ic e
v ir o n
T e m p
m e n t
e ra
T e
tu r
m p
e
.
2 1 .0
2 1 .0
C
C
In p u t M o d u le T e m p . 3 1 .6 C
O u tp u t M o d u le T e m p . 2 9 .6 C

H u m id ity 3 0 %

C a lib r a te d b y :
A n d y H a n d y m a n

N o te s :
J u s t T e s tin g

N e x t
B a c k M E N U
P a g e

Environmental data
71

Calibration History

To choose which which calibration run is shown, 2 5 .0 1 .2 0 0 5 1 0 :3 8


O u tp u t E rro r [% o f s p a n ]
press D/MENU, 1/Calibration History. +
0 .4 0 C a lib r a tio n
H is to r y

P r in t

Select the calibration run to be viewed from P ro to c o l

displayed list. 0

R e m o v e
R e s u lt

R e m o v e
-
0 % 5 0 % A ll R e s u lt 1s 0 0 %
M a x im u m E rro r -0 .1 7 0
M a x im u m H y s te r e s is 0 .0 6 4
M a x im u m U n lin e a r ity 0 .0 4 4
S p a n E rro r -0 .1 2 6
Z e ro E rro r -0 .0 4 4
N e x t
B a c k M C El o N s U e
P a g e M E N U

Calibration History
72

Deleting Calibration Results

When viewing Calibration results, press 0 .4 0


2 5 .0 1 .2 0 0 5 1 0 :3 8
O u tp u t E rro r [% o f s p a n ]
C a lib r a tio n
D/MENU. +

0 .4 0
H i 2 s 5 t o. 0 r 1 y . 2 0 0 5 1 0 : 3 8
O u tp u t E rro r [% o f s p a n ]
C a lib r a tio n
+ P r in t H is to r y
P ro to c o l

To remove the current calibration run, press


P r in t
0 P ro to c o l

3/Remove Results. 0 R e m o v e
R e s u lt
R e m o v e
R e m o v e
- R e s u lt
A l l R e s u l t 1s 0 0 %
To remove all calibration results (for the current
0 % 5 0 %
M a x im u m E rro r -0 .1 7 0 R e m o v e
-
M a x im u m H y s te r e s is 0 .0 6 4 A ll R e s u l t 1s

instrument), press 4/Remove All Results.


0 % 5 0 % 0 0 %
M a x im u m U n lin e a r ity M a x i m u m 0E . r 0 r o 4 r4 -0 .1 7 0
S p a n E rro r M a x i m u m - 0H . y 1 s 2 t 6 e r e s i s 0 .0 6 4
Z e ro E rro r M a x i m u m - 0U . n 0 l 4 i n 4 e a r i t y 0 .0 4 4
N e x t S p a n E r r o r C lo s e -0 .1 2 6
B a c k Z e r o E r r o r MM EE NN UU
P a g e -0 .0 4 4
N e x t
B a c k M C El o N s U e
P a g e M E N U

Deleting calibration results


73

MC5 Training Advanced Operation, Part 1


Display Modes
74

Changing the Display Mode

By default, MC5 displays all measurement, 2 7 .0 1 .2 0 0 5 1 3 :3 4


1 P re s s u re
generation and simulation data in engineering P 1 : IN T 2 C /-1 .0 ... 2 b a r g
Q u a n tit
[P re s s u
y
re ]

units. In Basic Mode you can also select some 0 .7 0 8 3 g a u g e


bF au r n c t / P o rt
[P 1 : IN T 2 C ]
special Display Modes. E n g . U n it D is p la y M o d e
S c a lin g [E n g . U n its ]
P e rc e n ta g e
To change the Display Mode, select the Window 2
E r r o r V a lu e
C u rre n t
U n it
[b a r]
setup menu Commands: E : C u rre n t M e a s u re m e n t
H A R T
D/Menu (and B/Window 2 Setup, if needed) 0 .4 3 7 4 m A
P re s s u re
3/Display Mode. T y p e
[g a u g e ]
Z e ro
P re s s u re
M o d u le
C W a il ni b d r o a wt i o 1 n W in d o w 2 O th e rs M C lE o N s U e
MS e o t d u e p S e tu p M E N U
75

Scaling, Using
Existing Quantities and Units

In this case a Quantity already existing in MC5 2 6 .0 4 .2 0 0 2 1 4 :0 2

and a related unit is used when scaling. Q u a n tity


S C A L IN G
P re s s u re
P o rt P 1 :IN T 2 C

Enter both the measured and the scaled range M e a s u re d R a n g e


0 % 0 .0 0 0 k P a

as well as the transfer function. 1 0 0 % 1 0 0 .0 0 0 k P a

S c a le d M o d e C u rre n t
S c a le d U n it m A
S c a le d R a n g e
0 % 4 .0 0 0
When scaling is active in a Basic Mode window, 1 0 0 % 2 0 .0 0 0

the window contains the text Scaling.


The true value is displayed in Extra Info row. T r a n s fe r F u n c tio n L in e a r

C a n c e l F ie ld S E t d o i pt S ta rt
R a m p in g

Configuring Scaling Display Mode,


using existing Quantities and units
76

Scaling, Using
Custom Units

Select Custom Units as the value for the 3 0 .0 5 .2 0 0 2 1 0 :0 9

Scaled Mode field. Then any unit may be Q u a n tity


S C A L IN G
P re s s u re
manually entered as the Scaled Unit. P o rt
M e a s u re d R a n g e
P 1 :IN T 2 C

0 % 0 .0 0 0 k P a

Otherwise the configuration is similar to the 1 0 0 % 1 0 0 .0 0 0 k P a

situation when existing Quantities and units are S c a le d M o d e C u s to m U n its


used. S c a le d U n it G a llo n s /m in
S c a le d R a n g e
0 % 1 .0 0
1 0 0 % 5 .0 0
D e c im a ls 3

T r a n s fe r F u n c tio n L in e a r

C a n c e l F ie ld S E t d o i pt S ta rt
R a m p in g

Configuring Scaling Display Mode,


using custom units
77

Percentage

In Percentage Display Mode the 2 6 .0 4 .2 0 0 2 1 4 :2 1

measured/generated/simulated values are Q u a n tity


P E R C E N T A G E
P re s s u re
displayed as percentage of a user defined range. P o rt
M e a s u re d R a n g e
P 1 :IN T 2 C

0 % 0 .0 0 0 k P a

The true value is displayed in the Extra Info row. 1 0 0 % 1 0 0 .0 0 0 k P a

C a n c e l F ie ld S E t d o i pt S ta rt
R a m p in g

Configuring Percentage Display


Mode
78

Error Value

In Error Display Mode one of Basic Modes 2 6 .0 4 .2 0 0 2 1 4 :2 9

windows corresponds with the instrument's input W in d o w


E R R O R D IS P L A Y
1
signal and the other with the instrument's output. IN P U T
P o rt
P re s s u re
P 1 :IN T 2 C
M e a s u re d R a n g e

The window from which Error Display Mode is 0 %


1 0 0 %
0 .0 0 0
1 0 0 .0 0 0
k P a
k P a

invoked is considered as the instruments output W in d o w 2

signal and the other window as the input. O U T P U T


P o rt
C u rre n t
E : I(m e a s )
M e a s u re d R a n g e
0 % 4 .0 0 0 m A
The input value is considered as the correct 1 0 0 % 2 0 .0 0 0 m A

value and the output value should follow the E r r o r C a lc . M e th o d % o f s p a n

input value, while taking the Transfer Function T r a n s fe r F u n c tio n L in e a r


S E t d o i pt
into account. C a n c e l F ie ld
R a m p in g
S ta rt

Configuring Error Display Mode


The true output value is displayed in the Extra
Info row.
79

MC5 Training Advanced Operation, Part 2


Special Measurements
80

Starting Special Measurements

Special measurements are extra utilities that can 1


2 7 .0 1 .2 0 0 5
P re s s u re
1 3 :3 4
Q u a n tit y
be activated to perform a special function P 1 : IN T 2 C /-1 .0 ... 2 b a r g

0 .7 0 8 3 1
[P re s s u
2 7 .0 1 .2 0 0 5
g a u g e
P r e s Fb s ua u r n r c e t / P
re ]
1 3 :3 4
S p e c ia l N o n e

alongside the normal measurement.


o rt
P 1 : I N T [2 P C 1 / :- 1 I N. 0 T . 2. . C 2 ] b a r g M e a s u re m e n ts M in im u m R e a d in g
[N o n e ] M a x im u m R e a d in g
0 .7 0 8 3
D is p la y M o d e
[E n g . U n its ]
g a u g e
Ab a l a r r m M in a n d M a x R e a d in g s
[--(--)] R a te o f C h a n g e ...
F ilte r a n d R e s o lu tio n ...

All special measurements are started similarly: 2 C u rre n t


E : C u rre n t M e a s u re m e n t
U n it
[b a r]
S e c o n d P o rt
[P 2 : E X T 1 0 0 ]
D e v ia tio n M e
D iffe r e n c e M
R e d u n d a n t M
a s ...
e a s ...
e a s ...
H A R T F u n c tio n a n d
0 . 4 3 2E 7 : C 4 C u u r r r e m r n e A t n M t
Select D/Menu (followed by B/Window 2
P o r t In fo
P r e s se u a rs e u r e m e n t
T y p e

Setup if needed). 0 e. ] 4 3 7 4
[g a u g
Z e ro
m A
P re s s u re E x tr a In fo
M o d u le
C W a il ni b d r o a wt i o 1 n W in d o w 2 O th e rs M C El o N s U e

As the special measurement setting command


MS e o t d u e p S e tu p M E N U

C W a il ni b d r o a wt i o 1 n W in d o w 2 O th e rs M C El o N s U e

is located in the window setup's submenu,


MS e o t d u e p S e tu p M E N U

select the numerical key to open the Window setup menus


submenu.

Then select 1/Special Measurements and the


type of special measurement from the provided
pop-up list
81

Minimum and Maximum


Readings

Minimum Reading, Maximum Reading and 1 4 .0 1 .2 0 0 5 8 :3 6


1 P re s s u re
Min and Max Readings do not require any P 1 : IN T 2 C /-1 .0 ... 2 b a r g

additional information. They start immediately 0 .7 3 2 5 g a u g e


b a r

after they are activated.


M in / M a x 0 .0 0 0 0 1 .1 0 2 4

2 N o n e

-------

C a lib r a tio n
M E N U
M o d e

Minimum and maximum readings


displayed in a Basic Mode window
82

Rate of Change

Before the rate of change special measurement 2 2 .0 9 .2 0 0 0 1 3 :2 2


R A T E O F C H A N G E M E A S U R E M E N T
is started, MC5 prompts for the rate of change
Q u a n tity
unit which can be selected from the following
T /C -te m p e r a tu r e
P o rt E T : T C i(m e a s )

options: U n it C

R a te o f C h a n g e U n it 1 /s

1/s, 1/ms, 1/h and 1/min.

C a n c e l E d it S ta rt
83

Special Filtering
and Resolution

Select the Time Constant setting and how many 1 9 .0 5 .2 0 0 0 1 3 :5 5


S P E C IA L F IL T E R IN G A N D R E S O L U T IO N
decimals should be cut off from the actual
resolution of the current measurement. Q u a n tity
P o rt
T /C -te m p e r a tu r e
E T : T C i(m e a s )
U n it C

The filter is a 1st order digital filter. T im e C o n s ta n t 5 s

D e c im a ls -2

T h e s e ttin g - 1 , fo r in s ta n c e , m e a n s o n e
le s s d e c im a ls .

C a n c e l E d it S ta rt
84

Deviation Measurement
In deviation measurement the reading of the 2 2 .0 9 .2 0 0 0 1 3 :3 2
D E V IA T IO N M E A S U R E M E N T
main port is compared to a reference value. The
Q u a n tity
reference value is subtracted from the reading of P o rt
T /C -te m p e r a tu r e
E T : T C i(m e a )

the main port.


M e a s u r e d e v ia tio n fr o m th e v a lu e

2 2 .0 0 0 0 C

C a n c e l E d it S ta rt
85

Difference Measurement
Difference measurement uses both the main 1
1 4 .0 1 .2 0 0 5
P re s s u re
8 :3 6
Q u a n tit y

measurement and the Second Port found in the P 1 : IN T 2 C /-1 .0 ... 2 b a r g

0 .7 0 8 3
[P re s s u
g a u g e
Fb ua r n c t / P
re ]

o rt
S p e c ia l

windows setup menu. [P 1 : IN T

D is p la y M o d e
2 C ]
M e a s u re m e n ts
[N o n e ]
g a u g e
Ab a l a r r m C o n tr o lle r
[E n g . U n its ] [--(--)] S e ttin g s
g a u g e

In difference measurement, the difference 2 C u rre n t


E : C u rre n t M e a s u re m e n t
U n it
[b a r]
S e c o n d P o rt
[P 2 : E X T 6 C ]
Sb a e
[N
rn s o r S u p p l y
o n e ]

between the reading of the main port and the 0 .4 3 7 4


H A R T
m A
F u n c tio n a n d
P o r t In fo

second port is calculated. The reading of the P re s s


T y p e
[g a u g
u re

e ] m A

second port is subtracted from the reading of the Z e ro


P re s s
M o d u
u re
le
E x tr a In fo
m A
main port. C W a il ni b d r o a wt i o 1 n
MS e o t d u e p
W in d o w
S e tu p
2 O th e rs M C El o N s U e
M E N U

C W a il ni b d r o a wt i o 1 n W in d o w 2 O th e rs M C lE o N s U e
MS e o t d u e p S e tu p M E N U

C W a il ni b d r o a wt i o 1 n W in d o w 2 O th e rs M C lE o N s U e
MS e o t d u e p S e tu p M E N U

Window setup menus


86

Redundant Measurement
Redundant measurement uses both the main 1
1 4 .0 1 .2 0 0 5
P re s s u re
8 :3 6
Q u a n tit y

measurement and the Second Port found in the P 1 : IN T 2 C /-1 .0 ... 2 b a r g

0 .7 0 8 3
[P re s s u
g a u g e
Fb ua r n c t / P
re ]

o rt
S p e c ia l

windows setup menu (see adjacent picture). [P 1 : IN T

D is p la y M o d e
2 C ]
M e a s u re m e n ts
[N o n e ]
g a u g e
Ab a l a r r m C o n tr o lle r
[E n g . U n its ] [--(--)] S e ttin g s
g a u g e

In redundant measurement, the measurements 2 C u rre n t


E : C u rre n t M e a s u re m e n t
U n it
[b a r]
S e c o n d P o rt
[P 2 : E X T 6 C ]
Sb a e
[N
rn s o r S u p p l y
o n e ]

of the main port and the second port are 0 .4 3 7 4


H A R T
m A
F u n c tio n a n d
P o r t In fo

compared with each other. If the readings differ P re s s


T y p e
[g a u g
u re

e ] m A

more than the entered limit value, MC5 gives an Z e ro


P re s s
M o d u
u re
le
E x tr a In fo
m A
audible alarm. C W a il ni b d r o a wt i o 1 n
MS e o t d u e p
W in d o w
S e tu p
2 O th e rs M C El o N s U e
M E N U

C W a il ni b d r o a wt i o 1 n W in d o w 2 O th e rs M C lE o N s U e
MS e o t d u e p S e tu p M E N U

C W a il ni b d r o a wt i o 1 n W in d o w 2 O th e rs M C lE o N s U e
MS e o t d u e p S e tu p M E N U

Window setup menus


87

MC5 Training Advanced Operation, Part 3


Alarms
88

Opening the
Alarms Window

Each main measurement in a window may have 1


0 7 .0 5 .2 0 0 2 1 2 :5 3
P re s s u re A Q l u a a r nm t i t< y
alarm limits settings. MC5 supports "higher P 1 : I N T 2 C / - 1 . 0 0 0 0 0 0 . . . 2 [b P a r r e s s u re ]

0 .7 0 8 3 4 (g a u g e ) A S lp a e r c m i a < l
than", "lower than", "high rate" and "low rate"
F m u b n ac rt / P o r t M e a s u re m e n t
[P 1 : IN T 2 C ]
[N o n e ]
(g a u g e )

alarms.
D i s p 0 l . a 5 y Mb a o r d e A m l ab r am r A C l o a n r mt r o < l l e r
[E n g . U n its ] [--(--)]
S e ttin g s
U n it S e c o0 . n 5 d b p a o r t (g a u g e )
S m e b n a s ro r S u p p l y
2 C u rre n t [b a r] [P 2 : E X T 1 0 0 ] [N o n e ]

Alarm settings can be found in the windows E : C u rre n t M e a s u re m e n t

0 .4 3 7 4 H A R T
F u n c tio n a n d
P o r t In fo
0 .5 b a r

setup menu (see adjacent picture). P re s s


T y p e
[g a u g
u re

e ]
S e c o n d
P o r t In fo
Z e ro
P re s s u re E x tr a In fo
M o d u le
W in d o w 1 W in d o w 2 O th e rs C lo s e
S e tu p S e tu p M o d e M E N U

W in d o w 1 W in d o w 2 O th e rs C lo s e
S e tu p S e tu p M o d e M E N U

W in d o w 1 W in d o w 2 O th e rs C lo s e
S e tu p S e tu p M o d e M E N U
89

Alarm Settings

The Alarms activated field makes it possible to 2 2 .0 9 .2 0 0 0 1 5 :2 1


A L A R M S E T T IN G S
deactivate the alarm limits without clearing the
Q u a n tity
limits.
R T D -te m p e ra tu re
P o rt E T : R T D (m e a )

A la r m s a c tiv a te d Y e s
Individual alarm limits may be disabled by
clearing the alarm limit using the B/Clear Limit A la r m , w h e n th e M e a s u r e m e n t
< 2 4
a n d /o r C
Function Key. > 1 9

A la r m , w h e n th e R a te o f C h a n g e
< ? ? ? ? ? ? ?
a n d /o r 1 /h
> 1

C le a r
C a n c e l E d it O K
L im it

Configuring alarms
90

Alarm Symbols

The alarm settings can be seen as symbols in 1 8 .0 1 .2 0 0 5 1 3 :5 0


1 P re s s u re A la r m < > (- -)
the Basic Mode window and in the window setup P 1 : IN T 2 C /-1 .0 ... 2 b a r g

menu as follows: 0 .7 4 0 9 g a u g e
b a r

Symbol Definition
[- - ( - - )] No limit settings
[< - ( - - )] Low limit is set 2 C u rre n t
[- > ( - - )] High limit is set E : C u rre n t M e a s u re m e n t

[< > ( - - )] Low and high limits are set 0 .4 3 7 4 m A


[> < ( - - )] Low and high limits are set. Low > High
[- - ( < - )] Low rate limit is set
[- - ( - > )] High rate limit is set
[- - ( < > )] Low and high rate limits are set C a lib r a tio n
M E N U
[- - ( > < )] Low and high rate limits are set. Low > High M o d e

[< > ( < > )] All limits are set.

An active alarm is underlined in the Basic Mode


window.
91

Acknowledging alarms
When an alarm limit is exceeded, MC5 emits an
audible alarm.

The alarm can be acknowledged from all other


keys except the following: , and
or any key combinations including one or more of
the previously listed keys.

The active alarm limit is underlined in the


measurement window for as long as the alarm
limit is exceeded.
92

MC5 Training Advanced Operation, Part 4


Special Generations
93

Opening the Step or Ramp


Configuration Window
MC5 includes Stepping and Ramping as 1
1 2 .0 4 .2 0 0 2
V o lta g e
1 2 :5 1

standard utilities for creating automatically


Q u a n tity
E T : V o lta g e G e n e r a tio n
1 2 . 0 [ V 4 . o 2 l0 t 0a 2 g e 1 ] 2 : 5 1

? ? ? ? ? 1E? T : ? V V o o l l tt FV aa ugg ene c G t i e o n n e / rP a o t i r o t n S p e c ia l

alternating output signals (special generations). [E T : V (g e n )]


M e a s u re m e n ts
[ 1 N 2 o. 0 n 4 e . 2 ] 0 0 2 1 2 : 5 2
? D ? i s ? p l a ? y M ? o d ? e ?1E V AV o l l a t a r m g e
T : V [ - o - (l t - a - ) g ] e G e n e r a t i o n S te p
[E n g . U n it]
M e a s u re m e n t, V -------

Step and Ramp configuration can be found in the U n it


[m A ]
? ? ? ? ? ? ?
S te p , R a m p RV a m p
2 S w itc h M e a s u re m e n t, V -------

windows setup menu (see adjacent picture).


E : S w itc h S e n s e
F u n c tio n a n d
H A R T
O p 2E e : S n S w w i t ic t h c h S e n s e
P o r t In fo
M e a s u re m e n t, V -------
S e c o n d P o rt

O p e n e d @ V
O p e n 2 S I nw f i o t c h
- - - - - - - E : S w itc h S e n s e
C lo s e d @ V ------- E x tr a In fo

W in d o w 1 W in d o w 2O p e n e d @ V C lo s e
O - - - -p - - - e n
S e tu p S e tu p C l o s O e dt h @ e r s V M E N U -------

Note. W in d o w
S e tu p
1 W in d o w
S e tu p
2 O O p t e h n e er s d @ V C l o s e
C lo s e d @ V M E N U
-------
-------

To be able to start the definition of a special B a c k


C lo s e
M E N U

generation, a suitable generation/simulation field


has to be active in the corresponding Basic
Mode Window.
94

Stepping, Example 1
Note that because of the Starting point setting, 2 2 .0 9 .2 0 0 0 1 4 :2 2

the first repeat is limited to a small starting step


S T E P P IN G

and only two full steps.


Q u a n tity C u rre n t
P o rt E T :I(g e n )

S te p p in g M o d e U p

Generally: if the stepping settings limit the first S te


R e
p T
p e a
im e
ts
1 .0 0 s

repeat to only a fraction of a full repeat it is still 0 =


S ta
c o
r tin
n tin u o u s
g P o in t
3
F ro m p re s e n t

calculated as one repeat. S te p S iz e 1 .0 0 m A


N u m b e r o f S te p s 4
D iv is io n o f S te p s L in e a r

R a n g e 0 % 8 .0 0 0 0
1 0 0 % 1 2 .0 0 0 0 m A

C a n c e l F ie ld S E t d o i pt S ta rt
R a m p in g

C u rre n t, m A

1 2

1 1

1 0

8
-1 0 1 5 1 0 1 4 T im e , s

S ta rt
95

Stepping, Example 2
You may enter either the Step Size or the 2 2 .0 9 .2 0 0 0 1 4 :3 9

Number of Steps. Because these values


S T E P P IN G

depend on each other, the value of the other field


Q u a n tity C u rre n t
P o rt E T :I(g e n )

is calculated whenever you make any changes. S te p p in g M o d e U p /D o w n

S te p T im e 0 .5 0 s
R e p e a ts
You can only enter an integer as the Number of 0 =
S ta
c o
r tin
n tin u o u s
g P o in t
0
F ro m 0 %

Steps value. If you enter a Step Size that results S te p S iz e 1 .5 0 m A

in a non-integer value for the Number of Steps N u m b e r o f S te p s


D iv is io n o f S te p s
-------
L in e a r

field, MC5 displays a dashed line instead of a R a n g e 0 % 8 .0 0 0 0

value for the Number of Steps field. 1 0 0 % 1 2 .0 0 0 0

S E t d o i pt
m A

C a n c e l F ie ld S ta rt
R a m p in g

This indicates that the size of the step that C u rre n t, m A

reaches the 100 % range limit is smaller than the 1 2

defined step size.


1 1

1 0

8
-1 0 1 5 1 0 1 4 T im e , s

S ta rt
96

Ramping, Example 1
You can enter a Rise Time and a Fall Time as 2 2 .0 9 .2 0 0 0 1 0 :4 3

well as two Wait times for ramping. R A M P IN G

Q u a n tity C u rre n t
P o rt E T :I(g e n )

Ramping always starts from the ranges 0 %


value. W a it
R is e
in 0 %
T im e 1
2
s
s

W a it in 1 0 0 % 1 s
F a ll T im e 2 s

R e p e a ts
0 = c o n tin u o u s 2

R a n g e 0 % 8 .0 0 0 0
1 0 0 % 1 6 .0 0 0 0 m A

C a n c e l F ie ld S E t d o i pt S ta rt
R a m p in g

C u rre n t, m A

1 6

1 4

1 2

1 0

8
-1 0 1 5 1 0 1 4 T im e , s

S ta rt
97

Ramping, Example 2
When Repeats field is set to zero, ramping 2 2 .0 9 .2 0 0 0 1 0 :5 2

continues until C/Stop Ramping Function Key R A M P IN G

is pressed. Q u a n tity
P o rt
C u rre n t
E T :I(g e n )

W a it in 0 % 0 s
R is e T im e 5 s
W a it in 1 0 0 % 0 s
F a ll T im e 0 s

R e p e a ts
0 = c o n tin u o u s 0

R a n g e 0 % 8 .0 0 0 0
1 0 0 % 1 6 .0 0 0 0 m A

C a n c e l F ie ld S E t d o i pt S ta rt
R a m p in g

C u rre n t, m A

1 6

1 4

1 2

1 0

8
-1 0 1 5 1 0 1 4 T im e , s

S ta rt
98

MC5 Training Advanced Operation, Part 5


Transmitter/Switch Simulation
99

Transmitter/Switch
Simulation
To start a Transmitter or a Switch simulation, 1 8 .0 4 .2 0 0 2 1 5 :1 5
1
configure Window 1 to measure a signal P re s s u re
P 2 : IN T 2 0 C /- 1 0 0 ... 2 0 6 8 k P Sa e gt t i n g s

(transmitter/switch input) and Window 2 to either 0 .0 2 g


kM
a u g e
P a ai n t e n a n c e

generate or simulate a signal (transmitter/switch H A R T


output), then select C o m m u n ic a tio n
S e ttin g s

D/Menu, 2 C u rre n t
C/Others and E : C u r r e n t S o u r c e /S in k
D a ta
6/Transmitter/Switch Simulation. 4 .0 0 0 0 L o g g in g
m A
T r a n s m itte r /
T r a n s m itte r F u n c tio n
S w itc h F u n c tio n S w itc h
S im u la tio n
M e a s u re m e n t m A 4 .0 0 0 0

W in d o w 1 W in d o w 2 C lo s e
E n d P a u s e O th e rs M E N U
S e tu p S e tu p M E N U

Note.
Configure the Basic Mode windows accordingly
before starting the Transmitter/Switch
Simulation.
100

Configuring
Transmitter Simulation
Enter the input and output spans, the saturation 2 6 .0 4 .2 0 0 2 1 4 :3 0

limits and also the transfer function. T R A N S M IT T E R S IM U L A T IO N


IN P U T P re s s u re
P o rt P 1 :IN T 2 C

If the saturation limits differ from the range limits, 0 %


1 0 0 %
0 .0 0 0
1 0 0 .0 0 0
k P a
k P a
MC5 extrapolates the output value based on the
input value and the Transfer Function until the O U T P U T
P o rt
C u rre
E : I(c
n t
o n tr o l)

saturation limit is reached. Then if the input 0 %


1 0 0 %
4 .0 0 0
2 0 .0 0
0
0
m A
m A

signal drifts farther from the input range, the S a tu r a tio n L im it O u tp u t V a lu e s


output value stays at the saturation limit. M in im u m 3 .5 0 0 0 m A
M a x im u m 2 3 .0 0 0 m A

T r a n s fe r F u n c tio n L in e a r

C a n c e l F ie ld S E t d o i pt S ta rt
R a m p in g
101

Transmitter
Simulation In Use
The adjacent picture shows the Basic Mode 3 0 .0 5 .2 0 0 2 1 0 :3 1

window while Transmitter Simulation is active. 1 IN P U T


P 2 : IN T 2 0 C / (0 .0 0 ... 1 0 0 .0 0 k P a g )

0 .0 6 g a u g e
k P a
The second row of both windows display the
active port and the input/output range of the
simulated transmitter. T R A N S M IT T E R S IM U L A T IO N
2 O U T P U T
E : I( c o n tr o l) / ( 4 .0 0 0 0 ... 2 0 .0 0 0 m A )
When pausing the Transmitter Simulation, you 4 .0 0 0 0 m A
can change the generation/simulation signal as
in Basic Mode. M e a s u re m e n t m A 4 .0 0 0 0

E n d P a u s e M E N U
102

Configuring
Switch Simulation
Enter the actuating point and the deactuating 2 6 .0 4 .2 0 0 2 1 4 :3 8

point of the input signal. S W IT C H S IM U L A T IO N

IN P U T P re s s u re
P o rt P 1 :IN T 2 C
Also enter the output signal levels for an N o m in a l P o in ts
A c tu a tin g 5 0 .0 0 0 k P a
actuated output and a deactuated output. D e a c tu a tin g 4 0 .0 0 0 k P a

O U T P U T V o lta g e
P o rt E T : V (g e n )
S w itc h L im its
A c tu a te d 5 .0 0 0 0 V
D e a c tu a te d 0 .5 0 0 0 V

C a n c e l F ie ld S E t d o i pt S ta rt
R a m p in g
103

Switch
Simulation In Use
The adjacent picture shows the Basic Mode 3 0 .0 5 .2 0 0 2 1 0 :3 2

window while Switch Simulation is active. 1 IN P U T


P 2 : IN T 2 0 C / (5 0 .0 0 /4 0 .0 0 k P a g )

5 2 .0 3 4 g a u g e
k P a
The second row of the upper window displays
the active port and the actuating and deactuating
points. S W IT C H S IM U L A T IO N
2 O U T P U T
E T : V (g e n )/ (5 .0 0 0 /0 .5 0 0 V )
The second row of the lower window displays the 5 .0 0 1 0
actuated and deactuated output of the switch.
M e a s u re m e n t V 5 .0 0 1 1

E n d P a u s e M E N U
104

MC5 Training Advanced Operation, Part 6


Data Logging
105

How to Start
Data Logging
The Data Logging utility is a firmware option that 1 9 .0 1 .2 0 0 5 1 0 :4 4

allows you to collect data using MC5. The 1 P re s s u re


P 1 : IN T 2 C /-1 .0 ... 2 b a r g S e ttin g s

collected data may then be viewed, transferred 0 .7 0 8 3 g a u g e


bM a a r i n t e n a n c e
to a personal computer (PC) and printed using a H A R T
utility shipped together with MC5, if the Data C o m m u n ic a tio n
S e ttin g s
Logging option is purchased. 2 C u rre n t
E : C u rre n t M e a s u re m e n t
D a ta
To start configuring the Data Logging (starting 0 .4 3 7 4 L o g g in g
m A

from Basic Mode), press T r a n s m itte r /


S w itc h
S im u la tio n
D/Menu,
C/Others, C W a il ni b d r o a wt i o 1 n W in d o w 2 M C lE o N s U e
5/Data Logging. MS e o t d u e p S e tu p
O th e rs
M E N U
106

Configuring
Data Logging
The upper part of the Data Logging configuration 2 0 .0 8 .2 0 0 2 1 4 :5 7

window is reserved for seven measurement N o n e


D A T A L O G G IN G S E T U P

1
channels and the lower part for general settings. 2 N o n e
3 N o n e
4 N o n e
The measurement channels may be configured 5
6
N o n
N o n
e
e
using the C/Edit Function Key. 7 N o n e

T r ig g e r P e r io d ic

To remove all channel settings, select S a v e M e th o d


In te rv a l [s ] 2
In s ta n t V a lu e

D/MENU, S a m p le s 1 0

1/Clear All Channels. T im e [h m in s ] 0 0 2 0

B a s ic D a ta
E d it M E N U
M o d e L o g g in g
107

Assigning Measurement
Ports to Channels
Select the measurement port first. The other 2 0 .0 8 .2 0 0 2 1 5 :0 1

settings depend on the port setting. D A T A L O G G IN G S E T U P

C h a n n e l 1

To clear a channel, select None from the list of P o rt P 3 :B

available ports. P re s s u re T y p e A b s

U n it b a r

C a n c e l E d it O K
108

Starting
Data Logging
When all channels intended to be used in Data 2 0 .0 8 .2 0 0 2 1 5 :0 2
D A T A L O G G IN G S E T U P
Logging are configured, select B/Data Logging 1 P 3 :B b a r a b s

in the Data Logging configuration window.


2 E T :R T D (m e a ) C P t1 0 0 a 3 8 5
3 E T :T C i( m e a ) C K N iC r /n iA l
4 N o n e
5 N o n e
6 N o n e
A new window opens where all channels are 7 N o n e 2 0 .0 8 .2 0 0 2 1 5 :3 2
D A T A L O G G IN G

shown and the actual logging can be started T r ig g e r


S a v e M e th o d
P e r io d ic
I n s t a n t V a l u e1 1 .0 1 1 3 A b s
b a r

using the C/Start Logging Function Key. In te rv a l [s ]


S a m p le s
6 0
7 2 0 2 2 4 .1 7 P t1 0 0 a 3 8 5
C

T im e [h m in s ] 1 2 0 0
3 2 4 .3 9 K N iC r /N iA l
C

Then, to stop/interrupt the logging, select B a s ic D a ta


4 ----- -- P o s it. k a n tti
k H z

C/Stop Logging.
E d it M E N U
M o d e L o g g in g
5 ----- -- P o s it. k a n tti
k H z

6 ----- -- P o s it. k a n tti


k H z

7 ----- -- P o s it. k a n tti


k H z

C o u n t 6 4 7 3 8 8 0 2 s
S to p
B a c k M E N U
L o g g in g
109

Data Logging
Results
Opening the Log Results window can be done 2 0 .0 8 .2 0 0 2 1 3 :1 9

both from the configuration window and the J u s t T e s tin g - A ll v a lu e s


T im e 1 : In s ta n t 1 : A v e ra g e
Logging window. Select 0 4 .0 9 .2 0 0 0
1 4 :1 9 :3 8
P re s s u re
[b a r]
P re s s u re
[b a r]

D/MENU, 1 4
1 4
:2
:2
0 :
1 :
3 8
3 8
1 .0
1 .0
1 1
1 1
3
3
1 .0
1 .0
1 1
1 1
3
3
2/View Datalog Results. 1 4
1 4
:2
:2
2 :
3 :
3 8
3 8
1 .0
1 .0
1 1
1 1
2
2
1 .0
1 .0
1 1
1 1
3
2
1 4 :2 4 : 3 8 1 .0 1 1 2 1 .0 1 1 2

If you have logged using more than two 1 4


1 4
:2
:2
5 :
6 :
3 8
3 8
1 .0
1 .0
1 1
1 1
1
1
1 .0
1 .0
1 1
1 1
2
1

channels, use the left and right cursor keys to 1 4


1 4
:2
:2
7 :
8 :
3 8
3 8
1 .0
1 .0
1 1
1 1
1
1
1 .0
1 .0
1 1
1 1
1
1

scroll through all the columns. Alternatively: use 1 4


1 4
:2
:3
9 :
0 :
3 8
3 8
1 .0
1 .0
1 1
1 1
0
0
1 .0
1 .0
1 1
1 1
1
0

the numeric keys to quickly jump to a column. 1 4


1 4 :3
:3
1 :
2 :
3 8
3 8 1 .0
1 .0
1 0
1 1 0
9
1 .0
1 .0
1 1
1 0
0
9
1 2

B a c k M E N U
110

Transferring the
Results to a PC
MC5_Datalog.exe displays the Data Logging
results in a table and a graph. The results may
also be exported to, e.g. Excel.

All communication between the PC and MC5 is


initiated from MC5_Datalog.exe.
111

MC5 Training Advanced Operation, Part 7


Generating Signals Using External Devices
112

Connecting to the
External Device
To select an external device, open either 1
0 6 .0 5 .2 0 0 2 1 5 :3 2
P re s s u re A Q l u a a r mn t i t< y

Window 1 or Window 2 setup menu (select P 1 : IN T 2 C / -1 .0 0 0 0 0 0 ...2

0 .7 0 8 3 4
[b P a r r e s
(g a u
F m u b n ac
s u re ]
g e )
rt / P o r t A S lp a e r c m i a < l

D/Menu and B/Window 2 Setup if needed). [P 1 : IN T 2 C ] M e a s u re m e n t


[N o n e ]
(g a u g e )
D i s p 0 l . a 5 y Mb a o r d e A m l ab r am r A C l o a n r mt r o < l l e r
[E n g . U n it] [--(--)] S e ttin g s
U n it S e c 0o . n 5 d b p a o r t (g a u g e )

Make sure the quantity setting of the window


S m e b n a s ro r S u p p l y
2 C u rre n t [b a r] [P 2 : E X T 1 0 0 ] [N o n e ]
E : C u rre n t M e a s u re m e n t
F u n c tio n a n d 0 .5 b a r
corresponds to the external device's quantity 0 .4 3 7 4 H A R T

P re s s u re
P o r t In fo

S e c o n d

requirements. Continue to the third submenu by


T y p e
[g a u g e ] P o r t In fo
Z e ro

pressing the numerical key twice. The


P re s s u re E x tr a In fo
M o d u le
W in d o w 1 W in d o w 2 O th e rs C lo s e

adjacent picture displays all the submenus


S e tu p S e tu p M o d e M E N U

W in d o w 1 W in d o w 2 O th e rs C lo s e

available when the selected Quantity is S e tu p S e tu p M o d e

W in d o w 1 W in d o w
M E N U

2 O th e rs C lo s e

Pressure. S e tu p S e tu p M o d e M E N U

Note.
To be able to utilize devices connected to the
auxiliary instrument interface (AUX) you need to
have the required firmware option.
113

Controller Settings
Select the Controller Type. The list of available 2 0 .0 8 .2 0 0 0 1 5 :0 5
C O N T R O L L E R S E T T IN G S
controllers varies depending on installed options
and the selected quantity. The other settings C o n tr o lle r T y p e
C o n tro l M o d e
P O C 4
S h u t O ff
vary depending on the Controller Type setting. D e v ic e A d d r e s s 0

S e r ia l C o m m u n ic a tio n S e ttin g s
9 6 0 0

C a n c e l E d it O K
114

Controlling the External


Device in Basic Mode
When an external device is configured for use, 2 5 .0 9 .2 0 0 0 9 :4 8

the device name is displayed in the window it 1 P re s s u re


P 2 :IN T 2 0 C /-1 .0 0 0 0 ...2 0 .6 8 4 0 b a r g
was assigned to, e.g. POC4 in the adjacent 0 .0 0 0 3 g a u g e
b a r
picture.
P O C 4 0 .0 0

Enter target values (set points) for the connected


external device in the numerical field following 2 C u rre n t
the device name. E : C u rre n t M e a s u re m e n t

4 .0 0 7 3 m A

Note.
If a field indicator cannot be moved to the
external device's set point field (or the field
indicator is not seen at all), make sure the C a lib r a tio n
M o d e
M E N U

connections and the settings are in order. When


MC5 is able to communicate with the connected
device, a communication symbol is shown in the
status bar's rightmost end (see picture).
115

External Devices and


Calibration Mode
The Settings window for external devices may 1 9 .0 1 .2 0 0 5 1 3 :3 6
I N S T R U M E N TV i e w
also be opened from Calibration Mode's 1 1 2 -T T -0 0 3 .1 In s tr u m e n t
D e ta ils
Instrument overview window. In ta k e T e m p e r a tu r e o n F e e d 1
M a k e a C o p y
T r a n s fe r F u n c tio n L in e a r
V ie w
More of Calibration Mode in sections dedicated C a l. P o in ts 5 C a lib r a tio n
R e s u lts
to calibration.
IN P U T 0 .0 0 ... 1 0 0 .0 0 C
M e th o d S i m u l a t e Dd e l e t e
S e n s o r T y p e P t 1 0 0 a 3 I8 n 5 s t r u m e n t

O U T P U T 4 .0 0 0 0 ... 2 0 .0 0 0 0 m A
M e th o d M e a s u re d
C o n tr o lle r
S e ttin g s
C lo s e
B a c k E d it C a lib r a te M E N U
M E N U
116

MC5 Training HART, Part 1


General
117

What is HART
HART (Highway Addressable Remote
H A R T s ig n a l
Transducer) is a digital communication protocol
that uses sine waves superimposed on the + 0 .5 m A

standard 4 - 20 mA current. 0

2 0 m A -0 .5 m A
HART is a master-slave communication 1 2 0 0
H z
2 2 0 0
H z
protocol, which means that during normal " 1 " " 0 "

operation, each slave (field device)


R
communication is initiated by a master A n a lo g C
communication device. s ig n a l
C
R

R
C
Two masters (a primary master and a secondary C = C o m m a n d
master) can connect to a HART loop. 4 m A R = R e s p o n s e

0 T im e
118

HART Message Frame

PREAMBLE Preamble 5 to 20 bytes, hex FF


START Start Character, 1 byte
ADDR Addresses: source and destination,
1 or 5 bytes
COMM Command: 1 byte
BCNT Byte count (of status and data): 1 byte
STATUS Status: 2 bytes, only in slave response
DATA Data: 0 to 25 bytes*
CHK Checksum: 1 byte

*25 bytes is a recommended maximum data length


119

HART Command Structure


Universal Commands Common Practice Commands
Model ID Read Variables
Tag ID Calibration (zero, span)
Description Initiate Self Tests
Message Serial Number
Date Time Constant Values
Range Values
Instrument Limits
Process Measurements
Device Status Device Specific Commands
Model Specific Functions
Start, Stop, or Clear Totalizer
Select Primary Variable
Special Calibration Options
PID - SP, Tuning,
Etc.
120

Burst Mode
Normally a slave (instrument) answers only after 2 0 m A

a request from the master.

A request to an instrument can be sent max 2 R


times / second. A n a lo g
s ig n a l
R
R
R
When instrument is set into Burst mode, it will C
R

send all the time (3-4 times / second) C = C o m m a n d


4 m A
R = R e s p o n s e (B u rs t M o d e )

0 T im e
121

MC5 Training HART, Part 2


HART Related Connections
122

Point-to-Point Connection
24V supply mA
Transmitter
Resistor
+ HART
_
mA
Polling address = 0 !
=> HART + mA signal

HART
Communicator
123

Multidrop Connection
24V supply
Resistor
+
_
4 mA
(fixed)

HART ...
Communicator
Addr=1 2 15
Transmitters (max 15 pcs), fixed 4 mA current,
Only PV output, different addresses (not 0 !)
124

Primary/Secondary Master
Connections
24V supply
Resistor
+
_

...
Primary Secondary 1 2 15
HART HART
Master Master
125

MC5 Internal Connections

24V supply Transmitter


Resistor mA
+24V
+
_ Ground

HART mA

HART
Communicator
126

HART vs. Conventional


Transmitter, Part 1
Conventional Transmitter

Input 4 20 mA

Adjustments:
Zero
Span
127

HART vs. Conventional


Transmitter, Part 2
HART Transmitter

Input section Conversion Output section

PV AO
Input 4 .. 20 mA
A/D D/A
Digital trim Range Analog trim
Xfer func.

PV AO
128

MC5 Training HART, Part 3


MC5 and HART, Basics
129

HART Communication
Settings, Part 1
Before attempting to communicate with HART 2 0 .0 1 .2 0 0 5 1 5 :2 3
1 P re s s u re
instruments it is advisable to review MC5's P 1 : IN T 2 C /-1 .0 ... 2 b a r g S e ttin g s

HART communication settings. 0 .7 5 4 9 g a u g e


bM a a r i n t e n a n c e

H A R T
Open the settings window by pressing C o m m u n ic a tio n
S e ttin g s
D/Menu, 2 C u rre n t
C/Others and E : C u rre n t M e a s u re m e n t
D a ta
3/HART Communication Settings 0 .4 8 9 7 L o g g in g
m A
T r a n s m itte r /
S w itc h
S im u la tio n

C W a il ni b d r o a wt i o 1 n W in d o w 2 O th e rs M C lE o N s U e
MS e o t d u e p S e tu p M E N U
130

HART Communication
Settings, Part 2
Master Address 0 6 .1 0 .2 0 0 0 1 5 :3 2
H A R T C O M M U N IC A T IO N S E T T IN G S
HART communication supports the use of two
masters, a Primary and a Secondary master. M a s te r A d d re s s S e c o n d a ry

Select a unique master address for MC5.


P r e a m b le s u s e d 5
N u m b e r o f R e tr ie s
Preambles used 3

Can be a value between 5 to 20. The smaller the


value is, the faster the communication is, but
instruments requiring a larger amount of
preambles may not communicate.

Number of Retries
C lo s e

Can be a value between 1 to 10.


Determines the amount of retries when
communication errors occur.
131

Connecting to a
HART Transmitter, Part 1
To communicate with a HART instrument, press: 2 0 .0 1 .2 0 0 5 1 5 :3 9
1 P re s s u re
D/Menu (and B/Window 2 Setup, if needed) Q u a n tity
P 1 : I N T 4 0 0 M C / - 4 1 . 4 0 0 . . . 4 1 .[ 4 C 0 u 0 r rk e P n a t ]

5/HART. 0 .0 8 9 g a u g e
kF P u an c t i o n / P o r t
[E :I(M e a s )]

D is p la y M o d e
[E n g . U n its ]

U n it
2 C u rre n t [m A ]
E : C u rre n t M e a s u re m e n t
H A R T
3 .9 8 5 5 m A
Note.
The HART menu option is disabled if the other
window already reserves the E module. Change W in d o w 1 W in d o w 2 C lo s e
the setup of the other window to enable the S e tu p S e tu p
O th e rs
M E N U

HART menu option.


132

Connecting to a
HART Transmitter, Part 2
MC5 searches for HART devices connected to 2 0 .0 1 .2 0 0 5 1 5 :4 3
H A R T
the current loop and opens a window similar to
the one seen to the right. S e a r c h in g fo r H A R T D e v ic e a t: 1 2

D e v ic e s fo u n d :

Up to 16 instruments may be connected to the 1 - P T 1 5 5 - N C


2 - P T 3 1 1 - A R G
T T R
J T
same current loop. 3 - P T 4 1 3 - W B A 4 N T B
4 - T T 2 1 0 - P L G R N S W F
5 - T T 9 2 1 - F P N IL L

Choose the appropriate instrument by pressing 6 - T T 1 4 8 - T B B N E

4/Select.

B a c k R e s ta rt S e le c t
133

Connecting to a
HART Transmitter, Part 3
The standard Window Setup menu is replaced 0 6 .1 0 .2 0 0 0 1 5 :5 7
1 P re s s u re
by a HART menu. This menu is automatically D is c o n n e c t
P 1 : I N T 4 0 0 M C / - 4 1 . 4 0 0 . . . 4 1 . D 4 e0 v0 i ck Pe a

opened after the HART instrument is selected. 0 .0 8 7g a u g e


kD P e a v i c e
In fo r m a tio n

D e v ic e
The measurement quantity of the window where S e ttin g s

the HART instrument was selected to is 2


U n it
[k P a ]
P re s s u re
automatically changed to the quantity of the E : H A R T /P V

HART instrument's Digital Output (PV). 0 .0 8 4 k P a


T r im
In s tr u m e n t
A d d H A R T
In s tr u m e n t
to D a ta b a s e
K W a i ln i b d r o o w i n 1t i W V i n a d l i ot s w e 2 C lo s e
O th e rs V a lik k o
S t e i lt a u p kS e e n t ut t p M E N U
134

HART Device
Information
To view/edit HART device information, select 0 6 .1 0 .2 0 0 0 1 6 :0 1
H A R T D E V IC E IN F O R M A T IO N
D/Menu and T a g 3 0 5 1 C
B/Window 2 Setup, if needed D e v ic e ID 1 1 0 9 9 4 9

2/Device Information. S e
M a
r ia l N u m b
n u fa c tu re
e r
r
7 2 9 2
R o s e
3 0 1
m o u n t
D e v ic e T y p e 3 0 5 1 C
S o ftw a re V e r s io n 5 .3 .1 7 6
The following fields are editable: H a rd w a re R e v is io n 1 .0

Tag, D a te 0 7 .0 6 .1 9 9 9

Serial Number, D e s c r ip to r B E A M E X

Date, M e s s a g e J U S T T E S T IN G

Descriptor and
Message. C lo s e
135

HART Device
Settings
To view/edit HART device settings, select 0 6 .1 0 .2 0 0 0 1 6 :0 9
D E V IC E S E T T IN G S
D/Menu and T a g 3 0 5 1 C
B/Window 2 Setup, if needed P o llin g A d d r e s s 0

2/Device Settings. O u tp u t
P V /R a n g e U n it m b a r
L o w e r R a n g e 0 .0 0
U p p e r R a n g e
The following fields are editable:
1 0 0 0 .0 0
D a m p in g (s ) 0 .4 0

Tag, T ra n s fe
B u rs t M
r F u n c tio n
o d e O ff
L in e a r

Polling Address, S e n s o r

PV/Range Unit, S e r ia
L o w e
l N u m b e r
r L im it
1 1 0 9 9 4 9
0 .0 0

Lower Range, U p p e
M in im
r L im it
u m S p a n
2 4 8 6 .4 1
2 4 .8 7
Upper Range, C lo s e
Damping,
Transfer Function,
Burst Mode and
Sensor Serial Number.
136

Disconnecting a
HART Instrument
To disconnect a HART instrument, select 0 6 .1 0 .2 0 0 0 1 5 :5 7
1 P re s s u re
D/Menu and D is c o n n e c t
P 1 : I N T 4 0 0 M C / - 4 1 . 4 0 0 . . . 4 1 . D 4 e0 v0 i ck Pe a

B/Window 2 Setup, if needed 0 .0 8 7g a u g e


kD P e a v i c e

1/Disconnect Device. In fo r m a tio n

D e v ic e
S e ttin g s

U n it
2 P re s s u re [k P a ]
E : H A R T /P V

0 .0 8 4 k P a
T r im
In s tr u m e n t
A d d H A R T
In s tr u m e n t
to D a ta b a s e
K W a i ln i b d r o o w i n 1t i W V i n a d l i ot s w e 2 C lo s e
O th e rs V a lik k o
S t e i lt a u p kS e e n t ut t p M E N U
137

MC5 Training HART, Part 4


MC5 and HART, Advanced Use
138

Adding a HART Instrument to


MC5's Instrument Database
MC5 has a semi-automatic utility to add HART 1 1 .0 4 .2 0 0 2 9 :5 7
1 P re s s u re
instruments to MC5's instrument database. D is c o n n e c t
P 1 : I N T 4 0 0 M C / - 4 1 . 4 0 0 . . . 4 1 . D 4 e0 v0 i ck Pe a

1 .1 g a u g e
Dk P e a v i c e

Select 7/Add HART instrument to Database In fo r m a tio n

D e v ic e
from the Basic Mode's HART Menu. Continue by S e ttin g s

selecting which output is to be added, analog or 2 P re s s u re


U n it
[m b a r]
digital. E : H A R T /P V

1 .1 4 2 k P a
T r im
In s tr u m e n t
The following instrument data is acquired from A n a lo g O u tp u t (A O ) A d d H A R T
D ig ita l O u tp u t (P V ) In s tr u m e n t
the HART instrument: W in d o w 1 W in d o w 2
to D a ta b a s e
C lo s e
O th e rs
S e tu p S e tu p M E N U

Input Quantity, Input Unit, Input Range, Input


Method, Output Quantity, Output Unit, Output
Range, Output Method, Transfer Function,
Device ID and Setpoint Delay.
139

Calibrating a HART
Instrument
Calibrating a HART instrument does not differ 0 6 .1 0 .2 0 0 0 1 7 :2 1
In p u t P r e s s u r e [P 1 :IN T 1 C ]
from the calibration of a non-HART instrument.
5 1 2 .1 5 6 G a u g e
m b a r

O u tp u t P re s s u re [E : H A R T ]

5 1 0 .2 4 m b a r

Note.
E rro r
Calibrating the Analog Output (AO) of a HART +
1 .0 0 -0 .2 1 % o f s p a n

instrument does not necessarily require HART


communication. MC5 treats it as a standard 0

analog instrument during the calibration


procedure.
-
0 % 5 0 % 1 0 0 %
A c c e p t
S to p
R e a d in g s
140

Trimming a
HART Instrument, Part 1
To start the HART adjustment in Calibration 1 5 .0 5 .2 0 0 2 1 3 :2 2
In p u t P r e s s u r e [ P 1 : I N ST 2t a 0 r C t ]
Mode, select H A R T
0 .0 4 3 4 gAa d ua j g u e s t m e n t
D/MENU and k P
P re s c a n
[O n ]
1/Start HART Adjustment. O u tp u t C u rre n t [E : M e a s ]

4 .3 3 7 4 m A
V ie w
C a lib r a tio n
E rro r
1 .0 0 - 0 . 0 6 R% e o s fu sl t p s a n
+

Z e ro
P re s s u re
Note. -
0 % 5 0 % M o d u le 1 0 0 %
C lo s e
There is also the possibility to trim a HART M E N U

instrument in Basic Mode, but this training


material concentrates on trimming in Calibration
Mode.
141

Trimming a
HART Instrument, Part 2
Before trimming MC5 opens the instrument data 0 6 .1 0 .2 0 0 0 1 6 :2 8
IN S T R U M E N T IN P U T

pages allowing you to check if the instrument is P re s s u re

as supposed. In p u t M e th o d M e a s u re d

P o rt 1 IN T 1 C

1 9 .0 4 .2 0 0 0 1 5 :2 9

After checking the instrument data, press A/Trim P re s s u re ty p e G a u g e


C u rre n t
IN S T R U M E N T O U T P U T

to start the adjustment. U n it m b a r


O u tp u t M e th o d M e a s u re d
R a n g e 0 % 0 .0 0 0 0 0 0
1 0 0 % 1 0 0 0 .0 0 0 0 P o rt E

U n it m A

R a n g e 0 % 4 .0 0 0 0 0 0
N e x t
T r im E d it 1 0 0 % 2 0 .0 0 0 0 0 0
p a g e

T r a n s fe r F u n c tio n L in e a r
N e x t
T r im E d it
p a g e
142

Trimming a
HART Instrument, Part 3
In the HART Adjustment menu, the Digital 2 9 .0 5 .2 0 0 2 1 5 :4 2
In p u t P r e s s u r e [ P 1 : I N TT r2 i 0 m C L] o w e r
Output (PV) selected by default.
0 .0 4 3 3 g0a%u gR e
k P a
a n g e

T r im U p p e r

To change to Analog Output (AO), use the menu O u tp u t C u rre n t [E : M e a s ]


1 0 0 % R a n g e

option 4/Select Output. 4 .3 3 6 2 C h e c


m V Aa l u e
k A n y

A n a lo g O u tp u t (A O ) S e le c t
O u tp u t
E rro r D ig ita l O u tp u t (P V )
1 .0 0 - 0 . 0 4 [% A O o f] s p a n
+
D e v ic e
In fo r m a tio n

0 D e v ic e
S e ttin g s
E n d
- H A R T
0 % 5 0 % A d j u s t m e n1 0 t 0 %

Note.
Since Digital Output trim affects the Analog
Output trim, make sure you check/trim both
outputs.
143

Trimming a
HART Instrument, Part 4
When trimming the Digital Output (PV), the 1 6 .0 5 .2 0 0 2 1 5 :5 8
In p u t P r e s s u r e [P 1 :IN T 2 0 C ]
lower part of the Output window includes an
1 .0 2 9 G a u g e

additional trim field.


b a r

O u tp u t P re s s u re [E : H A R T \ P V ]

1 .0 2 1 b a r
T r im 0 % to : 1 .0 2 9
E rro r
1 .0 0 -0 .0 4 % o f s p a n
+

-
0 % 5 0 % 1 0 0 %
E n d
F e tc h M E N U
T r im
144

Trimming a
HART Instrument, Part 5
When trimming the Analog Output (AO), the 2 0 .0 8 .2 0 0 2 1 5 :4 4
In p u t C u rre n t [K e y e d ]
lower part of the Output window includes an
additional trim field. 4 .0 0 0 0 m A

O u tp u t C u r r e n t [E : M e a s .]

3 .9 9 8 2 m A
E n te r m e a s u re d 0 % 3 .9 9 8 2
E rro r
1 .0 0 N /A % o f s p a n
+

-
0 % 5 0 % 1 0 0 %
E n d
F e tc h M E N U
T r im

You might also like