Acumulative Time Service/Special 6100 Par. Adjustment Service/Mechanic 6161 Factory Mode Service/Special 6666 then press F12 Del.Instrument Information In Fact.Mod. press F11 6666 Continuous Run In Fact.Mod. press F9 6100 Pulse graph in Fact.Mod. press F5, F6, F7
Command Procedure
Flush High Voltage On Aperture
Enhanced Flush Cleaning program With Strong Cleaner (600 sec) - Probe Cleaner Cleaning Flush aperture from oposite side Enhanced Cleaning Concentrated cleaner cleaning procedure (at least 10 min or more) EA Cleaner
Phoenix NCC 1211 Linux
Function Code
Mechanic Test Service/Mechanic 6100
Factory Mode Service/Mechanic 6666 Del.Instrument Information In Fact.Mod. Setting/System set Continuous Run In Fact.Mod. Setting/System set/Aging Pulse graph in Fact.Mod. Setting/System set/Pulese
Command Procedure
Flush High Voltage On Aperture
Enhanced Flush Cleaning program With Strong Cleaner (600 sec) - Probe Cleaner Cleaning Flush aperture from oposite side Enhanced Cleaning Concentrated cleaner cleaning procedure (at least 10 min or more) EA Cleaner