Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Radiol Phys Technol (2017) 10:155160

DOI 10.1007/s12194-016-0378-x

Radiation dose optimization for the bolus tracking technique


in abdominal computed tomography: usefulness of real-time
iterative reconstruction for monitoring scan
Yuya Ishikawa1 Atsushi Urikura1 Tsukasa Yoshida1 Keisuke Takiguchi1

Yoshihiro Nakaya1

Received: 12 February 2016 / Revised: 14 September 2016 / Accepted: 14 September 2016 / Published online: 30 September 2016
 Japanese Society of Radiological Technology and Japan Society of Medical Physics 2016

Abstract The purpose of this study was to optimize the Keywords Computed tomography  Iterative
monitoring dose, obtained using the conventional filtered reconstruction  Radiation dose  Bolus tracking  Contrast
back projection (FBP) method and iterative reconstruction media
algorithms, for the bolus tracking technique. A phantom
study was performed to assess the effect of the scan start
time in patients grouped according to different body 1 Introduction
weights. An oval torso phantom was used for simulating
the time enhancement curve of the bolus tracking tech- Technical advances in multi-detector row computed
nique. To reproduce image noise levels in the two body tomography (CT) have made it possible to obtain detailed
weight groups, the phantom diameter was adjusted with a volume data with faster scan times [1, 2]. In multiphase
water-equivalent material. The tube currents were 10, 20, dynamic scanning for the detection of hepatic masses,
30, and 50 mA. The monitoring scan was performed with faster scanning contributes to improvement in diagnostic
the conventional FBP method and real-time adaptive iter- imaging by reflecting the blood flow state of the liver [3].
ative dose reduction by three-dimensional processing Dynamic CT requires that the scan timing be optimized to
(AIDR 3D). The results at different doses were compared the patient. Test bolus injection and bolus tracking (BT) are
with those at 50 mA. The volume computed tomography the two methods most commonly used to determine the
dose index was 1.31, 2.65, 3.93, and 6.56 mGy at tube optimum scan timing for dynamic CT [35].
currents of 10, 20, 30, and 50 mA, respectively. The scan BT detects the arrival time of the contrast medium by
start time, reconstructed using FBP, was significantly faster scanning arbitrary slices continuously to monitor the image
at 10 and 20 mA in group A (5059 kg) and at 20 mA in in real time. Scans are performed either intermittently or
group B (C80 kg). The CT values in the region of interest continuously, and automatically detect the arrival time of
could not be measured at 10 mA because of artifacts. With the contrast medium either visually or by using a region of
real-time AIDR 3D, both groups showed no significant interest (ROI) in blood vessels to measure CT values.
differences between the measurements obtained at 30 or Mehnert et al. [5] reported that BT that automatically
20 mA and those obtained at 50 mA. Our study demon- detects the arrival time of contrast medium is effective in
strated that the real-time AIDR 3D algorithm improved the determining the timing of the hepatic artery phase in
accuracy of the CT measurements with the bolus tracking multiphase imaging. In general, monitoring scans using BT
technique. are performed at low doses of B50 mAs [58]. However,
as continuous scanning results in increased local doses, the
dose should be set as low as possible.
& Yuya Ishikawa Recently, various technologies have been installed on
y.ishikawa@scchr.jp the latest CT scanners to reduce radiation doses during
1 examinations [911]. Iterative reconstruction (IR) is an
Department of Diagnostic Radiology, Shizuoka Cancer
Center, 1007 Shimonagakubo, Nagaizumi, Sunto, image reconstruction technique that uses an iterative
Shizuoka 411-8777, Japan algorithm. It first attracted attention for its ability to
156 Y. Ishikawa et al.

maintain nearly the same image quality as that of images addition, because the contrast agent arrival times in the
obtained by using the conventional filtered back projection ROI differed between the patients, the contrast media
(FBP) method while reducing both the dose and image detection time was normalized to 0 s. In this study, we
noise; thus, it is widely used in clinical settings [12]. In defined the arrival times as the time immediately before the
some advanced CT scanners, real-time IR can be used to CT values increased to more than two times the variations
obtain BT monitoring scan images, which is expected to in the precontrast CT values. The contrast agent arrival
reduce BT monitoring scan doses. Almost no study has time was obtained from the DICOM header data recorded
been conducted to determine the optimum dose to improve on the images using an injector synchronization system.
the accuracy of BT, and no study has been performed on
dose reductions made possible by IR. 2.2 Relationship between subject body weight
The purpose of this study was to assess and optimize the and image noise
effects of patient body weight and an IR algorithm on the
monitoring dose used to obtain valid trigger timing when To reproduce the relationship between subject body weight
using BT. and image noise when using a phantom for dynamic liver
CT, we divided the subjects described in Sect. 2.1 into five
body weight groups (4049 kg, n = 8; 5059 kg, n = 12;
2 Materials and methods 6069 kg, n = 4; 7079 kg, n = 3; and C80 kg, n = 3).
Image noise was obtained for all body weight groups by
2.1 Measurement of the time enhancement curve measuring the standard deviation (SD) of the CT values
measured prior to contrast agent detection in the ROI
This study was approved by our institutional review board described in Sect. 2.1.
(No. 28-J70-28-1-3). To reproduce the time enhancement
curve (TEC) for BT during abdominal dynamic CT using a 2.3 TEC phantom
phantom, we retrospectively measured the TECs of
dynamic liver CT cases where BT was used. Between April An overview of the phantom is shown in Fig. 1. We
and May 2014, 30 consecutive patients (male-to-female enclosed a water-equivalent material (agar) inside an
ratio: 21:9, mean age: 68.8 9.8 years) underwent elliptical phantom (Kyoto Kagaku Co., Ltd., Kyoto,
dynamic liver CT at our institution. The patients with body Japan) that mimics the human trunk. Cylindrical holes
weights of up to 75 kg received intravenous non-ionic were positioned on the dorsal side of the phantom to
iodinated contrast media at a dosage of 630 mgI/kg body reproduce the TEC in the monitoring scans by allowing a
weight. Those with body weights of [75 kg received the simulated blood vessel to pass through the scan slices.
fixed dosage of 47.25 gI. The infusion time was stan- The simulated blood vessel rod had a concentration gra-
dardized at 30 s. A circular ROI was set in the abdominal dient of continuous CT values (from 0 to 250 HU).
aorta at the origin of the celiac artery for the monitoring Movement of the rod was controlled by a power injector,
scan, which commenced 10 s after contrast agent infusion and it was passed through the scan slice at a constant
was started. Real-time monitoring scans were performed speed. In addition, to reproduce two types of noise levels
intermittently until the CT value of the ROI reached 80 (group A: 5059 kg and group B: C80 kg), we adjusted
Hounsfield units (HU) and then switched automatically to image noise by adding a uniform layer of water-equiva-
continuous scan. lent material to the outside perimeter of the phantom. The
All data were obtained using a 320-row area detector CT circular ROI with an area of 50 mm2 was placed at the
(Aquilion ONE, Toshiba Medical Systems, Tokyo, Japan). simulated blood vessel slices.
The BT software used was Real Prep (Toshiba Medical The ROI values were measured five times under each
Systems). The scan parameters were as follows: tube condition, and the TEC and image noise were compared
voltage, 120 kVp; tube current, 30 mA; rotation time, with the results described in Sect. 2.1. The data acquisition
0.5 s; detector configuration, 0.5 9 4 mm; slice thickness, parameters used were the same as those described in
2.0 mm; and reconstruction kernel, FC11 (FBP). Sect. 2.1.
Retrospective image reconstruction was performed for
the monitoring scans for all the patients. The ROI was set 2.4 Relationship between monitoring dose
in the abdominal aorta, and CT values were measured with and trigger timing
an interval of 0.5 s using Digital Imaging and Communi-
cations in Medicine (DICOM) information. The measured To assess the relationship between the monitoring dose
CT values were plotted to calculate the enhancement unit and trigger timing in BT, we measured the main scan
(EU) by subtracting the precontrast CT value [13]. In start time at different monitoring doses. Phantom slices
Radiation dose optimization for the bolus tracking technique in abdominal computed tomography 157

Fig. 1 A schematic diagram of the phantom used in the analysis. An oval torso phantom was filled with a water-equivalent material (agar).
A TEC was reproduced by scanning a simulated blood vessel passing through the scan cross section in the z-direction

were continuously scanned and monitored. The ROI 2.6 Statistical analyses
threshold was set at an EU of 100 HU. Differences in
trigger start times were compared by measuring the start We used the MannWhitney U test for the statistical
times of the main scans. Accurate scan start times were analysis of the standard deviation of the CT values in
obtained from the DICOM header data recorded on the Sect. 2.2. We used the paired t test for the statistical
images by an injector synchronization system. Four tube analysis of the scan start times mentioned in Sects. 2.4 and
currents were used as follows: 10, 20, 30, and 50 mA. 2.5. The significance level for all the statistical analyses
The monitoring dose was defined as the CTDIvol dis- was \5 %. All statistical analyses were performed using
played on the console. the EZR 1.31 software [14]. EZR is a graphical user
The other scan parameters were the same as those interface for R (The R Foundation for Statistical Comput-
described in Sect. 2.1. Each condition was measured five ing, Vienna, Austria).
times, and the mean values of the results were compared
with the values measured at 50 mA. The data were
obtained using the same parameters as described in 3 Results
Sect. 2.1, except for tube current.
3.1 Measurement of time enhancement curve
2.5 Effect of adaptive iterative dose reduction
by three-dimensional processing on noise level A graph showing the median TEC values obtained via
and trigger timing monitoring scans using BT on dynamic liver CT is shown
in Fig. 2. The time from contrast agent detection to an EU
To assess the effect of real-time adaptive iterative dose of 100 HU was *4 s.
reduction by three-dimensional processing (AIDR 3D,
Toshiba Medical Systems) on trigger timing under the 3.2 Relationship between body weight and image SD
conditions described in Sect. 2.4, we measured the main
scan start time for each dose. All the results were compared The relationship between the subjects body weights and
with the values measured at 50 mA. the image SDs is shown in Fig. 3. The median image SD
158 Y. Ishikawa et al.

Fig. 2 A graph showing the TEC plot acquired from the BT scan Fig. 4 A graph showing simulated TEC plots acquired from the
images of abdominal dynamic CT phantom measurements (square plots). The dotted line with diamond
plots indicates the results presented in Sect. 2.1

Table 1 The trigger times of the monitoring scans for the tube cur-
rents obtained using FBP
Trigger time (s)
Tube current (mA) Group A p value Group B p value

50 16.7 0.05 16.6 0.20


30 16.5 0.19 n.s. 16.1 0.40 n.s.
20 16.0 0.15 \0.01 14.6 0.48 \0.01
10 15.6 0.43 \0.01 u.m.
n.s. no significant difference u.m. unmeasurable
* p values were calculated for 50 vs. 10, 20, and 30mA in the two
groups

3.4 Relationship between the monitoring scan doses


and scan start times

The volume CT dose indexes (CTDIvol) for the monitoring


Fig. 3 Comparison between patient body weight and image noise scan tube current settings of 10, 20, 30, and 50 mA were
(the standard deviation of the CT values from the abdominal aorta) 1.31, 2.65, 3.93, and 6.56 mGy, respectively. The scan start
time in group A was slower at 30 mA but did not represent
values were 39.0, 49.1, 63.3, 79.5, and 129.8 HU for the a significant difference. The scan start times for 10 and
40- to 49-, 50- to 59-, 60- to 69-, 70- to 79-, and C80-kg 20 mA were significantly faster than that at 50 mA
body weight groups, respectively. (Table 1). The scan start time in group B did not signifi-
cantly differ at 30 mA, but the scan start time at 20 mA
3.3 Measurement of the TEC and SD was faster. CT values could not be measured at 10 mA.
with phantoms
3.5 Effect of AIDR 3D processing on scan start
The TECs obtained from the TEC phantoms are shown in times
Fig. 4. The phantom TECs were nearly the same as the
results in Sect. 3.1. All of the EU plots up to 100 HU were The scan start times when real-time AIDR 3D was used are
within two SD. The image SDs for the phantoms simulat- shown in Table 2. In group A, the scan start times at 20 and
ing the two weight groups (groups A and B) were 30 mA did not significantly differ from those at 50 mA, but
48.1 5.3 and 127.6 15.4 HU, respectively, and a significant difference was observed at 10 mA. In group B,
showed values that were similar to the results in Sect. 3.2. no significant difference was observed at any dose.
Radiation dose optimization for the bolus tracking technique in abdominal computed tomography 159

Table 2 The trigger times of the monitoring scans acquired using the B20 mA, which cause marked changes in CT values in the
IR algorithm ROI, and the threshold may be reached faster. In low-dose
Trigger time (s) monitoring (10 mA) in the high body weight group, streaks
and dark band artifacts caused by photon starvation and
Tube current (mA) Group A p value Group B p value
beam hardening were observed, which made it difficult to
50 16.7 0.05 16.6 0.20 measure CT values [15].
30 16.7 0.21 n.s. 16.6 0.18 n.s. The accuracies of the scan start time measurements
20 16.4 0.28 n.s. 16.4 0.41 n.s. under all dose conditions were improved through the use of
10 16.2 0.30 \0.05 16.3 0.75 n.s. real-time AIDR 3D in performing monitoring scans. At
C20 mA, the scan start times were the same as those at
n.s no significant difference
50 mA, regardless of patient body weight. Figure 5 shows
* p values were calculated for 50 vs. 10, 20, and 30 mA in the two
groups examples of clinical images of a monitoring scan obtained
from a 62-year-old man (body weight, 80 kg; height,
161 cm; and body mass index, 30.9 kg/m2), which were
4 Discussion
reconstructed with FBP (a) and AIDR 3D (b). The mean
CT (SD) values in the abdominal aorta (arrowhead) were
The TEC phantom that we used was effective at repro-
162 (104.3) and 153 (32.8) HU for FBP and AIDR 3D,
ducing CT value changes caused by BT on dynamic CT. It
respectively. The CT values and SD showed that AIDR 3D
also faithfully reproduced image noise changes resulting
was extremely effective when used in monitoring scans of
from different patient body weights.
clinical images and in suppressing apparent elevations in
FBP obtained equivalent scan start times regardless of
CT values caused by noise amplitude. Based on the pre-
body weight at all tube currents of C30 mA, under the
vious studies on physical image quality, IR changes fre-
imaging conditions used in this study. However, the scan
quency characteristics, unlike FBP [1618]. However, as
start times were shortened at B20 mA. The CT values in
IR maintains CT values [19, 20], the use of IR in moni-
group B were impossible to measure when the tube current
toring scans is effective in reducing noise. The monitoring
was set at 10 mA. Many previous clinical studies that used
scan doses used in the BT method used in this study could
BT set the tube current to B50 mA [58]. However, the
be reduced by 40 or 60 % by changing the tube current
tube current settings for BT monitoring scans often do not
setting from 50 to 30 or 20 mA, respectively.
account for patient body weight. To obtain optimum scan
This study has several limitations. First, the image noise
timing when using BT, stable CT values should be mea-
level was not necessarily identical between the real-time
sured. Noise increases at low tube current settings of
monitoring images and the retrospective reconstructed

Fig. 5 Clinical images of a monitoring scan obtained from a 62-year- rotation time of 0.5 s, and tube current of 30 mA. The mean (SD) CT
old man (body weight, 80 kg; height, 161 cm; and body mass index, values in the abdominal aorta (arrowhead) were 162 (104.3) and 153
30.9 kg/m2), which were reconstructed with FBP (a) and AIDR 3D (32.8) HU for FBP and AIDR 3D, respectively
(b). Both images were obtained with a tube voltage of 120 kVp,
160 Y. Ishikawa et al.

images acquired by FBP and AIDR 3D. The accurate 7. Sultana S, Awai K, Nakayama Y, Nakaura T, Liu D, Hatemura
image noise level of the real-time images might be dif- M, Funama Y, Morishita S, Yamashita Y. Hypervascular hepa-
tocellular carcinomas: bolus tracking with a 40-detector CT
ferent from our results. Second, the phantom was made of a scanner to time arterial phase imaging. Radiology.
uniform material that did not include bone or intestinal 2007;243:1407.
tract structures. Therefore, noise and artifacts originating 8. Tatsugami F, Awai K, Takada H, Yoshikawa S, Takeda Y,
from these peripheral structures may not have been Morita H, Narumi Y. Reduction of interpatient variability of
arterial enhancement using a new bolus tracking system in
reproduced. Third, as the phantom did not take patient 320-detector computed tomographic coronary angiography.
respiration and body movements into account, we could not J Comput Assist Tomogr. 2013;37:7983.
know the effect of these factors on the clinical images. 9. Catalano C, Francone M, Ascarelli A, Mangia M, Iacucci I,
Fourth, this study did not assume differing thresholds and Passariello R. Optimizing radiation dose and image quality. Eur
Radiol. 2007;17(Suppl 6):F2632.
contrast conditions. Finally, this study investigated a single 10. Matsubara K, Sugai M, Toyoda A, Koshida H, Sakuta K, Takata
device supplied by one vendor and did not investigate CT T, Koshida K, Iida H, Matsui O. Assessment of an organ-based
scanners and IR algorithms from other vendors. tube current modulation in thoracic computed tomography.
J Appl Clin Med Phys. 2012;13:3731.
11. Singh S, Kalra MK, Hsieh J, Licato PE, Do S, Pien HH, Blake
MA. Abdominal CT: comparison of adaptive statistical iterative
5 Conclusion and filtered back projection reconstruction techniques. Radiology.
2010;257:37383.
Applicable real-time IR algorithms for dynamic liver CT 12. Willemink MJ, de Jong PA, Leiner T, de Heer LM, Nievelstein
RA, Budde RP, Schilham AM. Iterative reconstruction techniques
BT monitoring increased the accuracy of CT measurements for computed tomography Part 1: technical principles. Eur
by reducing high noise levels. This indicates that the use of Radiol. 2013;23:162331.
IR with monitoring scans allowed for reducing doses to 13. Yamaguchi I, Kidoya E, Suzuki M, Kimura H. Evaluation of
below those used with conventional FBP. required saline volume in dynamic contrast-enhanced computed
tomography using saline flush technique. Comput Med Imaging
Compliance with ethical standards Graph. 2009;33(1):238.
14. Kanda Y. Investigation of the freely available easy-to-use soft-
ware EZR for medical statistics. Bone Marrow Transplant.
Conflict of interest The authors declare that they have no conflict of
2013;48:4528.
interest.
15. Barrett JF, Keat N. Artifacts in CT: recognition and avoidance.
Radiographics. 2004;24:167991.
16. Richard S, Husarik DB, Yadava G, Murphy SN, Samei E.
References Towards task-based assessment of CT performance: system and
object MTF across different reconstruction algorithms. Med
1. Hein PA, Romano VC, Lembcke A, May J, Rogalla P. Initial Phys. 2012;39:411522.
experience with a chest pain protocol using 320-slice volume 17. Urikura A, Ichikawa K, Hara T, Nishimaru E, Nakaya Y. Spatial
MDCT. Eur Radiol. 2009;19:114855. resolution measurement for iterative reconstruction by use of
2. Rogalla P, Kloeters C, Hein PA. CT technology overview: image-averaging techniques in computed tomography. Radiol
64-slice and beyond. Radiol Clin North Am. 2009;47:111. Phys Technol. 2014;7:35866.
3. Hammerstingl RM, Vogl TJ. Abdominal MDCT: protocols and 18. Smith EA, Dillman JR, Goodsitt MM, Christodoulou EG,
contrast considerations. Eur Radiol. 2005;15(Suppl 5):E7890. Keshavarzi N, Strouse PJ. Model-based iterative reconstruction:
4. Sandstede JJ, Tschammler A, Beer M, Vogelsang C, Wittenberg effect on patient radiation dose and image quality in pediatric
G, Hahn D. Optimization of automatic bolus tracking for timing body CT. Radiology. 2014;270:52634.
of the arterial phase of helical liver CT. Eur Radiol. 19. Love A, Olsson ML, Siemund R, Stalhammar F, Bjorkman-
2001;11:1396400. Burtscher IM, Soderberg M. Six iterative reconstruction algo-
5. Mehnert F, Pereira PL, Trubenbach J, Kopp AF, Claussen CD. rithms in brain CT: a phantom study on image quality at different
Biphasic spiral CT of the liver: automatic bolus tracking or time radiation dose levels. Br J Radiol. 2013;86:20130388.
delay? Eur Radiol. 2001;11:42731. 20. Schindera ST, Odedra D, Raza SA, Kim TK, Jang HJ, Szucs-
6. Dinkel HP, Fieger M, Knupffer J, Moll R, Schindler G. Opti- Farkas Z, Rogalla P. Iterative reconstruction algorithm for CT:
mizing liver contrast in helical liver CT: value of a real-time can radiation dose be decreased while low-contrast detectability
bolus-triggering technique. Eur Radiol. 1998;8:160812. is preserved? Radiology. 2013;269:5118.

You might also like