Professional Documents
Culture Documents
Kolb 1992
Kolb 1992
0009-5893/92/9 0235-06 $ 03.00/0 9 1992 Friedr. Vieweg & Sohn Verlagsgesellschaft mbH
and has already been applied with various modifica- "vapor phase calibration t e c h n i q u e - VPC", f u l f i l ~
tions. In looking for the best procedure for an requirements for automation and is described ano ,,is
automated technique, none of those briefly discussed cussed in detail below. It should be mentioned that tt~e
below have been found ideal, i.e., sufficiently simple for . variant
a specxal . of. the general
. prmc~ple wher e 1( Call-e~ ~
this purpose, hence a new version of the headspace tech- derived from the resulting difference in peak atto6
nique, termed "vapor phase calibration (VPC)", had when the sample volume is varied, since with this *"
therefore been developed. technique the volume of the sample matrix in the refer'
Jones et al. [3] determined the partition coefficient of ence system becomes zero9
ethyl acetate in water at O ~ from independent
analyses of the total amount of ethyl acetate in the
headspace vial and from the amount in the gas phase9 "Vapor Phase Calibration (VPC)"
The gas phase was calibrated against an external liquid Technique
standard injected by a microsyringe. Kieckbusch et al.
[4] also used a microsyringe but for the analysis of the The VPC technique is based on the nrinciple that t!eo
9 .
liquid phase. The combination of headspace injection concentrataon of the volatile analyte r.]n the g a s P_I~a~'
/f
by gas syringe and liquid injection by microsyringe can be determined by external standard calibrations'^',.
needs very careful calibration between both sampling the total amount present in the vial is known, the co~
devices and is hard to automate 9 centration in the sam P le P hase at e q uilibrium is fott]j~
from the difference.
Two determinations using different sampling techni-
ques can be avoided if the same instrumentation is used This technique, where the distribution of a volatile c.~ I
and if, in the prevailing two-phase system, one of the in- pound between a two-phase system in a headspaCe *.'~s
fluencing parameters is varied. For such an approach is determined bv usinz a pure vapor as a reference, w~
. . . . . tllu
the total amount in the two-phase partition system can originally applied and performed with an autOna~...
- -coP"
be altered if, after headspace analysis, the equilibrated headspace
. . . .
sampler by". Kolb for
.
the
.
determination u'.,~l
9 l/,j
gas phase is removed and replaced by fresh inert gas. tnbutlon coefficients m gas-hqmd [11] and gas-soltd [..,
After re-establishing equilibrium, the amount of analyte systems and later by Schoene et al " [13] to deterr~, iJI0
in the gas phase becomes smaller. The partition coeffi- solubility coefficients of vapors in both solid and llq.,
cient can be derived from this difference9This technique polymers. Although these investigations have t,
. . . dtstnbutxon
focused on non-hnear . as found both in coO" .,
was first applied by McAuliffe [5] followed by others
[6-8] but needs a gas-tight syringe to separate mechani- centrated solutions and in gas-sohd adsorption sys~ .,,
cally and completely both the liquid and the gas phases9
It is obvious that these techniques, although they have
become quite popular, are difficult to automate 9
The gas phase, however, need not be replaced totally to of partition coefficients. After having screened alte[~.~
shift the equilibrium. It can be removed partially simply tives, it has been found very useful for a fully automat~
by venting a pressurized headspace vial to atmosphere. procedure. Analytical determinations are p e r t o r w
It is thus possible to derive K from the resulting dif- automatically under constant instrumental cond~t~o,
ferences of the peak areas in the sequential headspace while the necessary variation of the sample compositl~
chromatograms determined in a similar way [9]. This is manual when preparing samples.
MHE (multiple headspace extraction) procedure,
described in detail in a recent paper [10], can be carried
out automatically, but again various practical variants Fundamentals of VPC
are possible. One of these variants is used here in com-
bination with MHE, where the sample volume and thus When an amount m ~ of the volatile analyte is injeCte!
the phase ratio is varied. into a vial containing the sample matrix with the voltlt~~
Vs, it is distributed between the gas phase and t
It is obvious for the purpose of automation that the sample as described by Ea (1), and the amount ~n ~,
same sampling technique should be used. With head- sample (ms) after equilibration results from the o,
space sampling instrumentation only the gas phase is ac- ference, Eq. (2):
cessible for analysis of the volatile analyte. Whenever
physico-chemical data are determined, the original CS ms VG (1)
composition of the sample is usually well defined, In the CO mo VS
case of partition coefficient determination, it is the gas
phase composition which is determined analytically. ms = m ~ - mo (~)
From the total amount of volatile analyte, exactly
known from the preparation of the original sample, the The peak area As from a headspace analysis co[
concentration in the liquid or solid sample under the responds to the gas phase concentration C6 with @ ~
prevailing temperature conditions is derived from the calibration factor which includes instrumental Park
difference. Quantitative analysis of the gas phase can meters such as injected volume of headspace gas ~tl0
very simply be carried out by external standard calibra- detector response but no specific headspace paramete~
tion. This procedure, which therefore is termed here: such as phase ratios or matrix effects. @ can therefo rel~
rn~ = @ AC V v (4)
~'he - Table |. Water-air partition coefficient K of MEK at 70 *C by VPC,
COncentration ,~f
t0110..... , , , &~ e a n a lyt e in the g as P hase , C G, Eq. (9).
"~ xrom Eq. (5) and the amount, mG, by Eq. (6):
A d d e d volume of M E K [/aLl 2.0 3.0 4.0 5.0
CG= m G / V o = @ A s (5) peak area A c 35202 51357 66903 83657
peak area As 4868 7005 9207 11434
k'IG .~ @ As VG (6)
K 47.29 48.04 47.58 47.92
~'~ am~ ms, of analyte in the sample phase is deter- mean value for K: 47.70 +/- 0.7 % C.V., N = 4
ecl from the difference, Eq. (7):
rns = @ [Ac V v - As VG] (7) conditions: vial volume Vv = 22.3 mL, matrix volume Vs = 3.0 mL,
gas volume Vc~ = 19.3 mL.
Introduction of E^s r2~ and (73 in E (1) gives Eqs. (8)
antlrm . q.~ J , , q"
t~) Where @ is finally eliminated.
Results by Linear Regression
K = ms VG @[AcVv-AsVG]VG
- (8) Whenever differences in peak areas are used, it is good
Vs mG @ A s VG Vs
practice to check whether both areas are still in the
range of headspace linearity; this is a linear function of
K -- A c V v - As VG (9) concentration in the sample and its partial vapor pres-
As Vs sure expressed by a constant K, independent of con-
centration. Having the wide range of partition coeffi-
~,ie~cpC technique has been applied to 2-butanone
cients in mind and the corresponding variation of
be~ K) in water at 70 ~ This model system was chosen solubility and resulting concentration, it is advisable to
res~ use the partition coefficient of roughly 47 is a rep- include such a linearity test e.g., by linear regression.
o;^entative value in the wide range of partition coeffi-
~'Cnts The data in Table I can also be subjected to linear
f0r L from K = 0 to 104, where a reasonable volatility regression. K is derived by replacing AC and As in Eq.
"eadspace applications ends. (9) by the values of the corresponding slopes ac and as
according to the modified Eq. (9a):
EXPerimenta ! ac V v - as V o
K= (9a)
The. as Vs
fie- wa!er-air partition coefficient of M E K at 70 ~ was
terramed by VPC under the following conditions. ac = slope of regression line for vapor standard
as = slope of regression line for aqueous sample
~,L.YOther amount, not necessarily m* of the analyte can be used. Accuracy of VPC Technique
t].~s makes the VPC technique flexible enough to adjust the some-
-ues big differences in peak areas for the external vapor standard The accuracy of the VPC technique was tested by two
and the Sample in the case of a high K value. independent procedures. The M H E technique was used
3 354:3 I~'~
f~erttransformation (cf. Figure 2) K can be d e r i v e d for *conditions: sample volume Vs = 3.0 mL, vial volume Vv = 22.3 mL,
ern- m p e r a t u r e b y linear regression, at least in the 3.0/.tL mixture of three compounds added.
sev~erature r a n g e w h e r e K has b e e n d e t e r m i n e d at
ro, er.al t e m p e r a t u r e s . A n e x a m p l e of the linear
~ a h o n s h i p in a s e m i l o g plot is s h o w n in Figure 2 with
t h e V P C t e c h n i q u e with linear r e g r e s s i o n calculations.
~ a r t o l , M E K a n d d i c h l o r o m e t h a n e at five t e m p e r a -
A m i x t u r e o f these c o m p o u n d s was p r e p a r e d and 1.0,
~es in the r a n g e 40 ~ - 80 ~
3.0 and 5.0 p L w e r e injected into 5.0 m L of d e i o n i z e d
water. T h e resulting partition coefficients are listed in
T a b l e III. T h e t e m p e r a t u r e d e p e n d e n c e was deter-
SUPplemen t mined. B o t h coefficients a a n d b f r o m the linear regres-
sion, as s h o w n in Figure 2, are also listed to allow recal-
~v~ ition coefficients of several volatile c o m p o u n d s
culation for o t h e r t e m p e r a t u r e s .
tier" Wtdely differing solubilities in w a t e r h a v e b e e n
errained in t h e t e m p e r a t u r e r a n g e 40 ~ to 80 ~ by
compound 40 *C 60 *C 70 ~ 80 *C b* a*
*cf. Figure 2.