Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

betermination of Partition Coefficients by Automatic

qUilibrium Headsnace Gas Chromatography by Vapor


ase Calibration r

I~.Kolb* / C. Welter / C. Bichler


I~adenseewerk Perkin-Elmer GmbH, 7770 12/berlingen, Germany

tion technique, the value of K is in general not known


l(ey Words for most practical headspace applications and with few
exceptions is not used for quantitative calibration tech-
~as Chromatograph niques. Its knowledge, however, would be very useful
bleads . Y
Pa..... Pace samphng for quantitative determination, particularly for compar-
'tttlon coefficients ing headspace samples of differing sample volume. The
Multiple headspace extraction sample volume in a headspace vial influences the
headspace sensitivity as does the partition coefficient
and the sensitivity depends strongly on the value of K
SU~naary and the phase ratio 13,as discussed recently by Ettre and
Kolb [1]. The main application of E-HSGC is in the
~qu!librium headspace gas chromatography has been quantitative analysis of volatile compounds in liquid or
c~JPlled to the determination of the partition coeffi- solid samples, but there are also other interesting ap-
teeLts,of volatile compounds in water-air systems. Only plications for the determination of physico-chemical
sp~t~n~ques that are suited to a fully automatic head- data, based on phase partitioning systems. It is of inter-
nil_~ Procedure usin~ the oneumatic headspace sam- est, particularly for environmental analysis to obtain in-
~""g-t,~ ,- -
s ~ . "-Camque have~ been considered. Particularly
- formation on the solubility and adsorptivity of hazard-
~pU~l,,e is the technique of "vapor phase calibration - ous compounds in water and soil and their release to the
Cal:L Where an external vapor standard is used to atmosphere. Also for biomedical purposes, especially
he,Orate the concentration of the volatile analyte in the industrial hygiene, the partitioning of noxious com-
fou~Space, while the concentration in the sample is pounds between blood and air in the lung needs to be in-
.1,~.."'~,tram the difference in the total amount in the vial. vestigated for measuring BEI values (biological ex-
s technique is described in detail for 2-butanone in posure index) or for BAT values (biological tolerance
~,.,cr. Finally, the water-air partition coefficients of values for working materials). Owing to this increasing
ie;e ral selected volatile compounds at different interest in such partition systems, there is a need to
e,. Peratures are listed together with their temperature measure a large number of partition coefficients and an
"UtlCtlons. automatic technique is highly desirable for this purpose.
Rohrschneider [2] first used an automated headspace
instrument to measure the partition coefficients of six
reference compounds with different polarities in 80 sol-
vents. These measurements were performed on a rela-
Introduction tive basis, a particular system with a known partition
coefficient (n-octane in squalane/air at 25 ~ being
~qr~!!!brium headspace GC (E-HSGC)is based on the used for reference purposes. This procedure, however,
Ors~t,!~ of a volatile compound between the liquid is not generally applicable, as it depends on the
b,, ,,ad Sample and the surrounding gas phase followed availability of data for any desired temperature and
thJetransfer of the volatile compound in the gas phase to
their accuracy depend on the reproducibility of the ref-
nr,,gas chromatograph for analysis. The partitioning erence system.
~vh'~ess is governed by the distribution coefficient,
the ~' re.the case of gas-liquid systems is usually termed The partition coefficient is the ratio of the concentra-
nr^ Parhtion coefficient K. Although this distribution tion of a volatile compound in two phases and although
~'eadess is the most important parameter, both for the it can be derived from quantitative analysis of both
USpace sensitivity and for any quantitative calibra- phases, there is a variety of procedures in the literature
which are briefly discussed below.
. ediCaledto Professor Leslie S. Ettre on the occasion of his 70th
birthday. Of all these techniques the equilibrium headspace
method offers the best prerequisites for automation,
(2hrornatographia Vol. 34, No. 5-8, September/October 1992 235

0009-5893/92/9 0235-06 $ 03.00/0 9 1992 Friedr. Vieweg & Sohn Verlagsgesellschaft mbH
and has already been applied with various modifica- "vapor phase calibration t e c h n i q u e - VPC", f u l f i l ~
tions. In looking for the best procedure for an requirements for automation and is described ano ,,is
automated technique, none of those briefly discussed cussed in detail below. It should be mentioned that tt~e
below have been found ideal, i.e., sufficiently simple for . variant
a specxal . of. the general
. prmc~ple wher e 1( Call-e~ ~
this purpose, hence a new version of the headspace tech- derived from the resulting difference in peak atto6
nique, termed "vapor phase calibration (VPC)", had when the sample volume is varied, since with this *"
therefore been developed. technique the volume of the sample matrix in the refer'
Jones et al. [3] determined the partition coefficient of ence system becomes zero9
ethyl acetate in water at O ~ from independent
analyses of the total amount of ethyl acetate in the
headspace vial and from the amount in the gas phase9 "Vapor Phase Calibration (VPC)"
The gas phase was calibrated against an external liquid Technique
standard injected by a microsyringe. Kieckbusch et al.
[4] also used a microsyringe but for the analysis of the The VPC technique is based on the nrinciple that t!eo
9 .
liquid phase. The combination of headspace injection concentrataon of the volatile analyte r.]n the g a s P_I~a~'
/f
by gas syringe and liquid injection by microsyringe can be determined by external standard calibrations'^',.
needs very careful calibration between both sampling the total amount present in the vial is known, the co~
devices and is hard to automate 9 centration in the sam P le P hase at e q uilibrium is fott]j~
from the difference.
Two determinations using different sampling techni-
ques can be avoided if the same instrumentation is used This technique, where the distribution of a volatile c.~ I
and if, in the prevailing two-phase system, one of the in- pound between a two-phase system in a headspaCe *.'~s
fluencing parameters is varied. For such an approach is determined bv usinz a pure vapor as a reference, w~
. . . . . tllu

the total amount in the two-phase partition system can originally applied and performed with an autOna~...
- -coP"
be altered if, after headspace analysis, the equilibrated headspace
. . . .
sampler by". Kolb for
.
the
.
determination u'.,~l
9 l/,j

gas phase is removed and replaced by fresh inert gas. tnbutlon coefficients m gas-hqmd [11] and gas-soltd [..,
After re-establishing equilibrium, the amount of analyte systems and later by Schoene et al " [13] to deterr~, iJI0
in the gas phase becomes smaller. The partition coeffi- solubility coefficients of vapors in both solid and llq.,
cient can be derived from this difference9This technique polymers. Although these investigations have t,
. . . dtstnbutxon
focused on non-hnear . as found both in coO" .,
was first applied by McAuliffe [5] followed by others
[6-8] but needs a gas-tight syringe to separate mechani- centrated solutions and in gas-sohd adsorption sys~ .,,
cally and completely both the liquid and the gas phases9
It is obvious that these techniques, although they have
become quite popular, are difficult to automate 9
The gas phase, however, need not be replaced totally to of partition coefficients. After having screened alte[~.~
shift the equilibrium. It can be removed partially simply tives, it has been found very useful for a fully automat~
by venting a pressurized headspace vial to atmosphere. procedure. Analytical determinations are p e r t o r w
It is thus possible to derive K from the resulting dif- automatically under constant instrumental cond~t~o,
ferences of the peak areas in the sequential headspace while the necessary variation of the sample compositl~
chromatograms determined in a similar way [9]. This is manual when preparing samples.
MHE (multiple headspace extraction) procedure,
described in detail in a recent paper [10], can be carried
out automatically, but again various practical variants Fundamentals of VPC
are possible. One of these variants is used here in com-
bination with MHE, where the sample volume and thus When an amount m ~ of the volatile analyte is injeCte!
the phase ratio is varied. into a vial containing the sample matrix with the voltlt~~
Vs, it is distributed between the gas phase and t
It is obvious for the purpose of automation that the sample as described by Ea (1), and the amount ~n ~,
same sampling technique should be used. With head- sample (ms) after equilibration results from the o,
space sampling instrumentation only the gas phase is ac- ference, Eq. (2):
cessible for analysis of the volatile analyte. Whenever
physico-chemical data are determined, the original CS ms VG (1)
composition of the sample is usually well defined, In the CO mo VS
case of partition coefficient determination, it is the gas
phase composition which is determined analytically. ms = m ~ - mo (~)
From the total amount of volatile analyte, exactly
known from the preparation of the original sample, the The peak area As from a headspace analysis co[
concentration in the liquid or solid sample under the responds to the gas phase concentration C6 with @ ~
prevailing temperature conditions is derived from the calibration factor which includes instrumental Park
difference. Quantitative analysis of the gas phase can meters such as injected volume of headspace gas ~tl0
very simply be carried out by external standard calibra- detector response but no specific headspace paramete~
tion. This procedure, which therefore is termed here: such as phase ratios or matrix effects. @ can therefo rel~

236 Chromatographia Vol.34, No. 5-8, September/October 199~


~etern~ined simply by external standard calibration with Sample Preparation and Results
Por Sample of the anab,te For this nurt~ose a small
Four vials, each containing 3.0 mL water, were prepared
~OUnt of the analyte, e.g. m~ * is ~njected into an
9-ply vial either a - and 2.0, 3.0, 4.0 and 5.0 laL of M E K added by means of
Whe- 9 s the pure compound or as a solutmn a microsyringe. Identical volumes of M E K were added
tUrore ~t vaporizes totally under the prevailing tempera-
to the empty reference vials.
.calib,^.
.e conditions 9 T h. .~. . .r .o.~. n. l t i n ~~,,~
n e a k area A C is used to
in E ,~te the concentration CG ~ of the analyte vapor as The results of the VPC-technique according to Eq. (9),
q. (3), from which Eq (4) is derived. where the partition coefficient is determined from the
area difference of each pair of samples, are listed in
C~ ~- m~ / V v = @ A c (3) Table I.

rn~ = @ AC V v (4)
~'he - Table |. Water-air partition coefficient K of MEK at 70 *C by VPC,
COncentration ,~f
t0110..... , , , &~ e a n a lyt e in the g as P hase , C G, Eq. (9).
"~ xrom Eq. (5) and the amount, mG, by Eq. (6):
A d d e d volume of M E K [/aLl 2.0 3.0 4.0 5.0
CG= m G / V o = @ A s (5) peak area A c 35202 51357 66903 83657
peak area As 4868 7005 9207 11434
k'IG .~ @ As VG (6)
K 47.29 48.04 47.58 47.92
~'~ am~ ms, of analyte in the sample phase is deter- mean value for K: 47.70 +/- 0.7 % C.V., N = 4
ecl from the difference, Eq. (7):
rns = @ [Ac V v - As VG] (7) conditions: vial volume Vv = 22.3 mL, matrix volume Vs = 3.0 mL,
gas volume Vc~ = 19.3 mL.
Introduction of E^s r2~ and (73 in E (1) gives Eqs. (8)
antlrm . q.~ J , , q"
t~) Where @ is finally eliminated.
Results by Linear Regression
K = ms VG @[AcVv-AsVG]VG
- (8) Whenever differences in peak areas are used, it is good
Vs mG @ A s VG Vs
practice to check whether both areas are still in the
range of headspace linearity; this is a linear function of
K -- A c V v - As VG (9) concentration in the sample and its partial vapor pres-
As Vs sure expressed by a constant K, independent of con-
centration. Having the wide range of partition coeffi-
~,ie~cpC technique has been applied to 2-butanone
cients in mind and the corresponding variation of
be~ K) in water at 70 ~ This model system was chosen solubility and resulting concentration, it is advisable to
res~ use the partition coefficient of roughly 47 is a rep- include such a linearity test e.g., by linear regression.
o;^entative value in the wide range of partition coeffi-
~'Cnts The data in Table I can also be subjected to linear
f0r L from K = 0 to 104, where a reasonable volatility regression. K is derived by replacing AC and As in Eq.
"eadspace applications ends. (9) by the values of the corresponding slopes ac and as
according to the modified Eq. (9a):

EXPerimenta ! ac V v - as V o
K= (9a)
The. as Vs
fie- wa!er-air partition coefficient of M E K at 70 ~ was
terramed by VPC under the following conditions. ac = slope of regression line for vapor standard
as = slope of regression line for aqueous sample

Instrun~ental Conditions The necessary values to derive K from the data in


Table I and Eq. (9a) are the slopes of the regression line
t" e tig.tio. we e.rriod out wit
g"a~'lO.1 Automatic Headspace Sampler and Model 8700
a erk n- mor from the reference 'empty' vial (ac - 16091.10; correla-
tion coefficient r = 0.99990) and that of the sample vial
stY. e~romatograph. G C conditions: 2 m x 1/8" packed (as = 2190.0; r = 0.99974) together with the volume of
(2~Inless steel column, 1 5 % Carbowax 1500 on the sample matrix (Vs = 3.0 mL) and the vial volume
l~i~~ W, NAW, 60-80 mesh at 60 ~ detector: (Vv = 22.3 mL). From these data K = 48.2, which is in
Jec,'" Headspace conditions: thermostat time 60 rain, in- good agreement with the mean value of K = 47.7,
sa;l~n time 0.05 rain, pressurization time 2.0 min, +/- 0.7 % C.V. listed in Table I.
Pie matrix: deionized water.

~,L.YOther amount, not necessarily m* of the analyte can be used. Accuracy of VPC Technique
t].~s makes the VPC technique flexible enough to adjust the some-
-ues big differences in peak areas for the external vapor standard The accuracy of the VPC technique was tested by two
and the Sample in the case of a high K value. independent procedures. The M H E technique was used

{2hr~ Vol. 34, No. 5-8, S e p t e m b e r / O c t o b e r 1992 237


to determine the partition coefficient of MEK at 70 ~
and the result compared with that obtained by VPC. For
a quite different approach the partition coefficient of
cyclohexane in squalane was determined by VPC and A
compared with the corresponding literature value.

3 354:3 I~'~

Linear Regression Calculation by the M H E Technique Flllht

The value of K for MEK in water-air at 70 ~ deter- B


mined by VPC, was compared with that obtained by the
M H E technique. The fraction of the gas phase removed
from a pressurized headspace vial during the M H E pro-
cedure depends on the pressure difference between the
headspace pressure Ph and the final lower pressure, for
which atmospheric pressure Pa is usually assumed ac-
cording to Eq. (10). If this process of venting the pres- Figure 1
surized headspace gas is repeated several times the Determination of water-air partition coefficient (K = 46.5)otl011~
decrease in the peaks is described by the slope q at 70 *C by a 4-step MHE procedure according to Eq. t j i~'tle5
(q = An+l/An). The relationship of q to K, the phase derived from different slopes q (q = An,l/An) of regressi~
ratio, 13, and the pressure difference between the pres- from two vials with different samples volumes but containing~'
sure of the pressurized headspace vial Ph and the end total amount (3/aL) of MEK.
pressure Pa, is given by Eq. (10): A VsA = 1.0 mL, VGA = 21.3 mL, qa = 0.815, 1~a = 21.3
B Vs8 = 5.0 mL, Von = 17.3 mL, q~ = 0.959, [5n = 3.46
( K / ~ ) + (Pa/Ph)
q = (K/I~) + 1 00) o,ds
small K values (K < 100). The slope, however, dep~o
K can be derived if the absolute values of both Pa and Ph not only on K but on 15 as well and for systems wl~.
are known. Usually atmospheric pressure is assumed for high K value, e.g. alcohols-water, a lower VS is reC~'~
Pa. Onda et al. [14] have used this absolute procedure to mended to decrease q. This can be done without ~"~
determine distribution coefficients of flavor compounds loss in headspace sensitivity [1].
in solid coffee samples. Aqueous samples, however, In contrast to VPC, where quantitative analysis bY~.~e
have to deal with the problem that the end pressure ternai standard calibration is carried out using a v~os
after venting the pressurized gas phase in the vial is the vapor standard, the M H E technique applied here ~'~
prevailing water vapor pressure rather than atmos- two aqueous samples with different phase ratios.
pheric pressure. Since other parameters such as time of
venting, may also affect the volume of the expanded gas
phase due to the internal water vapor pressure, the cal-
culation based on Pa/Ph was not considered reliable Comparison with Literature Values
enough. These problems with internal water vapor pres- 9 I b~sis
sure can be eliminated if, again, two determinations are Although partition coefficients are the phys~ca d"
carried out under the same instrumental conditions. and thus the most important parameter for any heo~
The reference vial must contain an aqueous sample to space application, it is surprising to find little publiS~us
avoid any additional corrections for the water vapor at data9 The reason might be that most determinatt~
the prevailing temperature. Two equations are obtained based on headspace measurements have been carrt~
. J {fl~
with the same Pa/Ph, while the phase ratio 1~ is varied. out usmg a manually-operated, gas-tight syringe ano ,t
Two vials with 1.0 mL and 5.0 mL water respectively measurements have thus mainly been carried out [.
were prepared to which the same volume of 3.0 pL room temperature. Most practical headspace aPpllC.~.
MEK was added. The result of a 4 step M H E procedure tions, however, are nowadays carried out with attt
is shown in Figure I together with the analytical and in- mated headspace samplers and at elevated tenaper~"
strumental conditions. From the slopes of both sere/log tures to increase sensitivity. The data are, ther ef~
regression lines the partition coefficient was determined hardly comparable.
by Eq. (10) with the same pressure drop (Pa/Ph) for both A different technique has been used here to test th.e.~d0
measurements. K = 46.5 was obtained, which agrees curacy of VPC with proven literature values. Partlti A
fairly well with those from VPC (47.7, +/- 0.7 % C.V and coefficients have been used for this purpose deteraaiPT~
48.2 by linear regression). This agreement is surprisingly by chromatographic measurements from retention dat,.'
good, considering the small difference in the slopes, par- Very precise measurements were carried out in the e~tfl)
ticularly the very slow decrease in the 5 mL sample with days of GC by Porter, Deal and Stross t[151 9 us1 g..|t'
q = 0.96, where the differences in two successive areas squalane column. From their data the partition coefA
-- t0
approaches the statistical variation. This, therefore, cient of cyclohexane in squalane has been cornPaf,, s
demonstrates that the M H E approach works well in with that obtained by VPC. The conditions and rest~t~
principle, but is restricted to partition systems with are listed in Table II.

238 Chromatographia Vol. 34, No. 5-8. September/October 299~ :


"rahleII. Partition coefficients K of cyclohexane in squalane. 8
Ln K
~ [15]~ headspace gas chromatography* t 7
1~5 KK=ll2 K = 109.6, +/- 0.4 % C.V., N = 3 6
*C = 60.1 K = 61.2, +/- 4.2 % C.V., N = 3
5
Itgas .Phase: helium.
~rad~Pace gas chromatozraphy by vapor phase calibration ac- 4
10~otng !o Eq. (9); system: "2.6mL squalane + 2.0 ~L solution
Stala.~':ycl~ in toluene in 22.3 mL headspace vial, reference
air, uara: 2.0 ~tL eyclohexane solution in empty vial, gas phase: 3

cTyhc0agreement b e t w e e n the partition coefficients for


n: aexane in s q u a l a n e d e t e r m i n e d by different tech- 4'0
3'.z
sb
t
6'0
~
3.o
7'0
1.9
8b
i
3
toc]
3
[ ozra
t~ques is SUrprisingly good, thus showing that the V P C
eehrlique is in fact a reliable m e t h o d . Figure 2
Temperature function of water-air partition coefficient K of ethanol
(1), MEK (2), and diehloromethane (3) from 40 ~ - 80 *C by linear
TenlPerature D e p e n d e n c e of Partition Coefficient regression: y = ax + b with y = In K, x =lfI" (T in K), a = A/R
and b = -B according to Eq. (11).
~he d e p e n d e n c e of K on t h e a b s o l u t e t e m p e r a t u r e , T, is
g~vea [7] by Eq. (11): ethanol(l) MEK (2) CHzCIz(3)
a 5083 4298 2778
laK- A b -9.02 -8.62 -7.14
RT
B (11) r(corr, eoeff.) 0.9992 0.9994 0.9902

f~erttransformation (cf. Figure 2) K can be d e r i v e d for *conditions: sample volume Vs = 3.0 mL, vial volume Vv = 22.3 mL,
ern- m p e r a t u r e b y linear regression, at least in the 3.0/.tL mixture of three compounds added.
sev~erature r a n g e w h e r e K has b e e n d e t e r m i n e d at
ro, er.al t e m p e r a t u r e s . A n e x a m p l e of the linear
~ a h o n s h i p in a s e m i l o g plot is s h o w n in Figure 2 with
t h e V P C t e c h n i q u e with linear r e g r e s s i o n calculations.
~ a r t o l , M E K a n d d i c h l o r o m e t h a n e at five t e m p e r a -
A m i x t u r e o f these c o m p o u n d s was p r e p a r e d and 1.0,
~es in the r a n g e 40 ~ - 80 ~
3.0 and 5.0 p L w e r e injected into 5.0 m L of d e i o n i z e d
water. T h e resulting partition coefficients are listed in
T a b l e III. T h e t e m p e r a t u r e d e p e n d e n c e was deter-
SUPplemen t mined. B o t h coefficients a a n d b f r o m the linear regres-
sion, as s h o w n in Figure 2, are also listed to allow recal-
~v~ ition coefficients of several volatile c o m p o u n d s
culation for o t h e r t e m p e r a t u r e s .
tier" Wtdely differing solubilities in w a t e r h a v e b e e n
errained in t h e t e m p e r a t u r e r a n g e 40 ~ to 80 ~ by

Table III. Partition coefficients of selected compounds in water-air system.

Partition coeff. K regression coeff.

compound 40 *C 60 *C 70 ~ 80 *C b* a*

tetrachloroethene 1.48 1.27 0.78 0.87 -5.13 1736


1,1,1-trichloroethane 1.65 1.47 1.26 1.18 - 2.52 950
o-xylene 2.44 1.31 1.01 0.99 -7.61 2647
cyclohexane 0.07 0.05 0.03 0.02 -12.9 3228
n-hexane 0.14 0.043 0.012 - -28.6 8375
ethyl acetate 62.4 29.3 21.8 17.5 - 7.28 3563
n-butyl acetate 31.4 13.6 9.82 7.58 - 9.27 3971
toluene 2.82 1.77 1.49 1.27 -6.03 2198
isopropyl alcohol 825 286 179 117 - 10.6 5413
methyl isobutyl ketone 54.3 22.8 16.2 11.8 - 9.56 4237
dioxane 1618 642 412 288 - 7.94 4798
n-butanol 647 238 144 98.9 -10.2 5234

*cf. Figure 2.

Qhr~ Vol. 34, No. 5-8, September/October 1992 239


List of Symbols and Abbreviations References
Ac Peak area of ana~yte from external vapor standard [1] L.S. Ettre, B. Kolb, Chromatographia 32, 5 (1991).
As Peak area of analyte from headspace sample /2] L. Rohrschneider, Anal. Chem. 45, 1241 (1973)- :,~
ac slope of regression of analyte from external vapor [3} W.J. Jones, M.J. Egoville, E.O. Strolle, E.S. DellamOn~'
standard J. Chromatogr. 455, 45 (1988).
[4] T.G. Kieckbusch, C.J. King, J. Chrornatogr. Sei. 17, 273
as slope of regression of analyte from sample vial (1978).
Cs Concentration of analyte in sample [5] CA. McAuliffe, Chem Tech, 46-51, Jan. 1971.
Co Concentration of analyte in gas phase of sample [6] A.G. Vitenberg, B.V Ioffe, Z.S. Dimitrova, t.L. Btaae~
vial J. Chromatogr. 112, 319 (1975). ja~lalr
C o C Vapor concentration of analyte in external vapor [7] B . V..Ioffe,
. . A G Vitenberg, Head-Space Analysis an~ ~."~c,',.
Methods ~,nGas Chrot~atography. John Wiley & 5on~,
standard
K Partition coefficient York, NY, 1984. Chromatogr.~gl,
[8] R. Guitart, A. Puigdemont, M. Arboix, J.
m ~ Total amount of analyte in vial 271 (1989). ,det
m o Amount of analyte in gas phase [9] P Vospisil, B. Kolb, paper presented at the "At~ce'_'0r
ms Amount of analyte in sample Kolloqu[um tiber die Gaschromatographische Damplr~
analyse", l~ad Nauheim, 3.-4. October 1983.
Pa Atmospheric pressure [10] B. Kolb, L.S. Ettre, Chromatographia 32, 505 (1991)'
Ph Headspace pressure in pressurized vial (Ph = Pa + [11] B. Kolb, J. Chromalogr. 112, 287 (1975). ,-,~,001~'
PGC) with POC as excess pressure given by pres- [12] B. Kolb, in: B. Kolb (Ed,), Applied Headspace Oas~;080/.
sure of G C tograpy. Heyden & Son Ltd., New York, NY, pp. 1-tl t~',55,
[13~ K. Schoene, J. Steinhanses, A. KOnig, J. Chrornatog"
VG Volume of gas phase (headspace) in vial 57 (1988).
Vs Volume of sample in vial [14] N. Onda, A. Shhwhara, 1-1. Ishii, A. Sato, HRc&CC 14,
V v Volume of headspace vial 357 (1991). Sor ~'a,
13 Phase ratio ([~ = VGIVs) [~5] P.E. Porter, C.H. Deal, EH. Stross, J. Am Chem. '
@ Calibration factor 2999 (1956).

Received: April 1,1~2


Revised manuscript~ ~
received: May 8, ~9i9~
Accepted: May 15,

240 Chromatographia Vol. 34, No. 5-8, September/October 199~

You might also like