Professional Documents
Culture Documents
Calibration & Alignment Manual
Calibration & Alignment Manual
Calibration &
Alignment Manual
Table of Contents
Requirements 4-171
Welcome!
Quality Vision International, Inc. (QVI) has produced this Calibration & Alignment
Manual as part of its ongoing effort to provide users with useful, comprehensive
documentation. This manual has been developed using documentation standards
and a design that enhances readability and makes information easier to find.
Section Description
System Calibration & Describes how to calibrate and align the system.
Alignment
Warns you of the possibility of personal injury due to electric shock when
performing a task related to the subject matter of this manual.
Warns you of the possibility of personal injury related to the subject matter
of this manual.
Alerts you to the potential for damage due to electrostatic discharge. Special
instructions may be included to minimize this risk.
All ICE calibration functions are wizard-based to guide you step by step, and to
ensure consistency between users. ICE also includes diagnostic tools to help
monitor system status and troubleshoot system performance.
Access to calibration functions using ICE is restricted to protect the integrity of the
system, and to prevent unauthorized users from attempting calibration. Authorized
users have the ability to perform routine calibrations.
NOTE: This manual is current with release 4.0.x of the ICE software.
Key Terms
l Artifact
l Calibration
The process of determining the deviation from a standard and making the
necessary adjustments or applying correction factors so that a measuring
machine meets the standard.
A point within the measuring volume of the machine that all sensors are offset
from. If the machine has volumetric corrections enabled, the CSRP is the
reference point for the machine error map. See page 4-116 for more
information.
The absolute datum point where the X, Y, and Z axes positions in absolute
machine coordinates are all zero.
For a given set of tools on a probing system, this is the reference point for
data gathering.
o Optical Sensor The intersection of the optical centerline (which is
aligned to the center pixel of the camera) and the focal plane.
o Touch Probe Sensor The location of the center of the calibrated
probe tip.
o Laser Sensor The intersection of the optical axis and the center of
the capture range.
The signed distance (X,Y,Z) from the Measuring Point (MP)of each sensor to
the Common Sensor Reference Point (CSRP).
A sensor type supporting one or more interchangeable tools. Each tool has a
separate and unique MPO. Examples of different types of probing systems
are optics and touch probes.
A fixed point on the probing system where different tools mount. See page 4-
116 for more information.
o For an optical sensor, this is the center of the area where lenses are
attached.
o For a probe, this is the center of the end of the wand.
The signed distance (X,Y,Z) from the Probing System Interface (PSI) to the
Common Sensor Reference Point (CSRP).
l Qualifying a Sensor
The process of determining the XYZ offset from a sensor and tool
combination to the system reference point (CSRP). The system does not
consider a sensor and tool combination fully calibrated until it knows the
offset from the CSRP.
l Restore Point
l Tool
Any object that attaches to a sensor. For example, a replacement lens, lens
attachment, laser lens, or probe stylus.
l Verification
OGP offers customer service and technical support contracts that are tailored to
meet your specific needs and to protect the value of your investment.
OGP also offers Software Maintenance Agreements (SMA) which provide updates
of software products, application assistance, and discounts on purchases of
metrology software upgrades.
l Contact the QVI channel partner who provided your OGP product
l Visitwww.ogpnet.com
NOTE: Contact information for technical support is on the serial number label on
your OGPproduct.
For your own protection, back up the UCP database to removable, write-protected
media on a routine basis, and before performing any of the calibration or alignment
functions in ICE. This will allow you to restore a previous configuration of known
good system parameters and calibration data if inadvertent changes are made to the
system, or if the system controller suffers a hard drive failure.
NOTE: For information on how to restore a backup of the UCP database, see
"Restoring the UCP Database" in the ICE User Guide, which is accessible from the
Help menu in ICE.
2. If the Windows taskbar is not already displayed, press the Windows key on
the keyboard.
3. Right-click the UCP icon in the Windows notification area (to the right
of the Windows taskbar) and click Backup Database.
NOTES:
5. Click OK.
NOTES:
l We also recommend creating multiple backups, and store all backup files in a
secure place.
NOTE: It is assumed that you already started up the system, and the Portal
Navigation Panel is on the screen.
NOTE: This window only appears if power has not been removed from the
machine since the last time it was initialized.
NOTE: If you selected the Force Cold Start check box in the previous
window (shown on the previous page), the software automatically selects the
Home Stage-Zoom check box.
7. If the system has a touch probe sensor, do one of the following depending on
whether or not a stylus is attached:
l If a touch probe stylus is not attached to the touch probe sensor and a
stylus appears in the Tools list, either manually attach the correct
stylus to the sensor or select the Unload Current Tool check box
before continuing the initialization procedure.
The left panel of the ICE application window contains a tree view of components on
the system. Some of the items in the tree represent calibrations and others represent
physical hardware known to the system. Click any node to view information related
to that node.
Sensors/Tools node Under this node are listed all the sensors and tools known
to the system.
Sensor node Name of the sensor. Under this node are listed all the tools
available for use on the sensor.
l Optical sensors have two (2) levels: Sensor > Lenses
l Laser sensors have two (2) levels: Sensor > Laser lenses (in the case of
TTL lasers) or Laser tools (in the case of off-axis lasers)
l Touch probe sensors have four (4) levels: Sensor > Modules > Styli >
Tips
Tool node (bold text) Name of the tool currently attached to the sensor, and its
calibration level (see Calibration Levels for more information) double-click the
tool node and follow the on-screen instructions to detach the tool.
Change Racks node Under this node are listed all the change racks known to
the system, and the calibration status of each.
Stage node Under this node are listed all the stage calibrations, and whether or
not calibration data exists for each.
NOTES:
Access to ICE functions is restricted to protect the integrity of the system, and to
prevent unauthorized users from attempting calibration and making changes to the
system.
l Open Access (default) enables basic calibration and alignment functions that
may need to be performed on a routine basis, such as optics calibration,
camera alignment, sensor offset calibration, etc. No special access code is
needed for Open Access.
l Secure Access enables access to all stage and sensor calibrations, new
sensor configurations, and diagnostics. Secure access is enabled by entering
the ICE Secure Access code contained in the machine shipment notification
for the system.
l Full Access enables access to all ICE functions. It is intended for internal
QVI personnel only.
Log History
Configure Rotary
Configure Tool
Configure Artifacts
Configure Camera
Configure Sensor
(read (read
only) only)
Configure Corrections
(read (read
only) only)
Configure Machine
Configure Stage
Configure Zoom
Configure Lights
Configure Offsets
Configure PRL
Diagnostics (all)
Calibrate Stage
Seed Machine
NOTE: Any function not listed is assumed to be available for Open Access.
3. Click Change.
4. Wait for the Current Access indicator to change and stop flashing. Then close
the dialog box.
NOTE: For information on adding sensors and other accessories to the system, see
the ICE User Guide, which is accessible from the Help menu in ICE.
1. On the ICE Navigation Tree, right-click the node of the optical sensor, and
then select Configure > Add Tool.
Replacement lenses
Laser lenses
1. On the ICE Navigation Tree, right-click the node of the desired touch probe
sensor, and then select Configure > Add Body.
3. In the Name field, use the default name or type a different, unique name.
Example: TP20
4. Click Save.
5. Click the body you just added and then click Create on the toolbar.
a. To add a connector:
b. To add an extension:
For each tip, on its Tip tab, define the tip properties.
9. Click Save.
NOTE: Use the Laser Tool Configuration wizard to configure tools for an off-axis
laser (e.g., DRS laser and Rainbow Probe). To configure a laser tool (laser lens) for
the Flash TTL laser, use the Lens Configuration wizard.
1. On the ICE Navigation Tree, right-click the node of the off-axis laser sensor,
and then select Configure > Add Tool.
Example: DRS-300
l Ring gage
NOTE: ICE only supports the ground glass version of the QVI Alignment Reticle
(P/N 623970, revision G and higher).
1. On the ICE Navigation Tree, right-click the Artifacts node, and then select
Configure > Alignment Reticle.
Example: Alignment_Reticle
4. Click Finish.
1. On the ICE Navigation Tree, right-click the Artifacts node, and then select
Configure > Sphere.
3. In the Serial Number field, type the serial number of the artifact.
TIP: Set the shaft diameter to 10 mm, which will help make it easier to
calibrate a multi-tip probe.
1. On the ICE Navigation Tree, right-click the Artifacts node, and then select
Configure > Laser Reticle.
3. In the Serial Number field, type the serial number of the artifact.
4. Click Finish.
1. On the ICE Navigation Tree, right-click the Artifacts node, and then select
Configure > Rotary_Calibration.
Example: Rotary_Calibration
4. Set Gage Block Present to True, and enter the gage block thickness.
l Number of spheres
l Certified diameter
l Shaft diameter
l Shaft length
6. Click Finish.
1. On the ICE Navigation Tree, right-click the Change Racks node, select
Configure, and then select the type of change rack you want to add.
2. In the Name field, type a unique name for the change rack.
Example: 4_Port
3. Use the Add and Remove Last Port buttons to add/remove ports.
For example, you can assign touch probe styli to specific ports in the change
rack.
TIP: You may need to increase the Dust Cover Height value if the system is
unable to dock a particular probe.
2. On the ICE Navigation Tree, right-click the Rotaries node, and then click
Configure.
4. Click Initialize.
5. Click the Move Rotary tab, and then use the rotary move controls to move
the rotary to the desired home angle.
Current rotary angle you can use the knob perform an absolute move
Moves the indexer in the clockwise direction by the increment degree value
specified in the Degrees box
Moves the indexer in the counterclockwise direction by the increment
degree value specified in the Degrees box
Moves the indexer to the absolute degree value specified in the Degrees
box
8. Click Finish.
Requirements 4-171
l Properly configured QVI Laser Reticle (P/N 526685) laser systems only
l QVI Z Axis Calibration Gage (Step Gage) and its accompanying diagnostic
software routine
l Factory Data Backup disc (has an ENABLE.CAL file on it, which you will
need to verify XY using MeasureMind 3D or Measure-X)
Procedure Specification
Calibration Reports
l All reports are saved in HTML format, and can be viewed with your default
browser.
l All reports are saved in the following folder on the system controller:
C:\Users\Public\Public Documents\QVI\Reports
The calibration history log provides a summary of when each calibration task was
last performed and who performed the calibration. After you finish calibrating and
aligning the system, we recommend that you review the calibration history log to
confirm that you performed all the required calibrations.
To access the calibration history log, select File / Log History > Calibration
History.
To view the status of each applicable calibration for each sensor and tool
combination known to the system, click the Sensors/Tools node on the ICE
Navigation Tree.
l To view the latest calibration report for a specific calibration, click its View
button.
l To clear the existing calibration data for a specific calibration, click its Clear
button and then click Yes in response to the displayed confirmation message.
l Verify the active sensor icon appears next to the node of the
optical sensor. If necessary, double-click the node to activate the
sensor.
l If the 1X lens is not attached, double-click its node. Then follow the
on-screen instructions and click OK. (There may be multiple
instructions.)
4. Place a clean and properly configured QVI Alignment Reticle on the stage
glass.
Parfocal alignment sets the camera focus so the part image stays in focus through
the full zoom range, from high to low magnification.
When to Perform
Required Artifact
NOTE: It is assumed that you performed all the steps in the Before You Start topic
on page 4-14.
1. On the ICE Navigation Tree, right-click the node of the optical sensor, and
then select Alignment > Coaxial.
3. On the Controls tab below the video window, click Focus Target.
Be sure to adjust the illumination so you can see a clear image of the
surface without saturating any pixels.
Only perform this procedure after verifying that an adjustment needs to be made.
2. Remove the black tape covering the rectangular access slot in the optics
housing (see Figure 4-7). Do not discard the black tape. You will need to
reinstall it after you are done adjusting parfocal.
3. Insert a 2 mm Allen wrench into the rectangular access slot in the optics
housing, and into one of the holes in the zoom lens collar (see Figure 4-7).
Camera
Collar
2 mm Allen wrench
6. Cover the rectangular access slot in the optics housing with the black tape
removed earlier.
NOTE: You may want to leave the optics cover off until after you complete all
the required calibration and alignment procedures.
Use the Coaxial Alignment wizard to verify that the centerline of the video camera
is coincident with the centerline of the optics at all zoom positions.
During the coaxial alignment check, the software measures the movement of the
upper-right corner of the of the QVI Alignment Reticle throughout the entire zoom
range in each of the four quadrants of the field of view, starting in the upper-left
quadrant and ending in the lower-right quadrant. In each quadrant the motion is from
the corner toward the center as the optics zoom down.
At the end of the check, the software calculates the best-fit mechanical centerline by
comparing the position of the centerline of the camera with the best-fit centerline of
the optics throughout the entire zoom range.
When to Perform
Required Artifact
NOTE: It is assumed that you performed all the steps in the Before You Start topic
on page 4-14.
1. On the ICE Navigation Tree, right-click the node of the optical sensor, and
then select Alignment > Coaxial.
6. On the Controls tab below the video window, click Crosshair Target.
7. Align the upper-right corner of the reticle square to the Crosshair target.
Coaxial alignment changes could affect Flash TTL laser alignment (if equipped).
After adjusting coaxial alignment be sure to recheck all laser alignments.
Only perform this procedure after verifying that an adjustment needs to be made.
1. If you have not already done so, remove the optics cover.
2. Using a 3 mm Allen wrench, slightly loosen (do not remove) the three
mounting screws on top of the camera alignment plate so they are finger tight
only (see Figure 4-8).
Mounting screw
Figure 4-8: Video Camera Alignment Plate Mounting Screws, Top View
All adjustments are done at low magnification and in each case you need to
adjust the push/pull screws as described below and shown in Figure 4-9. You
may need to do this several times until the upper-right corner of the reticle
square is aligned.
l To move the image to the left or right, use the push/pull screws on the
side of the alignment plate, as shown in Part A and Part B.
l To move the image up or down, use the push/pull screws on the front
of the alignment plate, as shown in Part C and Part D. Be sure to
adjust both sets of push/pull screws equally to avoid rotating
the alignment plate.
NOTE: Use a 2.5 mm Allen wrench to adjust the pull screws and use a 2 mm
Allen wrench to adjust the push screws.
5. Retighten the three mounting screws on top of the camera alignment plate
(see Figure 4-8 on page 4-25).
6. Click Again in the Coaxial Alignment wizard, and perform the coaxial
alignment check again. If the alignment check fails, repeat Steps 2 through 5
to make the required adjustment.
7. Repeat Step 6 as many times as necessary until the coaxial alignment check
passes.
NOTE: You may want to leave the optics cover off until after you complete all
the required calibration and alignment procedures.
Use the Camera Rotational Alignment wizard to verify that the X axis of the
video camera is parallel to the X axis of the stage.
During the rotational alignment check, the software measures the Circle (dot) or Line
(top edge) using the specified number of points from the left edge to the right edge of
the field of view. You can also perform this check using the Interactive method,
which provides real-time feedback as you adjust the physical orientation of the
camera.
NOTE: The Interactive method is only available after you check rotational alignment
using either the Circle method or Line method at least once.
When to Perform
Required Artifact
NOTE: It is assumed that you performed all the steps in the Before You Start topic
on page 4-14.
1. On the ICE Navigation Tree, right-click the Optical Sensor node, and then
select Alignment > Rotational.
3. Click Next.
l Select the Line option (shown below) to measure the top edge of the
reticle square.
6. [Optional] Specify the number of points you want the system to measure.
Rotational alignment changes could affect Flash TTL laser alignment (if equipped).
After adjusting rotational alignment be sure to recheck all laser alignments.
Only perform this procedure after verifying that an adjustment needs to be made.
4. Click Next.
5. If you have not already done so, remove the optics cover.
6. Using a 3 mm Allen wrench, slightly loosen (do not remove) the three
mounting screws on top of the video camera alignment plate so they are
finger tight only. The screws are labeled 1, 2, and 3 in Figure 4-10.
When adjusting camera rotation, make very small changes and verify the
results. This is a very sensitive adjustment.
NOTE: Screw 1 should be slightly tighter than the other two, which allows the
video camera to pivot around screw 1.
10. Repeat Step 9 as many times as necessary until the camera rotation
alignment check passes.
NOTE: You may want to leave the optics cover off until after you complete all
the required calibration and alignment procedures.
The patented AccuCentric assembly inserts a reticle image into the optical path that
is used to recalibrate the optical system whenever you change the zoom position
(magnification).
When to Perform
1. On the ICE Navigation Tree, right-click the node of the optical sensor, and
then select Alignment > Coaxial.
Only perform this procedure after verifying that an adjustment needs to be made.
1. If you have not already done so, remove the optics cover.
2. Using a 1.5 mm Allen wrench, loosen (do not remove) the two setscrews that
secure the reticle housing to the assembly (see Figure 4-12).
Setscrews (Qty 2)
Reticle Housing
Polarizer
3. Slide the reticle housing in/out to focus the image of the AccuCentric LED
reticle.
4. Hold the reticle housing in place as you retighten the two setscrews.
TIP: If you are performing this procedure immediately after checking AccuCentric
reticle focus and you did not exit the Coaxial Alignment Helper, you can skip Steps 1
through 4.
1. On the ICE Navigation Tree, right-click the node of the optical sensor, and
then select Alignment > Coaxial.
4. Use the Acc LED control to increase the intensity of the AccuCentric LED
until you see a clear image of the AccuCentric reticle.
6. Resize the Double Circle target and observe whether or not the AccuCentric
reticle image is centered within the target.
Only perform this procedure after verifying that an adjustment needs to be made.
1. If you have not already done so, remove the optics cover.
2. Using a 1.5 mm Allen wrench, adjust the two centering setscrews in the
reticle housing until the image of the reticle appears centered within the
displayed target (see Figure 4-13).
Setscrews (Qty 2)
Reticle Housing
NOTE: You may want to leave the optics cover off until after you complete all
the required calibration and alignment procedures.
The optional grid projector is used to display a grid for measuring low contrast
features that are difficult to image. In order for this grid projection technique to work,
a focused grid image with high contrast is required.
When to Perform
Required Artifact
NOTE: It is assumed that you performed all the steps in the Before You Start topic
on page 4-14.
1. On the ICE Navigation Tree, right-click the node of the optical sensor, and
then select Alignment > Coaxial.
3. On the Controls tab below the video window, click Focus Target.
6. Position the Focus target completely over the glass, with no part of the reticle
square inside the target.
Only perform this procedure after verifying that an adjustment needs to be made.
1. If you have not already done so, remove the optics cover.
2. Turn off the grid projector by clearing the Grid check box in the video window.
3. Position the Focus target over the upper-right corner of the reticle square.
4. Turn off the surface (coaxial) light, and set the backlight to approximately
50%.
6. Click the Z-axis readout at the bottom of the video window to zero the
readout.
7. Turn off the backlight, and set the surface (coaxial) light to approximately
50%.
8. Select the Grid check box in the video window to turn on the grid projector.
10. Using a 2 mm Allen wrench, adjust the grid projector focus adjustment screw
until the grid image is focused. When finished, retighten the grid projector
lock-down screws.
11. Check the grid projector focus to verify it is within specification. If necessary,
repeat this procedure to make the required adjustments.
12. Repeat Step 11 as many times as necessary until the grid projector focus is
within specification.
NOTE: You may want to leave the optics cover off until after you complete all
the required calibration and alignment procedures.
If your system has the optional TTL laser pointer, perform the procedure below to
check its alignment to the optical system. You will need a blank sheet of white
paper.
NOTE: It is assumed that you performed all the steps in the Before You Start topic
on page 4-14.
2. On the ICE Navigation Tree, right-click the node of the optical sensor, and
then select Alignment > Coaxial.
4. Manually focus the surface of the paper using surface (coaxial) light.
5. On the Controls tab below the video window, click Focus Target.
8. On the Controls tab below the video window, click Double Circle
Target.
9. Turn on the laser pointer by holding down the button on top of the joystick.
l If the laser spot is not centered, perform the laser pointer adjustment
procedure.
1. If you have not already done so, remove the optics cover.
2. Using a 2.5 mm Allen wrench, loosen (do not remove) the three laser pointer
lock-down screws (see Figure 4-15).
3. While holding down the button on top of the joystick, center the image of the
laser spot in the Image window by adjusting the laser pointer centering
screws (see Figure 4-15). Use a 1.5 mm Allen wrench.
4. After centering the image of the laser spot, release the button on top of the
joystick and tighten the three lock-down screws.
5. Recheck the laser pointer alignment to verify the image of the laser spot is
still centered within the Double Circle target. If necessary, repeat this
procedure to realign the laser pointer.
6. Repeat Step 5 as many times as necessary until the laser sport is centered in
the video window.
7. When finished, reinstall the optics cover and remove the sheet of paper from
the stage glass.
NOTE: You may want to leave the optics cover off until after you complete all
the required calibration and alignment procedures.
Use the TTL Laser Alignment wizard to check (and adjust if necessary) the
alignment of the Flash TTL laser (if equipped) to the optics.
When to Perform
Required Artifact
1. Switch to Support Access if you have not already done so. (See Changing
the Access Level for instructions.)
l Verify the active sensor icon appears next to the node of the TTL
laser sensor. If necessary, double-click the node to activate the
sensor.
l If the laser tool (laser lens) you want to use is not attached, double-
click its node. Then follow the on-screen instructions and click OK.
(There may be multiple instructions.)
3. On the ICE Navigation Tree, right-click the node of the Flash TTL laser
sensor, and then select Alignment > QVI Laser.
5. Click Next.
6. Systems with a grid a projector: Position the Focus target completely over
the reticle square.
9. Click Next.
10. Wait for the system to turn on the grid projector (if equipped) and auto adjust
the lights.
11. Click in the video window to perform an autofocus. Then click Enter to
accept the focus point and enable the Next button.
l If the laser spot is not centered on the Crosshair target, perform the
Laser Spot Centering Adjustment steps on the next page.
a. Using a 2.5 mm Allen wrench, loosen (do not remove) the laser pull
screws shown in Figure 4-16 one full turn.
b. Using a 2 mm Allen wrench, adjust the laser push screws to center the
laser spot on the Crosshair target.
c. Tighten the laser push and pull screws until snug. Do not
overtighten.
d. Verify the laser spot remains centered on the Crosshair target at high
and low magnification. If necessary, repeat the above adjustment
steps.
e. Apply LOCTITE 242 to the screw heads to prevent the screws from
loosening.
l If the laser spot does not remain centered on the vertical centerline,
perform the Laser Spot Drift Adjustment steps on the next page.
a. Using a 2.5 mm Allen wrench, loosen (do not remove) the three laser
alignment plate mounting screws just enough to be able to turn the
vertical adjustment screws (see Figure 4-17).
b. Adjust the position of the laser spot as necessary using the vertical
adjustment screws.
l To move the laser spot to the left, turn the vertical adjustment
screws to the right (clockwise direction).
l To move the laser spot to the right, turn the vertical adjustment
screws to the left (counterclockwise direction).
d. Verify the laser spot remains centered on the vertical centerline of the
Crosshair target as you raise and lower the optical assembly. If
necessary, repeat the above adjustment steps.
e. Click Back and repeat Step 13 to verify the laser spot is still centered
on the Crosshair target. Readjust if necessary.
17. Systems with a grid projector: Perform a grid focus on the surface of the
reticle square.
l If the laser range indicator is not centered within the laser range,
perform the Laser Zero Adjustment steps on the next page before
continuing.
a. Adjust the laser detector push/pull screws to center the laser range
indicator within the laser range (see Figure 4-18). Typically, only minor
adjustments need to be made.
NOTE: The interactive laser range tool on the Controls tab updates in
real time as you adjust the laser detector push/pull screws.
24. Click Test to test the slope and deviation of the laser.
l If the test fails, perform the following adjustment and then click Test
again. Repeat until the test passes. Verify the laser spot is still
centered on the Crosshair target before clicking Finish.
o If the Slope is greater than 18,500 (for the 2X Laser Lens) or
105,000 (for the 5X Laser Lens), loosen (do not remove) the
three laser alignment plate mounting screws (see Figure 4-19)
and use the push/pull screws to move the laser approximately
0.33 mm toward the front of the machine. When finished,
retighten the laser alignment plate mounting screws.
o If the Slope is less than 18,500 (for the 2X Laser Lens) or
105,000 (for the 5X Laser Lens), loosen the three laser
alignment plate mounting screws (see Figure 4-19) and use the
push/pull screws to move the laser approximately 0.33 mm
toward the back of the machine. When finished, retighten the
laser alignment plate mounting screws.
Use the Center Pixel Calibration wizard to calibrate the centerline of the video
camera so it is coincident with the centerline of the optics at all zoom positions.
During calibration, the software measures the movement of the upper-right corner of
the QVI Alignment Reticle in each of the four quadrants of the field of view
throughout the zoom range.
When to Perform
Required Artifact
You may need to recalibrate all lenses and sensor offsets after performing this
procedure.
NOTE: It is assumed that you performed all the steps in the Before You Start topic
on page 4-14.
1. On the ICE Navigation Tree, right-click the node of the optical sensor, and
then select Calibrate > Center Pixel.
3. Click Next.
6. On the Controls tab below the video window, click Crosshair Target.
l Artifact is clean
Use the Optics Calibration wizard to calibrate the zoom lens and field of view at
each zoom position to ensure an edge can be accurately located anywhere within
the field of view. You need to do this for each configured lens.
NOTE: If optics calibration fails, the software saves calibration data for all calibrated
zoom positions and only calibrates the uncalibrated zoom positions when you restart
the calibration.
When to Perform
Required Artifact
1. Switch to the optical sensor and tool (lens) combination you want to
calibrate.
l Verify the active sensor icon appears next to the node of the
optical sensor. If necessary, double-click the node to activate the
sensor.
l If the optical tool (lens) you want to calibrate is not attached, double-
click its node. Then follow the on-screen instructions and click OK.
(There may be multiple instructions.)
2. On the ICE Navigation Tree, right-click the node of the attached lens, and
then select Calibrate > Optics.
4. Click Next.
l Find Z Focus Offset: Finds the Z focus offset so the system can
apply a parfocal correction at each zoom position
l Find Rough FOV: Finds the rough field of view instead of using an
estimated field of view
l Find Grid Projector Focus Offset: Finds the offset between a grid
focus and an optical focus during the measurement sequence
8. On the Controls tab below the video window, click Focus Target.
10. On the Controls tab below the video window, click Crosshair Target.
13. If you selected the Distortion option on the previous screen, select the zoom
positions you want to skip (if any).
NOTE: The highest zoom position is not listed because you cannot skip it.
NOTES:
l If the calibration failed, verify the following and then perform the
calibration again:
o Stage glass is clean
o Artifact is clean
o Artifact setup is correct (the upper-right corner of the reticle
square is in focus and it is aligned to the Crosshair target)
l If the calibration failed, but the system was able to calibrate the first
two zoom positions (starting at the highest), partial calibration data is
displayed and the Finish button is enabled so you can save the partial
calibration. Otherwise, the Finish button is disabled and you have to
perform the calibration again.
Use the Autofocus Calibration wizard to minimize the effect of any residual
optical astigmatism errors in the zoom lens assembly. If residual errors are not
corrected, Z values returned from an autofocus may vary depending on the type and
orientation of the part being measured.
Required Artifact
1. Switch to the optical sensor and tool (lens) combination you want to
calibrate.
l Verify the active sensor icon appears next to the node of the
optical sensor. If necessary, double-click the node to activate the
sensor.
l If the optical tool (lens) you want to calibrate is not attached, double-
click its node. Then follow the on-screen instructions and click OK.
(There may be multiple instructions.)
2. On the ICE Navigation Tree, right-click the node of the attached lens, and
then select Calibrate > Autofocus.
4. On the Controls tab below the video window, click Focus Target.
6. On the Controls tab below the video window, click Circle Target.
NOTE: The reticle dot does not have to fit entirely in the field of view. You
can select a point, move the XY transports, select another point, move the
transports again, and then select the third point.
Use the Focus Diagnostic wizard to evaluate the repeatability of the Focus target
or find the step size of the Focus target at the defined zoom positions.
This task requires Support Access or Secure Access (described on page 2-11).
Contact the QVI channel partner who provided your system for assistance.
When to Perform
Required Artifact
1. If you have not already done so, switch to Support Access or Secure
Access. (See Changing the Access Level for instructions.)
2. Switch to the optical sensor and tool (lens) you want to perform this
procedure with.
l Verify the active sensor icon appears next to the node of the
optical sensor. If necessary, double-click the node to activate the
sensor.
l If the optical tool (lens) you want to use is not attached, double-click
its node. Then follow the on-screen instructions and click OK. (There
may be multiple instructions.)
3. On the ICE Navigation Tree, right-click the node of the attached lens, and
then select Diagnostics > Focus.
4. Place the QVI Alignment Reticle on the stage glass. Secure the reticle so it
will not move.
7. Click Next.
9. To find the focus step size at each zoom position, type 10 in the second
Zoom Range box.
11. Wait for the measurement sequence to finish, and then click Save to save the
new step size values.
Be sure to click Save before clicking Finish or else the new focus step size
values will not be saved in the UCP database.
Since the Z-axis calibration function in ICE requires Secure Access (described on
page 2-11), it is not covered in this manual. However, if your system has
MeasureMind 3D or Measure-X installed, you can use either measurement software
to verify the measuring accuracy of the Z axis.
NOTE: For information on how to calibrate and verify Z using ICE, see "Z Scale
Resolution" in the ICE User Guide, which is accessible from the Help menu in ICE.
2. Use the Z Axis Calibration Gage, also known as a Step Gage, and its
accompanying diagnostic software routine to check and verify the linear
measuring accuracy of the Z axis. The documentation provided with the Z
Axis Calibration Gage describes how to set up and use the artifact.
If the verification is out of tolerance, you will need to recalibrate the Z axis.
Contact the QVI channel partner who provided your system for calibration
assistance.
Since the XY stage calibration function in ICE requires Secure Access (described on
page 2-11), it is not covered in this manual. However, if your system has
MeasureMind 3D or Measure-X installed, you can use either measurement software
to verify the measuring accuracy of the X and Y transports.
NOTE: For information on how to calibrate and verify XY using ICE, see "XY Stage
Calibration" in the ICE User Guide, which is accessible from the Help menu in ICE.
Stage calibration ensures maximum stage travel accuracy over the entire XY
measuring area by applying correction factors to compensate for normal wear over
time. The correction factors vary depending on the location of the X and Y
transports.
NOTE: If the verification fails, contact the QVI channel partner who provided your
system for calibration assistance.
1. Carefully place the 25 intersection grid reticle on the stage glass so the
chrome lines face up. Make sure the grid reticle is centered on the stage and
the line between Intersections #11 and #15 is parallel to the X axis.
5. Using the joystick, position Intersection #13 in the center of the Image
window.
11. Using the joystick, position each corner of the grid reticle in the Image
window and make sure Intersections #1, #5, #21, and #25 can be centered
within their field of views this ensures the grid reticle is within the stage
travel limits.
12. Verify the grid reticle is level by performing an autofocus on Intersections #1,
#5, #21, and #25. At each location, note the Found Z Location reading
displayed in the Image window. It should be as close as possible to 0.
15. Adjust (tram in) the rotation of the grid reticle so the horizontal line of the grid
reticle remains aligned to the Box target when you move the X transport from
side to side along the full length of X travel.
16. Repeat Step 11 to make sure the grid reticle is still within the stage travel
limits; adjust if necessary.
17. Align Intersection #13 to the Box target and repeat Steps 8 and 9 to refocus
the intersection.
18. Verify the grid calibration file and coordinates, as described on the next page.
After staging the grid reticle, you need to load the grid reticle calibration file. This file
contains the nominal grid reticle intersections. The name of the file is
GRIDxxxx.DAT, where xxxx is the grid serial number. You will have to load the file
from the disc or USB drive that came with the grid reticle, or you can create a new
grid file as described on page 4-97.
NOTE: If the grid reticle does not have an accompanying file defining X and Y
coordinate values or accompanying literature, contact OGP or the QVI channel
partner who provided your system.
To verify the existence of the grid calibration file and the accuracy of the grid
coordinates in the file, and to begin the verification, follow the steps below.
2. Insert the disc that came with the grid reticle into the system controller or plug
in the USBdrive if one was supplied.
3. Make sure the measurement units in the software match the units used in the
table on the printed sheet provided with the grid reticle; change the units in the
software if necessary.
The software searches for the specified grid calibration file on all available
drives on the system controller.
l If the system finds the grid calibration file, it displays a dialog box
(similar to the one shown below), which lists all the nominal X and Y
coordinates.
o Verify the accuracy of the displayed coordinates by comparing
them to the coordinates on the printed sheet provided with the
grid reticle.
o If necessary, change any incorrect coordinates in the dialog
box to match those listed on the printed sheet.
o After verifying the accuracy of the grid calibration file, select
the tolerance specifications and output options (see page 4-98).
l If the system cannot find the grid calibration file, you can create a new
grid calibration file, as described on the next page.
If the system cannot find the specified grid calibration file, something similar to the
following appears:
Click Yes, and then follow the steps below to create a new grid calibration file.
1. Refer to the literature that came with the grid reticle, and enter the correct X
and Y coordinates for each intersection. Values can be entered in inches or
millimeters depending on which units you selected in the software.
2. Verify the accuracy of the coordinates in the dialog box by checking them
against the coordinate values on the printed sheet.
After entering and/or verifying the grid coordinates, you must specify either an E2
tolerance specification (recommended) or an XY tolerance specification. You can
also indicate whether the measured data is output to a printer and/or statistics file.
The software displays a dialog box with a position repeatability value and a
distance uncertainty value. The values displayed are based on the
characteristics of the machine when it was calibrated at the factory.
3. Click OK.
When you run the verification routine, the software calculates the straight line
distance from each intersection to every other intersection. This calculation is based
on the formula: A + [L (mm) / B], where:
The software displays a dialog box with X and Y location tolerance values.
The values displayed are based on the characteristics of the machine when it
was calibrated at the factory.
3. Click OK.
When you run the verification routine, the system calculates the deviation
between the nominal and actual X and Y locations for each intersection.
To output all measured intersections to a printer, select the Print check box in the
Enter Grid Coordinates dialog box. The printout is similar to the routine output;
however, it provides information about each intersection rather than each step.
To output all measured intersections to a statistics file, select the Stats check box in
the Enter Grid Coordinates dialog box.
The system displays a dialog box for you to specify the statistics file name if the
statistics destination is File. The software outputs the desired values using the
current statistics templates and tags. (The software looks for the statistics
configuration file NLSTATS.CFG and loads it if it is found; otherwise it loads the
statistics configuration from INPUT.CFG.) The intersection number or distance
number is output for the step number. The tag output is the tag for X or Y location, or
straight line distance for E2. However, if there is no data for a particular output
character, nothing is output. This is the case for routine name, run number, feature
comment and prompt and header lines.
If you want the system to refocus any failed intersections, select the Focus check
box in the Enter Grid Coordinates dialog box.
If you are using a dark grid reticle with light crosshairs, select the Inverse check box
to reverse the direction of the dark-to-light search. If you are using the standard grid
reticle, which has dark crosshairs, leave the Inverse check box unchecked.
NOTE: This procedure assumes the grid calibration file has been loaded and verified
to be accurate.
The Box target appears in the Image window and the following message
appears in the Measurement window: Set to Intersection 11
2. Using the joystick, position Intersection #11 in the Image window and make
the following adjustments:
a. If the Box target does not match the size of the reticle lines, resize the
target using the Target Spacing slider.
The system automatically measures the intersection and sets the axis
alignment. The system then measures Intersection #13 and resets XY zero.
This is the only time Intersection #13 is used to reset XY zero; subsequent
measurements of this intersection are an indication of machine repeatability.
4. Click No if you want to end the verification, or click Yes to perform the
verification procedure again.
This subsection describes how to use Measure-X to verify stage calibration. If you
want to use MeasureMind 3D instead, see Verifying Non-Linear XY Stage
Calibration Using MeasureMind 3D on page 4-93.
Stage calibration ensures maximum stage travel accuracy over the entire XY
measuring area by applying correction factors to compensate for normal wear over
time. The correction factors vary depending on the location of the X and Y
transports.
NOTE: If the verification fails, contact the QVI channel partner who provided your
system for calibration assistance.
1. Carefully place the 25 intersection grid reticle on the stage glass so the
chrome lines face up. Make sure the grid reticle is centered on the stage and
the line between Intersections #11 and #15 is parallel to the X axis.
5. Using the joystick, position Intersection #13 in the center of the Image
window.
11. Using the joystick, position each corner of the grid reticle in the Image
window and make sure Intersections #1, #5, #21, and #25 can be centered
within their field of views this ensures the grid reticle is within the stage
travel limits.
12. Verify the grid reticle is level by performing an autofocus on Intersections #1,
#5, #21, and #25. At each location, note the Found Z Location reading
displayed in the Image window. It should be as close as possible to 0.
15. Adjust (tram in) the rotation of the grid reticle so the horizontal line of the grid
reticle remains aligned to the Box target when you move the X transport from
side to side along the full length of X travel.
16. Repeat Step 11 to make sure the grid reticle is still within the stage travel
limits; adjust if necessary.
17. Align Intersection #13 to the Box target and repeat Steps 8 and 9 to refocus
the intersection.
18. Verify the grid calibration file and coordinates, as described on the next page.
After staging the grid reticle, you need to load the grid reticle calibration file. This file
contains the nominal grid reticle intersections. The name of the file is
GRIDxxxx.DAT, where xxxx is the grid serial number. You will have to load the file
from the disc or USB drive that came with the grid reticle, or you can create a new
grid file as described on page 4-106.
NOTE: If the grid reticle does not have an accompanying file defining X and Y
coordinate values or accompanying literature, contact OGP or the QVI channel
partner who provided your system.
To verify the existence of the grid calibration file and the accuracy of the grid
coordinates in the file, and to begin the verification, follow the steps below.
1. Start Measure-X.
5. Click Open.
NOTE: The dialog box shown above lists the XY coordinates of every
intersection on the grid, with Intersection #13 having XY coordinates of (0,0).
7. On the printed sheet provided with the grid reticle, locate the XY coordinate
table for the grid reticle you are using. Then verify the accuracy of the
coordinates in the dialog box by checking them against the coordinates on the
printed sheet.
If your grid reticle does not have an accompanying file defining X and Y coordinates,
but does have accompanying literature, you can create a new grid calibration file and
manually enter the appropriate X and Y coordinates.
NOTE: If the grid reticle does not have an accompanying file defining X and Y
coordinate values or accompanying literature, contact OGP or the QVI channel
partner who provided your system.
1. If you have not already done so, select System / Calibration > Non-linear
> XY to display the Automatic XY Non-linear Calibration window.
2. Click Open.
3. In the File name box, type gridxxxx.dat, where xxxx is the serial number on
your grid reticle.
4. Click Open.
The system displays the Non-Linear Calibration Grid File dialog box. The XY
coordinate values for every intersection are (0,0).
6. Refer to the literature that came with the grid reticle, and enter the correct X
and Y coordinates for each intersection.
7. Verify the accuracy of the coordinates in the dialog box by checking them
against the coordinate values on the printed sheet.
8. Click OK.
After entering and/or verifying the grid coordinates, you must specify either an E2
(recommended) or XY tolerance specification.
This enables the Position Repeatability and Distance Uncertainty boxes. The
displayed values are based on the characteristics of the system when it was
calibrated at the factory.
When you run the verification routine, the system calculates the straight line
distance from each intersection to every other intersection. The calculation is
based on the formula A = (L (mm) / B), where:
This enables the X Tolerance and Y Tolerance boxes. The displayed values
are based on the characteristics of the system when it was calibrated at the
factory.
When you run the verification routine, the system calculates the deviation
between the nominal and actual X and Y locations for each intersection.
NOTE: It is assumed that you already staged the grid reticle, loaded the grid
calibration file, and verified the coordinates in the grid calibration file.
1. Verify the grid calibration file and select the tolerance specification as
described on the previous pages. Then select the desired options in the
Automatic XY Non-linear Calibration window.
l If you are using a dark grid reticle with light crosshairs, select the
Inverse check box to reverse the direction of the dark-to-light search.
If you are using the standard grid reticle, which has dark crosshairs,
leave the Inverse check box unchecked.
2. Click Start.
3. Using the joystick, position Intersection #11 (left column, middle row) in the
center of the Image window.
6. Click OK.
The system measures the intersection, sets X and Y to zero, drives to the
nominal location of Intersection #15, and displays the following:
If this occurs, make the necessary corrections and click Retry to remeasure
the intersection.
8. Click OK.
l If you selected the Print report when done check box in Step 1 of
this procedure, the software sends the verification report to the printer
automatically.
NOTES:
9. Close the verification report. The system displays a prompt to repeat the
verification. Click Yes to repeat it or click No to end it.
As you perform the XY verification (not E2), results from the verification appears at
the bottom of the Automatic XY Non-linear Calibration window. This provides useful
data about the individual intersection points in the grid.
l Green = in tolerance
To calibrate a tool, you must first "teach" the system the location of an appropriate
artifact by measuring it with a tool that has already been calibrated. In most cases
the 1X lens is used to locate the artifact, but you can use any calibrated tool.
Once the system knows the exact location of the artifact, you then need to
remeasure the artifact (without moving it) with the "uncalibrated" tool to determine
the sensor offset, which is then stored in the UCP database.
After locating the artifact with the calibrated sensor and tool combination, do not
move the artifact before measuring it with the unqualified sensor and tool
combination. Doing so will result in an inaccurate sensor offset.
Required Artifact
The required artifact for sensor offset calibration varies depending on the tool you
want to use to locate the artifact and the tool you want to calibrate. You need to use
an artifact that can be measured by both tools.
-OR-
Ring gage
Optical tool (lens) Laser tool (TTL or off-axis QVI laser reticle
laser)
Touch tool (stylus) Laser tool (TTL or off-axis QVI calibration sphere
laser)
-OR-
Ring gage
The Common Sensor Reference Point (CSRP) is a point within the measuring
volume of the machine that all sensors are offset from.
NOTE: If the machine has volumetric corrections enabled, the CSRP is the
reference point for the machine error map.
The diagrams on the following pages illustrate the relationship between the CSRP
and sensor offsets.
The calibration level of each sensor and tool combination indicates whether or not
the offset from the CSRP is known. The system does not consider a sensor and
tool combination fully calibrated until it knows the offset from the CSRP.
1. On the ICE Navigation Tree, right-click the Sensors/Tools node, and then
select Calibrate > Sensor Offsets.
2. Switch to the calibrated optical tool (typically, the 1X lens) you want to
use to locate the artifact.
l Verify the active sensor icon appears next to the node of the
optical sensor. If necessary, double-click the node to activate the
sensor.
l If the calibrated optical tool (lens) you want to use is not attached,
double-click its node. Then follow the on-screen instructions and click
OK. (There may be multiple instructions.)
5. Click Next.
7. Position the upper-right corner of the reticle square in the field of view.
Manually focus the video image.
TIP: If the artifact has not been moved since the last time it was used to
calibrate the sensor offset, you can click Set Position to position the sensor
automatically. (The Set Position button only appears if location data for the
artifact exists.)
If you are unsure whether or not the artifact has been moved, do not click
Set Position position the sensor manually to be safe.
9. On the Controls tab below the video window, click Focus Target.
10. Perform an autofocus on the upper-right corner of the reticle square using
backlight.
11. On the Controls tab below the video window, click Crosshair Target.
l Verify the active sensor icon appears next to the node of the
optical sensor. If necessary, double-click the node to activate the
sensor.
l If the optical tool (lens) you want to calibrate is not attached, double-
click its node. Then follow the on-screen instructions and click OK.
(There may be multiple instructions.)
18. Select the Calibrate option and leave the default settings as-is.
21. Perform an autofocus on the upper-right corner of the reticle square at high
magnification using backlight.
22. On the Controls tab below the video window, click Crosshair Target.
26. If you are done calibrating the offsets for your lenses, click Finish.
1. Verify the configuration settings for the calibration sphere you want to use.
b. Click Edit.
l Serial number
l Material
l Certified diameter
l Shaft diameter
d. Click Finish.
3. Switch to the calibrated optical tool (typically, the 1X lens) you want to
use to locate the artifact.
l Verify the active sensor icon appears next to the node of the
optical sensor. If necessary, double-click the node to activate the
sensor.
l If the calibrated optical tool (lens) you want to use is not attached,
double-click its node. Then follow the on-screen instructions and click
OK. (There may be multiple instructions.)
4. Select the Calibration Sphere you want to use during the calibration.
6. Click Next.
TIP: If the artifact has not been moved since the last time it was used to
calibrate the sensor offset, you can click Set Position to position the sensor
automatically. (The Set Position button only appears if location data for the
artifact exists.)
If you are unsure whether or not the artifact has been moved, do not click
Set Position position the sensor manually to be safe.
9. On the Controls tab below the video window, click Focus Target.
10. Perform an autofocus on the right edge of the calibration sphere using
backlight.
11. On the Controls tab below the video window, click Crosshair Target.
l If the touch tool you want to calibrate is not attached, double-click its
node. Then follow the on-screen instructions and click OK.
18. Select the Calibrate option and leave the default settings as-is.
20. Position the tip of the touch tool directly above the top of the calibration
sphere as shown.
22. If you want to calibrate the offset for another touch tool, click Again. Then
click Yes in response to the displayed message (if it appears) and go back to
Step 15.
23. If you are done calibrating touch tool offsets, click Finish.
l Verify the active sensor icon appears next to the node of the TTL
laser sensor. If necessary, double-click the node to activate the
sensor.
l If the laser tool (laser lens) you want to use is not attached, double-
click its node. Then follow the on-screen instructions and click OK.
(There may be multiple instructions.)
2. On the ICE Navigation Tree, right-click the node for the attached TTL laser
tool (laser lens), and then select Calibrate > Laser Range.
4. Click Next.
5. Wait for the system to turn on the grid projector (if equipped) and auto adjust
the lights.
9. Click Enter.
NOTE: If the calibration fails, verify the following, and then perform the
calibration again:
l Verify the reticle is resting directly on the stage glass (must be parallel
to the stage glass).
l Verify the laser is properly aligned (see Flash TTL Laser Alignment on
page 4-52.)
1. On the ICE Navigation Tree, right-click the Sensors/Tools node, and then
select Calibrate > Sensor Offsets.
2. Switch to the calibrated optical tool (typically, the 1X lens) you want to
use to locate the artifact.
l Verify the active sensor icon appears next to the node of the
optical sensor. If necessary, double-click the node to activate the
sensor.
l If the calibrated optical tool (lens) you want to use is not attached,
double-click its node. Then follow the on-screen instructions and click
OK. (There may be multiple instructions.)
3. Select the Laser Reticle you want to use during the calibration.
4. Place the selected Laser Reticle on the stage glass. Secure the reticle so it
will not move.
5. Click Next.
TIP: If the artifact has not been moved since the last time it was used to
calibrate the sensor offset, you can click Set Position to position the sensor
automatically. (The Set Position button only appears if location data for the
artifact exists.)
If you are unsure whether or not the artifact has been moved, do not click
Set Position position the sensor manually to be safe.
8. On the Controls tab below the video window, click Focus Target.
10. On the Controls tab below the video window, click Crosshair Target.
15. Switch to the laser sensor and tool combination you want to calibrate.
l If the laser tool you want to calibrate is not attached, double-click its
node. Then follow the on-screen instructions and click OK.
17. Select the Calibrate option and leave the default settings as-is.
19. Select the Real Time Measure check box to turn on the laser.
20. Position the laser so you can see the laser spot on the surface of the artifact,
in the location shown.
21. Use the sensor graph to focus the laser spot manually on the surface of the
artifact.
22. Clear the Real Time Measure check box to turn off the laser.
c. Click Next.
28. If you are done calibrating off-axis laser tool offsets, click Finish.
If you want to dock probe modules and styli, you must first calibrate a properly
configured change rack with a calibrated probe sensor and tool combination. The
Change Rack Calibration wizard teaches the system the actual and clearance
locations of each port in the change rack.
NOTE: To assign touch tools to ports in the change rack, use the Change Rack
Configuration window. For more information, see Configuring a Change Rack on
page 3-21.
1. Switch to the calibrated probe sensor and tool combination you want to
use to calibrate the change rack.
l Verify the active sensor icon appears next to the probe sensor
node If necessary, double-click the node to activate the sensor.
l If the touch tool you want to use is not attached, double-click its node.
Then follow the on-screen instructions and click OK. (There may be
multiple instructions.)
Make sure you can access all ports with the probe.
3. Retract all the dust covers on the change rack and verify all the ports are
empty.
Then right-click the node of the change rack you want to calibrate, and select
Calibrate > Change Rack.
6. Carefully center the DSM in X and Y next to the middle of the first port in the
change rack.
You may need to move the probe slightly in X, Y, and/or Z to ensure smooth
movement. Be sure to drive the module all the way to the back of the port
without touching the port.
9. Click Next.
12. Measure a point on the front surface of the change rack, in the location
shown.
Do not move in Z after measuring the first point on the front surface of the
change rack.
17. Lower or raise the touch probe sensor so the stylus tip is between the upper
and lower lips of the port as shown.
18. Drive the stylus tip to the back of the port until the probe triggers.
Use the Single Rotary Calibration wizard to determine the location and
orientation of the rotary axis on a single rotary indexer.
NOTE: For information on how to calibrate a dual rotary, contact the QVI channel
partner who provided your system.
Required Artifact
l Verify the active sensor icon appears next to the node of the
optical sensor. If necessary, double-click the node to activate the
sensor.
l If the calibrated optical tool (lens) you want to use is not attached,
double-click its node. Then follow the on-screen instructions and click
OK. (There may be multiple instructions.)
3. On the ICE Navigation Tree, right-click the node of the rotary you want to
calibrate, and then click Calibrate.
4. Select the artifact you want to use from the list of configured artifacts.
NOTE: It is assumed that the rotary indexer has been moved since the last
time it was calibrated.
7. Click Next.
10. Mount the rotary calibration artifact on the rotary, with the calibration sphere
positioned at the top of the rotary.
24. To save the sphere calibration data and skip the thickness calibration, select
the Skip Width Calibration check box.
25. To perform the thickness calibration, leave the Skip Width Calibration
check box unchecked.
Light balance calibration allows you to balance the light sources on multiple systems
by setting the desired offset for each available light source.
NOTES:
l This task only applies to users who have more than one system.
l Perform light balance calibration to verify and adjust light levels if any of the
illuminators are replaced on any of the systems.
Requirements
l Make sure the optical configuration (standard or optional lenses) and software
revision level are the same on all systems.
As detailed in the procedure on the next page, designate one of the systems as the
master system, set up a part on it and create a three-feature routine. Then set up
the same part on each of the other systems, run the same routine, and use the Light
Balance Calibration function in ICE to set the light offsets as needed.
NOTE: By default, the Lights Calibration function is used to set light offsets for each
individual zoom position. Use the Zoom slider to select the desired zoom position
and then adjust the desired light offsets in the Calibrate Lights window. To apply the
selected offsets to all zoom positions, select the Apply Lights to all Zoom
Positions check box.
3. On the master system, create a sample routine that measures multiple points
on the part. Measure each point with a different light source.
l Without moving the part, repeat the above steps to measure a point on
the surface of the part using surface (coaxial) illumination.
l If the system has a grid projector, repeat the above the steps to
measure a point on the surface of the part using the grid projector.
l If the system has a fiber optic ring light, repeat the above steps to
measure a point on the surface of the part using the ring light.
4. Run the routine on the master system and note the light levels for each
feature. Be sure to position the mouse pointer at the center of Crosshair when
measuring each point.
6. Run the routine on each of the other measurement systems using the same
part and lens configuration.
If any of the light levels differ, perform the adjustment procedure on the next
page. Do this on each system that needs to be adjusted.
7. Repeat Step 6 until the light levels on each system are within five points of
the levels you noted when you ran the routine on the master system.
2. On the ICE Navigation Tree, right-click the tool node for the attached lens,
and then select Calibrate > Lights.
5. Click Finish.
Please Note
The information contained herein is based on the experience and knowledge relating to the subject matter gained by QVI
prior to publication. No patent license is granted by this information. QVI RESERVES THE RIGHT TO CHANGE THIS
INFORMATION WITHOUT NOTICE AND MAKES NO WARRANTY, EXPRESS OR IMPLIED, WITH RESPECT TO THIS
INFORMATION. QVI shall not be liable for any loss or damage, including consequential or special damages, resulting from
the use of this information, even if loss or damage is caused by negligence or other fault on the part of QVI.
AccuCentric, AMF, AutoCheck, Avant, Basic-X, Basic Bench, Benchmark, Blaze, CCP, CC-14, CC-14L, CC-16, CC-16 5X,
CC-16L, CC-20, CC-30, CC-30S, CC-V, Cobra, Contour Projector, DRS, eBx, eCAD, eChek, eGx, Elements, Feather Probe,
FeatureFinder, Flare, Flash, Focus, GageFit, Gage-X, Hi-Eye, Innova, InSite, Kotem, MeasureFit, MeasureMenu,
MeasureMind, MeasureMind 3D MultiSensor, Measure-X, MicroLine, Micro-Metric, MicroTheta, MSR, Native Video, OGP,
OQ-30B, OVP, Pinnacle, Precis, Projectron, QC-Calc, Q-Check, QL-20, QL-30, Quest, QuickView, QVI, Rainbow Probe, ROI,
Scan-X, SmartCAD, SmartCheck, SmartFeature, SmartFit, SmartLink, SmartProfile, SmartRing, SmartReport, SmartScope,
SmartScope ZIP, SmartScript, SoftGage, SoftSectioner, Sonic, Specialist, Sprint, SprintMVP, Starlite, StarLite, TeleStar, Top
Bench, Vantage, VectorLight, VDM, VIEW, View Summit, and ZONE3 are registered trademarks, trademarks, or copyrights of
Quality Vision International, Inc. All registered trademarks are registered in the USA, and may also be registered in other
countries.
Due to the nature of this material, a number of hardware and software products may be mentioned by name. In most, if not
all, cases these product names are claimed as trademarks by the companies that manufacture the products. It is not our
intent to claim these names or trademarks as our own.
No part of this document may be reproduced or disclosed in any form or for any purpose, other than personal use, without
the written permission of Quality Vision International, Inc.
Publication Number 790881 Revision B May 2016