P/N 062-0026 Power On Power Off

Protg Quick Setup & Guide Rev E 11/09 Hold either button until Both buttons must be held together until timer
LCD says RELEASE counts down and LCD shuts down.
WARNING
This guide is not a substitute for the Protg owners manual. All individuals OR
who have or will have the responsibility of using or servicing the Protg
must read and understand the contents of the manual prior to operation.
Failure to do so may result in serious injury or death.
Quick Setup DAILY BUMP TEST: If installed, remove optional pump.
1) Connect cables and Shoe Pad to Protg. Fully charge the Protg prior to
Attach the calibration adapter and thread regulator to the
first use.
calibration gas cylinder. Connect tubing. Apply gas.
2) Install Scott Gas Monitor Software on PC. Protg should quickly respond to the target gases CAUTION
3) Use software to and begin alarming. Verify proper operation of alarms. ! Do not operate
the monitor with
customize alarm settings If monitor fails to respond (or is slow) or readings
and write data to monitor. are not with in 10% of the actual gas concentration, the calibration
4) Remove Shoe Pad and SPAN calibration is required. adaptor attached.
Power On the Protege.
Perform a ZERO Cal, DAILY MONITOR ZERO CAL: In clean atmosphere, press
SPAN Cal, and daily and hold the RIGHT operation button. Release when
BUMP test. display reads PLEASE WAIT. Unit will automatically
perform ZERO calibration. Press LEFT button when
Alerts & Messages display reads APPLY GAS to exit..
ALARM Alarms: Protg will alert user of alarm conditions through flashing
LEDs, an audible alarm, and vibration. The specific sensor will flash with SPAN GAS CALIBRATION. The Protgs calibration gas concentration
a HIGH or LOW indication on the LCD. Alarm intensity increases from a as set in the Gas Monitor software setting must match actual gas concentration
LOW to a HIGH condition. used. Perform ZERO calibration procedure connecting the calibration adapter
Low Battery: The battery icon will flash and activate the alarms. with hose and regulator when APPLY GAS is displayed. Once each sensor is
+++ Over-Range Conditions: Monitors sensor(s) have been exposed to calibrated, Protege will display CAL PASSED then REMOVE GAS. Remove
very high gas concentration exceeding sensor range. calibration adaptor and wait for the monitor main
PUMP Pump: Displayed when pump is connected to Protg. Icon flashes display to return before operation.
along with activating alarms when a low flow situation occurs.
OPTIONAL PUMP. Connect pump as shown. Power on
Sensor Failures: FAL will be indicated along with the sensor type
the Protg or push the left button to activate the pump.
for each sensor that fails to ZERO or SPAN calibrate.
When BLOCK PUMP is displayed, place a finger over Align
Tighten
Calibration Expired: Indicates monitor calibration is overdue. A SPAN Thumb
the probe until CLEAR PUMP is displayed. To Tab Screw
calibration must be performed prior to operation of the Protg
disconnect pump, press and hold LEFT button until Pump
Heart Beat: Blinks to indicate Protg is actively monitoring properly Insert at Inlet
PEAK exposures are displayed. Remove pump. Angle Rev. E 11/09
maintained sensors.