PreciControl Anti-HCV
03290379 190 16 x 1.3mL

English Avoid foam formation in all reagents and sample types (specimens,
calibrators and controls).
Intended use
PreciControl AntiHCV is used for quality control of the AntiHCV and Handling
AntiHCVII immunoassays on the Elecsys and cobase immunoassay The controls are supplied readyforuse in bottles compatible with the
analyzers. system. The controls should only be left on the analyzer during
performance of quality control. After use, close the bottles as soon as
Summary possible and store upright at 28C.
PreciControl AntiHCV is a readyforuse control serum based on human
serum both in the negative and positive concentration range. The controls Due to possible evaporation effects, not more than 7 quality control
are used for monitoring the accuracy of the Elecsys AntiHCV and Elecsys procedures per bottle should be performed.
AntiHCVII immunoassays. Please note: Both the vial labels, and the additional labels (if available)
contain 2 different barcodes. The barcode between the yellow markers is
Reagents - working solutions for cobas8000 systems only. If using a cobas8000 system, please turn
PC AHCV1: 8bottles, each containing 1.3mL of control serum the vial cap 180 into the correct position so the barcode can be read by the
system. Place the vial on the instrument as usual.
Human serum, negative for antiHCV; preservative.
Target range for cutoff index: 00.3 Storage and stability
Store at 28C.
PC AHCV2: 8bottles, each containing 1.3mL of control serum
Store controls upright in order to prevent the control solution from adhering
AntiHCV (human) in human serum; preservative. to the snapcap.
Target value for cutoff index:
Stability:
AntiHCV: approximately8
unopened at 28C up to the stated expiration date
AntiHCVII: approximately4
The exact target ranges (target value 30%), given in the form of a cutoff after opening at 28C 8weeks
index (COI), are encoded in the barcodes as well as printed on the on the analyzers at 2025C up to 5hours
enclosed (or electronically available) value sheet.
Target values and ranges Materials provided
The target values and ranges were determined and evaluated by Roche. PreciControl AntiHCV, 2barcode cards, control barcode sheet
They were obtained using the Elecsys AntiHCV and Elecsys AntiHCVII Materials required (but not provided)
assay reagents and analyzers available at the time of testing.
Elecsys2010, MODULAR ANALYTICSE170 or cobase immunoassay
Traceability information is given in the Method Sheet of the relevant Elecsys analyzers and assay reagents
assay.
See the assay Method Sheet and the operators manual for additionally
Results must be within the specified ranges. In the event that increasing or required materials.
decreasing trends, or any other suddenly occurring deviations beyond the
range limits are observed, all test steps must be checked. Assay
When necessary, measurement of the patient sample tested should be Treat the control serum in the system-compatible labeled bottles for
repeated. analysis in the same way as patient samples.
Each laboratory should establish corrective measures to be taken if values Read the data into the analyzer.
fall outside the defined limits. Ensure the controls are at 2025C prior to measurement.
Note: Run controls daily in parallel with patient samples, once per reagent kit, and
For technical reasons re-assigned target values valid only for a specific whenever a calibration is performed. The control intervals and limits should
reagent and control lot combination, must be entered manually on all be adapted to each laboratorys individual requirements.
analyzers (except for the cobase602 analyzer). Therefore always refer to Follow the applicable government regulations and local guidelines for
the value sheet included in the reagent kit or PreciControl kit to make sure quality control.
that the correct target values are used.
When a new reagent or control lot is used, the analyzer will use the original References
values encoded in the control barcodes. 1 Occupational Safety and Health Standards: bloodborne pathogens.
(29CFR Part 1910.1030). Fed. Register.
Precautions and warnings
2 Directive 2000/54/EC of the European Parliament and Council of
For in vitro diagnostic use. 18September 2000 on the protection of workers from risks related to
Exercise the normal precautions required for handling all laboratory exposure to biological agents at work.
reagents.
Disposal of all waste material should be in accordance with local guidelines. For further information, please refer to the appropriate operators manual for
Safety data sheet available for professional user on request. the analyzer concerned, the respective application sheets, the product
All human material should be considered potentially infectious. information and the Method Sheets of all necessary components (if
available in your country).
All products derived from human blood are prepared exclusively from the
blood of donors tested individually and shown to be free from HBsAg and A point (period/stop) is always used in this Method Sheet as the decimal
antibodies to HCV (PCAHCV1 only) and HIV. separator to mark the border between the integral and the fractional parts of
a decimal numeral. Separators for thousands are not used.
The testing methods applied were FDAapproved or cleared in compliance
with the European Directive 98/79/EC, AnnexII, ListA. Symbols
The serum containing antiHCV used for the positive control (PCAHCV2) Roche Diagnostics uses the following symbols and signs in addition to
was inactivated using propiolactone and UVradiation. those listed in the ISO 152231 standard.
However, as no inactivation or testing method can rule out the potential risk Contents of kit
of infection with absolute certainty, the material should be handled with the
same level of care as a patient specimen. In the event of exposure, the Analyzers/Instruments on which reagents can be used
directives of the responsible health authorities should be followed.1,2 Reagent
The controls may not be used after the expiration date.
Calibrator

2015-12, V 9.0 English 1/2


ms_03290379190V9.0

PreciControl Anti-HCV
Volume after reconstitution or mixing
GTIN Global Trade Item Number

COBAS, COBASE, ELECSYS and PRECICONTROL are trademarks of Roche.