INTRODUCTION
In a biological tissue crossed by an electric current are shown the following effects:
- thermal, faradic, electrolytic.
By using HF electric curr ent the last 2 und isired and useless ef fects are eliminated and it is utilized above all the thermal e ffect. In fact
when an electric current having such characteristics flows, from the active electrode to the neutral one, with suffi cient density the cellular
liquid of the tissues warms it and produces the following effects:
1) heating is so quick that the pressure of the vapour created in the cells breaks their membranes (cutting);
2) heating is lower, so the liqu id slowly evaporates allowing the coagulation of the coagulable components of the tissues (coagulation or
haemostasis);
3) the effect is a middle way between the two above-described ones (cut with coagulation).
The ESU is HF device that may destroy the cells of biological tissues and therefore it should be used by expert staffs in el ectrosurgery
and respecting strictly all the given instructions.
The ALSATOM SU-MPC series can be used for every kind of monopolar/bipolar cut and coagulation/microcoagulation in:
GYNAECOLOGY, DERMATOLOGY, PLASTIC SURGERY, DENTAL AND MAXILLO-FACIAL SURGERY, ANGIOLOGY,
GENERAL SURGERY, ORL, GASTROENTEROLOGY, VETERINARY.
To optimize the running of the units follow the following general indications:
1. Do not a ctivate the unit before touching the tissues by the active electrode (otherwise they will creat electric arcs able to p roduce
eschar on tissue and preventing them from good cicatrization).
2. Keep the active electrode as clean as possible. The patin a insulating on an elec trode doesnt permit a good contact with the ti ssues,
thus it lowers the output power and causes sparks or superficial carbonizations.
MN SU-MPC ING March 2008 Page 2 of 9
3. Using the pure cut (especially with the loop/ conization electrodes) if the first effect is not satisf actory (with a slight sticking of the
tissue on the electrode), to have the best result increase the power of 1015 W each time.
Pure cut (for biopsy, laparoscopy, cut or skin incision, uterine conization in gynaecology etc., in general for any
case needing cut without coagulating effect).
1. Use small size electrodes, such as for ex.:
thin needle electrode (from 34 W onwards)
SAD, SAD/1, SAD/2, SAD/3 extra-f ine needle electrod es no t insulated w ith diam. from 0.10 to 0.40 mm (from 34 W
onwards)
fine loop electrodes of different measures and shapes (from 8W onwards).
long type electrodes and LLETZ type electrodes for gynaecology (from 10W onwards)
2. Select CUT current
Cut coagulating (for laparoscopy, polypectomies or papillotomies in endoscopy, fistulas, haemorrhoids and ect. In
general for any case needing cut combined with an effective coagulating effect).
1. Use the electrodes already mentioned for pure cut or, if possible, the ones with larger section, such as for ex.: knife and thick needle
electrode (from 34 W onwards)
2. Select BLEND current and sl ow a little th e sliding of the el ectrode on the tissue . If coagulati ng effect is not enough, use even
COAG. current.
Micro coagulation
1. Use extra fin e needles: AID in sulated needle (the best for depilation), all the above mentioned SAD long ty pe, not insulated
needles for depilation, telangiectasia, spider naevi, pointformed red-ruby angiomas and ect.
2. Use ball electrodes to obtain deep effect avoiding as much as possible a surperficial sparking.
3. Select COAG MICRO current (from 0.5 Watt onwards)
Coagulation
1. Use ball electrodes, surgical forceps for coagulation with effective deep effect and good surperficial effect (fulguration).
2. Use small ball, needle, loop or for polipectomy electrodes, etc. to obtain strong superficial effect and limited deep effect.
3. Select COAG current (from 2-3 W onwards).
Bipolar coagulation
1. Use bipolar forceps or bipolar electrodes (also for laparoscopy or endoscopy, eventually supplied by the other manufacturers).
2. Select BIP current (from 2-3 Watt onwards)
3. Please remind that, to reduce the sticking phenomena of the tissue on the tips of the bipolar forceps during the operation, it is very
useful to clam them as little as possible, reduce at maximum the time of activation and moisten them by physicological solution (or
plunged inside a cup or on an imbibed gauze).
AUTOTEST
The running of the unit is completely controlled b y microcontr oller both durin g the working and when switching on (starting aut otest
which, if regu lar, ends with short acous tic signal). If an y failure the s ystem get blocked the output giving specific error cod es (acoustic
signals).
NO. OF
TYPE OF FAILURE ACOUSTIC SOLUTION
SIGNAL
Turn off and re-switch on the unit (if the problem
1. me mory RAM 1
continues, contact the technical assistance)
2. CRC control software 2
3. variables of the system 3
4. supply voltage of microcontroller 4
5. act ivation circuits 5
6. output power higher than the selected value 6
7. incorrect use of the selector 4 7 Check the position of the selector
8. got broken control 5 8 As specified for Fault 1
9. continuous activation for more than 20s 9 Deactivate and reactivate immediately
10. R.F. modulation signal 10 As specified for Fault 1
11. monitoring of the output current 11
12. monitoring of R.F. supply 12
13. watchdog timer 13
Check the connection of the plug and the cable (bend
intermittent and pull it, especially near the plug and the electrode).
14. intervention of neutral plate safety circuit
(grave) If the problem continues, contact the technical
assistance)
15. usage error (eg. Activation of cut function if
Intermittent
selecting BIP or activation of 2 switches Eliminate the cause
(acute)
comtemporally)
Furthermore the following cases may occur:
1. the unit is ON (autotest OK), but when you press the pedal switch, it does not work (without acoustic or ligh t signals) or work s
irregularly.
a) check if the pedal is well connected;
b) check if the pedal is broken b y activating the un it pushing th e central ho le of th e relevant socket with a roun d
point. If the unit works regularly it is probably a problem of pedal.
2. the unit is ON (autotest OK), all the running is OK, but there is not output power or it is lower than the normal value.
a) check the good contact of the neutral plate (if the contact is bad or absent there is no power). Remember that the
hair of animals is insulating.
TECHNICAL FEATURES
Electronic generator in compliance with the Safety Standards IEC 601-2-2 3^ed.
Monopolar and Bipolar working frequency: 475 kHz
Classification IEC: I type CF Classification EC MDD: IIB
Output circuit: floating out protected against the use of the defribrillator
Mains and Absorption: see rating on the back of the unit
Mains Fuses: see rating on the back of the unit
Neutral plate safety circuit with acoustic signal (strong, intermittent) and luminous signal (red)
Output power : setting by rotary switches
Running control: by microcontroller with autotest, output error control, error codes
Protection against liquids: common, not-protected casing
Cooling by convection without ventilator
Activation: discontinuous, 10s ON/30s OFF
Dimensions and weight:
cm (LxDxH) 23x24x10 Kg 4.75 (modd. SU 140-MPC and SU 140/D-MPC)
cm (LxDxH) 21x24x10 Kg 4.50 (modd. SU 50-MPC and SU 100/D-MPC)
Conformity EMC/Directive 89/336/CEE: Category A
Suggested distances to keep from not vital devices
Source of the Current RF Typical Power (W) Distance (m)
Microcellular telephones CT1,CT2,CT3 0.01 0.4
Mobile telephones DECT, Wireless devices (modems, LANs) 0.25 2
Mobile telephones (USA) 0.6 3
Hand mobile telephones (GSM, NMT, Europe) 2 6
(DECS 1800) 8 11
Walkie-talkie (police, firemen , protection, maintenance) 5 9
Bag mobile telephones 16 16
Mobile radio (police, firemen, protection) 100 40
For broadcasting stations which use frequencies less than 800MHz, the distance can be established by using the equation: A: d = 4P
For broadcasting stations which use frequencies between 800MHz and 2.5GHz, the distance can be established by using the equation:
B : d =2.3P
P = Nominal power of the transmitter in watt (W), established by the manufacturer.
Output power, Vpp-open circuit, Crest factors (SU 100-MPC, SU 140-MPC, SU 140/D-MPC)
Cut: 140 WRMS at 500 Ohm (Vpp 1200, cf 1.7)
Blend: 120 WRMS at 500 Ohm (Vpp 1450, cf 2.8)
Coag: 120 WRMS at 500 Ohm (Vpp 1952, cf 6.9)
Coag Micro: 60 WRMS at 200 Ohm (Vpp 1420, cf 2.8)
Bipolar: 100 WRMS at 100 Ohm (Vpp 400, cf 1.4)
ATMOSPHERIC CONDITIONS
usage transport and storage
Temperature (C) +10 +40 Temperature (C) -40 +70
Humidity 30% 75% Humidity 10% 95%
Pressure (hPA) 700 1060 Pressure (hPA) 500 1060
STANDARD ACCESSORIES
MPE/E - Sterilizable pencil with connection cable.(Rated voltage = 4000 Vp)
SEL/VI - Set of 6 electrodes. (Rated voltage = 4000 Vp)
EIP/9 - Neutral plate with connection cable.
FFE - Fixing rubber belt for neutral electrode.
D-STOP/P - Double pneumatic pedal switch (mod. SU 140/D-MPC)
STOP/PN - Pneumatic pedal switch (modd. SU 50-MPC, SU 100-MPC, SU 140-MPC).
11 10 9 11 10 9
4 4
5A 5B 6 7A 7B 5A 5B 6 7A 7B
11 10 9 11 10 9
8 8
4 4
5 7 6 5 7 6
At the back
1- Main switch . symbol: (switching on) (switching off)
2- Power entry module with double fuse-holder
3- Connection pedal switch symbol
only 140/D model:
A = cut (yellow signal)
B = coagulation (blue signal)
In the front
4 Function selector
5 Output power setting (modd. 140 and 140/D:A = pure cut and blend cut; B= coagulation, micro coagulation, bipolar coagulation)
6 Neutral plate socket ................................. symbol:
7 Active electrode socket (modd. 140 and 140/D: A= hand switch handle...... symbol:
( B= pedal switch handle ..... symbol:
(modd. 100 and 50 ..................................................... symbol:
8 Bipolar electrode socket ................................................................................... symbol:
9 Output activation/coagulation: (blue light) with display (only modd. 140 and 140/D) ....................... symbol:
10 Output activation/cut: (yellow light) with display (only modd. 140 and 140/D) ................................. symbol:
11 Neutral plate safety circuit (red light) .............................................................. symbol:
Be careful: read the annexed documentation: .......................................................... symbol:
Apparatus of Class I - Type CF protected against the effect of the defibrillator ... symbol:
Alternating current ................................................................................................... symbol:
90 90 900
TAGLIO PURO 80 80 800
MONOPOLARE
70 70 700
60 60 600
MONOPOLAR PURE CUT
50 50 500
SECTION PURE 40 40 400
MONOPOLAIRE 30 CUT Pos."5" 30 300
20 20 200
10 10 100
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
10
10
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
1900
2000
IMPOSTAZIONE IMPOSTAZIONE
SETTING SETTING
REGLAGE REGLAGE
Vp-p
(V) 1500
1400
W RMS
130 1300
W RMS
120 (@ 500 ) 120 1200
90 90 900
TAGLIO BLEND 80 BLEND Pos."10" 80 800
MONOPOLARE
70 70 700
diagram corresponds to BLEND/COAG position of the mode selector)
50 50 500
SECTION AVEC 40 40 400
COAGULATION 30 30 300
MN
MONOPOLAIRE BLEND Pos."5"
20 20 200
10 10 100
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
10
10
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
1900
2000
IMPOSTAZIONE IMPOSTAZIONE
SU-MPC
SETTING SETTING
REGLAGE REGLAGE
Vp-p
(V) 2000
W RMS
1900
140
1800
130
1700
W RMS
120 (@ 500 ) 120 1600
1500
110 110
1400
100 100
Page 6 of 9
20 20
200
POWER OUTPUT DIAGRAMS ( 20% ) Modd. SU 100-MPC, 140-MPC, 140/D-MPC (NOTE: in model SU 140-MPC, BLEND
10 10
100
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
10
10
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
1900
2000
IMPOSTAZIONE IMPOSTAZIONE
SETTING SETTING
REGLAGE REGLAGE
Vp-p
(V) 1500
W RMS
70 1400
1300
W RMS
60 (@ 200 ) 60 1200
1100
50 50 1000
900
COAGULAZIONE MICRO
40 40 800
MONOPOLARE
700
20 20 400
MICRO COAGULATION
300
MONOPOLAIRE
10 10 200
COAG MICRO Pos."5"
100
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
10
10
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
1900
2000
IMPOSTAZIONE IMPOSTAZIONE
SETTING SETTING
REGLAGE REGLAGE
Vp-p
(V) 500
WRMS 400
110
MN
WRMS
100 (@ 100 ) 100
90 90 300
COAGULAZIONE BIPOLARE
80 80
70 70
BIPOLAR COAGULATION
SU-MPC
60 BIPOLAR Pos."10" 60 200
COAGULATION BIPOLAIRE 50 50
40 40
30 30 100
20 20
BIPOLAR Pos."5"
10 10
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
10
50
10
10
100
200
300
400
500
600
700
800
900
IMPOSTAZIONE IMPOSTAZIONE
SETTING SETTING
REGLAGE REGLAGE
Page 7 of 9
W RMS W RMS
(@ 500 )
90 90
80 80 Vp-p
TAGLIO PURO (V)
1100
MONOPOLARE 70 CUT Pos. "10" 70
1000
MONOPOLAR 60 60 900
PURE CUT 800
50 50
700
SECTION PURE
40 40 600
MONOPOLAIRE
500
30 30
400
CUT Pos. "5"
20 20 300
200
10 10
100
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
10
10
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
1900
2000
IMPOSTAZIONE IMPOSTAZIONE
SETTING SETTING
REGLAGE REGLAGE
Vp-p
(V) 1500
W RMS
(@ 500 ) 1400
W RMS 90 90
1300
80 80 1200
POWER OUTPUT DIAGRAMS ( 20% ) SU 50-MPC
1100
TAGLIO BLEND 70 70
BLEND Pos."10" 1000
MONOPOLARE
60 60 900
MONOPOLAR 800
50 50
BLEND CUT 700
40 40 600
SECTION AVEC 500
COAGULATION 30 30
BLEND Pos."5" 400
MONOPOLAIRE
20 20 300
MN
200
10 10
100
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
10
10
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
1900
2000
IMPOSTAZIONE IMPOSTAZIONE
SETTING SETTING
SU-MPC
REGLAGE REGLAGE
Vp-p
(V) 2000
W RMS W RMS 1900
(@ 500 )
90 90 1800
1700
80 80 1600
1500
70 70 1400
COAGULAZIONE COAG Pos."10" 1300
COAGULATION 700
30 30 600
MONOPOLAIRE
COAG Pos."5" 500
20 20 400
300
Page 8 of 9
10 10 200
100
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
10
10
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
1900
2000
IMPOSTAZIONE IMPOSTAZIONE
SETTING SETTING
REGLAGE REGLAGE
Vp-p
(V) 1500
W RMS
70 1400
1300
W RMS
60 (@ 200 ) 60 1200
COAGULAZIONE 1100
MICRO 50 50 1000
MONOPOLARE 900
40 40 800
MONOPOLAR
MICRO COAG MICRO Pos."10" 700
COAGULATION 30 30 600
500
MICRO COAG MICRO Pos."5"
20 20 400
COAGULATION
300
MONOPOLAIRE
10 10 200
100
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
10
10
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
1900
2000
IMPOSTAZIONE IMPOSTAZIONE
SETTING SETTING
REGLAGE REGLAGE
Vp-p
(V) 500
MN
400
W RMS
W RMS (@ 100 )
60 60 300
SU-MPC
50 50
COAGULAZIONE
BIPOLARE 40 40 200
BIPOLAR 30 30
BIPOLAR Pos."10"
COAGULATION
20 20 100
COAGULATION
10 BIPOLAR Pos."5" 10
BIPOLAIRE
10
50
10
10
100
200
300
400
500
600
700
800
900
1000
IMPOSTAZIONE IMPOSTAZIONE
SETTING SETTING
REGLAGE REGLAGE
Page 9 of 9