at Ast Tgo Ifcc
at Ast Tgo Ifcc
www.biolabo.fr
AST GOT (IFCC)Single vial
MANUFACTURER: Reagent for quantitative determination of aspartate amino transferase activity (AST)
BIOLABO SAS, [EC 2.6.1.1] in human serum or plasma
Les Hautes Rives
02160, Maizy, France
REF 80025 R1 20 X 10 mL REF 80125 R1 8 x 30 mL
|
Tel: (33) 03 23 25 15 50
IVD IN VITRO DIAGNOSTIC USE
Fax: (33) 03 23 256 256
Mix. Start a timer. Record initial absorbance after 1 minute at 340 nm.
QUALITY CONTROL Record the absorbance again every minutes during 3 minutes.
BIOLABO EXATROL-N Level I REF 95010 Calculate absorbance change per minute (Abs/min).
BIOLABO EXATROL-P Level II REF 95011
Other assayed control sera referring to the same method. Note: Specific procedures are available upon request for automated
External quality control program. instruments. Please contact BIOLABO technical support.
It is recommended to control in the following cases:
At least once a run.
CALCULATION
At least once within 24 hours.
Calculate the result as follows:
When changing vial of reagent.
After maintenance operations on the instrument.
If control is out of range, apply following actions: With theoretical factor:
1. Repeat the test with the same control.
2. If control is still out of range, prepare a fresh control serum and
repeat the test. Abs/min) x 1746
IU/L = (
3. If control is still out of range, verify analysis parameters:
Wavelength, temperature, specimen/reagent ratio, time counting,
calibration factor. IU/L
Kat/L =
4. If control is still out of range, use a new vial of reagent and 60
reassay
5. If control is still out of range, please contact BIOLABO technical With serum multicalibrator:
support or your local Agent.
(Abs/min) Assay
AST Activity = x Calibrator Concentration
EXPECTED VALUES (1) (2) (Abs/min) Calibrator
UI/L at 30C at 37C
LINEARITY
The assay is linear up to 350 IU/L.
If Abs/min > 0.200, reduce specimen volume or dilute specimen with
saline solution and reassay taking into account the dilution factor to
calculate the result. Linearity will depend on the specimen/reagent
ratio.