Professional Documents
Culture Documents
Chem Lab Report 2
Chem Lab Report 2
APPARATUS
Analytical balance
Bunsen burner or hot plate
Crucible
Crucible tongs
Clay triangle
Glass rod
Measuring cylinder 10 mL
Porcelain or white tiles
Tripod stand
Wire gauze with asbestos center
CHEMICAL
Zin powder
HCL 6.0 M
PROCEDURE
QUESTIONS
ANSWER
Firstly, remember that you it should be read at the number that is at the
bottom of the meniscus with 1/10 of the smallest scale marked on the
burette. The meniscus is created by the surface tension of the liquid and
the extent that the liquid wets the wall. Part of the structure of meniscus
involves reflection of light. Therefore the meniscus, as you see it is
dependent on the background. The precision tap, or stopcock, on the
bottom is used to dispense known amounts of a liquid in experiments for
which precision is necessary for instance, Titration experiment. The
other factor that should be considered is the human error or parallax
error. The above two views of the burette taken with the eye at the same
level as the liquid level. If your eye is either higher or lower than the
liquid level, you will make an error in reading the liquid level. This is
referred to as a Parallax error. Since, except for the 90 degree angle
where no parallax error occurs, you probably will not reproduce an
arbitrary angle which you read the meniscus. Therefore this will cause
an error.
RESULTS
Data
b. Calibration of Pipette
Calculations :
0-5 4.47
5-10 9.54
10-15 14.49
15-20 20.82
DISCUSSION
This experiment was carried out to learn the qualitative and quantitative
aspects of common laboratory equipment and to expose students to the
factors that affect the accuracy of an experiment. The experiment was
carried out at laboratory which was provided a certain equipment that
involving measurement and transferring volumes of liquid. The
equipment that were been carried for the experiment are such as
Volumetric Flask, Pipette, and Burette. All the equipment are been
tested by students for the accuracy of the measurement and to study the
measurement skills from the students. By that mean, calibration from all
three equipment were be carried out and tables were formed based on
the result of the experiment consisting 3 calibration of three equipment
respectively.
Secondly, the calibration of pipette was carried out with more variation
where the reading were took more than once (2 times) using the same
pipette. Based on the results on (b), the first reading of empty beaker
was 107.15 and when added the distilled water the reading became
118.15 due to the additional mass from the distilled water which is 11
gram. However, the second reading was not equivalent with the first
reading. On the second reading, using the same pipette, the mass of
empty beaker was 107.20. The mass was changed when distilled water
is poured into the beaker resulting additional amount of mass to the
reading that is 118.18, based on this, the total mass of distilled water
was 10.98. According to both reading, it is shown that there variety of
reading took places. Along with the theory, this is due to limitations and
sources of errors during the experiment take place although with
constant temperature and density of water (from Table 1).
Lastly, the calibration of Burette was carried with more precise and
accurate reading where students provide a reading for the mass of
distilled water for each 5 mL burette reading (g) to lessen the source of
error. Students were took different reading of burette which is 5mL,
10mL, 15mL, and 20mL. Based on the result, it is shown that the higher
the reading of burette were resulting by the higher of the mass of
distilled water. As a result, the volume of water increase as the reading
of burette increase. The calibration of burette also been carried out with
the constant of temperature and the density of water.
CONCLUSION
The experiment that been carried out has exposed to the student about
the limitation and the sources of error when take measurement of
accuracy using Volumetric Flask, Pipette, and Burette. The discussion
has proven that human error or parallax do exist when taking
measurement where can cause false reading and errors during
experiment.