Dear Customer, MEASUREMENT PROCEDURE: CALIBRATION PROCEDURE:

USER MANUAL Thank you for choosing a Martini product.


This manual will provide you with the neces- 1 Turn the meter on by pressing 1 Turn the meter on by pressing ON/OFF.
sary information for the correct use of the ON/OFF. When the LCD displays - - -, the meter is
MI 415 instrument. Please read it carefully before
2 When the LCD displays - - -, the
ready.
using the meter. 2 Press CAL key and CAL will blink on the
meter is ready.
Turbidity Meter This instrument is in compliance with
3 Fill a clean, dry cuvet with sample
display.
directives.
up to the mark.
4 Replace the cap.
SPECIFICATIONS: 5 Clean the cuvet with a lint- 3 Press CAL key again. The date of calibra-
free cloth to remove any fin- tion in MM.DD format appears on the LCD.
Range 0.00 to 50.00 FNU gerprints, dirt or waterspots.
50 to 1000 FNU
Resolution 0.01 FNU and 1 FNU
6 Place the cuvet into the
Precision 0.5FNU or 5% of read- 4 To toggle between month and day
holder and ensure that the
ing, whichever is greater press DATE/ key.
notch on the cap is posi-
Light Source High emission infrared LED tioned securely into the
groove. 5 To change the date press READ/
Light Detector Silicon Photocell key.
Method Detection of scattered light 7 Press READ/ and SIP will blink during
Environment 0 to 50C (32 to 122F); measurement. 6 To confirm the date values press
100% RH max. CAL key.
Battery Type 1 x 9 volt A blinking ZERO message will appear on
Auto-Shut off After 5' of non-use the LCD. Insert the 0 FNU standard and
press CAL key. SIP will blink during mea-
Dimensions 192 x 104 x 52 mm 8 The instrument directly displays turbidity in surement.
(7.5 x 4.1 x 2) FNU units.
Weight 380 g

ISTMI415 07/05
7 At the end the blinking 10.00 message GUIDE TO DISPLAY CODES: BATTERY REPLACEMENT
will appear on the LCD. ACCESSORIES:
This prompt appears for 1
Insert the 10 FNU standard and press CAL second each time the instru- Battery replacement must only take place in a MI0004 Tissue for wiping cuvets
key. SIP will blink during measurement. ment is turned on. non-hazardous environment. (4 pcs.)
The dashes - - - indicates Simply rotate the battery cover on the back of MI0005 9V battery (1 pc.)
that the meter is in a ready the meter. MI0011 10 mL glass cuvete (2pcs.)
state and measurement can Detach the battery from the terminals and MI0012 cap for 10 mL glass cuvet
be performed.
attach a fresh 9V battery while paying atten- (2 pcs.)
Sampling In Progress. Flash- tion to the correct polarity. Insert the battery MI0013 Stopper for 10 mL glass
ing SIP prompt appears each and replace the cover. cuvet (2 pcs.)
8 The blinking 500 message will appear on
time the meter is performing a
the LCD. MI515-000 Calibration standards: 0,
measurement.
10, 500FNU (3 x 30 mL
Insert the 500 FNU standard and press CAL
The presence of battery icon dropper bottles)
key. SIP will blink during measurement.
on the display indicates that
the battery voltage is getting
low and the battery needs to
be replaced.
-BA-, the battery is dead and
must be replaced. Once this
9 At the end of the reading the instrument indication is displayed, the
goes directly to measurement mode. Now meter will lock up. Change the
the instrument is calibrated and ready to battery and restart the meter.
use.
CAL, the meter is in calibra-
tion mode. If the CAL key is WARRANTY
not pressed within 6 seconds, This instrument is warranted against defects in mate-
the meter will automatically rials and manufacturing for a period of two years
from the date of purchase.
switch to idle mode.
Note: A monthly calibration is recommended. If during this period the repair or replacement of parts
is required, where the damage is not due to negli-
The instrument can be checked with the CL appears on the lower part gence or erroneous operation by the user, please
supplied standard solutions each day. of the display when the meter return the meter to either distributor or our office in
is in calibration mode. the original packing and the repair will be free of
To check the date of the last calibration charge.
simply press DATE/ key. Damages due to accidents, misuse, tampering or
lack of prescribed maintenance are not covered.