User Manual: Measurement Procedure: Calibration Procedure
User Manual: Measurement Procedure: Calibration Procedure
User Manual: Measurement Procedure: Calibration Procedure
ISTMI415 07/05
7 At the end the blinking 10.00 message GUIDE TO DISPLAY CODES: BATTERY REPLACEMENT
will appear on the LCD. ACCESSORIES:
This prompt appears for 1
Insert the 10 FNU standard and press CAL second each time the instru- Battery replacement must only take place in a MI0004 Tissue for wiping cuvets
key. SIP will blink during measurement. ment is turned on. non-hazardous environment. (4 pcs.)
The dashes - - - indicates Simply rotate the battery cover on the back of MI0005 9V battery (1 pc.)
that the meter is in a ready the meter. MI0011 10 mL glass cuvete (2pcs.)
state and measurement can Detach the battery from the terminals and MI0012 cap for 10 mL glass cuvet
be performed.
attach a fresh 9V battery while paying atten- (2 pcs.)
Sampling In Progress. Flash- tion to the correct polarity. Insert the battery MI0013 Stopper for 10 mL glass
ing SIP prompt appears each and replace the cover. cuvet (2 pcs.)
8 The blinking 500 message will appear on
time the meter is performing a
the LCD. MI515-000 Calibration standards: 0,
measurement.
10, 500FNU (3 x 30 mL
Insert the 500 FNU standard and press CAL
The presence of battery icon dropper bottles)
key. SIP will blink during measurement.
on the display indicates that
the battery voltage is getting
low and the battery needs to
be replaced.
-BA-, the battery is dead and
must be replaced. Once this
9 At the end of the reading the instrument indication is displayed, the
goes directly to measurement mode. Now meter will lock up. Change the
the instrument is calibrated and ready to battery and restart the meter.
use.
CAL, the meter is in calibra-
tion mode. If the CAL key is WARRANTY
not pressed within 6 seconds, This instrument is warranted against defects in mate-
the meter will automatically rials and manufacturing for a period of two years
from the date of purchase.
switch to idle mode.
Note: A monthly calibration is recommended. If during this period the repair or replacement of parts
is required, where the damage is not due to negli-
The instrument can be checked with the CL appears on the lower part gence or erroneous operation by the user, please
supplied standard solutions each day. of the display when the meter return the meter to either distributor or our office in
is in calibration mode. the original packing and the repair will be free of
To check the date of the last calibration charge.
simply press DATE/ key. Damages due to accidents, misuse, tampering or
lack of prescribed maintenance are not covered.
ISTMI415 07/05