Professional Documents
Culture Documents
Manual D00339937 2
Manual D00339937 2
The column header cell displays the Additive device. The green
back ground indicates which additive device has been selected.
The Name cell allows an operator the ability to enter in a name for
the additive that is being delivered by this device. Up to 28 alpha-
numeric characters may be entered in this cell. To save the name cells as a blank cell,
enter in a space using the space bar. Press the delete key to erase the contents of a name
cell. The name will be saved on application shutdown and used on subsequent
application runs.
The Rate cell displays the current delivery rate of the additive
device. The blue bar graph at the top of the row represents output
drive. The bar graph is scaled from 0 to 100 %, the maximum
potential drive for this additive device. The back ground color indicates what Loop
Control mode the device is currently using; brown represents Open Loop Control Mode,
and white represents Closed Loop Control Mode.
The Conc cell displays the concentration being delivered from the
device. There are two bar graphs displayed, both of these are scaled
from 0 to maximum rate delivery for this additive device. The gray
bar represents the rate set point. Where as the blue bar represents the
current rate. During the job these two bars should match with each other.
The cell row will display the Master Rate for the selected device
when toggled to MRat.
The cell row will display the Error percentage as a function of set
point versus actual concentration for the selected device when
toggled to Error. The scale is 0 to 100 %.
The A/M cell displays the control mode the additive device is
currently running. The cell has three possible displays; OFF, AUTO,
MANUAL.
The Job cell displays the accumulated Job volumes for the additive
device.
Cell Help Function The mouse cursor will display the function of
the cell when the mouse cursor is moved to that particular cell.
A control window for a specific additive can be selected by touching or clicking the
mouse cursor on the respective column. In the example below the LA 1 device was
selected. Touching a cell in the name row will not select a control window.
The Summary window may be shifted right or left to select other devices that are not
displayed on the screen. The window below illustrates how the Summary display has
been shifted to the left by selecting DA 1. To shift the Summary Display back to the
right; select the furthest left column, i.e. the SS 2 additive.
The rows and columns may be resized by using the cursor. Move the cursor over to a
row or column line separator, the cursor will change to a cross hair cursor. Click and
hold the cursor when it changes to the cross hair and resize the row or column to the
desired width. The example below illustrates before and after.
On some jobs a particular additive device may not be used, in that case the column may
be resized so that it is no longer visible. The example below illustrates that LA 3 column
has been shrunk so that LA 3 is no longer visible. To return the Summary Display back
to normal close Blender Display and start a new Blender Display, refer to Chapter 2 for
more detailed information.
Stage Volumes
Job Volumes
Actual Concentration
Choose Auto,
Actual Manual or OFF
additive rate
Desired Manual
Pump Rate drive
adjust
Enter desired
Normally Clean Rate Im finished
concentration
OFF This will turn off the control delivery system. The command
drive percentage will be set to zero. The additive device will no longer
deliver chemicals via the ACE control system.
Set Point The desired concentration that will be delivered when the
Additive device is set to AUTO CONC control mode. The Set Point is
entered in gallons of additive per thousand gallons of clean fluid. The
number shown above the Set Point is the Actual Concentration at this
time. When in AUTO CONC mode the control system will drive the
pump until the Actual Concentration equals the Set Point Concentration
and then maintain that concentration. Press the Set Point button to bring up a pop up
window to enter in a desired value.
The Set Point for the Sand Delivery system is handled in a different way. Sand
Screw delivery is considered a group function that is all the Sand screws are handled by
the Sand Screw Group (SS Group). Sand concentration set point is entered in pounds per
gallons of clean fluid on the Sand Screw Group. An operator can enter in a Sand
concentration on an individual screw but the number will be shown as zero on that screw
and the number that was entered will be entered on the SS Group Concentration. Then
when the SS Group is placed in Auto Conc. the sand screw or screws that are in Auto will
have the correct Sand set point entered with respective to their Load Factor set up. Refer
to section 4.8.1 Sand Screw Loading.
The Master rate for the additive device is displayed on the Select
Master Rate Button. Selecting the Master Rate button will select a
pop up window to allow the operator to choose which rate the
additive device will be slaved as a ratio. The Selected Suction Rate is
the default setting on application start up.
4.3.5 How do I know the volume of chemical that has been delivered?
The Stage and Job volumes for a particular additive are displayed on the
green volumes button. The top number displayed is the volume for the
particular stage of the job. As the job advances through stages then this
value will be reset to zero on every stage advance. The bottom number
displayed is the Job volume for this particular additive. This number
accumulates through the job. Job totals are not reset at stage advances.
Pressing the displayed Stage totals button will select a pop window to set the volumes
values.
The Done button closes the window, any changes made have all ready taken effect.
4.4.1 How do I select Micro Motion or TACH pickup and calibrate for LAs?
Micro motions and Tachometer sensors work very differently. Micro motion is the
standard for Liquid Additive feedback when installed. Micro motions are Coriolis meters
that measure mass flow and are typically U shaped tubes. Micro motions are known to be
more accurate than Tachometers. The downside to the Micro motion is that they are not
tolerant to air infiltration. The Liquid Additives must be kept primed while running
Micro motions to insure the Micro motions do not allow air entrainment. Micro motions
will give very erratic feedback to the controller causing erratic drive to the pump if the
Micro motions are allowed to have air entrainment. Micro motions also require good
cleaning after the job to insure they are not clogged with chemicals.
Tachometer sensors are our backups to the Micro motions. Tachometer sensors
work by sending a pulse out to the controller every time a tooth on the sprocket passes by
the magnetic pickup. The sprockets are mounted to the pump's shaft. So the controller is
not aware of the chemical rate flowing through the pump or if chemical is even going
through the pump. All the controller knows is how fast the pump is turning. The
controller then calculates the rate based on the speed of the additive pump.
Remember the controller is sending a drive signal out to the pump based on the
feedback. A very common and disastrous problem can occur when the magnetic pickup
or Micro motion fails. If the device is in AUTO with a set point and clean rate, the
controller will drive the pump until it sees adequate feedback. If it sees zero for feedback
because of a bad sensor it will drive the pump wide open. Lets say the magnetic pickup
failed while you were on a job and you were pumping cross linker at .5gal/1000. If the
device is in Forced Closed mode then once the controller sees the feedback is zero it will
keep increasing the pumps speed until it sees .5. With a bad pickup it will never see .5
and pump the entire cross linker away, if you dont catch it. It is always good practice to
get a visual on the pump if the feedback fails. Tachometers are installed as a backup to be
used if the Micro motion fails or is inaccurate, always switch to Tachometers never adjust
Micro motion calibration during the job. Bucket testing before the job and strapping
during the job will eliminate many problems.
Pressing the actual additive rate button on a Liquid Additive Pump window will
display the additive Select flow meter/calibrate window. The actual rate from the Micro
Motion and Tachometer pickup feedback sensors are displayed in the green buttons. To
select the sensor to use, the operator will need to click on either of the blue Tack Pickup
or microMotion buttons.
ACE blender Calibration - press the green displayed rate button next
to the particular sensor to enter in a calibration value for a sensor. A
Calibrate Transducer pop up window will be displayed. Typically the
operator will enter in the pulses per gallon (pul/gal) by selecting the
Slope button. Micro Motion sensors have fixed values, typically 6000
pulses per gallon and should never be changed unless the size or type
of the Micro Motion sensor changes. The Adjust High button should
never be used. The % Efficiency button is Not Applicable (N/A) for
this calibrate window. The Output button displays the current additive
rate for this device; do not press the output button as it has no function for this window.
The Select Flow meter/calibrate window displays the maximum rate and speed for the
selected additive device, in the example the maximum rate is 3 gpm and the maximum
speed is 1387 rpm. The current
calibration value for the Tachometer
sensor is displayed as gallons per
revolution (gal/rev). The partial library
listing that is shown is the pump type
selected from this GUI. It is possible
the pump type does not match the pump
when changed from another GUI.
Other devices can be selected from a library listing, select the down cursor button and a
drop down list will be displayed. Use the scroll down bar or up/down button to move
the list up or down, select an item to
change the device. This operation is to
be used when changing out a pump
device and selecting nominal start up
calibration values. Then a bucket test
would be run to finalize the calibration
values. Use the Enter Selected
Additive Pump button that has changed
to Accept Pump Type Change to
confirm the selected device.
The information that is shown in the list is such; Descriptive name of device, Maximum
rpms, Maximum rate, gallons per revolution, pulses per gallon for Tachometer sensors.
The device that was selected will be saved during application shutdown and will be used
on the next power cycle.
Once the selected device is chosen these actions will occur immediately.
Displayed MAX speed (rpm) and rate (gpm) will be updated.
The calibration for the Tachometer sensor will automatically be updated.
The Max Display data will be automatically updated on the PID and Setup
window.
Multiple instances of this bucket test calculator can be run at the same time. Thus
several windows can be used to measure for Liquid Additives, Dry Additives, Sand
Screws, and Suction and Discharge devices at the same time. Closing out the window by
using either the DONE button while the timer is running will not terminate the calculator
process thus continuing the timer and volume accumulation. However, using the X
window control button will terminate the calculator process.
Use the newly acquired slope value fro the respective Tank. The new
calibration values will be automatically entered into the controller.
Calculated time for selected correction volume, informs the user how much time is need
for the selected finish volume. The user can run to the time fro the selected volume and
confirm fluid metered.
.
Done Closes the window is the timer is not running. Minimizes
the window if the timer is running
HTML example
In the first example below the calculator has been run for 1 minute. No error is reported
since no volume value has been selected.
MicroMotion
selected
Selecting the Micro Motion Calculated Volume button as shown in example 2 does
several things;
Sets the Finish volume for the Calculator to use as a value
Sets the % error to 0 for this meter
Returns the Calibration factor to the original value for the Micro Motion
The % error for the Tachometer is calculated using the Finish volume
A new calibration factor for the Tachometer is calculated
The % error for the Tank level sensor is calculated using the Finish volume
A new calibration factor for the Tank Level is calculated
The operator could then press the Accept New Calibration button and the new
Tachometer calibration would be used by the controller (0.01gal/rev). The Micro Motion
calibration is not set. Optionally the new Tank level calibration could be set.
In the example below, the finished volume has been set to 5 gal by using the drop
down list. The resultant error calculation is shown for the respective devices.
The new calibration for the Tach pick up could be accepted. The Micro motion error is
shown as a reference, the micromotion calibration are not set by using the accept new
calibration button. The Tank level sensor may also be set to the correct value by using
the accept new slope button.
Note: to remove the fixed 5 gal value the operator should select the drop down error then
release without selecting a value.
Add a new chemical to the library. The user will need to enter a
name and density of the chemical in pounds per gallon
Maximum deliver rate for this dry additive device
Pressing the DONE button will close the Dry additive calibration
window.
Set Additive Cal and Max This button will change when a change
is made to the calibration factors in the display. Press any of these
buttons to enact the changes that have been selected on the window.
This will automatically enter in a new calibration factor for the device type and chemical
selected. Device type selected and chemical selected will be automatically saved on
application shutdown. The calibration factor will not automatically be saved; the
operator will have to use the Save the Blender File button on the Local Only Tab to save
this calibration factor permanently.
The library list details the available chemicals that can be selected. The chemical list is
composed of names and calibration factors for that particular chemical. An operator can
enter in the pounds per revolution or pulse per pound instead of using the library.
The current calibration factor being used is displayed in the Current Cal
box; the values are displayed in pounds/revolution (lb/rev) button and
pulses per revolution (pul/rev) button. Press either of these buttons to
enter a new value so that a new calibration factor can be used
instantaneously at the blender. The top most value is Sand screw 1,
second is sand screw 2 and lastly is sand screw 3. The operator will
have to use the Save the Blender File button on the Local Only Tab to
save this cal factor permanently.
The center of the window is a temporary holding area for calibration factors. The
operator can preload values into the holding area and then at the operators discretion
accept those values. Accepting the values will instantaneously load those values into the
blender controller. The Sand Specific Absolute Volume will also be set on the
Densometer.
The calibration ratios that are internally used by the controller are displayed in these gray
fields. The feedback sensors on the sand screws devices are typically Optical Encoders
with calibration factors of 600 pulses per revolution.
This button serves as a visual reminder that a value has been changed.
This button will change to yellow and display a prompt when ever a
value has been change. Pressing this button enacts all items selected
on this window, including Sand Screw size, Sand type and Calibration Factors.
When a Sand Type has been selected, the button will display Accept
Sand Type Change. Select this button to accept all changes that have
been made.
When a Specific Volume has been selected, the button will display
Accept SpecVolume. Select this button to accept all changes that
have been made.
When a Screw Size has been selected, the button will display Accept
Screw Size Change. Select this button to accept all changes that have
been made.
Closes the window and returns back to the Control window if that window
is still open. This button will perform the same function as the DONE
button.
Additive devices such as Sand Screws have additional control factors associated
with these devices that the Liquid Additive pumps do not. Slurry Processors can have
several Sand Screw sizes; 9, 10, 12, 14 inches. Some units have a 12 inch screw and a 14
inch screw; in order to use these different sizes efficiently the controller has the ability to
set up what sand load each screw will take. This loading is called the Load Factor. The
Load Factor is defined as to how the desired Sand delivery will be divided among the
Sand Screws as a ratio of percentage when both Sand Screws are in Auto.
Example; using a 14 inch and a 12 inch screw, enter in load factors of 60% for a
14 inch and 40% for a 12 inch. Then entering in a desired Sand Concentration set point of
1 pound will result in the 14 having a set point of 0.6 lbs. and the 12 having a set point
of 0.4 lbs.
When only one Sand Screw is used the loading is 100% irregardless of Load
Factor. Thus using the previous example; if a 1 lb. set point was entered and only one
Sand Screw was in Auto then the Sand screw set point would be 1 lb. even though the
load factor was 60%.
The Load factor is Not Applicable (N/A) for this device; typically this
is shown for a Liquid Additive device which does not use Load
Factors.
Typically the Sand screws will need to be pre-filled before starting sand delivery;
pre-filling will help with preventing delays with bringing the Sand Concentration up in
the mixing Tub. Some locations will preload the Sand Screws by setting the Slurry
processor control switch to manual on the override control panel, and then adjusting the
Sand Screw rate until the screw is full of Sand. Other locations may desire to use the
Auto load feature of the Ace control system. The Auto load process is put into place
when a Sand Screw is placed in Auto, if the particular Sand Screw is empty then the
controller will output a drive command to turn the screw the desired number of
revolutions. If the Sand screw is marked full when placed in Auto then the Auto load
process will not occur.
Press this button to inform the controller that the Sand Screw device is full of
sand. At ACE application start up the default setting is full.
Press this button to inform the controller that the Sand Screw device is empty
of sand.
The Sand Screw is empty. The controller is expecting the Sand screw to
be empty when this field is displayed.
The Sand Screw is full. The controller is expecting the Sand screw to
be full when this field is displayed.
I = Integral Term The integral time has units of seconds, and is a very
critical part of the tuning of the PID controller. A simplistic, but fairly
accurate way to think of integral term is that the integral term should
be equal to the time that the controller is expected to take to bring the
process variable equal to the set point. For example, one would expect a change of set
point on a liquid additive to take approximately 2 seconds. The integral time does not
have to be exact, but should be within 50% of the closure time. The integral time must
also be within a physically possible response time for the device.
D = Derivative Term, the derivative term is not currently being used on any of the
blender control systems.
4.8.3 Closed vs. Open loop, Rejecting Noise - Command drive and feedback
sensor
The controller typically outputs a drive signal to an electronically driven hydraulic valve.
The hydraulic valve then turns a hydraulic motor that turns a device. A signal is then sent
back to the controller from a sensor, which informs the controller of the additive rate. We
call this return signal, feedback. In the automatic mode the controller can always keep a
steady set point on a given Additive pump using this Closed Loop Control. The controller
will vary pump speed (rpms) to keep the concentration set point tracking with the clean
rate. In some situations the controller will need to have the necessary information to
output a calibrated drive signal without expecting feedback sensor information, this is
called Open Loop Control. This calibrated drive signal equates to a certain drive output
which will result in a certain pump speed for that device, thus producing a known
additive delivery rate. This is called a mapped drive signal, an example is listed below.
0%, 0rpm
11.8571%, 0.4rpm
23.7143%, 1.5rpm
35.5714%, 70.8rpm
47.4286%, 140.2rpm
59.2857%, 210.9rpm
71.1429%, 281.8rpm
83%, 358.3rpm
100%, 392.7rpm,
This Mapped drive signal is created when all components are functioning and an
Auto Self Tune is run. An example of where this is used is the Sand Screws, the
Feedback sensor for these devices are Optical Encoders. If the Optical Encoder fails then
the controller will detect this and switch to Open Loop. The controller will then use this
mapped drive signal to deliver the correct amount of Sand.
OPEN if OOR Open if Out of Range, the controller will toggle from Closed
Loop to Open Loop mode if the feedback sensor information is out of range of
the mapped drive signal. The feedback sensor information is calculated and
displayed.
OPEN if Zero The controller will toggle from Closed Loop to Open Loop
mode if the selected sensor has failed and the controller has no information
(zero) from the Feedback sensor. The feedback sensor information is
calculated and displayed.
Forced Close The controller will output a command drive signal and always
use the feedback signal to correct the command drive signal. The feedback
sensor information displayed is the actual data from the sensor.
Forced Open The controller will output a command drive signal and never
use the feedback signal to correct the command drive signal. The feedback
sensor information is calculated and displayed. The controller will use the
mapped drive signal to deliver the correct amount of substance.
The toggle function control mode the additive device is using at the
current time. This does not indicate that it is for example running in
OOR but that it has the potential to switch states from Closed Loop to Open Loop.
The control mode the additive device is using at the current time.
Forced Open and Forced Close are being used when this message is
displayed.
Zero Mode The controller will not transmit a feedback signal to other external units
(TCC) when the command drive percentage is zero. Thus the TCC will observe no rates
or concentrations when there is no command drive and a false feed back signal is being
detected such as noise or vibration.
Normal Mode - The controller will transmit a feedback signal to other external units
(TCC) at all times. This will occur even when there is no command drive and false feed
back signals such as noise and vibration are detected.
NOTES