Professional Documents
Culture Documents
Electro Voltage 2
Electro Voltage 2
a r t i c l e i n f o a b s t r a c t
Article history: A systematic study of the effect of design and operation conditions of an electrochemical reactor on
Received 26 November 2008 the treatment time for arsenic (As) electro-removal from underground water (GW) was carried out to
Received in revised form 9 July 2009 analyse the energy and electrode consumption. The effects of four factorscurrent density, interelectrode
Accepted 10 August 2009
distance, electrode areavolume ratio, and liquid motion driving modewere evaluated. The response
Available online 18 August 2009
variables were the energy and the electrode consumption and the treatment time to reduce the GW
residual As concentration to 10 g L1 , which is the maximum contaminant level (MCL) established by
Keywords:
the World Health Organization (WHO) in drinking water. The results obtained in this study showed that
Electrocoagulation
Underground water
the factor that had the greatest effect on most of the response variables was the liquid motion driving
Arsenic electro-removal mode. The best residence time was 20 s, which favoured low energy consumption (58.78 Wh m3 ) and
low electrode material loss (9.59 g m3 ).
2009 Elsevier B.V. All rights reserved.
0304-3894/$ see front matter 2009 Elsevier B.V. All rights reserved.
doi:10.1016/j.jhazmat.2009.08.044
1618 J.F. Martnez-Villafane et al. / Journal of Hazardous Materials 172 (2009) 16171622
Table 1
Characteristics of the underground water.
Characteristics Value
pH 7.22
Conductivity 1067 (S cm1 )
Hardness 250.35 (mg L1 )
Ca 56.82 (mg L1 )
Mg 43.32 (mg L1 )
Na 106.48 (mg L1 )
K 20.28 (mg L1 )
SiO2 59.38 (mg L1 )
SO4 2 203.04 (mg L1 )
Cl 60.50 (mg L1 )
F 0.934 (mg L1 )
Table 2
Comparison of the calculated theoretical metal dissolution value wt with respect to the measured experimental value we .
d LM A/V i
(6 Wh) devices is considered. The theoretical amount of iron liber- for shorter d values. Hence, for larger d values the movement of
ated from the anode is given by Faradays law: the participant species (including those in reactions (1)(4)) in the
As removal process is slower, the thickness of the concentration
ItM
w= (6) boundary layer in the interelectrode gaps is larger, and the reaction
nF
rate decreases. This behaviour indicates the relative importance of
where n is the number of electrons transferred (2) in the reaction, F water motion and turbulence on the treatment time and their large
is Faradays constant (96,486 C mol1 ), M is the iron atomic weight dependence of the interelectrode gap.
(55.85), and t is the treatment time in seconds. From Eqs. (5) and Fig. 3 illustrates the energy consumption as a function of the
(6) it is clear that the energy and electrode consumption increase charge density (Q) and the measured iron electrode weight loss
with longer treatment time and higher applied current. when d increases from 3 to 5 and 9 mm. As shown in this gure,
the E value increased almost linearly from 61 to 184 Wh m3 as the
4. Results and discussion electrode material loss increased from 4.76 to 10.27 g m3 when d
increased from 3 to 9 mm. This behaviour was associated with a
4.1. Effect of interelectrode distance (d) t increase corresponding to an increase in d, which, as mentioned
before, was due not only to the slower movement of the partici-
The interelectrode distance was varied at values of d = 3, 5 and pant species in the As removal reaction but also the fact that the
9 mm while the rest of the factors remained unmodied at the distances traversed by the participant species are larger when d
following values: LM = aw, A/V = 0.466 cm1 and i = 1.50 mA cm2 . is increased. Additionally, the interelectrode distance also has an
Fig. 2 shows the effect of d on the residual As concentration as important effect on the cell voltage U according to Eq. (7):
a function of the EC process time. As can be seen in Fig. 2, as U = EA + A EC + C + dielectrolyte (7)
d increases the treatment time to achieve the MCL of arsenic
(10 g L1 ) increases. The treatment time values increased from 20 where EA and EC are the reversible anodic and cathodic potentials,
to 30 and 45 s when d increased from 3 to 5 and 9 mm. The better respectively, and A and C are the corresponding overvoltages,
t value obtained with the shortest d is attributed to an increase in respectively. The last term is related to the electrolyte IR drop,
the mass transfer during the EC process as a result of the turbulence where is the electrolyte resistivity. According to this equation the
generated. During the experiments with smaller d values, the tur-
bulence at the base and at the top of the electrodes was enhanced
Fig. 2. Effect of d on the residual As concentration in GW as a function of the EC Fig. 3. Effect of d on the energy consumption as a function of the charge density and
treatment time. The EC treatment was carried out at i = 1.5 mA cm2 , LM = aw and the electrode weight loss for As removal from GW by the EC treatment carried out
A/V = 0.466 cm1 . at i = 1.5 mA cm2 , LM = aw and A/V = 0.466 cm1 .
1620 J.F. Martnez-Villafane et al. / Journal of Hazardous Materials 172 (2009) 16171622
Fig. 6. Effect of the A/V ratio on the residual As concentration as a function of the EC Fig. 8. Effect of i on the residual As concentration as a function of the EC treatment
treatment time. The EC treatment was performed at i = 1.5 mA cm2 , d = 3 mm and time during As removal from underground water. The EC treatment was performed
LM = a. at d = 3 mm, LM = a and A/V = 0.466 cm1 .
by air injection is sufcient to achieve efcient As removal; thus sumption increases signicantly as a consequence of decreases
this LM mode will be used for the following experiments. in A/V. It increases from 7.20 to 12.97 g m3 when A/V decreases
from 0.700 to 0.350, respectively. The energy consumption remains
4.3. Electrode areavolume ratio (A/V) effect approximately constant (63 Wh m3 ) for A/V values of 0.700 and
0.583 cm1 and decreases to 55 Wh m3 when A/V is decreased to
The effect of the ratio between the total electrode area and the 0.466. However, for the lowest A/V value (0.350 cm1 ) the energy
volume of water in the reactor (A/V) was studied in an interval from and electrode consumptions increased abruptly to 81 Wh m3 and
0.350 to 0.700 cm1 by adding electrodes to the reactor from N = 3 to 12.97 g m3 , respectively, due to an increase in t as a consequence
N = 6, maintaining the GW volume (1.2 L) and increasing the applied of the smaller electrode area in the reactor. Moreover, an increase
current from 0.630 to 1.255 A to maintain a constant current den- in N not only promotes a decrease in the treatment time but also
sity of 1.50 mA cm2 . To determine the effect of this parameter on increases the total cell voltage to values of 1.35, 2.41, 4.49 and 6.65 V
the EC process response variables, the rest of the factors remained when N is increased progressively from 3 to 6. Consequently, E
unmodied: d = 3 mm, i = 1.50 mA cm2 and LM = a. Fig. 6 shows the increased according to Eq. (5). These results show that the addi-
residual As concentration as a function of t for each A/V ratio used. tion of electrodes into the reactor does not assure the optimal
As shown in this gure, t is reduced progressively from 75 s to 40, 30 energy performance of the process. Therefore, the most favourable
and 20 s when the A/V ratio is increased from 0.350 to 0.466, 0.583 A/V ratio, 0.466 cm1 (N = 4), was selected for the following exper-
and 0.700 cm1 , respectively. This is associated with the larger FH iments.
dosage generated due to an increase in the A/V ratio; therefore,
higher As removal is produced. When A/V 0.350 cm1 , the arsenic 4.4. Current density (i) effect
removal kinetics changes from the typical exponential decay to
a sigmoidal prole, which results in a substantial increase in the To determine the effect of i on t, E and w, experiments were
treatment time. carried out with i values of 1.50, 2.25 and 3.00 mA cm2 , while
Fig. 7 shows the energy consumption of the EC process as a maintaining the rest of the investigated parameters: d = 3 mm,
function of the charge density and electrode weight loss for each A/V = 0.466 cm1 and LM = a. Fig. 8 shows the residual arsenic con-
A/V ratio investigated. As shown in this gure, the electrode con- centration in GW as a function of the treatment time as i was
varied. As shown in Fig. 8, when i increases, the arsenic removal
rate increases. This is consistent with the results reported by sev-
eral authors [2,3,6]. The treatment time decreases from 40, to 30
and 20 s when i is increased progressively from 1.50 to 2.25 and
3.00 mA cm2 , respectively. This was associated with the higher
dissolution of Fe2+ ions and the shorter time period for FH gen-
eration when i increased, with the consequent arsenic removal
increasing.
Fig. 9 illustrates the energy consumption as a function of the
charge density and the electrode weight loss for different values of
i. As shown in this gure, an intermediate i value of 2.25 mA cm2
caused the highest energy and electrode consumption; this is in
part due to the increase in U, since when i increases progres-
sively from 1.50 to 2.25 and 3.00 mA cm2 , U increases to values
of 2.41, 3.52 and 4.54 V, respectively, which is consistent with Eq.
(7). Additionally, t was not short enough to compensate for the
increment in the energy and electrode consumption with respect
to that required at i = 1.5 mA cm2 . However, the highest i value
of 3.00 mA cm2 produces a small increase of energy and elec-
trode consumption with respect to those obtained with the lowest i
Fig. 7. Effect of the A/V ratio on the energy consumption as a function of the charge
density and the electrode weight loss for the As removal from underground water value of 1.50 mA cm2 . This behaviour is due to a signicant reduc-
by the EC process performed at i = 1.5 mA cm2 , d = 3 mm and LM = a. tion of 50% in t (20 s) achieved at i = 3.00 mA cm2 with respect to
1622 J.F. Martnez-Villafane et al. / Journal of Hazardous Materials 172 (2009) 16171622
Acknowledgements
References
[1] World Health Organization (WHO), Guidelines for drinking water quality, vol.
1, 2006, p. 306.
[2] S. Chaudhari, P.R. Kumar, K.C. Khilar, S.P. Mahajan, Removal of arsenic from
water by electrocoagulation, Chemosphere 55 (2004) 12451252.
[3] H.K. Hansen, P. Nunez, R. Grandon, Electrocoagulation as a remediation tool for
wastewaters containing arsenic, Miner. Eng. 19 (2006) 521524.
[4] H.K. Hansen, P. Nunez, D. Raboy, I. Schippacasse, R. Grandon, Electrocoagula-
tion in wastewater containing arsenic: comparing different process designs,
Electrochim. Acta 52 (2007) 34643470.
[5] D.L. Cocke, J.A. Gomes, M. Kesmez, Daida, M. Weir, H. Moreno, J.R. Parga, G.
Irwin, H. McWhinney, T. Grady, E. Peterson, Arsenic removal by electrocoagu-
lation using combined AlFe electrode system and characterization of products,
J. Hazard. Mater. B139 (2007) 220231.
[6] A. Maldonado-Reyes, C. Montero-Ocampo, O. Solorza-Feria, Remediation of
drinking water contaminated with arsenic by electro-removal process using
Fig. 9. Effect of i on the energy consumption as a function of the charge density and different metal electrodes, J. Environ. Monit. 9 (2007) 12411247.
the electrode weight loss during the EC process for As removal from GW by the EC [7] C.Y. Hu, S.L. Lo, W.H. Kuan, Effects of co-existing anions on uoride removal
process carried out at d = 3 mm, LM = a and A/V = 0.466 cm1 . in electrocoagulation (EC) process using aluminum electrodes, Water Res. 37
(2003) 45134523.
[8] X. Meng, S. Bang, G. Koratis, Effects of silicate, sulfate, and carbonate on arsenic
that obtained at i = 1.50 mA cm2 (40 s), which compensates for the removal by ferric chloride, Water Res. 34 (1999) 12551261.
energy and the electrode consumption. These results show that [9] J.R. Parga, D.L. Cocke, V. Valverde, J.A. Gomes, M. Kesmez, H. Moreno, M. Weir, D.
the best i value for As removal by the EC process in the interval Mencer, Characterization of electrocoagulation for removal of chromium and
arsenic, Chem. Eng. Technol. 28 (2005) 605612.
of current densities investigated in this work was 3.00 mA cm2 . [10] N. Balasubramanian, K. Madhavan, Arsenic removal from industrial efuent
through electrocoagulation, Chem. Eng. Technol. 24 (2001) 519521.
5. Conclusions [11] J.R. Parga, D.L. Cocke, J.L. Valenzuela, J.A. Gomes, M. Kesmez, G. Irwin, H. Moreno,
M. Weir, Arsenic Removal via electrocoagulation from heavy metal contami-
nated groundwater in La Comarca Lagunera Mxico, J. Hazard. Mater. B124
According to the results obtained in this work, the factor that (2005) 247254.
had the highest effect on most of the response variables was the [12] P.K. Holt, G.W. Barton, C.A. Mitchell, The future for electrocoagulation
as a localised water treatment technology, Chemosphere 59 (2005) 355
liquid motion driving mode, and the factor that had the smallest 367.
effect in the response variables in all cases was the current den- [13] D.L. Cocke, M.Y.A. Mollah, P. Morkovsky, J.A. Gomes, M. Kesmez, J.R. Parga, Fun-
sity. The factors that generated a greater effect on the treatment damentals, present and future perspectives of electrocoagulation, J. Hazard.
Mater. B114 (2004) 199210.
time in decreasing order are LM > A/V > d > i. The factors that sig-
[14] Y. Avsar, U. Kurt, T. Gonullu, Comparison of classical chemical and electrochem-
nicantly affected the energy consumption follow the sequence ical processes for treating rose processing wastewater, J. Hazard. Mater. B148
d > LM > A/V > i, while the factors that generated a greater effect on (2007) 340345.
[15] G. Chen, Electrochemical technologies in wastewater treatment, Sep. Purif.
the electrode weight loss in decreasing order are LM > d > A/V > i.
Technol. 38 (2004) 1141.
The results of this study show that low values of energy consump- [16] H. Ayedi, G. Mouedhen, M. Feki, M. Petris, Behavior of aluminium electrodes in
tion (58.78 Wh m3 ) and electrode material loss (9.59 g m3 ) as electrocoagulation process, J. Hazard. Mater. B150 (2008) 124135.
well as the shortest residence time of 20 s were achieved during [17] W.A. Pretorius, W.G. Johannes, G.G. Lempert, Electrolytic iron occulant pro-
duction with a bipolar electrode in series arrangement, Water SA 17 (1990)
the EC process for the arsenic removal from underground water 133138.
under the following conditions: d = 3 mm, LM = a, A/V = 0.466 cm1
and i = 3.00 mA cm2 .