UT801/802 Operating Manual

6. Do not overload voltage or current on EITHER Functions

UT801/802 between terminal and terminal OR between terminal


and grounding which indicate on meter limitation.
Symbol Terminal Input
V V COM
Explanation
DC Voltage Measurement
7. The rotary switch should be placed in the right
V V COM AC Voltage Measurement
Operating Manual position and no any changeover of range shall be
made during measurement is conducted to prevent V COM Resistance Measurement
damage of the Meter. V COM Diode / Continuity Buzzer
8. Do not use or store the meter in an environment Measurement
of high temperature, humidity, flammable and kHz V COM Frequency Measurement
electromagnetic environment. The performance of A mA A COM mA/A DC Current Measurement
the meter may deteriorate after dampened. 10A COM ADCCurrent Measurement
9. The internal circuit of the meter shall not be
altered at will to avoid damage of the meter and any A mA A COM mA/A AC Current Measurement

accident 10A COM A AC Current Measurement


10. Replace the battery as soon as the battery F V mA A Capacitance Measurement
indicator Appears. With a low battery, the (Use Multi-Purpose Socket)
meter might produce false readings that can lead to V mA A Temperature Measurement
electric shock and personal injury. (Use Multi-Purpose Socket)
11. Turn the meter off when it is not is use and take hFE V mA A Triode Enlargement Factor
out the battery when not using for a long time. (Use Multi-Purpose Socket) Measurement

General Specifications Functional Buttons


1. Maximum Voltage between terminal input and Turn the power on and off Turn the display
COM: 1000V( except 200mV, 230V) backlight on and off(suitable for battery
Bench Type Digital Multimeters 2. A mA terminal input protection: (CE)250mA 265V LIGHT supply, backlight turn on about 10 seconds
auto recovery fuse after auto shut down.)
3. 10A terminal input protection: (CE)F1 (10A H Press Hold to enter and exit the hold mode
HOLD
250V) Fast type melted fuse 5x20mm in any mode, the meter beeps.
4. Resistance input protection: PTC/250V
Overview
5. Capacitance input protection: (CE)F2, F3 (0.5A H Operational Measurement Guide (see Diagram 1, 2,
Digital Bench-Type Multimeter Model UT801 is the
250V) Fast type melted fuse 5x20mm 3)
maximum reading 1999 and 3 1/2 digits and UT802
6. Frequency input protection: PTC/250V
is the maximum reading 19999 and 4 1/2 digits,
7. Temperature input protection: (CE)250mA 265V
both models are in manual range, DC / AC current
fuse
type digital multimeter, This is also the extra large
8. terminal input protection: PTC/250V
characters in LCD display backlight with full function,
9. hFE input protection: (CE)250mA 265V auto
full measurement and full overload protection as well
recovery fuse, F3 (0.5A H 250V) Fast type melted
as a good product design outlook, In addition to all
fuse 5x20mm
the conventional features include DC/AC voltage,
10. Display: LCD full function signal display,
DC/AC current, resistance, frequency, capacitance,
maximum reading is 1999(UT801) , 19999(UT802)
temperature , Transistor hFE diode and
Updates 2-3 times / second
continuity buzzer.
11. Range: Manual
This operating manual covers information on
12. Polarity Display: Auto
safety and cautions. Please read the relevant
13. Overload indication: 1
information carefully and observe all the Warnings
14. Battery Deficiency:
and Notes strictly.
15. Operating Temperature: 040(32104)
16. Storing Temperature: -1050(14122)
Unpacking Inspection
17. Relative Humidity: 030 below 75%
Open the package case and take out the Meter.
3040 50%
Check the following items carefully to see any missing
18. Electromagnetic Field: Under 1V/m the influence
or damaged part. If you find any missing or damage, Diagram 1
of radiated radio-frequency electromagnetic field
please contact your dealer in your country.
phenomenon, Total accuracy= specific accuracy+
Operating Manual 1 piece
measurement 5%, Over 1V/m radiated radio-
Test Lead 1 pair
frequency electromagnetic which do not have any
Alligator Clip 1 pair
reference data on this topic.
K Type Temperature Probe 1 piece
19. Power: AC(external power adapter AC220V/
( For the temperature under 230testing)
DC9V-200mA) or DC(internal battery type 2 R14/1.5V
Multi-Purpose Socket 1 piece
6 pieces)
Power Cord 1 piece
20. Product size: (300x245x105)mm
(AC220V 50Hz DC9V/200mA)
21. Product Net Weight : About1500g(without the
accessories)
Safety Information
22. Safety Compaliances : IEC 61010: CAT1000V
This Meter complies with the standards
IEC61010-1 in pollution degree 2, overvoltage
LCD Display
category (CAT II 1000V) and double insulation. If you
1. Manual Range Indicator of manual range
can not follow up this operating instruction to use the
2. Warning Indicator for Warning signal
meter and it reduces the chance to have an using
3. The battery is low
protection.
4. Indicator for high voltage signal
1. Before using the Meter and Test Leads inspect
5. Indicator for Negative reading Diagram 2
both items. Do not use the Meter and Test Leads
display
if it is damaged or the case (or part of the case) is
6. AC Indicator for AC voltage or current
removed or no reaction on LCD display. Prohibited
( DC indicator do not display)
to use the meter without housing or housing without
7. Data hold is active
screw fix up in order to avoid possible electric shock
8. Test of diode
or to avoid possible damage to the meter or to the
9. The continuity buzzer is on
equipment under test.
10. Number Indicates testing reading
2. If the damage of test leads, use only the same
11. Units of measurement:
model number or identical electrical specifications
mV, V Unit of Voltage: The millivolt, volt
replacement parts.
3. Do not use your finger to touch on any testing Unit of current: Microampere, milliampere,
A, mA, A
cable, connector, unused terminal input or circuit ampere
during the testing stage Unit of electrical resistance: Ohm, thousand
, k, M
4. When the meter working at an effective voltage ohms, trillion ohm
over 60V in DC or 30V rms in AC, special care should Unit of electrical capacity: Accepts the farad,
nF/F
be taken for there is danger of electric shock. the microfarad
5. Selecting the correct terminal input and turn the kHz Unit of Frequency: Kilohertz
rotary switch to select the measuring function. In Unit of Temperature: Degree Celsius Factor
case of no any idea on the value input of the current, Unit of Triode enlargement: Times Diagram 3
just simply test from the high value to low one.
P/N: 110401101618 UT801/802 Operating Manual

UT802 is about 2M . 10. Continuity Test


Warning: Maximum Voltage Input: 750Vrms Range Resolution Remarks
1. Selecting the correct terminal input and turns the Frequency: 45Hz400Hz UT801 UT802 Open circuit voltage is
rotary switch to select the measuring function. If fails 0.1*
Display: True RMS 1* approximate 3V
to do that, the buzzer beep and the warning signal
When circuit disconnected with resistance value
flash on.
Range Alarm Alert On False Terminal Input 3. DC Current >100, buzzer does not beep.
V Hz 10A mAA Range Resolution Accuracy Tolerance:(% Reading+Digits) When circuit is in good connection with resistance
mAA hFE F 10A UT801 UT802 UT801 UT802 value 10, buzzer beeps continuously.
10A mAA 200A 0.1A 0.01A
2mA 1A 0.1A 11. Transistor hFE
2.DC or AC Voltage Measurement
20mA 10A 1A (0.8%+2) (0.5%+20) Range Resolution Remarks
To avoid harms to you or damages to the Meter
from electric shock, please do not attempt to measure 200mA 0.1mA 0.01mA UT801 UT802 Ib0 is about 10A, Vce is
voltages higher than 1000 V although reading may 10A 10mA 1mA (2.0%+4) (1.5%+40)
hFE 1* 0.1* about 2.5V
be obtained. * When 5A, Continuous measurement less than 10
The Meter has an input impedance of
seconds at an interval more than 15 minutes. Replacing the battery(see Diagram 4)
10M(exceptUT802/ACV input impedance of 2M)
This loading effect can cause measurement errors in
high impedance circuits and so you need to take a 4. AC Current
note on it. Range Resolution Accuracy Tolerance:(% Reading+Digits)
R14 1.5V*6
UT801 UT802 UT801 UT802
3.DC or AC Current Measurement 2mA 1A 0.1A
20mA 10A 1A (1.0%+3) (0.8%+40)
Before connect the Meter in serial with the tested
return circuit, closed 200mA 0.1mA 0.01mA
the return circuit current to avoid the dangerous of 10A 10mA 1mA (2.5%+5) (2.0%+40)
sparking. Frequency: 45Hz400Hz
Do not use over >10A current measurement. Power adapter
* When 5A, Continuous measurement less than 10
Although the meter can work on below 20A current
testing but for the avoiding any harms to you or seconds at an interval more than 15 minutes.
damages to the Meter.
4.Measuring Resistance, Diodes, Continuity or 5. Resistance Fuse
Capacitance Range Resolution Accuracy Tolerance:(% Reading+Digits)
To maintain the measurement accuracy, discount UT801 UT802 UT801 UT802
circuit power and discharge all the high voltage Diagram 4
200 0.1 0.01
capacitors during the measuring resistance.
When measuring high resistance on 1M or 2k 1 0.1
above, it is normal to take several seconds to 20k 10 1 (0.8%+3) (0.5%+10) Specifications and other information shown on this
obtain a stable reading. In order to obtain stable 200k 100 10 instruction manual are subject to change without
reading, choose shorter test lead to carrying out 2M 1k 100 notice
measurement.
20M 10k (1.2%+5)
The test leads and the Meter inside wire will
bring around 0.10.2 of error to resistance 200M 10k (5%+40)
measurement when measuring low resistance. To When >100M resistance measurement as reference
obtain accurate readings in low-resistance, short purpose.
circuit the test leads beforehand and record the
reading obtained, call this reading as X. Then use
6. Capacitance
the equation: measured resistance value (Y) (X) =
Range Resolution Accuracy Tolerance:(% Reading+Digits)
accurate readings of resistance.
During measurement, Diodes is in a good silicon UT801 UT802 UT801 UT802
junction drops between 500mV800mV as the 20nF 10pF 1pF
(4%+3) (4%+10)
normal value. The continuity measurement, the poles 2F 1nF 100pF
between resistance is >100. it is a short circuit, but 200F* 100nF 10nF (5%+5) 5%+10)
on the poles between resistance is 10, it is a good
connection, buzzer is continually beep on, and the *>40F capacitance measurement as reference
reading value is nearly to the circuit resistance value, purpose.
Unit is .
7. Frequency
Range Resolution Accuracy Tolerance:(% Reading+Digits)
Accuracy Specifications
Accuracy: (% reading + digits), guarantee for 1 year UT801 UT802 UT801 UT802
Operating temperature: 1828 2kHz 1Hz 0.1Hz (1.5%+5) (1.2%+10)
Environmental humidity: Less than 75%RH 200kHz 100Hz 10Hz
Input Amplitude a:
1. DC Voltage
Range Resolution Accuracy Tolerance:(% Reading+Digits)
(2kHz range) 50mVa30Vrms
UT801 UT802 UT801 UT802 (200kHz range)150mVa30Vrms
200mV 0.1mV 0.01mV (0.1%+5)
2V 1mV 0.1mV 8. Temperature
20V 10mV 1mV (0.5%+2) (0.1%+3) Range Resolution Accuracy Tolerance:(% Reading+Digits)
UT801 UT802 UT801 UT802
200V 100mV 10mV
-40-20 -(8%+5) -(8%+40)
1000V 1V 0.1V (0.8%+3) (0.2%+5)
>-200 (1.2%+4) (1.2%+30)
Input Impedance: is average on 10M 1 0.1
>0100 (1.2%+3) (1.2%+25)
Maximum Voltage Input: 1000V (Except 200mV,
>1001000 (2.5%+2) (2.5%+20)
250V)
* Thermocouple: It is suitable to use K type
thermocouple. This include point contact K type
2. AC Voltage
Range Resolution Accuracy Tolerance:(% Reading+Digits)
thermocouple can only be used on less than 230
UT801 UT802 UT801 UT802 temperature measurement.
2V 1mV 0.1mV (0.8%+3) (0.5%+20)
9. Diode Test
20V 10mV 1mV
Range Resolution Remarks
200V 100mV 10mV (1.0%+4) (0.8%+40) UT801 UT802 Open circuit voltage is around 3
1000V 1V 0.1V
1mV 0.1mV V, Silicon junction drops between
Development District,Hu Men Town,Dong Guan City,
Input Impedance: UT801 is average on 10M , 0.50.8V as the normal value. Guang Dong Province,China