EVO series
Scanning Electron Microscope
EVO
Scanning electron microscope
Original instructions
354706-0780-006
This document or any part of it must not be translated, reproduced or transmitted in any form or by any means, electronic or
mechanical, including photocopying, recording, or by any information or retrieval system. Violations will be prosecuted.
The use of general descriptive names, registered names, trademarks, etc. in this document does not imply, even in the
absence of a specific statement, that such names are exempt from the relevant protective laws and regulations and therefore
free for general use. Software programs will fully remain the property of Carl Zeiss Microscopy. No program, documentation
or subsequent upgrade thereof may be disclosed to any third party, unless prior written consent of Carl Zeiss Microscopy has
been procured to do so, nor may they be copied or otherwise duplicated, even for the customers internal needs apart from a
single back-up copy for safety purposes Due to an ongoing process of improvement Carl Zeiss Microscopy reserves the right
to make modifications of this document without notice.
Contents
Table of contents
2. Safety .......................................................................................................... 13
2.1. Intended use ................................................................................................... 13
2.2. Prevention of accidents and of improper use ............................................. 13
2.3. Safety summary ............................................................................................. 14
2.3.1. Hazards related to personal injury ......................................................................... 14
2.3.2. Hazards not related to personal injury ................................................................... 17
2.3.3. Fire risk summary .................................................................................................. 19
2.4. Safety equipment ........................................................................................... 20
2.4.1. Safety devices ....................................................................................................... 20
2.4.1.1. Protective cover panels ................................................................................................. 20
2.4.1.2. Column interlock............................................................................................................ 20
2.4.1.3. Vacuum interlock system .............................................................................................. 20
2.4.1.4. Beam deceleration kit interlock system ......................................................................... 21
2.4.1.5. Main circuit breaker ....................................................................................................... 21
2.4.1.6. Main shut-off valves ...................................................................................................... 21
2.4.2. Safety labels .......................................................................................................... 21
2.4.2.1. At the rear of the electron optical column (with optional Ion Pump system only) .......... 22
2.4.2.2. At the rear of the instrument.......................................................................................... 23
2.4.2.3. Inside SEM .................................................................................................................... 24
3. Description ................................................................................................. 25
3.1. Overview ......................................................................................................... 25
3.1.1. EVO MA and LS .................................................................................................. 26
3.1.2. EVOHD ............................................................................................................... 26
3.2. Control elements ............................................................................................ 27
3.2.1. SmartSEM user interface .................................................................................... 27
3.2.2. Dual joystick .......................................................................................................... 28
3.2.3. Optional control panel ........................................................................................... 29
3.3. Principle of operation .................................................................................... 30
3.3.1. Vacuum system ..................................................................................................... 30
3.3.1.1. High Vacuum mode....................................................................................................... 32
3.3.1.2. Variable Pressure mode................................................................................................ 33
3.3.1.3. Extended Pressure mode.............................................................................................. 35
3.3.1.4. Pressure ranges and apertures..................................................................................... 36
3.3.2. Specimen stage ..................................................................................................... 38
3.3.3. Electron optics ....................................................................................................... 39
3.3.4. EVO column Optibeam modes ............................................................................ 41
5. Installation ................................................................................................... 65
6. Operation ..................................................................................................... 67
6.1. Switching on the SEM ....................................................................................67
6.2. Starting the SmartSEM user interface ........................................................68
6.3. Finding your way in the SmartSEM user interface ....................................70
6.3.1. Showing or hiding toolbars .................................................................................... 70
6.3.2. Showing or hiding the data zone ........................................................................... 71
6.3.3. Showing and hiding a full screen image ................................................................ 71
6.3.4. Using Docking panels ............................................................................................ 72
6.3.5. Opening the Panel Configuration bar .................................................................... 74
6.4. Obtaining a first image ...................................................................................75
6.4.1. Preparing the specimen holder .............................................................................. 76
6.4.2. Loading the specimen chamber ............................................................................ 77
6.4.3. Locating the specimen .......................................................................................... 81
6.4.4. Switching on the gun ............................................................................................. 82
6.4.5. Switching on the EHT ............................................................................................ 83
6.4.6. Generating, optimising and saving images ............................................................ 85
6.4.6.1. Generating an image..................................................................................................... 85
6.4.6.2. Optimising the image .................................................................................................... 87
6.4.6.3. Saving the image .......................................................................................................... 90
6.4.7. Using automated or semi-automated functions ..................................................... 91
6.4.7.1. Using the auto saturation function................................................................................. 91
This instruction manual is considered to be part of the scanning electron microscope (SEM).
Read the instructions carefully. Keep the instruction manual near the SEM and hand it over to
future owners of the instrument.
This instruction manual is designed for users who have been trained to operate the SEM by an
authorised ZEISS service representative. Operators of the SEM must not deviate from the in-
structions provided in this document.
Reference to For details on optional equipment refer to the respective manuals delivered with the SEM. You
related will find these manuals in a plastic bag.
documents
For detailed information regarding the operating software refer to the Software Manual Smart-
SEM.
For details on technical data refer to the documents Product Specification and
Installation Requirements.
1. About this manual Explains function and structure of this instruction manual
10. Parts and tools Lists consumables, spare parts, tools, and accessories
14. Index Alphabetical list of key words that are referred to in this instruction
manual
The safety instructions in this manual follow a system of risk levels, that are defined as follows:
DANGER
This safety symbol and signal word indicates an imminently hazardous situation.
Disregarding this warning WILL result in death or serious injury.
WARNING
This safety symbol and signal word indicates a potentially hazardous situation.
Disregarding this warning COULD result in death or serious injury.
CAUTION
This safety symbol and signal word indicates a potentially hazardous situation.
Disregarding this warning MAY result in minor or moderate injury.
CAUTION
This signal word used without a safety symbol indicates a potentially hazardous situation.
Disregarding this warning MAY result in property damage.
IMPORTANT
This symbol and signal word draws your attention to important and useful
information.
For the description of software, the following typographical conventions are used:
Typography Meaning
Type <key1, key2> Type key 1 first, then type key 2 on the keyboard.
Type <Ctrl + Alt + Del>. Simultaneously type CTRL key, ALT key and DEL
key on the keyboard.
Click the Magnification icon. Icons, buttons, and menus are printed in bold.
Select File/Exit from the menu.
Enter 10 kV in the EHT target field. Values to be selected are printed in italics.
ZEISS service representative Specially trained service expert, either ZEISS staff or authorised serv-
ice partner of ZEISS.
2. Safety
The SEM is a microscope that scans a focused beam of electrons across the specimen to
generate an image.
The SEM is suited to analyse surface structures and near-surface structures of appropriate spec-
imens. For that purpose the specimen has to be located in the specimen chamber with controlled
pressure environments.
Commercial The SEM is to be used in a laboratory environment for commercial purposes only. Do not use the
use only SEM for any other purpose. Use the SEM only as instructed in this manual.
Using the SEM for any other purpose is not allowed and could be hazardous.
CAUTION
Risk of injury or damage due to improper operation of the SEM.
Operator Within the scope of initial start-up the ZEISS service representative will deliver a basic operator
training training. The basic operator training consists of fundamental operation procedures including safe-
ty instructions. An introduction to basic maintenance tasks will be given for an administrator of the
system.
The training performed shall be documented appropriately.
Special application training is offered on request.
IMPORTANT
Any persons undertaking maintenance, service and repair tasks not described in this in-
struction manual have to be authorised ZEISS service representative.
Follow the safety instructions given in this instruction manual. This is essential to prevent damage
and to protect yourself and others against accidents and unsafe practices. Do not deviate from
the instructions provided in this instruction manual.
This section summarises possible hazards and recommended safety procedures.
Service tasks
DANGER
Danger to life: Hazardous voltage inside the SEM.
Only service engineers trained and authorised by ZEISS service representative are al-
lowed to service the SEM.
WARNING
Magnetic fields can interfere with medical implants.
Ion getter pumps generate a magnetic field that can interfere with medical implants, such
as cardiac pacemakers.
The magnetic field is also present if the SEM is switched off.
When performing any maintenance tasks at the column, such as replacing filaments, it is
possible to get close to the ion getter pumps and their magnetic field.
If you have a medical implant fitted, always keep a safety distance of 10 cm to the ion get-
ter pumps.
WARNING
Magnetic fields
Strong magnetic fields can disturb electronic devices like heart pacemakers or impair
their function.
Maintain a safety distance of 10 cm between the magnet and the
heart pacemaker.
Radiation X-rays are produced within the SEM during operation. This is unavoidable since accelerated elec-
protection trons hit material thus generating radiation.
WARNING
Radiation hazard: X-rays are generated inside the SEM during operation.
Only authorised ZEISS service representatives are allowed to service the SEM.
Do not remove any parts. Do not disable any parts of the interlock system.
Use genuine ZEISS parts exclusively.
Observe all safety and X-ray protection regulations.
In the UK, the operation of the SEM is permission-free as the following requirements are fulfilled:
The acceleration voltage is limited to 30 kV.
The local dose rate at a distance of 0.1 m from the accessible surface of the SEM does not
exceed 1 Sv/h.
A respective label is attached to the SEM.
Outside the UK, the user of the SEM has to comply with the local regulations of the country where
the SEM is operated.
The SEM is equipped with several radiation protection devices, which ensure - under regular op-
eration conditions - that the SEM operates in accordance with German X-ray protection regulation
(RV) as well as with the EC Directive 96/29/EURATOM.
Electrical
connections
CAUTION
High leakage current
Ensure proper grounding.
Do not operate the SEM without separate ground connection.
Gases Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange.
Compressed air is used to operate several valves and the auto levelling system.
CAUTION
Suffocation hazard due to lack of oxygen, since the specimen chamber is vented with
gaseous nitrogen. Inhaling nitrogen may cause unconsciousness.
During specimen exchange, keep the chamber door open as short as possible.
Avoid inhaling the air from within the specimen chamber.
Ensure the area around the SEM is sufficiently ventilated.
IMPORTANT
Concerning the hazards of nitrogen installations and associated safety precautions refer
to guideline IGC Doc 44/xx/E: Hazards of inert gases, published by EIGA (European Indus-
trial Gases Association) which can be found on the EIGA homepage www.eiga.org/Publi-
cations/Documents. Check regularly if new documents concerning nitrogen installations
have been uploaded.
CAUTION
Risk of injury or damage due to the high internal pressure in gas cylinders (e.g. containing
nitrogen or compressed air).
Observe all safety labels on the gas cylinders and all safety instruction given by the gas
cylinder manufacturer.
Operation There are risks of injury and of damage to equipment if you operate the SEM incorrectly.
CAUTION
Risk of injury
Fingers could be trapped in the moving specimen stage.
Always close the chamber door before you move the specimen stage.
CAUTION
Pinch hazard when closing the chamber door.
Use the recessed grip to close the chamber door. Ensure not to get your fingers caught in
the chamber door gap.
CAUTION
Risk of damage to environment due to aggressive or toxic chemicals.
When disposing of waste that has been generated during a service operation (e.g. used
rotary pump oil) comply with all national and local safety and environmental protection
regulations.
IMPORTANT
We recommend to observe regulations for ergonomic work environment, such as environ-
mental and local health and safety rules (VDU work place assessment).
You should maintain a healthy posture and take regular breaks.
Maintenance LaB6 only: Baking out the gun head has to be performed as a regular maintenance procedure
procedures and when needed.
Only advanced operators are allowed to perform the bakeout procedure.
CAUTION
Burn hazard
Some parts inside the SEM will get hot during the bakeout procedure.
Do not place any combustible objects on the top of the electron optical column.
Only authorised ZEISS service representatives is allowed to service the equipment. Dis-
connect power and let surfaces cool before opening.
CAUTION
Burn hazard
The firing unit gets hot during operation.
Allow time to cool before handling during exchange procedure.
CAUTION
Risk of property damage
SEM or specimen could be damaged if specimen stage is at a short working distance when
opening the chamber door.
Always move specimen stage to a long working distance before opening the chamber
door.
CAUTION
Risk of property damage
Connect ZEISS-approved equipment only.
Ensure the total load connected to the SEM does not exceed 10 A.
IMPORTANT
Fingerprints can cause vacuum leaks.
Always wear lint-free gloves when touching the specimen or inner parts of the specimen
chamber.
In case of fire, small amounts of fumes may be emitted from various materials within the SEM.
In summary, it can be stated that there is only a low probability of fire originating from the SEM.
In order to prevent any risk of hazard to human health or of property damage, the SEM is
equipped with several safety and protective devices.
WARNING
Hazardous voltage inside the SEM. Contact may cause burn or electric shock.
X-rays are generated inside the SEM during operation.
Do not remove any parts.
The SEM must not be operated with removed protective cover panels.
Appropriate safety labels on the SEM warn you of possible hazards. Each safety label is fixed
near the point where a particular hazard exists.
Do not remove labels from the SEM. If damage occurs to any label(s) so that you cannot read it
fully, contact ZEISS to request a new label. Fit the new label in place of the damaged label.
2.4.2.1. At the rear of the electron optical column (with optional Ion Pump system
only)
B Safety
information CAUTION
Hot Surface
Skin burns under contact.
Do not touch.
This area may get hot.
C Safety
information WARNING
Hazardous voltage inside
Contact may cause electric shock or burn.
Disconnect power before opening.
D Safety
information CAUTION
Magnetic Field
Interaction with metallic objects may produce pinch
hazards.
Persons with medical implants keep back 12 inches.
E Legal
information
F Safety
information CAUTION
X rays produced when energized
The microscope should be approved by the national
regulations.
Do not remove any parts.
3. Description
3.1. Overview
5
4
2 Plinth 5 Monitors
3.1.2. EVOHD
EVOHD is equipped with a high brightness source, a LaB6 emitter, that increases the gun
brightness at low accelerating voltages. Images taken at 1 kV appear 100 times brighter than
a standard tungsten and 10 times brighter than a standard LaB6 source. This is achieved with a
factory pre-aligned firing unit that is supplied by ZEISS.
The operation of a HD gun is almost identical to tungsten and LaB6, any user familiar with the
standard EVO family of machines will operate the HD gun by simply changing probe current and
acceleration voltage as required.
All EVO series microscopes provide the possibility to use the EasyVP feature (except EVO18
SEMs); the ability to seamlessly switch between High Vacuum and Variable Pressure.
EasyVP allows using the mid-column 20 m aperture as the beam defining aperture. The aper-
ture is aligned using focus wobble in both High Vacuum and Variable Pressure. This enables the
use of all Optibeam modes when operating the SEM under Variable Pressure conditions.
A maximum pressure of 133 Pa can be achieved using any of the charge compensating gases,
air or water.
The SEM is controlled by the SmartSEM software. The software is operated via a graphical user
interface.
slider
data zone Panel Configuration bar
status bar
arrow button
MiniBar annotation bar image area mouse assignment
The dual joystick is used for stage control and specimen navigation.
All axes are deflection-compensated: When the joystick is only moved slightly, the respective axis
will move slowly. However, major movements of the joystick will result in a faster movement of
the stage.
The X-, Y-, and Z-axes are magnification-compensated. When working at a low magnification, the
stage moves relatively fast. At higher magnifications the stage movement is slower.
The different axes can also be moved simultaneously.
The control panel is optionally available. It integrates a full sized keyboard and allows direct ac-
cess to 14 of the most frequently used functions on the SEM.
The following functions are available through:
Shift Y (14)
Brightness (15)
Contrast (16)
Focus (18)
5 7 9 11 13 15
4 6 8 10 12 14 16
2 17
18
1
IMPORTANT
All these functions are available via the SmartSEM software and can be controlled by ap-
propriate mouse movement and MACRO execution.
The SEM uses a beam of electrons to generate an image or to analyse the specimen. The
electron beam scans the specimen surface.
For operation of the SEM, gun head (1), column and specimen chamber have to be under vacu-
um. The vacuum is essential to operate the gun and to prevent collision of electrons with gas
molecules.
1
2
3
4
5
6
7
8
System Rotary pump (7) and turbo pump (8 ) pumps the specimen chamber. The system vacuum is
vacuum measured by a Penning gauge (2). As long as the detected pressure in the specimen chamber
has not reached the operating target, the chamber shut off valve (5) is closed.
Gun vacuum An ion getter pump is used to keep the gun area under high vacuum (2).
The vacuum in the gun head is called Gun vacuum. It should be well below 5 x 10-7 mbar.
The specimen holder is fitted in the chamber on the stage dovetail part when the chamber is at
air (3). The specimen can be exchanged after selecting the Vent command via the SmartSEM
user interface.
Venting When receiving the Vent command, venting gas flows into the specimen chamber via the vent
valve (6). As soon as the pressure equilibrium is obtained, the chamber door can be opened to
change the specimen.
Pumping to In order to continue operation, the Pump command pumps the specimen chamber using the pre-
Vacuum pump (rotary) and turbo.
As soon as the vacuum in the specimen chamber is ready for operation, EHT Vac ready
message appears in the SmartSEM user interface. Gun and EHT can then be switched on.
Both gun and chamber are at high vacuum (1). DPA (differential pumping aperture)(2). Isolation
valve (3) open.
Application High Vacuum mode is the regular operation mode for standard applications.
Variable Pres- The VP 100 m aperture is fitted under the final lens when the SEM is configured for VP. The 750
sure configura- m mid-column aperture is used for imaging in this configuration. The beam first goes through
tion the 750 m mid-column aperture and then passes through the 100 m VP aperture (pressure lim-
iting aperture) before landing on the specimen surface.
In VP configuration, only the Normal and Analysis Optibeam modes can be used for imaging. It
is though possible to switch between HV and VP modes when using this configuration. The max-
imum pressure that can be achieved in the VP configuration is 400 Pa for EVO W systems, 273
Pa for EVO LaB6 and EVOHD SEMs.
EasyVP config- In the Easy VP configuration, the EasyVP, 750 m aperture, is fitted under the final lens and the
uration 20 m mid-column apertures is used for the beam control. In the EasyVP configuration, the beam
first goes through the 20 m mid-column aperture and then passes through the 750 m EasyVP
aperture (pressure limiting aperture) before landing on the specimen surface. The maximum
pressure that can be achieved in the EasyVP configuration is 133 Pa for all EVO series SEMs.
Optibeam modes such as resolution, field, depth, fisheye and analysis modes can be used for
imaging, in the EasyVP configuration. It is possible to switch between HV and VP modes and the
beam alignment is optimum in both conditions without changing any gun or column parameters.
Operating The residual gas atmosphere in the specimen chamber creates an interaction region of electrons
principle and residual gas molecules between the final lens and the specimen. In this region, high-energy
electrons in the primary electron beam hit the residual gas molecules and ionise them. The ions
generated in these collisions contribute to the compensation of negative charge on the specimen.
However, another effect of these collisions is to scatter the electron beam. This is called the skirt
effect. The electrons that are lost from the primary beam by this effect provide only a resolution-
limited background signal for imaging purposes. Although it is possible to tolerate these leakage
losses at chamber pressures up to a few hundred Pa, it is necessary to carefully select and con-
trol the important factors such as acceleration voltage, chamber pressure and beam path. Al-
though operating in VP mode compensates for the charging effects, the increased chamber
pressure reduces the signal-to-noise ratio. This can be compensated by using the noise suppres-
sion features of SmartSEM.
3
2
Gun at high vacuum (1). Chamber at variable pressure (2). Single DPA (differential pumping ap-
erture) fitted in the bottom of objective lens; Variable Pressure (100 m aperture) or EasyVP (750
m aperture) (3) Isolation valve closed (4).
+ or
2
Gun at high vacuum (1). Chamber at extended pressure (2). Two differential pumping apertures
are fitted in the bottom of the objective lens (3). Isolation valve closed (4). The rotary pumps are
directly pumping the chamber (5).
Min/Max
SEM Application Hardware Filament
pressure*
Min/Max
SEM Application Hardware Filament
pressure*
Min/Max
SEM Application Hardware Filament
pressure*
MA 10 Charge 10-400 Pa W
MA 15 compensation
MA 25 VP aperture
10-273 Pa LaB6
VP aperture
Min/Max
SEM Application Hardware Filament
pressure*
* Both air and water vapour can be introduced, as a charge compensating gas, into the chamber.
Mid-column The mid-column click aperture assembly design allows using 4 apertures. However, only three
click aperture apertures are fitted as standard. The position of each aperture is marked, as shown below, on the
mid-column aperture assembly.
EVO
1 30 m
2 20 m
EVOHD
1 20 m
2 20 m
Standard specimen stage is a 5-axes motorised Cartesian stage that is controlled by the Smart-
SEM software. The stage can be operated by the dual joystick controller or by using the soft
joystick in the SmartSEM user interface.
The five axes are called:
X X-axis
Y Y-axis
Z Height
R Rotation
T Tilt
The focus on the Cartesian stage is not maintained when the specimen is tilted.
This can be compensated for by using the Compucentric function in SmartSEM.
The EVO column is the area of the SEM, where electrons are emitted, accelerated, deflected,
focused, and scanned. Main characteristics of the EVO optics are the condensor and objective
lenses.
1
2
UEHT
3
5
6
7
8
2 Gun alignment and emission display coils 6 Objective lens with through the lens
pumping
Gun alignment The electron beam is centred by the gun alignment and emission coils (2).
and display
coils
Condensers A double condenser system (3) allows the continuous regulation of the probe current.
Apertures The electron beam passes through the currently selected aperture in the mechanical mid-column
aperture changer (4) to ensure optimum quality of beam is maintained.
Scanning and The scanning coils (5) move the electron beam in a point-by-point scan over the specimen
stigmator coils surface.
The stigmator coils (5) compensate for astigmatism, so that the electron beam becomes
rotationally symmetrical.
Objective lens The electron beam exits the objective lens (6).
Signal When the primary electron beam hits the specimen, certain interaction signals are generated,
detection which are recorded by various detectors, such as the CZ BSD (7).
Optibeam
Characteristics Beam path
modes
Analysis mode The specimen remains in focus for any change in probe
current.
Field mode A larger field of view for navigation and a large depth of field.
Optibeam
Characteristics Beam path
modes
Resolution mode Smallest probe diameter for a chosen probe current at any
working distance or keV.
Depth mode Largest depth of field for a chosen probe current, at any
working distance or keV.
Fisheye mode Largest field of view for navigation and a very large depth of
field.
The interaction products most frequently used for the generation of images in scanning electron
microscopy are secondary electrons (SE) and backscattered electrons (BSE).
Detected
Standard detectors Typical application
signals
Detected
Optional detectors Typical application
signals
*By means of measuring current produced by ionisation of gas by SE Electrons generated around
specimen surface.
IMPORTANT
For more details, refer to the document Product Specification EVO.
Secondary Secondary electrons are ejected from the outer atomic shell of the specimen material upon
electrons impact by the primary electron beam.
The three main categories of secondary electrons are based on the origin of the secondary
electrons and on the distance from the PE impact point,
where they leave the specimen.
SE1 electrons are generated and leave the surface of the specimen directly at the spot where
the primary electron beam impacts on the specimen surface.
SE2 electrons are generated after multiple scattering inside the interaction volume, and leave
the specimen at a greater distance from the primary beams impact point.
SE3 electrons are generated by backscattered electrons colliding with the chamber walls or
the lens system.
Secondary electrons have low energy (less than 50 eV).
Backscatter All electrons with energy higher than 50 eV are known as backscattered electrons (BSEs).
electrons
BSEs are generated by elastic scattering in a much deeper range of the interaction volume and
carry depth information.
The backscatter coefficient increases with increasing atomic number of the elements within the
specimen. This allows the BSE detector to generate atomic number contrast-, or compositional
contrast images.
2 1
1 Specimen 3 SE detector
Operating When the high energy electrons hit the scintillator layer (2), photons are generated inside the
principle scintillator.These photons are directed towards the light pipe (3) and are transferred to the pho-
tomultiplier (5). The photomultiplier multiplies the flashes of light and outputs a signal that can be
used for imaging.
5 3
2
2 Scintillator 5 Photomultiplier
3 Light pipe
Collector bias The Detectors tab of the SEM Controls panel allows adjusting the collector bias voltage in the
voltage range of -250 V to +400 V in steps of 1 V. This voltage generates a bias in front of the detector.
The bias attracts the low energy SE electrons and accelerates them towards the detector. For
allstandard applications the collector bias voltage is usually set to 300 V.
It is also possible to set the collector bias voltage to a negative value. This generates a field that
deflects the secondary electrons, preventing them from reaching the scintillator and contributing
to the signal. Backscattered electrons are not affected significantly by the negative bias voltage
and reach the scintillator to contribute to image information. This allows the generation of a pseu-
do-backscattered image, which shows enhanced topographical information.
Surface images that show enhanced topographical information can also be generated using BSE
detectors, but they do not show the shadows that can be created using the SE detector.
The working distance has a significant effect on the efficiency of the SE detector. Shadowing ef-
fects occur when the working distance is too short. If the specimen is too close to the final lens,
most of the electrons will be deflected by the field of the electrostatic lens or move to the final lens
itself. This means they cannot be detected by the SE detector.
Depending on the specimen material and on the specimen geometry, a minimum working
distance of approximately 4 mm should be used. Extreme signal loss is likely to occur if shorter
working distances than this are used. Conversely, the SE detector is very good when used for
imaging at long working distances. This is particularly important for low magnification imaging that
is necessary for adjusting the orientation of the specimen holder or locating a specific area on the
specimen.
Optimal The following settings provide a good field of view for navigating on the specimen at low
initial settings magnifications:
Initial working distance in the range of 10 mm to 20 mm.
Acceleration voltage of approximately 10 kV.
VP/EP:
750 m For VP/EP mode only.
1 VPSE G3 detector
Chamber When using a VPSE G3 detector, it is important to remember that a certain pressure is required
pressure in the specimen chamber to generate enough photons to be detected. If the number of residual
gas molecules is too low, the collision probability and therefore the efficiency of the detector
reduces. Depending on the specimen used and the operating parameters, optimum detection
occurs in the pressure range of 20 Pa to 60 Pa.
IMPORTANT
Increasing the chamber pressure also increases the scattering process, and reduces the
resolution of the SEM. It is therefore essential to find optimum conditions for each
particular application.
Dwell time Another key parameter is the length of time that the primary electron beam dwells on the
specimen. If this dwell time is too short at a fast scan rate, there is not sufficient time for an ion
cascade to develop or for the residual gas molecules to generate the imaging photons.
This reduces the detector efficiency.
If the dwell time is too long at a slow scan rate, very high energy is applied to the specimen per
time unit. This can result in charging artefacts on the obtained images. Optimum scan rate for
different specimens might vary and can only be determined by experiment.
IMPORTANT
To reduce charging effects, use the Frame Averaging function in SmartSEM.
This function uses fast scan speeds and increases the number of frames.
Collector The collector bias sets the electric field between the specimen and the VPSE G3 detector.
bias If the collector bias is too low, the VPSE G3 detector efficiency is significantly reduced.
Too high collector bias values result in saturation of the VPSE G3 detector depending on the
specimen, the acceleration voltage, the probe current and the pressure in the specimen chamber.
IMPORTANT
The banding effects can be eliminated by reducing the collector bias or by reducing the
pressure in the specimen chamber. There is also a calibration dialog available that can
help you to further improve the image quality (Tools/Goto Panel/VPSE-G3 Calibration).
The HDBSD is a highly sensitive BSD detector avialable with either four or five diode segments.
The HDBSD provides the ability to visualise exceptionally fine surface detail in high contrast using
low incident beam energies.
Life sciences
Mineral deposits in plant structures
Bone structures
IMPORTANT
For detailed information on EVO applications please search for
Electron & Ion Beam Microscopy on www.zeiss.com/microscopy and select Applications.
3.4. Specification
3.4.1. EVO
Chamber 310 mm () x 220 mm (h) 365 mm () x 275mm (h) 420 mm () x 330 mm (h)
X = 80 mm X = 125 mm X = 130 mm
Y = 100 mm Y = 125 mm Y = 130 mm
Z = 35 mm Z = 50 mm Z = 50 mm
5-Axes Motorised T = 0 - 90 T = 0 - 90 T = 0 - 90
Specimen Stage R = 360 (continuous) R = 360 (continuous) R = 360 (continuous)
Stage control by mouse or Stage control by mouse or Stage control by mouse or
optional joystick and control optional joystick and control optional joystick and control
panel panel panel
Maximum Specimen 100 mm (with the ZTR mod- 145 mm (with the ZTR mod- 210 mm (with the ZTR mod-
Height ule removed) ule removed) ule removed)
Image Framestore max. 3072 x2304 pixel, signal acquisition by integrating and averaging
3.4.2. EVOHD
1.9 nm @ 30 kV SE
3 nm @ 30 kV SE VP mode
Resolution 10 nm @ 30 kV 1 nA
5 nm @ 3 kV SE
8 nm @ 1 kV SE
10 - 273 Pa
Pressure Range
10 - 400 Pa (with optional TTL upgrade)
X = 130 mm X = 125 mm
Y = 130 mm Y = 125 mm
Z = 50 mm Z = 50 mm
5-Axes Motorised T = -10 - 90 T = -10- 90
Specimen Stage R = 360 (continuous) R = 360 (continuous)
Stage control by mouse or Stage control by mouse or
optional joystick and control optional joystick and control
panel panel
145 mm (with the ZTR module 210 mm (with the ZTR module
Maximum Specimen Height
removed) removed)
IMPORTANT
For more details refer to the document Product Specification EVO.
C D
B
E
A
F
I II
III
IV
E Nitrogen supply
F Exhaust line
1
2
1
1
Size (mm)
Weight (kg) Distribution
No Description approx. Footprints
approx. of load
length x width x height
Location requirements
Installation category II
Electrical supplies
Protection class I
Nominal frequency 50 - 60 Hz
Circuit breaker 25 A
Protective ground High leakage currents can be present in the SEM. Therefore, the SEM has
to be connected to an equipotential bonding bar. An exclusive grounding
connection to earth must be provided, i.e. the grounding terminal must not
be common to other electrical equipment.
A grounding wire AWG10 (5 m long) is delivered with the SEM.
Gas supplies
Nitrogen
Compressed air
Quality oil-free
Environmental requirements
Stability 0.5C/h
Pollution degree 2
IMPORTANT
Also refer to the documents Product Specification EVO and Installation
Requirements EVO.
3.6. Options
A range of options and accessories is available for use with the EVO SEM. Examples of
the available options and accessories are as follows:
Detectors VPSE, BSD/HDBSD, CL, STEM, SCD
Additional Chamberscope or Stubscope
Specimen holders and stage accessories, including Coolstage, Right-hand tilt, Faraday cup
Software add-ins and enhancements
Column maintenance kit, O-ring kit
For full details of the available options and accessories, please contact your local ZEISS service
representative or sales representative.
For customer service, please contact your local ZEISS service representative.
A list of ZEISS locations and authorised service partners can be found at:
www.zeiss.com/microscopy
IMPORTANT
To maintain best possible performance of the SEM it is essential to perform preventive
maintenance on a regular base.
Moreover, it is recommended that you conclude a service contract with your local
ZEISS service organisation or representative. This will ensure a continuous trouble-free
operation of the SEM.
4.1. Transport
CAUTION
Crushing hazard while load is being moved or lowered.
Maintain a safe distance. Do not walk or place your hands or feet under the load while it is
being lowered. Wear safety shoes and gloves.
CAUTION
Risk of injury due to heavy components.
Rotary pumps and vibration blocks are heavy and can cause injuries when lifted by one
person alone.
Be careful when transporting the rotary pump and the vibration block.
At least two people are required to lift the rotary pump or the vibration block.
CAUTION
Risk of damaging the SEM
The SEM may only be transported in air-suspended vehicles. Moving parts must be se-
cured during transport to prevent them from slipping or tipping over.
Avoid rocking the crates back and forth.
Devices for transporting the SEM must be rated to handle its full weight and dimensions.
Note the weight information on the package and on the shipping document.
Check that none of the items has been damaged during shipment.
4.2. Storage
5. Installation
Unpacking, installation and first start-up are carried out by an authorised ZEISS service repre-
sentative.
6. Operation
Preconditions:
The SEM is switched on.
The Windows operating system has been loaded.
The EM Server opens while loading various software and firmware. The function of the EM Serv-
er is to implement the internal communication between software and hardware of the SEM.
The EM Server is minimised to a small element (icon) on the right side of the Windows task bar.
Several toolbars such as user toolbar, status bar, and annotation bar are available for easy
access to the SmartSEM functions.
The data zone is used to display current parameters. You can also include a micron marker to
show the base magnification. The default toolbar is shown below.
You can take advantage of the full monitor size by showing a full screen image.
The maximum size of the image is fixed at 1024 x 768 pixels.
It is possible to dock various control panels onto the main window. The purpose of the docking
panel is to keep the area of the image completely clear, as the docking panel is outside the main
image display.
3 To move a docking panel to the docking area, pick up the panel by clicking the title bar and
drag and drop it into the docking area.
4 To stick a control panel to the docking panel, click the title bar of the control panel and drag
it to the docking panel.
The following section summarises basic sequences to quickly obtain an image using the SE de-
tector. To simplify the procedure, the method described mainly uses SEM Controls panel and
status bar functions.
Prerequisite:
SmartSEM has been started and is ready to control the SEM.
lint-free gloves -
IMPORTANT
Contamination caused by fingerprints can lead to vacuum deterioration or prolonged
pumping times. Always wear lint-free gloves when touching specimen, specimen holder
or stage.
CAUTION
Risk of damaging the objective lens and/or your specimen
Ensure not to hit the objective lens while driving the stage. Change to TV mode to observe
the moving stage.
CAUTION
Suffocation hazard due to lack of oxygen, since the specimen chamber is vented with
nitrogen.
After the specimen exchange, keep the chamber door open as short as possible.
Avoid inhaling the air from within the specimen chamber.
Ensure the area around the SEM is sufficiently ventilated.
IMPORTANT
Contamination caused by fingerprints can lead to vacuum deterioration or prolonged
pumping times.
Always wear lint-free gloves when touching specimen, specimen holder or stage. Keep the
chamber door open as short as possible.
CAUTION
Pinch hazard when closing the chamber door.
Use the recessed grip to close the chamber door. Ensure not to get your fingers caught in
the chamber door gap.
CAUTION
Risk of damaging the objective lens and/or your specimen.
Ensure not to hit the objective lens while driving the stage. Watch the moving stage in TV
mode.
2 Move the specimen by using the dual joystick (optional) or by calling the Soft Joystick from
the menu bar.
IMPORTANT
After loading the specimen, there is the possibility to automatically set a number of key
parameters. Refer to section 6.4.7.5. for details.
EHT stands for Extra High Tension. This voltage has to be applied to the gun in order to make
it emit electrons.
The following workflow describes how to obtain images with the SEM. To make it more compre-
hensive, the workflow is separated into generating, optimising and saving images.
IMPORTANT
The following procedures describe how to manually perform the necessary steps.
There are also several automated functions available that can help you with these tasks.
For details, refer to section 6.4.7.
13 Set the focus: Press the middle mouse button and drag the mouse to focus. Alternatively, use
the Focus knob on the control panel.
IMPORTANT
The following procedure describes how to manually improve the image quality. There is
also an automated function available that can help you with this task. For details, refer to
section 6.4.7.3.
1 Set the mouse or control panel adjustment to Coarse by toggling the Coarse/Fine button in
the status bar. This can also be achieved by pushing <TAB> on the keyboard.
2 Select the Apertures tab.
3 Activate the Focus Wobble checkbox (1).
IMPORTANT
Once the alignment is optimised, DO NOT change the position of the X and Y micrometer
gauges on the mid-column click aperture. It is now possible to work at any EHT, Iprobe
and working distance without the need to change the position of the X and Y micrometer
gauges on the mid-column click aperture.
12 Slowly increase the magnification to the value required (e.g. Mag 50.000 x). It is prudent to
refocus the image if and when required whilst increasing the magnification.
13 On the toolbar, click REDUCE.
It is now possible to change the EHT, Iprobe and working distance to any value without
changing the position of the X and Y micrometer gauges on the mid-column click aperture.
When the acceleration voltage (EHT) or the Iprobe values is changed during the imaging, use the
automatic aperture align for aligning the beam (see section 6.4.7.3.).
3 Set the preferred imaging conditions (example 10 keV, 8.5 mm WD and 300 pA) and obtain
an image.
4 From the Panel Configuration bar, select the Aperture Align dialog.
5 Click Auto Aperture Align.
The software starts aligning the beam and the progress bar appears on the panel.
A full size image is displayed when the alignment is completed.
4 to improve the absolute accuracy of the drift correction, select a higher store resolution than
1024x768.
5 To save the reference image, click Create
Reference.
The drift correction process will use the imaging conditions set when the reference was cre-
ated and when it is complete the imaging settings will be restored.
7 Once the reference rectangle is selected and the reference image has been saved, click Hide
Rectangle to continue working on the specimen.
If a drift or movement of the specimen is encountered during the imaging process, i.e. the dis-
tinct feature moves away from the reference area then click Do SEM Drift Corr.. This will
modify the beam shift such that the area of the specimen scanned matches the original posi-
tion. The command will be executed from a macro.
Parameters in Maximum Pixel Error specifies the amount of pixel mismatch the system tolerates when
the Settings identifying the reference feature. The default value is 1 pixel.
group In general a smaller value is better as it results in higher accuracy of the drift correction. Using
a value of zero can however cause the drift correction to fail to match a valid image. Using a
value of 1 is recommended for most applications. This value is the minimum accuracy of a
successful drift correction but the corrected drift may obviously be more.
Minimum Confidence specifies the system certainty in identifying a feature on the specimen
as being the reference feature. Choosing too high numbers may lead to the system failing to
locate the reference feature whereas choosing too low values may lead the system errone-
ously identifying other features as being the reference feature. The default setting is 50%.
Drift Maximum Tries specifies the maximum number of times that the system should loop
through the drift correction algorithm in order to identify a feature that matches the reference
feature. The default is 5 times but a good match is usually found in one or two tries.
Period Drift Correction is used when an area is imaged over a prolonged period of time and
the time interval between the corrections (periods, seconds) can be predefined and entered
in the field.
Additional system and imaging conditions can be selected under the options.
A dialog appears asking you to provide more information about the sample under analysis.
When the information regarding the specimen status is provided, the software runs macros
that are assigned to the different specimen types, with predefined parameters and vacuum
settings.
Editing or
adding new It is possible to edit or add new sample types to the Sample Type Selection menu.
sample types
1 In order to add a new sample type to the menu obtain an image from the desired specimen.
2 From the Panel Configuration bar, select Sample Type Save.
The SEM conditions, vacuum mode and imaging parameters will then be saved for the new
specimen type.
The Sample Type Save dialog displays the new sample type.
You have successfully added the new sample type. It can now be selected from the Sample
Type Selection menu.
3 Click on the External Image and load a camera picture of the specimen holder with speci-
mens.
Follow the wizard (defining 3 points on the specimen holder and correlating those points to
the camera picture).
6 Finish the manual registration setup by selecting the Yes option in the dialog, to save the cur-
rent image and registration data.
IMPORTANT
It is also possible to save the Image and the specimen holder position in a different folder.
7 To save the specimen holders camera picture in a chosen folder, click Save Image in the
Registration Image section.
8 To save the specimen holder position, click Save in the Manual Registration section.
Procedure:
1 From the Panel Configuration bar, select Automated Imaging.
You can switch between Multi mode and Add mode on a stub or stubs to build up a list any
way you like.
Specimens that are studied in the SEM can be divided into two main categories, namely
conductors and non-conductors and these will be discussed in greater detail later.
However there are several factors to consider during all specimen preparation and these
are given below.
1 The size and weight might necessitate some reduction of the specimen to fit holders and to
ease specimen manipulation for observation.
2 Mineral and metallurgical specimens may require polishing. Etching and electropolishing may
also be required.
3 The specimen should be able to withstand the vacuum of the SEM as it might become dam-
aged or deformed.
4 The specimen should be clean and dry if it is to be used in high vacuum mode, i.e. free of
dust, moisture, oils and grease, as their presence can lead to charging, contamination and
longer pump down times. Otherwise use VP or EP vacuum mode.
5 Porous samples will also take a long time to pump out in high vacuum mode.
Coat non-conductors to prevent charging if allowed or VP mode is not available.
6 Observation of the specimen at low kV (start at 1 kV) is another good alternative
where charge balance can be obtained.
7 The specimen should be firmly attached to the specimen stub or holder either
mechanically or by gluing. Silver dag and carbon dag can be used as conductive glues for
small samples. Carbon tabs/tape are also convenient but may not provide good mechanical
stability.
8 There should be good electrical connection between the surface of the specimen and speci-
men stub or holder for non-conductors.
9 Health and Safety procedures regarding the handling of the specimen during its preparation
should be observed.
The mid-column click aperture is used for changing the mid-column apertures. There are three
apertures fitted in the assembly (20 m, 30 m and 750 m) and the position of each aperture is
fixed and labelled. The apertures can easily be changed by rotating the manual adjustment knob.
Suggested The conditions given below are suggestions and the operator may find that different values will
settings give better information for the types of specimens that are being investigated.
MC = Mid-Column Aperture
PLA = Pressure Limiting Aperture
MC = 30 m
Cycle time to reduce noise = 20
EDS
EHT = 20 kV for metals and minerals
EHT = = 8 to 10 kV for semiconductors and organic materials
Filament I set to first peak for Qualitative analysis
Filament I set to second peak for Quantitative analysis
Iprobe = 1000 pA or adjust for 30 % deadtime
WD = 8.5 mm for a 35 take off (elevation) angle
Detector = BSD
Final aperture = 30 m
Cycle time = 20 or longer for X-ray mapping
High resolution
EHT=30 kV Iprobe = 10 pA WD = 5 mm
Filament I set to 2nd peak and untick the Lonflife Mode
MC = 20 m
Detector = SE with collector bias + 400 V
Cycle time to reduce noise =1.3 mins or longer
Warning remove the BSD to its parked position
EVO can be operated in different Optibeam operation modes, depending on the type of appli-
cation. Optibeam takes the requirements for probe current, working distance, etc. and determines
the optimum lens settings to achieve the best performance from the column.
Some Optibeam modes might not be available when the fixed aperture is fitted.
Fisheye mode requires the optional SmartSEM software licence FISHEYE.
Analysis mode:
Specimen remains in focus for any change in
probe current.
Field mode:
A large field of view for navigation with a long
depth of field.
Resolution mode:
Smallest probe diameter for a chosen probe
current at any working distance or keV.
Depth mode:
Largest depth of field dor a chosen probe cur-
rent at any working distance or keV.
Fisheye mode:
Extreme field of view for navigation and a very
large depth of field.
IMPORTANT
For some MA EVO microscopes the option to select all the possible EP and VP aperture
variants is not an option. When this is the case selecting the Change Apertures tab will not
open the full aperture selection page but will instead toggle between No Aperture, Easy
VP and VP Aperture Without Beam Sleeve.
IMPORTANT
When using EVO in EasyVP mode it essential to align the beam through the 20 m mid-
column aperture. The resolution being determined by the 20 m mid-column aperture.
IMPORTANT
For some MA EVO microscopes the option to select all the possible EP and VP aperture
variants is not an option. When this is the case selecting the Change Apertures tab will not
open the full aperture selection page but will instead toggle between No Aperture, Easy
VP and VP Aperture.
IMPORTANT
When using EVO in VP mode it is essential to align the beam through the 750 m mid-
column aperture. The resolution being determined by the 100 m (pressure limiting) VP
aperture.
18 Adjust contrast to make the outer region of the emission image visible around the brighter,
central part of the Emission Image. It should look like a fried or poached egg.
20 Adjust the emission image with gun shift and gun tilt so that the bright centre is centred within
the emission image and that the emission image is centred on the screen.
It is recommended to use the arrow buttons instead of the sliders for easier adjustment.
22 Focus on a feature on the specimen and still keep the WD to 8.5 mm.
23 Select a fast scan speed e.g. 1.
24 On the Apertures tab, turn on the focus wob-
ble function and use the fine adjustment
screws on the mid-column aperture to align
the beam.
25 To perform a hysteresis correction, type
<Shift+F2>.
26 If necessary, repeat steps 20 to 25 to realign
the beam accurately.
27 To complete the alignment, continue with the steps described in section 6.4.6.
IMPORTANT
From now on do not touch or move the fine adjustments screws on the mid-column aper-
ture assembly.
Procedure:
1 Load the Faraday cup into the specimen
chamber.
2 Pump the specimen chamber.
3 Switch on the electron beam.
4 Set a magnification that allows transmission of
the complete electron beam into the cavity
through the aperture orifice.
5 Open the Panel Configuration bar.
The specimen current is measured permanently. The measured value is displayed in the field
Specimen I =.
The BSD Control panel allows you to change the polarity of the segments, select BSD modes
and set the BSD gain.
5-segment BSD
BSD Imaging Stage biasing (beam deceleration) is a method used for decelerating the electron beams landing
with Stage energy by applying a negative potential to a specimen. The lower landing energy of the primary
Biasing electrons provides the possibility to obtain more specimen surface detail especially at low mag-
nifications. Better image resolutions can also be obtained at high magnifications.
Beam deceleration is generally used when imaging specimens with BSE detectors. The stage
bias functionality can only be applied in the high vacuum mode in the range of 0 to -5 kV. The
effectiveness of the stage biasing and beam deceleration will solely depend on the specimen
type. Imaging improvements can easily be achieved when using metallic specimen; more surface
detail will be revealed.
Landing energy is the difference between the energy of the primary beam electrons and the bias
applied to the stage. If the EHT voltage is set to 5kV and a stage bias potential of -4kV is applied,
the landing energy would be 1kV.
The beam deceleration option can also be used for applying low voltages on specimens in the
range of +20V to -20V. The SE detector is normally used in combination with low voltage speci-
men biasing. Although, the low specimen biasing effects are subtle, an improvement in the image
contrast can easily be seen. The low voltage specimen biasing can be applied under the HV and
VP conditions.
The beam deceleration kit is mounted through a plate on the stage door. A specially designed 9
stub specimen holder is generally used when imaging specimen with the stage biasing function-
ality. This is to isolate the specimen holder, i.e. the bias is only applied to the specimen.
1 2 3
The HDBSD images below show the effect of applying a bias to an aluminium fracture spec-
imen. The images are taken at 5 kV accelerating voltage, in HV and at a working distance of
8.5 mm.
1 2
Fig. 6.1: Al fracture surface imaged at 0 V bias (1). Al fracture surface imaged at -4400 V
bias (2).
When the stage bias is 0 volts the landing energy of the electron beam is 5 kV. The landing
energy of the electrons is decelerated to 600 V as the stage bias of 4400 V is applied.
More morphological information is revealed as the landing energy of the electron beam is re-
duced.
IMPORTANT
If the Beam deceleration carousel checkbox is activated, then the Stage bias low voltage
option is not available, but the beam deceleration option is available.
Despite sharing the same electronics, low voltage and high voltage are mutually exclu-
sive.
Low voltage demands the beam deceleration carousel is removed (the carousel is a spe-
cial sample holder designed to withstand the high voltage and also improve the imaging).
High voltage demands the beam deceleration carousel is fitted. The user tells the system
this via the checkbox. This checkbox is used to enforce rules.
Using stage
bias low
voltage
IMPORTANT
LV Biasing can work with any standard, conducting, dovetail mounted carousel; it will not
work with lens mounted sample holders and other non-dovetail mountings.
Preconditions:
Stage has been initialised
HV mode is active
Standard conducting dovetail-mounted sample holder fitted
Procedure:
1 From the Panel Configuration bar, select Stage Navigation.
CAUTION
Risk of stage collision
If you choose a voltage between -5 V and +5 V, a warning is displayed saying that touch
alarm is disabled. In this voltage range, the touch alarm becomes unreliable and there is
a risk of stage collision going undetected.
Always move the bias voltage to a larger value before moving the stage.
A Peltier cooling stage and an Extended Pressure (EP) EVO system is required for imaging hy-
drated specimens. This will enable to study hydrated specimen in their native state with little or
no loss of water in the SEM (the sample is kept fully hydrated during the pump down).
CAUTION
Risk of damaging the Peltier stage pipe by using stage rotation.
Only move the stage in X and Y.
28 Select the USB connection and tick the Humidity Option checkbox.
The aim of purging is to remove air from the water bottle and to fill the chamber with just water
vapour.
How to If the water kit has recently been fitted, filled with water, or has not been used for a while, con-
continue tinue with procedure 1.
If the system has recently been used in the wet mode, continue with procedure 2.
6.5.5.4. Setting SEM parameters for imaging the specimen in wet mode
The SEM operating parameters depend on the type of specimen being examined and the nature
of the detail you are expecting to see.
It is recommended to start with:
EHT = 20 kV
I Probe = 400 pA or higher
WD = 6.0 mm
Detector = BSD or EPSE
5 Select a pressure around 100 Pa higher than the Purge Min value and enter that as
Purge Max value. The Purge Cycle can now be changed to 3.
6 Vent the chamber without changing any of the parameters.
7 Remove the plain stub and dry the area underneath the stub on the Peltier coolstage.
Carefully mount the specimen on a new stub, place and tighten the stub on the Peltier
coolstage.
8 Use a pipette for putting a few droplets of distilled water on the Peltier coolstage holder.
IMPORTANT
Do not put any water droplets on the stub or directly on the specimen as the water droplets
can cover the specimen surface and prevent viewing the areas of interest.
9 Move the Peltier coolstage slightly to the right or left of the chamber before closing the cham-
ber door. This will prevent any droplets splattering up on the EPSE, BSD or the final lens area
during the pump down process.
10 Close the chamber door and pump down the system.
11 Run up the gun and move the sample under the beam using chamberscope and/or image
navigation panel.
12 Start imaging in wet mode; if necessary increase the pressure but in very small steps. If nec-
essary, adjust the humidity level values to achieve 100% humidity. But the latter will change
both the Peltier temperature and the chamber pressure.
13 Increasing the pressure in large steps results in accumulation of water on the specimen and
prevents the imaging.
The SmartSEM user interface offers a multitude of help texts containing information on the
operation of the SEM, the optimization of the images and the handling of accessory
options.
1 Press <F1>.
Alternatively, select Help/SmartSEM help from the menu.
1 Press <SHIFT+F1>.
Alternatively, select Help/Whats This from
the menu.
Many functions and menus which are often used in the SmartSEM user interface can also be
opened using the keyboard. A list of short cuts (key combinations) can be displayed in the Smart-
SEM help.
1 Press < F9>.
Alternatively, select Help/Keys help from the menu.
It is recommended that when using a Tungsten filament the gun is turned off when the system is
not being used for any significant period of time.
It is recommended that when using a LaB6 filament that the gun is not turned off during its service
life time. To obtain the maximum life time of a LaB6 emitter the EHT must be switched off when
not in use and for long periods of time the gun should be switched to STANDBY.
The service life time can be further extended by Long Fil Life in the Gun tab of the SEM Controls
panel. This will reduce the filament current to First peak conditions which will give satisfactory
operating conditions for general microscopy.
If optimum performance is required the
Long Fil. Life checkbox should be unticked.
6.8.2. Exiting
Change to STANDBY mode when the SEM is not operated, even for longer periods.
The filament will continue to be heated, and the vacuum in electron optical column and specimen
chamber will be maintained.
IMPORTANT
Before switching on the SEM, ensure that the reason for the emergency off does not exist
any more and that it is safe to switch on the SEM.
This procedure completely cuts off the SEM from the electrical main supply.
WARNING
Magnetic fields can interfere with medical implants.
Ion getter pumps generate a magnetic field that can interfere with medical implants, such
as cardiac pacemakers.
The magnetic field is also present if the SEM is switched off.
When performing any maintenance tasks at the column, such as replacing filaments, it is
possible to get close to the ion getter pumps and their magnetic field.
If you have a medical implant fitted, always keep a safety distance of 10 cm to the ion get-
ter pumps.
CAUTION
Risk of injury. Risk of property damage.
Perform only such tasks described in this instruction manual.
All pursuing tasks of maintenance, service, and repair not described in this instruction
manual have to be performed by authorised ZEISS service representatives.
IMPORTANT
To maintain best possible performance of the SEM it is essential to perform preventive
maintenance on a regular basis.
Moreover, it is recommended that you obtain a service contract with your local
ZEISS service organisation or representative.
This will ensure a continuous trouble-free operation of the SEM.
Scheduled downtime is the time when the SEM is not available to perform its intended function
due to planned downtime events.
Change of consumables and chemicals
Basic preventive maintenance performed by the operator
Preventive maintenance performed by authorised ZEISS service representatives
The isolation mounts lose air over time. Therefore it is necessary to adjust them from time to time.
CAUTION
Crushing hazard.
The gap between isolated table and plinth is designed to allow the specimen chamber to
swing freely.
The size of the gap varies during operation and can cause crushing injuries.
Do not place your fingers between isolated table and plinth.
1 To check if the air feet need to be adjusted, give the column a gentle nudge.
- If it is unable to move freely, adjust the air feet as described in the following steps.
- If the column is able to move freely, no further action is required.
There are three inlet valves at the back of the plinth. Each of the inlet valves corresponds to a
different section of the levelling system.
2 To adjust the air feet, remove the caps from the inlet valves at the back of the plinth.
Refer to the instruction manual of the pre-vacuum pump provided by the pump manufacturer.
You will find this documentation in the SEM document folder.
Possible reasons:
Filament has come to the end of its normal service life
IMPORTANT
You can display the operating hours of the filament by selecting View/SEM Status/
Select/Filament Age. The Filament Age will be displayed in the Display tab.
7 2
4
5
CAUTION
Risk of contamination.
Dust particles and skin grease can cause bad vacuum and flashovers.
Always wear gloves when changing the filament.
IMPORTANT
Where a column interlock is fitted as described section 2.4.1.2. the requirement to fully
switch off the SEM is removed. The firing unit change is the same process for machines
with and without the column interlock, however the image below shows how the interlock
is separated when the column lid is opened and care must be taken when closing the gun
to ensure the interlock is correctly lined up when the gun is closed.
1 Shutdown Gun.
CAUTION
Burning hazard
The EO column gets hot during operation.
After switching off the SEM, wait for fifteen minutes before opening the top section of the
EO column.
8 Wait the firing unit to cool down for at least fifteen minutes.
IMPORTANT
The tensator spring washer is not designed for pre-aligned filaments. If the tensator spring
washer is installed with a pre-aligned filament, the height of the filament will be wrong.
Do not reinstall the tensator spring washer when installing a pre-aligned filament.
IMPORTANT
The tungsten filament deposits tungsten on the firing unit over time.
These deposits have to be removed after every filament exchange.
1 2
CAUTION
Risk of contamination.
Dust particles and skin grease can cause bad vacuum and flashovers.
Always wear gloves when changing the filament.
IMPORTANT
Where a column interlock is fitted as described section 2.4.1.2. the requirement to fully
switch off the SEM is removed. The firing unit change is the same process for machines
with and without the column interlock, however the image below shows how the interlock
is separated when the column lid is opened and care must be taken when closing the gun
to ensure the interlock is correctly lined up when the gun is closed.
If the filament holder shows tungsten deposits, remove the deposits as described in
section 7.2.7.
IMPORTANT
The tungsten filament deposits tungsten on the filament holder over time.
These deposits have to be removed after every fifth filament exchange.
This also applies to the anode (refer to section 7.2.7.2.).
2 3
Table 7.1: Setting the filament distance by turning the brass retaining washer
CAUTION
Risk of contamination.
Dust particles and skin grease can cause bad vacuum and flashovers.
Always wear gloves when changing the filament.
IMPORTANT
Where a column interlock is fitted as described section 2.4.1.2. the requirement to fully
switch off the SEM is removed. The firing unit change is the same process for machines
with and without the column interlock, however the image below shows how the interlock
is separated when the column lid is opened and care must be taken when closing the gun
to ensure the interlock is correctly lined up when the gun is closed.
If the filament holder shows tungsten deposits, remove the deposits as described in
section 7.2.7.
IMPORTANT
The LaB6 filament deposits sublimed material on the filament holder over time.
These deposits have to be removed after every filament exchange.
2 3
IMPORTANT
Accurate centring and height adjustment are essential to ensure optimum performance
and maximum service life.
IMPORTANT
The filament runup will take a while for the first few times when a new filament is installed.
This is quite normal and is designed to minimise damage due to thermal shock when the
filament is new.
IMPORTANT
To achieve optimum performance, the emission image must be aligned to the centre on
the image area. This is achieved by using Beam Tilt and Beam Shift on the Apertures tab.
CAUTION
Risk of contamination.
Dust particles and skin grease can cause bad vacuum and flashovers.
Always wear gloves when changing the filament.
IMPORTANT
Tungsten and LaB6 firing units are not compatible with each other.
IMPORTANT
It is not possible to install a LaB6 firing unit on an instrument built for Tungsten.
This can only be achieved by carrying out an instrument upgrade which requires return of
the unit to the factory.
THIS UPGRADE PROCEDURE IS NOT POSSIBLE AT CUSTOMER LOCATION.
IMPORTANT
Where a column interlock is fitted as described section 2.4.1.2. the requirement to fully
switch off the SEM is removed. The firing unit change is the same process for machines
with and without the column interlock, however the image below shows how the interlock
is separated when the column lid is opened and care must be taken when closing the gun
to ensure the interlock is correctly lined up when the gun is closed.
1 Shutdown Gun.
CAUTION
Burning hazard
The gun at the top of the EO column gets hot during operation.
After switching off the SEM, wait for fifteen minutes before opening the top section of the
EO column.
8 Wait the firing unit to cool down for at least fifteen minutes.
9 Open the gun.
If the filament holder shows tungsten deposits, remove the deposits as described in
section 7.2.7.
IMPORTANT
The tungsten filament deposits tungsten on the filament holder over time.
These deposits have to be removed after every filament exchange.
This also applies to the anode (refer to section 7.2.7.2.).
2 3
Table 7.2: Setting the filament distance by turning the brass retaining washer
IMPORTANT
The filament runup will take a while for the first few times when a new filament is installed.
IMPORTANT
To achieve optimum performance, the emission image must be aligned to the centre on
the image area. This is achieved by using Beam Tilt and Beam Shift on the Apertures tab.
If a stored stage position cannot be approached or if absolute stage movement is required, the
stage needs to be initialised.
Executing this function requires the Stage initialise privilege in the user profile.
Prerequisites:
Specimen chamber has been vented.
All large specimens and large specimen holders are removed.
Procedure:
1 Select Stage/Stage initialise from the menu.
2 Confirm by clicking on Yes.
IMPORTANT
If initialisation of the stage does not solve the
stage problem, contact your local
ZEISS service representative.
For maintenance of the pre-vacuum pump refer to the instruction manual of the pump provided
by the pump manufacturer.
You will find this documentation in the SEM document folder.
For optimum use of the LaB6 filament it is essential that the gun vacuum is always in the
10 -7 mbar region (10 -8 mbar for EVOHD). If the vacuum falls below this value it is strongly
recommended that the gun chamber bakeout procedure is carried out which would typically be
an overnight procedure.
IMPORTANT
In order to maintain peak operating condition, it is recommended to perform a weekend
bakeout at least twice a year.
4 Double-click Bakeout.
CAUTION
Risk of injury from hot surfaces during bakeout Parts of the enclosure in the upper range
of the column may become hot during bakeout, particularly after a long bakeout cycle. Do
not touch any parts of the cover panel or place any combustible objects on the electron
optical column.
The pressure will be changing due to residual heat retain in the gun area and a cooling period
is necessary to allow the vacuum to improve. The cooling period depends on the room tem-
perature. The gun monitor will monitor the pressure improvement as the cooling time
progresses and reached the pressure 10 -7 mbar region (10 -8 mbar for EVOHD).
O-rings are necessary to keep up the vacuum and protect the column and the chamber from
dust particles and grease. The column and chamber door O-rings have to be checked and
cleaned
regularly.
Possible reasons:
Regular maintenance
Bad vacuum
CAUTION
Risk of damaging the O-rings.
O-rings can be damaged when treated with any solvents such as acetone or alcohol.
Do not use any solvents to clean the O-rings.
IMPORTANT
Grease on O-rings can attract dust causing it to build up on the vacuum seal which can
lead to vacuum leaks. Some greases can migrate into the vacuum chamber leading to poor
vacuum and contamination of samples, detectors and other SEM components.
Do not apply grease to O-rings.
CAUTION
Suffocation hazard due to lack of oxygen, since the specimen chamber is vented with
nitrogen.
Avoid inhaling the air from within the specimen chamber.
Ensure the area around the SEM is sufficiently ventilated.
IMPORTANT
Contamination caused by fingerprints can lead to vacuum deterioration or prolonged
pumping times.
Always wear lint-free gloves when touching specimen, specimen holder or stage. Keep the
chamber door open as short as possible.
Procedure:
1 Vent the specimen chamber.
2 Open the chamber door.
CAUTION
Danger of damaging the sealing surface when using metallic tools.
If required, use a plastic or wooden tool to remove the chamber door O-ring.
3 Check visually if there are any dust particles or grease on the O-ring (1).
CAUTION
Risk of damaging the aperture.
The filament deposits material on the firing unit and the anode over time.
This can damage the aperture and reduce the filament life time.
Clean the firing unit after every filament change.
8 Rinse the parts contaminated with polish with warm water containing a few drops of mild de-
tergent.
CAUTION
Risk of property damage.
Residual polishing compound can cause flashovers and bad vacuum.
To remove residual polishing compound, grease and other hydro carbon contaminates it
is essential to clean all the firing unit parts using a weak solvent such as isopropanol
(propan2ol CAS No67635).
8 Further clean the anode by placing the tray containing the anode and the retaining screws in
an ultrasonic bath as described in section 7.2.7.2.
IMPORTANT
Contamination caused by fingerprints can lead to vacuum deterioration or prolonged
pumping times.
Always wear lint-free gloves when touching the cleaned anode and retaining screws.
9 Remove all parts from the cleaning solvent and allow to dry. It is essential that the anode and
retaining screws are handled using clean lint free gloves from this point of the procedure.
Cleaned anode
The annual preventive maintenance is performed by the local ZEISS service representative.
It includes
Inspection
Preventive actions
Equipment test
Verification run
7.4. Repair
The repair tasks described in the following section can be performed by the operator.
IMPORTANT
All pursuing tasks of repair not described in this instruction manual have to be performed
by authorised ZEISS service representatives only.
Possible reasons:
Chamber door does not close tightly, bad chamber vacuum
CAUTION
Suffocation hazard due to lack of oxygen, since the specimen chamber is vented with
nitrogen.
Avoid inhaling the air from within the specimen chamber.
Ensure the area around the SEM is sufficiently ventilated.
IMPORTANT
Contamination caused by fingerprints can lead to vacuum deterioration or prolonged
pumping times.
Always wear lint-free gloves when touching specimen, specimen holder or stage. Keep the
chamber door open as short as possible.
Procedure:
1 Vent the specimen chamber.
2 Open the chamber door.
CAUTION
Danger of damaging the sealing surface when using metallic tools.
If required, use a plastic or wooden tool to remove the chamber door O-ring.
Possible reasons:
Circuit breaker is tripped (lower position)
Procedure:
If this does not solve the problem, contact your local ZEISS service representative for assistance.
IMPORTANT
Tripped circuit breakers may be a hint for an electrical problem in the SEM.
If a circuit breaker is tripping again, switch off the SEM completely and contact the
ZEISS service for assistance.
8. Troubleshooting
DANGER
Danger to life: Hazardous voltage inside the SEM.
Only service engineers trained and authorised by ZEISS are allowed to service the SEM
and to perform work on the electrical system of the SEM.
SEM SEM does not operate. Circuit breaker is tripped (lower Check the circuit breakers.
position). Refer to section 7.4.2.
Vacuum Vac ready = OK is not shown System vacuum is bad due to a Check the chamber door seal for
after specimen exchange. vacuum leak at the chamber cleanliness.
door. If required, replace the chamber
door seal.
Vac ready = OK is shown very Gas ballast at rotary pump or Deactivate gas ballast at the pre-
late after specimen exchange. scroll pump is activated. vacuum pump.
System Vac ready = OK is indicated Penning gauge has not been Restart the SEM:
vacuum abnormally fast. identified correctly. If this does not solve the problem,
call your local ZEISS service rep-
resentative.
System Bad system vacuum. Chamber door seal does not Replace the chamber door seal.
vacuum seal tightly. Refer to section 7.4.1.
Specimen Stage does not move. Stage needs to be initialised. Initialise the stage. Refer to sec-
stage tion 7.2.3.
If this does not solve the problem,
call your local ZEISS service rep-
resentative.
Specimen Stored position cannot be Stage needs to be driven to a Initialise the stage.
stage approached correctly. well-defined position. Refer to section 7.2.3.
Drift Specimen seems to be moving. Charging effects. Ensure proper conduction of the
Non-conducting specimen. specimen.
Optimise specimen preparation.
Apply a charge compensation
method.
Image quality Image is bad at low EHT Working distance is too long. Reduce working distance to a
(e.g.1 kV) maximum of 7 mm.
SE2 image SE2 image is noisy Scintillator is used up. Call your local ZEISS service
engineer to have the scintillator
replaced.
After emer- Stored stage position cannot be Stage needs to be initialised. Initialise the stage.
gency off or approached correctly. Refer to section 7.2.3.
power failure
If the SEM will not be used for an extended period of time e.g. several months, it should be put
out of operation.
Contact your local ZEISS service representative to have the SEM put out of operation.
9.2. Disposal
The operator must ensure that solid waste (consumables) are disposed of and recycled in a
responsible manner.
The operator must ensure that waste products are disposed of and recycled in a responsible
manner.
Refer to EC directive 2002/96/EC on waste electrical and electronic equipment (WEEE).
The SEM consists of several modules. Be careful to separate the materials properly when you
dispose of the SEM.
Materials: e.g. metals, non-metals, composite materials, process materials
Electronic scrap material: e.g. transformers, circuit boards, cables
CAUTION
Risk of injury or property damage when using inappropriate parts or tools.
Use genuine ZEISS parts only.
Order parts and tools at your local ZEISS service organization.
For customer service please contact your local ZEISS service representative.
A list of ZEISS locations and authorised service partners can be found at:
www.zeiss.de/microscopy
Small pliers -
Stubs -
Tweezers -
Gloves, lint-free -
11. Abbreviations
EC European community
H Height
M M-axis
PC Personal computer
PE Protective Earth
R R-axis
SE Secondary Electron
T T-axis
U Voltage
W Width
WD Working distance
X X-axis
Y Y-axis
Z Z-axis
ZTR Z-Tilt-Rotate
12. Glossary
Aperture Small opening in the beam path that forms and limits the electron or ion
beam.
Astigmatism Lens aberration that distorts the shape of the electron beam, compen-
sated by the stigmator.
Backscattered Electrons High energy electrons that are liberated from the specimen surface when
the specimen is hit by the primary electron beam.
Condenser Device that collects and focuses the electron beam onto the specimen.
Depth of field Distance along the optical axis which an object in the specimen can be
moved while remaining in focus
Faraday cup Small Insulated metal container, equipped with an aperture where elec-
tron can enter but not escape. Used to measure the specimen current in
the SEM.
Focus wobble Function that sweeps the focus of the objective lens backwards and for-
ward through the focus on the specimen plane. When the aperture is mis-
aligned a lateral shift is observed.
Optibeam The Optibeam program takes the requirements for probe current, working
distance, etc. and determines the optimum lens settings to achieve the
best performance from the column. Optibeam operates in several differ-
ent Modes that can be selected on the SEM Control Apertures tab.
Some Optibeam modes are not available when the fixed aperture is fitted
Penning gauge Device for measuring high vacuum in the vacuum system.
Primary electrons Narrowly focused beam of accelerated electrons that hit the specimen
surface.
Secondary electrons Low energy electrons that are liberated from the specimen surface when
the specimen is hit by the primary electron beam.
Secondary electrons are generated by inelastic scattering.
X-ray Type of ionising radiation that is generated during the operation of elec-
tron microscopes
Model: EVO
ZEISS Ltd.
511 Coldhams Lane
Manufacturer
Cambridge CB1 3JS
UK
This is to declare that the machinery mentioned above fulfils all the relevant provisions of the
Directive 2006/42/EC.
CE marking The CE conformity marking is located on the type plate of the machinery.
14. Index
A Fuses 230
Abbreviations 237
Acceleration voltage 39 G
Administrator 13 Gas ballast 231
Analysis mode 41 Gun vacuum 31
Anode 39
Apertures 39 H
Auomated image acquisition 99 Help 146
Auto aperture alignment 93 High leakage current 16
Auto gun alignment 92
Auto saturation 91 I
Automated functions 91 Image navigation 98
Initialise stage 215
B Installation 65
Bakeout 216, 239 Intended use 13
Beam deceleration 126 Interlocks 20
C K
Chamber door interlock 21 Keyboard 29
Chamber door seal 228
Charging 232 L
Chemicals 159 Location requirements 60
Compressed air 231
Conformity 241 M
Consumables 159, 235 Main shut-off valves 21
Control panel, optional 29 Maintenance overview 159
Conventions, typographical 11 Material Safety Data Sheet 24
Cover panels 20
Customer service 62 N
Nitrogen 16, 231
D
Declaration 241 O
Detectors 43 Operating modes 41
Disposal 233 Operation, safe 215
Documents, related 9 Operator 12
Dose rate 15 Operator training 13
Downtime 159 Operator, advanced 12
Drift 232 Operator, basic 12, 13
Drift correction 94
Dual joystick 28 P
Penning gauge 231, 239
E Power failure 232
EHT 39 Preventive maintenance 159, 215, 227
Emergency off 154, 232
Evacuating 31 R
Radiation 14, 15
F Radiation protection 14
Faraday cup 119, 239 Resolution mode 42
Field mode 41
Filament 39 S
Fisheye (column mode) 42 Safety 10, 13, 14
Fisheye mode 42 Safety devices 20
Safety instruction 10
Safety labels 21, 215
Sample type selection 96
Scintillator 160, 239
Seismic protection 21
Service 14
Shutdown 233
SmartSEM 12
Software 27
Software Manual 9
Solid waste 233
Spare parts 236
Specimen chamber 25
Specimen stage 38
Stage bias low voltage 127
Stage biasing 124
Start-up 65
Storage 64
Switching off 156
System vacuum 30, 231
T
Terms 12
Tip seal 160, 162
Tools 236
Transport 63
Troubleshooting 231
U
User 12
User interface 27
V
Vacuum 231
Vacuum leaks 18
Vacuum system 30
Ventilating 31
Voltage, hazardous 14
W
WEEE 234
X
X-rays 15
www.zeiss.com/microscopy