Download as pdf or txt
Download as pdf or txt
You are on page 1of 14

Designation USBR 6515

Procedure For
Using Piezometers to Monitor Water
Pressure in a Rock Mass
This procedure is under the jurisdiction of the Materials Engineering and
Research Laboratory, code 86-68180, Technical Service Center, Denver,
Colorado. The procedure is issued under the fixed designation USBR 6515. The
number immediately following the designation indicates the first year of acceptance
or the year of last revision.

1. Scope from American Society for Testing


and Materials, International Society
1.1 Explanation.-This designation of Rock Mechanics, Bureau of
covers the description, application, Reclamation, and other sources.
selection, installation, data collection,
and data reduction for using 2. Applicable Documents
piezometers to monitor water pressure
in a rock mass. 2.1 USBR Procedures:
USBR 3000 Using Significant Digits
1.2 Application.-The procedure in Calculating and Reporting
contained in this designation applies Laboratory Data
to both hard rock and soft rock.
USBR 3910 Standard Terms and
1.3 Units.-Stating the values in Symbols Relating to Rock Mechanics
SI/metric (inch-pound) units is to be USBR 9300 Checking, Rounding, and
regarded as standard. Reporting of Laboratory Data

1.4 Caveats.-This designation 2.2 ASTM Documents:


does not purport to address all the ASTM D4750 Standard Test Method
safety issues associated with its use for Determining Subsurface Liquid
and may involve use of hazardous Levels in a Borehole or Monitoring
materials, equipment, and operations. Well (Observation Well)
The user has the responsibility to
establish and adopt appropriate safety ASTM D 5092 Standard Practice for
and health practices. Also, the user Design and Installation of Ground
must comply with prevalent Water Monitoring Wells in Aquifers
regulatory codes while using this
procedure. ASTM D 5093 Standard Test Method
for Field Measurement of Infiltration
1.5 Sources.-This designation Rate Using a Double-Ring Infiltro-
reflects the information available meter with a Sealed-Inner Ring
USBR 6515-09

2.3 ISRM Suggested Methods: 3. Summary of Method


Suggested Methods for Pressure
Monitoring Using Hydraulic Cells, 3.1 Select.-Based on the
"Rock Characterization Testing and evaluation of available geological
Monitoring," ISRM Suggested information, carefully select the
Methods, Brown, E.T. (ed.), locations of boreholes and the
Pergamon Press, 1981, pp 203-211 collection zones for the boreholes
where piezometers are to be installed.
2.4 Other Publications:
Concrete Dam Instrumentation 3.2 Drill.-Carefully drill the
Manual, Pressure Measuring Devices, boreholes straight with no drilling
Sections 2.3, 2.4, 2.5 and 2.6, Bureau mud. If the boreholes cave use
of Reclamation, pp 20-36, 1987. biodegradable drilling fluids. Drill
the boreholes about 300 mm (12 in)
Embankment Dam Instrumentation below the proposed location of the
Manual, Pressure Measuring Devices, piezometer tip. Wash the borehole
Sections 27, 28, 29, 30, 31, 32, 33 and clean of drill cuttings.
34, Bureau of Reclamation, 1987, pp
17-56. 3.3 Backfill.-Backfill the bottom
of the borehole with clean fine sand to
Dunnicliff, John, Measurement of a point about 150 mm (6 in) below
Groundwater Pressure, Chapter 9, the piezometer tip position.
Geotechnical Instrumentation for
Monitoring Field Performance, John 3.4 Place.-Lower the piezometer
Wiley and Sons, 1988, pp 117-164. into position and place clean fine sand
around the piezometer to a point 150
Drilling of Exploratory Holes, millimeters (6 in) above the
Observation Wells and Piezometers, piezometer.
Section 7-7, Ground Water Manual,
U.S. Department of the Interior, 3.5 Seal.-Immediately above the
Water and Power Resources Service, sand packed zone, seal the borehole
Denver, 1981, pp 201-204 with either (1) alternating layers of
Groundwater Data Acquisition bentonite and sand backfill, tamped in
Methods, Chapter 9, Engineering place for about 300 mm (12 in)
Geology Field Manual, Bureau of followed by common backfill or (2)
Reclamation, pp 167-183. an impermeable bentonite-cement
grout mix.
Instrument Installations, Designation
E-27, Instructions for Installing and 3.6 Tamp.-If multiple piezometers
Reading Hydraulic-Type Twin Tube are to be placed in a single borehole,
Piezometers in Earth Dams, and tamp the bentonite-sand plugs in
Designation E-28, Instructions for place below and above the
Installing and Reading Porous-Tube piezometers and also at intervals
Piezometers, Earth Manual, Bureau between the piezometer zones.
of Reclamation, Denver, 1974, pp
650-699.

2
USBR 6515-09

3.7 Record.-From the initial and 4.5 Absolute Pressures.-A


subsequent recorded readings of vibrating wire piezometer is the only
piezometer data over a period of time, device that can measure negative pore
calculate and report water pressure pressures (because it indicates
variations. absolute pressures).

4. Significance and Use 5. Description of Terms Specific


to This Designation
4.1 Groundwater Presence.-The
presence of ground water or other 5.1 Aquifer.-A body of rock that
liquids and the depth of water table or is sufficiently permeable to conduct
liquid level below the ground surface ground water or other liquid and to
have considerable significance in yield sufficient quantities of water or
geotechnical, geological, other liquid.
hydrological, waste-management
planning/storage, and other design 5.2 Aquitard.-A confined bed that
investigations and studies. These retards but does not totally prevent
investigations or studies typically the flow of water or other liquid to or
may include completion of from an adjacent aquifer.
observation wells, performing aquifer
tests, and monitoring of the water or 5.3 Aquiclude.-A geological
other liquid levels with the passage of formation that confines the flow of
time. Alternatively, frequently read water.
piezometers can provide similar
information at substantial cost and 5.4 Collection Zone.-The aquifer
time savings. zone that influences the flow of water
or other liquid toward a piezometer
4.2 Stratum Characteristics.- and also the pressure exerted on the
Observation wells have hydraulic installed piezometer diaphragm.
connectivity vertically of different
stratum and cannot yield the 5.5 Piezometer.-Any of several
hydrologic characteristic of a single geotechnical instruments that are used
stratum. for measuring the pore pressure in a
soil or rock stratum.
4.3 Pressures.-Piezometers can
monitor hydrostatic water pressure in 5.6 Other Terms.-See USBR
a single stratum or a small stratum 3910.
zone and yield pore pressure data of
that single stratum. 6. Apparatus

4.4 Strength.-Piezometer data can 6.1 System Components.-A


be used to indirectly estimate the piezometer system has three main
effective stress and thereby the index components: (1) a piezometer, (2) a
of rock mass strength. conductor cable, and (3) a readout
box and several auxiliary
components.

3
USBR 6515-09

twin tube piezometers, open stand


6.2 Piezometer.-A vibrating wire pipe piezometers, pneumatic
piezometer, commonly used by piezometers, electrical resistance
Reclamation, is shown on figure 1. piezometers, push-in piezometers, and
Its stainless steel housing nests the heavy liquid piezometers are usually
different components of the not used by Reclamation.
piezometer such as (1) the filter, (2) Reclamation uses twin-tube
the diaphragm, (3) the vibrating wire, piezometers and open stand pipe
(4) the plucking and pickup coils, (5) piezometers for monitoring pore
the internal bulkhead seal, and (6) the pressures in embankment and
cable connectors. The vibrating wire concrete dams.
converts the fluid pressures into
equivalent frequency signals that are 6.2.4 One end of the vibrating
then recorded. The piezometer wire (figure 1) of the piezometer is
pressure range could be 0.034 to 34.5 connected to a sensitive stainless steel
MPa (5 to 5,000 lbf/in2) with an diaphragm, and the other end is
accuracy of 0.5 percent. The connected to an internal block. In
operating temperature range could be use, the changing water or other fluid
-29 to 65 C (-20 to 150 F). Also, pressure on the diaphragm causes the
vibrating wire piezometers can be diaphragm to deflect. The deflection
used to measure fluid temperature (if of the diaphragm changes the tension
thermistors are installed in the and thereby the frequency of vibration
housing) and existing negative pore of the vibrating wire element. The
pressures. square of the frequency of vibration is
directly proportional to the pressure
6.2.1 The filter (figure 1) could applied to the diaphragm.
be a standard 50-micrometer sintered
stainless steel or a high entry 2- A coil and magnet assembly, located
micrometer alumina. close to the vibrating wire (figure 1)
is used to both pluck the wire and
6.2.1 Vibrating wire sense the resultant vibration
piezometers are hermetically sealed at frequency. In use, a pulse of varying
the factory and their performance is frequency is applied to the coil and
not affected by use of long cable magnet assembly, causing the wire to
lengths. They have long-term vibrate at its resultant frequency.
stability in difficult environments. This resultant frequency in the form
of an induced signal is transmitted
6.2.3 Vibrating wire and recorded using the read-out box..
piezometers come in different models
such as standard, drive point, small 6.3 Electrical Cable.-A 4-
diameter, vibro/pneumatic, low conductor cable is a standard item for
pressure, pressure transducer, and a vibrating wire piezometer (figure 1)
auto resonant models. Two of the conductors of the cable
can be used with the built-in vibrating
Note 1.-Other types of piezometers, wire temperature gauge (thermistor)
such as hydraulic cell piezometers, to measure temperature of water or

4
USBR 6515-09

Figure 1. A vibrating wire piezometer.

any other liquid in the rock pores. gauge reading is displaced on a five-
The other two cables are used to read digit liquid crystal display. Both fluid
dthe frequency and thereby the pore temperature and pore pressure are
pressure. The cable has a thick PVC read by switching to proper controls
outer jacket. The conductors in the in the readout box.
cable are 6.4 millimeters (0.25 in) in
diameter. 6.5 Battery.-A sealed, gel-type 8-
volt rechargeable battery with a
Note 2.-Because vibrating wire battery life of 10 to 16 hours is used
piezometers convert pore pressure with the readout box. A battery
into an equivalent frequency, the charger requires 115 volts 10
resistance changes of the conductor percent at 50 to 60 hertz. A 230-volt
wires brought about by water or fluid battery charger is optional.
penetration, temperature variations,
and contact resistance do not affect 6.6 Biodegradable Drill Fluid.-
the frequency output of the signals. This fluid is composed of an organic
polymer that self destructs through
6.3.1 Direct Burial Cables.- enzyme breakdown with time,
Armored and gel-filled electrical reverting to water. This type of
cables that are used for direct burial drilling fluid provides the necessary
of long conductor cables for remote stabilization for the drilled hole and
sensing of piezometer readings. eases cutting removal during drilling.
Upon reverting to water, this fluid
6.4 Readout Box.-A portable leaves the borehole free of the organic
vibrating wire readout box (figure 2) polymer.
is battery operated, waterproof, and
suitable for cold weather operations. 6.7 Lightning Protection.-
Vibrating wire piezometers are read Protection can be built into the body
out by connecting them to the readout of the piezometer that protects the
box using the input plug or patch cord piezometer against low energy
provided with the readout box. The voltage spikes across the input loads
box is switched on and set to an to the plucking coil. Alternatively, , a
appropriate display setting. The metal pipe or metal rod can be driven

5
USBR 6515-09

into the ground for lightning for pore pressures to equilibrate.


protection at the surface, as close to
the piezometer as possible. 9.2 Enzyme Breakdown.-If
biodegradable drill fluid is used
7. Auxiliary Items during borehole drilling, allow
sufficient time to elapse for enzyme
Borehole drilling and viewing breakdown (consult manufacturer's
equipment biodegradable drill fluid data) before
Grout mixing and injection taking piezometer readings.
tools
Wrenches 10. Precautions
Electrical connectors
Couplings 10.1 Saturation.-Check that the
Casings piezometer filter stone is fully
Hand tools saturated. Because of surface
Bentonite tensions caused by improper
Wire mesh screen saturation, unsaturated filters can
Sand seriously affect the readings.
Cloth bags
10.2 Isolation.-Check proper
8. Calibration and isolation of the collection zone.
Standardization
10.3 Leakage.-Check that no
8.1 Manufacturers Settings.- electrical leakage occurs at splices
Calibrate the piezometer in and connections and that the cable is
accordance with the recommend- not severed or cut at any location.
ations obtained from the piezometer
manufacturer. Calibration data are 10.4 No Dirt.-Protect tubing and
supplied by the manufacturer. Also, cable ends from intrusion of dirt
the manufacturer supplies the zero during installation. Use dust caps on
reading of the piezometer at a specific all end fittings after installation.
temperature and barometric pressure.
10.5 No Couplings.-Minimize the
8.2 Zero Readings.-Zero readings use of tube couplings or avoid them
of the piezometer at the site should altogether if possible.
coincide with the manufacturer's
furnished zero readings within a few 10.6 Straight Lines.-Avoid
digits (not more than 20) after zigzagging of piezometer tubes if
barometric and temperature used, and introduce some slack in the
corrections are made. tubing.

9. Conditioning 10.7 No Snags.-Be sure to remove


all burrs and sharp corners that may
9.1 Seal Swell.-Before taking come in contact and damage the
piezometer readings, allow sufficient piezometer tubes or cables, if used.
time after the seal swell is complete

6
USBR 6515-09

10.8 Seals.-Check that the seals discontinuities that are permeable in


isolating the collection zone are nature, (2) procedure, timing, and
effective and are not punctured or sequence of construction activities
damaged. before, during, and after installing
piezometers, and (3) the nature of
other geotechnical instruments
installed or likely to be installed in the
vicinity of the piezometer.

11.2 Selection of the Piezometer


System.-Select the best piezometer
system consistent with the scope and
objectivity of the piezometer program
that has proper range and precision.
The selected piezometer should have
long-term reliability, excellent
stability, high sensitivity, good
repeatability, and robust, corrosion
resistant construction. Vibrating wire
piezometers have proven to provide
best long term service.

Figure 2. - Vibrating wire readout box. 11.3 Time of Installation.


To obtain a complete history of pore
pressure variations, install
10.9 Shock Resistance.-In blasting piezometers at the very start of the
zones, ensure use of shock resistant construction activity for the project.
piezometers.
11.4 Documentation of Field
11. Specifications Work.- Properly document the
piezometer data on a real-time basis
11.1 Location Selection.-After
thoroughly evaluating the geological 12. Procedure
data, select the location, orientation,
length, and number of piezometer 12.1 Drilling Boreholes:
boreholes, and numbers and positions 12.1.1 Use a drilling method
of piezometers in each borehole. A that is consistent with the existing
thorough evaluation of the geology and the availability of the
construction or maintenance type of drilling equipment. The
operation sequence and the Rotary method of drilling is
geotechnical features of the host rock recommended. Using percussion
is essential to plan an intelligent and drilling has to be evaluated with
effective piezometric monitoring caution due to issues that may result
program. Important factors during in a core hold with unsatisfactory
this evaluation are: (1) estimates of parameters, such as diameter or
directions and magnitudes of rock alignment.

7
USBR 6515-09

12.2.2 Check the zero reading


12.1.2 Check that the size of of the piezometer at the site by
the borehole is consistent with the connecting the piezometer to the
size of piezometer unit that is likely to conductor cable, the conductor cable
be used (use manufacturer's to the readout box, and the readout
recommended diameter for the box to the recharged battery. Allow
boreholes or select the borehole about 30 seconds for warming up
diameter to be 50 mm [2 in] larger before taking the zero reading.
than the external diameter of the
piezometer unit). 12.2.3 The zero reading at the
site should coincide with the factory
12.1.3 Verify the location and reading within a few digits after
orientation and drill the boreholes to correcting the reading for barometric
their designated depths. The depth of and temperature variations from the
each borehole is to be at least 300 mm factory conditions (supplied by the
(12 in) deeper than the tip of the manufacturer). In case of large
piezometer (tip of lowest piezometer differences in zero reading (more than
in the same borehole). 20), have the piezometer repaired.
Only use a properly calibrated
12.1.4 Inspect the drilling piezometer.
operations from start to completion of
the borehole. Keep notes on the 12.2.4 Remove the sintered
drilling rates, drill casing installations stainless steel filter (figure 1) from the
(avoid using casings if possible; use piezometer housing and place it in a
biodegradable mud instead of the bath of boiling water. Boil the filter
casing), detected geologic soft and until all the air is purged.
permeable zones, cavings, pluggings
of boreholes, or other drilling 12.2.5 Reassemble the
difficulties. piezometer while keeping it
submerged in a cool water tank. The
12.1.5 Log the cores recovered tank should be large enough to allow
from the boreholes. It is important to reassembly of the piezometer and
record the rock lithology, joint enclosure of the piezometer in a sand
orientation, degree of weathering, bag (section 12.2.6) while keeping the
location of water bearing seams, and piezometer completely submerged at
other pertinent information. all times.

12.1.6 Clean and wash the 12.2.6 Place a cloth bag in the
borehole to remove the drill cuttings. tank and partially fill the cloth bag
with well graded sand (grading to be
12.2 Preinstallation Procedure for recommended by the test engineer).
Vibrating Wire Piezometer: Place the reassembled piezometer in
12.2.1 Record carefully the the center of the bag and thoroughly
particular piezometers identification pack the sand around the piezometer.
number for future use. When full and compacted, tie the top
of the bag and wrap the entire bag

8
USBR 6515-09

with a wire mesh screen (size of mesh 12.3.6 If multiple piezometers


screen to be recommended by the test are installed in the same borehole,
engineer). The entire operation must repeat the steps in sections 12.2 and
be completed while keeping the the steps in sections 12.3.1 through
piezometer completely saturated at all 12.3.5 for each piezometer unit.
times until the piezometer is installed
in the borehole. 12.3.7 Following compaction
of upper bentonite seal, remove all
12.2.7 Using the readout unit, slack in the electrical conducting
obtain and record an initial reading on cable(s) and backfill the rest of the
the piezometer while it is still in the borehole with a sand-bentonite
water tank. This reading is the zero mixture.
or initial reading for the particular
piezometer. 12.3.8 If a casing was used,
then pull the casing as filling
12.3 Piezometer Installation in progresses or after backfilling is
Boreholes: complete.

12.3.1 In the cleaned and 12.3.9 If a biodegradable drill


washed borehole, place well graded fluid was used, then wait for complete
and saturated sand up to an elevation biodegradation before taking
150 mm (6 in) below the piezometer piezometric readings.
tip.
12.4 Monitoring Piezometer
12.3.2 Using suitable tamping Readings:
or compacting techniques, thoroughly
compact the saturated sand. 12.4.1 Determine the
piezometer readings at intervals as
12.3.3 Lower the cloth bagged recommended by the instrumentation
piezometer into the borehole to the engineer. Usually, more frequent
desired elevation and take a reading readings are required in the beginning
to ensure that the piezometer is of a monitoring program.
operating properly.
12.4.2 Determine the
12.3.4 If the piezometer is temperature of the piezometer.
operating correctly, backfill the
collection zone or influence zone with 12.4.3 Determine the pore
saturated, well graded sand up to pressure and its variations with time
about 150 mm (6 in) above the top of using the equations shown in section
the piezometer. Compact the sand 13.
properly.
Note 3.-Temperature and pore
12.3.5 Place a bentonite pellet pressure readings are taken by
seal about 300 mm (12 in) thick over switching channels to respective units
the compacted sand (see section on the read-out box.
12.3.4).

9
USBR 6515-09

12.5 Report: Write a report using 13.2 Temperature Correction:


guidelines stated in section 14. 13.2.1 Obtain the temperature
correction using equation 3:
13. Calculations
P(T1) = G(R0-R1) - M(T0-T1) (4)
13.1 Pore Pressure Calculation:
where:
13.1.1 The diaphragm C(T1) =temperature correction in pore
deflection, , is given by the pressure, MPa (lbf/in2)
following equation: M =temperature correction factor
supplied by the
N
= k 2 = (1) manufacturer (values of M
T2 depend upon the
temperature measurements
where: in C or F
k =constant, manufacturer supplied or T0 =initial temperature at the time
determined by calibration of pore pressure
=frequency, cycles/s measurement, C (F)
N =another constant, manufacturer
supplied or determined by 13.2.2 The corrected pore
calibration water pressure is given by equation 4:
T =period of vibration, s
C(T1) = - M (T0-T1) (3)
13.1.2 Because pore water
pressure is proportional to diaphragm
deflection, the relationship between
where:
the pore water pressure and the
reading at the read-out unit is
P(T1) = pore water pressure at
expressed by equation 2:
temperature T1, MPa (lbf/in2)
Other symbols were defined
P = G (R0 - R1) (2)
previously.
where:
13.3 Correction for Barometric
P= pore water pressure, MPa (in Pressure:
lbf/in2)
G=gage constant, adjusted depending 13.3.1 Because the zero
upon whether SI or reading of the hermetically sealed
inch-pound units are used piezometer was calibrated at the
Ro=initial zero reading at the read-out elevation of the manufacturer's
unit with zero pressure factory, the elevation difference at the
applied on the piezometer at site has to be calculated with respect
the time of installation to the elevation at the factory.
R1=reading under actual site
conditions of pore pressure

10
USBR 6515-09

Figure 3.- Calibration graph for a vibrating wire piezometer

11
USBR 6515-09

13.3.2 The barometric


pressure changes at a rate of about 14.3.2 Log of drilling,
0.0034 MPa (0.5 lbf/in2) per 305 including description of important
meters (1000 ft) of elevation geological features and difficulties in
difference. drilling, if encountered.

13.3.3 The corrected pore 14.3.3 Borehole coordination


pressure, allowing for the barometric and collar elevation.
pressure change, is obtained by
adding (for higher elevations than the 14.3.4 Elevations of installed
factory) or subtracting (for lower piezometers.
elevations than the factory) the
change in barometric pressure (MPa 14.4 Monitoring.-The monitoring
[lbf/in2]) with respect to the factory report shall include the following:
conditions, to the calculated pore
pressure P(T1). 14.4.1 A set of tabulated field
data (see table 1).
Note 4.-Vented piezometers
designed to eliminate barometric 14.4.2 A plot of variations of
effects are available but not used by pore water pressure (uplift pressure)
Reclamation. For vented with time (figure 4).
piezometers, barometric pressure
corrections (section 13.3) are not 14.4.3 An interpretation of the
required because the space inside the piezometric data.
piezometer is no longer isolated and
evacuated but is connected via a tube 14.5 Results. The laboratory test
to the atmosphere. But with time, results shall be reported in accordance
water may enter the tube and destroy with Reclamation designations USBR
the connection to the atmosphere, 3000 and USBR 9300.
making the piezometer readings
erroneous.

s14. Report 15. Precision and Bias

14.1 The report shall include: The precision and bias for this
designation have not been
14.2 Background.-Project name, determined. Any variation observed
project location, scope and purpose of in the data is just as likely to be
the piezometer program. caused by specimen variation as by
operator or laboratory testing
14.3 Installation.-The piezometer variations. Because of the variability
installation report shall include the of rock, this test procedure has no
following: reference value.

14.3.1 Types and details of


drilling equipment used.

12
Designation USBR 6515

Table 1. - Tabulated field data.

VIBRATING-WIRE PIEZOMETERS

Dam _ __ _ _ _ _ _ _ _ _ _ _ _ _ Observer(s) _ _ _ _ _ _ _ _ _ _
Reservoir Elevation _ _ _ _ _ _ _ Date _ _ _ _ _ _ _ _ _ _ _ _ _
Tailwater Elevation _ _ _ _ _ _ _ Sheet _ _ _ _ _ _ of _ _ _ _ _
Borehole number _ _ _ _ _ _ _

Temperature
Correction
Resistance Piez. EL
Initial Current Piez. Piez. Temperature After
Piez. Serial Gage Reading Reading Head Tip EL Piez. EL ( x 103) Correction
No. No. Factor (Hz) (Hz) m (ft) m (ft) m (ft) C (F) (ft)

G Ro R1 H* A H+A

H* = 2.308 G(Ro - R1)


Designation USBR 6515

Figure 4. - Data plots on vibrating wire piezometers.

You might also like