1. A Brief Introduction to HR-150D Motorized Rockwell Hardness Tester
1.1 An automatic testing pressure loading / unloading mechanism, as well as a
potentiometer for total testing pressure holding time adjustment, are equipped on
THR-150M Motorized Rockwell Hardness Tester, and the testing pressure change can
be obtained by turning of the load-selection handwheel, making its operation easy and
simple. There is no human error except zero setting of the Panel .The tester enjoys
high sensitivity and stability.It has stable function,widespread use,good economy and
practicality thus is designed for use in workshops and laboratories.
1.2 Scope of application: Quenched steel, tempered steel, aluminum alloy, malleable cast
iron, and annealed steel and so on.
1.3Operative norm : GB/T230.1、 GB/T230.2、 JJG112
(Table 1):
Scale Hardness of Standard Blocks Display Tolerance
(20~75)HRA ±2HRA
HRA
(>75~88)HRA ±1.5HRA
(20~45)HRB ±4HRB
HRB (>45~80)HRB ±3HRB
(>80~100)HRB ±2HRB
HRC (20~70)HRC ±1.5HRC
(40~70)HRD ±2HRD
HRD
(>70~77)HRD ±1.5HRD
(70~90)HRE ±2.5HRE
HRE
(>90~100)HRE ±2HRE
(60~90)HRF ±3HRF
HRF
(>90~100)HRF ±2HRF
(30~50)HRG ±6HRG
HRG
(>50~75)HRG ±4.5HRG
2
(>75~94)HRG ±3HRG
HRH (80~100)HRH ±2HRH
(40~60)HRK ±4HRK
(>80~100)HRK ±2HRK
(Table1)
The most commonly used scales for Rockwell hardness test are A, B and C.
4.1.2 The relative humidity in test room shall not be over 65%;
3
4.2 Tester unpacking
4.2.1 Remove nails, open the packing box cover and take out all cushion materials.
4.2.3 Lift the bottom plate; unscrew the two M10 bolts under the bottom plate with a
spanner, to separate the hardness tester from bottom plate, take out the tester. All
4.2.4 After unpacking, the tester shall be placed on a stable bench with levelness
the bench (see figure 1) to enable the Lifting Screw (8) to operate properly.
(Figure 1)
4.3 After the hardness tester is properly placed (see figure 2), remove the top cover (1)
and the rear cover (2). Untie the fastening white gauze tape on the extension rod
(Pag11, figure 4), remove the white gauze tape on moveable parts. Then cover the top
4
+ - START
C
90 10
B
20 40
80 20
10 50
0 60
70 30
90 70
80
60 40
50
(Figure 2)
1. Top Cover 2. Rear Cover 3.dial plate 4. Panel 5. Indenter Locking Screw
6. Indenter 7. Test Stock 8. Lifting Screw Cover 9. Turn Wheel
10. The pressure selecting handwheel 11. Power Switch
Take the weights out of the accessory box and rub them clean. Turn the pressure
selecting handwheel (10) to position 588, take out the hanger (14) from rear cover (2),
and lead the hanger into the hole on weight A (18), then tighten the two M10 nuts (19) at
the end of hanger, hook the hanger onto the hoist eye (13) at the rear of beam (12), and
put weight B (17) and weight C (16) on two resting forks (15) respectively. Then turn
the pressure selecting handwheel a whole cycle and ensure that the weights’ pins shall
fall into the groove on the resting forks. When weight is hang in the air, it should not touch
5
12. Beam
13. Hanger
14. Resting Fork
15. Weight C
16. Weight B
17. Weight A
18. Nuts
19.Tray
(Figure 3)
(table 3)
Value of pressure
Scale Test Pressure selecting Applied Weights
handwheel
5.1.1 Surface of the specimen to be tested must be smooth and clean, free from dirt,
surface and testing bench of the specimen should be clean and ensure good tight.
5.1.2 Minimum thickness of the specimen shall be larger than 10 times the depth of
6
hardness test. The relationship between minimum thickness of the specimen and
Table 4
Hardness (HR)
5.1.3 The specimen to be tested shall be placed in stable state on top of the test stock. It
shall be fixed during the test process. Ensure that the test pressure is
5.1.4 In case of a cylindrical shape specimen to be tested, the V type test stock shall be
used.
5.1.5 When testing the value of HRC, HRA, diameter of the specimen is less than 38mm;
When testing the value of HRB, diameter of the specimen is less than 25mm.The
result should be revised, and the revised value shall be a positive value.(table5)
5.2.1 Plug the power cord (11) and the digital tube of the Panel (4) is on.
5.2.2 Rotate the dial plate (3) l to set the long pointer to the up position and the small
7
pointer to the black spot.
5.2.3 Choose a scale according to table 2 based on the extent of soft and hard of the
material to be tested. Turn the pressure selecting handwheel (10) clockwise to set
5.2.4 Push the indenter(6)into the hole on indenter rod till it rests neatly on the supporting
surface, then make the plane of the indenter handle gap face the indenter locking
screw (5), slightly tighten the indenter locking screw (5) and place the sample on test
stock (7).
5.2.5 Rotate the panel (3)to the set position “C” to the above position.
5.2.6 Rotate the turn wheel (9) clockwise to let the lifting screw move upward, ensuring
that the specimen to be tested slowly contact with the indenter with no impact, till the
short pointer on the panel turn from the black point to the red one, and by this time
the long pointer shall have passed three turns and point to position “C”. The
deviation of long pointer shall not exceed 5 grades, otherwise the pointer shall not
be allowed to turn back, instead the test spot shall be changed to restart the
operation.
5.2.7 Rotate the panel (3) to the set the long pointer on position “C” or position “B” in case
of HRB determination
5.2.8 Press the start button, motor starts to exert test pressure automatically. When the
dwell time is up, the motor begins to operate and remove test pressure
automatically.
.5.2.9The value pointed by the long pointer on panel at this time shall be taken as the
determined hardness value of the sample (in case of HRB determination, value shall
5.2.11 Rotate the lifting wheel counterclockwise to lower the testing table. Repeat above
5.2.12 No less than 5 spots shall be tested for each specimen (first spot will not be
inspection.
Table 5
8
Hardness Diameter of Cylindrical Specimen (mm)
Value
6 10 13 16 19 22 25 32 38
(HRC)
(HRA) Offsets (HR) to Rockwell Scale C,A,D
20 2.5 2.0 1.5 1.5 1.0 1.0
25 3.0 2.5 2.0 1.5 1.0 1.0 1.0
30 2.5 2.0 1.5 1.5 1.0 1.0 0.5
35 3.0 2.0 1.5 1.5 1.0 1.0 0.5 0.5
40 2.5 2.0 1.5 1.0 1.0 1.0 0.5 0.5
9
The accuracy of the tester is calibrated at factory. If any error is caused by improper
handling during transportation, operator can make adequate adjustment but in prior to
do that he shall thoroughly understand the structure and working principle of the tester
apparatus. The procedure is as following: Remove the top cover (1). If displayed value
is lower than the actual hardness of standard block, loosen two boults and push
adjustment sheet forward. If adjustment sheet goes forward for 1mm, the hardness
value will improve 2 HR, and then screw boult. Test the hardness value until it is
adjusted in the required error range (table1) .If the tested value is higher than it of the
standard block, you should move the adjustment sheet in the opposite direction. If
after adjusting the hardness value, the long pointer deviates a little bit, then you
should loosen the nut, slightly turn the screw to make the long pointer point to the up
position. But the small pointer must point to the black spot, and then tighten the nut.
Nut
Boult Adjustent Sheet
Dial
Figure (4)
manual. Frequent calibration against standard blocks is necessary before and after
actual tests. For tester not frequently used, several pre-tests shall be done on the
standard blocks after the tester is turned on. Test of the specimen should begin
6.2 During hardness test, it is strictly prohibited to turn the pressure selecting handwheel
6.3 Use of the standard blocks can only be done on their working surface. A minimum
10
distance of 3 mm between two successive test spots or between the center of indent and
edge of the specimen is necessary. Life period of a standard block is limited to 2 years.
6.4 When handling the hardness tester, the extension rod shall be fixed, and the weights
and hanger shall be removed after pull out the power plug.
6.5 The hardness tester shall be kept in clean state. After completion of testing procedure,
the tester shall be covered with a dust cover. Hardness block and indenter shall be
coated with antirust oil after use to prevent rust and corrosion.
6.6 Periodical verification shall be done, at least once a year, on the tester so as to ensure
its accuracy.
It is recommended that the user contact the manufacturer if problems occur with the tester.
However, the following table may help the user to detect and solve some common problems
(table 6)
Table 6
Problem Possible cause Suggested remedy
Indicator lamp do not 1. No power 1. Check if the power cord is
light up when tester is 2. Fuse blows out properly connected.
turned on 2. Replace the fuse with the
new one in accessory box.
Refer to figure 4. Slightly
deviates from the up Adjustment of displayed value in small increment to align the
position. The small may cause deviation of the long pointer onto the up
pointer deviates from the long pointer. position and the small pointer
1. Damaged indenter.
1. Exchange diamond or ball
11
2. Wrong order of weights indenter.
hardness values and inner wall of the tester 3. Use leveling instrument to
Lifting Screw (8) got Fit clearance of Lifting Screw Take off Lifting Screw Cover
stuck (8) is small; the end of a (8), clean the screw by clean
thread or dirty things may cloth, and then hold the Turn
make it get stuck. Wheel (9), pull the screw up
and down. (Don’t use sand
paper to rub the screw)
12
7. Accessories
Item No Description & Specification Quantity
6 HRB 1
8 Fuse (0.5A) 1
9 Power Cord 1
10 Weights A, B, C 3 in total
13 Instruction Manual 1