Professional Documents
Culture Documents
Cardiac Pacemakers
Cardiac Pacemakers
All rights reserved. Nopart of this bookmaybe reproducedin any form, nor mayit
be stored in a retrieval systemor transmittedin any form, withoutwritten
permissionfrom the publisher.
ISBN0-7803-1134-5
IEEE Order Number: PC5597
Libraryof Congress
Cataloging-in-Publication
Data
Edited by
John G. Webster
University of Wisconsin, Madis.on
~, IEEE
PRESS
TECHNICAL
ACTIVITIES
Editorial Board
J. B. Anderson, Editor in Chief
3.1 Automaticity 35
3.2 Triggered rhythms 41
3.3 Slowed conduction and block 47
3.4 Reentry and unidirectional block 52
3.5 Simultaneous impulse generation and conduction abnormalities 61
3.6 References 62
3.7 Instructional objectives 63
4 Diagnosing Arrhythmias
Valtino X. Afonso
Battery 161
John G. Webster
Glossary 471
Index 476
Preface
techniques for amplifiers, filters, and comparators. It also discusses techniques for
minimizingelectromagnetic interference.
Artificial pacemakers contain microprocessors. Chapter 9 describes the
rhythm required of the artificial pacemakerand details the functions of all the
timers that are required to either wait for durations to expire or to reset whenthe
heart provides intrinsic waveforms. ChaPter 10 shows howthese sensed waveforms
control the timers and provide safety features whenmalfunction occurs.
Chapter 11 shows how the cardiac muscle responds to output pulses of
varying durations and amplitudes~ This~, chapter also details the circuitry that
provides this output pulse. Physicians cani~se telemetry to changeparameters in an
implanted pacemaker. Chapter 12 describes howthese telemetry systems operate
and the safety codes that prevent inadvertent modification of stored parameters.
These programmingdeVices also extract stored data such as histograms of beat
intervals.
Early pacemakers operated at a fixed rate. The next four chapters address
methods that permit rates to vary. Chapter 13 outlines several methods for
designing rate-~esp0nsiv e pacemakersthat increase the heart ~ate in response to
exercise. The next tb_/ee chapters illustrate howthe rate c~ be increased through
the use of sensors: in Chapter 14 with piezoelectric sensors to sen§e body naotion;
in Chapter 15 with a thermistor to sense increase in blood temperature; in Chapter
16 to sense events in the electrogram or to use intracardiac impedance,in Chapter
17 to sense th6racic impedanceand calculate minute ventilation.
The heart maylapse into the undesii~ably high rateof tachycardia: Chapter 18
shows howcarefully timed pacings from the antitachy pacemakercan interrupt the
tachycardia and restore normalrhythm. If the ~heart rliyttim should proceed tO
ventricular fibrillation, there is a wayto restore it to normal.Chapter19 showsthat
the implantable cardioverter-defibrillator can provide a Stepped response of shocks
to restore normal rhythm.
Testing cardiac pacemakers is important. Chapter 20 shows the-detailed
design of a simple pacemakertester. The tester can v~r), its rate to electronically
exercise the response .of the pacemaker. The boisk appendix provides the design
and software of a microprocessor-based pacemaker. A glossary provides
definitions of manywordsin both Cardiology and electronics:
I would welcomeyour suggestions for improvementof future editions.
John G. Webster
Madison, Wis~ot~sin
NormalConduction System of the Heart
James R.. Bowers
K~" 4 mM
0 mV
Figure 1.1: The modelcell membrane is only permeable to potassium ions. The concentration
gradient tends to drive potassiumout of the cell (shaded arrow) and the potential difference
drives potassiuminto th~ cell. At equilibrium the two forces are balancedahd no net potassium
flows. The potassiumconcentration is expressedin millimoles (mM):
DESIGN OF CARDIAC PACEMAKERS
The answer is that the ions are not capable of leaving the cell due to the
development of a charge separation across the membrane.The charge separation
creates a membranepotential, which hinders further diffusion. As the ions leave
the cell, positive charges accumulate on the outer surface of the membraneand an
excess of negative charge remains on the inner surface. Whenthe membrane
potential becomes sufficiently large; further net movementout of the cell is
stopped. The concentration gradient equals the electrical gradient at this point and
electrochemical equilibrium has been established. Ions still pass back and forth
across the membrane,however, there is no net movement~of potassium.
The question then arises as to howlarge the potential must be to cancel the
chemical gradient. The potential necessary to achieve electrochemical equilibrium
is defined as the potassium equilibrium potential. The potential depends on the
difference between the logarithms of the extracellular concentration [Ko+] and
intracellular concentration [Ki+]:
This is the Nernst equation for potassium wh~re R is the thermodynamic gas
constant, Tthe absolute temperatt~e, ~ the valence of the ion (in this case +1), and
F the Faraday constant. At mammalianbody temperature"(37*C) the expression
RT/zF is approximately 27 mV.Figure 1.2 shows typical ion equilibrium potentials
for cardiac muscle cells. The resting potential 0f cardiac muscle cells is about
-90 mV.
Note that the electrostatic and electrochemical forces tend to pull sodiumions
into the cell. The actual inward flux is small, however, since the sodium
permeability is very small at the resting potential. This is also true for calcium.
Given normal potassium concentrations found in cardiac cells,~ Fxt. (1.1) gives
goodapproximationof the resting potential. Tiffs indicates that the resting cardiac
cell is mostly permeable to potassium. The resting potential..can be measured
experimentally by using a glass microelectrode with.a tip diameter.of 0.1 #m to
puncture ~the membraneof4he cell without greatly damaging it~ The membrane
then seals around the glass electrode and~a voltage measurement can be made
relative to the external medium.
+ a[Na
] + l[Cl ]
Era=
-- ln ÷ ÷
(1.2)
Equation (1.2) differs from the Nernst equation in that it includes the ionic
permeabilities in addition t6 concentrations. If is knowna~ the ConStant field
equation because it is derived assumingthat theelectric field within the membrane
is.~,gd: ~ ,F_zluation 1.2 is flso knownas the Goldman-Hodgkin-Katz(GHK)
p¢~i~i~i!iy
eqU~0fi ~t~r’ coefficient for ion
its, developers: emx.represents
The iarger._the permeability,,
the membranepotential the
andgreate~ it~
Px istile
Ix = gx (Vm-Ex)
(1.3)
Intracellular
Extracellular --"
Embedded "
Figure 1.5 showsthe basic structure of a channel protein. The protein forms a
pore, through whichions can move.A region of the pore acts as a selectivity filter,
regulating ion permeation according to size and molecular structure. The channel
gate:allows or prevents ionsTm~m crossing the membrane.,Theopen or closed state
of the :gate depends on the magnitude of the voltage across the membrane.This
control mechanismis knownas voltage-activation.
NORMAL CONDUCTION SYSTEM OF THE HEART
~F~.~)
,,,,nslsts of a lectivi
se .. ty filter (S) and_a gate ((3). Thoylterdrestricts ~o~a_.~ormnt~
+ .... ~t~ A)lowsions .to ~ass through the memoranean canbe regulated by membran
voltage.
~ .Efficient operation of the heart depends on the ability of fibers to work
synchronously~ Figure 1.6 shows the connection of mdivxdual umnucleate cells to
forn~_~_.a branched network of muscle fibers. The connection is formedat the cell
endsby intercalated disks. These disks consist of accumulations of dense material
on the insides of the two cell membranesto fix the cells iogether and allow the
filaments of the contractile apparatus in one cell to pull on those of the next cell in
the line. The flow of electrical signal from one cell to another.is facilitated by the
presence of gap junctions between the cells. The junctions allow a low,resistance
path for the current to flow and resemble the membranepores discussed earlier.
Moregap junctions are present wherethe fibers are in contact longitudinally than
side to’side. Electrical conductiontherefore proceedsfaster parallel to the long axes
of the fibers than in the direction perpendicularto the fibers. Thefiber i~ striated in
appearance due to bands of thick myosin and thin actin filaments aligned
tz~nsverseto the long direction.
Intercalated Gap
disks Fiber bands junction "
Nucleus
/(enlarged)
%~___ / ~. ~//
E -60-- ~-
-90-- ]--- T T
0 0.1 0.2 0.3 0 0.1 0.2 0.3
......... Time,(ms)
Figure 1.7 shows that the resting potential is more :: negative in the fas!
response cells and the upstroke is significantly steeper. ~A!s0, the,ampfi~deand the
overshoot is greater in the fast response: Furthermor~e,~the~amp~md-e of:tlie’action
potential and the,slope of th~:upstroke are ~po~anf factors dete~ining
conduction velocity downa muscle fiber. Therefore, the conduction velocity in
tissue characterized by the slow response is slower th~ in tissue exhibiting fas!
response ~tion p~tentials. This is"~portant since stoP’velocity often!~ads t~
conduction block.
NORMAL CONDUCTION SYSTEM OF THE HEART 7
_~, ~ Similar to the ionic basis of the resting potential~ there is an ionic basis to the
action potential. ~Figure1.8 showsa typical fast response action potential. Therapid
upstroke of the action potential is designated, phase 0. Immediately after the
upstroke, there is a brief period of partial repolarization (phase I), followed by
plateau region (phase II). The potential then becomesmore negative (phase III)
imtil~the resting state is attained. Theinterval fromthe completionof repolarization
until, the beginningof the next action potential is phase IV.
II
0 III
Time (ms)
1
’ ~’~gNa
I
I
0
I
..~
"0
|_j " .
0
Time(ms)
I Strong ’
c¯
.-
x~
, timu
us’..t,..
! , . Normal
’ imulus
~
-90- L
l~(.., / .
---’ 7
f~ Reponariza~:ladzed Resting ~
The connection between cardiac action potentials and muscle contraction depends
on ,internal calcium concentrations activating .filaments within the cell. Duringthe
plateau region of the action potential calcium is flowing into the eell.Thi~ causes
in~rnal stores of calcium to be released into tbe cell cytoplasm., The:con~ntration
of internal calcium increases fifty times its normal restinglevel~ This increase in
calcium concentration results in myosinheads attaching to the thin actin filaments
at a particular angle. This cross-bridge undergoes a structural change where the
10 ¯ ~DESIGN OF CARDIAC PACEMAKERS
heads:tilt from 90~to45*, drawing the thin filament along with them.-This
produces an overall shortening of the muscle fiber. Active ion pumpsthen move
the calcium out of the intracellular space and the muscle relaxes. Both contraction
and relaxation phases require ATPhydrolysis. Figure 1.11 shows this actin and
myosin contraction mechanism.
ldyosin
¯ = ~ % . / Myosin head
Figure1.11 Myosinheadsbondto the actin fdamentand the h~dstilt from90" to45°, drawing
the thin actin filamentalong.Thisresults in a museularcontraction.
The heart consists of two pumpsin series, one to propel Moodthrough the lungs
for exchange of,oxygen and carbon dioxide (pulmonary. circulation) and the other
to, prope!bloodto al!:other tissues of the body(systemic circulation). Figure 1", t
sliows.th~ basic st~eture of the heart and the ~ direction of blood"flow2 The
ventricles provide the pumpingof the blood and the contractile phase is Called
systole. The resting phase of the heart is called diastole. The atria function as
fillingand holding Chamberswhile the ventricles, contract.and,also prime the
ventricles.
The pericardium ~ a sac-that encloses the entire hea~ and the cardiac portion
of ~e blood vessels~ ...This sac contains a small amountof fluid which .provides
lubtication for continuoOsmovementof.the heart, The material.,of the pericardium
is mechanicallyStiff so.that it prevents large and-rapidincreases in cardiac..size.
This helps to prevent overdistension of:the chambers:.
The. cardiac valves eonsist.-~f thin flaps of. flexible,, tough fibrous gssue,
Movement of the valve flaps is.essentially passives.and the~otient~tion of.the valves
is responsible for the unidirectional flow of blood in the heart. The tricuspid -~alve
is composedof three cusps and lies between the tight atrium and tight ventricle.
The mitral valve has two cusps and separates the left atrium and left ventricle.
Strong fitaments (da0rdae tendin~ae)~e~attacltedi to, thor free .ends of the val~es and
prevent the valves from opening when the ventricles contract. The pulmonary
valve and aortic valve are semilunar valves and consist of three cup-like cusps
attached ~to the ’valve:tings.-These,’valves provide unidirectional flow ,out, -of ,the
ventricles. Revergal:0f blood~ flow ;towardthe-ventricles, causes ~the: euspsto snap
flow intothe venlrieles. .~~ ,-~:
~Tlae (1) eells~that
initiate ~ and eohdtiet impulsesl arid (2) :eells that\ibesideseonducting, respond
stimuli~b3/e~eting. The latter ConstitUte the Worldngmuseulatureof the hearVor
ttie my6cardium.The myocardiumof the vemtieles is aftincti0nal syncytium, that
NORMAL, CONDUCTION SYSTEM OF :’I~IEHEART 11
is, ,the cells are not electrically insulated Or mechanically separated from one
anoti~r~A stimulus, arising at any point in the ventricle spreads to cause complete
contraction of both ventricular chambers. The same applies to the atria.-The atria
and ventricles are.not connected except for the AVnode.
To head’
and
From
head
and
Right
lung.
Left
lung
Right
coronary Right
artery Atrium
To trunk
andlegs
artery
Fromtrunk
andlegs (1) Tricuspidvalve
(2) Pulmonaryvalve
(3) Mitral valve
(4) Aortic valve
Fi~ ~1~j2S~hematic the direction
showing ofblood
flow the heart:
through
Excitation of the heart normally proceeds: from the sin.oatrial ~($A) node which.is
the physiological pacemaker of the heart. From the SAnode, excitation spreads
12 , i, DESIGN OF CARDIAC PACE~KE~
through both atria to the atrioventricdar (AV)node,whence,, via the bundle o~ His
and its two brancbes, it reaches the Purkinje network, w~ehcarries the impulse to
the ventricular muscle: Activation of the ventricular muscle takes place from
endocardium (inside of heart)to epicardium (outside of the heart) and from
apexof the ventricles to the base.
II
0
Threshold potential
waveforms show an ad&fional wave ..aft~ er the T wave. This is called the U wave
and its origin is attributed to Slow repolarization of ventricular papillary muscles.
1.8 REFERENCES
Berne, R. M., and Levy, M. N. 1988. Physiology. 2t~d Ed. St. Louis: C.V. Mosby.
Dangman,K. H., and Miura, D. S. (eds.) 1991. EleCtrophysiology and pharmacologyof the
hea~. NewYork: Maw,el Dekker.~1991.
Despbl~odl0~;~A:; and Silb~rha~l~S. C61oratlas ofphysioi~gy. 4th Ed~: NewYork: Thieme
MedicalPublishers.
Keynes, R. D., and Aidley, D.L 1991. Nerve and muscle. 2rid Ed. Cambridge:: Cambridge
University Press.
Nieholls, J. G., Martin A. R., and Wallace, B. G. 1992. Fromneuron to brain. 3rd Ed.
Sunderland, MA:Sinaner Associates.
Webster~,~I. G, (ed.) 1992. Medicalinstrumentation: application and design, 2rid Ed. Boston:
~H0ugliton~Mifflin.
aftedoad, and contraction frequency. Preload refers to the initial tension prior to
a contraction of the myocardial fibers. The preload mechanism is volume
~
dependent and is the basis for the understanding of the Frank-Starling effect~
Therefore, afterload refers to the tension exerted on the myocardial fibers
immediately after contraction. This is measured by ventricular elasticityma
parameter that is difficult to quantify as a result of complexgeometry. The last of
the inducers is contraction frequency, otherwise knownas heart rate. This is a
well knownand measurable parameter. It plays a key indicating role for the
condition of the heart and is easy to precisely measure, providing information on
the heart’s condition.
Intrinsic adaptations
I Extrinsicadaptations
I
I
’madl
I c mic,*fa=oI ,+a
I Neural
ors
!
~ ~r----], Hormonal II r-.I SYmpatfieticl
I Parasympathetic
~Card;ac performance ~
I ~’as~mP=het~l
I Syn-e=h=~c
activityI I Syrnp~c
activit~
I activitv~tohearlI I t~ h~J~rt I I to arterioles |
02
I
I ,~pirat~ movement~
Osmoladty
-Eicanoids~
Bmdykinin
Myogenicresponse
Substancesrelease
dudnginjury
~ , :
8 I End-diastofic
~olurne : -- - ? "~""
I,I I :~ *.e,o,.,co,=~on
I
II ~! I I Anaiotensinll I I , I acuityI I I 1~o.~) I
t ili,,’ ..... ’
~_.
I ./~,~,~,vo~,~l
, .1 ’ .I
.I .,~r,r~to " ,
~1 "P’ . -J
~/ I c.~c
o.t~ut
I
Meani
cardiovascular system, These factors interact with each other to maintain bulk
flow conditions throughout the body. At any given time, at least 5% of_the total
circulating blood travels through the capillaries. This is actually the goal and
most important aspect of the system. For it is this 5%that provides the essential
nutrients and removal of metabolic end products at the cellular level, keeping the
individual alive.
In Figure 2.2, the top third represents "the neural input factors, the middle
third represents the chemical and physical influences, and the bottom third of the
diagram characterizes cardiac performance. A comprehensive .systems flow
18 DESIGN OF~ CARDIAC PACEMAKERS
200
100
Ventricularend-dib.stolic volume
(ml)
Figure 2.3 The Frank-Starling law of the heart. As the amount of exercise or stress on the
cardiovascular system inere ~a~es, movementalong the curve to the right occurs. (From Rhoades,
R., and Pflanzer, R. 1992. Humanphysiology. Saunders.)
The solid line arrows indicate directly proportional affects upon the
succeeding factor. Thus, an increase in sympathetic activity to the heart resulting
from a decrease in arterial baroreceptor firing, causes an increase in heart rate.
Likewise, the dashed line arrows indicate inversely proportional affects upon
succeeding factors. An increase in parasympathetic activity to the heart causes a
decrease in heart rate. Onecan easily verify the regulatory effects by following
any of the pathways. For example, hemorrhagingor other disturbances that cause
a decreas~ in the meanarterial blood pressure (MABP)would cause a directly
proportional decrease in the firing by the arterial baroreceptors. This sudden
alteration within the body would then have multiplicative effects. They would
either be excitatory or inhibitory. Hence, the decrease in arterial baroreceptor
firing would cause increases in the sympathetic activity to the veins, heart, and
arterioles and a decrease in the parasympathetic activity to the heart. These
changes, after causing further alterations of their respective succeeding
parameters, would ultimately create an increase in cardiac output and total
peripheral resistance. These alterations reestablish the biological norm for the
MABP which is approximately 90 mmHg2 Any other effects causing disturbances
in the model result in. integrative cooperation on all levels, restoring normal
conditions. However,often certain parts of the body are either degenerated or
dysfunctional, making it extremely’-difficult for this natural-autoregulatory
process to succeed. In these cases, the engineer must provide solutions.
Figure 2.2.~ by no meansattempts to represent all’influential factors of the
heart. Other influences may exist that have not been.-determined to date. Any
significant factors that exist are compiledin Figure 2:2. Significant factors are
CONTROL SYSTEMS, OF THE HEART 19
The i~nnervation of all tissues other than skeletal muscle is through the ANS.~The
ANScoordinates bodily functions needed for, survival~and acts tO regulate the
e~nvir00me~nt necessary for cells to function properly, Although inherent
rbyth.~lmcity of.the heart is due to its natural pacemaker, the sinoa~ial node,
cgn.tinuouscontrol of heart rate is dependent on the relative balance between
sympathetic, parasympathetic, hormonal, and chemicalimpulses delivered frpm
the brain to the sinus node.
2.2.1 Physiology
Domalroot
CentralANS PeripheralANS
Sensorynervefrom
visceralorgan
Target
organ
Ventralroot
ganglion
nerve
cell Postganglionic
nerve
cell
~ ganglion
Autonoticm
The function of the central division of the ANSis less integrative than the
peripheral division. ActiVity of the hypothalaiiaus is more~direcfly ,related .to
control of the adrenal gland. Hence, the discussion of its regulatory process is
included in the section on humoral control. The organization in the brain stem is
actually not well understood’ withrespect to eardiac Control..The, processes
occu~ng in the region-are extremely complex and laci~ ~ precise definidbn.
However,information regarding the spinal cord .struCture is available iti~detail:
The preganglionic nerve cells are locatext in the interomedial lateral cell ~61~n~
They have sensory fibers that enter-the spinal cord from the d0rsalroots e~g
information from the target, organ, Whilefibers-leave fr0m the Ventral roots. ~The
cell bodies of the sensory fibers are located in the dorsal ’root ganglia. Both the
sympathetic and parasympathetic subdivisions exhibit this organization.
The peripheral division exhibits distinct and intricate modesof operation:
This is essential due to the unpredictable actions of the cardiac system. The two
branches of the ANSoriginate from different’level~ 0f ~he’ spinal cord. The
sympathetic nerves that innervate the heart leave the th0raciJ region of the spinal
cord between’T1 & T4 (Figure 2.5). They te~itate a~ ~esinus bode, conduction
system, atria, ventricles, ~ and coronary: v~els. All ~bl00dvessels recei#e
s~mpathetic fibers. In addition; Organ~yst~msare activated ~Simultaneously,,
piaying"an important role during stress ~r~spbnses: ~ In compariso~ tli~
and
parasympathetic nerve fibers leave the s~infl cord from the br~in stem
terminate at the sinoatrial and atrioventficul~ nodeS, atrial and ven~cul~
CONTROL SYSTEMS OF THE HEART 21
musculature and coronary vessels. Not all vessels receive input from this branch.
It alsoaetivates ~fferent organ systems more independently.
. The two subdivisions of the autonomic nervous system also differ in location
of ganglia. Most .of the sympathetic ganglia lie close to the spinal cord and form
two chains,, one rOn each side. They are knownas the sympathetic trunks. The
celiac,-superior mesenteric, and inferior mesenteric ganglia make up the
collateral ganglia, They lie closer to the innervated organs, farther awayfrom the
spinal~.~¢0rd (Figure 2.5). In contrast, the preganglionic fibers of the
parasympathetic subdivision have long myelinated and unmyelinated axons. As a
result, they innervate postganglionic nerve cells grouped near or in the target
organ. This provides for very short axons of the postganglionic parasympathetic
nerve cells. These ganglia can act either as automatic relay stations wherebythey
do not change the information they transmit to the target organ, or as important
integrating centers capable of generating individualized responses. Figure 2.5
demonstrates the origin and pathway of the autonomic nerves. The specific
attachment of neural networks to-the heart is complexand still being researched.
This specificity, once determined, will enhance opportunity for cardiac control.
Dorsal motor
nucleusof vagus
~ Superiorcervical /~.
~idbrain~ ganglia I
I1 g~lnfed°r--.~--~-~
~cervical
=’~.I~’ "¯
’ : ’ I I I/
~ Spinal cord ’,
,
,
~ ~ = Poslganglionicnerves
The ’physiological differences are more complexin detail (Figure 2:6). This
difference lies in the types of neurotransmitters and receptors available for
synaptic activity in each of the ANSbranches. The two major neurotransmitters
in the ANSare acetylcholine and norepinephrine (NE). Cholinergic and
adrenergic fibers are those that transmit acetylcholine and NE respectively:
Acetylcholine is released at the synaptic terminals ofall preganglionic neurons. In
postganglionic nerve cells, NEis found in the terminals of sympathetic fibers,
while acetylcholine is found in the terminals of the parasympathetic fibers. ’The
sympathetic branch sometimes contains acetylcholine in the terminals of
postganglionic fibers that innervate the skeletal muscleblood vessels. ~¯
Muscadnic
Pmasympathetic
¯ recept°r
I
"
Preganglionic Postganglionic
Nicotinic neurons ce~s
neuroll=$ Ganglia ~
Sympathetic
I):-recptor
Manyof the drugs that stimulate or inhibit various componentsof the ANS
affect receptors for acetylcholine and NE. The acetylcholine receptors on the cell
bodyof postganglionic neuronsare different from the -recept~ors on target cells~,of
the p~sympathetic branch,. Cholinergic receptors ard.,eitht~ nicotinic, those 6n
pos~ganglio~ic neurons,: or muscarinic~ thos~ on.~a~et c~lls such las.-cardia~
musele: This classification evolves from receptor~activation by ~nieotineLor
museadne.Similarly, there are two types of adrenergic :receptors, alpha (tz),and
beta (fl). Generation of a second messenger and :a greater sensitivity
CONTROL SYSTEMS OF THE HEART 23
Figure 2.7 Effects of autonomic nervous activity throughout the coronary system.
Neurotransmitter
receptortypesdiscussedin section2.2.1 are also listed.
The ANS has four basic effects on cardiac performance. They are
chronotropic (heart rate), inotropic (heart muscle contractility), dromotropic
(conduction delay in the AVnode), and bathmotropic (excitability). Inotropic
effectSare easily measurable and thus are useful in: monitoring sympathetic and
parasympathetic activity. Positive inotropic effects are exhibited by the
sympathetic branch. Overall, the enhancementof calcium into the effector cell
during the plateau phase of the action potential occurs (see Chapter 1). The
parasympathetics cause negative inotropic effects. The sympathetic branch also
experiences positive Chronotropic and dromotropic effects. Likewise, the
parasympathetic branch expedence~negative’ effects.
Ba~thmotropic changes are-minor influences as a result of sympathetic
stimulation and offer no means for control. The sympathetic nerves ~act to
increase the excitability of the myocardial cells by lowering the threshold
potential. This parameter is not only minor in it~ ~ influential aspects but
conceivably unmeasurable as well.
A number of methods exist that facilitate the monitoring of ANSactivity.
Systolic time intervals, ventricular ihotropic parameters, and stroke volumecan
all be monitored by intracardial impedance measurements. The pre-ejection
period (PEP) represents the time interval between the onset of ventricular
excitation and the opening of the semilunar valves and is also a measure of ANS
activity. Sympathetic activity acts to decrease the PEP duration, while the
parasyfiapathetics lengthen it (Schaldacl~, 1992).
In any Control system i~ is essential to determine what is being controlled and
what is doing the controlling. In the case of the cardiac system it is ex~ident that
the meanartedolar blood pressure is the factor being controlled in a closed loop
24 DESIGN OF CARDIAC PACEMAKERS
In examining the control system, we must consider the proper and most effective
parameters. These are no longer only cardiac output, medal pressure, and heart
rate. Metabolic alterations, chronotropic, inotropic, dromotropic, and
bathmotropic changes are reflex response mechanisms that neural components
effectively regulate. Of these, inotropic responses appear to provide the most
useful measure of sympathetic activity (Schaldach, 1992). The duration of the
pre-ejection period (PEP) facilitates this reasoning. However,we must ponder all
of the alternatives in order to gain enoughinsight towards the establishment of an
efficient Control parameter.
The sympathetic contribution to the overall operation of the heart appears to
be substantial. The vast amountof research that focuses on developing the proper
control pathwaysindicates that direct nervous control of myocardial effector cells
provides a specialized and extremely efficient regulatory action (Randall, 1973).
Hormonaland heterometric (intrinsic) controls are no longer beiieved to be the
major or predominant control mechanisms.The neural innervation specifically at
local control levels remains unknowndue to the intricate meshworkof the nerve
fibers. However,information obtained so far provides an optimistic foresight in
achieving this goal; characterization of local neuron pathways for control
purposes by using stimulation techniques. Until this occurs, specific control
options are limited.
chronotropic effects indicate the high probability of a sinus rhythm due to the
stellate excitation. The left side predominantlycauses an increase in contractile
force (positive inotropic effects). Stimulation of the stellate nerve can produce up
to a 35-45%increase in fight atrial inotropic activity. The left atria exhibits no
response to such stimulation. Hence, if a neural rate-adaptive pacemakeris used,
it wouldnot be effective to sense the left atria if stimulating or pacing the stellate
nerve. Subsequently, this is an important consideration for lead placement.
Another sympathetic nerve called the recurrent cardiac also elicits positive
inotropic activity in the atria and considerable positive inotropic activity in the
fight ventricle (Woolseyet al., 1967). Notice that the left ventricle does not show
such activity: Similar circumstances exist here that were just mentioned above.
The craniovagal nerve is believed to contain only parasympathetic fibers.
Hdwever,~.’Woolseyet al. (1967) measured inotropic, chronotropic, and blood
pressure increases in the fight and left atria after the administration of atropine,
an inhibitor, of parasympathetic neurotransmitter activity. Ventricular increases
up t030%were also obtained, thus indicating a strong likelihood that sympathetic
fibers are present in the craniovagalnerve.
Wodseyet al. (1967) also indicate that the left thoracic vagus and the
ventrolateral cardiac nerves contain no sympathetic innervation of either, the SA
or AVnode. This is a result of insignificant increases in heart rate or contraction
force upon~stimulation. They did find that norepinephdnecaused distinct positive
inotr0pic Changes throughout the myocardium. This is an extremely common
finding and considered well known. Therefore, the specificity of myocardial
innervation is regionally distinct as well as sensitive to electrode pulse location.
as the Bachman’sbundle, inferior left atrium and coronary sinus contribute to the
overall control of heart rate. Positive chronotropic regulation occurring in
alternative locations is beneficial for engineering purposes. These locations
provide supplemental opportunities for sensing’ in order to improve cardiac
performance.
It is.rare and difficult to observe heart rate and conduction changes due to
parasympathetic stimuli. De Beer (1977) successfully measured the effects on
conduction using step changes in atrial pacing rate and parasympathetic
stimulation. The results show an exponential time course of AVconduction after
a step.change in atrial pacing. Maximalparasympathetic stimulation increases the
exponential time constant by a factor of four. These effects are independent of
sympatheticactivity. _
Levy and Martin (1977) also provide a focus on the interaction between the
effects of AVconduction and heart rate. In order to measure the combined
effects, stimulation to an unpaced and paced heart is necessary. Comparingthe
recordings of the heart rate and AVconduction time provides interesting results
from ~similar parasympathetic stimuli. Their results indicate considerable
nonlinear interactions betweenthe direct and indirect vagal effects.
Twomechanismscause these results. First, the vagal stimulus altered the
atrial activation patterns indicating a shift in the cardiac pacemakingsite. Second,
by increasing the cycle length, the depressive effects of acetylcholine on AV
conduction improve (De Beer, 1977). The details which remain unclear are the
interactions betweenthe heart rate and parasympatheticactivity.
Acetylcholinesignificantly reduces the atrial action potential. This results in
a reduction of the atrial refractory period upon parasympathetic stimulation
(Zipes, 1974), Acetylcholine effects on ventricular refractory periods remain
unclear. It is certain that acetylcholine antagonizes adrenergic effects. The dispute
exists over the degree of independent control of refractoriness by the
parasympathetics. Regardless, the lengthening of refractory periods is beneficial
in terminating arrhythmias whichare sustained by a reentrant circuit. Conduction
can be stopped if the refractory period is lengthened enough. This allows for the
interruption of the reentrant feedback that maintains the arrhythmia.
Vagal., stimuli also control inotropic effects both in the ventricle and atria. De
Geest (1965) successfully shows negative inotropic effects up to 25%in the
ventricles during maximalvagi stimulation. The regional distribution of efferent
vagal,fibers is highly specific. Theinotropic effects are stronger at the base of the
ventricle than at the apex. This is useful in determining the placement of
electrodes if pacing by neural rate-adaptive mechanismsfor maximaleffects.-
Atrial inotropic effects are more sensitive to vagal influences than
ventricular effects are. This is due to higher concentrations of cholinesterase and
richer intrinsic and extrinsic parasympatheticinnervations. These factors result in
more rapid onset and termination responses. Also, the diffusion distance of
acetylcholinein the atria is mucli smaller than in the ventricle. However;these
can not be independent factors as Priola et aL (1980) indicate. "The,ventricular
responses to intracoronary injections of acetylcholine have a slower onset and
slower time course than do the atrial responses" (Randall, 1984),
Ventrieular,standstill occurs: during vagal stimulation having an effect on
cardiac function:This is usually-followed by a brief period of positive
chronotropy. Overdrive suppression is the mechanismthat .accounts for these
events: The sinus rate is much faster than the ventrieular pacemakers- alone.
Therefore, continuous overdrive of the idioventricular pacemaker cells occurs
30 DESIGN OF-CARDIAC PACEMAKERS
The maintenance of blood pressure is necessary for blood flow to the organs and
cells of the, body. Althoughit is~difficult to measurethis parameter as a means,of
cardiac control, it warrants focus as an influential factor in the cardiac control
system (Figure 2.2). The desired mean arterial blood pressure is 90 mmHg.
Baroreceptors located on strueturesin the heart such as the aortic arch and
carotid sinus sense the blood pressure. Pressures above 90 mm,Hgcause the
stretch receptors to expand and discharge moreaction potentials. If~the pressure
drops below 90,mmHg; the inverse occurs.-
The response to changes in the~ meanarterial blood pressure is complex,
involving several, areas of the brain. , Whenthe pressure rises above 90 mmHg;
CONTROL SYSTEMS OF ~THE HEART 31
the increased action potential activity in the fibers from baroreceptors activates a
cardiovascular regulatory center in the brain stem. This is the depressor center
andcauses a reduction in blood pressure (Figure 2.8). Neurons in the depressor
center,synapse with preganglionic sympathetic nerve cells to inhibit their
activation, This inhibition of sympathetic’nerve cells i~educes the heartYs force of
contraction and chronotropic activity. It also reduces the vasoconstriction of
blood vessels. The depressor center also activates the nucleus of the vagus,
another region of the brain stem which directly reduces heart rate. The net result
of these actions produces .a reduction in the meanarterial blood pressure. Figure
2.2 shows a complete pathway of events.
Heart
Higher
centers
center
Vessels
Rp
Figure 2.8 The hypothalamus control of MABP.CO, Pa, and Rp are cardiac output, arterial
pressure, and total peripheral,resistance respectively.
The catecholamines that circulate in the blood affect the same receptors and
target organs as postganglionie nerve cells. However, their actions are not as
discrete as those generated by sympathetic nerve cells because they enter the
blood. Circulating catecholamines have positive inotropic and chronotropic
effects, causing the arteries of the skeletal and coronary musclesto dilate. The net
effect of these actions is to provide more blood to skeletal muscles, heart, and
brain and to reduce blood flow to the vegetative organs in the body.
Several factors influence the function of the adrenal medulla. Anindividual’s
emotional state is a primary factor. The hypothalamus operates as the brain
center for emotion. Stress-induced increases in nerve cell activity in the
hypothalamus activate preganglionic sympathetic neurons. These preganglionic
neurons then innervate and activate the adrenal medulla; Causing the secretion of
epinephrine and norepinephrine into the blood stream. Other factors such as the
extreme loss of blood, hypothermia; ~and hypoxia will also activate the adrenal
medulla, attempting to correct functional disorders and maintain a constant
internal environment.
For control purposes, it is:difficult to design a pacemaker that is rate-
adaptive to the kinetic turnover rate of catecholamines ~in the adrenal medulla.
However, they do play an integrative role in sympathetic activity, which can
cause secondaryeffects at the cardiac level.
2.4 REFERENCES
Levy, M.N. and Vassalle, M. (eds.) 1982. Excitation and neural control of the heart. Bethesda,
MD:AmericanPhysiology Society.
Levy,M. N., Iano, T., and Zieske, H., 1972.Effects of repetitive bursts of vagal activity on heart
rate. Circ. Res., 30: 286-295.
Little, R. C. (ed.) 1980. Physiblogyof atrial pacemakersand conductive tissues. MountKisko,
NY:Futura.
Martin, P., 1977. The influence of the parasympathetic nervous system on atdoventricular
conduction.Circ. Res., 41: 593-599.
Porter, R., 1974. The physiological basis of Starling’s law of the heart: A Ciba Foundation
symposium.NewYork: Associated Scientific.
Pdola, D. V., 1980.Intrinsic innervation of the canineheart. Effects on conductionin the atrium,
atrioventricular node, and proximalbundlebranch. Circ. Res., 47: 74-79.
Randall, W. C. (ed.) 1984, Nervous control of cardiovascular function. NewYork: Oxford
University Press.
Rhoades, R., and Pflanzer, R. 1992. Humanphysiology. NewYork: Saunders.
Rosenblueth,A. and Simeone,F. A. 1934. The interrelations of vagal and accelerator effects on
the cardiac rate. Am.J. Physiol., 110: 42-55.
Schaldach, M., 1992. Electrotherapyof the heart. NewYork: Springer-Verlag.
Sideman,S. and Beyar, R. (eds.) 1985. Simulation & control of the cardiac system (1, 2, 3).
Boca Raton, FL: CRCPress.
Strackee, J., and Westerhof,N. 1993. Thephysics of heart andcirculation. Philadelphia:Institute
of Physics Publishing.
Szentivanyi,M. J., and Pace, J. P. 1967. Localizedmyocardial,responses to stimulationof cardiac
¯ ~sympatheticnerves. Circ. Res., 21: 691-702.
vander," A. 1990. Humanphysiology. 5th Ed. NewYo~k: McGraw Hill.
Wallick, D.’W., and Levy, M. N., 1979. Effects of repetitive bursts of vagal activity .on
atrioventricular junctional rate in dogs. Am.J. Physiol., 237:275-281.
Warner,H. R. and Cox, A., 1962. A mathematicalmodelof heart rate control by sympatheticand
vagus ~fferent information. J. AppLPhysiol., 17: 349-355.
Wo01~y’,M. D., Brody, D. A., and Arzbaecher, R. C., 1967. Measurementof spontaneous
rn0rphologiealvariations in the electrocardiographic p-wave.Comp.in Biomed.Rest, 1: 265-
275.:
Zat].chetti,: A,, and Bartorelli, C. (eds.) 1970. Cardiovascularregulation in health and. disease.
Milan:Instituto di RicercheCardiovascolari.
Zipes, ’D., Mihalick, M., and Robbins,G., 1974. Effects of selective vagal and stellate ganglion
stimulationon atrial refractoriness. Cardiovasc.Res., 8: 647--655.
As one can tell from the often repeated quote above, manypeople over the years
have placed a great deal of emphasis on the heart’s rhythm, but none have hadso
great an interest as those who suffer physical ailment: the cardiac patient.
Deviation in the :heart’s rhythm from physiologically normal behavior is often
associated with reduction in the quality of life--if: it doesn’t result in death.
Clearly, restoration of the heart’s natural rhythmis desirable. To meet this goal,
the physician or pacemaker engineer must first understand the underlying
mechanisms of ~he’ cardiac arrhythmia. This chapter provides a,,working
knowledgeof the causes for and mechanisms.of cardiac rhythm disturbances..
three . groups:,
¯ Cardiac arrhythmias
arrhythmias of classified’by
may be abnormal impulse
theirinitiation,
underlying abnormalities of
mechanisms into
3.1-: AUTOMATICITY
Cardiac cells that are normally automatic are called pacemakercells, and Figure
3.2 shows a hypothetical membrane potential recording. The cell does not
maintain a constant resting membranepotential but slowly depolarizes from the
maximaldiastolic potential until reaching the threshold. Uponreaching threshold,
the cell depolarizes rapidly; the result is initiation of an action potential. This
slow depolarization is called phase 4 depolarization or diastolic depolarization,
and Chapter 1 describes the mechanismsthat give rise to this behavior.
(a)
(b)
Tpl"-"
(c)
(d)
Unlike the specialized pacemaker cells discussed above, normal atrial and
ventricular myocardial cells do not exhibit-automaticity, Workingatrial and
ventricular cells do not normally have spontaneous ~astolic depolarization nor do
they spontaneously.initiate impulses even.after .haviog been unstimulated ,for long
periods of time. Whena cell’s resting potential has been lowered below its
normal levels (made more, positive), however, spontaneous diastolic
depolarization mayoccur and cause repetitive impulse initiation. This is called
MEC~NISMS OF CARD~C ARRHYTHMIAS 37
0 Musclefiber
potential
B
-80 mV
C
A A
Microelectrode
current
Normally the pacemaker of the heart, the sinus node maintains its dominanceby
depolarizing all other potential pacemakersbefore, they have.a chance to initiate a
beat. Figure 3.5 shows that the sinus node simply outpaces all other latent
pacemakers, resetting them before’ they reach threshold on their own. This is not
the only mechanism, however, by which the pacemaker hierarchy is enforced.,
Overdrive suppression and cell-to-,cell interactions also work to suppress latent
pacemakers.
Sinus node 0 ~.
B
Latent 0 ~_
pacemaker T ......
A
Overdrive 0 ....
suppression
T-~-
Overdrive suppression
The~bottomtrace of Figure 3.5 indicates that not only ~-are the latent pacemakers
prevented fromreaching threshold~when being driven by the-sinus node, the
slope of their diastolic depolarization is actually inhibited. This effect is called
overdrive suppression and can be demonstrated by stimulating a Purkinje fibe!~.at
a rate of 90 impulses/min. It takes more than 20 s after :termination of~ the
stimulation before the first spontaneous impulse arises. The r~ate ~gradually
increases until normal intrinsic rates have been restored (Van Capelle, 1987).
The mechanism of overdrive~ suppression is ion-based. Na+ enters a cell
during each action potential, and, when. a pacemakercell is driven at rate faster
than its intrinsic rate, the intracellular concentration of Na+ +.
increases. The. Na
K+~ pumpactivity,increases in an attempt.to ~re8tore normal, transmembrane
electrolyte gradients. Since the pump-extrudes 3 Na+for each 2 K+ brought in,
there is a,,net positive~ current out of~the ~ell w~chhyperpolarizes th~membro~e
and- counteracts diasto!ic~depolarization (Frame~and- Hoffman~1984). Although=
this hyperpolarization is slight, only_ a few~extra millivolts are enoughto cause
slowing or even completf,, suspension of spontaneousdepolarization. ~
MECHANISMS OF CARDIAC ARRHYTHMIAS 39
Because the sinus node is the dominant pacemakerof the heart, alterations in its
rate maylead to an’hythmias. Sinus tachycardia results whenthe sinus node fires
at rates.in excess of 100 beats per minute. Conversely, sinus bradycardia occurs
whenthe, sinus node fires, at less than 60 beats ~per minute, .Note that ~either of
these two conditions maybe either normal or abnormal. Sinus.tachycardia can be
the appropriate response to external factors ~such as exercise, fever, or
hypotension. In well,conditioned athletes, sinus bradycardia can be the normal
response to exercise-increased parasympathetic stimulation.. In the average
person, however, sinus bradycardia more often reflects an abnormality of the
sinus node and escape pacemakers due to disease, .abnormalities in
parasympathetic stimulation, or outside factors such as drugs (Myerburg et al.,
1994)
3.1.5 Arrhythmias arising from ectopic pacemakers -~
ArrhythmiaS due to automaticity also result when the site of dominant
pacemakingshifts from the sinus node to any ectopic (nonsinus) pacemaker. Sinus
40 DESIGN OF CARDIAC PACEMAKERS
The second cause for ectopic rhythms is the removal of inhibitory influences of
the surrounding myocardium. Decreasing the coupling between latent pacemaker
cells and the surrounding nonpacemakertissue mayremove the inhibitory current
flow described in section 3.1.3 and allow the latent pacemakersto fire at their
intrinsic rates. Coupling can be reduced by fibrosis, which can separate
myocardial fibers, or by factors that increase the intracellular Ca2+levels (Waldo
and Wit, 1994). Fibrosis in the atrial portion of the AVjunction mayfreethe
nodal pacemakersfrom suppression by the atrial tissue and allow them to control
the ventricular rhythm. Purkinje fiber pacemakers may be separated from
damagedventdcular muscle cells during myocardial infarct allowing the Purkinje
fibers to fire at their normal, intrinsic rates (Waldoand Wit, 1994), Note that
someinhibition of the sinus node is still necessaryfor impulseinitiation to shift tO
uncoupled pacemaker sites because they still have slower intrinsic rhythms than
the sinus node and are subject to suppression:
can also
Early occur during the rapid
afterdepolarizations often rep01adzation ~ Of
arise from the phaseof
eplatea~ 3 (see Figurepotential
an action 1.8). Figur
b’ut
0
-90 mV
(a) (b)
Early afterdepolarizations can occur in almost any type of cell but have been
most studied in cardiac Purkinje fibers and ventricular muscle calls. EADsand
triggered activity have been produced experimentally under a variety of
conditions, most causing marked delays in repolarization of heart cells. Slow
heart rate and drug toxicity are.:two examples (Bigger, 1994). Events that
increase inward current Componentsor decrease outward currents are expected to
result in conditions favorable to EADs(Wit and Rosen, 1989).
Whenthe rate of stimulation is markedly slowed, the outward current
+~
generated bythe Na+-K÷pumpis ~reduced, especially whenthe extracellular K
is lower than normal. At physiological ranges of cycle lengths (1000-700 ms),
EADshave occurred rarely in the studies on isolated cardiac fibers, but as cycle
lengths increase and repolarization is prolonged, EADsare more likely to occur.,
Another important observation is that a longer drive cycle results in a greater
number of impulses triggered by. EADS(Waldo and Wit, 1994)
Once EADshave reached a steady-state amplitude ata constant drive cycle
length, ,any event that shortens the drivecycle reduces the EADamplitude (Waldo
and Wit, 1994). A single .premature,~depolarization accelerates repolaxization and
will reduce the accompanyingEAD.Triggered activity, therefore, is not likely to
follow a premature stimulus. Treatments ~that shorten action potential ’duration
are therapeutic. Rapid pacing, increasing extracellular K+ concentration, and
drugs that increase K+ conductance tend to eliminate EADand triggered activity.
+
Note that each of these methods shares common property: increased K
conductance in heart muscle (Bigger, 1994).
I tt I t
t
second release of Ca2+ causes the TI is still unclear, however. After one or more
afterdepolarizations, intracellular Ca2+ levels drop because Na+-Ca 2+ exchange
extrudes Ca 2+, and the stops oscillating. (Waldo and Wit,
1994).
One cause of DAD-tri is digitalis toxicity. One possible
mechanismis that di pumpresulting in a measurable
+
increase in Na +
doses, The increased intracellular Na
concentrations decrease the Na+ gradient that drives the Na+-Ca2+ exchange,
reducing Ca2+ extrusion a~d~increasing intracellular Ca2+ levels. The result is a
net inward Na+ current which C~ lead to DADs(Cranefield and Aronson, 1988).
Catecholamines are also an accepted cause of DADs.They mayc~iuse delayed
afterdepolarizations by increasing the slow inward Ca2+ current as well as
enhancing sarcoplasmic uptake (Wit and Rosen, 1989)~
Delayed afterdepolarizations and triggered activity mayalso occur without
pharmacological agents, catecholamines, or increased levels of Ca2+. Triggerable
fibers have been found in animal models and in apparently normal humanatrial
myocardial cells (Wit and Rosen, 1989).
Because the transient inward current that causes delayed afterdepolarizations
is at its maximum around -60 mV,the possibility of triggered activity depends on
the level of the membranepotential. In digitalis toxic Purkinje fibers, there is a
i:ahge of maximaldiastolic potentials near-70 mVok, er which the amplitude of
DADsis greatest (Wasserstrom and Ferrier. 1981). WhenDADsoccurat these
favorable membranepotentials, any action that hyperpolarizes ordepolarizes the
membrane tends to reduce the amplitude of the DADsand suppress any DAD--
triggered rhythms (Waldo and Wit, 1994). Simil~ly, if there are no DADsin the
presence of digitalis and the membraneis outside this Wiridow, any ~/ction that
iJrings the’membrane into this range often induces DADs. A similar DAD
dependende on membrane potential has been shown in atrial fibers of the
coronary sinus and Purkinje fibers from infarcts (Waldoand Wit, 1994).
: Delayed afterdepolarization amplitude is affected by the duration of the
action potential. Longer action potentials allow more Ca2+ to enter the cell,
favoring DADs.Drugs such as quinidine, which prolong action potentials (APs),
may increase DADamplitude while those such as lidocaine~ which shorten the
action potential duration, maydecrease DADamplitude (Waldo and Wit, 1994).
Delayed afterdepolarization amplitude also depends on the number of action
potentials preceding it. After a period of inactivity, a single APwill result in a
smallDADor noneat all. With continued stimulation, the DADsincrease in
amplitude and triggered rhythms may occur (Waldo and Wit, 1994).
The period of stimulation~ however, also affects DADamplitude; triggered
activity can arise if this cycle length is ~?educedbelowsome~Cridcal value. Figure
3.8 shows this cycle length dependencyin an atrial fiber located in the canine
~oronary sinus. Stimulation with a cycle period of 2000 ms resulted in a 5 mV
DAD f611owingthe last stimulated action potentialS. In the center, stimulating the
c611s every 1500 ms led toa final afterdep~larization amplitude of 15 inV. On the
right, a drive cycle period of 1200 ms ended in triggered activity with a 20 mV
¯ ~kDjust before initiation. A decrease in the len~ ~f even~ single driVe cycle,
~ premature impulse for example, results in an increase in the amplitude of the
DAD that follows the premature beat, and arrhythmias caused .by DADscan be
eXpeCtedtb be initiated by aspontaneous or paced increase in heart rate (Wit and
Cr~mefield, 1977).
46 DESIGN OF CARDIAC PACEMAKERS
7500 msec
------’----4
50 mV
Thus far, the mechanismsfor arrhythmia have been defects in impulse formation.
Now,abnormalities of impulse conduction will be examined.
Delay of cardiac impulses can occur anywhere in the heart, and it can be
general or localized. Therapy witha typic IC antiarrhythmic ~g, fo~ exam~.ple,
leads to delay in all cardiac tissue, while tissue injured in infarction mayconduct
impulses slowly even though the surrounding tissue conducts normally. On the
other hand, conduction delay maybe normal: one example is the ~increased~time
for AVnodal conduction of a premature beat. Conduction delay Will play alarg~
role in tachycardias Causedby reentry (section 3.4)
If the delay is extreme, the impulse maybe blocked. There are .two ~forms of
block: bidirectional and unidirectional. Because unidirectional block is so
intimately~ tied with reentry, discussion of this will be also postponeduntil section
3.4.
48 DESIGN OF CARD~C PACEMAKERS
Areas of slow conduction maybe anatomically normal; the AVnode, for example
introduces an electrical delay between the atria and ventricles that allows
ventricles to fill completely before theycontract. Areas of slow conduction, on
the other hand, mayexist where they are not normally expected. This latter type
of Slow conductionis not present during normal sinus rhythm but is functionally
present during an arrhythmia. Tissue. damaged by myocardial infarct is one
example.
There are several causes of slow conduction that maylead to arrhythmias:
changes in membrane current, changes in the cable properties of the cell,
anisotropy, or changes in gap junction resistance.
Changes in current
-90 mY
Figu~?3.9. Therelationship between actionpotentialslewrate. vs. membr.ane potentialat the time
of actwatlon. (a) Slewrate decreasesw~threducedmembrane potentials. (b) Theeffect
prematureStimulusoncardiactissue that has previouslybeenactivated.Thedashedline at stimulus
Aghowsthat the rate of depolarizationis reducedbecausea stimulus occurredat a reduced
membranepotential.. Apremature stimulusat Bresults in a fastexaetionpotentialthanthe stimulus
at. A,but it is not fully normalbecausethe tissue has not fully repolarized.FromWaldo,A. L. and
A. L. Wit. 1994. Mechanisms of cardiac arrhythmias~andconductiondisturbances. In R. C.
SchlantandR. W.Alexander(ed.) Theheart, "arteries, and:~,eins.Sthed. Withpermissionof
McGraw-Hill; Inc.
Figure 3.9(a) shows that the action potential slew rate decreases with
decreasing resting membranepotentials. Figure 3.9(b) shows that the amplitude
MECHANISMS ~OF CA~C AR~THMIAS 49
of the resulting action potential depends on the degree to which the cell has
repolarized (greater repolarization results in increased amplitude of the second
action potential). Because these impulses are conducted by partially inactivated
fast Na+ channels, they are called depressed fast responses.
There is also a slow response inward current carried by Ca2+ channels,
which contributes to the upstroke.of the action potential. The threshold for the
Ca2÷ channels is about-30 to -40 mVcompared ÷
with the -70 mVof the fast Na
channels; therefore, this current inactivates muchmore slowly. In cells with
resting membranepotentials smaller than -60 mV(when membraneconductance
is very low or whencatecholamines are present) this normally weak current may
produce .action potentials that propagate slowly and are prone to block. Slow
response action potentials can occur in diseased cardiac fibers but can also occur
in some .normal tissue of the heart such as the cells of the sinus and AVnodes
where:the maximal diastolic potential is normally smaller than -70 mV(Waldo
and Wit, 1994).
Membrane Resistance
Internal + External Resistance (3.1)
Physically, a change in potential on the fiber Will cause the potential one
length constant away to change by about lie ~of the applied potential. With a
greater length constant, a point fUrther out along the. fiber can be stimulated. The
Velocity of conduction, therefore, increases with,an increasing 2L (the distance
covered per unit time is greater). If all other influences are held constant, the
conduction velocity will increase when the membraneresistance increases and
decrease wheneither or both of the longitudinal resistances increase (Cranefield
and Aronson, 1988):
The length constant typically ranges from 0.5 to 225 mm,meaning that a
100 mVaction potential will cause cardiac tissue two ~,away (1 to 5 mm)
depolarize to threshold. In a passive cable, a 100-mV impulse causes a
depolarization at two ~, of 100/e2 or approximately !3.5 inV. Cardiac fibers are
not passive, however. The effective length constant is reduced during .the
upstroke of an action potential’ in an excitable fiber so that the actual ~voltages
experienced at two length constants will be somewhatdifferent (Cranefield and
Aronson, 1988).
The external longitudinal resistance (resistance to current flow in the
extracellular space) can rise significantly if the extracellular space is reduced
during swelling of injured cells, This swelling mayoccur, during ischemia; in
fact,..., merely~interruptingperfusion,, which. ~. causes. . a loss of "stiffness" in the
tissUes, can also cause an increase in intracellular resistance. Ischemia mayalso
me the lntracellular longitudinal resmtance by uncouphngat the location of
the intercalated discs or ischemia may. lower membraneresistance by increasing
external K+ (Cranefield and Aronson, 1988).
50 DESIGN OF CARDIAC PACEMAKERS
A nisotropy
The cable properties just mentioned also form the basis of cardiac anisotropy:
conduction velocity depends on the direction in which it is being measured.
During conduction, axial current flows from one cell to the adjacent cell through
the gap junctions of the intercalated disks. These disks form a major source of
resistance along~the fiber bundle. The structure of the myocardiumthat governs
placement of these disks, therefore, has a large effect on fiber resistance and
conduction.
Muscle fibers are arranged in unit bundles in which cells are tightly
connected with each other. Each unit bundle is connected with others but the
connections only occur in a lateral direction at intervals.of 100-150 #m (Waldo
and Wit, 1994). The result is that the myocardiumis better connected along the
long axis because of the low frequency of connections between the unit bundles
and the high frequency of connection within the bundle itself. The result is a
reduced axial resistivity relative to a transverse direction and a corresponding
difference in conduction velocity predicted by cable theory.
There are two types of anisotropy: uniform and nonuniform. Uniform
anisotropy (Figure 3.10(a)) is characterized by an advancing wavefront that
smoothin all directions (longitudinal and transverse .to fiber-direction). In one
experiment in normal.septal muscle, conduction in the longitudinal direction was
found to be 0~51 rrds and in the transverse ~ direction 0.17 rrds (Spach and Dolber~
1985). As the direction of propagation changed between the two, the conduction
velocity changed monotonicly (Waldo and Wit, 1994).
M-usclefib-er I uscle
fiber\
Longitudinal Longitudinal
axis axis
(a) Uniform anisotropy (b) Nonuniform anisotropy
Normally, the action ,potential from the sinus node dies out after orderly
depolarization of the atria, AVconduction system, and the ventricles. The
impulse does not usually conduct backwardsbecause the tissue just stimulated~is
refractory and, therefore, unable to generate a second action potential. As a
result, the normal heart must wait for new sinus pulse for each subsequent heart
beat.
Reentry occurs whenthe action potential does not die out but continues, to
propagate and-reactivate the heart. Almost all clinically, important
tachyarrhythmias are due to reentry (Waldo and .Wit~ 1994). Reentry can occur
almost anywhere in the heartand can assume manysizes and shapes.
Section 3.4.1 explains the concept of reentry by example. Section J3~4.2
outlines reentry classifications. Section 3.4.3 explains the general requirements
for .reentry. while section 3.4.4: describes in detail one of the most important
requirements: unidirectional block. Finally, section 3.4;5 introduces reflection: a
special case of reentry.
or if-the loop is large enough that even at normal conduction speeds, the tissue
has time to recover its excitability. The counter-clockwise pulse can then reexcite
tissue that it has already passed through. This process continues, with the pulse
circulating to, stimulateitself.
./-lo~e
Jellyfi~,,~hRing
If the block is bidirectional, each pulse circulates until it reaches the block
and dies. The cells behind the wavefront are refractory and ,unable to be
stimulated againwhile thecells ~ahead are incapable of being stimulated. Thus, the
unidirectional block is essential for initiating the reentrant rhythm.
The wavefront will continue to circulate provided a few. prerequisites have
been met; reentry requires a region of unidirectional block, a central unexcitable
area .~ound which to circulate, and an action potential wavelengththat is shorter
than the length of circuit. Wavelengthis defined as the product of the conduction
velocity of the circulating wavefront and the effective refractory period of the
tissue ~0f the. reentrant circuit (Waldoand Wit, 1994). The effective refractory
period is defined as the time between the beginning of depolarization and the time
at which be earliest impulse is,capable of further propagation in the conduction
system (Alpcrt~ 1980).
3.4.2 Classification
Ordered reentry
also cause ordered reentry if these properties are conf’medto a specific location
and reentry occurs .only in that location.
Examples of ordered reentry are atrial flutter, most monomorphic
ventricular tachycardias, AV nodal reentrant tachycardia, AVreentrant
tachycardia with an accessory pathway, and sinus node reentrant tachycardia.
Random reentry
3.4.3 Prerequisites
tissue through which the waveis propagating. Recall that the effective refractory
55
period is defined as the time between the beginning of depolarization and the
earliest time.at which an impulse is capable propagation. (Alpert, 1980). For
reentry to occur, the wavelengthof reentrant excitation must be shorter than the
length of the pathway. This guarantees that the impulse always finds excitable
tissue.
Note that an area of slow conduction is not absolutely required. Reentry can
occur at normal conduction velocities so long as the path length is long enough.
The jellyfish ring is one example. For virtually all clinic~ally relevant reentry
circuits due to ordered reentry, however, the path length would be too small (or
the reentrant wavelength would be too. large) given a constant speed of
cOnduction. The action potential would travel too quickly around the circuit and
reach tissue that is still refractory; thus, almost all the reentrant arrhythmias have
regions of slow conduction (Waldoand Wit, 1.994);
Areas of slow conduction arise by the mechanismsdiscussed in section 3.3.
Conductionslow enough to cause reentry mayoccur in diseased cardiac cells with
persistently small membraneresting potentials of about -60 to -70 mV.At these
potentials, about 50%of the Na+channels are inactivated. Slow response action
potentials can also occur in somenormaltissue of the heart such as the cells of the
sinusand AVnodes where the maximumdiastolic potential is normally .smaller
than,.-70 mV.Slow conduction is a normal property of these nodes, and either
node maybe a critical portion of a reentrant circuit, for example AVreentrant
tachycardia.
Slow conduction in reentry mayalso be caused by other factors besides those
that affect current. Anisotropic circuits usually remain fixed and can cause
ordered reentry. Anisotropy can slow conduction enough to cause reentry in
small anatomic circuits or in functional pathways: circuits in nonuniform
anisotropic bundles can be as small as 2 to 4 mm2.Also~:the~oircuits are elliptical
or rectangular because of the directional differences in ~0nd~tion with the long
axis in the direction of the fast, longitudinal direction. Circuits with this shape
can have a smaller size than circular circuits such as the leading Circle (see below)
(Waldo and Wit, 1994).
Changes in the effective refractory period of the cardiac tissue mayalso
facilitate reentry~ A decrease in the refractory period decreases the wavelengthof
reentrant impulse, thereby decreasing the necessary size of the reentrant circuit.
If the wavelengthdecreases, the need for slow conduction also decreases.
The effective refractory period of cardiac, fibers maybe shortened during
rapid tachycardias because of rate-dependent shortening of action potential
duration. In atrial muscle the refractory period is shortened by acetylcholine
released during vagal stimulation makingatrial fibrillation easier to induce.
Action potential durations and effective refractory periods are also decreased in
the ventricle during reperfusion following small periods of ischemia or in some
of the ventricular muscle ceils in chronic i~chemia (Waldoand Wit, 1994).
Changesin gap junctional resistance and other factors that affect the fibers"
cable properties are also expected to contribute to the slow conduction necessary
for most forms of reentry.
Reentrant tissue can be located anywherein heart, and the circuit can have a great
numberof shapes, sizes and types of cardiac tissues. The reentrant ~substrate can
56 DESIGN OF CARD~C PACE~KERS
The central area of block necessary for reentry may be anatomical (static),
functional (dynamic), or combination of the two. Anatomicblock is the result
a nonconductive region .in the center of the reentrant, circuit. One example is
atrial flutter found in patients whopreviously have had a ,Mustard procedure to
repair transposition of the great vessels (Waldo and Wit, 1994). The procedure
involves makinga large incision in the atrial wall:
Functional block occurs when there is .a block of impulses in otherwise
excitable cardiac tissue. During the healing phase of myocardial infarction, a
dynamic block may be formed in the ischemically damagedbut surviving cells
(Bigger, 1994). The ventricular excitation wavefront enters the ischemic zone
find an arc of refractoriness. The excitation wave movesslowly around the edges
of the arc and around the back until it reaches a point near the. center which by
this time has recovered its excitability. The wavefront then conducts through .the
center and back downthe sides for a reentrant circuit causing tachycardia. Figure
3.12 shows the reentrant pathway.
Figure3.12 Dynamic blockleadingto reentry. (a) Theimpulsefinds the central portionof the
isehemicarea refractoryandis blocked.Theimpulsetravels slowlyaroundthe central blockand
aroundthe backside.(b) Bythis time the central portionhas regainedits excitability andthe
wavefrontconductsbackto whereit Started.Frtm Bigger, J. T. 1994.ElCetropl~ysiology,
diagnosis,and management.In R. C. Schlantand R. W.Alexander(ed.)Theheart, arteries, and
veins. 8th Ed. Reproduced
with permissionof McGraw-Hill, Inc.
If there are differences in the duration of the effective refractory period that
occur in adjacent areas, conduction of an appropriately timed premature impulse
maybe blocked in the region with the longest refractory period.
In Figure 3.13(a), a premature beat arises in a region of short refractory
period. Tl~etop t~a~e shows a quick decline from the maximally depoladze~l
MECHANISMS OF CARDIAC ARRHYTHMIAS $7
level~ Becausethe tissue has been stimulated previously, conduction downthe left
path blocks: That tissue on the left has a long refractory period, and, since the
strength~of:the action potential is related to the membranelevel at which it is
evoked(section 3.3.1), the second potential ~is too weak to initiate another
depolarization. Conduction precedes downthe-right path, however, because that
tissuehas a short refractory period and has had time to repolarize almost fully.
Figure 3.13(b) completes the reentrant circuit. By the time the wavefront
circulates to the left branch, the tissue has had time to recover its excitability.
Since. the tissue at the top and the right of the-circuit have short refractory
periods, they too are excitatory. If the time around the loop is long enough for
the left branch to repolarize sufficiently before the next stimulation, reentry will
continue.
(a)
.~ ~_..Premature beat
Blocked / / Conducted
period period
returns after the area has recovered (section 3.4.4). Figure 3.14 shows that the
excitation circulates around a central area ,that is kept refractory because it is
constantly depolarized by the circulating wavefront (a:functional central area of
block). The circumference of theleading circle may be as small as6 to 8 mmand
represent a pathway that is just long enough to allow for stimulation even though
the tissue is still partially depolarized. Conduction through the depolarized
circuit, therefore, must be slowed.
The different refractory periods of the atrial fibers in close proximity with
each other madethe reentry possible, and canine modelsindicate that atrial flutter
may operate by.the leading-circle mechanism(Waldo and Wit, 1994).
For reentrant arrhythmias due to regional differences in refractory periods,
therefore, both a trigger (the premature impulse) and the appropriate substrate
(reentrant circuit) are needed. The cause of the trigger is quite different from the
arrhythmia it initiates.~ .~It mayarise s~ontaneouSlyby ai~t0maticity or maybe
result of triggered activity. It mayeven be induced by electrical stimulus during
programmedstimulation.
The differences in the effective refractory periods (called the dispersion)
required for a reentrant circuit maybe quite small.In the atria, the minimal
dispersion could be in therange 11 to 16 ms, well within normal physiological
range of variation. In the ventricles, where the refractory periods are much
larger, the difference betweenthe longest and shortest refractory period duration
is on the order of 40 ms. Unlike. the atria, the differences in refractory periods is
not large enough to allow reentry initiated by premature impulses. Differences in
refrac.tory periods must be increased to 95 to l~5.ms before a premature stimulus
can trigger a reentrant circuit (Waldoand Wit, 1994).
The refractory period’s dispersion is not the- only factor that determines a
premature impulse’s ability to stimulate reentry, however. Relatively long and
short refractory areas must be reasonably dose to each other so that, a premature
stimulus can reach the areas of long action potentials in enough time to cause
block. The size Of block is also critical. If it is too small, an impulse traveling
through the area of unidirectional block maynot be~delayed sufficiently to allow
the site of the prematurebeat to repolarize.
DecreasingExcitability ...."
Reflection ~’~~
II ~/~ Depressed
Many6f the arrhythmias already discussed rely on the interaction of’ several
simu~eous abnormalities; however, often, they were unrelated. Reentry, for
exami~i~, requires certain areas of delayed conductionas well as a trigger to start
the reentrant rhythm, The source of the trigger is not ,relevant; it on|y:..matters
that~a trigger occurs. In the next two examples, the arrhythmia arises from a
specific combination of both abnormal impulse generation and conduction:
3.5.1 Parasystole
A ’ lat~rit
.... pacemaker maybe protected from overdrive suppression if it is
surrounded by a region that blocks sinus impulses (entrance block). This block
mustbe unidirectional, however, so that the latent pacemaker impulses can
propagate into surrounding myocardium. A protected pacemaker is called a
parasyst~o!ic focus. An impulse mayexit the focus and excite ~e heart ’if the
outsidetissue is not refractory. The results can be premature, beats or even
tachycardia.
62 DESIGN OF CARDIAC PACEMAKERS
Block of an impulse may occur if an impulse arrives at a-location (the His bundle
or bundle branches for example) that is partially depolarized during spontaneous
phase 4 depolarization (diastolic depolarization) but has not yet reached
threshold. This spontaneous diastolic depolarization can depolarize the tissue
sufficiently to inactivate the fast Na+ channel, resulting in failure of propagation
(Waldo and Wit, 1994).
3.6 REFERENCES
Diagnosing Arrhythmias
Valtino X. Afonso
The main step before using artificial pacemakersto managean abnormality in the
pacing or conduction system in the heart, is the diagnosis of the arrhythmia. The
effectiveness of the pacemakeras a therapeutic device depends on the accuracy of
the diagnosis of the abnormality. ~
The electrocardiograph noninvasively measures a standard I2-1ead
electrocardiogram, which is used to analyze the electrical activity in the heart.
Diagnostic criteria obtained .from the 12-lead electrocardiogram provide
discriminatory information to diagnose different abnormalities. Literature on the
subject explains the various techniques used to diagnose arrhythmias and also uses
the word interchangeably with dysrhythmia. Automated analysis to compute
features of the electrocardiogram has improved the ability to diagnose
arrhythmias. Features of the morphologyof these signals, which are summarized
in tables within this chapter, help diagnose abnoi~aaalities that mayrequire ,the
insertion of a pacemaker as therapy Ambulatorymonitoring provides the ability
to monitor patients on a long-term basis in order to detect transient arrhythmias.
During stress testing, patients are monitored for arrhythmias while performing
physical activity. In order to supplement diagnosis based on the
electrocardiogram, invasive electrophysiologic studies use transvenous
multielectrode catheters to record electrical signals directly from the cardiac
muscle. These invasive techniques include electrical stimulation of the
myocardium to assess functionality of the cardiac system.
This chapter provides information on the techniques, signals, and
interpretation of the 12-lead electrocardiogram and intracardiac signals used for
diagnosing arrhythmias.
The electrical current generated by the heart is conducted through the pairs of
electrodes and leads, and is amplified, recorded, and processed by the
electrocardiograph. The wires connecting the pairs of electrodes on the surface of
the body to the electrocardiograph are called leads. The different features and
modules of a typical electrocardiograph include the protection circuitry, lead
selector, calibration signal, preamplifier, isolation circuit, driver amplifier,
memorysystem, microcomputer, and recorder or printer (Neuman, 1992).
The protection circuit prevents any damagedue to high voltages that may
a~ear as inputs to the electrocardiograph. This is useful in situations such as
wt~n a patient is being defibrillated. ~ The lead selector can be controlled by an
operator or automated, and provides the ability to record from various
combinations of the leads connected to the patient. The electrocardiograph is
usually calibrated before an acquisition session by a 1-mVcalibration signal
Whichis applied to each channel recorded. The output for the calibration signal
should be the same for each channel recorded.
The preamplifier stage consists of a differential amplifier with a high CMRR
(common-mode-rejection ratio). The high CMRR ensures that any potential
the patient’s body that is common to both inputs of the differential amplifier is
not amplified by the electrocardiograph. The preamplifier stage feeds an isolation
circuit, which acts as a barrier by preventing more than 10/~Aof 60-Hz current
flowing through the patient to ground even under the fault condition of the
patient accidentally comingin contact with 120 V from the power line.
The input of the driver amplifier circuitry is at-coupled and this prevents
the output of further amplifier stages from saturating due to the offset in the
input signal. The ECGis then filtered with an upper comer frequency (3 dB)
150 Hz and amplified sufficiently so that it can be recorded. Modern
electrocardiographs include an analog-to-digital converter (ADC)to digitize the
signal. Data segments of the ECGfrom each lead, and other relevant information
of the patient can be stored in memory. A microcomputer in the
electrocardiograph also enables the operator to select leads to record, process
ECGsignals, perform preliminary arrhythmia analysis and other related tasks.
Twelve leads usually comprise a diagnostic ECGrecording: six limb leads (three
bi ,1~ lar and three unipolar), and six unipolar precor:dial leads. Theinstantaneous
¢~diaC scalar voltages resulting from the electrical activity in the ~eart is
measured in each of the 12 lead~, Since the cardiac vector varies in m~ignitode
wi~ fime~ overa three:dimensional space, it is important to knowits presentation
(i.e. appearanceor projection) in each of the 12 leads of the ECG. .~
¯ t~igure 4.1 sh0ws the lead placement to acquire the 12-lead ECG.The leads
Can be Categorized into the frot~tal i~ads (I, II, ~I, aVR,aVL, and aVF), and the
transverse leads (V1, V2, V3, V4, V5, and V6~). The frontal leads measure the
projection.of the Cardiac vector on :the frontal, plane of the bi0dy. The frontal
plane i s parallel to tile floor whenlying supine. The transverse or precordial
leads measure the projection of the cardiac Vectoron the h0dzontaI plane, (i.e.
the plane that is parallel to the floor whenstanding).
.Le~ads I, Ii, and iii of the frontal plane ~ .are bipolar. They record the
differdnces between two points on the body. Figure 4.1 ~hows that lead I is
66 DESIGN OF CARDIAC PACEMAKERS
measured between an electrode on the left arm (the positive electrode) and
electrode on the fight arm. The three-dimensional cardiac vector projects into
each of the bipolar leads, indicating the strength and direction of the
instantaneous cardiac vector (Luna, 1993).
v!. -l’wc’r
Figure4.1 Theinstantaneous
cardiacvectorprojectsinto eachof the leads, resultingin different
morphologies.ECGsketches are that of anormalmorphology.LeadsV1-V 6 use the Wi!son’s
centralterminal(WCT),whichis formedby the three resistor network.
Leads aVR(on the rightarm), aVL(on the left arm), and aVF(6n the
are Unip0!ar leads. They measure the potential difference on the limbs with
respect ’t0a reference point formedby the ~two resistors between limb electrodes
(Figure 4.1). For example, lead aVRis measured between an dectrode’ on the
right arm, and a reference point formedvia a resistor to the left arm and ~mother
resistor to the 16ft foot. These leads ’Show°the cardiac vector pr0jection ~ on the
frontal plane, and are amplified by about50%(i.e. aUgrnented) so that their
amplitudes are comparableto those of the bipolar leads.
’ The six precot:dial leadS~’ Vi tO V6,: ~a~e u~p01ar and measure the cardiac
vector projefh0n ~on ~e tiorizOntaFpl~e. ’These pree~rdial leads are measured
with respecf tb the e Wilson central te~al which is formed by a three.tesist0r
ne~twork ~ shown i n Figure 4A. V1 and V2 areplaced on thefqurth intercostal
space to the right and left, .respectively, I of. tbe= sternum~The V4 electrode, i~s
placed on the fifth intercostal space at the left ~idclavi~ular iine~ The V3
DIAGNOSING ARRHYTHMIAS 67
electrod
’ e lies between V2 and V4. Electrode V5 is placed to the left of V4 on the
anterior axillary line, and V6 is placed on the same’level as V5onthe midaxillary
line;~:.It is-important to account for the position of these electrodes when
interpreting the ECGon leads V1 through V6. The precordial leads measurethe
potential between each of V1 through V6 and the Wilson’s central terminal
formed as shownin Figure 4.1.
, The 12-lead ECGprovides ,various viewpoints of the three-dimensional
instantaneous cardiac vector that are somewhatredundant, and this is ltelpful in
providing discriminatory information for diagnosing abnormalities in the pacing
and conduction system of the heart.
Theelectrical activity due to the specialized cells in the heart results in an electric
potential on the surface of the body. Each cell can be modeledby a dipole, and
the superposition of the potentials from the dipoles for all of the cells in the
myocardiumresults in a three-dimensional cardiac vector for the heart at each
instant in time. The cardiac vector at each instant of time represents the net
electrical activity in the heart.
Figure 4.2 shows the Hexaxial reference system which is used in the
diagnosis of certain abnormalities. The Hexaxial system shows the orientation of
the front.al-plane leads. The various orientations of each lead result in a different
projection of the cardiac vector onto that particular lead. A meanelectrical axis
as a function of time, during the depolarization and repolarization phases of a
cardiac cycle can be calculated. Forexample, the electrical axis of ventricular
depolarization, ,~QRS,represents the a~,erage of the instantaneous cardiac vectors
during ventricular depolarization. The ~Ql(S,Usually lies between aVL and aVF
in Figure 4.2. It is easy to diagnose left and right axis deviation, LADand RAD
respectively. In LAD,lead I is predominantly positive (i.e, R waveis positive)
and both leads II and III are predominantly negative (i.e. R wave is small or
absent). Both II and III must be predominantly negative, i.e. if in lead II the
°
-90
Normal range
~
of QRSa~
RAD
wave is smaller than the R wave, LADis not present. If lead II is equiphasic (R
and S waves have equal magnitudes), then there is borderline LAD.In RAD,lead
I is predominantly negative and both II and III are predominantly positive
(Bennett, 1989).
Figure 4.3 shows the Einthoven triangle superimposedon the locus of points
formed~in the frontal plane by a normal instantaneous cardiac vector, the
vectorcardiographic loop during one cardiac cycle. The measured ECGon each
lead is a projection of the instantaneous cardiac vector. The P loop corresponding
to atrial contraction, projects onto leads I, II, and III as an upwarddeflected
wave. However, the S wave is projected onto lead III remarkably more than~.in
leads I and II. The instantaneous orientation of the cardiac vector, and the
orientation of the lead determine whether there is a positive going or negative
going waveform on the lead. Thus the different leads of the 12-lead ECGshow
various projections of the phases in the cardiac cycle~
.,\ /
¯ QRS
e
" T
~~ "~~’ P ’
Time (II
I~ ~1 Time
S’T
segment
I
! interval ’ interval___~.i
0.28-0.43
~.4.4 Thefiducial points on a typical ECG provide diagnostic informationsuch as the QRS
wi~Ith,for evalttatingthe pacingandconduction
phasesof the heart.
¯ " TP
Sensitivity of a criterion (%) - TP + FN (100)
TN
~.~Specifieity of a criterion (%)~ ;I’N + FP (100)
t by using
of~the criteria.
DIAGNOSING ARRHYTHMIAS ~ 71
heart rate with different activities. It can also be useful to detect any malfunctions
in the pacemaker pacing or sensing functions. Current ambulatory monitors
include a separate channel for recording the pacing stimulus, and can
automatically~ provide information concerning failure to capture, failure to sense,
failure to generate an impulse, and percent of beats paced. Even though these
features are "currently reliable for single-chamber pacemakers, additional work
needs to be done for the automatic evaluation of dual-chamber pacemakers
analysis (Greenspon and Waxman,1992).
without heart failure, exercise stress tests are important in that ECGchanges
provide the information about the course of the disease. Exercise testing has
limitations as a diagnostic method. Besides the false result possibility mentioned
earlier, death, ventricular fibrillation, myocardial .infarction, or serious
arrhythmias could occur during the exercise (Luna, 1993).
Over the past few years there has be4man increased trend toward processing, of
electrocardiogram (ECG) data by microcomputers. Many systems have been
designed .and implementedto perform, signal processing tasks such as 12-lead off-
line and real time ECGanalysis, Holter tape analysis, and real time patient
monitoring. All of these applications.require an accurate detection of the QRS
segment of the ECG.A few of the techniques that have been developed to process
the ECGinclude QRScomplex detection, signal averaging, and S-T-segment
analysis.
QRS detection
Signal-averaging
Life-threatening ventricular arrhythmias; ~following myocardial infarction are
most often due-to re-entry of aslow and fractionated wavefront in the ventricle.
Studies performed on electrograms from the border,zone between ~normal and
infarcted, myocardiumshow,that these,, areas ~depolarize ~late, reflecting :slow
conduetion:i’~.Thus abnormal, :~fragmented electrical acti~vity is due to the slow
inhomogeneousconduction in this,zone. These fragmented electrical signals were
recordedin patients with sustained;ventricular tachycardia. These eltetrical
signals, the late potentials, occur, after the QRScomplexor,in the S-T segmentof
th~ surface ECGas low-amplitude, high-frequency signals. Signal averaging of
these late potentials.
involves selecting anormal cardiac:cycle as a ,,template,
comparing subsequent waveforms with the template, and averaging subsequent
waveforms which are normal., As more beats are-averaged the noise in theECG
is minimized. After signal averaging~ the averaged waveformis high,pass filtered
74 DESIGN OF CARDIAC PACEMAKERS
at about 40 Hz to remove, theS-T segment and the T wave, and thus the late
potentials are enhanced.
Computer processing of the signal-averaged ECGinvolves computing the
QRS duration, mean square root voltage of the last 40 ms (MSR), and the
duration of the low-amplitude signal (LAS) (amplitude less than 40/,tV). A
duration greater than 110 ms, MSRless than 20 ~tV, and LASgreater than 38 ms
is considered a positive test for late potentials (Macfarlaneand Lawrie,’: 1989).
Frequency analysis is also performed to characterize the ECGnear the QRS
complex. Studies have ~dem0nstrated an increase in high-frequency components,
20 to 50 Hz, of the signals from ventricular’taChycardia patients comparedto that
from normal patients. Frequency-domain techniques may demonstrate the
preSer~ce of late potentials in situationg Whentime-domain techniques are not
usefulsuch as when bundle:~branch block is present (Macfarlane and Lawfie,
1989; Luna, 1993).
S-T segment analysis
Tompkins (1993) reports on techniques used in biomedical digital~ signal
processing. It is important to design signal processing algorithms to make
measurementsof any changes.in the S-~T segment.
Weisner et al. (1982).designed a microprocessor-based device for analyzing
the S~-T segment,-Anyeffieient ~technique is used to detect the approximate
location Of the QRS:waveform.~The Rwaveis then defined as the maximal, value
within 60 ms of the ~QRSdetection mark,. The Q waveis the first inflection point
before the ~RwaVe;An. inflection point, is detected by a Change in the sign of
slope, zero slope, or a significant changein ~slope. TheS ,point is located as the
first inflection point after_the R, wave,~AHanning digital-filter can be applied to
smooth noisy ECGdata before ~ealculating the slope. By searching between the P
and Q wavesfor, a 30~ms ~ interval~ of zero slope, the isoelectrie line of the ECG
can be located ands.measured. Measurements of;the QRS duration, R peak
magnitude relative, to the isoelectrieline,-and the R-R interval can then .be
obtained. .........
The ~T point is~.defmed at the onset~of the T wave. The J point is the first
inflection point after the S point, or maybe the S point itself in certain ECG
waveforms. The T point is found by first locating the T wave peak (the maximal
absolute value, relative to the isoelectric line, between J + 180 ms and R +
400 ms). If a noisy signal does not permit the T point to be found, the upper
search limit sample point is used. The T point is then found by detecting a 35-ms
period on the R side of theT, wave which.has values within one sample anit of
each other: If theT point cannot,be detected it defaults to J + 120 ms,
With these EI2G features; S-T segment measurements can: be ~made by a
windowedsearch method~ Two-boundaries, the J ÷ 20 msand theT point,define
the windowlimits~The,point Of maximaldepression or elevation in the~windowis
then identified: S:T~segment levels can be expressed as :the absolute.change
relative to the isoelectrie line, The S-T slope is defined as the amplitude
difference between:,~the S-T-segment point ’and the T point, divided by the
corresponding time interval. The~S-T area is~also ~caleulated ~by summing,.all
sample values between the J and T points.’after subtracting the~ isoelectric.line
from eaehpoim.~ A S-T~,index is ,calculated as the~sum of the S-~T segment and
one-tenth of~the S-T sl0pe.. -~ ,~ ~
: ~,:.~Based on,the S-T,segment measurements obtainedin the above analysis;
changesin the’blood supply to themyocardiumcan be analyzed~
DIAGNOSING ARRHYTHMIAS- ~ 75
Techniques for recording from and stimulating specific sites within the r heart
have been developed. The ability to record the electrical activity from the sinus
node, His-bundle, and fight ventricle, and .perform endocardial cardiac mapping
and programmedelectrical stimulation have improved remarkably. Intracardiac
electrocardiographic techniques are usefutin the diagnosis :and the management
of arrhythmias. Diagnosis based on the surface ECG has been greatly
supplemented because of the correlation with data available from intracavitary
studies.
Catheters
with the rest. The resultant waveform then includes the His-Purkinje related
signal (Macfarlane and Lawrie, 1989).
Utilizing an esophageal pill electrode the electrogram from the atrium can be
recorded and atrial pacing can also be performed. Burack and Furman (1969)
reported a case study in which a patient was paced in an asynchronousf’Lxed,rate
pacing mode for 36.h, and then paced in the demandpacing mode for 24 h. A
transvenous~ electrode, 58 cm long, was passed through the nose into the
esophagus, and was placed posterior to the left ventricle about 4 em above the
esophageal-gastric junction,. This technique can also be used :in the case of a
patient who cannot perform the exercise stress test protocol.. The esophageal
electrode is used to pace the left atrium and increase the heart rate (Heger et al.,
1993).
In order to study the factors that initiate, terminate and sustain Ventricular
tachycardias. (VT)~it is importar~t to reproduce these as reliably a~ possible.
Using programmed electrical stimulation (PES), it is possible to itiitiate clinical
tachycardia i~ patients whohave the arrhythmia spontaneously..The therapy for
the induced atrhythmia depends on the medical history of the pafient~
PES was initially used to study the mechanismsof elinica!. - arrh~as. For
example, re,entry was found to be the operative meeh~ism.of::~ecurrent
Sustained ventricular tachycardia. The activation pattern of the arrhythmias can
be mappedusing end0cardial and epicardial catheter ,re~ordingg;.~d the site Of
origin of the arrhythmia identified: Further studies, in the area demonstrate that
PEScan be used to select antiarrhythmic drugs for therapy, but this is not often
reliable. The ability to induce the clinical arrhythmia allows the activation
sequence during the VT to be recorded and analyzed, and the influence of drugs
on the VT to be studied (Greenspon and Waxman,1992)/~
PESinvolves stimulation of the heart muscle, usually at the right ventricle
apex. S’tudies in the literature showthat performing PESat a seeondSite in the
right ventricle might help to induce VT. The North American Society~of Pacing
and Electrophysiology recommendsa protocol involving a pulse duration of 1.5
tO 2 ms, at twice the diastolic threshold; with up to three extrastimnli (Greenspon
and Waxman,1992),.
Different ventricular arrhythmias can result from PES. :The-induced
arrhythmias are classified according to their duration and morphology.. A
sustained (greater than 30 s) VT is one witha rhythm faster than 100 bpm.
monomorphicVT is one with a constant morphology, whereas-a pblymorphie VT
is one where the morphology changes frequently. Monomoi’phie arrhythmias
suggest that there is only one arrhythmie focus that has a stable entrance °and exit
site.
PESis increasingly being Used in clinical~studies.~Based on PESstudies
patients are selected for antitachycardia ~pacing, automatic iraplantable
defibrillators, and surgical ablative techniques.
The SA node, under normal conditions, acts as the .heart’s natural pacemaker by
generating impulses at a rate of about 60 to 100 times/min in most adults at rest.
78 DESIGN OF CARDIAC PACEMAKERS
Diagnostic criteria
Type HR (bpm) P-R (ms) Morphology Particulars
Brad. < 60 N ’~ ~ P:N; QRS:N " Low HR
Tach. > 100 Sh/Lo P:N; QRS:N
Block N, varies N p:dropped . PP: (shortens beat
~
.... 2* ~ ~ sometimes dropped)i(related
In sinus bradycardia the~heart rate is less than60 bpni. This heart rate is .the
distinguishing criteria from normal sinus rhythm, The impulse is generated in the
sinus node, and is propagated through the horror conduction system of the heart.
Sinus bradycardia .may be seen in healthy adults who.have~exercised
regularly during their life and thus have increased cardiac muscle tone. Increased
parasympathetic activity., certain .types of medication, and organic disease of the
heart also result in sinus bradycardia.
The distinguishing feature from the surface ECGis the lower heart rate. The
morphology, of~the different components in the ECGare the same .as.that for
normal
a. sinus rh~. Figure 4.8 shows, an ECGtracing of sinus bradycardi
Diagnostic:criteria: ,The rate is less than ,60 bpm; rhythm is regular;
interval is normal; all P waveshave the same shape and are upright in leads I, II,
aVF, ~and V4 to V6: QRScomplex is normal,, but may be different due to other
abnormalities (Conover, 1986).
Because of the lower heart rate, the cardiac output may decrease and this
results in a lower supply of blood to the brain and other organs of the body. If
the heart rate is not very low in this arrhythmia, then the patient will not have
any major symptoms. However, if the~:heart rate isvery low, s~ptoms .that are
commonto marked low cardiac output are observed. These symptoms include
dizziness, weakness, chest pain and pressure, shortness of breath, low blood
pressure, and lower pulses detectable at the;limbs.
DIAGNOSING ARRHYTHM|AS~ ~ 79
, If the, symptomsare persistent and the patient cannot tolerate them; then
tre~ent is warranted. Medication is reduced if it caused .the bradycardia, or
~nisteled to. increase heart rate. A tempo~a,~ pacemaker ’is often ~sed when
there has been damage to the cardiac muscle such as with acute myocardial
iiifarctipn, if the symptomsdo not disappear then a ¯permanent’ pacemaker is
required (Catalano, i993)..~ ’
Treatment is required to reduce the heart rate until good cardiac output is
obtained. The cause of the tachycardia (fever, anxiety, exercise) should
alleviated.
4.4.3 SA block
1" SAblock
2" SAblock
Diagnostic criteria: The rate is normal but it varies because of pauses; the
rhythm is irregular; P-R interval is normal; QRScomplex is normal but may
change as a .result of other pathologies; in type I block, P-P intervals shorten
tmtil,a~P waveis dropped and the resulting pauses are twice the shortest cardiac
cycle; in type II block the P-P intervals are a multiple of the cardiac cycle
(Conover, 1986; Chou, 1986).
Patients do not have any symptomsif the blocks are transient. In more
persistent cases patients complain of skipped beats. If the pauses between two
heartbeats are3 s or more, patients maybegin to feel dizzy, weak, chest pressure
or pain; or even have a transient syncope. If the patient had a recent myocardial
pathology then the patient should be monitored (Catalano, 1993).
Treatment is warranted in situations when the blocks are severe and
persistent and there is hemodynamic compromise. If medication caused the
arrhythmia then the dosage is reduced. Atropine IV is used to stimulate the SA
node and to improve the conduction through the atria. If the block is caused by
myocardial infarction then a temporary pacemakeris used until injury due to the
infarction is healed. If the sinus block appears to be a permanent condition, and
symptoms attributable to bradycardia are persistent, then an. implantable
pacemaker is mandated (Catalano, 1993). Studies in the literature. report that
sudden death dug to :bradyarrhythmia were-more related to depressed
automaticity and (or) SA block, than to AVblock.(Luna, 1993),
Sinus~ arrest occurs whenthe SAnode-ceases activity and .does not generate the
impulses necessary to maintain a minimal rate. In this,, situation subsidiary
pacemakers within the natural pacing system (AV junction or ventricles) take
over and result in a heart beat, i.e. an escape rhythmis generated.
Since the sinus node does not generate an impulse in: this abnormality, the
surface ECGdoes not manifest a P wave, or. P-R interval, .There is a long pause
in,the ECGuntil a subsidiary pacemakergenerates an impulse that causes a heart
beat.A, slower rhythm is then generated by the escape rhythm from a pacemaker
in ’the AVjunction or ventricles. Sinus arrest should be differentiated from SA
block. Pauses in sinus node activity ofup.to 2.0 s due to high vagal tone maybe
found infit, young people (Bennett, 1989). Figure 4.10 shows a prolonged period
of sinus arrest.
Figure 4.10 The ECGshows periods of sinus arrest. AVblock is also present. From Chou, T.
C. 1986. Electrocardiography in clinicalpractice. 2nd Ed. Grune &Stratton.
Diagnostic criteria: The rate is normal, but it varies due to the’pauses; the
rhythm is irregular; all P waves, when present have the same shape; QRS
complex is normal unless other pathologies are present; there is an abrupt
interruption of the sinus rhythm. The P-P interval betweenthe pause has no fixed
relationship to the normal P-P interval (Chou, 1986; Conover, 1986).
Some patients may have no symptoms and very little hemodynamic
compromisedue to sinus arrest. However,there is the potential for sinus arrest to
develop into asystole (no activity in the cardiac ~muscle). Thus these patients
should be monitored if not treated. In other patients who manifest symptoms,
dosage of medication which may have induced the arrhythmia should be reduced.
Medication to stimulate the sinus node is administered to increase SAactivity. If
the sinus arrest is caused by cardiac disease, such as that of the coronary arteries
or acute infection;, then a permanent demandpacemaker is implanted (Catalano,
1993; Conover, 1986).
On the surface ECG.the sick sinus syndrome (SSS), manifests itself as severe
sinus bradycardia, sinus arrest, SA block, bradycardia,~alternating with
tachycardia i.e. tac-hy-brady syndrome,chronic atrial fibrillation whichcontinues
even after cardioversion, and AVjunctional escape rhythm (Chou, 1986).
junctional escape rhythm results whena pacemaker in the His bundle triggers a
heart beat if the SAimpulse is abnormally impeded.
Twenty-four-hour ambulatory ECGrecordings are often used to obtain
diagnostic information of episodes of bradycardia or tachycardia. If symptoms
and rhythm disturbances are infrequent, then intracardiac electrophysiological
testing maybe helpful (Bennett, 1989). Sinus bradyeardia and tachycardia are
normal during sleep and physical exertion, respectively:
Diagnostic criteria: The rate is either fast or too slow; the rhythm is
irregular; the-P-R interval may be abnormal; the, QRScomplex is~ normal
~although other pathologies mayprolong it; distinguishing features mayinclude
severe sinus bradycardia, sinus arrest, atrial standstill,, SA block, sinus
bradycardia with recurring atrial fibrillation, bradycardia-tachycardia, transient
asystole following tachycardia or cardioversion, paroxysmalatrial fibrillation, or
inability of sinus node to accelerate in respOnseto, exercise (Conover,1986).
Intracavitary electrophysiologic studies of the sinus node ’function and
sinoatrial conduction time are performed to diagnose and select therapy for
patients. The sinus node canbe suppressed if artificially stimulated at rates higher
than its intrinsic rate. The sinus node recoverytime is the time between the last
artificial pacing of the sinus node and the first spontaneous firing of the sinus
node. The sinoatrial cor~duction time can be measured directi~using the sinus
node electrogram, which records the electrical activity close to the sinus node,
and the atrial electrogram which records atrial activity (Macfarlane and Lawrie,
1989). Thb criteria of sinus recovery time and-sinoatrial conduction time are
sensitive, but not specific to SSS. Morethan 40%of pacemakers implanted are
due to SSS (Luna, 1993).
Patients with SSS have symptoms, and are treated, according to the
particular type of the arrhythmia.- Medicationsthat increase the heart rate (for
bradycardiaCondition) would tend to increase the heart rate of the tachycardia
condition in a tachy-brady syndrome. For ~tacliy-’brady syndrome, an e~fective
tre:atment is to administer cardiac depressant medications that control the
DIAGNOSING ARRHYTHMIAS 83
Diagnostic criteria
Type HR (bpm) P-R(s) Morpholo[y Particulars
1". N > 0.2 P: N; QRS:N i P followedby QRS
2",i r, ’ N Increases P: N/blocked R-R with block is
until QRS: N shorter than twice P-
locled ¯ P
(a)
Ill
(b)
(c)
’,
:::
Figure 4.12 (a) First-degree AVblock. (b) Type I second-degree AVblock with the
Wenckebachphenomenon. (c) TypeHsecond-degreeAVblock. (d) Third-degreeAVblock.
Chou.T. C. 1986.Electrocardiography
in clinicalpractice.2ndEd. Grtme&
Stratton.
Diagnostic criteria: The rate depends on the SA node and is normal; rhythm
is regular; QRScomplex is normal; the P-R interval is greater than the normal
value of 200 ms; each P wave is followed by a QRScomplex.
First-degree AVblock by itself does not cause patients to have any
symptoms,and is not usually treated (Catalano, 1993). Howeverif the arrhythmia
is associated with myocardial infarction the situation could progress to a higher
degree block. The patient needs to be monitored in these situations. If the patient
has persistent symptoms(due to a coexisting abnormality) then medication
prescribed. If the H-Vinterval is between 55 ms and 75 ms, a pacemaker should
be used on patients with persistent symptoms.If the H-Vinterval is greater than
75 ms, a pacemaker must be used (Macfarlane and Lawrie, 1989).
Type II second degree AVblock (Mobitz II) is less commonthan type I and
usually involves abnormal conduction in the His bundle or the bundle branches.
In ~second degree block there is intermittent failure of conduction of the
impulses to the ventricles, and this is manifested on the surface ECGas P waves
without its associated QRScomplex. The abnormality lies in the His bundle or
bundle branches as comparedto in the AVnode as with fn’st-degree AVblock. If
the block is in the bundle branches the QRScomplexes are broad (Bennett 1989;
Chou, 1986). Figure 4.12(c) shows an ECGtrace manifesting type II second,
degree AVblock.
A block can be .in the His bundle itself, in which case the QRScomplex~is
narrow. -The HBEthen shows the A and H waves following the P wave.of the
ECG~It is also possible to have conduction abnormalities within the His bundle
(Helfant~~1974).
Diagnostic criteria: The rate is normal; the rhythmis irregular; P-R interval
is normal (120 - 200r ms), fixed and consistent, but some P waves are-not
86 DESIGN OF CARDIAC PACEMAKERS
conduction through the AVnode and reduce the degree of the block. If the block
is transient a, temporary pacemaker is prescribed, otherwise a permanent
pacemaker is warranted (Catalano, 1993; Conover, :1986).
The .,el~ctric~ impulse normally, generated at the SAnode, travels through the AV
~o.de, ~His bundle, and then through.th e fight and left bundle branches, which
sp~ad~0ut into the Purkinje system.. The left bundle branch divides into the
~a~rior ~ ~and superior, subdivisions called fascicles. If these pathways behave
no~ail~, i~e fight and left ventricles are activated at the same time, resulting in
a Synchronous contraction of both ventricles. During right (or left) bundle
branch block, RBBB(or LBBB),there is an abnormal conduction of the impulse
through the fight (or left) bundle. The impulse can also be blocked in the anterior
or (and) posterior fascicle resulting.in anterior hemiblock, posterior hemiblock,
bifascicular, or trifascicu!~ bl0~k. Figure 4.13 describes someof the features of
the surfaceECGUsed in di~ignosing bundle branch block.
T~e,
N .: QRS>120ms R’ in right: PCL
Broad S
aVL,V6.~Delay in
¯ deflection.in right
PCL > 50 ms ¯
LBBB N QRS>120 ms Broad, notched, R in
NoQ:~ I, V6 I, V5, V6. Broad,
negative QRSin ~V6.
N QRS: (N/> LAD of QRS axis
by 20 ms) .
QRS: (N/> ;RAD of QRS axis
by 20 ms)
being directed anteriorly and rightward. The cardiac vector during the delaye~
RVcontraction phase is unopposed by left ventricular forces. Thus a prominen
secondary R wave, R’, in right precordial leads (V1 and V2) is recorded. The let
precordial leads (V5 and V6) as well as lead I record a slurred S wave. The
wave is slurred because of the slower conduction through the cardiac muscl~
during the abnormal spread of the impulse to the RV. V1 is usually the mos
important lead for diagnosis of RBBB(Chou, 1986; Bennett, 89). Figure 4.14(a
shows an ECGpattern manifesting R.BBB.
Diagnostic Criteria: A QRSwidth greater than 120 ms; ~an R’ in righ
precordial leads with rsR’ or rSR’ morphology; delay in deflection in righ
precordial leads greater than 50 ms; and wide S wavein I, V5, and V6; the rate it
normal; rhythm is normal; P-R interval is normal; in leads I, aVL, and V6 thert
is a broad S wave; or secondary T-wave morphology changes (Chou~ 1986
V
3 V
4
have lesions which result in various degrees of heart block, i.e. intra-His bundle
block. These are manifested as a split H wave, H and H’, separated by 10 to 20 ms
or more (Helfant, 1974).
Patients with LBBBhave no symptoms that are associated with the block
itself. If the arrhythmia has developed recently in association with acute
myocardial infarction, there is an increased probability of complete block and a
higher mortality rate, and these patients should be monitored closely and
prescribed a temporary pacemaker (Catalano, 1993; Conover, 1986).
4.6.3 Left anterior hemiblock
The anterior division of the left bundle branch conducts impulses to the
anterosuperior (frontwards and upwards) region of the LV, and a block in this
pathway results in left anterior hemiblock (LAH). LAHresults in a delay in the
activation of the anterosuperior portion of the LV, but activation of the
posteroinferior portion (caused by the posterior division of the left bundle)
unaffected.
In LAH,the impulse initially spreads inferiorly through the posterior
division of the left bundlebrancli.. ~his results in an initial positive deflection in
infedor!y oriented leads (II, III, and aVF)and in an initial negative deflection
the lateral leads (I and aVL)..The excitation then begins to spread from the
posterior region of the LV t0"~the anterosuperior region, The cardiac vector
during this phase Will thus be superiorly directed resulting in an R wavein I and
aVL~mdan S wavein leads II, liI, and ~VF. The forces during this phase are also
very prominent and result in the meanQRSto be shifted leftward and superiorly,
i.e. a left axis deviation.
The conduction through Ordinary, slow conducting myocardiumrather than
specialized conducting tissues results in delayed activation of the anterosupedor
region and consequently unopposed electrical activity from the rest of the
ventdcles~ Thus the superiorly directed waveis larger than the initial inferiorly
directed waves. To diagnose LAH,two criteria must be satisfied: there must be
left axis deviation, and the inidal direction of ventricular activation must be
inferior and to the right, i.e. there must be an initial r wavein leads II, III, and
aVF. Figure 4.15(a) ~hows an ECGmorphology during LAH.
Dh~gnostic criterM: The rhte is normal; rhythm is normal;. P-R interval is
normal; QRScomplex is normal (or lengthens by 20 ms); left axis deviation
the QRSaxis in the frontal plane between -30* and -90*; qR or R waves in leads
I and AVL; rS complex in leads II, HI, and aYF; terminal R wave in aVRand
aVL; increased QRSvoltage in the limb leads (Chou, 1986; Conover, 1986).
In left posterior hemiblock, LPH,the posterior division of the left bundle branch,
which conducts impulses to the posteroinfedor regions of the LV, is blocked.
The situation with LPHis opposite to that with LAH.The activation in the
posteroinferior region of the LVis delayed. The electrical forces during the
delayed activation results in a cardiac vector directed inferiorly and rightward.
The delayed forces, as in LAH,are unopposed, and this results in a rightward
displacement of the mean QRSaxis, i.e. right axis deviation RAD.Since the
anterolateral region of the LVis activated early there is an initial R wavein leads
I and aVL and a q wave in the inferior leads (II, III, and aVF) (Chou, 1986;
Bennett, 1989). Figure 4.15(b) shows the ECGmorphology during LPH.
DIAGNOSING ARRHYTHMIAS 91
(a) V
1 V
2 V
3 V
4 V
5 V
6
(a)
V
I V
2 V
3 V
4 V
5 V
6
If conduction is blocked in only two of the three fascicles (among the right
bundle’ branch, anterior fascicle, and posterior fascicle), tile functional fascicle
.will cbnduct ~itrial impulses to the ventricles and maintain sinus rhythm. A block
m the remaining third fascicle will lead to complete AVblock.
92 DESIGN OF CARDIAC PACEMAKERS
tO be determined.: The long P-R interval maybe due to a proximal lesion of the
bundle of His. Studies of the His bundle electrogram are needed to determine if it
is a true :trifascicular block (Luna, 93; Chou, 1986). For example, Figure 4.16
shows: how, the His bundle electrogram is used to differentiate in the case of first
degree heart block associated with RBBBor LBBBand determine if the
conduction abnormality lies in the AVnode or the His-Purkinje system. Figure
4.16 shows situations of two patients with a prolonged P-R interval. Patient 1 has
an abnormal His-Purkinje conduction time (H-Q = 92 ms), and a normal P-H
time (115 ms). Patient 2 has a normal H-Q~time (39 ms) with an abnormal
nodal conduction time (P-H = 175 ms) (Helfant, 1974).
Preexcitation arrhythmias result when the impulses ~0f the SA node reach the
ventricles earlier than ~normal. The enhanced conduction may be due to
conductive muscle (beside the normal conduction pathway) anywherebetween the
atria, and.ventricles (Wolff-Parkinson-White.syndrome), a seg, .mem:. -ofthe-AV
node which conducts the impulse faster than usual (Lown-Ganong-Levine
syndrome), or Mahaim fibers which are abnormal pathways.between the AV
node~ His bun~e, or bundle branches and the ventricles.
on the fight side of the heart conducts impulses prematurely to the right
ventricle. The fight ventricle thus depolarizes before the left ventricle (similar to
a left bundle branch block). The delta wave and the QRSare negative in V 1 and
V2 but are upright in the left precordial leads (Chou, 1986; Catalano, 1993).
Figure 4.17 shows the ECGmanifesting type B WPW syndrome.
Figure 4.17 Type B WPWsyndromeresembles the morphologyof left bundle branch block.
FromChou,T. C. 1986.Electrocardiography
in clinicalpractice.2ndEd. Grune&Stratton.
The delay of impulses in the AVnode serves to protect the ventricles from
rapid electrical activity due to atrial fibrillation (350 to 600 bpm). The bundle
Kent howeverprovides a route for high rate impulses to activate the ventricles.
Thus patients with atrial fibfillati0n and preexcitation can have a ventdcular
response, thatean lead toheart failure or shock and ventricular fibrillation.
If an ectopic beat in the atria occurs~dufing-normal Sinus rhythm~reenla’y
tachycardia cotdd ’.result due to the accessory pathway and AVn6de~ having
different refractory periods. This arrhythmia is an example of paroxysmal
supraventricular tachycardia; PSVT, and is manifested on the surface ECGas
narrow ventricular complexes in rapid, regular succession. There are no delta
wavesin this situation. Treatment of this situation is directed at incre~a~ng the
refractory period of the AVnode and includes medication and cardioversion, the
latter resulting in the refractoriness of the entire myocardium.
Somepatients"with PSVTdo not present any evidence of pre~xcitation on the
surface ECi3.:In thesesituations inVasive~ electr~physiologic studies mayreveal a
concealed bundle of Kent~ This accessory path~ does no~ conduct impulses
from the SAnode either because of its distance fro~nthe SAnode, the length of
the bundle; ~ direction of conduction possible, o~’~onducti~An~timeinvolved.
Hbwever,impulses :from the ventdcles ~an~ be~con~lUctedretr0grad~ to the atria,
throug h the concealed pathway. Thus there is no delta wave or ~P.-qR interval
shotterling in the normal sinus rhythmof the,an-hythmia. A patient with a normal
ECGmorphology during sinus rhythm¢ and an inverted P wave between QRS
complexes during tachycardia, should be suspected for concealed preexcitation
(Bennett, 1989).
inti:acavitary electrophysiologi~ studies-that include pacing the atria at
different rates~’~fid recording theletect~o~i!am help to 10~ai~ize the ~site’ of the
anomalous pathway (Luna, 1993~ Maefaflane and Lawrie; 1989)? Right atrial
pacing results ~n greater Prc~citation for ~ type B~Wsyndrorn6 tha~.,ifor ~ type
AW-PW. Similarly it is possible to see inore preexcitation in patients with type A
DIAGNOSING ARRHYTHMIAS 95
WPW syndromeif the left atrium is paced, suggesting that the accessory pathway
is on the left. Electrogramsfrom the left ventricle record activity coinciding with
the delta wave on:the ECGfor type. A WPW syndrome. Recording activity from
the right and left atria simultaneously helps determine the location of the
retrograde ventricle-atria conduction in patients with WPW and supraventricular
tachycardia (Chou, 1986).
,Diagnostic criteria for typical WPW syndrome: the:.rate is normal;, the
rhythm is regular; P-R interval is less than 120 ms; QRScomplex width is
greater than 110 ms; distinguishing features include a short P~-R, broadQRS,
delta wave, and a tendency towards PSVT, and atrial fibrillation or flutter;
Secondary T waves may be present.
Diagnostic criteria for concealed WPW syndrome: the~rate is normal; the
rhythm is regular; P-R interval normal; QRScomplex normal; the distinguishing
features include normal P-R, normal QRS, and a tendency to PSVTand atrial
fibrillation or flutter.
WPW by itself does not produce any symptoms on patients. However, if
supraventricular tachycardia is present with rapid rates,then symptomspresent
are due to low cardiac output. The patient mayfeel weak; dizzy, palpitations in
the chest, pressure or pain in thechest, or shortness of breath (Catalano; 1993).
Treatment for induced PSVTwould involve use of vagal maneuver, and if
this is unsuccessful, cardioversion or medication is administered based on the
hemodynamicstatus of the patient. Treatment of atrial fibrillation and a rapid
~,entricular rate:would include cardioversion and avoiding digitalis (Conover,
1986).
Vl V
2 V
3
.......... : ......... ~.
Diagnostic criteria: the rate is ~normal; the rhythm .is regular; the P-R
interval is < 120 ms; QRScomplex normal; distinguishing features include a
short P-R, normal QRS,and tendency to PSVTand atrial fibrillation or flutter;
no delta waveis present.
The patient does not have any symptoms due to LGL syndrome itself.
However, if LGLsyndrome is associated with supraventricular tachycardia,
symptomsof low. cardiac output are present. Treatment is crucial in ~ the case of
associated atrial arrhythmias or supraventricular tachycardia. LGLsyndrome is
treated well with cardiac medications and surgical intervention is dangerous and
not often effective (Catalano, 1993).
Mahaimfibers from ,the base of the AVnode, or top of the His bundle, to the
ventricles conduct impulses originating from the SA node and traveling via the
AVnode directly to the ventricles.
These Mahaim fibers may :originate in the AVnode, a nodoventricular
connection; or in the~.,His bundle or bundle branches, a fasciculoventricular
connection. The impulse from the SA node is thus delayed normally at the AV
node and then .directly travels.to, the base of the ventricles via the Mahaim
fibers.
The surface ECGthus manifests anormal P-Rinterval and the delta wave due to
slow.conduction via the accessory pathway.
Diagnostic criteria: The rate is normal; the rhythm is regular; P-R interval
may be short or normal; QRScomplex varies in morphology; distinguishing
features include a tendency to PSVTand atrial fibrillation or flutter. The ECG
morphologydepends on the origin and insertion of the extra fiber, as well as:its
length.
Treatment is same as for the situation of supraventricular tachycardia with
WPWsyndrome.
The premature ectopic beat may occur very early in the cardiac cycle and
overlap with the T wave of the preceding cycle or it mayoccur just before the
normal sinus P wave and result in a fused beat. Ectopic beats generated from the
same ectopic site have intervals between them that are approximately the same.
Most PACsare followed by an incomplete compensatory pause. The interval
containing the PACis shorter than twice the basic sinus interval because the
ectopic beat depolarizes and resets the SA node. Figure 4.19 shows an ECG
tracing of the PACabnormality.
Diagnostic ~criteria: The rate is that of the intrinsic rhythm; the rhythmis
irregular because of the PAC; the P-R interval is normal; the P waves are
up~gtit and ha#e the same shape in leads I, II, aVF, and V4 to V6; the P’-R
interval maybe no~, longer, or shorter than the P-R interval (where P’ is the
atrial depolarization due to the ectopic beat); the P’ maynot be conducted to the
ventricles; distinguishing features include a normal QRSand a P’ wave that is
premature and has a different morphology than the normal sinus P wave
(Catalano, 1.993; Conover,_!~86). -
Most patientswitth~this abnormality do not havelany symptoms.Howeverfew
patients mayfeel skipped beats if the PACsare frequent. The cardiac output is
not significantly compromisedbecause the ventricles contract normally. The
abnor~aality is t~eated’:~ith ~medicationif the PACsare frequent:
Whena single area of the atria discharges spontaneously at rates greater than
250/min,then atrial flutter results.
Atrial flutter is manifested on the Surface ECGas rapid and regular F waves,
i.e. a sawtooth morphology.This pattern is best Seen in .leads II, III, aVF, and
V1. In most situations, the F-waveaxis is directed superiorly, andthe F waves
are thus inverted in leads II, III, and aVF. The downwarddeflection of the F
wave has a small slope, while that of the upward deflection has a higher slope,
The F waves may not be noticeable on leads perpendicular to the F whve axis,
e.g. lead I. Variations of these criteria are possible (Chou, 1986).
The ventricles cannot contract at rates as high as that possible in the atria.
The AVnode helps prevent high rate impulses of the atria stimulating the
ventricles by blocking someof the excessive impulses generated in the atria in
ratios such as 2: 1, (i.e. one out of every two atriallimpuises are c~nducted)~~3:
and 4:1. Figure 4.20 shows an ECGpattern manifesting atrial flutter.
Diagnostic criteria: The atrial rate is 250-300 bpm; and that of the QRS
complexes 125-150 bpm; the QRScomplex is normal; theT’-R intervalis 260-
460 ms, where P’ is a nonconductedP wave; distinguishing characteristics .include
a sawtoothpattern in leads II, III, and aVFi a sharp, positive P’ in V1(Catalano,
1993; Conover, 1986).
Most patients do not have any~ or have very insignificant/symptoms, The
symptomsinclude a rapid fluttering feeling, palpitations ~ in the chest, or-~the
feeling of skipped beatsJ If the arrhythmia has just d6veloped¢then~ medication is
used to convert it to normal sinus rhythm. If medications are:trot effeCti:ve then
cardioversion is administered. If the arrhythmia’ is persistent due ~to~a cardiac
disease, ~then medication is usually prescribed to keep the ventricular rate,below
100 bpm (Catalano, 1993):
This arrhythmia maybe caused by multiple reentry circuits within the atria
and results in the atria not filling up during the cardiac cycle. All the impulses
from the atria, generated at a rate of 300-650/min, are not conducted to the
ventricles because the AVnode cannot conduct during its recovery period. Thus
the ventricles have an irregular depolarizing rate.
The rapid atrial activity is manifested on the ECGby ’f’ waves, which are
irregular in amplitude, and frequency. These waves are not detectable in all leads
but are best .seen in lead V1.. If these wavesare not detectable, then the irregular
ventricular depolarizing rate is indicative of atrial fibrillation: Atrial fibrillation
can exist with other abnormalities such as WPW syndrome and bundle branch
blocks, and this will result in additional changes to the morphologyof the ECG
(Chou~1986). Figure 4.21. shows the ECGmorphologyduring atrial fibrillation.
4.9 V NXaICULAR
i CXOI’ICS
A ventricular ectopic beat results whenan area of the ventricle captures a heart
beat due to enhanced activation, reent ~, ~r naturally. Someof the ventdcular
ectopics arrhythmias are manifested on the ECGas premature ventricular
cont~aetions~.ve~tricular tachycardia and flutter, ventricular fibrillation, and
vet~ ~ "¢ular, escape:.
100 DESIGN OF CARDIAC PACEMAKERS
An area of the ventricle, below the His bundle, may discharge spontaneously
before the impulse from the sinus node has captured the heart. The result is a
premature ventricular contraction (PVC), and the basic rhythm of the cardiac
cycle is disrupted.
The ectopic beat is triggered by a location in the ventricles, and thus
conduction will be abnormal. This abnormal conduction is manifested on the
ECGas a wide QRScomplex. If the ectopic area is in~.the ~left ventricle,, the
impulse spreads towards the right ventricle, and the leads oriented ~in the
direction of the.spread of this impulse will show a positive deflection. The
repolarization phase, is also altered, resulting in changes in the S-T segment and
T wave. The S-T segment is depressed and the T wave inverted if the major QRS
deflectionis itp right,
A PVCis as u~ lly followed by a compensatorypause, i.e. the R-R interval
containing the PVCis approximately equal to twice that .of the normal sinus
rhythm. This is usually observed if the sinus node is not..affected by impulses
from the ectopic site (Chou 1:986). Figure 4.22 shows an exampletff ia PVC.
Figure.4~2--The-eompensatory
pauseis present after the prematureventricul~rContraction,The
R-Rinterval containingthe PVCis greater than the normalR-Rinterval.FromChou;
T. C. 1986.
in "clinicalpractice.2ndEd:-Cimne
Electrocardiography &-Stratton.
PVCswith the same shape and interval~ between them are ~ssumed~to arise
from the same ectopic site, The R on T phenomenon occurs when .a PVC
coincides with the T wave of the previous cycle. Manypresentations of PVCsare
possible (Chou, 1986; Catalano 1993).
Diagnostic criteria: The rate is that of the intrinsic.sinns rhythm; the rhythm
is irregular because of the ,PVC; the P-R interval is measured from the normal
beat and not the PVC; the QRScomplex of the PVCis broadand premature, and
that of the normal cardiac cycleis normal; there is no P wave associated with the
PVC;the T waveof the PVChas an opposite polarity .to ~that of. the QRS;there
are variations in the frequency and distribution of,the PVCs;there is a. full
compensatory pause following a PVC(Chou, 1986; Conover, 1986).
Most patients do not have any symptoms. However, with frequent PVCs,
patients mayfeel skipped beats or palpitations in the che~t.’lsolate ] ~ncidents of
PVCsare not dangerous, but medication is prescribed if the patient has frequent
PVCsor has myocardial infarction (Conover, 1986; Catalano, 1993).
results whenthere is an area of the ventricles that takes over as the pacemakerof
the heart.
During sustained VT, the abnormality lasts longerthan 30 s. The onset of the
VTprovides prognostic information, and thus it is important to classify the VT
according to howlong it lasts. Mechanismsfor sustained VT include reentry and
aneetopic focus (Luna, 1993).
During ventricular flutter, only one area of the ventricle is spontaneously
discharging pulses spontaneously and activating the heart. During ventricular
fltitter,.the heart beats faster-than in VT, and thus can be:differentiated by the sine
Wavem0rphology of the’ECG; The ECGis rounded on the.top and the bottom in
ventricular flutter, as compared to in VT where a sharp angular configuration
due to ventricular depolarization is noticeable, as shownin Figure 4.23.
(a)
(b)
Diagnostic.criteria for VT: The rate is 100-170 bpm; the rhythmis regular
m0~t of~the-time; the P-R interval cannot be measured; the QRS~complexis
greater than 120 ms; distinguishing characteristics include left axis deviation and
a monophasic R, QR, qR; or RS morphology in lead V1; AVdissociation may be
present; there is an abrupt onset and termination of theabnormality(Chou, 1986;
Conover~ 1986). -
Diagnostic criteria for ventricular flutter: The rate is faster than 250 bpm;
the P-R interval cannot be measured; the QRScomplex is broad aiad not
presented well; the morphologyof the ECGresembles a sine wave, i.e. without
d~finedQRS complexes; it is not possible to differentiate T ware’from QRS
complexes; tlie ventricular complex is not recognizable in this situation as
e0~pated to in VT (Catalano, 1993; Conover, 1986). * -\ -~i
" ’ Patients with VT have symptomsthat mayor not be tolerated~depending on
the h~’ rate. Patients with-heart rates of 100-125 bpmcan tolsrate~ it for along
time:They eXperience fluttering in the chest, shortness of breath,, intolerance to
activi~, low blood: pressure, and a low pulse. Treatment is based on whether the
patiei~’ean tolerate the symptoms or not..Medication may be preSCribed.
Cardi0~efsion is administered for a rapid termination of ~/T. Patients with
v~ntri~ular flutter have the. same symptomsas: those ~with VT. However,these
symptomsare not tolerated well and are often treated with a tiered implantable
cardft~erter-defibrillator. Ventricular flutter could progress to ventricular
fibrillation if not treated quickly (Catalano 1993).
I02 DESIGN OF CARDIAC PACEMAKERS
Whenthe pacemakersites in the region above the ventricles fail to trigger, or ~in
the presence of a conduction block, i there maybe an interval Of time ~hat i’s
grea~er than the intrinsic firing rate of the natural pacemakersin the v~ntdcieS,
whenthere is no ventricular depolarization. A natural pacemakerin the ventricle
theninitiates theheart to beat, resulting in a ventr~cular escape beat, (VEB)..
The impulses from the higher sites in the conduction system do not activate
the ventricles. Thus the ECGshows the VEBwithout a P wave before it..It~-i~
possible to have a P wave dissociated with the.following VEBor a retrograde one
after the VEB. The QRS complex has a morphology similar to that of a
ventricular ectopic beat. ~ Figure 4,25 shows an example of an ECGtracing with
VEBs.
Diagnostic criteria: The rate is normal ~but usually slower ~.than usual; the P-
R interval is measured from the normal cardiac, cycle because .the VEBdoes not
DIA(;NOSING ARRHYTHMIAS 103
have an associated wave before it; the QRScomplex of the escape beat is broad; a
distinguishing feature is a broad QRS complex followed by a long pause; the
escape beat has an attached T wave in a direction opposite to that of the QRS
complex (Catalano, 1993; Conover, 1986).
Figure 4,25 A blocked premature atrial contraction results in a ventriculareseape beat. From
Chou,T. C. t986. Electrocardiographyin clinical practice. 2nd Ed. Grune&Stratton.
Patients with VEBsdo not have any specific symptoms but may feel skipped
beats during the long pause before the escape beat. The cardiac output is reduced
because of the slower ventficular rate, and thus patients also have symptoms
resulting from low cardiac output. Treatment of VEBs involves increasing the
rate of the basic rhythm by prescribing medication (Catalano, 1993).
4.10 REFERENCES
Artificial Pacing
Mohammad H. Asgarian
The previous chapters described how cardiac pathology may affect the patient.
After diagnosis, the patient seeks proper therapy. The first avenue of therapy is
to use .~ ~drugs. Manyofthe abnorm]~lit]es~of the heart, can be treated bydrugs. If
drugs alone are not sufficient, then temporary(artificial) pacing is considered.
the disease is not chronic, in manycases,, acOmbinatioii of temporary pacing and
drugs-helps the patient. But if temporary pacing is inconvenient and the patient
needs, regular pacing, an (artificial) permanent pacemaker is eonsiderod. This
option involves a surgical operation and the related cost.
Implantable pacemakersimprove the quality of life and give the patients near
normal life functionality. This chapter considers general considerations for
artificial pacing. Section 5.1 reviews basic componentsof an artificial pacemaker,
the lead and pulse generator. Section 5.2 describes different types of pacing
(synchronous, asynchronous, and rate-adaptive). Section 5.3 describes temporary
pacemakers. Section 5.4 reviews indications for pacing (whether tem~rary o~
pe~anen0. Sectipn 5.5 defines the methods of designating pacemaker modes.
~ecfion 5Ji defines Several functional pacing modes and ~inally section 5.7
eons~d~ ,the criteria for selection of proper pacemakers. .
5.1~.~:ARTIFICIAL PACEMAKERUNIT
The pacemaker unit delivers an electrical pu~e with the proper intensity to the
prolTer location in, the :heart to stimulate the heart at a desired rate and~ thus
the patient with a functional heart. Figure 5.1 shows a functional
of an artificial pacemaker, which requires a pulse generator and a lead.
The p~ generator .houses electrical components responsible for generating be
pulse, .(~a-output circ~ts) at the proper time (via (timing)control circuits) base,
on events sensed (via sensing circuits). It also contains the power supply and may
include other elements such as telemetry for testability and programmability and
ROM or RAMto store data for diagnostic purposes. The lead contains a wire to
deliver the pulse to its destination in the heart and to sense and carry back
info~tion tO the sensing u~ts in the pulse generator.~ ,This ~ section~ reviews
major c~mp~entsof an arZtficial pacemaker.
5.1.1
Sensing
unit
Lead
Powersoume
Control
unit
Pulsegenerator
The power source provides.the energy required for the 0perafi0 ~ Of the’ circuitry
of a pacemaker, which includes the control, ~ensing and puls~:gener~tting units:
For implantable pacemaker~at present the p6We~s~0ur~e is usually’ a ~aemical
battery. Several other sources (such as a nuclear battery and a biological battery)
have been developed, but their use~ in pacemakersis considered, less practical.
Modempacemakers typically use batteries with lithium as the anode element and
iodine as the cathode. These form a battery that does not produce gas and can be
hermetically Sealed,to prote~t’the~battery-from body’ti~sues (Monde"1983~: ’A
main concern in Using a battery ~ is itslongevity. Longevity of a battery can’,be
determined"knoWingbattery :capacity antt~current drain. Battery
usually defined in term of ampere-hour~~while’ ei~r~ent drain isin ~ terms 6f
microamperes.The ~Ui’tSnt drain:.is~depend~nton tlae typ’e::0f electrode as Well~ as
the ciret~itry and typeofpulse getieration ofthe pacemaker: Chapter 7 discusses
power,supplies.
5.1.3 ¯ ’Sensing
,The- sensing unit ,amplifies. and filters the information.received via the electrode
and lead from activities inside the heart. To avoid attenuation.of signals,~Op~amps
with a high input impedance amplify the signal. Bandpass filtering remgves
unwanted signals. A comparator determines whether the QRSis detdetedqn 6rdet
to reset a timing circuit. Modem pacemakersalso include noise reversion circuits
tO Change the pulse generator tb ~’an:asynelironoi~s paeitlgmode :when ~e~nc~i~
ARTIFICIAL PACING 107
level surpasses the noise reversion threshold. This prevents inhibition of pacing in
the presence of noise. The sensing circuits also include circuits for protection of
the electronic, circuitry of the pulse generator in the clinical situations where
excessive voltage maybe applied to the sensing circuit. Such a case would happen
in the presence of defibdllation,~ensing is explained in Chapter 8. The sensing
element relays the processed information to the control unit where it is analyzed
and decisions are made.
5A.4 Codtroi
The control unit is responsible for determining whento send a pulse for pacing,
to change the mode, and to save data. The control unit for the most.part is ,a
timing device. The first control units were simple timers madeof resistors and
capacitors. Presently the timing circuit is madeof a crystal oscillator, which
generates accurate signals. Using the clock pulses, the control logic determines
whento trigger the output pacing pulse, the °blanking and the refractory intervals
and the AVdelay and to reset the escape interVals of an inhibited pacing system
or~ trigger initiation, of ~ AVdelay for triggered pacing modes, The control also
contains a rate-limiting circuit: that sets. an upper,limit ,(runawayrate) for pacing
in~case of a failure. ~Chapter 9-discusses the timing logic.. Chapter 10 covers_
CMOS.,technology used for implementation of control. In,-.many of today,s
pacemakers telemetry, circuits are available :to allow programmingthe control
functions and transfer of collected information(if any) for diagnosis purposes.
addition, a magnetic field detector is .included to~ intentionally, interrupv the
normal’ functions of :a pacemaker tot test purposes, Chapter 12 covers
programmingand telemetry. ~
5.2.2 Synchronous
5.2.3 Rate-adaptive
Despite the fact that the normal sinus node is the ideal rate-adaptive generator,
the. occurrence of sinus node dysfunction and atrial fibrillation in many
individuals limits the use of atrial sensing for rate adaptation(Kay, 1992). Rate~
adaptive pacing refers to the case whenthe base rate of pacing is influenced by
events occurring outside the heart. Modulation of the pacing rate based on these
events, whichcorrelate indirectly or directly with the level-of :metabolic ~demand,
satisfies the metabolic .needs of a-patient as the situation changes (e.g. ~from
resting to exercise). These : events ’. include changes in mechanical vibration,
ventilation, right ventrieular stroke volume and systolic time, intervals,
temperature, paced depolarization integral, and mixed venous oxygen saturation.
Chapter 13 explains rate-adaptive pacemakers.:. Chapters 14, 15, and 16 cover
motion-based rate-adaptive pacemakers, temperature-based rate-adaptive
pacemakers, and fight ventdcular stroke volumeand systolic time intervals-based
rate-adaptive pacemakers;
pacemakers,the size of pulse .generator is not a major issue and no major surgery
is involved, It can be used for diagnostic purposes as well as in therapeutic
applications where artificial pacing is needed immediately or temporarily.
Therapeutic uses of temporary pacing include any transient and reversible
situation such as tachyarrhythmias due to effects of antibradyarrhythmia drugs,
temporary symptomatic arrhythmias, open heart surgery, or while a patient is
waiting for.:~an iimplantable pacemaker. Diagnostic ap,p_lications of temporary
pacing inchide amongothers the assessment of a patient s tolerahee to stress -and
assessment of coronary artery diseases by increasing the heart rate artificially.
This section describes different techniques in applying temporary pacing. The
techniques differ in the delivery of the pacing pulse and include transcutaneous
cardiac pacing, transvenous pacing, transthoracic pacing, and transesophageal
pacing.
interval
<c,A A I :A
Figure5.2 Diagramshowjag-ventricular triggered and inhibited synchronousmode.(a)
representsa naturallyoccurringbea~tbeatsituation,(b) representsthe ventriculartriggeredpacing,
while(c) indicatesventrieular
inhibitedpa~.ing:,’,
Back
¯
In transvenous endq~ardial pacing, pacing leads are usually "lnserted~htravenously
under fluoroscopy !~nd local anesthesia to b~ " "
posmoned safely, a~d accurately
(Figure 5.4). This methodof pacing is the most reliable of all temporary methods
and can be used for a long time; however, it is time-consuming and has some
complications (such:i as venous spasm and chance of lead displacement with body
position changes). Both atria and ventricles can be paced (asynchmnously and
synchronously) aadi~t is used mainly for cardiac arrest and prophylactic pacing.
Pulse generator
Atrial
ball tip
cathode
Ventricular
cathode
Figure 5.4 TemporarytransVenouspacing. The 16ads ~are passed throtigh Veins under
fluoroscopy.
ARTIFICIAL PACING 111
In transthoracic pacing, pacing leads are inserted into ventricles directly through
percutaneous cardiac needles inserted into the chest wall. This method is the
fastest and simplest of all temporary pacing ’methods; however, placement of the
needle’can cause coronary vessel laceration leading to cardiac tamponade(Purcell
et al., 1986) and its efficacy is unproved(Woodet al., 1992). Only ventricles
easily be paced, and it is used mainly for cardiac arrest.
In transes0phageal pacing, pacing leads are placed into the esophagus; which.is in
the proximity of the left atrium. This methodis simple and safe whenpacing the
atrium and allows reliable atrial capture; however, it provides poor ventricular
capture and maybecomeintolerable (when the output current is high). It is used
mainly in prophylactic atrial pacing, diagnostics and termination of
sui~raventricular tachycardias (Woodet al.; 1992).
Indications for pacing are constantly evolving and it is difficult to define absolute
indications for pacing. Further, pacing, in particular permanentpacing, is costly
and overuse of pacemakersmustbe avoided. To address .these issues and to design
a standard for pacing indications, the American College of Cardiology and the
American Heart Association collectively developed and published guidelines of
indications for permanent pacemakers, antitachy pacemakers, and implantable
defibrillators in 1984. The guidelines also included the recommendedpacing
modes and were accepted widely, .In 1991,~the guidelines were revised to
incorporate new developmentsin artificial pacing.
The guidelines classify indications of pacing for different groups of
arrhythmias based on the degree of their acceptabilities and lack of
contraindications. Figure 5.5 shows the three classes defined by ACC/AHA for
permanent p~cing: This section discusses indications’for antibradycardia
permanent and’~temporary.pacing. SUbsections 5.2t.1 to 5.4.6 address indications
of permanent pacing while 5.4.7 summarizes indications of temporary pacing.
Chapter 17 covers indications for antitachyarrhythmia pacing and Chapter 18
addresses indications for implantable cardioverter defibrillators.
Class I
A. Completeheart block, permanentor intermittent, at.any anatomiclevel, associated with. any
one of the followingcomplications:
1. Symptomaticbradycardia. In the presence of complete heart block, symptomsmust be
presumedto be due to the heart block unless provedtobe otherwise.
2. Congestiveheart failure.
3. Ectopic rhythms and other medical conditions that require drugs that suppress the
automaticity of escape pacemakers and result in symptomaticbradycardia.
4. Documented periods of asystole _> 3.0 s or any escape rate < 40 beats/rain in symptom-free
patients.
5. Confusionalstates that clear with temporarypacing.
6. Post AVjunction ablation, myotonicdystrophy.
B. Seconddegree AVblock, permanentor int~rmittent,~ regardless of the type or the site of
~lock, with symptomaticbradycardia
¯ Atrial fibrillation, atrial flutter or rare cases of supraventricdartachycardiawith complete
heart block or advancedAVblock, bradycardia and any of the conditions described under IA.
Thebradycardiamustbe unrelated to di[i~alis or druids known’tO impair AVconduction.
Class lI ¯
A. Asymptomatic completeheart, block, permanentor intermittent, at any anatomicsite, with
V~ntdcular rates of 40 beats/rain or faster.
B. Asymptomatic type II seconddegree AVblock, permanen t or intermittent.
l
C. Asymptomatic type I secondd%,reeAVblock at intra-His Or infra-HiS levels.
A. First degree AVblock.
B. Asymptomatic
Class llI
type I second de~ree AVblock at the supra-His (AVnode) level. ,
Figure 5.6 1991 ACC/AHA guidelines for permanent pacing in acquired AVblock in adults.
FromDreifus, L. S., Fisch, C., Griffin, J. C.~ Gillette, P. C., Mason,J. W., ahd Pars0nnet, V.
1991. Guidelinesfor implantation of cardiac pacemakersand :antiarrhythmia devices. JACC18: 1-
13.
Class I
A. Persistent advancedseconddegreeAVblockor completeheart blockafter acute myocardial
infarctionwithblockin His-Purkinje system(bilateral bundlebranchblock).
B., Patientswithtransient advanced AV,block andassociatedbundlebranchblock.
: CI.ass
~II
A.Patientswithpersistentadvanced blockat the AVnode.
Class III
A. TransientAVconduction disturbances in the absenceof intraventficularconduction
defects.
B. TransientAVblockin the presenceof isolatedleft anteriorhemiblock.
C. Acquiredleft anteriorhemiblock in the absenceof AVblock.
D, Patieot~:,with persistent first degreeAVblockin the presenceof bundlebranchblocknot
dbmonstr~it~ed’previougly.
Class I
A. Bifascicular block with intermittent complete heart block associated with symptomatic
bradyeardia(as defined).
B. Bifaseicular or trifascicular block with intermittent type II seconddegree AVblock without
symptoms attributable to the heart block.
Class II
A. Bifascicular or trifascicular block with syncopethat is not provedto he due to completeheart
block, but other possiblecausesfor syncopeare not identifiable.
B. Markedly-prolonged HV(>100 ms).
C. Pacin$-inducedinfra-His block.
Class III
A. Fascicular block without AVblock or symptoms.
B. Fascicular block with fkrst de~ree AVblock without symptoms.
Figure 5.8 1991 ACC/AHA guidelines for.pacing in bifascicular and trifascicular block
(chronic). FromDreifus, L. S., Fisch, C., Griffin, J. C., Gillette, P. C., Mason,J. W., and
Parsonnet, V. 1991. Guidelines for implantation of cardiac pacemakersand antiarrhythmia
devices. JACC18: 1-13.
Class I
A. Sinus node dysfunctionwith documentedsYmptomatic bradycardia. In somepatients this will
occur as a consequenceof long-term(essential) drug therapy of a type and dose for whichthere
are no acceptablealternatives.
Class II
A. Sinus node dysfunction, occurring spontaneouslyor as a result of necessary drug therapy,
with heart rates < 40 beats/minwhena clear association betweensignificant symptoms consistent
with bradycardiaand the actual presenceof bradycardiahas not beendocumented.
Class III
A. Sinus node dysfunctionin asymptomaticpatients, including those in whomsubstantial sinus
bradyeardia(heart rate< 40 beats/rain) is a consequenceOf long-termdrug treatment.
B. Sinus node dysfunction in patients in whomsymptomssuggestive of bradycardia are clearly
documented not tohe associated with a slow heart rate.
Figure 5.9 1991 ACC/AHA guidelines for permanent pacing in sinus node dysfunction. From
Dreifus, L. S., Fisch, C., Griffin, J. C., Gillette, P. C., MasS,a,J. w., and Pa~sonnet,V. 1991.
Guidelinesfor implantation of cardiac pacemakersand antiarrhythmiadevices. JACC18: 1-13.
ARTIFICIAL PACING 115
Class I
A. Recurremsyncopeassociated with clear, spontaneousevents provokedby carotid sinus
stimulation;minimalcarotidsinus pressureincludesasystoleof > 3 s durationin the absenceof
_anymedicationthat depressesthe sinus nodeor AVconduction.
Class II
A. Recurrent syncope without clear, provocative events and with a hypersensitive
cardioinhibitory
response:
B. Syncopewith associated bradycardia reproducedby a head-up tilt with or without
isoproterenolor other formsof provocativemaneuversand in whicha temporarypacemaker.and
a secondprovocative
test canestablishthe likely benefitsof a permanent
pacemaker.
Class III
A. A hyperactivecardioinhibitoryresponseto carotid sinus stimulation in the absenceof
symptoms.
B. Vaguesymptoms,such as dizziness or light-headedness, or both, with a hyperactive
cardioinhibitory
responseto carotidsinus stimulation.
C. Recurrentsyncope,light-headednessor dizziness in the absenceof a eardioinhibitory
~response.
Figure 5.10 1991ACC/AHA guidelines for permanentpacing in hypersensitive carotid sinus
andneurovascularsyndromes.FromDreifus,L. S:, Fisch, C., Griffin, J. C., Gillette, P, C.,
Mason.J. W., andParsonnet,V. 1991.Guidelinesfor implantationof cardiac pacemakers and
antiarrhythmiadevices.JACC18: 1-13.
Class I
A. Secondor third degree AVblock with symptomaticbradycardia, as defined.
B. Advancedsecondor third degree AVblock with moderateto markedexercise intolerance.
C. External ophthalmoplegia with bifascicular block.
D. Sinus node dysfunction with symptomaticbradycardia, as defined.
E. Congenital AVblock with wide QRSescape rhythmor with block below the His bundle.
F. Advanced secondor third de~,e AVblock persistin[ 10 to 14 days after cardiac surgery.
Class II
A. Bradycardia-tachycardiasyndromewith needfor an antiarrhythmic drug other than digitalis
or phenytoin.
B. Secondor third degree AVblock within the bundleof His in an asymptomatic patient.
C. Prolongedsubsidiary pacemakerrecovery time.
D. Transientsurgical secondor third degreeAVblock that reverts to bifascicular block.
E~ Asymptomaticsecond or third degree AVblock and a ventricular rate < 45 beats/min when
awake.CompleteAVblock whenawake,with an average ventricular rate < 50 beats/rain.
F. CompleteAVblock with double or triple rest cycle length pauses or minimal, heart rate
variability.
G. Asymptomatic neonate with congenital completeheart block and bradycardia in relation to
age.
H. Complexventricular arrhythmiasassociated with gecondor third degree AVblock or sinus
bradycardia.
I. Long QT syndrome.
Class III
A. Asym_ptomatic postoperativebifascicular block.
B. Asymptomatie postoperativebifascicular block with first degree AVblock.
C. Transientsurgical AVblock that returns to normalconductionin < I week.
D. Asymptomatic type I seconddegree AVblock.
E. Asymptomatie congenital heart block without profoundbradycardiain relation to age.
Figure $.11 1991 ACC/AHA guidelines for permanentpacing in children. FromDreifus, L. S.,
Fiseh, C., Griffin, J. C., Gillette, P. C., Mason,J. W., and Parsonnet, V. 1991. Guidelines for
implantation of cardiac pacemakersand antiarthythmia devices. JACC18: 1-13.
paci,ng~ This code, however, was not intended to replace the generic code. In
1987~ NASPE.and the British Pacing and Electrophysiology Group (BPEG)
ad0pted a generic code for pacemakers based on the two earlier recommended
versions of pacemaker codes (1974 and 1981 ICHDcodes). This new generic
code called NASPE/BPEG Generic Pacemaker (NBG) Code allowed inclusion
additional features such as rate-adaptive (in whichthe escape rate is controlled by
events other ~than that of heart) and antitachyarrhythmia devices including
eardio~;erters and defibrillators. The generic code is used for conversational tasks
and th, e~:specific code is used whenthe generic code can.not provide enough
ilnf0~tiOn (such as the case in some dual pacemakers). In effect; the NBGcode
is a ebaci~e code andthe NASPE Specific code is a precise code.
~ Three-letter pacemakercode ICHD
-! . ":Position I H m
-~ Cate$or~
¯ " Chamber(s) paced Chamber(s)sensed Modeofresponse(s)
’ V= Ventricle V= Ventricle T = Triggered
’ A = Atrium A = Atrium I = Inhibited
D = Double D = Double D = Double*
’,.......L:~: - , . . O = None. ..... O =,None
~Atrialtriggeredandventricular
inhibited,
Figure 5.15 shows that the NBGcode is composedof five letters: The first letter
corresp6nds to the chamber(s) paced. In this position, O refers, to :no bradycardia
pacing, V refers to pacing the ventricle while A denotes pacing the Atrium and D
corresponds to Dual whenpacing both chambers.
The second letter corresponds to chamber(s) sensed.~ The lettei~s are the same
as for the first position.
The. third letter corresponds to the response to sensing: An O is used when
there is no ’r6spons~ to the sensed event. A ~T is used Whenthe sensed evem
triggers the paced event and an I whenthe sensed event inhibits the paced event.
A D i s used if in responseto the sensedevent(s), pacing in chamber(s) is inhibited
orltriggere d basedonsensed eVent(s). D in this’position stands for Dual (T +
""The fourth letter corresponds to programmability and rate modulatiom An
O in this’ position indicates :the lack of programmability-andrate modulatiori. A’ P
stands f6~ simple programmable, in which case one or two features can be
programmed: Usually, however, it indicates the programmability in rate and
118 DESIGN OF CARDIAC PACEMAKERS
Figure 5.14 The revised five-letter version of the 1974 ICHDcode, was,recommended in 1981.
In the first twoletters, there wasno change.In the third position an additional, letter wasaddedto
indicate pacemakerswhich start pacing when-the heart beats becometoo fast. This modeof
response was designated by R for Reverse. Letter position four correspondedto programmable
functions and containedthree letters: P for (simple) programmable (either rate or output or both
can be programmed),Mfor multiprogrammable (capable of changing morethan two features)
Ofor no functions. Letter position five eorrespoudedto special tachyarrhythmiafunctions. In this
position, O stands for none, B for bursts of impulses, N for normal-rate competition(such as .in
dualrdemandpacemaker),S for scanning response (such as timed extrasystoles, etc.), and E
external control (pacemakeractivated by a magn,~tof by radiofrequenc)/)~ Telemetricf_~nctions
were not’included: in the code as they did not involve pacing or s~nsing,modalities i From
Parsonnet, V., Furman, S., and Smyth, N. P. D. 1981. A revised code for pacemaker
identification. PACE 4: 400-402.
Figure 5.15 The NBGpacemakercode; internationally accepted and adopted Since 1987. From
Bernstein, A. D., Camm,A. J., Fletcher, R. D., Gold, R. D., Richards, A. F., Smyth,N. P. D.,
Spielman, S. R., and Sutton, R. 1987. The NASPE/BPEG generic pacemaker code for
antibradyarrhythmia and adaptive-rate pacing and antitachyarrhythmia devices. PACE 10: 794-
799.
NASPEspecific code
General format Atrial-channelfunctions/ ventricle,channelfunctions
a-c abef (a-c ate f) / v-c abef (v-cate
Antibradycardiafunctions Antitachycardiafunctions
.Basic functions O=-none U=underdrive
pacing B=burst
S=sensing ¯ 0R=ramp
I--inhibited X=extrastimulus
T--triggered¯ l C=cardioversion
D=defibrillation
Source of sensing a=atrium. a=atrium~
~--ventdcle v=ventricle
e=extemal e=extemal
~igu~ 5.16 NASPE specific code. The code ¯provides a detailed -description of functions of a
ember,a-c: atrial~hannel; v-c: ventricle-channel;abe: antibradycardia;ate: antita¢~ycardia;f:
funeti0hS~ FromBernstein, A~ D.,’C~ A. J., Fletehef~ R. D., G01d,R. D., Ridhardg~A. F.,
Smyth~N. P. D., Spielman, S. R:, and Sutton, R. 1987. The NASPE/BPEG generic pacemaker
code for~antibradyarrhythmiaand adaptive-rate pacing and antitachyarrhythmiadevices. PACE 10:
794-799.
Examples
NBGcode NASPEspecific code
VOO O/P
O/PSIv
O/PSIv
VVIC O/PSIv
O/PSIv
PSla/O
DVIM PIv/PSIv
S/PSIvTa
PS~aIv/PSIvTa
DDDM PSTaIv/PSIvTa
DDDR PSTaIv/PSIvTa
DDDMS PSIaIv/PSIvTa
DI)IX~ PSIaIv/PSIvTa(D)
VVIMP O/PSIv(Bv)
OOOOP
DDDCP PSIv(BaBvUv)/PSIvTa(Uv)
Figure 5.17 Examplesof NBG and NASPE specific code. NASPE explains the functionality of
a pacingmodewith a greater degreeof precision.However, it doesnot holdinformationsuchas
programmability,
telemetry,or rate-adaptation.Thusin the NBG examples,the fourthpositionhas
no bearingon the NASPE specific code. Theabsenceof the fifth letter indicates the absenceof
antitachycardiafunctions. Unlikethe NBG code, NASPEspecific codespecifies antitachycardia
functionsandthe chamberin whichantitaehyarrhythmia functionsexist. FromBemstein,A. D.,
Carom,A. J., Fletcher, R. D., Gold,R. D., Riehards,A. F., Smyth,N. P. D., Spielman,S. R.,
and Sutton, R, 1987.TheNASPF_,/BPEG generic pacemakercodefor antibradyarrhythmiaand
adaptive-ratepacingandanfitachyarrhythmia devices.PACE 10: 794--799.
Pacing E=escapeinterval
parameters: R=refi’actoryperiod
AV=atdoventricularinterval
Controlling
variables: ps---pace or sense . ~
qt=stimulus-to-Twaveinterval
r=respiration
t=temperature ¯
vpb=ventricularprematuredepolarization
Figure 5.18 Rate-adaptive parameters suggested by Bemstein in 1991 tobe added to NASPE
specific code for inclusion of rate:adaptivity in the code. From Bernstein, A. D: 1991.
Classification of cardiac pacemakers.In EI-Sherif, N., and SametP. (eds.) Cardiacpacing and
electrophysiology. 3rd Ed. Saunders.
Figure 5d9 The NASPE/BPEG Defibrillator Code (long version). From Bemstein, A. D.,
Carom,A. J., Fisher, J. D., Fletcher, R. D., Mead,R. H., Nathan, A. W., Parsonnet, V.,
Richards, A. F., Smyth, N. P. D., Sutton, R., and Tarjan, P. P. 1993. The NASPE/BPEG
defibrillator code. PACE16:1776~1780.
Positions I and II are needed in all cases. Position III should be included if
antibradycardia function is present or hemodynamic sensing is present. For
device labeling and record keeping, when antibradycardia pacing is present, the
fourth letter is replaced with a hyphen followed by the 4 letter NBGcode
corresponding to that antibradycardia pacing mode. For example, a ventricular
hemodynamic defibrillator with rate-adaptive ventdcular antibradycardia pacing
could be labeled VOH-VVIR.
Figure 5.20 shows a version of the NBDcode, called the short form, which
is intended for conversational usage.~ This code allows distinguishing among
devices that are cardioverters or defibrillators only and the ones that incorporate
antitachycardia and antibradycardia pacing as well. Figure 5.21 provides a few
examples of short and long forms of the NBDcode.
Figure 5.20 The NASPE/BPEG defibrillator code (short form). From Bemstein, A. D., Carom,
A. J., Fisher, J. D., Fletcher, R. D., Mead,R. H., Nathan, A. W., Parsonnet, V., Richards, A.
F., Smyth, N. P. D., Sutton, R., and Tarjan, P. P. 1993. The NASPE/BPEG defibrillator code.
PACE16: 1776-1780.
Figure 5.21 Examples of defibrillator code (NBD). From Bernstein, A. D., Camm,A. J.,
Fisher, J. D., Fletcher, R. D., Mead,R. H., Nathan, A. W., Parsonnet, V., Richards, A. F.,
Smyth, N. P. D., Sutton, R., and Tarjan, P. P. ¯1993. The NASPE/BPEG defibrillator code.
PACE16: 1776-1780.
Manufacturers use the NBGcode to give the description of pacemakers. The five
letters designate the mode and capabilities of a pacemaker. At least the first four
letters are needed to give product descriptions; however, in clinical usage, only
three are needed if the fourth letter does not indicate a rate-adaptive pacemaker
(Bernstein et al., 1987). Each different pacing mode can be indicated for certain
abnormalities and may ’be harmful for , some others. (Improvements in
implementations may change the view.) This section reviews different modes of
pacing, their usage, and their contraindications.
ARTIFICIAL PACING 123
In’this mode, the atrium is paced if a P wave does not occur within a certain
interval. The occurrence of a naturally occurring P-wave will inhibit the
pacemakerpulse..In addition, the length oft he interval is changed in response to
physiological events. AAIRis used in treatment of symptomaticbradyeardia as a
result of sinus node dysfunction in the chronotropically incompetent patient and a
predicted high level of physical activity and whenAVconduction can be proven
normal. This mode will restore rate responsiveness and maintain AVs~nchrony.
AAIRhas been contraindicated whenadequate atrial sensing can not be attained
and with sinus node dysfunction with associated AVblock eider demonstrated
spontaneously or during preimplant testing (Hayes, 1991), .....
block with normal sinus and atrial electrical function, but it should be avoided
whenan abnormality of sinus or atrial function is present. Due to the atrial
sensing refractory period, the maximal ventricular rate is limited (to
approximately 125 beats/min), which is disadvantageous in young patients but is
protective in situations whereatrial sensing mayoceursuch as in supraventricular
tachycardia and atrial flutter. Whena unidirecti’onal AVblock with slow
retrograde conduction is present, it may result in taehyarrhythmias (Mond,
1983).
In this mode, occurrence of certain events (P or R wave) start cycles. Since only
ventricular activity is tracked in pacing ventricles, its rate of pacing never
exceeds the programmedrate. This mode is used in patients with Pa.roxysmal
supraventricular tachycardias that could result in rapid ventricular pacing if the
patient were programmedto the DDDmode (Hayes, 1991), It is also used with
chronotropically competent patients with AVblock and sinus node dysfunction in
the presence of significant PSVT.It has .been contraindicated, in chronotropically
incompetent patients with a demonstrated need or improvement with rate
responsiveness (Hayes, 1991).
After the patient has been diagnosed properly and the need for permanent .pacing
has been confirmed, the proper pacemakerrnustbe identified.~ I,~ selecting the
p..roper~pacing mode, whether synchronous,~Or~rate~adapti,~e (asynchronous mode
is obsolete as an implant time mode), severalfactors must: be considered. These
factors include the patient’s physical condition (Hayes, 1991), the presence
coronary heart disease and angina pectoris (Dreifus et al., 1993), the anticipated
level of activity, exercise capacity, chronotropic response to exercise, and the
cost. For example, in th~ absence Of chronotr~pic i/~crmpetence, one may not
need a rate-adaptive pacemaker, which (at present) costs more than a synchronous
pacemaker.
Somebelieve that single chamberrate-adaptive pacemakers can replace dual
chamberpacemakers in terms of physiological effectiveness. However,the single
chamber rate-adaptive pacemakers do not maintain AVsynchroriy~ This may not
be an issue at fast heart rates, but at slow heart rates, maintaining AVsynchrony
is desired. Absence of AVsynchrony may lead to atrial fibrillation and stroke
and reduces patient life expectancy,.,especially !f impaired ventricular functions
are present. Figure 5.22 shows whe~ier pacing ~aodes .listed preserve. AV
synchronyand are rate-responsive to exercise.
Finally, in selecting the proper pacemaker, one needsto have a:’ complete
knowledge of advantages and disadvantages Of using a pacing mode an’d be
fa~iiiar with abnormalities and ~eir cfffiseq~ences~ Figure 5.23 shows an
example of an algorithm for selecting the ~roper pacemaker for a patient
possessing symptomatic bradycardia while FigureS.24 shows a logic diagram for
selecting the proper pacing mode~
128 DESIGN OF CARD~C PACEMAKERS
Figure 5.22 Hemodynamic effects of pacing mode. N/A: not applicable; *: only if AV
conduction is preserved; **: only if sinus response to exertion is, preserved; ***: unless sinus
bradycardia is present. From Naccarelli, G. V. 1991. Cardiac arrhythmias: a practical approach.
Mount Kisco, NY: Futura Publishing,
Figure5.23 Alogical sequencein selecting the ~proper pacing modefor a patient with
symptomatic bradycardia.FromBenditt, D. G. 1993.Current pacing, modes:a brief reviewof
their features andindications. In Benditt, D. G. (ed.) Rate-ad~ptivepacing.Cambridge,
MA:
BlackwellScientific.
5.8 REFERENCES
DDDR I
DDIR ]
DDD I
Figure 5.24 A logical diagram of relationship between rhythm disturbances and therapeutic
opfacing modesBedim
the heart. for selecting the proper" pacing mode.FromSchaldach, M. M. 1992. Electrotherapy
Springer-Vedag.
Bernstein, A. D., Cairn, A. L, Fletcher, R. D., Gold, R. D., Richards, A. F., Smyth, N. P. D.,
Spie!man., S. R., and Sutton, R. 1987. The NASPE/BPEG generic pacemaker code for
" at~til~rad#arrhy~a and adaptive-rate pacing and antitachyarrhythmiadevices. PACE 10: 794-
799.
Brownlee, R. R., Shimmel, J. B., and Dcl Marco, C. J. 1981. A new code for pacemaker
Operating modes. PACE4: 396-399.
Dreifus, L. S., Fisch, C., Griffin, ]. C., Gillette, P. C., Mason,]. W., and Parsonnet, V. 1991.
Guidelines for implantation of cardiac pacemakersand andarrhythmiadevices. JACC18: 1-
Ell~n-bogen, ~K. A., and Peters, R. w~1992. Indications for permanentand temporarycardiac
i ~ pacing. In Ellenbogen,K. A. (ed.) Card’~.~pacing:
BoSton:BlackwellScientific.
Furman, S. 1993. Pacemakercodes. In Furman, S., Hayes, D: L.; and Holmes, D, R: (eds.)
’ ’p/a~tice Of cardiacpaci~lg3rd Ed. MountKisco, NY:ilutura Publishing.
Eurman,S., ,Hayes, D. L.,~imd Holmes,D. R. Jr. 1993~A practice of cardiac pacing. 3rd Ed.
MountKisco, NY:Futura Pul~lishing;
Garrett, A. E., and Adams,V. 1986. Common cardiac arrhythmias, recognition and treatment.
Pl~i,~’ladelphia:
J. B.Lippincott.
Griffin, C;IIG{1993. Indications for the use of implantedarrhythmia devices: comments on the
i991 ACC/AHA Task Force report. In Barold, S. S. and Mugica,J..(eds.) Newperspectives
in carcl~.,~ pacing.3. MountKisko, NY:Futura publishing~ . ,, ,
Hayes;D~E.~ 1993. Indications for permanentpacing. In F~an, S., Hayes, D. L., and rtonnes,
. D.R. A practice of cardiac pacing 3rd Ed. MountKisco, NY:Futura Publishing.
Hhy~s,D. L., and Holmes,D. R. 1993. Temporarycardiac pacing. In Furman,S., Hayes, D. L.,
and Holmes, D. R. (eds.)A practi~e of cardiac piecing 3rd Ed. MountKisco, lqY: Futura
Publishing, ""
~.~.#,’G. N; !992. Basic aspects ofciirdiac pacing. In Ellenbogen,K. A.
~ (ed.) Cardiac pacing.
~ ,>"B~stoniBlackwellScientific; " .... . " " "
M6~id~H. G~ 1983. The cardiaC pace~r,’function and raalfunction: 1st Ed, NewYork: Grune
and S~atton. ’ ’ ......
Naccarelli~ G. V. 1991. Cardiacarrhythmias: a practical approach~MountKiaco, NY:Futura
Publishing.
130 DESIGN OF CARDIAC PACEMAKERS
Parsonnet, V., Furman, S., and Smyth, N. P. D. 1981. A revised code for pacemaker
identification. PACE 4: 400-402.
Purcell, L A., and Burrous, S. G. 1986. Fundamentalsof pacing. In Riegel, B., Purcell, J. A.,
Brest, A. N., and Dreifus, L. S. (eds.) Dreifus" pacemakertherapy: an interprofessional
approach?Salem, MA:F.A. Davis.
Purcell, J. A., Kloosterman,N. D., and Miller, L. K. 1986. Care of the hospitalized patient
undergoingpacemakertherapy. In Riegel, B., Purcell, J. A., Brest, A. N., and Dreifus, L. S.
(eds.) Dreifus’ pacemakertherapy: an interprofessional approach.Salem,MA:F. A. Davis.
Reynolds, D. W. 1992. Hemodynamics of cardiac pacing. In Ellenbogen, K. A. (ed.) Cardiac
pacing. Boston:BlackwellScientific.
Schaldach,M. M. 1992. Electrotherapyof the heart. Berlin: Spdnger-Verlag.
Wood,M., Ellenbogen, IC, and Haines,D. 1992. Temporarycardiac pacing. In Ellenbogen, K.
A. (ed.) Cardiacpacing. Boston:BlackwellScientific.
5.11 ~ esignation in both NBG and NASPE specific codes if it were madetoday.
xplain cardioversion~defibrillation, underdfive,and burst.
5.12 Write the code for each of the following pacing mOdesin both NBG arid NASPE specific
codes: (I) ventricular pacing inhibited by sensing in ventricle and tachycardia detee~on
ventriclesresults in bta:sts in ~/entriele(2) pacingin ventricleinhibited~); sensingin ventricle
and includes these anfitachyarrhythmia functions: burst, underdri~,e, ~amlJ, b~t plus
ex_tr;astimulus, extrastimulusin ventricle aeti~,ated externally (3) pacing of b~th’eh~bers
alter an event sensed in the atrium’(andinclude multiprogran~iJility).
5.13 De~cribeeach of these pacing mOdes:(1) PSIalv/PSIv’~a(D),(2) O(Ua)/O;(3) P/O
PIv/S’. .... ’ ....
5.14 Describethe operation of each DDD modelisted ~ Figure 5.17.
5.15 Write the cOdefor these VVIRpace.makers using NASPE specific code and Suggest~drate-
adaptiveaddition: (1) varies e~eapeiiiterval in responseto the changesin (a) te~ra~
ventricular prematuredepolarization(e) ventilation, (2) varies refi’dctory pedodm response
to changesin (a) activity level(b) stimulus’to-Twaveinterval, and (3) varies atfioventdeu!ar
interval in responseto.(a) temperature(b) ventilation.
5.16 Write the code.for aDDD pacemakerWhoseAVinterval differs after paced and sensed
events and whose atrial refractory period is extended after a Vent~ular premature
.d.e.polarizatiOn. Givethe advantages~f ~sing this p~maker.
5.17 write the code in NBD format (short form,~long form and label) for the followingdevices:
(1) ventricle-only ICD wlthVVIR~antibrad}e~dla pacing, (2yICD with ventri~ular
cardioversiorddefibdllation, dual chamberantitaehyca~y~din paein~~1. DDDR antibt;ady~dia
pacing, and (3) ICDwith X;entrieular cardi0Ve~sion/defibrillafion, AFconversion,
hemodynamic sensing, and WIRantibradyeardia pacing. Note that, in each Case, more than
one long form mayb~ p6ssible.
ARTIFICIAL PACING 131
5.18 Describe each of the following NBDcodes, indicate whetherthe correspondingcode is for
general description, for conversational purposes, or for book-keepingpurposes and give
each code in its conversational format if not already in that form: (1) VOEO (2)
(3) AAEA (4) DDED(5) DDE-DDDC(6) DDH-DDDR (7) AAE-AAICICD- T
.(9) ICD-S(10) ICD-B.
5.19 List the potential problemswhenusing VOO mode.Describe howa failed pacemakersensor
wouldaffect the patiem. Identify the groupof patients for whichAOO maybe used. Identify
the modethat preserves AVsy~Chrony. "
5.20 Explain the pacemakersyndrome.Give the indications for VVIpacing. List the conditions
for which the use of DVImodecan be beneficial. Identify the modethat preserves AV
¯ synchrony.
5.2i Identify the situations in whichAATor VVTmaybe used. List the disadvantages of using
these modes.Identify the modethat preserves AVsynchrony.
5.22 DiStinguishwhich of these pacemakersmaybe used in chronotropieally competentpatients
and which in ehronotropically incompetent patients: VVI, VVIR,DVI, AAI, AAIR,DDI,
~d DDDR. List the modesthat preserve AVsynchrony.
5.23 Comment on the rate responsiveness to exercise of each modementionedin section 5.6 and
, hot listed in Figure 5.22. Comment on whether ornot the rate-response can be considered ¯
~’ ~te in the presence of atrial arrhythinias in each case. ’ ~ ’
52~4 E~pihin~/htth~r or not dual chamberpacemakerscan replace single chamberrate,adaptive
: .: ,’ pacemakers,ExplainwhyAVsynchronyis important.
5.25 S~gest th~ proper.pacingmode,for the following patients using,Figure 5.24: (1) a patient
! : ,, Wi~often ~ ~uare~lia,b.le atrialrhythms(2) a patientwitli sometimesumllab!eatrial rhythSmsbut
’~ Witii no n~ for atrial syn~hrony(3) a patient with unreliable atrialrhythmsin needof rate
~ - ~dapt3vepaeihg. Ifmor6than one modeis possible~ select only one as the proper modeand
. , ?~justify your;answerassumingthat you have a programmablepacemaker.Whatwouldhave
youselectedif cost werea factor in determiningthe propermode..’?Justify youranswer.
6
The modempacemaker electrode often fulfills two major roles. The first is the
introduction of cathodal stimuli, produced by the pulse.generator, into excitable
myocardial muscle. If implemented, the second role is to optimally sense
intracardiac electrocardiograms and conduct them back to the pulse generator for
signal processing and algorithm control (Sinnaeve et al., 1987).
Figure 6.1 shows an example of a modemporous steroid eluting electrode.
The implementation of porous electrodes began in the late 1970s at Cardiac
Pacemakers, Inc. [CPI, St. Paul, MNU.S.A.] (Mugica et al., 1988). Various
porous and steroid eluting electrodes have provided significant advances in the
pacemakerindustry. Porosity utilizes the principle .that the ratio of the electrode
ELECTRODES, LEADS, AND BIOCOMPATIBILITY 133
tip’s electrically active surface area to the tip’s overall size should be large
(Mugica et al., 1988; Schaldach, 1992). Steroid elution designs increase pacing
efficiency and sensing sensitivity by reducing encapsulation of the electrode tip.
The primary design criterion for electrode design is safe cardiac stimulation.
A patient’s safety is of utmost importance. Mini~zing energy loss from a small
pulse generator battery source is another~important electrode design
consideration. Charge consumption from_, the.battery is reduced, to provide
extended life. Electrodes used with today s du~. ghamber and rate-responsive
pacemakers, which inherently draw increased c~,urrent from the power source,
must especially reduce current drain. Because the heart beats an average
exceeding 85,000 times a day, the potential existsi~at a relatively large amountof
energy could be inefficiently used. During paci~gi~:--the pulse generator acts as a
power source while the heart acts as anelectriCal lo~’d. Wl~enthe pacemaker
operates in sensing mode, the heart becomesthe source of electrical~energy~and
the pacemakersensing circuit becomesthe load. These differing and alternating
roles sometimesrequire different design considerations.
Other-important electrode design criteria include: biocompatibility,
biostability, electrode size, invasiveness to cardiac and circulatory functions,
fixation! into heart tissue, and ease of clinical manipulation(implantation and if
necessary removal). While overall system biocompatibility i~s addressed
specifically in section 6.3, electrode tip biocompatibility is of special concern
becauseit directly correlates to electrode efficiency.
134 DESIGN OF CARDIAC PACEMAKERS
Endothelium
Figure ~6.3(a) shows a first order approximationof .the electronic and ionic
interactions occurring at the stimulating cathodal electrode-electrolyte interface.
More complex, models can and havebeen developed. An electrical phase
boundaryis defined as an interface whereon one side electrical charge is .carried
by electrons, while on the other, side charge is carried by ions, Electrons in the
electrode are drawn to the interface surface by their attraction to positive ions
ELECTRODES, LEADS, AND BIOCOMPATIBILITY 135
present in the bodily electrolyte (Na+ and H+, for example). Electrolyte cations
are drawnto the interface surface by their attraction to the electrode’s electrons.
Equal and opposite charge concentrations arise on each side of the electrode-
electrolyte interface and an electrical field is thus established (Deconinck,1992).
Interface particles
’Cation
Electron
M-O ~ Metal-oxide
J i complex .
-~: i ~ Helmholtzdoublelayer
C-H ÷
Figure 6.3(a) shows a primary layer of water molecules nearest the electrode
~surface, ,More complex models may incorpgrate many more primary layers.
~i]aese’i~d~ary layer water molecules neariy C0nii~leteiy co~ter the e!ec~0d~tip’s
~ufface. N0td that dipoled water moleculds tehd~tp ’ align themselves
~ Under.the
influence Of the inducedelectric field (Waltonet al.,"1987).
136 DESIGN OF CARDIAC PACEMAKERS
Figure 6.3(a) also shows the attraction of positively charged cations to the
electrode tip’s surface. Electric fields exerted by these cations are sufficient
enough to draw dipoled water molecules around them. Thus, these ions develop
hydration shells (Schaldach, 1992; Walton et al., 1987). These ion-water shell
complexes are knownas hydrated ions. Positive hydrated ions are drawn towards
the electrode-electrolyte interface. Together the hydrated ions comprise the
secondary water layer. The primary and secondary water layers comprise the
Helmholtz double layer, as proposed by.Helmholtz in 1879 (Schaldach, 1992;
Waltonet al., 1987).
Figure 6.3(b) shows that a simple Helmholtz double layer approximation
corresponds to a plate capacitor schematically represented byCH.The value of
this Helmholtzcapacitance is determined by three physical factors: the dielectric
constant of the primary water layer er; the active surface*area of the electrode tip
a; and the distance d betweenthe electrode’s charge and the electrolyte’s ions:
CH = e°era - (6.1)
d
The second method to obtain RCTis simply charging and discharging the
Helmholtz capacitance shown in Figure 6.3(a) as CH. In order to improve the
reversible charge transfer by means of CH, the capacitance must be increased.
Accordingto Equation 6.1, this can be achieved by increasing the active electrical
surface area a. Section 6.1.4 discusses howa can be madesubstantially large by
implementing surface porosity.
-i.
Anode :
Four major model components are shownin Figure 6.4: the pulse generator,
cathode, anode, and tissue. VBrepresents the pulse generator’s stimulation
votfage. CRis the pulse generator’s reservoir capacity. Cc is the coupling capacity
t~etween the pulse generator and the leads. Switches S1 through $3 represent
~ous switching elements designed in many pacemakers to permit charging and
¯ ~harging of various system ,capacitances.
The lead resistances for the cathode and anode are shown by RCand RA,
f~Sp~ectively. The cathode and anode specific Helmholtz capacitances CHCand
~, le~tie~tive ly, multiplied by their respective ~leetrode ~a~a andai result in
138 DESIGN OF CARD~C PACEMAKERS
R = RE + Rt (6~4)
RL = RC + RA (6.5)
(6.6)
.K,t~R ~’c aCHc
C2 = Cma (6.7)
y_~y2 -4x
Pl =- (6.9)
2x
P2 = _y÷__2~ (6.10)
2x
atid variables x ~d y are determined by:
(RLRS
+ RLe+&R)qC2 (6.11)
y (6.12)
(6.13)
[e p2T -I~(RLp2 + R)C2P2+ 1]]
Macrophagesand foreign body giant cells migrate into the pores, cracks, or
grooves of an electrode tip’s surface. In addition to the collagen capsule
surroundingthe tip, these cells increase the effective electrically active size of the
electrode. Indeed, an extremely small tip can be potentially biologically
destructive due to Faradic current. Yet, too large an electrode size results in a
decreased electric field density at the myocardium. Electric field density
decreases as a function of the square of the distance between the electrode’s
surface and the myocardium(Irnich, 1973; Irnich, 1975).
Pacing Sensing
Electrode polarization
Microscopic
surface area
Macroscopic
geometric size
capacitance CH. This second method requires that .the electrode tip’s active
surface area a be increased. Both methodsincrease pacing efficiency.
(a) (b)
Figure6.8 Electricalfield inducedbyan elec[r0detip. (a) In this ideal case, noencapsulation
the electrodetip has yet 0¢eurred.(b) In re~, h0~ever;p~ss .of encapsulationre~ultff in,a
collagenousnetworksU/’rounding tip’s surf/lee area. This effecfi~Veeiilargementof the tip s
geon~triCsize re, stilts in a r~aldi~r~of dn~ttedelectric field aadd~pacingefficacy.
Detection sensitivity
The same electrode used to pace myocardial tissue is commonlyalso designed to
sense, or detect, an electrogramsignal from’: within the herbert and deliver itto the
pulse generator. Somepacemakersutilize this natural cardiac signaLin algorithm
control for demandand/or rate,adapfve.pacing systems,
Because pacing and detection applications shar~e ~e same electrode-
electrolyte inteffa.ce~:electrode pola~rizati~n losses ~or,sensi~g are, reducedin the
same mannerthey are for pa~ii~g, ~Methodsav~lable for reducing these interface
losses are described in section 6,1.2~
Chapter 8 discU,s~s.eS methods6g enhancingdetection sensitivity by utilization
of sophisticated amplifier designs. Combinedwith highly selective filters, these
amplifiers commonlyhave both extremely high input resistances and gains, ~They
often provide high fidelity: reproductions of carc[iac signals to the p, ulse
generator. A major criteria for electrode design is minimizingcharge loss. from a
small battery source to increase dewce longewty. Muchof.a pacemakeg,.SYStem
charge loss is attributed to pacing. Enhanceddetection circuitry internal to the
pulse generator provides ~design flexibility for increasing:pacing efficiency and
extending device longevity,
ELECTRODES, LEADS, AND BIOCOMPATIBILITY 143
For traditional design, the electrdde tip’s radius should be less than or equal to
of that will i
If the thickness of the expected encapsulation layer is between 0.80 and 1.4 mm,
the surface area of the tip is calculated to be between 4.0 and 12 mm 2 as the
optimal, i: .tip ., : radius is assumedto be 0.80 to 1.4 mm,respectively. Effective
traditional .pacing cathodes commonlyexhibit geometric sizes within this range.
Not6 that as an encapsulation layer accumulates on the surface of the tip, the
effective radiu~ and size of the electric field emitting active surface increases.
This.reduces pacing efficiency. ~
Somemore contemporary designs have tip sizes smaller than 4.0 mm 2 due to
suppressed inflammation and reduced encapsulation techniques. The next two
sections will discuss how different tip materials and steroid eluUon techmques
provide ~e capability to design smaller, more efficient,electrodes.
" In addition to size, the evolution of pacemakerelectrodes has also included
several types of tips not having simple rounded surface~i During the past decade
electrode designers have developed electrode tips having a wide variety of
different geometric shapes. Sometips are nearly spherical or hemispherical.
Other tips are flat, annular, ring-shaped, or barbed (similar to that of
fishhook), lntermedics [Intermedics, Inc, Angleton, TXU.S.A.] has introduced
an IROX®model electrode implementing sharp edges and points for.l, ocalized
concentration of current. Debate still continues to whether any of-these designs
alone contribute to increased pacing efficacy (Adler et al., 1990;
Djordjevic et al., 1986; Karpawichet al., 1992; Mugicaet al., 1988; Pioger and
Rip:art, 1986).
To reduce electrode polarization losses_for pacing and ..sensing, the active
microscopic surface area of an electrode tip Canbe increased without necessarily
increasing thefip’s geometric size. The active s.urface ~ea can be greatly
increased if the tip’ s surface is designedto be conducivefor bodily electrolytes to
flow into any surface microcavities. Collectively knownas being porous, ~several
industrial processes have been developed to produce high active surface areas.
144 DESIGN OF CARDIAC PACEMAKERS
Figures 6.9(a) through (e) show electron microscope scans of five different
porous surfaces. Note that in all of the different porous surfaces shownin Figure
6.9 that microscopic grooves, crevices, or pores exist for electrolyte to flow into
or between. The microscopic ridges and edges generated by these various surface
processes increase the tip surface area. In modemporous electrodes, microscopic
surface areas have been reported to exceed macroscopic surface areas by factors
sometimes exceeding 1,000 (Schaldach, 1992). Other methods of creating porous
surfaces not sbown in Figure 6.9 include- sintedng--where metallic powder is
partially welded together by application of nonmelting beat--and chemical vapor
deposition (CVD)(Schaldach, 1992).
Figure 6.10 shows how pacing voltage loss can be reduced using a porous
electrode tip surface as opposedto a relatively smoothtip surface.
Electrode materials
I
Smoothelectrode surface - _ .
Porous, platinum
Electrode
body coatedtitanium tip
Silicon rubber
plug (impregnated
with DSPsteroid)
apparatus and protection of lead filament coils. In addition, lead flexibility is also
decreased (Cameronet al., 1990; Ormerodet al., 1988). Section 6.2 discusses the
physiological implications of both of these design disadvantages. Coated helical
tips, however, permit less hazardous transvenous passage without having
rotational apparatus incorporated into the lead (Ormerod et al., 1988).
nonretractable helical screw is encapsulated with a biocompatible gel that
dissolves after implantation, thus making insertion and handling muchsimpler
and less dangerous than other nonretractable models (Ormerodet al., 1988).
All-three active fixation methods result in extremely low dislodgement rates
(often lower than 1% of the total numberof implantations). However,invasively
puncturing heart tissue via active fixation, is theorized to induce increased
inflammation. It has already been determined that such an increase reduces pacing
efficiency.
A third method of implanting pacemaker electrodes has:also been developed:
passive fixation. Passively f’~xated electrodes utilize natural tissue encapsulation of
their physiologically foreign materials to anchor the devices in place. They do not
puncture the myocardium--theygently lie against the endocardium very near the
myocardium. This reduces inflammation. Several mechanisms have been
developed .to promote tissue passive fLxation, including: wings, crowns, flanges,
bristles, projecting wires, and tines. Often these fixation devices are comprisedof
polymers due to their relative biocompatibility and flexibility. While active
fixation methods still-exhibit lower dislodgment rates, improving technology
contributes significantly to increasing passive fixation reliability (Mondand
Sloman, 1980). Various tined fixation implementations currently exhibit the most
passive fixation clinical success, exhibiting dislodgement rates averaging between
3~4%. These designs typically ~include three or more tines in a variety of
configurations such as helically woundaround the electrode body, symmetrically
spaced in one or more rows spanning the length of the electrode body,, and
others.
.Figure 6.12 showsvarious examplesof active and passive f’Lxation electrodes.
6.2 LEADS
Th~ pacemaker lead often fulfills two roles: (1) delive~ng stimulation pukes
from the pulse generator to the electrode and, if implelndnted, (2) delivering
electrogram signals sensed by the electrode.to the pulse generator ....
Figure 6.13 shows two general types of modempacemaker leads: unlpoiar
and bipo!ar.~ Unipolar designs only require one lead conductor (knownas a coil).
The stimulating cathode is attached to the distal end of the lead. The pacemaker
casing is often utilized as the anode. An advantage of unipolar implementations
includes simple, single coil technology less prone to clinical and manufacturing
difficulties. In addition, unipolar leads are typically thinner than bipolar designs
(because they have only one coil as opposed to bipolar’s two coils). Unipolar
leads sometimesappear to ,induce less inflammation .than bipolar models because
bipolar leads are stiffer and result in increased pressure on cardiac tissue
(Cameronet al., 1990; Jacobs et al., 1993). In addition, lead compression damage
attributed.to medial subclavian caudal traction is commonlyreduced in unlpolar
leads because of increased flexibility(Cameron et al., 1990; Jacobs et al., 1993;
Magneyet al., 1993).
Ventdcular Ventdcular
Unipolar Bipolar
Figure6.13 Unipolarandbipolar implementationsof both J-shapedandnonpreshaped leads.
All modelshavedistal cathode.Bipolardesignstypically havea ring anodeproximal10-15nun
onthe lead.
150 DESIGN OF CARDIAC PACEMAKERS
Bipolar leads also have distal cathodal electrodes. In addition, each also
typically has a ring anode that floats in the heart cavity proximal on the lead. The
distance between these two electrodes varies by lead model from approximately
10-15’ram. Bipolar leads offer several potential advantages compared to
unipolar designs; including: reduction of far-field potential amplitudes (Aubert et
al., 1986; DeCaprioet al., 1977; Gdff’m, 1983), relative immunityto external
interference and myopotentials (Antoniucci et al., 1981; Breivik et al., 1983;
Daley and White, 1982; Levine et al., 1982; Levine and Klein, 1983; Secemskyet
al., 1982), signal-to-noise ratio improvement(Aubertet al., 1986; DeCaprio
al., 1977; Griffin, 1983), and decreased skeletal muscle stimulation (Cameron
al., 1990).
A physician typically decides whichtype of lead is, inthe best clinical interest
ofhis’or her patient (Hayes, 1992). Due to lifestyle variances, different lead
configurations are implanted in different people. Clinical advantages of bipolar
leads commonlyoutweigh the advantages offered by unipolar designs~(Hayes,
1992). Bipolar leads have been made even more clinically desirable as
technological advancements have madethem thinner and more flexible. In 1989,
76% of nonsurgeons and 60% of surgeons preferred bipolar configurations
(Bernstein and Parsonnet, 1989). However, pacing and sensing differences
between the two do not dramatically differ. ~
RL = pL (6.15)
ac
wherep is the coil material’s resistiVity, L the coil filament’s total length, and ac
the coil filament’s cross-sectional area. Onewaylead resistance can be lowered is
by shortening its length. However, lead length is determined primarily by the
physical characteristics of the patient, Especially in children, extra lead length
must be coiled somewhere to accommodate both bodily ~ flexion and physical
growth. In child implantation cases, researchers are nowable to determine within
ELECTRODES, LEADS, AND BIOCOMPATIBILITY 151
95%accuracy howtall a child patient will become. Thus, excess lead allocation
can be made(O’Sullivan et al., 1993). For both children and. adults, patients are
conimonly requested to maximally inhale, exhale, and exertbodily movements
typical, to their lifestyles. Fromboth expected height information and allowance
required for physical movement,required lead length can be determined. While
extra length can be coiled behind the pulse generator itself, considerable bulk
added tothe implant resulting in decreased cosmetic desirability and increased
risk of lead extrusion make,this a less attractive option (O’Sullivan et al.. 1993).
Excess lead .length is now commonly looped in the right atrium
(O’Sullivan et al., 1993).
Another method of lowering lead resistance is by choosing coil material
having low resistivity. Lead coils are commonlymanufactured from cobalt-based
alloys such as MP35N(35% Co, 35% Ni, 20%Cr, 10%Mo) having silver-filled
cores due to theirextreme flexibilities and low resistivities. - In .addition, they are
not difficult to manufactureor consistently coil (Cameronet al., 1990).
to200
no
the
Lead testing
determine if lead insulation and joints can withstand chemical conditions similar
to those found in the body.
After most lead tests have been performed, electron microscope scans (EMS)
are routinely performed to determine coil integrity. Figure 6.14 shows two
examples of coil deformation commonlyattributed to compression induced by the
scissoring effect betweena patient’s clavicle and first fib upon an implanted lead
(Brinker et al., 1991; Stokes and McVenes, 1988; Stokes et al., 1987). The
procedure used to implant leads experiencing such deformations is. knownas the
percutaneous subclavian vein approach. This procedure has recently accounted
for between 75-95%of pacemaker lead implantations (Bernstcin and.Parsonnet,
1989). While clinical ease and speed of implantation have. accounted for the
popularity of this procedure (Hess et al., 1982; Jacobs et al., 1993), increase
coil fracture occurrence, attributed to repeated scissoring compression have
required evaluation of b0th the implantation procedure and lead design (Alt et al.,
1987; Luck and Pae, 1991).
(a) (b)
Figure 6,14 Examplesof compressed leads. (a) Compression
damageto soft wire coil after
single applicationof compression
in medialsubclavianimplantation,(b) Coilfracture morphology
associatedwithrepetitive andcompressivescissoringbetween
a patientYsclavicle andfirst fib.
FromJacobs, D. l~I., Fink, A. S., Miller~R. P., Anderson,
W.R., McVen~s, R. D., Lessar, J.
F., Cobian,K. E., Staffanson,D. B., Upton,J. E., andBubrick,M. P. 1993.Anatomicaland
morphological evaluationof pacemaker lead compression.PACE,16: 434-444.
Lead stiffness
often has a pair of coaxial coils comprising the remainder of its proximal length.
This increased bulk can significantly increase lead stiffness. Intermedics has
introduced a thin bipolar lead having greater flexibility than observed in many
other bipolar designs (Adler et al., 1992). Figure 6.15 shows both traditional
coaxial and Intermedics ThinLine®lead models.
Lead insulation
6.3- BIOCOMPATIBILITY
----~
Fibrosis~
Corrosion and degradation ~
Inert, noncorrosive
Products of abrasion~ v materials .~
Infectioti~~ Atoxic matei’ials
-’’-~s.........../
Metabolicchange " ¯
There are fo~ general classes of biornaterials: metals, polymers, gl~ses, and
cei~cs. There are also composites of these materialS. All are knownas b~ing
al,10plastic, meaning not biologicalin origin. The chemical bonds holding each
ni~t~dal together geiiera!ly dete~e implantation Utility. - ~
Metals
Polymers
Glasses and ceramics are extremely hard and exhibit desirable properties relevant
to~,~hermal, coefficients of expansion, specific heats, insulation, and smoothness.
Tl!e~e ~e several pacemaker uses for glasses and ceramics.i One use includes
ee~c encapsulauon of the pulse generator for protection. Th~s sealant layer ~s
ex~mely smooth and very soft tissue biocompatible. Ceramics or glasses are. also
sometimes activated with metals or vitreous carbon to produce high active surface
areas desirable for electrode tips (Katsumoto et al., 1986; Mundet al, I986;
geti~d~ich, 1992). Due to absorption characteristics, eerie collars surrounding
electrode tips have been used for containment.and elution.of steroids into
surrounding tissue to minimize inflammation and collagenous accumulation
~iii:i~rson et aL, 1990; Anderson et al., 1991; Mathivatlar et al., 1990;
~i~sky et al., 1990; Wilson et al., 199I)~ Lastly, glass is commonlyused to both
seal=pulse generator can entry and comprise the connector block where lead
eon~eetion occurs.
6.4 REFERENCES
Adler, S:C., Foster, A. J., Sanders;: R. S:, and Wuu,E. 1992. Thinbipolar loads: a solution to
problemswith coaxial bipolardesigns. PACE,15: 1986-1990.
Adler, S. C., Spehr, P., Allen, J., and Block, W. 1990. Chronic animfl testing ofneweardiac
pacing electrodes. PACE,13: 1896-1900.
Alt, E., Volker, R., and Blomer, H. 1987. Lead fracture in pacemakerpatients. Thorac.
:. Cardiovasc.Surg., 35: 101--104. . ~~
Anderson, N., Mathivanar; R., and Skalsky, M. 1990. Reduction of threshold peaking and
chronic thresholds using a ceramicdrug eluting collar. (abstract) PACE,12: 108,
Anderson,N., Skalsky, M.~Mathivanar, R., Tunstelli A., Harman,D., and Ng, M, 199L-Active
fixation leads--long term threshold reductionusing a drug-infusedceramiccollar.- (abstract)
PACE,14: 639.
Antoniucci, D., Marehi, F., and Multinu, D. 1981. Musclepotential interference: a study with
ambulatory ECGmonitoring. PACE,4: A-30,
Aubert; A, E., Ect0r, H., Denys, B. G., and~,de,Geest, H~1986. Sensing characteristics of
unipolar and bipolar orthogonalfloating electrodes: morphology and spectral analysis. PACE,
9: 343.
Bemstein,A. D.; andParsonnet,V. 1989. Surveyof cardiac pacing in the United States in 1,989.
Am.J. Cardiol., 69: ,331~338. "~
Beyersdorf,F., Schneider,M., Kreuzer,J., Falk, S., Zegelman,M~,and Satter, P, 1988.: Studies
of the tissue reaction inducedby transvenouspacemaker electrodes. :I. microscopicexamination
of the extent of connectivetissue aroundthe electrode.tip in the humanright ventricle. PACE,
1~: 1753-1759.
Billmeyer, F. W. 1984. Textbook ofpolymer science. 2nd Ed. NewYork: John Wiley&Sons.
Bittence, J. C. 1983.Mate:rialsengineering/materials selector. Cleveland:PentOn/IPC.
Bockris, J. O’M,and Drazic, D. M. 1972. Electro-chemicalscience..London::Taylorand Francis,
~Ltd.
Bolz, A., FrOhtich; R., and--Schaldach, M. 1993. Elektrochemisehe aspekte der
elektrostimulation:-ein beitrag zur~senkungdes energiebedarfs. In M. Hubmann and R. Hardt
(eds.) Schrittmachertherapie und ~ik. Mflnehen: MMVvedag..
BOretos,J. W., anti’Eden, M: 1984~Contemporary biOmaterials. Park Ridge~:NJ:Noyes.
Breivik, K.; Engedal, H,~ and Ohm,O. J. 1983, .Long-termcomparisonof unipolar and bipolar
pacing and sensing, using a new multiprogrammable pacemakersystem. PACE,6: 593.
158 DESIGN OF CA~IAC PACEMAKERS
Brewer, G., Mathivanar, R., Skalsky, M., and Anderson,N. 1988.~ Compositetips containing
externally placed drug releasing collars. PACE,11: 1760-1769.
Brinker, J. A., Zimmem, S., and Gentzler, R. 1991. Coaxialbipolar leads---potential for internal
insulation problem.(abstrae0 PACE,14: 85.
Brown,S. A. 1988. Biomaterials, corrosion and wear of corrosion. In J. G. Webster (ed.)
Encyclopediaof medical devices and instrumentation. NewYork: John Wiley&Sons.
Cameron,J., Mond,H., Ciddor, G., Harper, K., and McKie,J. 1990. Stiffness of.the distal tip
of bipolar pacing leads. PACE,13: 1915-1920.
Charles, R. G., Clarke, L. M., and Drysdale, M. 1977. Endocardial pacing electrode design and
rate of dislodgement.Br. Heart. Z, 515.
Daley~ J. E., and White, A. A. 1982. Non-invasive analysis of simulated pacemakerfailure
available in multiprograrnmable pulse generators. PACE,5: 4.
DeCaprio, V., Hurzeler, P., and Furman, S. 1977. A comparison of unipolar and bipolar
electrograms for cardiac pacemakersensing. Ciwulation, 56: 750. ’
Deconinck,J. 1992. Current distributions and electrode shape changesin electrochemicalsystems.
In C. A. Brebbia and S. A. Orszag(eds.) Lecture notes in engineering. Berlin: Springer-
Verlag.
Djordjevie, M., Stojanov, P., Velimirovic, D., and Kocovic,D. 1986. Target lead’low threshold
electrode. PACE,9: 1206--1210.
Dymond, A. M. 1976. Characteristics 0f the metal-tissue interface of stimulation electrodes. 1EEE"
Trans. Biomed. Eng., ~ 23: 274,:
Fraker, A. C., and Griffin, C. D, 1985. Corrosion,anddegradation of implant materials. Second
international symposiumon corrosion and gradation of implant materials. Philadelphia:
ASTM, Bd. ~
Fraker, A.C., Ruff, A. W., Sung, P., van Orden, A. C., and Speck, K. M. 1980. Surface
preparation and corrosion behaviourof titanium alloys for surgical implants. Ti ’80 Science
and Technology, 2447.
Gillette, P. C,, Wampler,D. G., and Shannon, C. 1985. Use of atrial pacing in a young
population. PACE,8: 94-100.
Griffin, J. C. 1983. Sensingcharacteristics of the right a~al appendageelectrode. PACE,6: 22.
Hayes, D. L. 1992. Pacemakerpolarity configuration--what is best for the patient? PACE,
15:1099-1100.
Henson,P./v12. 197LThe immunologicrelease of constituents from neutrophil leukocytes II.
Mechanismof release duringphagocytosis and adherence ~o non-phagocytosable surfaces.
J. lmmunol., 107: 1547-1557,
Henson,P. M. 1980. Mechanismsof exocytosis in phagocytic inflammatorycells. Am~J:Path.,
101: 494-514.
Hess, D. S., Gertz, E. W., and Morady, F. 1982. Permanentpacemaker implantation in the
cardiac catheterization laboratory: the subctavianvein approach.Cathet. Cardiovasc.Diagr~,
8: 453-458.
Irnich, W~1973; Considerationsin electrode design for permanentpacing~In H. J. Thalen(ed.)
Cardiacpacing; proceedingsof the fourth international symposiumon cardiac pacing. Assen,
The Netherlands: Van Goreum& Co.
Irnieh, W. 1975. Engineeringeonceptsof pacemakerelectrodes. In M. Sehaldachand S. Furman
(eds.) Advancesin pacemakertechnology. NewYork: Springer-Verlag.
Jacobs, D. M., Fink, A. S., Miller, R. P., Anderson, W.:R., MeVeues,R. D., Lessar, J. F.,
Cobian, K. E.,, Staffanson,D. B., Upton, J. E., and Bubriek, M. P. 1993. Anatomicaland
morpbologiealevaluation of pacemakerlead compression. PACE,16: 434-444.
Karpawieh~P. P.; Hakimi, M., Arciniegas, E~, and Cavitt, D. L.. 1992. Improvedchronic
epicardial pacing in children: steroid contribution to porous platinized electrodes. PACE,
15:1.151-1157.
Katsumoto, IC, Niibori, T., Takamatsu,~ T.,’ and Kalbara, M. 1986. Developmentof glassy
carbon electrode(dead sea scroll) for low energy.cardiac pacing, PACE,9; 1220~-1229.
Levine, P. A., Caplan,C. H., and Klein, M. D. 1982. Myopotentialinhibition of unipolar lithium
pacemakers. Chest, 82: 461.
Levine,P. A., and Klein, M.. D..1983.Myopotentialinhibition of unipolar pacernakers:~a disease
of technologic progress. Ann,Intern. Med., 98: 101. ~
Llewellyn,M., Bennett, D., Heaps,C., and Slaven, Y. 1988. Limitation of early pacing threshold
rise using a silicone insulated, platinised, st.eroid-eluting lead. (abstract)PACE,
11: 496.
Luck, L C.; and Pae, W. E. 1991:. Pacemaker. complication. In J. A..Waldhausen (ed.)
Complicationsin cardiac surgery..St: Lonis:~Mosby Year~.Book. , ~
Magney,J. E., Flynn, D. M., Parsons, L A., Staplin,.D.H4 Chin-Purcell,M. V., Milstein~,S.,
and Hunter, D. W~1993.Anatomical mechanismsexplaining damageto pacemaker leads~
r ELECTRODES, LEADS, AND BIOCOMPATIBILITY 159
Stokes, K. B., and Stephenson, N. 1982. The implantable cardiac pacing lead--just a simple
wire? In S. S. Barold and J. Mugica (eds.) The third decade of cardiac pacing.
Mount Kisco, NY: Futura Publishing.
Szycher, M. 1983. Biocompatible polymers, metals, and composites. Lancaster: Technomic
PublishingCo., Inc.
Tarjan, P. P., and Gold, R. D. 1988. Implantable medical electrical devices. In J. Kline (ed.)
Handbookof biomedical engineering. NewYork: AcademicPress.
Walton, C., Gergely, S., and Economides,A. P. 1987. Platinum pacemakerelectrodes: origins
and effects of the electrode-tissue interface impedance.PACE,10: 87-99.
Williams,D. F. 1987. Definitions in biomaterials. Oxford:Elsevier SciencePublishers.
Williams, W.G., Hesslien, P. S., and Kormos,R. 1986. Exit block in children with pacemakers.
Clin. Prog. Electrophysiol. Pacing,4(5): 478-489.
Wilson, A., Kay, N., Padeletti, L., Michelucci, A., Ferri, F., Baird, D., Mathivanar, R., and
Skalsky, M. 1991. A multicentre study of steroid eluting collar leads. (abstract) PACE,
14: 629.
Zeidler, D. E. 1977. Stress fractures in pacing leads. Medtronicnews, 7:8-11.
6.1 State the twomost common electrical roles of a pacemakerelectrode. Explainbriefly which
is mostimportant~to extendingpacemakerduration of service and why.
6.2 Excludingelectrical roles requestedin Instructional Objective6.1, list six considerations
importantto electrode designphysically. Explainbriefly whyeachis important~
6.3 Sketch the fwst order approximationof the Helmholtzdouble layer formedby metal being
submersedin electrolyte. Includeelectrical schematicmodelcorrelated to ~xhibitedphysical
characteristics of this doublelayer~ Whyis metal-oxideformationdesirable?
6.4 Explainwhyreversible reactions are preferable to irreversible.reactions at the electrode-
electrolyte interface. Explainbriefly twopossiblq waysthat reversible chargetransfer (RCT)
is obtainedat this interface. .
6.5 Discussthe.origin of Faradiccurrent. Whyis Faradiccurrent undesirableS.
6.6 Explain whythe suppression of electrode induced inflammationis important to decreasing
both acute and chronic stimulation threshold change. Explain briefly howsteroid-eluting
designsare theorizedto accomplishthis.
6.7 Describethe difference betweenelectrode tip macroscopic size andmicroscopicsurface area~
Explainwhythe difference is importantto electrode tip design.
6.8 Explain whyporosity is important to electrode tip design. Discuss the theory behind its
Success.
6.9 List two general advantagesto active helical fixation,, methods.Withrespect to pacing
efficiency, state a disadvantageassociatedwith general active fixation methods.
6.10 For a general ventricular electrode implantation, explain whyapassive tined electrode may
be preferableto an active hefical tip implantation.
6.11 Sketchboth unlpolar and bipolar types of leads. List twoadvantagesfor each.
6.12 Whatis the difference betweenbiocompafibilityand biostability? Whyis each important?
6.13 List four uses for metals in pacemakertechnology.Explain whytitanium and its alloys are
so effectivein their respectiveapplications.
6.14 Discuss two primary concerns associated with implanting polymersin the body. List the
advantages polymershave over other biomaterials. List two. polymerscommonly used in
pacemaker technologyand their respective uses.
Battery
!0~ G. Webster
7,1 MATERIALS
~l!l’Early development
The first implant in 1958 was, designed by Elmquist and used a rechargeable
(secondary) nickel-cadmium battery. It was inductively recharged by the
transmissionof energy to the implanted receiver. The cell.voltage was 1.25 V and
the capacity was 190 mAh.It was unsatisfactory because thefifetime was not. long
enough compared to a (primary) nonrechargeable battery. Also, it was not
satisfactory to place the responsibility for recharging in the hands of patients~ who
in manycases were senile.
Z~n~c~.~mercu~,
batteries
In 1958-1960, the zinc-mercury battery ~ was modified for longer shelf life for Use
in electric watches. For pacemakers,three to six cells in series provided 4-8 V. It
became widely used in pacemakers, typically cast in epoxy, which was porous to
the evolved hydrogen and permitted its dissipation. However, even with
improvements,by 1970, longevity was only about two years.
Biological batteries
The concept of using power from within the body is attractive, Somehave tri~
biogalvanic cells in whichdissimilar electrodes and the body’s electrolytes yield
power. Others have tried fuel cell’s that utilize oxygenfrom the blood and hydrogen
162 DESIGN OF CARDIAC PACEMAKERS
from body proteins or glucose from the blood. Still others have converted body
movementthrough piezoelectric generators or self-winding watch escarpments.
Noneof these has proved practical.
Nuclear batteries
Fabrication
Since 1972, a variety of lithium batteries have been used. These include Li/SOCI2,
Li/Ag2CrO4, Li/CuS, Li/I2-Polyvinylpyridine (PVP), and, in more limited use,
Li/LiI(A12)3/PbI2,PbS, Pb. The Li/I2-PVP is the principal pacemaker b.a~e~: and
yields 2.8 V (Schaldach, 1992).
The cathode is a complexof iodine and poly-2-vinyl pyridine (P2VP). Neither
conducts electricity, but whenmixed and heated at 1497Cfor 3 days, they react
into a black viscous paste that conducts electricity. This is poured into the battery
whenmolten and cools to form a solid. Whenthis paste contacts metallic lithium, a
monomolecular layer ~of crystalline lithium iodine forms. It is a molecular
semiconductorthat passes lithium ions, as required, for current flow, but not iodine
molecules (Greatbatch and Seligman, 1988).
Reactions
Figure 7.1 shows that conventional current flows through a device from anode to
cathode. For a battery, the current flows from the negative~anode,~through~the
batte~, to the positive cathode. Oxidation of metal occurs at the anode,
t
12 + 2e- -~ 21-
R.~.uction
Load anode
Current
Resistance
~’ UNCOATEO
/
160OO
/
12OOO / /
8OOO
4OOO
0
o 200 600 ~10001400’1800 ’2200 26003000
Capacity(mAh)
, Figure. 7.3, phase I showsthat the voltage starts to decrease. In phase lI, th
dathode becomes S~ed .of iodine and de~elops a~ higher impedancb’tli~/he.,LiI
electrolyte, Causing the shoulder region of Figure 7.3. During phase"III; He
164 DESIGN OF CARDIAC PACEMAKERS
VDC
3.0V
2.8V
iodine in P2VP
matrix
3 Ah
7.2 MANUFACTURING
7.2.1 Construction
Figure 7.4 shows the internal construction of a central anode/case grounded cell
with corrugated anode. Lithium is easily formedinto sheets that can be cut to the
required sizes. It is easily pressed ~to specific anode shapes. The lithium anode is
coated three times with a solution of PVP. The solvent is evaporated to leave a
contiguous film of pure PVPon the anode surface. The precoated central lithium
anode is corrugated to increase its area ~and lower battery impedance. To obtain
lower impedance, newer designs use more concentrated active materials and larger
~n_heodesurface
complex ~as~,. Multiple
ofi0dine anode surfaces
and l~ly-2-vinyl pyritemay:be:used
~P2VP)is~t0pod
lower.~b_e
into theimpedance.
cath6de
c~e to cool
..... " dohe
bye fe through
or ceramicinsulator.
BAT~RY 165
GONNEGTION
TO CATHODE GLASS-TO-METAL
SEAL
(CASE)
CENTRAL
ANOOE
CATHOOE
7.2.1 Testing
Nondestructive examinations
Thermal cycling
High pressure
Mechanio.~lvibration
Temperature/humidity
Mechanical shock
Voltage/temperature
Seal terminal strength
Elevated temperature discharge
Destructive analysis
Solvent resistance
73;~ Ratechanges
As the battery of, a pacemakerprog ~ressively discharges, the internal resistance can
increase to the point that inadequate charg~gof the ou~ut capacitor occurs, If the
166 DESIGN OF CARDIAC PACEMAKERS
capacitor fails to store enough charge between pulses, then there is the chance a
stimulus pulse delivered could fail to cause contraction of the heart. Therefore, it is
wise to ensure that the stimulus amplitude is always above threshold.
Figure 11.6 shows one method for assuring adequate output amplitude
Elmovist (1984). This circuit releases stimulation pulses only if a specific minimal
voltage Vminis reached. Every pulse delivered will be above threshold, and the
heart is certain to be stimulated.
Battery
I
I Vb
i Monostable
one
t~0.8 ms
Mt~,OonOStable
t~,O.5,~
~,,~
ms ~ i -..--.I
7 J’-°
"
Ana!£gor, I ’, "" I
I cD~gimt~la~ator ~ivmin ~ "
(b)
Ri
Figure 11.6(a) shows that starting with both monostables low, the capacitor
charges from the battery through battery resistance Ri and a,~f~ed r, esis~tor. The
output capacitor is connected to a comparator which compares the capacitor
voltage value with a set minimalvoltage Vmin.If the voltage at the capacitor is
below Vmin,the comparator output is_ high and a logic J is delivered to the~:~OR
gate, resulting in an output from the NORgate whichis low. Oncethe, voltage .at
the output capacitor is greater than Vmin,the comparator output goes ~low;:.which
delivers a logic 0 to the NORgate. Assumingmonostable one is already low, the
output of the NORgate becomeshigh. This activates monostable two’,~,whi6hYliff
active for approximately 500 Its, and subsequently switches the transistor into its
Frost (1986) designed a circuit that indicates depletion of the battery prior to rate
changes. It momentarilyswitches the battery to a test capacitor. Figure 7,7 shows
that pacemaker 12is normally powered£rom battery, VC. Switch S1 is opened
momentarily and the pacemaker continues to be powered by filter capacitor CB,
Switch $3 has discharged test capacitor CTand opens. Switch $2 closes and CT
~hearges wi~’,from
termined a time constant z = RCT,
the Charging ,The internal
charaCteriStics resistance of the battery
shown in Figure 17.8 andR is
is
telemetered from the body.
VOLTAGE
AT
VBA VDD
T
VREF
St CLOSED
S2 OPEN
$3 CLOSE{
TIME
S3 OPENS
SI OPENS $2 OPENS
SI CLOSES
sz cLos~"
COMPARATOR
OUTPUT VDD
TIME
vssI
~.i~, ~ 7.8 VDDis common.When$3 opens, VBATgoes tO -Vc because voltage drop throughR
decreases to zero. When$2 closes, VBATis clamped to VDD,then exponentially charges through
R toward ~-Vc. Comparator14 detects time toxeach VREF,~whichis a function of R (Frost, 1986).
6
LI
["-,~X-lO0 ms -I
Figure 7.9 After a stimulating pulse, a smaller nonstimulating marking pulse is placed at time
X. 100 ms, where X is remaining battery capacity in years (Moherg, 1987)..
7.4 POWERPRIORITY
Pless and Stotts (1986) designed a priority switching circuit for extending
pacemakerlife while providing a minimalvoltage to a voltage-sensitive load. The
pa~makeroutput circuit requires a large current to recharge the output capacitor
between pulses~ This current flows through the battery internal resistance and
~reases the battery output voltage. This temporary decreased voltage can cause
~b~eiitially dangerous intermittent malfunction of the control circuit and
corresponding erratic operation of the pacemaker.
Figure 7.11 showsa powerdistribution controller that selectively switches the
battery betweenthe output circuit and the control circuit. A hold-up capacitor Chold
is connected in parallel with the control circuit in order to maintain at least a
minimal operating voltage Vminwhenthebattery is disconnected fromthecontrol
circuit andconnected topower theoutput circuit.
Figure 7.12shows thatwhenthebattery isnearendoflifeandRbatt
the.power distributioncontroller mostefficiently distributes thepower~ to maintain
controller voltagc~Figure7.13gives typical d~...ta fora power prioritysystem.
CONTROL
CIRCUITRY !
Figure
7.11
Whenvoltageon
theholdcapacitor
decreases
totheminimum,
switch
S1isclosed
torecharge
itandswitch
$2opened.
Tocharge
thepulse
output
capacitor,
switch
S I isopened
andswitch
$2closed
(Pless
andStotts,
1986).
7.5 REFERENCES
R~B~t’~IK
A
Tl-l"iml
8.1 REQUIREMENTS
Figure 8.2 illustrates unipolar and bipolar sensing and unipolar and bipolar
electrode systems,~ The.unipolar ~ease-utilizes a~ single-ended-amplifier whose
reference-is~thepacemaker case and a single dectrode lead.~hich~terminates in the
electrode tip ~whiehisin.eleetfiCal contact, with the~myocardium~Thebipolar
system utilizes a differential amplifier and a dual-conductor lead which terminates
DESIGN OF CARDIAC PACEMAKERS
in the tip and ring electrodes. The unipolar pacemakersense amplifier responds to
electrical potential differences whichappear betweenthe tip electrode and the case
of the pacemaker; the bipolar pacemaker responds to electrical potential
differences betweenthe tip and ring electrodes.
Blanking
Gaincontrol
Variable gain
Window ~
amplifier and comparator
bandpass filter
, -sense
Blanking
circuits Voltage
reference
’ " Telemetry [._~
To telemetry circuits
I~ amplifier /
Pacing/s
lead(s)
¯ Ring
Tip c
Lead [- I ’~
Case~’~"( ~_~~
Unipolar Bipolar
/3 IIII s-r
III
-4"
II
-6" I
0 200 400 800
Time, ms
¯ Ne
I atlvel ,-intleC :eo I t lave
2
~ o
-2
-6
0 200 .400 600 800
Time,ms....
Figure 8.4 Chronic unipolar right-ventficular intracardiae electrogram. Note the absence of
current of injury. (Adapted from Furman, S., et al., 1977a. The ventricular endocardial
Mosby Year Book Inc. Used by permission).
,electrogram and pacemaker sensing. J, Cardiovasc. Thorac. Surg. 73: 258-266. Copyright
. ~
The electrograms of Figures 8.3 and 8.4 are of the biphasic type (the intrinsic
deflection excursions are both positive and negative with respect to the isoelectric
baseline), but considerable variation exists in polarity and morphology.Furmanet
al. (1977b) state that 58%of acute unipolar ventricular electrograms are biphasic,
while 30% are monophasic negative and 12%are monophasic positive. The same
reference gives the characteristics of the chronic ventricular electrogram as 68%
biphasic and 32% monophasic negative with no S-T segment elevation. The
chronic electrogram shows amplitudes comparable to those of the acute
electrogram but somedecrease in slew rate (the rate of maximalvoltage change in
the intrinsic deflection) was noted as the tissue-electrode interface aged. A later
study by Furman(1993) reports Similar data.
The unipolar atrial electrogram is morphologicallysimilar to the ventricular
electrogram; 89%of atrial electrograms were found to be biphasic and 11%were
monophasicnegative with little change in the intraatrial signal as the electrode
matured (Furmanet al., 1977b). Furman(1993) states that 14%of acute intraatrial
electrograms showcurrent of injury. An exampleof the atrial electrogram is shown
in Figure 8.5.
The studies of Furmanet al. (1977a, 1977b) found wide variations in the
characteristics of the intracardiac electrogram. The amplitudes of the acute
intraventricular electrograms ranged from 2.0 mVto 36.4 mVwith a mean of
12.4 mV. Chronic intraventricular el~dtr0grams ranged from 1.2 mVto 26 mV
with a meanof 10.5 inV. Meanslew rates for intraventricular electrograms were
1;5 mVlms,for acute:,electrograms and 0.9 mV/ms:for chronic electrograms;
Amplitudes: of intraatrial elec~rograms were typically smaller than th6se of the
intr~ventricular electrogram (a meanvalue of 4.8i mVwasl,reportea). ¯
PACEMAKER SENSE AMPLIFIERS 175
O 0 ~ ""-’~ ~
Ventrl:ular( ~pola~
~zatioi~
s~er than~that of theventricular electrograms of Figures 8:2 and 8.3. The crosstalk from the
far-field = ventricular potential is also apparent. (Adapted from Furman, S., Hurzeler, P., and
DeCaprio, V. 1977b. Cardiac pacing and pacemakers HI. Sensing the cardiac electrogrmn. Am.
Heart J. 93:794-801. Copyright MosbyYear Book Inc. Used by permission).
with electrodes of fLxed area (11 ram2). The P waves were mostly biphasic (similar
to Figure 8.5); an unusually-long duration P wave was recorded from a patient
manifesting complete atrial-ventricular dissociation. The Pwavesrecorded in this
study ranged from a mean of 4.92 mV(representing 20 cases spanning a range
from 3.1 mVto 8.1 mV)with a terminating resistance Of 100 k.Q to a meanof
3.3 mV(representing 5 cases spanning a range from 1.8 mVto 4.9 mV)with
terminating resistance of 2 k.Q. The meanslew rates were observed to fall with
decreasing terminating resistance (from 0.915 mV/msfor a terminating resistance
of 100 k.Q to 0.63 mV/msfor a load resistance of 2 k.Q).
Irnich (1985) presented a modelfor the genesis of the intracardiac electrogram
in which the spread of depolarization in the myocardiumis viewed as the uniform
linear motion along the myocardial wall of a pair of orthogonal electric dipoles.
The longitudinal dipole is oriented parallel to the myocardial wall while the
transverse dipole is normal to the myocardial wall. The total potential at any
intracardiac observation point is the superposition of the potentials due to each
individual dipole. The unipolar intracardiac electrogram is represented as the time-
varying potential recorded at an intracardiac observation point relative to a
reference (presumedto beat infinite distance) as the dipole pair movesalong the
myocardial wall. The bipolar electrogram is represented as the difference between
the time-varying potential at two intr~icardiac observation points. This simple
model successfully explains the clinical observation that some intracardiac
electrograms exhibit .monophasic morphology while others exhibit biphasic
morphology;the motion of the longitudinal dipole past the observation point will
result in a biphasic morphology,whereas the motion of the transverse dipole will
result in a monophasic morphology. The observed morphology depends upon
which dipole is predominant. The model is somewhatless successful in predicting
signal amplitude as a function of electrode area; the modelpredicts a reduction in
signal with increasing electrode area, whereas Myers et ai. (1978) found the
converse. Experimental results cited in this study found 60% of 133 acute
intraventrieular electrograms were~biphasic while 30% of 20 chronic
intraventricular ~Iectrogra~ns were biphasic. MonophasiC positive intra~;entricular
morphologieswere not~xt only in case~ in which myocardial injury was present. All
acute intraatrial electrograms(7 cases) were biphasie.
Someof the studies previously cited also examined bipolar electrograms.
Furman et al. (1977b) compared the tip unipolar signals and bipolar signals
obtained simultaneously from the same intraventricular lead and found little
difference in either meanintrinsic deflection amplitude or slew rate. This study
found that 51%of the bipolar electrogram intrinsic deflections were smaller than
those of the simultaneous unipolar electrograms; the bipolar electrogram
amplitudes exceeded their simultaneous unipolar amplitudes i~ 43%of the cases.
Unipolar and bipolar amplitudes were found to be equivalent’ in the remainder of
the cases. It was reported that 2%of the cases manifested bipolar electrograms too
small to be sensed while the simultaneous unipolar electrogram was of adequate
amplitude for sensing. This finding appears to confirm the theoretical possibility
that the spreading myocardial depolarization could produce similar potentials at
each electrode of .a bipolar .electrode pair, .resulting in little or no po[ential
difference. Irnich (1985), however, claimed to have never encountered inadequate
bipolar sensing in Clinical practice; this agrees with the experience of Det~viler
(1 ~94). ’Parsonnet t al. ( 19~0)presented bipolar intraatrial .d ecti:ogram data fr
two cases with load resistances of 100 k.Q and 20 k~. The amplitudes were 4.4 mV
and 9.1 mVfora load resistance of’l~ k~; the figures for a terminating resistance
PACEMAKERSENSE AMPLIFIERS 177
of 20 1~ were 4.3 mVand 6.2 mV. Measured slew rates were 0.92 mV/msand
1.9 mV/ms for a terminating resistance of 100 kf~ and 0.82 mV/ms and
1.31 mV/msfor 20 1~. The principal benefit of bipolar sensing was the reduction
in the effect of far-field signal sources; Parsonnet et al. (1980) noted that
ventdcular crosstalk in the intraatrial electrogram was completely attenuated by
bipolar sensing without concomitantsacrifice in atrial signal. Furmanet al. (1977b)
reported that canine studies had shown’thatthe atrial bipolar endocardial electrodes
reduced ventricular potentials sensed in the atrium by 80%relative to the
corresponding simultaneous unipolar electrogram.
The results cited in the preceding paragraphs were obtained with endocardial
electrodes. Furmanet al. (1977b) reported observations of chronic right- and left-
ventricular epicardial electrograms; the meanleft-ventricular amplitude in 9 cases
was 18.0 mVwith a meanslew rate of 1.85 mV/ms,while the corresponding right-
ventricular results in 14 cases were 11.24 mVand 1.09 mV/ms.
Wesee from the studies cited above that intracardiac e!ectrograms exhibit
wide variation in amplitude, slew rate, and morphology;successful sensing over
such a wide range of signal characteristics represents a significant design
challenge. The desirability of pacemaker.devices whosesensing functions maybe
programmed in situ (to customize the pacemakerto the patient) is apparent.
Irnich (1984, 1985) indicated that widely-reported data concerning the spectral
.c09tent of signals encounteredby pacemakerswere erroneous..~Th,e figures cited by
Irnic’h indicate that T waves contain energy at frequencies beloW 10 HZ while
R waves contain components eXtending from 20 Hz to 45;Hz; the frequency
content of P waves extends from 50 Hz to 90 Hz, while that of electromyographic
signals extends from 80Hzto 1 kHz. Schaldach (1992) appears-to confirm these
figures, citing design frequencies of maximalresponse of 70 Hz fo~’atdal sensing
and 40 Hz for ventricular sensiag with asymmetric 10w- and high-frequency
responses (four poles of high-pass faltering and two poles of low-pass filtering,
providing a sharp rolloff at low frequencies to reject repolarization potentials).
Other studies, however, confirm lrnich’s claim that these data are incorrect.
Kleinert et al. (1979) analyzed the spectral energy density of intracardiac signals
and found maximaof approximately 30 Hz for the intraatrial sigaal and 20 Hz for
the intraventricular signal. The frequency of maximalT wave energy density was
below 3 Hz. The maximal energy density of the P wave was less than the maximal
energy density of the R wave by a factor of approximately 10, This study
concludedthat ventricular and atrial spectra are sufficiently similar that sensing
circuits of similar frequency response maybe used for both, a finding confirmed by
Olson,(1994a), Parsonnet et al. (1980) noted that the frequencies of P wavespectral
maximawere inversely related to pacemakeramplifier input resistance (rising from
approximately 25 Hz at a resistance of 100 k~ to 30 Hz with a resistance of
20 kf~). Ohmet al. (1977) concluded that the spectra of myopotential signals
endocardial signals have a significant degree of overlap.
The Fourier transform of the chronic ventricular electrogram of Figure 8.4 is
shown in..~Figure 8.6. The most prominent spectral peak is associated with
ventricular repoladzation (T wave); the next-larger spectral peak is associated with
ventricular depolarization (R wave).
178 DESIGN OF CARDIAC PACEMAKERS
wave
0 20 40 60 80 100
Freq., Hz
Pacemakersense amplifiers must be light and compact to fit within the massand
size constraints imposed by the pacemaker package. They must operate correctly
with battery voltages as low as 1.5 V and must be tolerant of battery-voltage drop
PACEMAKERSENSE AMPLIFIERS 179
as the battery ages; and they must consumeas little eloctdcal energy as possible.
The ~wide range of possible intracardiac signals ,which the pacemaker may
encounter highlights the desirability of being able tochangethe detection threshold
(sensitivity) through external programming.
WhereT(s)is the transfer function of the twin-T notch filter comprised of R1-R3
aiad~(1,7-C3; gin is the transconductance of JFETQi; hie2iS the common-emitter
input impedanceof Q2; hfe2and hfe3 are the small-signal current gains of Q2 and
Q3, respectively; and the zero and two poles are given by:
e c,
do so, we find that the twinoT notch filter has equal numbersof poles and zeros
with a pair of zeros at s = .+_jl2RC.The twin-Tcircuit exhibits unity gain for both de
and high frequencies while having a notch at a frequency of II4~RC. With the
values used by Greatbatch (R = 3001~ and-C = 0.02 ILtF), the notch frequency
approximately53 Hz. Theinclusion of the notch filter is particularly noteworthyin
that it demonstratesthat the possibility of interference from power-line sources was
being considered in pacemakeramplifier design as early as the late 1960s.
Figure 8.8 shows the results of SPICEanalysis of the circuit of Figure 8.7
using generic JFET and bipolar transistors models for Q I-Q3. The predicted
maximal gain of 36.5 dB occurs at approximately 10 Hz. Note the low-frequency
gain is approximately2.8 (9 dB), whichis close to the ratio of R6to R5. Theeffects
of the zero and the poles of the active amplifier circuit are evident, as is the notch
inserted by the twin-T network.
The linear amplifier of Figure 8.7 supplies its output voltage to a passive RC
differentiator connected to the base of a. 2N2450transistor normally biased in
cutoff. A ventricular depolarization wouldresult ina biphasic pulse at the base of
this output transistor; the negative-going portion of this pulse would turn on the
output transistor and provide a positive~going pulse at output Vo. This pulse would
inhibit the otherwise free-running oscillator whichproducesventricular-stimulation
pulses.
Schaldach and Furman (1975) give other examples of pacemaker-amplifier
design techniques in the early 1970s with diagrams of amplifier circuits from
instruments produced by Medtronic Inc., American.Optical Company,and Cordis
CorPoration, all of which utilized discrete circuit techniques like the circuit of
Figure 8.7.
PACEMAKERSENSE AMPLIFIERS 181
20
-2 -1 0 1 2 3 4 5
~og(f),
Figure8.8 Frequency
responseof the ventricular-sensingamplifierof Figure8.7 as calculated
by SPICEusing generic modelsfor QI-Q3.
The circuit of Figure 8.7 operates with fixed gain and sensing threshold. Later
developmentsincorporate circuit variations which permit adjnsmaentof sensitivity
by meansof variable-gain amplifiers or adjustable-threshold voltage comparators.
DeCote(1988) provides an example of the latter approach. A fixed-gain amplifier
provides an amplified and f’dtered cardiac signal to a pair of voltage comparators,
the ratio _of whosethreshold voltages is determinedby a resistive voltage divider. A
DAre’operating under microprocessor control sets the voltage across the divider
si~e~~the rhore sensitive comparatoris-triggeredby car~ae activity while the
less sensitive comparatoris not. A significant change in cardiac-signal amplitude
will,,ca~s e both comparators to trigger or neither to trigger, in which case the
~p~ocessor can take the appropriate corrective acti6n to fmda new s~ensing
Vd(
If the conditions of Eq. (8,2) are satisfied, the OTAused for U1 gives the same
performance~asa traditional operational amplifier. OTAsU2 and U3 .are employed
as high-gain voltage comparators; they-are operated~ open:loop anti’their outputs,
used to drive .the gates d f MOS
switch ~ansistors, ~are operating into virtually open
circuits. Threshold ~voltages for these eomparators are derived from the output
current Of Q 1 flowing through R2 and R4. Thus a commoncurrent source sets the
PACEMAKERSENSE AMPLIFIERS 183
operating point of U1, establishes the comparator thresholds, and provides bias
currents to the three OTAdevices.
+V
U2
R4 ,R5 I Ib
C2
R8
8.2.3 CMOS
operational amplifiers
Id=/doW
tu_ v_~..
T eq~,~_,T//nk (8.3)
L
Figure 8.11 shows a simple two-stage CMOS operational amplifier 0f the type
analyzed at length by Gray and Meyer(1982).
PACEMAKERSENSE AMPLIFIERS 185
Vdd
~..~.~, Figure 8.12 shows a small-signal model thereof, Analysis of the model of
Figure 8.12 yields the ,following transfer function:
(8.5)
wheregml is the transconductance of the first stage (differential pair Q1and Q2);
gin2 is the transconductan~e of the output stage (Q5); gol and Col are the output
condiictance.and capacitance, ,respectively, of thefirst stage; go2 and Co2ar e the
Output condfictance and capacttance (including 10ad conductance and capacit~ce)
ofth~ ~ec0nd ~tage; and ~t is the con~pensating (or pole-splitting) capaci~ce
Figurf 8..1!. Equation (8.5) shows the existence of a rightrhalf-plane zero, which
may pose a problem of excess phase shift for micropowerMOScircuits operating
at low .transconductances. In a similar two-stage bipolar amplifier, the higher
transconductance of the second stage wouldmovethe zero further from the s-plane
origin and thus diminish its effect. Gray and Meyer(1982) show that the voltage
gain of a common-source MOStransistor under open-circuit condi’tions is inversely
related to drain ~urrent until the drain ~urrent is redaged to.the point at whichth~
weak-inversion regime b~gins; the voltage gain then remains relatively constant at
a value .similar to that attainable.with a bipolar transistor. This wouldap~ar t0 b9
a~omalous, given that we have already se~n that the transconductance of an MOS
device~n wd~inversion is directly proportional to ~ain current (see Eq. (8.4)).
186 DESIGN OF CARDIAC PACEMAKERS
C1
gml(Vi) go1 =t go2
gm2(Va) CoZ
Vi = 1 = 1 (8.6)
Vdd
Q6
Vss~ Vss
R2
C(I+RI/R2)
tobea
clock frequency is f, the charge transferred per second isfC(Vi - Vo). The
switched capacitor appears tobe an effective resistance whosevalue is given by:
~)2 C ~2
~2
Switdhed-capacitor integrators
where AVi= V1 - V 2 and T is the sample period given by T = 1/f. The notation
AVi(nT- T/2) denotes the differential voltage AVi whichexisted across Ci at one-
half-switching period before time nT, i.e., at the time at which¢1 was active.
I ~ n T- T/2
nT-T..-..-....~I I I.,~---nT
= (8.9)
The transfer function of the lossy integrator is likewiS~ similar t6’ its lossless
counterpart except for an addi~onal denominator term which shows the effect of
Cx:
V°-= Ciz’0"5 (8.12)
z-0.5
AVi " Cf(l_z-l)+Cx
.......... = (8.13)
, -AVi Cf(1--e -jc°T) e-
+ jc°T/2
Cx
192 DESIGN OF CARDIAC PACEMAKERS
If we again make the approximation, that coT is small, e -jt°T = 1- jcoT and
e-Jf°TI2 __. 1; the frequencyresponseof the lossy integrator is then:
(8.14)
AVi jo)TCf + x
This is the transfer function of a single-pole low-pass filter with dc gain Ci/Cx and
(radian) comer frequency CxiTCf. One of the previously,mentioned advantages of
switched-capacitor networks is. nowmadeevident; both the d~ gain and the cutoff
frequency of the lossy integrator are determined by the ratios of capacitors, not
their absolute values. This makes possible the implementation of monolithic
switched-capacitor networks with capacitors in the picofarad range which are
nevertheless usable as filters at frequbncies characteristic of cardiac electrical
activity. A switched-capacitor low-pass filter utilizing Cx = ! pF, Cf = 8 pF,
Ci = 10 pF, and a~switch frequency of 2 kHz will manifest a de gain of 10.and a
comer frequency of 40 Hz; Figure 8.18 below compares, this circuit with its
continuous-time counterpart. Note the differences in the responses of the two
networks are virtually .identical for frequencies below300 Hz,~but that they begin
to diverge as the signal frequency rises. Also note that the response of the
switched-capacitor network between signal frequencies of 1 kHz and 2 kHz is the
mirror image of the response between de and 1 kHz; this symmetry, about half the
switch frequency is characteristic of sampled-datasystems.
2O
~hed-oapaeitor/
10
-10¯
~:~ f~i,]ossless and lossy integrators shownabove are adequately described in the
pr~ing equations only if the effects of the parasitic capacitances shown in
iTi~e~ 8.19 are ignored. Leakage capacitor Cx is assumedto have an associated
p~ificcapacitance to ground designated Cpx. Floating input capacitor Ci has two
~SO~.:i~d parasitic capacitances to ground designated Cpl and Cp2. Note that Cpx
has,~e effect of increasing the value of Cx, reducing the gain of the circuit and
increasing the. corner frequency. The role of.the parasitic capacitances associated
v~ith-;fl0ating input capacitor Ci is somewhat Lmore complex. The parasitic
capacitance Cpl charges to V1 on clock phase ¢1, but this capacitor is simply
discharged during ¢2 and none of the charge stored on this parasitic.capacitor
during~.~l is transferred-to integrating capacitor Cf. The presence of this particular
parasitic capacitance is thus seen to be benign insofar as operation of the circuit is
concerned.
Cpx
(~ ~~
Cx+
Cf
Ci :
Cpl
Cf Vo(nT) = Cf o ( nT- T) +
Figure8.20Differential-input
lossyintegratorin parasitic-insensitive
topologywithimplicitparasitic
capacitances.
Notethe topologicalchangesbetween this circuit andthat of Figure8.19.
(8.17)
6O
CMRRof differential.’i~put parm;itic-insen-
ve differen i~l integralor
¸5O
4O
30"¸
~oglf),
~t!z
wi~ Cx
2 kHz.Notethat the abscissais logarithmic
insteadof linear as wasthe casein Figure8.1
196 DESIGN OF CARDIAC PACEMAKERS
Parasitic.insensitive filter
bandpass
(8.18)
C2:20 pF
C3:1 pF
10’ C,4:1.5 oF
C5:5pF, ~=
C6:1.5pF i
-10
¯ . LL=2.4
Switchfreq~=ency= 2 kHz
2
log(f), Hz
~in
Figure8.23 Typicalfrequencyresponseof biquadswitched-capacitor filterofthe tTpe~ho~n
Fi~: 8it7. Thei~spon~b~ gins to deviate from~at of a c0ntinuous~fim~filter
for frequencies
~ is mo~tevident~in tbe~phaseresponse;~a~ntinuous-time
~b~approximately 100 Hz,~This
ne(v/brk’s phase response wouldasymptoticallyapproach~-90~for frequencies beyondthe
respon~pe, ak,~
Switches
The Switches
+Vdd
Clock signal
Q1
+Vdd OIil~
Vo
Q2 ,~
-Vss
’01tage
ion and
PACEMA~R SENSE AMPLIFIERS 199
~/_ Metal /
P÷ wo,, ’ I ’ ~ ¯
~
"~( n-substrate Thin°xlde! ’ (a)
Metal
Oxide Metal Oxide
Polysilicon
p-I-
(b)
~
"~ Oxide. Polysilicon .......
~
Metal -,~
’ Thin oxide- ~
~ substrate ~:
.....
,- . ,(c)
Figure 8.25 Typical MOS capacitor structures. The capacitors in (a) and,(b)Use_a buffed
we!l,¢oform~0n¢iplate of the capacitor; the structure in (c) !utilizes twodopedpolysilieonlayers
t~.formthe capacitor plates. ~ _~,-,
~of
~n but capac ~s
200 DESIGN OFCARDIAC PACEMAKERS
Operational amplifiers
The operational amplifiers described in section 8.3 are suitable for monolithic
switched-capacitor networks, including the operational t~ransconductanceamplifier.
It is importantthat the amplifier be able to deliver sufficient output current to re-
distribute the charge stored on switched capacitors to non-switched capacitors
during clock phase ¢2. Gray and Meyer(1982) state that the minimal open-loop
voltage gain required of an operational amplifier in a switched-capacitor networkis
on the order of several thousand. Stone et al. (1984) report that simple two-stage
CMOS operational amplifiers like that of Figure 8.11 have been used with success
in two-poleswitched-capacitor~filter circuits.
~2 ~)1
R1
-I c~
(a) Differential-inputamplifierandlow-pass
filter
C13
V/
C12
(b) Adjustable-gainhigh-passfilter
Figure8.26,Swi,tehed-eapaeitor
circuitsdescribedby,Morgan(1991). Part(a) is the diff.e .re.ntiai-
input amplifierandlow-passfilter; part (b) representsoneof three high-passstages wnienare
cascadedafter the low=pass
sectionabove.Onlyoneof the three high-passfilters has switchable
gain.Vbis a bias-voltage
bus.
~.Th. ~ alert reader will have already realized that somethingis not fight in the
description of th~ Circuits in Figure 8:26. ~e circuits as described by ~Morgan
C6~d~zotbe redu~ed~o monolithic fonni ~thb ca~it0~v~u~’,..are .mucli.tooia~ge
for ~stic monolithic implementation.and .th~ filter top010giesehosen are not
parasitic-insensitive. In addition, Morganshowspower-supplyvoltages of :E5 V for
the operational~ amplifiers in his circuit; these voltages ~are:,not available in the
typical implantable pacemaker. This .e~uit can nevertheless give someinsight into
the,frequeney-responsecharacteristies of a typie~ pacemakerand can allow ~us~to
practice-some of the analysis teehniquespreviously explored; for these reasons, it
is worth mentioning.
Stotts ~(1990, 1992) describes switched,capacitor amplifiers~similar to the
circuit of Figure 8~26(b), Switched-capacitor biquad topologies were employed
Castello et al; (1990) to build a sixth-order bandpassfilter~ for implantable-device
202 DESIGN OF CARDIAC PACE~KERS
applications: They present a filter fabricated in 2-pro CMOS technology which has
a’centerfrequency of 50 Hz, a bandwidth of 60 Hz, and draws only 500 nA from a
+1.2 V battery while being capable of’driving capacitive.loads of 30 pF. This
implementation utilized a 2-kHz clock and multiplexed two operational amplifiers
amongthe three biquadratic calls (the operational amplifiers are necessary during
the clock phases whencharge is being redistributed but not whenthe charges on
the nonswitchedcapacitors are simply being held). The oper.ating voltage range is
given as +1 V to +1,8 V. The total silicon area occupiedby the filter was given as
2.
2200 mils
The battery specifications cited by Castello et al. (1990) andthe use of a
bias-voltage bus in Figure 8.26 leads to the tentative conclusion that circuit ground
is defined at the midpointbetweenthe terminal voltages of the 2,8 V lithium-iodine
battery. The ground,potentialshown
in, preceding~
switched-capac~or;~~ ,.~ circuits need
not necessarily be identical with the potential of one end or tile other of the battery.
To sensing/pacinglead To senseamplifier
’ control
~ signal
8.5 OTHERCIRCUITS
Vs
$4
¯ ;Switchandclo~ksignalstatetable ~ CIkA
~ ,I
~l [ n 0 1 2 ~
3: J ~
~,. 1’ CIk B
4 3 4 3 ~: ~
5 Clk A 5 ClkB
Va=C2( (Ci+C2)
Vref-Vd, )_C1
’- +c2) (8.19)
(q
whereVais the voltage at the inverting input of the comparatorand Vs is the signal
voltage at the bandpassfilter output. Hip-flop F1 is clocked during phase 1, and its
/Q output will be driven high if the output of the comparatoris low. This occurs if
Va> 0, whichrequires that:
Capacitors C1 and C2 are charged in phase 2 as they were in phase 0 with the
exception that: C2 isinitially charged to Vref instead of Vdd. The output of the
voltage comparatorwill go high during phase 3 if:
Figure 8.29 shows circuits which are used to select either bipolar or unipolar
sensing or stimulation and to permit such selection to ’be madeautomatically. The
table in Figure 8.29 describes what combination of switch, FETsQ1-Q3is used for
each pacing and sensing mode.FETQ4is activated briefly after a stimulus pulse to
permit rapid repolarization of the electrode and discharging of the capacitor which
couples the pacing stimulus to the lead.
Id2=/doL2
W2-e-(V~2-Vr+Vnv)/nkT
in which VT is the threshold voltage of Q1; VG1 and VG2are the gate-to-source
voltages of Q1 and Q2, respectively; Wl and L1 are the channel width and gate
length of Q 1; W2and L2 are the channel width and gate length of Q2; and VBGis
the bandgap voltage of silicon (1.11 V @300 K). Other parameters of Eq. (8.22)
were previously defined (see Eq. (8.3)). If we compareEq. (8.22) with Eq.
we will note the introduction of negative exponents in Eq. (8.22); this sign change
accounts for the fact that p-channel devices Q I and Q2 show increasing drain
currents with increasing negativ e gate-to,source voltages. The current mirror
formed by Q3 and Q4 forces the drain currents of Q 1 and Q2 to be equal; by
equating the two relations in Eq. (8.22), wehave:
+V
04
+V
+ n+ gate
Vref / Ibias
Q3
Vss o
Circuits previously described have often had current sources that were integral
parts thereof. The circuit of Figure 8.31 shows howsuch current sources maybe
constructed. It depends upon the availability of a moderately-large resistance,
which may be implemented in high sheet-resistivity polysilicon as described by
Stone et al. (1984). Identical p-channel devices Q3and Q4 cause the drain currents
Vdd c Q~
Isouree ¯ ]
Isink
Vss c -
Figure8.31Micropower
currentreferencecircuit. Thiscircuit is :~apableOf prtvidingbothsink
and go~ ~urrents.
PACEMAKERSENSE AMPLIFIERS 207
Idl = e(V~l-Vr)/nkT
Ido
(8.24)
Id2 = Ido
e(Vo2-Vr)/nkr"
By<e6mbining
Eqs. (8.24)and Eq. (8~25), we find that the drain currents of Q 1
Q~ ~ given by:
These are ~the drain currents of Q3 and Q4 as well; the drain current of Q5 maybe
usedas a current source whosevalue is given by Eq. (8.26)if weassumethat Q5
identical; Thedrain current of Q6maybe similarly utilized as a ~cu_rrent sink; this
sink current will be described by Eq. (8.26) if Q6is identical to Q3: Notice that the
output currents of the circuit of Figure 8;31 are directly~proportional ,to
temperature. Werecall from Eq. (8.4) ~ ,that ~the transconductance of an FET
operating in the weak-inversionregime is directly proportional to’ drain current ’and
inversely proportional to temperature. ~Theuse of a current reference like that of
Figure 8.31 to determine the operating currents of FETs operating in weak
inversion can compensatetheir transconductances~ for variations in temperature.
V~ry small reference currents may be/generated With moderate-.values, of RS.
Assumethat nkTIq is 54 mVand that the width=to-length ratios ofboth Q1 and Q2
are equal; a resistance of 75 k,Q will produce reference Currents of. 720 hA. This
could be further reduced by increasing the width-to-length ratio of Q 1 or reducing
that, of Q2.
the body) sources of unwanted potentials to which the implanted pacemaker may
be exposed. Exposure of pacemakers to potential sources of interference is
unavoidable. Electrotechnology in various forms is widespread in industrialized
society, and pacemaker implantation is common;Sowton (1982) estimated that one
person in 800 in the United States and one person in 2000 in the United Kingdom
had an implanted pacemaker, with implantation rates in other western European
nations and in Canada being comparable to that in the US. Later figures give a
worldwidetotal of 1.5 million implanted pacemakers, of which 700,000 were in the
US (Smith and Aasen, 1992).
8.6.1 Endogenousinterference
8.6.2 Exogenousinterference
Low.frequency interference
The frequencies used in ac power transmission (50 Hz and" 60 Hz) lie close to the
ranges of frequencies characteristic of P and R waves(see section 8.1); it is natural
PACEMAKERSENSE AMPLIFIERS 209
that there be concernover the possibility that high voltage powertransmission lines
might cause interference with pacemakers.~ Butrons et al. (1982) described a series
of~xperiments in which volunteer subjects fitted with Medtronic 5985 Spectrax
unipolar pacemakers were exposed to 50 Hz electric fields with intensities of
20!kg/m, resulting incorporeal displacement currents of up to 300/zA: No
inhibition nor other abnormal pacemaker responses were noted. A follow-up study
0f~35 unipolar demand pacemakers (representing 16 pacemaker models from
different manufacturers) gavedifferent results (Butrous et al., 1983). No noticeable
effects, were seen ,in 11 units; 18 units revertedto asynchronous mode. The
remainder manifested irregular or inappropriately slow pacing. The corporeal
currents necessary to provoke reversion to asynchronous modewere found to range
fftm 18.4/zA to 250 ,uA.
~ Kaye et al. (1988) found similar results by utilizing 50 Hz current injection
into:the bodies of 18 subjects with implanted unipolar WI pacemakers. These
~pacemakers represented 12models from 6 manufaCturers; eleven pacemakers were
implanted in a left prepectoral ~ position~and,four, were implanted ina right
prOpectoral position. The remaining three devices were implanted in the abdomen.
bIormal pacing was seen in~ all pacemakersat low values of corporeal :~current.
~ents~ above~acertain threshold caused inappropriate pacing, erratic sensing, and
intermittent inhibition in most of~the pacemakers-used in this study; currents
beyonda second (higher) threshold caused the susceptible pacemakersto revert
asynchronous mode.,The three Medtronic units utilized in this study maintained
normalpacing up m corporeal currents of 600 gA(the limit of the current-injection
d~viee); units producedby Telectronics were found to be the most susceptible.
~,~:~:~.~stridge et al,~(1993)performed a similar study utilizing dual-chamber
p~cemakers with programmable lead configuration. Reversion to asynchronous
mixie~withbipolar lead configurations occurred at relatively high currents (170
to~550/~A); the onset of malsensing and~inappropriate function was found robe in
th~range of 80 #A to 500/~A~ Unipolar sensing was much more susceptible to
cbrporeal current; unipolar atrial sensing manifested the onset of malsensing at
currents between 10/~Aand 80/~A, while ventricular~ malsensing with inhibition
was found to occur at 40/~Ato 120/JA. No significant differences in susceptibility
were found among models produced by Siemens Pacesetter, Intermedics, and
Telectronics. Malsensing in bipolar modewas not observed in either of the two
Medtronic models which were tested; one model showed no abnormalities in
unipolar modeat - the maximal current of 600 #A, while the other reverted from
normal operation to asynchronous pacing at currents ex~g 120
.... Lrnich (1984) describes mechanisms, of magnetically-coupled interference.
Thearea subtended by a unipolar pacemakerand its lead arranged in a semicircle is
estimated at 570 cm2; a magnetic~field intensity of 21 gT ~rms) at 50 Hz:or
1,7,:5/~T (rms) at-60 Hz normal to that semicircle would produce a potential,
l~mV~peak to peak, a value Irnich takes to be the critical value at whichmalsensing
maybegin. These kinds of magnetic field intensities’may be found in the vicinity
of devices which draw large ~currents (such as arc welders and electric steel
furnaces). Bipolar sensing should prove muchless susceptible to magnetically-
coupled interference.
Radio-frequency interference
Early pacemakers were encapsulated in nonconductive epoxy resin and had few
interference-rejection components; as a result, the literature of pacemaker
210 DESIGN OF CARDIAC PACEMAKERS
~ interference sources
O~e~r
$.7 REFERENCES
Anonyhibus.
1975.Pacemaker Standard.Labelingrequirements,performancerequirements,and
terminologyfor impiantableartificial cardiacpacemakers.
Arlington,VA:.Assoeiati0nfor
the Advancementof MedicalInstrumentation(/~dkMI). ~
212 DESIGN OF CARDIAC PACEMAKERS
Astridge, P. S., Kaye,G. C., Whitworth,S., Kelly, P., Carom,A. J., and Perrins, E. J. !993. The
responseof implanteddual chamberpacemakersto 50 Hz extraneous electrical inte’rference~
PACE16: 1966--1974.
Baker, R~G., Jr., Calfee, R~V., Haluska,E. A., and Whistler, S. J. 1989. Implantablecardiac
stimulator with automatic gain control and ban@assfiltering in feedback loop. USPatent
4,880,004.
Barreras, F. J., and Manucc~,J. P. 1985. Cardiacpacer havinginput/output circuit programmable
for use with unip~lar ~andbipolar pacer leads. USPatent 4,558,702.
Belott, P. H, Sands, S., and Warren, J. 1984. Resetting of DDDpacemakers due to EML
PACE7:169-172.
Bossert, T., and Dahme,M. 1989. Hazardsfrom electromagnetic fields; influence on cardiac
pacemakersby po~;erful radio transmitters. In ElectromagneticComp~bility:’theUniversity
of York, 12-15 September1988. London:the Institution of Electronic and RadioEngineers’.
Butrous,G.S., Barton;D. G., Bonnell, J. A., Carom,A. J., Meldrum,S; J., :and Male,J. C. 1982.
Effects of high-intensity power-frequency elec~ic fields on implanted modern
. multiprogrammablecardiac pacemakers. J. Royal Soc. Med. 75:327-331.
Buttons, G. S., Male,J., Barton, D., Nathan,A., and Camm, J. 1983~Theeffect of high-intensity
powerfrequency electric fields on implanted cardiac pacemakers(abstract). PACE6:A53
Castello,"R., Grassi, A. G,~ and Donati, S. 1990. A 500-hAsixth-order bandpass switched-
capacitor filter, IEEEJ. Solid-State Circ. 25: 669-676.
DeCote, R., Jr. 1987. Unsaturable ~sense amplifier for pacer system analyzer. USPatent
4,677,986,
DeC0te,R., Jr. 1988. Automaticsensitivity control for implantable cardiac pacemakers.US
Patent 4,768,511.
Detwiler, Alan J.. 1994. Private communication.
EI-Kareh, B., and .Bom~b~,d,R. J. 1986. Infroduction to VLSISilicon Devices. Hh3gham,MA:
KluwerAcademicPublishers. " ¯
Furman~S., Hurzeler, P~, ~d DeCaprio,V. 1977a. The ventricular endocardial electrogram and
pacemakersensing. J. Cardiovascularand Thoracic Surg. 73: 258=266.
Furman,S., Hurzeler, P., and DeCaprio,V. :197To. Cardiac pacing and pacemakersHI. Sensing
the cardiac electrogran~.Am.HeartJ. 93:794-801~ ~ ,~
Furman,S. 1982~Electrgmagneficinterference(editorial). PACE 5: !-3.
Furman, S. 1993~ Sensing a~d timing the cardiac electrogram. In Fuji Si, Hayes, D. L., and
Holmes, ~r. D. R., A practice of cardiac pacing, 3rd edition. Mr. Kisco; NY: Futura
Pubfishing Co.
Ghausi, M. S., and Laker, K. R. 1981. Modernfilter design.: active RCand switched-capacitor.
Englewood Cliffs, NJ: Prentice-HalL
Grant, L. L 1993. Cardiac pacemakersand electromagnetic interferences. In IEE Colloquiumon
’EMCandMedi~ine’ (Digest No: 098). London:the Institution ~ of Electrical Engineers.
Gray, P, R., and Meyer, R. G, 1982. MOSoperational amplifier design--a tutorial overview.
, ~ IEEEJ. Solid-State Circ. SC-17:969-982.
Greatbatch, W. 1972. Multimodecardiac pacer with P-wave and R-wavesensing means. US
~ Patent 3,648,707.
Greatbatch, W. 1994. Private communication.
Hanser, R. G. 1994. Interference in modem pacemakers. Medtronic News22(1): 12-20.
Hayes, D. L. 1993. Electromagneticinterference, drug-deviceinteractions, and other practical
~ siderations. In Furman,S., Hayes, D, L., and Holmes,Jr. D. R., A practice of cardiac
ing, 3rd edition. Mt. Kisco, NYiFutura PublishingCo.
Herscovici, H, and Tarjan, P: P. 1984. Dual channel cardiac pacer isolation circuit. USPatent
4;470,418.
Huelsman,L. P. 1993, ACtive and passive analogfilter design. New.York: McGraw-Hill.
Irnich, W.1984. Interference in pacemakers.PACE7:1021-1048
Imich, W.1985. Intracardiae electrogramsand sensing test SignalS:eleetrophysieal, physical, and
techuieal considerations. PACE8:870-888.
Kaye, G. C., Butrous, G. S., Allen, A., Meldrum,S. J., Male, J. C., and Carom,A. J. 1988. The
effect of 50 Hz external electrical interference on implanted cardiac pacemakers.
PACE11: 999-1008.
Kleinert, M., Elmqvist, H, and Strandberg, H. 1979. Spectral properties of atrial and ventricular
endocardial signals. PACE3:4-6-417.
Laker, K. R., and Sansen, W.M. C. 1994. Designof analogintegrated circuits and systems. Hew
York: McCn’aw-Hill.
Mirth, S. IC, and Kurth, C. F. 1989. Miniaturized and integratedfllters. NewYork: John Wiley
and Sons.
PACEMAKER SENSE AMPLIFIERS 213
8.1 List and explain the constraints whichpacemakerapplications imposeon electronic design.
8,2 Explain the difficulties encountered in makingcontinuous-time circuits with frequency
responsessuitable for cardiac-sensingapplications.
8.3 Explain someof the difficulties encounteredin utilizing MOS devices as linear-amplifier
elementsoperating in weakinversion.
8.4 DeriveEq. (8.6) (capacitancemultiplication).
8.5 List the advantages and disadvantages of switched-capacitor technology for cardiac-
pacemakerapplications.
8.6 Calculatethe effective resistance of a 20-pFcapacitor switchedat a rate of 20 kHz.
8.7 Describe the operation of a noninvertinglossless switched-capacitorintegrator and derive
its charge-balanceequation.
214 DESIGN OF CARDIAC PACEMAKERS
In single chamber pacemakers, all timing cycles begin and end by pacing or
sensing’one chamber, In a ventricular based pacemaker (VO0, VVI, VVT), .the
Vent~ele detem~~’nes ~alltiming events. ~ Similarly, in an atrial based pacemaker
(AO0, AAI~AAT)the atrium determines all timing events. "
9.1.1 Definitions
Fi~ re 9,1 shows that the LRI (also known as the automaticinterval) is the
lo~ngest period be~tween consecutive paced or sensed events occurring in the
relevant chamberl. The pacemaker keeps track of t_he LRI by u~izing a timer. If
thefimer expires, before an event is sensed, the pacemaker will stimulate the
chamber and reset the timer. The programmed LRI value defines the minimal
number of beats per minute (bpm) as #bpm= 60/LRI (s). For example, if
LRI were equal to 600 ms, the minimal number of bpm would be 100. The LI~I
is determined by the physician and is based on the patient’s age, fitness level,
activity level andlifestyle.
Figure 9.2 shows that the URIis the shortest period allowable between paced or
sensed events, while still maintaining one-to-one (1:1) atrioventricular (AV)
synchrony. 1:1 AVsynchrony means that for every atrial beat there is a
corresponding ventricular beat. Once the URI is reached, 1:1 AVsynchrony can
no longer be maintained. At this point special algorithms are used in order to
pace efficiently. These algorithms are discussed in section 9.2.4. For sensor
driven modes (VVIR, AAIR) and dual chamber tracking modes (DDDR, DDD,
VDD), the URI is used to limit the maximal rate at which the pacemaker can
pace. This rate is also determined by the physician and depends on the patient’s
age, activity level, fitness level, and lifestyle.
The URI becomes a factor when ventrieular activity begins to occur too
rapidly. The ,URI, based on the VRPis: upper rate (bpm) = 60/VRP (s).
numberof algorithms for dealing with the URIare discussed in section 9.2.4.
URI ~
Refractory period
VRP ~
Figure 9.3 TheVRPbeginswhen a ventficular event is paced or sensed. It lasts for a set
amountof time, typically 200-350
ms.
The noise sampling interval has been integrated into the ventdcular refractory
period (VRP) of modern pacemakers. During the noise sampling interval, if
significant amountof electromagnetic interference is sensed, the pacemakerwill
pace the ventricle regardless. This is done because if there is a lot of EMI,the
true ventricular signal will be difficult to sense.
9.1.2 Modes
The VOOpacemaker is the most basic. This pacemaker paces the heart at a fixed
rate with no regard for underlying cardiac activities. Pacing at a constant, preset
rate is referred .to as asynchrq,nous pacing (also knownas fixed rate pacing),
single timing interval is employed in these relatively simple pacemakers. Each
time a pulse is delivered, the internal timer is reset. While the pacemakeris in
218 DESIGN ~ OF CARDIAC PACEMAKERS
this mode,if there does happen:to be underlying spontaneous heart ,activity, the
stimuli will be effective only outside of the cardiac refractory period, as the
cardiac tissue is.unresponsive to stimuli during the cardiac refractory period.
Figure 9.4 is a logical flowchart, representing a VOOpacemaker. The
diagram uses a counter to keep track of time, however, any circuitry capable of
tracking time could be used (i.e. a simple RCcircuit). Suppose a pulse has just
been delivered. The counter is reset and counting begins anew. The counter is
continually incremented until the LRI is reached. The LRI corresponds to the
programmed(or factory specified) lower rate interval. Oncethe LRI is reached,
the pacer delivers a pulse and the cycle is :repeated.
]I Resetcounter
~--Jl,cre~:nt
coun,er
J
Deliverpulse
Figure 9.4 Flowchart of events in a VOOpacemaker. Upon delivering a pulse, the counter is
reset, and a new timing interval begins. The counter is continually incremented until the count
value is greater than the LRI. At this point a pulse is delivered and the cycle is repeated.
The VVI mode of pacing delivers a ventdcUlar stimulus whenever the heart
neglects to do so. This pacemaker employs two timing cycles, the lower rate
interval (LRI) and the ventricular refractory period (VRP).
Figure 9.5 shows that a VVI pacemaker’s timing cycle always begins and
ends with either a paced or sensed ventricular event. Supposethat the pacemaker
has just delivered a pulse, its internal counters are reset and the pacemakerwaits,
continually incrementing its internal counters. The pacemaker monitors the
ventricular channd for a QRScomplex. If a QRScomplex is sensed, and the
ventricular refractory period expires, the counter is reset, and the cycle begins
again. If the LRI expires before an event is sensed, a pulse is delivered by the
pacemaker,and then the counter is reset.
which processes the sensor’s data. These details are covered in Chapters 14, 15~
16, and 17.
No
" No
~ DeliverpulseJ
The dual chamberpacemakerhas the ability to sense and pace both the atrium and
the ventricle. Ever since the late 1970s, the use of the dual chamberpacemaker
has been steadily growing. As of 19.86, an estimated 50-70%of all permanent
pacemakers implanted were dual Chamber. One probable explanation for their
adoption is that they most closely mimicthe function of the physiologic heart.
9.2.1 Definitions
Figure 9.6 shows that the ,AVI is the electronic analog to the P-R interval. The
AVI is the programmed~~nterval from a paced or sensed ~ atrial event to the
following paced or sensed ventricular event.
The importance of the AVIis thatjt gives the ventricle time to fill following
an atrial contraction. An optimal A~I, which is necessary for maximal
he~Odynamic benefit, may cause up to a 40% increase in ventricular end
di~/st61i’c volume~ri patients with~ sinu~~ rhythm (Benchim6i~eta!., 1965). If
p/i~e’itlaker has ~ ~p~grammableAVLand~ the physician has ~ equipment to
~oninv~/si~cely meaningcardiad output (i.e. puisea’ boppler); ~ai/6pti~al AVI
could be determined by pacing at a fixed rate~ varying the AVI, andmeasuring
220 DESIGN OF CARDIAC PACEMA~RS
cardiac output. However, if the above experiment can not be performed, it has
been found that an AVI of 150 or 200 ms provides a patient with maximal
cardiac output, whenpaced at 80 and 110 bpm(Haskell and French, 1986).
Figure 9.6 TheAVIis the programmed interval froma paced or sensed atrial event to the
followingpacedor sensedventrieularevent.
Figure 9.7 .shows that ther ventriculoatrial interval (VAI)--also knownas the
atrial escape interval (AEI)---iS the time between apaced or sensed ventdcular
event to the next atrial event. The VAI can be derived by subtracting the AVI
from the LRI (Barold, 1988). The VAI is important in that it provides li nk
between the atrium and the ventricle.
VAI
Figure9.7 TheVAIis definedas the time betweena pacedor sensedventricular event to the
nextconsecutive
atrial event.It canbe derivedbysubtractingthe-AVl
fromthe LRI.
PVARP ~
T~e TARPrepresents the total amount of time that the. atrial channel is
r~fract0ry. It Consists of two separate timing intervals previously discussed: The
TARPbegins witha paced or sensed- atrial event and extends through the AVI.
¯ The atrial sensing amplifier is always refractory for the duration of the AVI.
Finally, ~the TARPends with the completion of the PVARP.Therefore, Figure
919 shows that the TARPis equal to the sum of.the AVI and the PVARP.The
d~u’ation of the TARPdefines the shortest possible upper rate limit.
PVARP ~
Summary
Figure 9.10 summarizesall of the timing intervals that have been covered in the
chapter.
A DDDpacemaker senses and paces both the atrium and the v~ntricle. A simple
DDDpacemaker can be made by taking a VVI pacemaker and adding an atrial
channel (Bernstein et al., 1987). Indoing so the simple DDDpacemaker possesses
six timing intervals: the LRI, the VRP, the URI, the AVI, the VAI, and the
PVARP.
_:/~ ..~e. atrial, .and ventricular channels ofa DDDpacemaker are, intimately
fifiked and ~the events in.one channel influence theother (Barold et .al.,. 1989). For
example, suppose the ~trial channel has just paced theatrium or sensed an atrial
event. Nowthe AVI begins. Whenthe AVI timer expires, ifno~QRS complex has
222 DESIGN OF CARDIAC PACEMAKERS
The VBPand theVSP are two dining intervals used-to help prevent crosstalk and
are incorporated into-the current atdoventdcular interval (AVI). ~Grosstalk’-in
electromagnetics is the ,influence-of one signal onanother. Similarly, in cardiac
pacing, crosstalkis the incorrect:sensing 0fone channel by another, Thisoccurs
LOGIC FLOW AND TIMING DUGRAMS 223
Atdum
sensed
expired
AS VS AP VS AS VP AP VP
AVI
PVARP
TARP I I I I I ’t i
VAI I I ~\\\\%~ I I I
VRP
URI I I i "i I ,1 I
LRI
Figure 9.12 AS= atrium sensed; VS= ventricle sensed; AP= atrium paced; VP= ventricle
paced;stripedboxes= interval terminated earlydueto spontaneous
cardiacactivity. Aninve.rted.P.
waveor QRScomplex symbolizesthat it wasinitiated by the pacemaker. Thefirst sensedatrial
eventinitiates the first AVI.Sincethe first QRScomplexis spontaneous,
it causesthe first AVI to
be prematurelyterminated,andit also causesthe first VAIto begin.Atthe completion of the first
VAI,an atrial pulse is delivered.Dueto the spontaneous secondQRScomplex,the secondAVIis
prematurelyterminated.ThesecondVAIbegins; It is terminatedearly dueto the spontaneous P
wave(the third P waveabove).At the endof the third AVI,a QRScomplex is paced,since none
weredetected. Thethird VAItimesout, anda P waveis paced.At the completionof the AVI,a
QRScomplex is paced,since noneweredetected.
Following the VBP comes the VSP period. It encompasses the next
60-120-ms of the AVI. During the VSP period, any signal (crosstalk, QRS
complex, etc.) sensed by the ventricular channel will not inhibit the pacemaker.
Instead, the sensed signal will initiate a pulse to be delivered by the pacemaker.If
the sensed signal were a QRS-complex,the delivered pulse will fall in the
refractory period. If the sensed signal were indeed crosstalk, the AVI.will be
abbreviated due to the early release of the QRScomplex. Once the VSPperiod is
complete, any subsequent sensed signals will inhibit the pacemaker’soutput.
Hysteresis
Sincei:~arly~iin .the chapter, the upper rate, interval has been an integral timing
interval in many pacemakers. Brat, how does the pacemaker respond wben ~e
atrium beats faster than the programmed upper rate? There are a number of
algorithms used, none of which are ideal.
The two methods discussed here depend on the complexity and sophistication
of the pacemaker. The fixed ratio block method uses the AVI and the PVARP,to
define the upper rate li~i’t. The Wenekebach method has a separately
programmableupper rate limit, which gives these pacemakers a more desirable
upper rate response.
zxed~ratla ,.block
AS VP AU AS VP AU
~ AVI ~ PVARP
Figure 9.15 shows that in order for’a ~paeemaker to implement th6 Wenckebaeh
type of an upper rate response, a sep~ately programmableURI is necessary. In
addition, the URI must be programmed longer than the TARP. In these
pacemakers,as the atrial rate increases; the AVIis lengthened so as not to yield a
ventricular pacing pulse prior to the end of the upper rate limit. As atrial rate
increases further, and P waves begin to fall within the PVARP,fLXed-ratio block
then occurs. The amount of time that the AVI is lengthened is equal to the URI
minus the TARP. ¯ ~
The advantage with the Wenckebachresponse is that it avoids the sudden
reduction of ventricular rate (which commonlyoccurs in fixed ratio block).
addition, this methodalso helps maintain some degree of AVsynchrony at higher
atrial rates.
AS VP AS VP AS VP
~ AVI ~ URI
~ PVARP ~ Wait
Flgu 9A5S = sensed; P = paced, ExampleofWenekebach t~pper rate response. Note that
as the atriumbeginsto speedup, the AVIis extended~so
that aventficularpulseis not delivered
beforethe expirationof the URI.
LOGIC FLOW AND TIMING DIAGRAMS 227
9.3 REFERENCES
Logic Implementation
Surekha Palreddy
Control
Unit
Programmability I
~xternal
&telemetry programmer
PACEMAKER REQUIREMENTS
The basic requirement of a pacemaker isto pace the heart when needed. Other
functions "include~. the ability to sense-the cardiac signals. Multiprogrammability
adds flexibility in-choosing parameters of the pacing pulse and also to ~.monitor the
dat~ storage; Somepacemakersare provided with diagnostic capabilities that can
be’~sed for".therapy. VLSI technology makes it feasible to integrate alLthese
features of the pacemakeronto a small chip. Someof the basic requirements .of a
pacemaker are:
Basic funct.io~:-.Pace, sense, programmable,diagnostic, interactive
High ~eliabil~i ~O.~5%failures/month "
Adequate lon~ty:i’7=lO years . .
Snu!~ ![ si~: i O:~.thick, 25750 g "
Simple to be p~. grammedb~y a physician ~:
The functions of a pacemaker depend completely on the electroifie eonttol
system. Therefore, selection of electronic.technology that provides the pacemaker
requirements forms the basis~to pacemakertherapy.
Initialize refractory
periodtimers
P-wave
sensed R-wave
refractory
period exceeded
I watchdOg
timell I Interrupt handle~
I~
I Flags
ROMi
I Wakeup
I Pa~
II
I I Register
file
. ;amcounter
~!I InstrUctionregister IIntemalI-3..
~
clock ~1
I’ [Clock
. . control
I Backup ’I Extema
controller
. _ I II Digttall/O
I c’°c.l I r,sI
I
/ / tiiillll~
. ’ I Sens~rsll
Figure~ 10.4 Functional block diagramofa micml~rocessor-baSed pacemaker.Memory unit:
ROM i~ Usedto store the instructions of the program,~ to store the various programmabIe
parameters,
timersto keeptrackof the elapsed)ntervals,
registerfile to Store~ntermediate
values,
an ALU to performthe arithmetic calculations, and’otherauxiliary units that enhanc~the
performanceof a microprocessor-basedpacingsystem.
Load/storeinstructions
Load re$ addr Loadsa valuefromRAM into a resister
Load immediate Loadsa number specifiedinto the accumulator
Loadtimerre$ timer ~LOads a ~valuefroma:re~sterinto a timer
Store re[ addr Storesthe valuein a resister into RAM
Storetimer re$ timer Storesthe valuein timerfoa resister ¯ .
ALU instructions
Add Addavaluefroma resister to accumulato, r
Sub re$ Subtracta contentsof a resister fro.~ accumulator
Mult re$ Multipliesthe valuein a resister withaccumulator
Increment re$ increl~entsthe valuein regis~r by one: :
Decrement rel~ Decrements the valuein a resister byone
~ontrolinstructions
BGErel~l re$2 Branchwhen(re$1)>_(re$2)
BLEregl reg2 Branchwhen(regl) ~ (reg2)
8LT regl reg2 Branch,when(re$1) < (re~2)
lump addr lump,to ’~e.~ s~eih~laddress
~pecialinstruction
~LEEP : - Puts, the microprocessorto sleep
Instruction format.
The instruction format is decided based upon the total numberof instructioiis in
~e instruction set. ~:Th, e instructions fetched from memory,are 8 bits long, Each
instruction ha~ an Op~odefield(2 hits),, a register~ specifie~field (3-bits), a~nda
bit immediate,field."The opcodefield indicates ~e ~ype ofthe inst/’ucti0n that was
fetched. The’ ~egister specifier indicates. ~e ~add~es6of the reglsterin the r~gister
file on which the operations are pefformed~ The, immediate field i~: Shift~and
sign extended t6 0~tain :the address, of the memory.location in load/store
instructions, Similarly, in branch ~and jumpinstructions; the Offset field i8 hsedto
calculate the address~of the memoryi~ahbt(the,.9ontrol needs to. be transfe.rr~ ed
tO. ~e fgrmat of the instruction, set., ,chosen to implementthe AViate-respotlsi~ee
pacemakers is discussed in the Al~ndix~,, ’ .... ~
Register file , , .
A regf~er fde is a collection of r~isters in which any register can be read from
or written tO by s~ecffying the tiiJrnber of the register inth6 tilde. Base.d ~on’the
requirements of the design, the, s~ze of the register file is d¢¢ide~.’ Fgr the
purposeg of implementation ofa phce~aker algorithms, a registe/fi|~ of eight
registers is sufficien~t, with~ three .~pecial:purposeiegisters (0-2) ~d. fig~ ge,fi’era!
purpose registers (3-7), as showii in Figure’ 10.6. Register "0" always holds the
value "zero". Register ’T’ is dedicated to the sensed/paced flags. Register "2" is
an accumulator in which all the arithmetic calculations are p~rf~ed~.T~e
read/write address port provides a 3-bit address to identify the rel~i~ter:being
r~ or written into, The ~write ~ta PO~provides. 8.-bit data to he written into ~
registerseitherfrom ROM/R~or ~rs~ Read.~e.n~bie.~0ntr61~ .when assert6d
LOGIC- IMPLEMENTATION 2 35
enables the register file toprovide data at the read data port. Write enable control
enables writing of.data being provided at the write-data port into a ~register
specified by the read/write address.
Read
AVtimer
Pacetimer
Clock Write
enable enable
Figure 10.6 A sebelii’atie diagram of the register file,.timers and ROM/RAM. Three registers of
the register file ~ dedicated for Special purposes and the Test are general purpose registers. An
instruction register is used to load the instruction fetched ~frommemory.
Timers
Datapath, ,
Clocks Pulse
1 MHz Control¯ I/O portsI Atdalsensed/
100 Hz
ventriclesensed
Program
counter Register
file
Read Wdte
enableenable
Zeroflag
PaceInterval
timer
Wdte
Clock
Control=
clock cycle. Since the instruction fetch and decode stages-are commonto.all the
instructions, the initial two states are commonto all the instructions. Step 3 to
step 5 differ dependingupon the opcode. After the execution of the last step, the
finite state machinereturns to the fetch state.
State0 State1
Jump
Bmnoh
State State State
address
Fi~r~ t0.8 Finite state machine diagram indicating various states:of instruction types. Load
in~h3~cfign:takes 5 clock cycles whereas store-and A~.~ instm_ ctions take 4 clock cycles. Branch
instructionsare the shortestandare completed
in 3 clockcyc|e~. .
:A finite state machine can be implemented with a register ’that holds the
Current stage ~ind a block of eombinationaI16gicSuch as a PLA~it determines the
datapath sign’s that need t,o be aS~6rted as well as the next state. A PLAis
des6dbed as an array of ANDgates f~llbwedby an array of ORgates.. Since any
function can be computedin two levels of lo~ic, the two-level logic 0f PLAis
~s~~fot generating e~ntml signals. ’ ’ ~"~ ~:~ i ’’ " * ’ ¯ ~ .
~:’!,~ ,~e occurrence of a w~ehp event initiates a sfor~doperating routine
e~Sponding to the e~nt; In the time imervailbetw~n a~completed operating
r6~ne and a next wakeupevent; the~ internal logic components of the pr0ces~b~
ar~;’lieactivated and no energ~is being expended:in l~efforming an operating
r6iatine.
238 DESIGN OF CARDIAC PACEMAKERS
~ Pacemaker [
External Defibrillator
clock
32 kHz
1MHzclock with Sleep"
ON/OFFand
Clockdivider intemalinhibit
~ Intermpt,wakeup
~ i control logic
Sense/telemetry
andother inputs
Figure 10.9 A schematic block diagram showing the microprocessor-based pace ~n~. r system
with dual clockcontrol. Anexternal low-frequency clock(32 kHz)dividedinto requi ~redlower
frequenciesfor timers. Ahigh-frequency
internal clock(I MHz)is activatedin the "Wakeup"
state
of the microprocessor
(Russie,1991).
shownin the figure, the output of the clock is terminated whenthe "STOP-CLK"
signal goes fromhigh to low.
~ J Jstabl~RCoscillator
ii [ start (in_hibit)
Stop C_J-KDs ~F
To microprocessor
Stop CLKinhibit
Ds I I
Stop CLK \~
Start /
Figure 10.12 shows the "clock inhibit" mode employed to prevent the
microprocessor from reading the value of the timers while they are being clocked
by a 128-Hzclock. The 8:ms clock is the primary dock to the timers. This signal
is a pulse~one-half, the; width of the 32-kHz clock cycle which occurs every
7.8125 ms, The timers are .clocked onthe failing edge of the 8,ms iclock and the
inhibit circuit is inactive as long as the 8 ms is low. This: is accomplished by
LOGIC IMPLEMENTATION 241
holding FF1 and FF2 in reset states. Whenthe 8-ms signal goes high, the reset is
released. At the trailing edge of the first pulse of ASfollowing the transition
from 0 to 1 of the 8-ms clock, FF1 is clocked triggering Q1. to logic 1. Thus, a
logic 1 is applied to the date input of F’F2. At the second ASpulse~ the output of
FF2 (Q2) goes high, which holds the microprocessor in the address decode state
until the 8-ms clock goes to low. Thus, it is guaranteed that 3 gs will be provided
from the time the timers are clocked until the first .high-to-low transition of the
DS, when the data is written or read. FF1 and FF2 provide synchronization
between the 8-ms signal derived from the 32-kHz external clock and ASderived
from the 1-MHzclock of the microprocessol;
8 msinhibit
1 MHz
AS
8 ms /1/fill//
Inhibit
Accumulator
VCO counter
Up ~ f ~ Zero
count
I D~
In
Reset Up Down
V(x)
Eref ~7
CIk
Strobe NDcomplete
In "N"’output
counter
N(x)
where the accumulator count is a function of the input "Vin" and the length of the
time this voltage is applied to the VCO.
n
Tup - Folk (10.5)
- V(x)
N(x) = X Eref (10.6)
II0
interrupt
request reinitialization. The time required for execution of one complete cycle of
the program depends on the number of active tasks on the system and the time
taken for each task. The timer is usually initialized to the maximalcycle time
possibly required by all the defined active systems. In order to check whether a
program.’ segment is being executed at all, markers are introduced into those
segments that get set whentraversed. The main program studies the behavior of
the markers and takes diagnostic measuresto identify if a problem is suspected.
:~.~ The regulations of the Food and Drug Administration require that, whenthe
watchdogdetects a major failure, the implantable device be set in a state that is
safe for the patient. If necessary, that state can be nonfunctional.
10.7.2 Redundant pacemaker system
Redundantbackup systems are essential in any life-saving devices. The redundant
pacemaker system is a simple pacing circuit, which operates in a VVI mode
(Schaldach, 1992). In this mode, the ventricle is sensed and a pulse is applied
whenthe signal is not detected. The redundant circuit is activated, simultaneously
deactivating the microprocessor circuit, whena fault is detected in the operation
of the microprocessorcircuit.~ This circuit provides stimulation pulses at a fixed
predetermined rate.
Highpriodty interrupt i i~ .
Lowpdodtyinterrupt "
Lowpriority interrupt
Highpdod~interrupt
was proposed. The battery test circuit is provided :with a pulse counter and input
logic to measure the consumed charge from the operating parameters of the
pacemaker and the number of pulses delivered over a period of time. During
each test, the charge delivered since the last battery test is calculated based on the
count ,in the pulse counter, which is then summedto the contents of the charge
counter in memoryas shown in Figure 10.16. The content of the:charge counter
is a measure of the total charge consumed and provides information about the
remaining life of the battery (Moberg,1987).
Pulse parameters
cou~ter i ~1 1
I Input logic It,~l~ Moderegister
~1
I Charge counter
10.8WHY CMOS?
Vdd
Figure10.17CMOS
inverter configuration.When
In = Vdd,Ptran is openandNtranis closed,
so out = 0 V and vice versa .... ¯
410.9 REFERENCES
Amado,J. B., Belaza, J., Diaz, A., and Tur, J. B. 1985. Technologyof pacemakerselectronic
circuitry. In F. P. Gomes et al. (eds.) Cardiacpacing: Electrophysioiogy.Tachyarrhythmias.
Mt. Kisco, NY:Futura Publishing. " ~
Baker, R. G., Jr. 1989. A-Vresponsive rate adaptive pacemaker,USPatent 4,856,524.
Buffet, J., Gautier, J. P., and Jacquet, J. P. 1982. The software pacemaker.In G. A. Feruglio
(ed.) Cardiacpacing: electrophysiology and pacemakertechnology. Padova, Italy: Piccin
Medical Books.
Buffet, J., Gautier, J. P., and Jacquet, J. P. 1982. The software pacemaker- Feasibility of
recording pacemaker.In S. S. Barold and J. Mugiea(eds.)The third decodeof cardiac pacing:
advancesin technologyandclinical applications. MountKisco, NY:Futura Publishing.
Buffet, J., Meunier,J. F., Gautler, J..P., and Jacquet, J. P. 1982. Technology and reliability of
microprocessorsused in pacemaking,In S. S. Barold and J. Mugiea(eds.)The third decadeof
cardiac pacing: advancesin technology and clinical applications. MountKisco, NY:Futura
PuMishing.
Dassen, W. R. M., Dulk, K. D., and Wellens, H. J. J. 1988. Modem pacemakers: Implantable
artificial intelligence? PACE,11:2114-2120.
Dassen, W.R. M., Steld, A. V., Braam, W. V., Dulk, K. D., Gorgels, A. P. M., Bmgada,P.,
and Wellens, H. J. J. 1985. PACTOT: A repmgrammable software pacing system. PACE,8:
574-578.
Dassen, W., Steld, A. V., Dulk, K. D., Brugada,P., and Wellens, H. 1984. The soft pacemaker:
A new approach in pacemakerdesign? Computersin cardiology, 4: 529-532.
~uggan, S. R. 1992. Analogto digital converter, USPatent: 5,092,330.
inspruch, N. G, and Gold, R. D. 1989. VLSIelectronics microstructure science, VLSI in
medicine :17, Orlando, FL: AcademicPress.
Fromer, M., Shenasa, M., Kus, T., and Pag6, P. 1987. Management of a patient with recurrent
sustained ventricular tachycardia with a newsoftware-based antitachyeardia pacemaker.J.
ElectrophysioL, 1: 133-139.
Gaggini, G., Garberoglio, B., and Silvestri, L. 1992. Mixedmicroprocessor-randomlogic
approach for innovative pacing systems. PACE,15: 1858-1861.
Garside, R. G. 1980. The architecture Of digital computers.NewYork:OxfordUniversity Press.
Harrigal, C. E., and Walter, R. A. 1990. The development of a microprocessor controlled
implantable device. Proc. IEEENortheast BioengineeringConf. 3: 137-138.
Hartla{~b, J. 1982. Pacemakerof the future: Microprocessorba~ed or customcircuit? In S; S.
Barold and J. Mugica(eds.)The third decode of’cardiac pacing: advancesin technologyand
clinical applications. MountKisco, NY:Futura Publishing.
Kraft, G. D., and Toy, W. N. 1981. Microprogrammed control and reliable design of small
computers,Englewood Cliffs, NJ: Prentice Hall.
J. L., and Patterson, D. A. 1993. Computer organization & design: The
hardware/softwareinterface. San Mateo, CA:MorganKaufmann.
, L. 1987. Battery test circuit for a heart pacemaker.USpatent 4,715,381.
Russie, R. J. 1991. Implantable cardiac device with dual clock control of microprocessor. US
Patent 5,022,395.
250 DESIGN OF;CARDIAC PACEMAKERS
Thefunction .of the output circuit is to deliver periodic voltage pulses to the heart
muscle. High standards of quality and durability are required for the design and
implementationof the circuitry. A patient’s quality of life, and sometimeshis/her
life itself, dependupon the proper operation of the pacemaker.Malfunction of the
anit~generallyresults in an unhealthy and Unhappycustomer.
~Design of the output circuit has evolved since the first pacemaker was
implanted. Manyearly Units had RCoscillators and transformer-coupled outputs.
Quite a few of these devices were also constant current output sources. As time has
progressed, these units have fallen out of favor and been replaced by devices with
constant voltage capacitor discharge outputs and crystal oscillators. Other advances
in pacemaker design, such as triggered and inhibited sensing, have created new
problems which require solving. Someof these problems and their corresponding
solutions are addressed in this chapter, along ,with the basic fundamentals for
designingoutput circuitry ........ ~ ....
Cell
(a) Membrane
S
(b)
u
r
r
e
n ~eobase
t
t=ic +iR=C(aV
/ dt)+ V / (11.1)
The voltage drop across the membranedue to current flow can be found with the
equation
V = IR(1 - -t/Rc) (11.2)
(Geddes, 1984). To depolarize the cell, V must equal -20 mV.With R equaling
1 f~, the minimalcurrent whichwill achieve this is
’minimal current Value is the rheobase of the curve. The rheobase value is
depe,ndentupon the characteristics of the load (in this case the membrane).
.T0ill~s~ate the Cli~ge saved by increasing ~e amplitude, and shortening the
dBi~afipn of the.stimulus pulse; we Will use current levels of~25 mAand 40 mA
(rem~,-,~..ber that, ~ese values haveno ,relatioti to actual pacemaker-indue~xl
currents
~,heart). With a 25 mAcurrent mplittlde, the mininial dUrati~ia of the square
~av~ stimulus is
20mV = 25 inA x 1 fl(1 -t/10-6
~ e ~ ). (1L4)
This results ina stimulus time t of 1.60 ~s. Thetotal charge tised (current x time)
at tlfis’-stimulus amplitude is’ 40nQ.Now,if thb~40 mACurrent pulse is used, the
stimulus duration required to achieve depolarization is
;.-~
.: 20 mV= 40 mAx 1 £2(1 ~ e-~/i0-6 (11.5)
with t equaling 0.69//s. The .amount of charge used is,only 28 nC. Therefore, the
charge dissipated during excitation with the 40 mAstimulus amplitude is only 70%
of~that consumedwhenthe stimulus is25 mA.Itis easy to see that shortening the
duration of the stimulus pulse is effective in reducing charge consumption.
Although charge consumption is,minimized when the pulse duration is
shortened, energy is not. Assumingthe pacemakerload (i.e. the heart) is a fixed
value; energy usage can be shownby the: following equation
W = 12Zt (11.6)
whereZ is the entire load impedance. The minimalproduct of voltage (or current)
and duration is described by
WhereI0 =,voltage rheobase,, tc~is the chronaxie time, t is the duration.of stimulus,
and~I,is the, amplitude,of ,the..~stimuluslcurrent.~The,chronaxie,time,is,. the stimulus
dttration:when, the stimulus current magnitudeistwice the rheobase,~andis near the
point of minimal energy consumptiomCombiningEqs~ (11.6). and (11.7) results
the energy use at the chronaxie time tc being .~
254 DESIGN OF CARDIAC PACEMAKERS
W = 4.5IgZt~ (t = Zt e ) (11.11)
The energy used with t = 0.2 tc and t= 2tc is 180%and 113%, respectively, more
than what is used whent is equal to the chronaxie time. This energy consumption
definitely influences the longevity of the pacemaker.
The stimulus threshold curve results from the interaction of the pacemaker
patient’s heart and the electrod6 and i~ influeheed by ~any factors~Th~se include
the size and nature of the stimulating electrode, the nature Of the’ heart, the
placement of~e lead; epinephrine leyels in’~e body, etc’. ~ Since nunierou~’~actors
affect the threshold for cardiac padhg,a 100%safety margiia i~~ genelally set
(Furmanet al., 1993). This meansthat the energy delivered while pacing should
100%higher than the minimalamountof energy necessary to excite the heart. This
can be easily determined using the strength-duration curve and energy equations
previously mentioned..Thethreshold:for excitation increases by: a factor of two to
three after imp!antatio, n due to inflammationandthe deve~lopmentof scar tissue but
then stabilizes after a monthor so. ~.
Changing nowfrom working with current to voltage, the chron~xie time can
be found by determining the stimulus threshold at two fixed voltages VI and V2.
The stimulus threshold at fixed" voltages call Be found by varying the pulse width
until cardiac activity in response to pacing is barely present. The pulse widths
associated with V1 and V2are tland t2 respdetively; In equation form
with V0 being the voltage rheobase (the minimalvoltage, required for stimulation).
Manipulating Eqs. (11.11) and (11.12)~ we can fmd the chronaxie
The chronaxie time approximates the most efficient stimulation pulse duration
(Furmanet al., 1993). Thus, including the 100%safety factor, a preferable stimulus
would have an amplitude of twice VOand a ’duration of twice the chronaxie time.
The duration is lengthened to twice the chronaxie time to achieve the safety
margin. Such anadjustment is generally~easierthan changing thevoltage level,
since the timing,circuit and pulse width,seleetor:can-:be more easily: designed to
deliver a plurality of pulse widths than,are a~cailable from~oltageeonverters: Note
that the chronaxie.time calculations .in,thiS paragraph were done with voltages, as
all new cardiac pacemakersare constant voltage sources, -
PULSE OUTPUT 255
In a unipolar stimulating device, the electrode tip stimulates the heart, while the
pacemakerunit serves, as the reference. In a bipolar ~device, the lead has both a
Stimulating tip, the cathode, and a ring, the anode. The ring generally has a much
larger surface area. The separation of the ring and tip are 2-3 cm, dependingon the
pacemakermodel (Furmanet al., 1993).
.The current threshold of stimulation is the samefor both unipolar and bipolar
l~adS. The voltage threshold, however,is slightly higher for a bipolar lead because
of~the4ncreased lead resistance (Furmanet al.,.1993); :For instance, with a pulse
duration ~of 0.1 ~ms, the impedance of a unipolar lead was found to be around
489 ~2, while the bipolar impedance was 600 f~ (Furman et al., 1993). This
increased resistance during bipolar pacing is due to the ring area being much
smaller than the pacemaker case area..(which is the anode in unipolar pacing).
Since resistance is a function of condueiionpath area, the anode in bipolar pacing
has moreresistance than in unipolar pacing. The extra tissue betweenthe electrode
and can in unipolar pacing can be neglected becametissue is a good conductor.
A benefit of a bipolar lead configuration is that the signal-to-noise ratio of the
sensed heart signals is better.than that found with unipolar leads. The bipolar
sensing configuration eliminates muchof the noise resulting from nearby muscle
movement.However,as will be seen in the next section, bipolar stimulation can
also have its drawbacks. Most modemde;cites can be changed from unipolar to
bipolar configuration, and vice-versa, through telemetry (see Figure 8.23).
interior to exterior of a heart cell is about -90 mVwhenthe cell is in its resting
state. Loweringthis potential drop to approximately -70 mVwill cause to cell to
exhibit an action potential. In the case of cathodal stimulation, direct application of
-20 mVlowers the extracellular potential to -20 mV,and the entire potential drop
across the cell membraneis now-70 mV.This results in the cell firing an action
potential. Onthe other hand, if cathodal stimulus is used the extracellular potential
is raised in the positive direction, and the resulting drop across the cell membrane
isnow around -110 mV.This large voltag e drop results in hyperpolarization of the
cell, and no action potentials occur: Note that increasing the anodal stimulus level
will eventually result in the cell firing. However, the magnitude of stimulus
required from the anode can be two to three times what is required from the
cathode.
Threshold
(v)
Anodal
I
Cathodal
i
0 ms 200 ms 400 ms
Delay Time
Note also the myocardial response to stimulation during the refractory period.
In early bipolar devices, the anode and cathode were nearly equal in size (Furman
et al:, 1977). It was found the anode of bipolar electrodes may have been
responsible for someoccurrencesof ~,entricular fibrillation. In fact, Stevensonet al.
(1986) found that anodal excitation can occur during bipolar stimulation, producing
changesin local myocardialactivation,-potentially causing initiation of Ventricular
strength-interval curve is the corresponding ECGof the heart. During the T wave,
the anode can easily cause stimulation. Dueto this, multiple response (MR), which
is morethan one response to a single stimulation, and ventricular fibrillation (VF)
conditions are more sensitive to anodal stimulation than they ever are to stimulus
from the cathode (Furmanet al. 1977). Other factors also contribute to the heartrs
susceptibility toward MRand VF. The ratio of the surface area of the cathode to
the anode is one factor, and another is the proximity of the anode to stimulatable
tissue. One solution used to eliminate VFand MRis to increase the size of the
anode, while keeping the cathode small. Another solution is to movethe anode
further awayfrom the heart. Whilemovingthe anode helps to solve ~the problem of
anodal stimulation, it also decreases the signal-to-noise ratio of the evokedheart
responses, whichare sometimes; sensed to allow more effective stimulation of the
heart.
Pacemakers should ideally deliver a square wave pulse from the cathode with
characteristics similar to Figure 11.4(a). The duration andamplitudeof,the pulse
should be adjustable. All modempacemakers are of the constant voltage output
type. Thus, the load impedancehas a large efftct on the stimulus current. A load
withhigh impedance will have an excessivelyhigh threshold ofstimulafion, while
a 10wimpedancewilt result in large.currents and premature battery drain. Furman
et al. (1993)-describes the four different pacemaker output designs which have
been or areiavailable~
Vdd
Figure 11.4 (a) A representation of an ideal stimulation pulse from a constant voltage
stimulator.Thevoltageis measured fromthe stimulatingtipto.the reference’. (either the ring or
the pacemaker can). Vdd Can either be a fixed value, as in single ~voltagede~iees, or. a
programmable value, with variability dependentuponthe pacemakermodel. The period of
stimulationT is variablein all devices.(b) Arealistic depictionof a waveform appearingacross
the heart emittedfroma capacitordischargeoutputcirc~t. Nolethat the drop.in Prise voltage
magnitude is dependentuponthe size bf the ttitput capacitor:’ Alarger’eap~ito~~vill :havea
waveformwhichmore closely resembles an ideal constant voltage. The small rise and
exponentialdecayafter the stimduspulseis an afterpotenfial,whichis discussedin section11.3.
Devices with this designation have two output voltage-levels, 2.5 and 5.0 V. The
period of stimulation Tis also variable similar to single voltage devices. Formerly,
the 5.0 V setting was regularly used, and the 2.5 V setting was used in special cases
whenextremely low (at the time) thresholds of stimulation occurred. Now,with the
advent ofnewlow threshold electrodes, the 2.5 V setting is regularly used, and the
5.0 V setting used only in situatons where2.5 V is insufficient for stimulation.
Another limitation is the low voltage values available from the pacemaker
battery. The voltage of a modernpacer battery is about 2.8 V, which is below the
functional range of many standard componentsused in electronics. While the
voltage in pacemakerscan be stepped up, this should be avoided wheneverpossible
due to the losses incurred whenusing voltage multipliers. The circuit must also
function over a range of supply voltage values. As the lithium-iodine battery
depletes, the magnitudeof current whichit can effectively supply decreases as the
internal resistance of the battery increases. Since replacing or recharging the
battery is out of the question, the pacemakershould still be able to operate under
low battery conditions.
(11.!5)
wli6re T is the s~itehing period, fi~ the frequeoCy0fswitching,and V is’ithe peak-
t0~peak ripple v~ltage o~ Cp.. Since,the ch,~rge Supplied froth Cp .to Coisin turn
supplied to the load,
(11:. 16)
The average current from the battery is equal to the sum of the charge
supplied during both phases. During Phase I, the battery supplies the pump
capacitor with ~pV. ’During Phase :II, ~ the pumpcapacitor supplie~Co with CpV;
while the battery is still supplying CpV.The average battery current iS then
(a)
PhaseI
(b) +Vbat
¯ ""Vbat Cp Co lVout ¯
PhaseII
Figure 11,$ Circuit diagram of a voltage doubler. (a) Phase I. During this time, the pump
capacitor Cpis chargedto Vbat alld the output.capacitorCosupplies charge to the load. (b) Phase
II. Thepumpcapacitorcharge.sthe outputcapacitor,which.,iss, till supply,!ngt~,.e lo~a.dc,u ~. nt.
Note the voltage drop across the output capacitor is twice meoattery volu~ge (Le. meoagery
voltageplusthedropacross~ which alsoequalsthe.batteryv.ol~ge):
If .hi.’ghe.rm.u.l~ples,
of ~e
battery~oltageareneeded,tiffs doubling
circuitcanbec..asc.a~d~d,with
o~_r~o,,u~un.g
~h-~cm.is.
FromStotts, L. J. 1989.Introductionto implantible
biomedica~~ aes~gn.11~’~t, wcuztsuemces
Magazine,~ (1): 12-18.
In the ideal case that the battery current is exactly twice the load current (Eq.
(11.17)), ~ is exactly100%(i.e. Vout/(2Vbat)= 1). In reality, the energyefficiency
is actually very clos~to 100%with lowvaluesof ripple.
Whenthe battery resistance gets large, the pumpcapacitor does not fully
chargeduring PhaseI, and do~snot dischargeto the steady-state value in PhaseII.
This results in the pumpcapacitor chargingto an averagevoltage of
Vpum (11.20)
p = Vba
t -t 2 lL Rba
Theoutputvoltage is the result of, load resistance RLin series with voltagesource
2Vpump,and the output resistance Rm.Thus,
You (11.22)
t = 2Vicar "4RbatlL ~ I~ ICpf
When
the battery resistance is high, the outputresistance of the multiplier is
PUI~E OUTPUT 261
Rm=4Rb
t +l/Cpf (11.23)
Rm = n2Rbat +(n-l)/Cpf
p f>>I/RbatC (11.25)
Voltage
~.Time
electrolytic capacitor has a polarity opposite that of the + sign), and both transistors
are not in the conductingstate.
Vdd
Zener 8.2 V
~_lJ+/ _~ Output
:~t.3:lx F " to heart
(c)
Micro- I
I
processor~~_ ,,
If a pulse is applied at the base of Q1, with sufficient magnitudem place the
transistor in the conductive state, then the. voltage at the e011ector of Q1 will
effectively go to ground. Since capacitor A cannot discharge immediately, this
forces the emitter voltage of Q2 to be shifted below the ground potential of the
circuit. The base of Q2 is connected through the 15-162 resistance ~to groundand is
switched into the conductive state whenthe emitter voltage drops~ This changes the
voltage at the collector of Q2 toward -Vdd. Just prior to Q2 being switched, the
3.3-/.tF capacitor has a charge of positive Vddappearing at the negative side of the
capacitor. Whenthe collector voltage of Q2 is abruptly changed toward -Vdd, the
3.3-pF capacitor already has a voltage drop of Vddacross it. This causes a voltage
close to -2Vddto appear at the output electrode. The circuit acts as a voltage
doubler. The zener diode is in place to limit the output voltage fluctuations, while
the 33-/zF deeoupling capacitor B connected from the supply rail to ground is in
place to stabilize Vddwhena pulse is being delivered.
The duration of the output pulse can be controlled by adjusting the length of
time that Q1 is in the conductive state. A controller attached to the base of the
transistor could suP ply ......
pulses with the , duration
. .,. ~Jad’usted
.... internall b some1~ ic
¯ . ,Y Y
implementation, or externally through telemetry. As an even more basic design,
attachmentof a fixed rate oscillator to,Q 1 and the battery directly to the supply line
Vdd, would provide an output With a fixed voltage andfixed duration.
Vdd
TA
Switch ~Switch
Network~,,O__ ~2
Sensing
Vdd
Amps.
S
S
o
I
|
(a)
Ring Vdd
enable
Paceinput ~2
signal I
_ Ring
Vdd
Pac(
enable
Tip
enable
(b) Pace input-.-[TL--J-~
appears at Q2 and ground appears at Q3. A logic 1 applied to the tip enable
similarly results in Vddbeing connected to Q6 and ground connecting to Q7.
Setting the pace enable to logic zero allows the flip-flop to be clocked by the
pace input signal. Everypulse from the pace input will invert the values of the flip-
flop outputs. For instance, a pace input of logic 1 results in a logic 1 at Q (of the
flip-flop) and will place logic 0 at Q2andQ3.This sameinput, resulting in logic
at Q prime, places logic 1 at Q6 and Q%The final result of the pace input is that
Q2 and Q7are on, and Q3 and Q6 are off. With both ring andfip enablesat logic 1,
the,voltage Vddappears at the ring, and groundappears at the tip of the electrode.
A second logic 1 appearing at the pace input, signal will result in the flip-flop
outputs inverting. Thus, logic 1 is placed at Q2 undQ3and logic 0 is placed at Q6
and Q7. The end result is Vddconnecting to the tip and ground appearing at the
ring. A stimulus has been applied across the heart, as shownby the differential
electrode output in Figure 11.9.
After the pacemakerdelivers a pulse to the heart, there is a period of time where
sensing cannot occur. This effect is due to what is called the post-stimulus potential
or afterpotential. In order to sense signals accurately, it is first necessaryto remove
~ e afterpotential whicharises from charge stored at the electrode interface. ’Figure
L i0 qualitatively shows what occurs near the electrodes ~dudn’g and after a
stimulus pulse. ~i ~
- Measuring:" the re~sponse’of the heart to a.;stimulu~ pulse is:’desirable for a
nu~r’~of reasons. T~ackingthe threshold 0f stimulation, for. i6~sfance, requires
distinctionbetween an effective pulse (one whichcauses the lieartt~ contract~ and
a subthreshold pulse. Reliable sensing is impossible until the .charges resulting
from the stimulus dissipate sufficiently ~because the poststimulus potentialS are
muchlarger than those resulting from a heartbeat. In the case of a dual-chambered
pacemaker,whenan atrial stimulus occurs, detection of events in the ventricle is
necessary for the microprocessor to operate the pacemaker efficiently. The
stimulus at the atrium generally results in potentials appearing at the ventricle
sensing electrode, whichis called "cross,talk". Cross,talk must also be minimized
to allow effective sensing of intrinsic signals.
(a)
Ring
Tip
(b)
®
0
(c)
Ch
RI
Rf
Ch = er x e0 x A / d (11.26)
Z = R1 + Rf / (1 + ja~ChR
1) (11.27)
One’ method~ ~
Of active recharg6used to shorten the electrode recovery time ~is
referred to as charge dumping. Figure 11.12(a) shows an example 0fa charge
DESIGN OF CARDIAC PACEMAKERS
Applying a .biphasic stimulus has also been found effectivein reducing the
afterpotentials associated with cardiac stimulation. Figure 11.14 shows a method
for providing a biphasic stimulation. ’ Thecircuit consists of three transistors that
are used as switches; an-output capacitor Cout, a charge capacitor B, and capacitor
A which is a power-supply decoupling capacitor. Figure: l l.14(b) shows the
voltage that appears at the outputelectrode.
Assumeas an initial condition that the output capacitor and Capacitor B are
charged to Vdd/2. At time tl (see Figure 11.14(b)), Q3is turned on, and the output
capacitor is charged for .the period TI until the voltage across it is nearly Vdd.At
the same time, Q1 is turned on, and capacitor B discharges (its charge going into
the output capacitor) undl both sides are at Vdd. At time t2, QI and Q3are turned
off, and Q2 is turned on. This delivers a stimulation pulse of length Tst to the
myocardiumof the heart. At time t3, Q2 is turned off, and Q1 is ~ed on. for
period T2. The output capacitor and capacitor B recharge tt: Vdd/21Th~is prt’~ides
the second positive recharge pulse. The switches are left open until the next
stimulation is to occur.
PULSE OUTPUT 269
(a)
[Timing ]
Circultry
I
T~ T~
--
(b)
I IV I
I I I I
I T1 ITsd T2 1
Switch S 1 is then openedand $2 closed for a preset period. This supplies a voltage
pulse to node A, with the sample-and-holdcapacitor located at the input of Amp2
determining the magnitude and polarity of the pulse. Application of these
poststimulus pulses speeds the dissipation of the stored charge at the electrode
interface. Repeating the cycle a predetermined numberof times returns the voltage
at node A to a value close to Vref and the voltage at the stimulating electrode near
ground. This then allows sensing of evoked responses to occur with the same
electrodes used for stimulation.
~ ..... _ll
I
I
I
I Vref
lAmp 2 I I
Vdd
)ll I I
I
/
Node A
Hold Pt
During implantation, the pacemaker lead can receive damage which affects the
electrical insulation. This damagecan go undetected and not affect the system until
later:in the service life of the unit. Whenthe insulation is damaged,there can be a
logs of sensing or a loss of capture due to less energy reaching the heart muscle.
Since all new pacemakersare constant voltage output generators, changes in the
lead resistance will have a large effect on current flo~ through the load (Furmanet
al., 1993).
Other things mayoccur to alter the impedance characteristics of pacemaker
leads. Fracture in a conduction coil can affect operation by increasing lead
resistance. A complete fracture wouldresult in no energy reaching the heart due to
an infinite lead impedance.Impropercontact of the electrode tip with the heart wall
wouldalso result in a high impedance.
Measuringthe lead impedancecan indicate if any of the previously mentioned
problems have occurred. In some situations, whena pacemakeris being replaced,
the lead will be left in the patient, and only the pacemakerunit will be changed. In
this~case; it is importantto knowfit.he entire unit has failed or if the pacemakeris
working, but the impedanceof the lead(s) has risen to a level that prohibits proper
function ofthe unit. The physician should be able todetermine if the lead being left
in ~tli~ ’patient is in satisfactory conditioti{ Kn0w!edgeof the lead impedance~ght
alS0 possibly allow logic circuitry to. alter,the output voltage in response tO a
change in impedance. For example, the output voltage amplitude could be
increased if a high-impedance situation occurs, ensuring that enough energy
reaches the heart to provide stimulation. ~
1~.4.1 Lead-impedancemeasurementcircuit
(a) Vdd
I-clock
Counter~
~----Enable
(b)
~.-time
tknown tlead
the relationship between the capacitance and theresistances Rknownand Rlead are
shown by
C = tknown / ln(0.5)Rknow
n (11.31)
The value of Rknownis already known,so it can be used as a constant. The variable
lead resistance is then found by setting Eqs. (11.31) and (11.32) equal to each other
Rlead = Rknow
n (tlead / tknown ) (11.34)
The impedancesof the switches will generally be insignificant and can be ignored,
or they can be subtracted if their impedancevalue is known.
Figure 11.17 shows a different methodto measure the lead impedance using a
current mirror. This device consists of two transistors connected to the output after
the load (heart). Neglecting the effect of finite fl, the output current through the
load will be equal to the reference current. If the reference current is measured,
with our ___kn, owledgeof the output voltage wecan find the impedancevery simply
using Ohms Law. The equation for finding the reference voltage is
Iref = (Vou
t - VBE) / Rload . (11.35)
PULSE OUTPUT 273
Vout
iref
lout
I
protection circuitry, while Figure 11.18(b) and (c) are two current-limiting devices
.which can be placed in the box labeled Current Limiting, Device in part (a). Figure
11.18(d) showsthe current vs. voltage characteristics of the devices in parts (b)
(c).
The current-limiting device in Figure ll.18(b) is a symmetrical depletion
modedevice. The device is basically two field effect transistors connected in
series. The gate and drain of each FETare shorted to each other by a conductor.
Whenthe voltage increases at the drain, it also increases at the gate. This causes a
self-limiting effect wherethe current cannot becomegreater .than a certain value.
Referring to the current vs. voltage characteristics, whenthe drain voltage (i.e. the
voltage received by the electrode) is small, the device acts as a resistor. This
resistance should be minimalwhenin the linear range, in order to keep the load
resistance low. Increasing the drain voltage causes the depletion regions to pinch
off betweenthe gates. The current then levels off, since the gate voltage increases
in conjunction with the drain voltage. As the voltage increases, the current flowing
through the electrode stays at a maximum of 20 mA.If the voltage occurring at the
electrode is greater than +1500V, then breakdownoccurs, and the current limiter
fails. In this situation, large currents will flow through the lead. The pacemaker
circuitry is still safe though,as the zener diode will limit the voltage that app¢~ars.at
the output capacitor and shunt the current flow to ground. But the lead(s) ~nay
damagedwhenvoltages of such large magnitudes occur (in which case the patient
will probably also be damaged).
(a)
..--~ Electrode
Current
Limiting
Device ~
Generator~
(b)
(c) (d)
10 ~ 2KQ ~ 15~X)
~ osidve voltages, the transistor is the current limiter, and for negative voltages, the
iode limits the current. Both devices should be able to withstand voltage
magnitudes of 1500 V. Note that for this circuit to be effective, the pacemaker
cannot be emitting a stimulus pulse, This pulse would cause, the voltage~ at the
emitter of the transistor to go negative, and as a result, the BJT. would be in its
con~ctive state an allow large currents to flow. Thus, reception of large voltages
at the electrode should cause inhibition of the pacemaker in order for this setup to
be..effective.
IrOn"REFERENCES
11.1 For Figure ll.!(a), with the membraneresistance and capacitance 1 f~ and 1 RF,
respectively, calculate the time required to depolarize the membrane "with an applied
current of g0 mA.
11.2 Describewhyanodal stimulation of a cell requires a larger stimulus amplitudethan does
cathodal stimulation.
11.3 Explainwhyanodal stimulation has a greater ~hanceof causing ventricular fibrillation
than cathodalstimulation.
11.4 List twoconstraints involvedin output circuitry design.
11.5 Giveabrief; qualitative description of voltage multiplication using apumpcapacitor and
switching network.
11.6 Briefly explain whythe waveformdelivered by the circuit in Figure 11.9 maybe
considered undesirable.
11.7 Show,with a figure, the ion movements whichcreate afterpotentials.
11.8 Explainwhyan increase in the lead resistance is not a desirable methodfor speedingthe
dissipation of aftevpotentials.
11.9 Explainqualitatively howthe lead impedancemeasurement circuit in Figure 11.15 works.
For Rknown I kf2, tlead 0.5 ms, and tknown0.7 ms, calculate the lead resistance.
ll.lO Describe howthe device in Figure ll.18(b) can limit lead current during cardia~
defibriliation.
ll.ll Designan output circuit that includes deviceprotection and output voltage variability.
External Programming
Kevin M. Hugo
12.1 HARDWAREINTERFACE
Figure 12.1 shows the block diagram of the pacemaker programming and
telemetry interface Theleft half of the figure is the programmer--an external
device which communicatesprogrammingand teler~etr~ information with the
pacemaker. The sections of the pacemaker associated with programmingand
telemetry are shownon the right. The programmingsequence is initiated by
bringing ~i permanentmagnetin the proximity of the pacemaker,whichdoses a
reed switch in the latter. Information is then encodedinto a special error-
correcting pulse sequenceand transmitted electromagnetically through a set of
coils. Thereceived messageis decoded,checked.for errors, and passed on to the
unit’s logic circuitry. Modernpacemakersinclude the capability of bidirectional
communication.
Reed switch
Oscillator---
Pacemaker
Figure12.4 Programmer
headpositioningcircuit. AdaptedfromBaltina and White(1985).
I
. (a) 1 bit
1
8 bits
10010000
MSB LRB
8 bits
00101100 10010111I
M~B LRB I MRB 8L~B
8 b~ts
10111100
bits I[ M~B LSB
Start
fl
175kHzcarrier
(a) ~ Invalid
(d) ~ One
’~’~
Telemetry data may be either analog or digital. Digital signals are first
stream using an encoding such as shown in Figure
stream or the analog data is then frequency modulated for
282 DESIGN OF CARDIAC PACEMAKERS
An advantage of this and other encodings is that they provide multiple forms
of error detection. The coils and receiver circuitry are tuned to the modulation
frequency, eliminating noise at other frequencies. Pulse-position coding can
detect errors by accepting pulses only within narrowly-defined intervals. The
access code acts as a security key to prevent programmingby spurious noise or
other equipment. Finally, the parity field and other checksumsprovides a final
verification that the messageis valid. At any time, if an error is detected, the
entire messageis discarded (Gordonet al., 1982).
In order to increase the ~bit transmission rate, manypacing systems use a
more sophisticated type of pulse position modulation. In these, the position of a
pulse within a frame is encoded into one of a finite numberof values, e.g. 16. A
special synchronizingbit is transmitted to signal the start of the frame. Typically,
the frame contains a code which specifies the type of data contained in the
remainder of the frame.
Enable
decoding, the parameter value is placed in a RAM,at the location specified by the
parameter number. Some pacemakers have two copies of the RAMma permanent
set and a, temporary set---which make it easy for the physician ,to set the
pacemakerto a:temporary configuration and later reprogram it back to the usual
settings.
Decoder RAM
detector
Figure 12.9 shows the basic circuit used to receive telemetry data from the
cir~uit Of Figure 12,7. The coil, and Capacitor create a resonant .~ircuit tunedto
the!~,c~irrier frequency. :The signal ~,furthe¢~ hand-paSS fll~t~red and then
fre~ ueacy-demodulate~!using a phase-locked loop.
Band-pass Phase-
filter locked
loop
As stated earlier, the closure of the reed switch by the progra~mming, magnet
causes a nonmaskable interrupt to the pacemaker processor. The processor
suspends its normal operation and reverts to an asynchronous pacing mode. It
then waits for an incoming message and dispatches it. Programmingparameters
replace old valueS:in the paramete~ RAM:Telemetry can be started and st6pped
by speciat pro~ng instructions~;When the magnet ~ is removed, the process.or
restarts pacing ¢¢ith the~newparameters. ~
12.2, ’ SOFTWARE.INTERFACE
Figure12.10Typicalprogrammable
parameters.Items denotedwith asterisks (*) haveseparate
atrial
andventricular
settings.
The pacing modeis the basic operating mode for the pacemaker, as defined
by the NBGcode. In the absence of sensed beats, the pacer outputs at the sensor-
indicated rate or the lower rate, whichever is greater. The unit will not begin
pacing until the natural rate falls belowthe hysteresis rate. The upper rate is the
fastest rate that the pacemakerwill output; special options are sometimesavailable
to handle tachycardia. The refractory periodis the period after the last sense or
pace during which a new pace will not occur. The AVdelay is the interval after
an atrial sense or pace that the ventricle must be sensed; otherwise, a ventficular
pace may occur. It may be reduced with increasing heart .rate by setting the
dynamic AVdelay(Biotronik, 1993.). Chapter .9 gives further descriptions
these parameters.
EXTERNAL PROGRAMMING 285
-,~.,- Inorder to detect an atrial or ventricular event, the voltage received by the
sense,amplifier of that channel must exceed its sensitivity value. After an atrial
pace;~the ventricnlar sense amplifier is blocked for a given, duration to avoid its
m
sensing-of the atrial pulse. The polarity of the leads during,sensing and pacing
whether the voltage applied or measured is relative to ~he pacemaker case or
across a bipolar ,lead---can also be configured. See Chapter 8 for more
irfformation. Whena pace occurs, its peak amplitude and duration is determined
bythe~ pulse~width and pulse-height.parameters (Chapter. 11).
Additional functions can ,be enabled via programming. Some
microprocessor-based pacemakers allow executable code to be downloadedinto a
RAM.This allows the physician to patch the built-in code or provide a different
algorithm. This capability is-generally implementedonly in prototype or research
pacemakers (Ripart et al., 1984). The uploading of telemetry data is also
e0n~olled using special start and stop commands.
.... "Many pacemakers now have the capability of performing programmed
~finllilation through the pacemaker. The procedure spares the patient the trauma
Parameter Units ~
Prog~g parameter status Read back value "
B,atte~. vol~ge
Battery’impedance
Battery current
Low battery True/False
Lead impedance
Pulse voltase V -
Pulse current
Pulse width
Pulse charge
Pulse ener~
Sensed events
Paced events Count
Histogramof events vs. pulse rate Counts
Electr0gram . mV
Rate-adaptive sensor reading Varies
The pacemaker should be monitored closely during the first~few months after
implantation. In the U.S., Medicare Guideline I specifies coverage for pacing
systems with less than five years of clinical data, It provides for transtelephonic
follow-up once amonthfor the first six months, then e’(ery other monthuntil the
end of the third year after implantation, and monthlythereafter. Manyphysicians
use this schedule (Biotronik, 1993).
The visit begins by placing the programmer head--containing the reed
switch magnet and the programmer coils---over the pacem.aker l~ody (Figure
12.12). LEDs on the programmer head indicate when thehead ~and pacer are
within the proper range.
aker-(implanted)
Programmerhead
Several tests are routinely performed during these visits. The programming
parameters are verified and the battery status is checked. Real-time telemetry of
the electrogram and other sensor signals is acquired; a normal rhythmshOddbe
observed: Finally, the pacemaker should exhibit the normal operation of sensing
EXTERNAL PROGRAMMING 287
and pacing: Because programmers are often used by persons without engineering
of eoniput~r backgrounds,it is especiallyimportant to makeit easy to use.
~ ,’ Pacemakermanufacturers sell programmersthat are compatible with. their
ownpacers, but are rarely compatible with other pacemakers due to proprietary
modulation and encoding schemes. Programmers which are compatible with a
variety of pacemaker brands are also available. Multiprogrammerseaninterface
with, more than one manufacturerrs proprietary protocol. A single protocol for
all pacemakershas been proposed but not accepted; in that case, we could use a
single universal prograrntner to program any pacemaker. An emergency
programtner~has also been proposed which could :revert any pacemaker into a
basic VOOmodewith 70 beats/min at maximal output (Schoenfeld, 1993).
,~.~:Instead of makingan office visit for a check-up, a patient maysave time and
~oney by using :a telephone link from homeor a local Clinic to the physician’s
office:, The patient, with prior training, uses a portable programmingunit that
has the entry and display hardware replaced by a computer modem.Another
modemcgnnects the physician’s computer via local or long-distance telephone
(lrhakor et al., 1982).
12.4, REFERENCES
Man, K. C., Davidson, T., Langberg, J. J., Morady, F., and Kalbfleisch, S. J. 1993.
~-Interference from a hand held radiofrequency remote control causing discharge of an
"~ implantabledefibrillator. PacingClin. Electrophysiol.,16: 1756-1758.
Marco, D., Eisinger, G., and Hayes, D. L. 1992. Testing of work environments for
electromagneticinterference. PacingClin. ElectrophysioL,15: 2016-2022.
l~exl~nie, inc. 1990. Minix Multiprogrammable PacemakerTechnical Manual.Minneapolis, MN.
Muller-Runkel,R., Orsolini, G., and Kalokhe, U. P. 1990. Monitoring tbe radiation dose toa
multiprogrammable pacemaker during radiation therapy: a cash~ report. Pacing Clin.
ElectrophysioL, 13: 1466-1470.
Ripart, A., Fontaine, G., and Mugiea, J. 1984. Howshould the software pacemaker be
programmed during manufacturingand after implantation? PacingClin. Electrophysiol., 7:
1202-1206.
Schoenfeld, M. H. 1993. A primer on pacemakerprogrammers.Pacing Clin. ElectrophysioL, 16:
~ 2044-2052.
Thakor, N. V., Webster,J. G., and Tompkins,W. J. 1982. A battery-powereddigital modemfor
telephone transmission of ECGdata. IEEETrans. Biomed. Eng., BME-29:355-359.
Toivonen, L., Valjus, L, Hongisto, M., and Metso, R. 1991. The influence of elevated 50 Hz
dectric andmagneticfields on impl,,ml, te~ cardiacpacemakers:the role of lead configurationand
;!~: prOgramming ~ensitivity: PacingChn. Electrophysiol., 14:’ 2t14-2122.
INSTRUCTIONAL OBJECTIVES
Rate-Adaptive Pacemakers
Timothy Harvey
13.1 INTRODUCTION
Cardiac output is determined by two factors, the volume of blood pumpedout per
beat and the heart rate. Normally,the heart rate is the major factor in controlling
cardiac output. During strenuous exercise, heart rate can increase by 250%over
its resting rate, comparedto about 50%for stroke volume.If the heart rate is set
at a constant rate, for example by a VVI pacemaker, then the body will have to
adjust stroke volumein order to adjust cardiac output. A fixed heart rate places a
serious limitation on the ability of the body to adjust cardiac output to the
demandsof exercise or other physical activity. This abnormal situation puts a
heavy strain on the heart muscle. If there is insufficient cardiac output, the patient
maybecomefatigued and maynot be able to continue exercise. It is clear that,
especially for active patients, rate adaptation will increase the quality of life,
comparedto fixed rate pacemakers.
One way to determine if rate-adaptive pacemakers will help a patient is to
look at their heart rate reserves (Furmanet al., 1993). Generally, chronotropic
incompetenceis classified as whenthe range of heart rate is less than 80%of the
theoretical value. The theoretical maximal heart rate (MR) and the heart rate
range are determined from the patient’s age and resting heart rate (RR).
For example, consider a 70 year old patient with a resting heart rate of 65
bpm. Using Eq. (13.1), his maximalheart rate should
13.2 SENSORS
cost, complexity and perhaps some radiation side effects. Electrodes in the
bloodstream may also experience more motion artifacts during heart beats.
Sensors which do not need to be placed in the bloodstream are attached to the
outside of the heart. Typically these are implanted during open heart surgery,
which also adds trauma. In the body, some sensors and electrodes get coated with
more fibrotic tissue, makingthem less sensitive and perhaps less useful; Also,
some sensors require more power than others. This will lead to shorter
pacemakerlifetimes.
system like a closed-loop controller essentially let the pacemakeradjust the heart
rate to keep the pH constant.
The system would-not require any additional power consumption. However,
there are manyproblems with this system. First, there is a complexrelationship
between pH and heart rate. The electrodes are even more of a problem. The lack
of long-term pH reliability and stability in the body have led to problems. Also,
the specialized pacing lead with the pH sensor in it is a drawback. The electrode
gets covered with fibrotic tissue and maybecomeless sensitive to pH. The pH-
monitoringsystemswere unsuccessful in their initial clinical applications and are
not currently available.
de~ease the battery life of the pacemaker. Benditt (1993) estimated that these.
sensors will shorten the pacemaker life by six months. There is no SO2_sensor-
based, pacemaker available today, However, due to rapid progress with these
senators; they maybe important rate-adaptive sensors in the f~ture.
’~ ’ =~Oxygenated
~-~ ,.,~ ~’, J H~moglobin
III = ,Deoxygenated
~ ~: 1 ~ ~lemoglobin
l~ig~ire ,1~ Oxygen saturation sensor. Oxygenated hemoglobin m the right ventriclereflects the
red LED s ~ght. Conversely the dex~xygenated hemoglobi~ absorbs it. Both ~flect infrared light. ~
Venalation.~ rate
Since: effective SO2sensors: are not available yet, another ~me~odis to estimate
the oxygen, intake: Perhaps the most obvious parameter to estimate~oxygenintake
is~e ventilation rate (,Alt et al., 1993;~Benditt~,1993; ,Furmar~;etal., 1993). Since
~9156,ventilation rate has, been considered= for use ;with rate-adaptive pacemakers.
I~,has been shown~expedmentallythat.there ~is a good correlation: between heart
rate, ventilation rate, and oxygenintake. This is in both normalpatients and those
with~respiratory, diseases.
Ventilation~ rate can ;be measured by analyzing, the:imPedance between ~e
pacemaker electrode ,and the paeemaker,’s~=body. The, impedance measurement;
conlfigurations in Eigure 1,3;l(a~ b) are the techniques used for measuring
ve~i!lation~rate, with the extra lead ,method more eommon:,,As ,the patient
b~athes, the chest impedance changes. From~ichanges~,~i~, this-me~s,ured.
impedance,:ventilation rate,can !no,derived: The ventilation :sign~ has a relatively
low; frequency,= comparedto ~heartbeats and noise. The pacemaker~must select and
measure .the proper signal.component. Since .,the range of ventilation rate :and
ttem, rate :ovedap,, just simple bandpassf’dters cannot be used, Therange of heart~,
rates is ~normally 60-200 bpm, so the, pacemaker must be able,.to h~andle ~a
ventilation rate above 60,breaths/min~ Typieal~ly, the pacemakerwould filter o~!t
s~gnals,, above,60~eycles/min, ~ thus= losing~tra~k of~~the ,ventil,ation rate~ The
pacemakerwill hold the ~ensor , signal; to that correspondingtoga ventilation rate
of 60 breathslmin, until the rate. drops into the:readable range;,.,~, :
A final~,~way~todetect ~ventilation rate,is to use, envelopedetection o~ other
sensor signals (Chirife, 1991). Manysignals like the intracardiac,,ven~icular
296 DESIGN OF CARDIAC PACEMAKERS
Minute ventilation
A better approximation of the SO2sensor is the minute ventilation sensor (Alt et
al., 1993; Benditt, 1993) (see Chapter 17). Minute ventilation is the product
ventilation rate and tidal volume.This effectively measuresthe amountof air that
is being inhaled. Minute ventilation has been found to be an excellent parameter
for estimating metabolic .rate, and can detect some changes due. to stress and
fever. Minute ventilation ranges from about 6 L/min at rest to about 150 L/rain
for an athlete at maximal exertion. TO measure minute ventilation, impedance
measurementsof the chest, like those used by ventilation rate,, are used~ Envelope
detection of other sensor signals (like pressure~ and electrograms) can also
usedto obtain ventilation inftrmation in multiple-sensor systems, The frequency
of the ventilatory signal gives the ventilation rate, .and the amplitude gives tidal
volume. The bipolar lead in Figure 13.1(b) is the most commonimpedance
sensor.
The signal received by the sensor contains a lot of information and has to be
prbcessed. First, the signal will have in it a higher frequency componentdue to
the beartbeats. To alleviate this problem a signal-averaging process with a 30-s
averaging windowis used (Alt et al:, :.1993). This workslike a low,pass f’flter.
The system should also track~slow changes in minute ventilation, due. to things
like circadian rhythm and sickness. To dothiS a reference value is determined by
a second signal-averaging process, wi~.an-averaging~windowof about one hour.
The difference between the two averagers is the change in minute, ventilation that
is sent to the controller. This setup~allows the system :to track slow changes, as
designed, but can present a problem. If the ~patient participates-in prolonged
exercise, approaching one hour or more, the reference value will slide up to the
current,value. This would, canse the difference between the two averagers to
decrease eventually to zero. If the pacemaker was~allowedto do this, the heart
rate would erroneously decrease~: .To keep this from happening, minute
ventilation pacers of this-type freeze the long~ermaverager when.-the short term
value is above 50%its maximal value. Oncethe short.term value ~drops below
50%, both ~registers resume theiraveraging~. -,
RATE-ADAPTIVE PACEMAKERS 297
Q-T interval
The most successful physiological sensor for rate adaptation has been the Q-T
interval sensor (see Chapter 16). The Q-T interval is the same as the Stim-T
interval and is the time between the onset of the QRSwavet o the T wave. When
the body tries to increase the heart rate, not. only is the SA node frequency
increased, but catecholamines are also injected into the heart muscle. These
hormones shorten the contraction duration. ~During exercise or stressjthe Q-T
interval decreases. An appropriate heart rate response can then be implemented.
To measure these signals, the pacing leads used for stimulation can be used to
pick up the intracardiac ventficular electrogram:
Since this system has been successful, there isa good basis of clinical support
for it. This system also uses the standard lead, whichdoes notneed to be placed
inside the heart, They do not require any more power, except for the extra
processing, Q-T measurements respond to physiological changes~ in about one
minute. Onelimitation is that there are problemsdetecting repolarization signals.
Systolic indices ¯ . i ~ ~ ~.
S~stolic indices include stroke volumeand pre-ejecfion phase (see Chapter 16).
The pre-ejection pha~se (PEP) is the time bet~veen_~e ~oh~i ~0f ventricular
depolarization and. the opening of the ’abtedc. v~qe~ Sys[61iC indices are good
indicators f0r"rate adaptation. Impedance mea§t!re~ents ~ f¢om the p~cing
electro’de to the~.~acemak~r body c~tn-be used to mt~s~ 9hang~.,in blood pool
volume. S~roke’ vtlume will increase, and.the PEP will:decreast~ith exercise.
The pacem ~aker .can ~n adapt the heart rate to minimize the change in stroke
volume.
These pacemakers can use sta~dard:0r ;~i~hfly mbdified leads. They should
respond qtfickly to changes. However,like all impedancemeasuring systems, they
require more power and therefore shorten the battery life. These systems may
also pick up motion artifacts, " ~’
Pregsure " - ’
The cardiovaseula~ colati~0l system works to.keep,the mean. arterial blood
pressure (MABP)e0mtant (Air et al.,:1993;’Ber~ditt, 1993; Furmanet, al., 1993).
Therefore it is only natural totry to makea pacemaker that does the same. Both
RATE-ADAPTIVE ,!PACEMAKERS 299
dP/dt
Piezo
Because the major reason for changing the heart rate is exertion, th6 most
commonrate-adaptive pacemaker is the motion-detecting pacemaker (see Chapter
14). In these systems, either an accelerometer or a vibration sensor is used to
estimate activity. These elements are placed, inside, the pacemak~ercase. These
sensors detect motionand vibrations in the upper chest. After the level of actlwty
is estimated, the heart rate is adjusted according to an algorithm.
The prime advantage of activity sensors is the experience and confidence
people have in ,tbem, Furtber, placing~the rugged sensors inside ,~,e pacemaker
case makesimplanting, a rate-adapfive~pacemaker~.ofthis type no differe~t~ than a
normalpacemaker.This also protects the sensor, providing long-~termtell.ability.
Also, very little additional poweris needed to run these sensors. Activity sensors
are very fast and therefore can allow for very responsive- pacemakers,, ~e m~n
disadvantage of these sensors are their specificity. They may pick up some
undesirable,signals and~.miss other desirab|¢~ones. Imagine. a task ofwalking
staixs-~ Goings,up is harder~workthan~going~.down~!but~the,latter causes heavier
footsteps andthus stronger pressure~waves in:the cbest. ~ An activity sensor may
very well cause a higher heart rate going downstairs, comparedto going up~
300 DESIGN OF CARDIAC-PACEMAKERS
There are many other sensor techniques that may prove useful after more
research. Manyof these can be obtained from standard leads (Chirife, 1991).
Other possible cardiac indices which can be obtained by impedance measurements
are end-diastolic volume, end-systolic volume, ejection fraction dVIdt, and
ejection time. Someother possible signals that can be obtained from intracardiac
electrograms are T-waveslope, pacing threshold, and depolarization.
Manyof the sensors discussed above have good attributes and work-well, but
none is perfect (/kit et al., 1993; Furmanet al., 1993). The obvious solution is
combinemore than one of them. In almost all available single-chamber multiple-
sensor systems a fast activity sensor are combinedwith a slower, more accurate
sensor. Activity sensors have been combined wi~ .Q-T and minute ventilation
sensors. One available system that does not use. the activity sensor is a minute
ventilation and a ventricular depolarization gradient system. Here, minute
ventilation provides the prolonged response while vtntricular depolarization
gradient provides the quick response.
Activity, minute ventilation, and PEPare each used along with atrial sensors
in rate,adaptive dual-chamberedpacemakers. The first two are available, and the
PEPsystem is under investigation~
With one set of leads, manydifferent signals can be obtained (Chirife, 1991).
Sensor techniques such as intracardiac ventficular electrograms and impedance
measurements,along with. envelope detection, can. yield manydifferent signals.
Manyof these are listed in the previous section..Actually, a standard bipolar lead
could be used for both impedance measurements and electrograms. With
increasing processor power and speed, it~may.be possible that many more
combinations of sensors will be used in one pacemaker-in order to help eliminate
artifact. Artifact is erroneous signals that the pacemakerdetects and acts on. The
algorithm that runs the pacemaker may check the signals to see if they are
corrupted with motion artifact or other artifact. For example, .if the pacemaker
saw both Q-T and stroke volume decrease, something is wrong. Signal processing
could then be used to resolve the contradiction.
Refe~ence
Control Cardiac
algeritlnn Paeemake !~..~1 ~ systei~
" Sii~oe I "
I
Sensorsignal
Physiological
disturbances
Figure13.5 Disturbance
feedforward
controller. It calculatesa heart rate to cancelthe effect of
physiological disturbances. ,
allowance for heart rate. If the Q-T interval shortens due to exercise the
pacemakerwill increase the heart rate. ,~As heart.rate increases, the Q--T interval
shortens even more, causing the pacemakerto falsely detect more activity. If the
pacemakeris responsive enough this could yield anunstable system.
Combination controllers
There are somecontrollers that do not fiteither of the previous, categories. They
seem to fit somewherein between the two. For example, the minute ventilation
sensors use the sensor signal and a reference value to obtain the change in minute
ventilation, which resembles a closed-loop controller. Howeverthe algorithm
used, as shownin Figure 13.6, is an open-loop type algorithm. For any given
change in minut~ vei~tilation,, the algorithm yields one heart rate. A closed-loop
controller would .try to adjust the heart rate to keep the change in minute
ventilation value equal to zero. Ventilation rate, minute ventilation, and systolic
indices all fit in betweenclosed loop and disturbance ~feedforward.
Heart
Rate
(bpm)
Min
Change in MinuteVentilation
(impedance)
Figure 13.6 shows a multiple slope control function that;is similar tothose
used in minute ventilation pacemakers. The algorithm has about 60 different
preprogrammedslopes~ Tbe pacemakers, only need three parameters to be. set by
the physician ortechnician: ~These parameters. ~e the.~al and.maximal heart
rate, and the numberof the.desired slope. Somevariations of this-algorithm are
RATE-ADAPTIVE PACEMAKERS 303
to have three levels, or different attack and decay slopes. Other control
alg0tithms are presented in the following four chapters.
13.4 CONCLUSION
Frommany clinical studies, rate-adaptive pacers have been shownto allow the
recipient the potential for greater exercise capacity. For example, a Q-T sensor
can provide, twice the cardiac output as comparedto a fixed-ratepacer ~ Further
research into sensors like the SO2will provide:~Ore yiable ~ensb~r’ 0iJtions. The
future of rate-adaptive pacemakersis multiple sen~or, niore advanced algorithms,
and automatically tuning systems~
Several sensor techniques have been mentioned. Figure 13.7 summarizes
someof the key aspects of the different techniques.
304 DESIGN, OF CARD~C PACEMAKERS
Alt, E., Barold, S. S., and Stangl, K. 1993. -Rate adaptive cardiac pa~cin.g. Berlin! ~ Springer
Verlag.
Benditt, D. G. 1993. Rate-adaptivepacing. Boston:BlackwellSelentifiePublications.
Chirife, R. 1991, Acquisition of hemodynamie data and sensor signals for rate control from
standard pacing electrodes. PACE,14: 1563~-1566.
Furman,S., Hayes,D., Holmes,D..Jr. 1993. A prd¢tiCe of cardiac pacing. 3rd Ed.’Motmt Kisco,
NY:Futura Publishing. ~"
:":~
Lan, C., Antoniou, A., Ward,D. and Camm,A. 1988. Initial clinical experience ~with a minute
ventilation sensing rate modulated pacemaker:improvementsin. e~ercise..capacity and
symptomology. PACE,11: 1815-1822.
Lau, C., Lee, C., Wong,C., Chong, C. and Leung, W. 1990. Rate ~potisive pacing with a
minute ventilation sensing pacemakerduring pregnancyand delivery. PACE,I3~ 158-163.
Schaldach,M. 1992. Electrotherapyof the Heart. Bedim"Springe~,Verlag.
Sulke, N., Dritsas, A., Chamhers,L, Sowton,E. 1990. Is accuraterate response programming
necessary? PACE,13: 1031-1044.
Zegelman, M., Creslinski, G., Kreuzer, L 1988. Rate response during submaximalexercise:
comparisonofthree different sensors. PACE,11: 1886-1895~¯
13.6 INSTRUCTIONALOBJECTIVES
;~¢ most widely used method for rate-adaptive pacemakers is to sense body
~ovements. This approach has been met with clinical success because it does not
require additional leads (sensor is-inside the pacemakercan) so the system is easy
to implant, is relatively simple to program, and, as we will see later, rapidly
adjusts the pacing rate. A motion sensor inside the pacemaker senses body
movementswhichare primarily due to the ~patients’ foot ~fall. The pacemakeruses
this sensor signal to determine the appropriate pacing rate. Bodymotion sensing
does not directly sense metabolic changes such as anxiety, fever, or circadian
laeart rate-changes.
The motiowbasedrate-adaptive pacing rate accuracyis based on:
~’1: Sensitivity: modifying.the sensor, signal to makeit proportional to patient
~o exertion. ~
2~ .~- Specificity: pacing rate should be specific to activities that. normallyincrease
...... heart rate, in other words, pacing should not changedue.to~ activities that
normally do not cause heart rate increases (nonactivities); environmental
-factors such as vehicular travel, or other noise inputs.
3."- Rate of increase~decrease (speed of response):-appropriate pacing rate
;. in~tease/decrease following activity-changes~
To meet these requirements the motion-based pacemaker designer ~must
¢arefully~choose (Figure 14.1):
1~: A sensor that detects activity and uses nanoamperesof power~
2. ~ Signalprocessing that distinguishes between activities and nonactivities.
3~- An algorithm for generating an appropriate pacing rate based on the
Progf~ammable
rateof
increase
Programmable and decrease
Parameters
Activities Pacing
Noi3-activit|es rate
Noise signal
W = mv2
2 = 2 (14:1)
Alt et al. (1989) used tfiaxial external accelerometers strapped onto the pectoral
region of twelve subjects (six healthy volunteers and six pacemakerpatients)
emulate accelerometers inside a pacemaker can implanted in the pectoral region.
Mianulli et al. (1989) and Benditt et al. (1991) found strap-on acceleration-sensed
rate-adaptive pacemakers are valid representations of implanted .rate-adaptive
pacemaker response when standardized techniques are used. Vertical (up and
down for a standing human), horizontal (forward and backward), and lateral
(side to side) accelerations were recorded for a variety of exertion levels,
activities, and noise sources. The piezoresistive acederometers from Endevco
(2262 CA-200) had a measuring range of +~00 g, a much.wider range than they
measured (:~2 g).
The following sections describe the results from the study performed by Alt
et al. They performed a Fast-Fourier Transform on the data to provide
frequency and amplitude information. The subjects had a wide variability in age
(22-78), height (154-196 cm), and weight (58-100 kg) however, the
showedno direct influence due to these parameters.
Walking
For various exertion levels walking on a treadmill, the acceleration frequency for
each axis was linearly related~to the step:frequency (numberof times the foot hits
the surface per second) ~ The frequency of the higbest:aceeleration amplitude was
found to be in the range from 1 to 4. Hz for speeds varying from 3.2 km/h
(0% grade) [a normal walk] to 5.6 km/h" (16% grade) [a near run] (Alt et
1989).
Harmonics
The peak amplitude occurred at the step frequency: Harmonics of the step
frequency are also significant, but less than the fundamental step frequency. For
instance, the horizontal axis, acceleration third harmonic for a Walkingspeed of
4,2 km/h is about 40%of the p~ value, while the componentat 1.0 Hz is about
20%of the peak (Alt et al. 1989).
speeds were tested for ascending and descending a flight of stairs. Acceleration
variation with speed of ascent or descent was insignificant for the lateral and
horizontal axis: The horizontal axis showedsimilar meanacceleration levels for
ascent and descent, The lateral axis mean acceleration ascending was almost
dbuble that of descent. The vertical axis meanacceleration had a variation with
speed, but mean acceleration levels were similar during ascent and descent
~Alt et at. 1989).
Figure 14.2 Piezoelectric elementbondedto the inside of the pacemaker can. Bodymotion
causespressurefluctuationswhich~ausethe can to deflect whichbendsthe sensorto.producea
voltage.Theleads fromthe piezoelectricsensorare connected
to the pacemaker
electr0ni~s.Thisis
onepossiblelayoutfor the pacemaker components.
310 DESIGN OF CARDIAC PACEMAKERS
The body acts like a mass-damper system with a resonance around 10 Hz.
According to Deickmann(1957), the mechanical energy originating from the feet
propagates through the bodyas pressure waves, especially if the energy is in the 5
to 40 Hz region.
Wehave not found published raw data on the crystal bonded to the can like
the study done by Alt et al. (1989) on accelerometers. Manystudies have tested
the heart rate performance of piezoelectric-based pacemakersbut did not measure
raw sensor output. The studies will be presented later in the chapter. Anderson
and Brumwelt.(1984) reported that "data has been collected which shows that the
response of the typical human body which results from mechanicalactivity
related to the physical activity of the patient such as pedal impacts from walking
or running are centered around approximately 10 Hz."
The noise components, similar to the signals from the accelerometer, occur
at a variety of frequencies and amplitudes. A piezoelectric sensor does not
respond to de signals; however, direct pressure on the body around the can will
increase the effective mass of the can which will magnify signals. For instance,
the cardiac signals from the heart (-- 1 Hz) mayappear~ as an activity whenthe
patient is lying prone (i.e. sleeping).
The piezoelectric sensors like classical accelerometers should have a stronger
signal descending stairs than ascending because the feet are hitting the ground
harder. An activity like biking should produce amplitudes less than walking or
running because very little impact is present. However,there is body movement
and signals will be produced.
The type of footwear mayaffect the amplitude of crystal and accelerometer
signals because some shoes (i.e. running) mayabsorb the foot impact more than
other shoes. However,the step frequency is not affected by changes in footwear.
~
Compression1
Fome
Figure 14.3 A piezoelectric bender element induces a voltage (several millivolts) when
mech~.
~a~ically
deflected.
Accelerometers use piezoelements and a seismic mass and/or a cantilever
beam. It is analogous to a weight on a spring which is connected to a frame. As
the frame moves, the mass will tend to stay at rest until the spring, being
stretched, can exert enough force on the mass to makeit move. The piezoelectric
bender element is one example of a sensor used in accelerometers to measure
displacement changes of the base. A beam 6f piezoelectric or piezoceramic
material ig attached to the seismic mass as shownin Figure 14.4. Electrical
contact is madeto electrodes on the top and bottom of the beam. The base of the
accelerometer is attached to the test object whosemovement,say, is perpendicular
to the bender and will exert a force on the mass of the beam. The mass’s inertia
will resist movement,;causing the beamto bend. The electrodes detect the electric
signal which can be conveyedto an amplifier.
Cantilever beam
Brass
Sensed
Base accelerations
Piezoceramic
material
1 ACsl
fcH - 2/i:CflReffl - 2/1:C~ (14.1)
where, Reffl = llfsCsl,fs = the clock switching frequency, and Csl = leakage
capacitance.
Csl
The charge sensitivity for the Picochip accelerometer determines the charge
induced by an acceleration (g), and then the amplifier first stage converts charge
to voltage. The gain of the trust.stage charge amplifier is
GI = Cfl ~ (14.2)
RATE’ADAPTATION BY MOTION 313
1 fsCs2
feL- 2~’Cf2Reff2- 2~’Cf2 (14.3)
Ci V
G2 -- Cs2 V (14.4)
Solution: ~
The charge produced by the Picochip for an acceleration of 2 g is
2pC 2g = 4pC
g
For the high-pass filter, given fell = 1 Hz, choose fs =1000 Hz and
Csl = 0.25 pF to minimize Cfl so G1 is maximized.
Solving for Cfl in Eq. (14.1): "Cfl = 40 pF and
1 - V
G1- 40 pF = 0.025~
-5V V
G2-0.1 V - -50~
For the low-pass filter, given fcL = 5 Hz, choose Cs2= 0.25 pF and
fs =1000 Hz. Solving for Cf2 in Eq. (14.3): Cf2 = 8.0 pF.
Solving for Ci in Eq. (14.4): Ci = 12.5 pF.
Silicon
substrate
Sideview
Accelerations
I ~
~ ~.T°P view
Srees~n:.~O
rnsdu ct ° rMass
~
Integrated Semiconductor
Accelerations signal resistors
~J in and out ~ndlioning
of page
(a) (b)
(c)
R2
Vo= RI +R2 (14.5)
e0 erA and Cb - e0 er A
Ct-d-x d+x (14.6)
The difference between the top and bottom capacitor voltages is proportional
to the acceleration. Therefore, two capacitors can be used as the input impedances
to~a-differential switched-capacitor~amplifier (Figure 14J6(b)) whose~output
proportional~to acceleration.
:~ Piezocapacitive accelerometers, in the +_2 g range change capacitance by
about. 2-5%:with a total displacement around one 1 #m. This makes .the sensors
difficult to mountbecause a small change in the thin ceramic walls will produce
capacitive offsets. Piezocapacitive accelerometer signal conditioning is also more
316 DESIGN OF CARDIAC PACEMAKERS
~xI Accelerations
Siliconsprings
SwiChing
netwock
Ct
Switd
Vo o¢(Ct - Cb)/Cf= acceleration
Cb
(b)
Figure14.7 (a) Apiezoeapaeitiveaecelerometerwherean internal massis suppot~,.dby four
silicon springsin betweenthe upperandlowercapacitors.Thecapacitancesof the upperandlower
plate vary as a functionof acceleration.(b) Aswitched-capacitor
differential amplifierwhose
outputvoltageis proportional
to acceleration.
Chevalier (1985) reported clinical studies which suggested that the sensor could
be placed in either the pectoral or abdominal region. Andersonand Moore(1986)
performed a couple of activities with the sensor located at both locations. They
found a slightly larger signal from the abdominal location than the pectoral
region, but both were within the range of the threshold programming.
Pacemakermanufacturers are constantly trying to reduce the size and weight
of the pacemaker and its can. The size of the pacemaker can may vary the
coupling (how the can transmits movement)between the patient, sensor, and can.
Therefore, each modelwill be built with a slightly different internal gain so that
heart rate response does not change. The Coupling mayalso be affected by the
patient’s upper body structure or how the pacemaker rests in the patient
(pacemaker may movearound after implanted).
This section describes the first motion-based rate-adaptive pacer patent to show
the basic operation of a rate-responsive pacemaker. Then, the operation of
Medtronic’s various rate-adaptive motion-based pacemaker models are discussed.
In 1992, rate-adaptivepacing systems accounted for more than 60%of Medtronic
pacemaker units sold around the world (Medtronic, 1992);
Programmable
S!opes]
Piezo- [Bandpass
H Comparator~ Integrator I
st electric -"-’[ 1< f< 20 Hz
Activitie sensor Activity
Non-activities estimate I Voltage
noise
output
bandpass Threshold
~al
Pacing
I I\/Vl I\. t Time rate
output
comp-
arator
.I- .... Dashedline
output
integrator I I,. _ _ _ J,- I hashigher
!~3-’---" programmed
minimal ~ slope
pacing Time
rate
Activitrax ® "
Activitrax was released in 1984 as the first rate-adaptive motion-based
pacemaker. The motion-based rate, adaptive technology has been used by over
100,000 patients between 1984 and 1989 (Medtronic, 1989). Similar to the patent
by Anderson and Brumwell (1984), accelerations produce pressure to bend the
piezoelectric sensor which produces a voltage. The programmer selects .from
three minimal pacing rates ~(60,~ .70, or 80ppm),.and three’maximal: rates (100,
1~25, 150 ppm). A bandpass filter removes ventilatory, cardiac, and other tow-
frequency noise componentsfrom the raw, sensor, signal. A.threshold, detector and
counter eotmts excursions per second above a programmable threshold (Figure
14.9): Zero counts/sis no detected activity above the threshold and 15 counts/s is
the heaviest ~activity. The programmerhas the choice, of three threshold levels.
The counts/s are mappedto a target pacing,rate using one of ten.programmable
curves that mapan activity estimate (level) to~ a target pacing rate. The target
pacing rate is ~ the pacing rate that should be obtained for the-present activity
estimate (level),~ Thetarget rate~ is updated,everytwo seconds,: but the pacing rate
does not change as rapidly because the~pacing rate,is smoothedby the pacingrate
time constants,-which are,diseussed~in the following paragraphs. The activity
estimateto :target pacing rate.mapping could be implemented with a 4ook-up
table. Figure 14.10 showsthe shape of five rate-response curves.
RATE-ADAPTATION BY MOTION 319
Programmable
thresholds
Threshold , J Determine|
electric I [ Bandpass detector
nmssurel----’t 7 < f < 45 Mapping~ pacing
and rate
’ I /
A ~ensor [ ! (Hz) counter Activity
N0n-activies estimate Stimulations
noise (counts/s) per minute
Thresholds:
Outputof ~ .... _Threshold
level counts/s
bandpass". high 3
filter l’~/" ~-t’~- ~-I’~-7- Low 7
v ~- "~-tr - - - Medium medium
~_~V~i~ "f~"
low 9
V VV ~V~S time
Figure 14.9 Activitrax rate-adaptation block diagram.Countercounts activities abovea
programmablethresholdsetting. Rateresponsecurvesconvertthe activity estimate(counts/s)to
target pacingrate. Thepacingrate is calculatedbasedon presentrate, target rate andrate of
increase/decrease
whichdetermines howfast the pacingrate shouldchange.
The next pacing rate is calculated based0n the present pacing rate, target
pacing rate, and the rate of increase/decrease. The rate of increase/decrease is
howfast/slow the pacing rate can change as a function of rime. The~Acdvitraxhas
set acceleration and deceleration times, which determine the rate of
increase/decrease as a function of rime. The acceleration and deceleration times
vary the rate of pacing rate change as a ftmcdon of time (how fast/slow does the
pacing rate Change).’ The,acceleration/deceleration time is the time it takes to
achieve 90%of the difference betweenthe present rate and the sensor-iiidicated
target rate. The 90%difference is calculated using pacing intervals and not by the
pacing rate.
Target I Curve#
; pacing .150t
rate . - .... ~-~
~
;.=.=T-. ~-=---~: Maximal
(ppm) 140.]- 9f ~.~"~ 7 ~)
~ pacingrates
/ / .... j_25_ (selected by
120 - ..... , ..... -.--’Z: ....... physician)
110
100"
80"
-0 3 6 9 12 15 Activity
~ .¯ -: estimate
,. (counts/s)
~’,.’.gurei4.10Five of the~tenprogrammable
rate response
Curves. Thecurvesmapthe activity
ebtimate(c0ums/s)io a target pacingrate.: Reprintedwith p~rmissionfromM6dtronic,Inc.
©Medtronic, Inc. 1986.
320 DESIGN OF CARDIAC PACEMAKERS
After an increase in the activity estimate, the rate increase curve resembles a
relatively fast (smaller time constant) first-order system step response, while
after an activity decrease the pacing decrease curve resembles a slowly decaying
response. If activity level stabilizes, the pacing rate maintainsits steady state value
until the activity level changes. Equations (14.7) and (14.8) showthe response
dependent on the present pacing rate (ppr), target ,pacing rate (tpr) and the
constants. The time constants are proportional to the acceleration/deceleration
time and determine the. rate of increase/decrease. The increase (,i) and decrease
(~d) time constants are nonprogrammable for the Activitrax but have
programmable values for other Medtronic pacemakers such as the Legend®.
Figure 14.11 shows a first-order response to a step change in activity level. The
actual pacing rate is updated once per second and therefore would be constant
over the one second interval.
Targetrate
Pacing .dudngincrease,
rate
(ppm) _ Jp.resent rate
._._~w..-Present
rate dunngdecrease
dudngincrease
Time(s)
where
npr = next p~acing rate
ppr = present pacing rate
tpr = target pacing rate
*i = rate of increase time constant
*d = rate of decrease time constant
One disadvantage of a device depicted in Figure 14.10 is that, for maximum
activities along some curves, the rate can not reach the programmedmaximal
rate. An example of this is: curve #5 is programmed along with a maximal
pacing rate of 150 bpm. The maximal rate is not achievable because the maximal
curve flattens out at 140 bpm. The inability of the Activitrax to achieve the
maximal rate was confirmed in several studies including Mond(1993) and Lau
(1989). ~,h_is disadvantage was fixed by using~linear mapping curves
Medtronic s next gener~0h pacemakers.
RATE-ADAPTATION BY MOTION 321
Toff et al. (1987) tested the Activitrax with typical settings aboard
aircraft. They found that pacing rates were not unduly affected by travel in fixed
wing aircraft although large rate increases did occur transiently in the smaller
aircraft. Sustained rates were observed in rotary wing aircraft and hovercraft.
These rates are often well tolerated and often unnoticed as long as the patient is
awake.
Kubischet al. tested the Sensologi n a sports plane and reported that jostling
from turbulences causes rate increases. The affect is highly dependent on the
chosen programmingparameters. Further investigation of air travel with respect
to the programmedparameters Should be carried out.
Medtronic’s present rate-adaptive pacemakersdo not distinguish the intensity
of activities once they are over the minimal threshold. The amountof activity is
indicated by counting activity occurrences, even though one count may have a
muchhigher intensity over the threshold than another count. The effect on the
pacing rate may be minimal for most exercises because as the activity level
increases, so does the cyclic frequency. This was shownby Coates and Meadein
1960. Lau et al. (1988b) showedthat a better correlation ~betweenpacemakerand
sinus rates could be achieved by-using the strength of vibration rather than a
process of peak counting. However, an accelerometerwas used to determine the
strength ~ of vibration and this was compared to the Activitrax’s heart rate
response. Therefore, this study is somewhatinconclusive because two different
variables were sensed and different signal processing is used in the devices.
Lau et al. (1988a) found that the Activitrax responds rapidly and accurately
to variations in walking speed on a constant slope, but whenthe treadmill slope
~ increased the pacing rate did not change even though the energy level was
~gher.~:This ~effect is, similar to stair ascension and will probably be ~xperienced
by all motion sensing pacemakers because the impact frequency and magnitude do
n~t changeproportional to grade levels so the heart rate will’ not be dependenton
grade.
Synergyst ®
The Synergyst line uses Activitrax’s rate-adaptive technology, but is a dual
chamberpacemaker,pacing both the atrium and ventricle in resp9, nse to activity.
Programmable
Programmable rateof
thresholds increase/decrease
IPiezo- Bandpass
HThreshOlddetector
electric , I Determine/
, ,~ pressure 7 < f < 45 ._~-~
MappingS] pacing I
and rate
Activitiesl sensor (Hz) =
counter Activity ’l /
Nonactivites estimate Pacing
noise (counts/2s) rate
(ppm)
Outputof -’
bandpass.
~.. ~: V ~_ _ ,Jwindow
’~Sh~dedareais
lowthreshold
window
Figure 14.12 Legend®and Elite® rate-adaptive block diagrams. The activity estimate
(counts/2 s) is producedby counting the numberof peaks outside a programmed threshold
window.Theactivity estimate is mapped to a target pacingrate througha programmable rate
responselinear curve. Thepacingrate is determinedusingthe presentrate, target rate andthe
pro~ammablerate Of increa,~e/decrease
settings.
Target Curve#
pacing
rate 150’ Maximal
(ppm) 140- pacingrates
120-
100-
80
60
I I I I I
0 3 6 9 12 15 Activity
estimate
(counts/s)
Figure 14.14 shows that the Biotronik Ergos operates similarto Anderson and
Brtlmwell (198~t) because it uses a pressure sensor, bandpass filter, comparator
and integrator. The raw signal is filtered .with a second-order bandpas~ filter
centered at 4Hz. Figure 14.I5 shows two of three programmable thresholds;
which determine the level of a cornparator which outputs an’activity estimate.
This is a fixed height pulse width proportional to the area"of the filtered signal
above the programmedthreshold. Then, the activity estimate is integrated .to
produce a signal proportional to the pacing rate. The slope of integration is set by
a programmablerate of increase gain. In the absence of pulses, the pacing rate
d,~creases proportional to the Programmedrate 0f decrease. The programmerhas
a’ehoice of three decay’time Settings (30, 50, Or 100 s); which determine how
long it takes for the pacing rate-to de~rease downto the target pacing rate.
Programmable
Programmable ratesof
thresholds increase/decrease
Sensolog III®
The piezoelectric sensor, signal is first multiplie d by a programmable gain
(Figure~ 14.16) (Pacesetter,,1990). A bandpass filter~ with a range from 2-50
(Stangl et al., 1989) is used .after the gain, and then an actiwity estimate
produced after a programmablethreshold (choice of 5)i~ Used tO reject any low-
level signals. The activity estimate is mappedto a target rate using one of eight
Programmable
rate of increase
Programmableand decrease
Pro-rammable
~1 slopes ! I
IP
Piezo- I threshold. ~ I’ ~ /~ |
,~ ’ I- ¯ ,/ ! Determine/
I--~ Threshold ~ Mappingl’~-~I pacing ~
Activities] sensorI ’ I
/~ctivitv I Irate I Pa.cing
Target rate
Non-activities estimate
noise [
,ressure pacing
,rate
Viewfromfront
Viewfrom of patient,s,
right shoulder c~e~ ’ ~ r
~- Skin -, = Battery
/ =
Ribs
~ ~ Accelerations
Accelerometer
Acce~erometer
~
(senses
accelerations
intoor
out ,of pagb)
Figure14,17Accelerometers
detect upperbodyaccelerationsin the horizontalaXis(antefior~
posterior, forward-backwarddirection).
326 DESIGN OF CARDIAC PACEMAKERS
DC
offset .Modulation, I ~a<np<~is(~ ~__.
.I Accel- Amplification, I kl
-I e rati°n ~ Filtering Modulationl--- ~ H
Activities ! Sensor Harmonics,,and I ’
Non-activities Demodulation |
noise
Activity
estimate Arate
~ Targe
(~ ~ + I Rate I _ rate t Oeterminel
~l ’’’~J
mean t-*~ response I-T’*~ pacing" ~ ’rat~
~__1’- Inumber |~ .]+ Irate . /
Threshold Basel]
rate,
level ,~ ~" Rateof
increase
and decrease
Intermedics was assigned dual sensor patents, (Alt (1991); Adkins anti Baker
(1989)) that incorporate a piezoresistive accelerometer. Alt (1991) describes
RATE-ADAPTATION BY MOTION 327
use of a piezoresistive accelerometer and Adldns and Baker’s (1989) describe the
use of a piezoelectric accelerometer. Section 14.1.3 describes AltOs (1989) work
on the basis for detecting humanactivities with an accelerometer. Alt (1991) also
shows some time domain accelerometer raw data during walking, running, hitting
thorax, coughing, laughing, and during a car stopping. Frequency domain data
are also included during walking.
Both patents show the use ofa bandpass filter to isolate frequencies between
0.3-4 Hz. Various long algorithms to incorporate dual sensors have been
described in the patents. The acceleration algorithm relies’only on. acceleration
changes and not on absolute acceleration values (Alt etal. 1988). Alt (1991)
explains the details of this algorithm. ~ ~¯
Intermedics Dash® and Relay® pacemakers use accelerometers and pass
accelerations between 1 and. 4 Hz (Figure 14.19(a)). Then the signal is integrated
to produce an’ activity estimate. The activity estimate is mappedto a pacing rate
response using a triphasic curve (Figure 14.19(b)) (Intermedics, 1990).
curves have a steep slope at the onset of activity, level off, then have a steep slope
again during high levels of activity. Accordingto Lau eval. (1992) this maintains
rate stability during ordinary workloads. The programmercan .choose between
several onset slopes, ordinary workloadlevels, and ten high-activity-level:slopes.
"~Programmable
low and high
rate ,of increase
IAccel- Band
--
. --
. ~
ass I I =, IDi~terminel
P _, ,]-"llntegrator I’--’tpai~iiag -I
sensor ~1erati°n , ~ .Actlvitylrates.~ iva~!ng
ActivitiesI
.Nonractivities
:estimate 17ate
noise
(a)
Pacingrate
of increase
Tailoredrate
lowerrate of increas
e : _.,~:
.. . activity estimate
(b) -~ ~
Humenet al. (1985) showed that the Medtronic’s early motion-based pacemakers
significantly, improved cardiac performance and increased exercise tolerance.
This was confirmed by Faerestrand et ,al. (1987) and Beyersdoff et al. (1986).
Candinaset al. (1991) studied the patient well being during routine daily exercise
for patients with e~ther ma Activitrax or Sens~10gpacemaker. The conclusion was
that patients who were fully dependent on pacemakers showed greater benefit
from rate response pacing compared to those who only intermittently required
pacing.
was significantly less than the intrinsic heart rate. At 100 steps/min, the pacing
rate change was about one-tenth the intrinsic heart rate change. This test also
showedthat the pressure-sensed pacemaker inaccurately had higher pacing rates
descendingthan ascendingstairs for rates of 80 and 100 steps/re_in.
Res and de Boer (1990) compared the Ergos 2® and the Activitrax pacemakers
during a multiphase protocol. This study showed that the Ergos 2 followed the
intrinsic rate much better than the Activitrax. The Ergos 2 did perform much
better during bicycling. However, the Activitrax’s performance may have been
improved by programming the pacemaker better because it is biased about
25 beats/rain belowthe Ergos 2 and the intrinsic heart rate. The poor pacing rate
during bicycling and lower rate increases may have been improved by lowering
the Activitrax activity threshold or by choosinga quicker activity response.
Stangl .et al. (1989) compared the Sensolog and Activitrax pacemakers
during bench tests and in patients during various activities.~ The Sensolog
pacemaker increased its pacing rate for a constant speed, increasing slope test
performed on a treadmill. The Activitrax did not discriminate between slope
levels. This could be due to the Activitrax’s ~ algorithm of counting steps/s and
disregarding the intensity information over the set threshold. Sensolog’s
algorithm includes integrating the signal over’ the set threshold so that intensity
information is included. A disadvantage to this technique, as supported in this
Study, is that the pacing rate increases more whenthe patient is lying in a prone
~sition. Various noise sources from travel were found tO ~!Jave~e same minimal
effect on both devices. ~ ~
In contrast to’ Stangl et al. (1989), Kubischet al. (1988) did not find a
increase of the Sensolog during treadmill exercise ~with an-increasing gradient
while keeping the same speed. Kubisch et al. chose the same rate responsive
parameters found useful in the patients’ daily life while Stangl et al. did not
report the l~rogrammingparameters chosen.
Lau et al. (1992c) studied the effect of footwear on these two devices. The
Sensolog’s signal integration is more susceptible to the type of footwear because
the impact intensity over the threshold changes the pacing rate. Medtronie’s
threshold:, detection method is not used for amplitude.iiiformationbut.-to 1)
~
distinguish betweennoise and activities and 2) correlatethe activity frequency to
heart rate.
Bacharach et al. (1992) compared the Legend (pressure sensed) and Excel
(accelerat.i0n sensed)~during various activities using nominal ~programmed
settings. Excel’s pacing rates were Closer to the ii~tr~nsic heart rate than the
Legend’s rates du~ng treadmill walking, bicycle erg~¢try, and stair ~nsi0n.
Lau et al. (1992b) found that acceleration-sensed pacemakersrespond better
than pressure-sensed pacemakersduring walking, jogging, and standing..
Matula et al. (1992) comparedpacing reSgQa~Sof the Relay (accelerometer-
sensed), Sensolog, and Activitrax (pressure,sensed) during a ~ead~l!exercise
with changing speeds and slopes (Fibre 14.20). The basic rate for all pacemakers
was set at 70 bpm, and the pacemakerswere calibrated to yield a pacing rate of
330 DESIGN OF CARDIAC PACEMAKERS
.... Alt et al. (1989) showedthat acceleration signals from about.l-4 Hz should
be used for activity sensing, but a slightly higher cut-off Couldbe used effectively
because this would incorporate higher frequency activity signals~ The horizontal-
axis (anterior-posterior) accelerations are currently sensed. The lateral-axis also
shows distinguishable acceleration variations during biking and large variations
during stair ascent/descent and walking. The lateral and vertical axis has probably
not been used because it maybe harder to mount the sensor inside the pacemaker
s~eways and also if the pacemaker rotates inside the body the detected signal
levels maychange.
14.7 REFE~NCES ~ ~
Analog Devices, One Technology Way, P.O. Box 9106, Norwood,MA02062-9106. 1-800-262-
5645.
Anderson, K. 1986. Sensor pacing - research leads to major breakthrough in rate-responsive
pacem__aking. MedicalElectronics, 10:. 89-93..
Anderson,IC and Brumwell,D. 1984. Rate adaptive pacer. USpatent 4;428,3781
Anderson, IL and Moore,A. 1986. Sensors in pacing. PACE,9: 954-959. . .
Bacharach,D, I-tilden, R. Millerhngen,J:, et ’hi. 1992. Activity-ba~edpacing: comparisonof a
device using an aceelerometerversus a piezoelectric crystal. PACE,15:,188-196.
Barold, S., S. and Mugiea, J. 1993. Newperspectives in cardiac pacing. MountKisco, NY:
Futura Publishing.
Benditt, D. G., Mianulli, M., Fetter, J., etal. 1991. Anoffice based exercise protocol for
predicting chronotropie response of activity-triggered rate-variable’ pacemakers.Am.
CardioL, 64: 27-32.
Beyersdorf, K. 1986 Increase in eardiac output with rate-responsive pacemaker.Ann. Thoracic
Surg., 42: 201-205.’
Candinas, A., Gloor, H., Franz, W, et aI. 1991. Activity-sensing rate responsive versus
conventional freed-rate pacing: a comparisonof rate behavior and patient well-being during
routine dallyexereise. PACE,14: 205-213.
Chevalier, P. 1985. Improvedheart rate and exercise performance with an activity sensing
pacemaker.(abstract) PACE,8: 22.
Coates, J. E., and Meade,F. 1960. The energy expenditure and mechanical energy demandin
walking. Ergonomics 21~97-119,
Deiekmann, D; 1957. Ein fluss ventikaler maeehanisehen swingungen auf den:~Meneh.
Arbeitsphysiol, 16.
Faerestrand, S.;Breirik, K.,andOhm,O. 1987:Assessment of theworkcapa~eity and
relationship betweenrateresponseand ’exercise: tolerance associatedwithactivity-sensing rate
responsive ventricular pacing. PACE,10: 1277-1290.
RATE-ADAPTATION BY MOTION 333
14.1 List the three pacing rate accuracy factors for a rate-adaptive pacemakerwith a motion
sensor and give an exampleo~ each.
14.2 Givethe frequencyrange that shouldbe used for accelerationsensingaccordingto/kit et al.
(1989). State whetheror not the hodzontalaxis acceleration can distinguish betweenstair
ascent/descent.Repeatfor different exertionlevels duringbiking.
14.3 Designa switched-capacitoramplifier with a total gain >400and a baiidpass falter topass
the frequency range from 0.5-4 Hz for the Picochip piezoe[e_c.tric acceierometer by
Endevco:The clock frequency ’range is 1’5 kHz and capacitors values must range from
0.25-50 pF.
14.4 Describethe operationof a piezoresistive accelemmeter.
14.5 Describethe operationof a piezoeapacifiveaccelerometer.
14.6 Describeand illustrate howa piezoelectric cry~stal is used to sense vibration andpressures
in pacemakers.Explainhowthis configuration is diff,’rent fro musing an ac;cel~rometer.
List two advantage~and give two examplesofinapp~priate pacing respot~ses:
14.7 Explainwhatthe threshold levels are used for. Giveah examplefor whyone patient’s level
wouldbe set lowand anotherpatient’s level high.
14.8 ~De,scribehowa disadvantageof the ActiviWax~ Wasfixed in the next:generafignLegend.
14.9 Explainthe,differences betwe~,~nthe methods~ised~to.~nerateaft activity estimate for the
MddtronicLegend, Bidtrotfik Erg0s, and P ,a~settet Seiisolog llIi ~ .. ~ ¯
14.10 Explain tl/e differences betweenthe meth&lsus&l io generate an activity estimate for
Car~acPacemakerand Intermedics pacemakers.
14.11 List three factors that dete~ne the pacing rate for the MedtrOnie’Legend pacemakerand
explain howthe factors are determined.
15
Chapter 13 notes that patients whosuffer from atrial flutter or,fibrillation, SSS
(sick sinus syndrome)~:sinus bradycardia, .or other types of sinus node disease are
incapable of benefiting from atrial-sensed rate-responsive pacing,:;Although
according to Starling’s law, cardiac output can vary dueto an increase in stroke
volume,L sueh~ehanges are limited to a maximal increase of less than 50%,
whereas metabolic demandsmayrequire a four to fivefold increase (Houdas and
Ring,~ 1982; Furman,1990). Therefore, in order, for such patients’ heart rates to
vary, their pacemakers should be able to make,accurate and reliable changes in
pacing rate in response to direct or indirect metabolic indicators which parallel
normal heart rate (HR) modulation., One of these, indicators is eentrat venous
blood temperature (CVT).
In order to understand howCVTis used as a measurand for rate adaptation,
it is both useful and important to understand how the body produces heat and
regulates temperature.
Endocrineand
otherinfluences I I Brai~ I
I
Therm°sens°rs Craerd:Op~na~e~lar
]I I
constriction dilationVaS°"
[ I
I Shvedngand
metabolicheat
production
I
Swe
ringI
Body
temperature 12
I
I
I Respiration Environmenti
Figure 15.1 The basic components of the negative feedback thermoregulatory loop Tset
represents the hypothalamus’s changing thermostat. From Houdas, Y., and Ring, E. F. J. 1982.
RATE ADAPTATION BY TEMPERATURE 337
Circadian variation~
CVThas a circadian (diumal)~ fluctuation: the CVTfluctuates, approximately
x~0.25,C around the baseline, reachinga minimumat night and amaximal peak
during the day. The increase _ in HRduring the dayis at.least partially due to the
increased~ metabolic need while awake. Floweret, the CVTreaches a maximal
peak generally during mid to late afternoon whether or not a person is active or
resting. This peak has been shownto occur continually even during experiments
where the, subject was completely isolated from factors such as light, time, and
noise. Thus the circadian variation is not o~y a function of increased activity but
also seems to be an intrinsic property of the bod~ (Houdas and Ring, 1982).
~2imilarly, HRalso varies in paralle l, ,with CVT(Sellers,. 1987), approximately
O bea~s/min (~pm), reaching a minimumat ~ght, a rise during W~inghouri,
and a maximal peak during late afternoon. BothHRand CVThave a direct
correlation with metabolic rate (Hanser, 1984; Berggren.and Christensen, 1950;
Altet al, 1986). Figure 15.2 shows the circadian C~I" and FIR fluctuations over
a 24-h period with periods of exercise supedmpesed.
Exercise . " .
Aithoughcircadian variations in HRa~e significant, the .need for an in,reded HR
d~fig exercise in the~absen~e .of normal sinus function initiated. ~e development
0f.rat~-adapfive p~cemakers~HRs can increase between 25 and 90 bpm at ~owto
moderate exercise levels (20-150 W), which corresponds to exercises such
bicycle ergometry at approximately 50-60 revolutions/rain or trea ~dm.ill tests at
approximately, 2-5 km/h.. ~is value, of course~ can .vary depending on the
physical condition, of the individual. At a given workload, th~ increase in HRas a
~unCtion of time is linear to exponential in shape until a point at whichthe FIR
plateaus (Laczkovics, 1984; Fearnot, 1988). The rate of increase in FIR as well
the level of the plateau is dependent on the individual’s oxygenuptake at. that
workload relative to the maximaloxygen uptake (relative workload) (Aft et al.,
1986; Munteanu, 1986).
0.75--
~- 0.0 ~
~ O0 l: O0 ~: O0 S: O0 ~5:O0
TIHE (HOURS) ¯
Figure 15.2 Thecircadian fluctuations of CVT and HRof one subject, averagedover 10-rain
intervals, with periodsof exercisesuperimposed. TheHRandCVT are at a minimalvalueduring
the night andat a maximal
valueduringperiodsof exerciseandin the midafternoon.FromSellers,
T. D., Fearnot,N. E., Smith,H.J., DiLurenzo, D.~ M.,Knight,J. A., andSchmaltz,M.J. 1987.
Rightventricularbloodtemperature profiles for rate responsivepacing.PACE,10: ,467-479.
338 DESIGN OF. CARDIAC PACEMAKERS
The need for an increased HR.is not only to compensate for increased
metabolic demand,but it is also required in order to increase circulation to allow
for cooling of the body as suggested by abnormally elevated CVTsin persons
during exercise who have impaired cardiac output (Nielsen, 1966). Although the
need for an increased HRmaybe clear, the mechanismby which it changes is not.
The increase in HR,according to available evidence, is due to the activation of the
sympathetic nervous system by reflexes in the contracting muscle; and not due to
either a perceived drop in blood pressure due to vasodilation of the blood vessels
in the muscle or due to a drop in concentration .or partial pressure of 02 or
increase in concentration or partial pressure of CO2(Berne, 1981). In fact, the
baroreceptive reflex, which responds to changes in blood pressure, is reduced
during exercise (Berne, 1981). "
CVTalso increases during exercise. Due to ~e increase in metabolic rate,
additional ~heat is produced. This additional heat causes the CVTto rise. CVTs
can increase between 0.4"C and 1.4’C from an initial exercise-evoked dip at low
to moderate exercise levels’(20-100 W). Similar to HR, at a given workloadthe
inc.rease in CVTas a function of time is linear to’ exponential in shape
(Laczk0vics, 1984; Cook, 1984) until a point where heat production is balanced
by heat loss at which it plateaus. The rate of increase in CVTaswell as the point
at which it plateaus is set’according to the relative workloadof the individual~
The CVTrate of increase and its plateau at a sustained workload also correlate
with HR. Meantypical HRrate increases (d(bpm)ldt) and CVTrate rises (dT/dt)
for exercise levels described previously are approximately 10-15 (bpm)/min and
0.05:-0.11 C/rain, respecuvely (Sellers, 1985, ’Griffin, 1986). Studies companng
HRch~geto CVTchange during exercise at levels described above have shown
correlation coefficients between 0.95 and 0.9864 for both healthy and pacemaker
patients (Munteanu,1986; Alt et al., 1986).
Although the correlation between the CVTand HR is very good, it is
necessary to investigate the correlation further. Figure 15.3 shows a healthy
individual’s CVTand HRvariations during moderate exercise (Alt et al, 1993).
~ knvh
~ k~n~h
l S%
120 37.60
1011
37.10
36.60
Time
Fever ~
In.general, an increase of 10-15 bpm/*Cis associated with fever along with an
increase in metabolism(Alt et al., 1986). In addition to CVTincreases, often
times fevers have periods of chills. Studies involving such effects on HRand CVT
have not been reported.
Anxiety
EmOtionalanxiety, which is Sometimesexperienced before many Of the exercise
protocols, has :been shownto cause both initial HRincreases and CVTvariations
similar to the onset of exercise (dips followed by increases) (Fearnot, 1987;
Sellers, 1987; Feamotet al., 1989).
Ettvironment
A decrease in the ambient temperature, which initially decreases the CVT,will
trigger an effector response from the hypothalamuswhich will cause an increase
in thermal metabolism and/or vasoconstriction. If the effector is solely
340 DESIGN OF CARDIAC PACEMAKERS
Eating-and drinking
Studies "by Sellers et al. (1985, 1987) reported that eating causes transient
increases in CVTbetween 0.13-0.48"C. These studies reported that ~ drinking cold
liquids caused transient CVTdrops between 0.08-0.15°C over periods less than a
minute. Hot-liquids had no-effect on CVT. The effects on HRwere not reported.
Figure 15.4 summarizes the.effects of temperature change on heart rate.
Heart rate
Studies have shown that an increase in pacing rate does not cause the CVTto rise
(Jolgren, 1984; Laczkovics, 1984), Thus pacemakers using this technology have
negative feedback. Increased pacing rate, as mentioned earlier, will help regulate
the body temperature as a healthy person’s would.
RATE ADAPTATION BY TEMPERATURE 341
Stimulator
Endocardium
~at~piing
igure 15.5’ The;CVT ~ate-adaptive pacemakersystem. The microprocessorcontrols the
rate of the CVT.
In addition,the microprocessor
maycause~th~
powerto be ttirned off to
th~ evaluationcircuit wben~the~
temperature is not beingsampled.Also,in pacemaker
Systems
such
as~the IntermedicsCircadia® (see Figure15.6) the microprocessormayadjust the reference
temperatureof the amplifier/evaluation
circuit dueto component
tolerancesordrift.
Electrode
3 to 7 cml
Figure15.6 Pacing/sensingelectrode with integrated thermistor Adapt,#from:C00k
etal.
(1985).
Basic circuit
In order to determine whether or not the present pacing.r~te ’sho.~ld be changed,
the change in resistant6 of the thermistor must be~c0~tedt0 a voltage that
relates to temperature.-In order to eliminate unnecessary processor floating point
operations, it is useful to first lineafize the temperature/voltage relationship over
the entire range of the ADCbefore converting the analog representation to
digital. Figure 15.7 showssuch a circuit. Note that this Circuit is only shownfor
functional purposes, and that in practice it maybe done differently.
5/:/
1.2V~
R
Fo ADC
-1.2 V Rth
The inputs of the circuit, shownwith values of 1:2 V and -L2 V, are below
the battery’s voltage becauseit will gradually decrease ovtr ~time.
The thermistor, which is assumedto have anegative temperature coefficient
of -4%/*C, assumed to be linear over the temperature range 35-40°C, and
assumed to have a nominal resistance of 1.786R @25"C, in Figure 15.6 is the
RATE ADAPTATION B~ TEMPERATURE 343
resistor with the value Rth. Thus the resistance will change by -0.0714R/*C,
which corresponds to a 0.286R swing (between R and 0.714R) over the
temperature range of interest (36--40°C).
Equation (15.1) relates the output of the amplifier,
(15.1)
Thus the output voltage will vary linearly with CVTbetween 0 V (36°C) and
2.4 V (400C).
The ADCfollowing this Circuit is assumed to’be an 8-bit ADCwith a
reference voltage of 2.4 V. The resulting resolution is 0.02*C/bit, which is a
co,only reported value.
An improved approach
Because resistors have imprecise values and both the resistor’s and the
thermistor’s values maydrift over time, circuit compensation is desirable. The
circuit of Figure 15.8, which has been slightly modified from the corresponding
figure in Calfee et al. (1989), has.the ability to adjust for the aforementioned
tolerances: This circuit is used in the Intermedics Circadia pacemaker.
" I ladderl I J
~
~ Vl R! ~ V2 R~.. 2R1 (k~ /~th
/ Micropressor I l l Tomicro
rocessor
i at a ro s usI , ,,,
Fibre 15,8 A mo~ adv~ced tem~rat~e/voltag¢ line~izafion ~C u~d in ~e Inte~edics
Cima~a pacem~erw~ch ~lows for adjus~ent due to resistor tolerates ~d com~nent &i~.
The 8-bit R-~ ladder DACis us~ to dete~ne ~e cu~nt C~ while the 4-bit R-2R ladder
DACis ~ to adj~t for resistor tole~ ~d~. A~dfrom C~f~ et ~. (1989).
as 71.5 k~ (Calfee et al., 1989). Thesupply voltage, Vs, of the circuit is also the
reference voltage of the DACs.
For ease in analysis, the voltage outputs of the 8-bit and 4-bit DACshave
been labeled V1and V2, respectively.
Using superposition, Eq. (15.2) describes the voltage, Vn, of the
noninverting terminal of the comparator
v. = + v2 + 8Vs (15.2)
18
1 7
(15.3)
(15.4)
~ \ n=O
where
Vs is the reference voltage of the DAC
n is the bit placement
an is the value 0 or 1 of the individual bit of the 8-bit DAC
bn is the value 0 or 1 of the individual bit of the 4-bit DAC
Up .until now we have shown that the CVThas a strong correlation w~th heart
rate and that the additional hardware changes to nonrate-adaptive pacemakersare
minimal and easily implemented, Perhaps be most important--and developing--
aspect of the. CVTrate-adaptive ~pacemaker is the algorithm by which, the
pgcemaker changes its rate ba~d on CVT.
The best algorithm is. one which exactly, mimics the function of a normal
person’s heart. Any algoritllm relating CVTwith HRmust have programmable
parameters because of individual differences in CVT,HR, ,and rate-adaptive
needs. Some of the factors which need to be considered when developing an
algorithm include the magnitude of the temperature Change (both for the
e~ercise-evoked dip as well as for the increase in temperature during exercise),
file.rate of changein temperature (both positive and negative), the .rate of change
in HRfor a given magnitude or rate of temperature Changeduring both exercise
and circadian (slow) cycles, the length of time at an elevated pacing HR,and the
maximaland minimal pacing rates (commonin all types of pacemakers).
In addition, it is necessary to have an algorithm that will distinguish
exercise-evoked CVTchanges with those occurring due to other factors such as
circadian cycles or periods of fever. Also, during periods of exercise the
algorithm should provide a pacing rate adjustment that is as quick, accurate, and
smoothas possible.
This section describes four algorithms that have,been,used in CVTrate-
adaptive pacemakers. The first two algorithms give a general understanding of
how the aforementioned parameters are implemented in an algorithm. The last
two algorithms, which are more advanced, show the latest algorithms and how
they offer an improved pacing~response when compared to the two former
algorithms. Due to limited public iriformation, all aspects of the following
algorithms are true in their description insofar as information allows.
One of the first CVTrate-adaptive pacemakers was the Cook Model Kelvin 500
series (available in both unipolar and bipolar models). This type ~of pacemaker
bases itsrate,adaptation strategy on the following algorithm which~wasdeveloped
from tests on canines (Cooket al., 1985).
where
HRis the instantaneous HR(bpm)
A is the resting heart rate (bpm)
B is the slope of the HRversus temperature curve during exercise
T is the smoothedor f’dtered instantaneous CVT
TOis the resting CVT
C is the initial rise in HR(bpm) divided by the slope of the temperature-
versus-time curve during exercise
sign(dTIdt) is the sign (+) of the derivative of temperature with respect
time
Equation (15.7), however, is only a basis of the algorithm, not the actual
algorithm that is used. The algorithm present in the Kelvin 500 series pacemaker
not only modulates the heart rate according to the factors affecting Eq. (15.7),
but it also takes into account the initial dip in temperature as well as adjusts its
baseline HRand CVTthroughout the day (Sellerset al:, 1987; Fearnot and Evans,
1991; Bixler, 1994). Figure 15.9 lists a description of the symbolsused in Figure
I5.10, whichgives hysteresis curves (a and b) and a flowchart representation (c),
which describes the algorithm during each sample period, which happens once
every 10 s (Bixler, 1994). In addition to the instantaneous temperature, the rate
of temperature Changeversus tithe is needed for determining the required pacing
rate. This is calculated by taking the average of the six most recent CVTvalues
and subtracting it from the average ofthe six CVTvalues which preceded.
Therefore, a total of 12 CVTvalues (2 re.in) are neededto determine dTIdt. Both
Figures 15.9 and 15.10 were developed from available information (Cook et al.,
1984, 1985; Sellers et al., 1987; Volosin et al., 1989f Heggset al.,~ 1990; Fearnot
and Evans; 1991).
Using Figures 15~9 and 15.10, we can simplify Eq. (15.7) by the following:
where ....
HR(t) is the current
HR(t-1) is the previously determined
AHRis a positive~or negative transient rate increase
Although this equation appears to allow the HRto assume many different
values~ most of these values are only transient rates whichallow the HRto adjust
to one of the three distinct rates (Hr, Hi, He). The algorithm changes the pacing
rate based on both the difference in C, VTfrom the baseline temperature as wall as
the rate of change in the CVTversus time. The former of which decides which of
the two hysteresis curves (Figure 15(a) and (b)) the algorithm follows.
maximal and minimal decision point~ (Ln, Hx, Lx, Hn) of the two curves are set
in such a wayas to makeit more difficult for the HRto adjust to the exercising
rate, He, whenthe CVTis below Th and easier for HRto adjust to the resting
heart rate, Hr, whenthe CVTis below Th than to adjust to the intermediate rate
when the CVTis above Th.-These set points help reduce the possibility of an
unnecessary adjustment of the pacing rate from Hr to He as well as reduce the
required CVTversus time slope required to adjust the HRfrom Hi to He. This
reflects the natural HRand CVTrelationship at higher CVTs(see section 15.1.2).
RATE ADAPTATION BY TEMPERATURE 347
Figure 15.9 Parameters used in Figure 15.10, which describes the algorithm by which the
Kelvin500series pacemakeradjusts the stimulation rate. ¯ Note that all of the parameterslisted
abovewere.notspecifically namedas such but are providedhere for clarification of the algorithm.
*Valuenot given, but is probably less than Tr + 1 C. Values not g~venbut are probably a
function of the remaining programmableparameter called Kelvin. set~w~ieh ffetermines the
sensitivity of the algorithmto temperaturechanges(Adaptedfrom Cooket al., 1984~1985;B0alet
al., 1987;Sellers et al., 1987;Volosinet al., 1989;Heggset al., 1990;Feamotand Evans,1991).
dT/dt
(a) (b)
no
SetTr
yes Set HR= Hr Set HR= Hr
yes lies
I~et ~-IH = I
IDOSI(IVe
rate
no
Set &FIR=
negativerote
negative
rate Remove
oldest temperature
valuefrombuffer andshift valuesup negativerate
c)
3 rate adjustments (-AHR= 30 bpm), the HRadjusts to the Hi. Soon after, the
< Th and the slope of the dTIdt ramp decreases but is still less than Ln. The HR
will then begin to drop again according to the rate -AHR,eventually reaching Hr2
Nowassumethat the patient stops exercise instead of continuing (once again
at Hi on the dotted line of Figure 15.10(a)). The subject’s HRwould stay at
until ti has ’elapsed. After this time, the subject’s HRwould be decremented by
-AHRuntil it reached Hr. The purpose of such a limit is to ensure that the pacing
rate, HR, does not stay at an elevated rate for an extended period of time in case
the patient suddenly Stops exercising after inidaily Starting (which’ wouldcause
Sl0w return to baseline CVT)or in case thealgorithm mistakenly interprets
drop in CVT(due to circumstances other than exercise) as an exercise~evoked
dip.
Nowlet’s assume that rather than stop exercising, the patient continues to
exercise for’a period of 2 h. Usingthe left half of Figure 15:10(c), wesee ~that,
the HRstays at anyof the three HRlevels for more than 2 h, the Tr is ~ adjusted to
be..the second min~al temperature of 8 equally spaced samples over the 2-hour
period (T2m)..The algorithm correspondingly adjusts to Hr, the programmed,base
~acing late. PresumablyTh, Hi, He arealso adjusted. This of Coursg is ’aot the
~lesired physiological~e~ponse.If the patient cohtinues to extrci~e, h0wever, i~tis
likely that the cvT will rise in such a way to at least adjust for pacing/i/the
intermediate rate because of the resulting increase in CVTdue to the relationship
between HRand temperature regulation (see section 15.1.2).
Although this algorithm offers a great improvementto a need for increased heart
rate during exercise for patients who cannot benefit from the. traditional AV
synchronous pacemaker, its implementation is suboptimal because it.responds in a
stepwise fashion with only three distinct pacing rates. Also; the transitions
between the three pacing rates are abrupt, which does not correspond to the
naturally smoothtransition of the HRand also allows for high~levels.of~ unde~Sired
tachycardia in the event of false exercise detection. In addition, ,a!thougti: the
algorithm adjusts to circuit and circadian CVTfluctuations ~y adjt!sting the.
baseline rate) to allow for an improved response to exercise regardless of the
circadian cycle, this also allows for an unnatural respons~dudnglong periods of
sustained exercise. Asimilar undesirable response can occur during fever, which
is more.likely to last more than 2 h. Finally, this algorithm has a latency.~between
the algorithm’s response and the natural response as shownin studies by Sellers et
Another one of the first CVTrate-adaptive pacemakers was the Intermedics Nova
MR(both unipolar and bipolar models), which differs from the Kelvin 500 series
in that its pacing algorithm has a more dynamic HRresponse. Like the Kelvin
500 series,pacemake r , the Nova MRnot only modulates the heart rate according
to the derivative of the CVTversus time curve as well as the instantaneous CVT,
but it also takes into account the initial dip in temperature. It does this, however,
using an algorithm with a response that is more natural. Figure 15.11 gives the
parameters for Figure 15.12, which gives HRand CVTcurve relation (a) and (b)
and a flowchart representation (c) that describes the algorithm based upon which
curve the algorithm is currently working on. Figure 15.12(a) shows a series
curves (3 are shown)that are referred to as the:,exercisf curves and are used to
adjust the heart rat e as a linear function 0f CVTduring exergise. The. algorithm
uses two curves that have a linear relationship ~betweenC~Tand pacing rate, The
nrst ofthese curves,,~e basic curve orfever curve, has an ’adj/~stable slope of 6,
12, 18, or 24 beats/C increase. This curve follows teinpe~ture changes that
occur" more s, lowly such as those due to circadi~ fluctuations and ~fe~er. The
slope ~f thesecurves is programmableand is t,.yl~ic~lly aroui~dS0bpm/,Cinitially
followed by a decreasing slope (about 25 bpm/~C)at higher HRS(Alt ~ al., 19 86;
AI~, 1987; Alt et al., 1993).
. ~Pro~ammable
ba~ierate thresholdconstant ***-0.08*C/rain
Prol~ammable ~ time at HI
Maximaltime on an exercisecurve.... 30 rain’ ....
Figure15.11Parameters
usi~dinFigure15,12whichdescribes
thealgorithmbywhichthe
Intermedics
NovaMRpacemaker
adjust
thestimulation
rate.
Notethatall.of
theparameters
listed
abovewere
notspecifically
namedassuch,butareprovided
here
forclarification
ofthealgorithm.
¯ This**
vdue~
ismost
likely~
an....
average
ofa number
ofthe
mos,[
rec~nt
CVT
sawn..pl,e
s (ap~x~mat.ely"
8).Themethodbyw~chthisvalueisdetermined
isnotglven;’h0wever,
zt~smost hkely
taken
asthedifference
betweenthe
maximal
andminimalCVTfr6ma set~fthemostt’e~entiy
stored
samples.-***Value
notgiven;
estimated~from
available
information
(Adapte,
xlfrom Altetal.,
1986~
Alt~1987;
Aftetal.,1988;
Aftetal.,
1993).
,.sampii
gratgiven
as,
eve.:
ilis,
cos,or
scond
,Calt,
1988)
.b_.ut !~ reported to ~ every 4 or8 pacemaker.cycles (Siaeed, 1994,. The period at
hicn;’i, ~e. pacemak.e.radjusts the ~ac~ag’rate is identical, (Sn~<19,94),~Like; the
dK, ,~l, vih~i0~ns~dT’"~,thve N~)vaMR,~oa~hinalso. use;’k~yd..set; ’of CyT"~alues, fo¢
tete,rm.i iig th;,, CTrate.of 3hage, alt iSugh .the,atlmb r..of temperature
sam~"pies im;olved in th’si dete~nation is n0~ ind~ted~ in addition,the ~,+aMR
RATE ADAPTATION ~ BY TEMPERATURE 351
Increase HR
Transition
to HI
threshold
threshold
TransitiontoK1
from above. In addition, the patient will experience an initial brief decrease in the
pacing rate at the onset of exercise until the dip meets, the rate and magnitude
threshold criteria. Also during periods of sustained exercise longer than tt, the
pacing rate initially shifts from K1 to K2 when the natural response would
assume the higher rate. Also, transition between curves KI and K2 can often be
unnaturally abrupt.
"
110
~.e.
37.5
70
1. An algorithm that adjusts the pacing rate in a way whichbetter simulates the
rate of a normal, healthy heart under conditions such as rest, circadian
rhythm, exercise, and fever.
2. An algorithm that provides an improved detection of the beginning of
physical activity.
3. An algorithm that controls the slew rate of the calculated pacing rate to
avoid abrupt increases or decrease in HR.
4. An algorithm that provides a more optimal response after saturation of the
pacing rate whichcan occur during periods of strenuous activity.
5. An algorithm that, in addition to performing the above, will automatically
adjust the reference temperature used by the rate response algorithm.
The algorithm bases its pacing rate, which is calculated every 4 or 8 beats
(programmable), on Eq. (15.9) (Calfee et al., 1989; Sneed, 1994). The Circadia
pacemaker samples the CVTevery beat.
where
Reference Rate is the desired base pacing rate.
Natural Rate Response is the desired change in resting pacing rate with a
CVTchange due to natural causes such as fever and circadian cycles.
Dynamic Rate Response is desired change in the pacing rate due to CVT
changes associated with exercise.
Step Rate Response is desired incremental change in pacing rate associated
with the exercise-evoked dip in CVT.
The Reference Rate is the desired resting pacing rate and is initially
programmed by a physician (typically around 70 bpm). The ~remaining
parameters will be explained using applicable equationS’, ~t~bies, and pseudocode.
Figure 15.14 displays a general flowchart of this algorithm while Figure 15.15
illustrates somecharacteristic curves of the af~ementioned parameters (Calfee et
al., 1989).
where
KNATPis a programmablepositive natural rate coefficient with a typical
value of 12 bpm/*C(see Figure 15.15(b)).
KNATNis a programmablenegative natural rate coefficient with a typical
value of 6,bpnd*C. ~’
TAVGis the current average CVT(°C) calculated from’ the 4 or
(programmable) most recent temperature samples.
REFTMPis the weighted average reference t,e~perature which initially is
programmedby the physician (typically 37 C)~ but after implantation
modulated by Eqs. (15.10) -(15.12),
Find current
Start temperature
limit test
I Perform rate I
End
AN = K*.SN-1. (15.10)
DN = TA VG ÷ AN (15.11)
SN = SN-1 + DN (15.12)
where
AN is the current weighted average reference temperature over a
predetermined time period.
K is a constant coefficient, (1/2)n wheren is an integer between8 and 24.
DNis the difference between the current and weighted average temperatures.
SN is the current=sum.of the previous :. average~ weighted sum and the
difference between the current average, temperature and the weighted
average temperature; initially programmed with a value of
2n(REFTMP).
SN-1is the previous average weighted sum.
Tern)erature Naturalrate
72
response(bpm)
KNATP
-~FTMP
Time TAVG(’C)
(a)
Dynamicrate Steprate
response(bpm) response(bpm)
STEP_SLOPE
STEP_DURATION
STEP_
RESPONSE’
RADJ1" ~
TAW ¯ i ~ ~ i g g ~’ g Differential
cycle
(c)
Temperature
Calculatedpacing
rate w/osaturation
control&noratelimit
Time
(e)
l~i~ure 15.17PSeudocode
that determinesthe.Circadia Dynamic
Rate Response.Adaptedfrom
Calfeeet al. (1989).
~here
RTAVG isthe rate ofchangeof the average CVT(’C).
RBCRITis the minimal programmablerate of change threshold value, which
..indicates whether or not the subject is exercising (typically
0.02"C/differential~ cycle), A differential .cycle is given as 8 pacing
cycles.
TAVGis thecurrent averageCVT(’C).
DYNRFris a valuewhichtracksthe TAVGvalueuntilthe momentRTAVG
> RBCRIT at whichit remains constant (se~-Tl on Figure 15.15(a)).
KP0,KP1,andKP2arethefirst, second, thirdexercise coefficients with
typical values of90,66,and48 bpm/°C, respectively.
RADJ1,andRADJ2arethefirstandsecondconstant rateadjustments with
typical values of27 and46.8bpm,respectively
BPI,andBP2arethefirstandsecond temperature breakpoints withvalues
of0.3and0.6"C, respectively.
It should be apparent that the values KP0, KPI, KP2, RADJI, RADJ2,BP1,
and BP2are interrelated such that they will provide a smooth pacing rate..when
the two breakpoints are crossed (see Figure 15.15(c))~ These parameters
rprOgrammableinsuch a way as toretain the smooth transition: In’addition, the
subsequent decreasing slopes of KP0, KP1, and KP2closely follow the natural,
exponential HRas the HR(and CVT)increase.
~ .Onc~: the TAVGrate of ¢h~nge fal!s~ be!ow the value RBCRIT(fo!lowing
cessation bf exelcise; see ~ Orl Figure 15~i5(a)), DYNRFTdoes not ~ediately
tr~ck TANG;instead,~ D~ i~s incremented according to a programmable
~ func~0naccording ~t0 parameters~TSTP ~d.~ DELAYin o~der to
provide a smoothpacingrate decrease such as a normal, liealthy heart Would.
Rate
358 DESIGN OF CARDIAC PACEMAKERS
T_CTR is a counter that ensures that the patient has been at rest for a certain
period of time before the Step Rate Responsecan be added to the overall
pacing rate (initially set by the algorithm to MAX_COUNT, which is
typically about 22 diferenfial cycles).
TL is the average CVTfrom the previous differential cycle.
DIPSLOPEis the programmable CVT dip slope criterion (typically
0.006*C/differential cycle).
DIPSIZEis the dip magnitude criterion.
RATE is the present calculated pacing rate.
STPCRTR is a programmable criterion associated with the present pacing
rate, which assures that the Step Rate Response is not added to the
pacing rate when the CVTdip is not due to the onset of exercise
(typically 80 bpm).
STEP_RESPONSE the step rate adjustment contributed to the overall pacing
rate (equal to STEP_SIZE which is typically 15 bpm).
STEP2DURATION is the criterion that determines how lotig the StepRate
Response will contribute to the overall pacing rate (equal to
STEP_DURATION_MAX, which is typically 4 differential’ cycles).
TPEAKis a peak TAVG which is calculated by the following:
if(TAVG>TNL) then TPEAK = TAVG
where TNLis the previous TAVG value.
Figure 15.18 Pseudocode that determines the Circadia~tep Rate Response through the use of 5
criteria. Adaptedfrom Calfee et al. (1989).
Also similar to the previous algorithms is the limited amountof time that the
step: rate response is added to the overall pacing rate. The point-at whichthe Step
Rate Response begins to decay is given by the programmable parameter
STEP_DURATION, which has a typical value of 4 differential cycles after the
Step Rate Response, STEP_RESPONSE, has been added to the overall pacing rate
(see. Figure 15.15(d’)). The Step Rate Response decays during
STEP_DURATION period by the value STEP_SLOPE,Which has a typical value
of 4 bpm, until the Step Rate Responseno longer contributes to the pacing rate. -
510 Series algorithm uses the absolute magnitudeand derivative of the exercise-
evoked CVTdip as an indication of an amount the pacing rate should be
increased, ratherthan their values relative to thresholds. This results in an earlier
increase in pacingrate following the onset of exercise.
where
TRis the target pacingrate
LRis the base heart rate as well as the lower pacing limit (typically 70 bpm)
DIUis the desired addition to pacing rate with a CVTchangedue to natural
causes such as fever andcircadian cycles
DT is the desired addition to pacing rate due to the rate of change of the
CVT
STXis the desired addition to the pacing rate due to the magnitudeof the
exercise-evoked temperaturedip
TXBis the desired addition, to the pacing rate’ due to the increase in CVT
above a local minimum during exercise
These values are calculated according to the pseudocodeof Figure 15.19 and
the programmable coefficients for tau which are listed .in Figure 15.20. These
coefficients were developed by studies relating CVTand HRin both normaland
pacemakerpatients having normal sinus node activity. In addition, the above
pacing parameters are determined primarily by comparing a plurality of
temperaturedeviations from three movingbaseline temperature values that are
shownin Figure 15.21.
Figure15.19Pseudeeode
flaat determines
the pacingrote of the CookSensorModelKelvin510
Series pacemaker.
Adapted
fromFeamotandEvans( 1991).
~TE ADAPTATION IBY ~MPE~TURE 361
¯ The Kelvin 510 Series pacemaker samples the CVTonce every 10 s and
adjusts:fiae pacing rate every 5 s whenthe pacing rate is increasing and every 10 s
whenthe pacing rate is d~creasing. In addition;~ pacing is adjusted in increments
of 1 bpm for pacing rates between 50 and 130 and in increments of 2 bpm for
pacing rates between 130 and 160. The range of possible pacing rates is between
50 and 160 bpm (Cook Pacemaker Corporation, 1992).
It should be evident from Figure 15.19 that the target rate, TR, is adjusted in
steps. ~ first possible addition to the target rate is that due to the magnitudeof
a exereiffe~evoked CVT dip (STX).
greater than the previous dally minimal temperature, DIU will be added to the
target pacing rate due to the diurnal variation. If the value calculated for DIUis
negative, then DIUis 0 (see Figure 15.19)~
Buier
IDV- changed
2rid min. D value I every 24 h
CVT- sampled
every 10 s
- sampled every
1 min I
D/6
I - sampled
every 5 min. I
Diurnal
In addition to the features already described and to the features that are common
to all pacemakers, there are several additional features that are present in the
Cook Kelvin 510 Series pacem~er; arid presumably similar features in all CVT
rate-adaptive pacemakers. These include the following:
1. A short term :(up to 4Omin)snap shot histogram, which is initiated using the
programmer, that is most helpful for set activities, such as ,walking or
bicycle ergometry, during office .visits. Mile in this mode, every time the
pacemaker measures temperature, and updates the pacing rate, one of six
counters is incremented, The counters represent the .temperature ranges, of
~ 50-64, 65-79, 80-94, 95-109, 110-124, and 125-160bpm with a limit of
255 for each counter. ~
2.,: A long term histogram (up. tO 5 years), which is also initiated~ by the
programmer,that will ~tell the numberof times the pacing rate was above or
below a programmabledistribution rate.
3. A synchronized telemetry option is also available, which will cause the
pacemakerto telemetgr the.pacing rate and CVTdata.every 10 s. TM’sallows
the l~h~si~ian to test the effectiveness of ~hangingthe vai’ious programmable
values during an office visit.
Also, some .pacemakers, such as the Intermedics Nova and Circadia, have not
progressed, beyond clinical trials, not due to ineffectiveness, but due to business
decisions. These .business decisions were primarily , based on the fact that
physicians do not like to be forced to.~use.a certain type .of pacing lead (the
number of types of pacing leads with integrated~thermistors is extremely limited)
as well as the fact that rate-adaptive pacemakers using activity sensors, which do
not require changes to the conventional pacing leads, have shown acceptable rate-
adaptation during most types of. exercise. (Sneed, t99~). These .types of rate-
adaptive pacemakers, however, will not provide increased pacing rates due to
fever, anxiety, or inherent circadian type cycles.
15.8 REFERENCES
Fearnot, N. E., and Evans, M. L. 1988. H~art~ rate collation, response time and effect of
previous exercise using an advancedpacing rate algorithm for temperature-based rate
modulation. PACE,11: 1846-1852.
Fearnot, N. E., and Evans, M. L. 1991. Temperature-basedrate-modulated cardiac therapy
apparatus and method.USpatent 5,005,574.
Fearnot, N. E., and Smith, H. J. 1987. Six possible factors affecting intercardiac temperature
duringrest and exercisetesting (abstract). Circulation,76 (suppl. IV): 1447.
Fearnot, N. E., Osamu,K., Fujita, T:,’Okamura,H.; and Caldefini~ ~VI~1989. Case studies on the
effect of exercise and hot water submersionon intercardiac temperature’andthe performance
of a pacemakerwhichvaries pacingrate basedontemperature. Japanese Heart Journal, 30:
35~-363.
Furman,S..1990. Rate-modulatedpacing. Circulation, 82: 1081-1094."
Gillette, P. 1984. Critical analysis 0f.~en,so~ for physio!ogi~al responsive pacing. PACE,7:
1263-1265.
Griffin, J. C., Alt E.~ Nielsen, A, and Ca!fee, R. 1986, Venousbloodtemperatureto pacing rate:
importanceof the algorithm(abstract). C//~ PrOg.Electr~physiol.Pacing,4 (suppl):
Griffin, J. C:, Jutzy, J:’P., and Knutti, J.W. i983. Central body’temperamrea~ a guide to
optimal heartrate..PACE~6: 498-=501.
Hauser, R. G. 1984. Techniques,for imp[ovingcardiac performance’with.implantable devices.
PACE,7: 1234-1239.
Heggs, K~ S., Johnson, W. L,, and Stevens, D. A. 1990. Temperature-controlled cardiac
" pacemakerresponsive to bodymotion. USpatent 4,905;697:
Houdas,Y;, and Ring, E. F. J. 1982.~~ Human body temperature:Its tn~¢asurementand regulation.
¯ ~ NewY~k: Plenum PreSs.
Jolgren, D. !984. A rate-responsivepacemakercontro~lledby~right ventticular bloodtemperature.
PACE,7: 794-801. ¯ °
Laczkovics, A. 1984. The central venous blood temperature as a guide for rate control in
’ " pacemaker therapy: PACE;7: 822-830: ¯ ~’~ ". .... , ¯~ ’
Lekholm,A. 1993. Apparatus for in v~vo in~ardi ~ of a measuredsignal corresponding to the
physical activity of a subject and a he~ pacemakerhaving a stimulation rate controlled
thereby. USpatent 5,218,961. ’ ’ ’¯ ¯.
Morgane~.P.J., and Panksepp, J~ (eds.)’ 1980. BehaPioralStudies~ the HypothalamusVolurne 3
Part A. NewYork: Marcel Dekker.
M.~unteanu,~ J.;:.Alt, E.~,"Hirgstette~C.,~ ~andHeinz~,M: 1986. Central’venousbloodtemperature
’" represeriting metabolic rattatid~individuals exereise capab~ility (abstract). Clin. Prog.
Electrophysiol.Pacing,4 (suppl): 27.
Nielsen, B. 1966. Regulationof bodytemperatureand heat dissipation at different levels of energy
and heat production in man.Acta. Physiol. Scand, 68: 215.
Rhoades, R., and Pflanzer, R. 1992. HumanPhysiology. NewYork: Saunders.
Saltin, B., and Hermansen,L. 1966. Esophageal,rectal and muscletemperatureduring exercise.
J. Appl. Physiol., 21: 1757-1762.
Schaldach,M.1992. Electrotherapyof the heart. Berlin: Springer-Verlag.
Sellers, T. D., Fearnot, N. E., Johnson,R. E., Shirley, D. A., DiLorenzo,D. M., and Knight, J.
A. 1985. Right ventdcular blood temperature profiles for physiologic pacing (abstract).
Circulation, 72 (suppl. lll): 1730.
Sellers, T. D., Fearnot, N. E., Smith, H. J., DiLorenzo,D. M., Knight, J. A., and Schmaltz,M.
J. 1987. Right ventricular blood temperatureprofiles for rate responsivepacing. PACE,10:
467-479.
Sneed, R. 1994(Personal conversation with Intermedics engineer). March24, 1994.
Sugiura, T., Kimura, M., Mizushina, S., Yoshimura,K., Harada, Y. 1988. Cardiac pacemaker
regulated by respiratory rate and blood temperature. PACE,11: 1077-1084.
Sugiura, T., Mizushina,S., Kimura,M., Fulmi, Y., Harada, Y. 1991. A fuzzy approach to the
rate control in an artificial cardiac pacemakerregulatedby respiratory rate andtemperature:a
preliminary report. J. Med.Eng. Technol., 15: 107-110.
Volosin, K. J., O’Conner, W: H., Fabiszewski, R., and Waxman,H. L. 1989a. Pacemaker-
mediated tachycardia from a single chambertemperature sensitive pacemaker.1989. PACE,
I2: 1596-1599.
Volosin, K. J., Rudderow,1L, and Waxman,H. L. 1989b. VOOR ~ Nondemandrate modulated
pacing necessitated by myopotentialinhibition. PACE,I2: 421-424.
Zegelman,M., Winter, U. J., Alt, E., Treese, N., Kreuzer, ~., Henry,L., Mugica,J., Schroeder,
E., Klein, H., and Vulker, R. t990 Effect of different body-exercise modeson the rate
response of the temperature-controlled pacemakerNovaMR.Thorac. Cardiovasc. Surg.,
38: 181-185.
368 DESIGN OF CARDIAC PACEMAKERS
15,1 Explain whyCVTrate-adaptive pacemakerssample the blood temperature from the right
ventricle.
15.2 Explain howthe body tempera~ regulated.. ¯
15.3 Sketchthe CVT and HRas a person is at rest, then begins to exercise, and finally ceases
exexcise..Statethe r~asonsfor tl~ shapeof ~ ctu, ve.
15.4 Explain whyCVTrate-adaptivdpace~rs have ndgative fcedb_ack.
15.5 Drawa basic block diagramof the CVTrate-adaptive pacemakersystemand explain the
purposeof each of the-blocks.
15,6 ~)escribethe impo~~rt~. c~. of isolating the :the ~rn~tor frompossiblefluid intrusion.Explain
the resulting CVTrate-fidaptive pacemaker responseinthe event of fluid intrusion if it were
not detected.
15,7 Explain why~thecircuitof Figu~15.8 is moreoptimal tha~ that of Flgm¢15..
15,8 DesCribehi~wtbe prc~nt t~n~peratm¢is determinedusi~gthe Circuifof Figure 15.8.
15.9 Describe which factors need to’be considered whendeveloping an alg0rithm for a CVT
rate-adaptive pacemaker.,
15.10 Describe howthe Kelvin 500 series pacemakeralgori~ Changesthe pacing ~ate based on
the CVT.
15,11 Explain the advantages and disadvantages, of the Cook Kelvin 500 series pacemaker
algorithm.
15.i2 Describe howthe IntermedicsNova MRpacemakeralgorithrn~changes the,pacing rate
based on the CVT.
15.13 Explain the advantages anddi~advantages of the~Intermedies Nova MRpacemaker
algorithm. ¯ ~,
15,14 St~td the equation whichtheIntermedicsCircadia algorithmuses to adjust the pacing rate
and describe the proposeof each of the p~ameters,
15,15 Explain howthe Int¢~e~cs Circadia de~ermines whether or not an exercise-invoked
ten~peratur~
dip has ~d.
15,16 Explain~ hoWthe Sensor ModelKelvin 510 pacemakeralgorithm adjusts the pacing rate
following exercise.
15.17, Explain the importanceof the followrupfunctions of CVTrate’adaptive pacemakers..
15.18 Explainhowan additional sensor world increase the ¢ffectivenes~of the CVTrate-adaptive
pacemaker.
Rate Adaptation by Electrogramand Intracardiac
Impedance
Clark Hochgraf
The sympathetic tone responds to manyfactors that influence the heart rate
such as workload and emotional stress. In comparison, activity-based rate
adaptation schemes can only estimate the workload and cannot adapt for
emotional demands. The ability to respond to emotional and mental stress is a
feature of the methods based on cardiac signals. Impedance-based measurement
of stroke volumeand ejection~fraction mayliave additional utility~in the detection
of tachycardia. The sensors for EGMand impedance signals are~ simple and
robust.
Of the manysignals that maybe used for rate adaptation, intracardiac signals
have certain advantages. The signals are often obtainable using a standard or
slightly modified pacing lead. As such, they do not need additional sensor leads or
complex sensors such as required for detecting blood pressure or blood oxygen
levels. Rate algorithms based on the EGMrequire no additional sensor energy.
Rate algorithms based on impedance do require additional sensor energy. Both
the impedance and the EGMsignals contain information about the state of the
heart’s natural control system--the autonomic nervous system.
Derived parameters such as the Q-T interval,, stroke volume, ejection
fraction andothers give indirect information about the sympathetic .tone of the
heart. In a normal heart, an increase inactivity in the sympathetic,nervoussystem
would cause an increase in the heart rate. In the. damaged, chronotropically
insufficient heart, the normal mechanismsfor.changing the heart rate°are not
fully functional. However,the effect of sympathetic activity can still be observed
in the heart’s operation through other indications such as contractility.
The impedance-sensing scheme can use unipolar, bipolar or multipolar
electrode, configurations. Processing of the EGM and impedancesignals is easily
performed. One disadvantage of using cardiac signals such as-the Q,T interval is
that there is a large patient-to-patient~ variability, iwthe ,relationship of the ideal
rate to the detected signal. Thus to achieve a good, rate response, the pacemaker
must be trained during, an exercise regime, wherein the heart is taken ~throughits
paces, so to speak. For a patient with poor,,cardiovascular health, it may-berisky
or simply not feasible to perform the exercise training required.
To understand the relationship between sympathetic tone and the normat sinus
rate, a simplified control system model of the heart is used. Schaldach (1992)
describes a control system mode!:,~of the. cardiovascular system, with particular
focus .on the, heart~s regulatory:role. Figure 16.1 shows the-influence of the
sympathetic and parasympathetic tone on ,theEheart rate, Par~ympathetic nerve
activity decreases the heart rate via the SAnode .andAVnode.Sympathetic nerve
activity increases the hea~ rate via the SAandAVnodes as well as affecting the
heart muscle itself. Sympathetic nerve stimulation affects the heart muscle by
RATE ~ADAPTATION BY EGM AND IMPEDANCE 371
causing an increase in the .contractility of the heart, resulting in an increase ,in the
stroke volume. Stroke volumeis the, volume of blood,ejected from the heart in
each beat.
Highereffects
(stress,fight or flight, etc.)
I Medullan]circulatory
centers
Di =
variables:
(workload,etc.)
There is clinical evidence that rate response to the sympathetic tone produces
beneficial regulation of the MABP. Grubbet al. (1993) reported on the successful
application of a rate-responsive pacemaker in regulating the’MABPin a patient.
with severe refractory orthostatic hypotension. The pacing rate was modulated in
response to the changes in sympathetic tone resulting in improved regulation of
MABP.
The patient’s condition was such that standing would cause him to faint due
to a marked decrease in blood pressure. Drug therapy was ineffective in this
patient. Continuous tachycardia pacing would have prevented hypotension except
that upon lying down, the patient would suffer from hypertension. Therefore, a
fixed-rate pacemaker that boosted his standing blood pressure, would
overcompensate whenhe was in a supine position.
A pacemakerwith rate adaptation based on the right ventricular pre-ejection
interval (Precept DRModel 1200 from Cardiac Pacemakers Inc.) was employed
to help moderatehis blood pressure. Whenthe patient stood up, the pacemaker
detected a sudden decrease in the pre-ejection period and correspondingly
increased the heart rate. The MABP was thus controlled quite effectively,
allowing the patient to resumea normal life.
such stress conditions. The ideal would be to have a signal that has a constant
simple relation to the normalheart rate in all cases.
For someof the signals, the relationship betweenthe signal and the sinus rate
is ambiguous ’or conflicting during some of the stressor conditions. The
processing algorithm inside the artificial pacemakermust take this into account.
Positive feedback must be avoided, otherwise the pacing rate will start to respond
to itself and~notto the physiological demand.
One of the more severe examples of a methodprone to this positive feedback
is the use of the Q-Tinterval .signal. The Q-Tinterval shortens during exercise,
indicating the need for the heart rate to increase. However,Q-T interval also
shortens if the resting heart rate is artificially, increased through paced
stimulation. This positive feedback effect can cause pacemaker-mediated
tachycardia~ Careful adaptation of the rate can compensate for this effect and
pernut succdssful use ofthe Q,Tinterval in rate adaptation
The standard pace sensing leads are used to detect the Q-T interval and
sVentricular
ensedfrom thedepolarization
pacing !cad aregradient. Changes inthethepre-ejectlon
used to determine intraeardiacimpedance as
period, stroke
volumeand ejection fraction.
Tti~ intracardiac .electrogram EGM can be used for altering the paced heart rate
in response to sympathetic activity. The EGMis usually measured from the same
lead (bii3olar or unipolar) as used for pacing which makes these signals
c0i~venient to obtain. Measurement of the Q-T interval, and ventricular
depolarization gradient can be used. for detecting changes in metabolic demands.’
The use of these parameters for rate response is :described in detail in the
following sections.
I!6,3.1,Q-T interval
Figure 16.2 ~defines the Q-T interval, which ~reflects the timing ~of ventricular
activity. The. onset of ventricular activity can be markedby either a~ natural QRS
complex or a pa~ed QRScomplex, thus the Q--Tinterval is sometimes referred to
asthe Stimulus to T interval. The Q-T interval is strongly correlated with the
ievel~ofexereise. Forthese reasons, it has been used in several pacemakermodels
as a signal for rate adaptation. The most challenging aspect of using the Q--T
interval is not its detection but finding an algorithm that safely and correctly
converts the timing information into a suitable heart rate. Problems with the use
of this~Signal for rate response have largely been caused-by ~the algorithm used.
Early algorithms did not account for the coupling of changes in the-paced heart
rate back to changes in the Q-T interval.
PeakNegativeSlope
~ ""
%
of the T wave
Q-T interval - I
¯
!~p~oved algorithms
~e:motivationfo~ improvements to the linear~ s!0pe.alg0rithm e~e from a
numberof directions and reveal the influential factorsqn the refinement process.
Shortcomings of the algorithm were seen in performance tests of the pacemaker
Where the prOportionality of response was n0t~unifo~m withworkload. The
requirement for individual training to calculate a patient-specific slope was ’time
consumingand involved physically, stressing the patient, Studies revealed that the
relationship of Q:-T interval to h~art rate during exercise is better represented
by a negative exponential rather than a linear function as was used in the original
version of the pacemaker. In addition, it was obserVed that whenthe heart rate
reached its upper limit, the rate tended to oscillate around the limit. Whenthe
slope was .too high; ~ the risk Of inducing tachycardiaby the positive feedback of
heart rate t0’Q-T shortening was also a concern.
The essential problem with the Q-T-interval-based pacemaker is that the
paced h~art’rate has an effect on the Q’Tinterval. This positive’ feedback effect
from the’heart rate to the Q-Tinterval needs to be accounted for.
,In light of this additional information, an improved algorithm was
developed~ Heijer et ali-(1989) presented the new algorithm:anti its resulting
performance. The improved algorithm encompassed severaladditional featu~res~
First, the pacemakercould nowperform self training for the individual patient as
well as adapt its slope to slow changes over the course of days. Second, the Q-T
to heart rate relationship was no longer linear but rather reflected the actual
376 DESIGN OF CARDIAC PACEMAKERS
nonlinear relation that had been observed in other studies. ~Thechange in slope
with heart rate allowed the pacemakerto respond more rapidly to exercise.
To test the algorithm’s performance before actually building it into a new
model of pacemaker, the real-time bi-directional telemetry capabilities of
previously implanted pacemakerunits, was utilized. The new algorithm was tested
in 37 existing pacemakersby using an external computerto continually.update the
programmedslope parameter. The external computer received, in real time, the
heart rate and internally measured Q-T interval and then told the pacemaker
what slope to use. The new algorithm became known as the dynamic slope
algorithm.
where (Q-T)rest is the measured Q-T interval during rest. If the coupling
between the resting Q-T interval to heart rate is included, the formula is
modifiedso that the slope is a function of the rate. Theunits o~ thesI0pe are beats
per minute per millisecond of Q’T shortening. In Eq. (I6.2), the slope is now
partially a ~unetion 0f the’heart rate throu~i the ~oupling factor~.M, which
represents the-~nonexercise-related change in the Q-T interval with a change in
heart rate.
The adaptation term is the second term. Since the value of Q-Trest minus Q-
Tsensed is always positive, the numerator term is increased during exercise, but
the denominatoris also increased due to the resting Q,T to heart rate relation as
expressed in the factor M. As the Q.T interval gets shortgr, the adaptation, slope
is reduced by a factor of
1
(16.4)
1 + M x ((Q-T)rest ~.(Q-T)sensed)
RATE ADAPTATION BY EGM AND IMPEDANCE 377
The dynamic nature of the slope is expressed in this term. The coupling
factor Mis readily derived from observing the Q±Tshortening when the paced
heart rate is temporarily increased in a resting patient. The Q-Tshortening is due
entirely to the artificially induced changein the paced heart rate and not due to a
change in sympathetic tone. Thus the measured parameter M represents the
coupling between the paced heart rate and Q-T shortening. By knowing M, we
can minimize its effect in determining the physiologically appropriate heart rate.
Figure 16.3 shows the value of the rate,adaptation slope, and it showsthat if Mis
not zero, then the rate-adaptation slope should decrease as the heart rate
increases.
2.5
1.5
1
M=O
0.~
0’
60 80 100 120 140
Figure 16.3 Exampleof change in the rate-adaptation slope when the effec t of the heart rate on
the Q-Tinterval is included. Anincrease in tilt of the line correspondsto a larger effect of the heart
rate on Q--T shortening. The coupling factor Mis measuredwith the patient at rest.
Figure I6.4 shows~t’:ih’ large value of Mcauses the paced Q-Tto heart rate
characteristic to havea significant curvature to it.
At low .heart rates, the slope is l~ge, so that small changes in the Q-T
interval cause a rapid response in the paced heart rate. However,at high heart
rates, the slopeis small so that the sensitivity to Q-Tshortening is reduced.
The higher slope atonset of Q-T shortening reduces the delay in responding
to activity. Withthe dynamicslope algorithm,-the slope at near the upper rate
l~mit rapidly, .~tppr0aclies zero. By having the slope go to zero, the problemwith
upper rate oscillations is avoided.
Clinical Studies
Studies of the linear and,,dynamic slope algorithm have. shown that the
proportionality of the response ~to workload is quite good. Figure 16.5 shows the
dynamicslope algorithm-to have a quicker, onset and better proportionality or
response than the linear slope. Dynamicslopehas also reduced the:tendency for
the pacemaker to oscillate in frequency near the upper limit and assists in
avoiding positive feedback leading totachycardia.
378 DESIGN OF CARDIAC PACEMAKERS
i40
120
100
"~ 8O
6O
330 340 350 360 ~ 3?0 380 390 400
120
115
¯ o
o
o
7O
Tlme[secl
o U,~. ~.,~o~m ~
~ 16~ Pacing ram d~ng exerci~ for ~e L~e~ ~gofi~m and ~prov~ ~n~c slo~
flgori~m. Notice howtbe improv~flgofi~ has a quicker~ ~s~nse to~ exewise ~d ~tter
pro~onflity to exewi~. ~e l~e~ flgofi~ can~s a rapid inc~e in rate ~. ~e pac~g~te
ge~ ~gber ~d ~e~T~fl ge~ sho~er.FmmHeijer, P. D., Nagelkerke, D., Pe~ns, E. J.,
Horst~, E., V~W~em,R. J., Neiderlag, W., Jord~ns, L., Wilde~P. D., H~el~rS, W.
B., ~d Lie, H. 1989. ~proved ~ ~s~nsive flgofi~ in ~T ~ven pace~e~vflu~on
of ~ififl respon~to exerci~. PACE.,12:805~11.
algorithm helps to minimize this delay, bm the limitation of the method seems
tied to the fact that. changes in the Q-T waveform seem to lag changes in
noradrenaline levels by about 1 rain. The maximalheart rate was seen to occur
29.2 s to 69.5 s ~after the cessation of exercise, .depending on the programmed
rate
adaptation slope.
Self-training of slope
The slope parameters of the algorithm can be programmedby a physician during
exercise training or~they can.be programmedautomatically by the pat.croaker
through self-training-algorithms. Twoself-training modesof the pacemaker are
the Fast~ Learning ~algodthm and the Automatic Slope Adaptation algorithm. In
the fast learning mode,the lower heart rate slope is set by having the patient sitat
rest while :the pacemaker setsits rate to 70 bpm then 80 bpm then 70 bpm,
holding each for 1 minute, while recording the Q-T interval~ The coupling
parameter Mfor low heart rates is then directly calculated. The patient is then
exercised at high workload until the upper rate limit is reached. If the Q-T
interval continue~tO .get,shorter .once the rate limit is reached, then it is known
that the slope (for high heart rates) is too high. The’ Slop~ fo~high heart rates
then reduced.
The Automatic Slope Adaptation method sets the lower rate slope once every
night (asdetermined by its 24-h Clock). During sleep, the lower et~d slope
calculated in much the same way as for the Fast Learning algorithm. The slope
for upper heart rate is adjusted if theupper rate limit is reached, and the Q-T
interval ~ontlnues to decrease even further. This indicates that the limit w~smet
prematurely and that the upper heart rate slope Should be reduced by one
increment. If, on the other hand, the upper rate limit is not reached once in a
period of 8 days, the~ sl~3pe for the.upptr heart rate is increased by one increment
ti~all0w the full range of heart rates to be used.
Zhe ventrieular depolarization gradient (VDG)is the area ’under the QRS
complex.. It ~has been observed that. in normal heart operation, ~ the ventricul~
depolarization gradient stays fairly constant: During, exercise, the ventricular
depolarization gradient decreases. Whenthe pacing rate is ~increased, the VDG
38O DESIGN OF CARDIAC PACEMAKERS
The electrode design issues ~ complex and there ,have~ been numerous
arrangements proposal for achi~ving~ good leveI of sensing. Unipola~,i~ip01ar,
trtpol~r, quadr~polar~ and multiple ~onopolar leads have been mentioned in the
~t~frits.
literature and it ~ee~s that With ~ppropria~e SignalP~essing, ~chhas its
The standard leadg are unipolar or bipolar. If the pacemaker is used as a
replacement then it is undesirable to removean already implanted lead. The feel
of the lead also affects the probability that the lead will be used. Somephysicians
maynot want to use a lead that has morewires init becaus~itis less’flexible than
they are accustomedto.
-~ As an example of two arrangements, Figure 16.6 shows two conftgurations.
¯ he first~arrangemerntUseS-tworing electrodes for current injection ~and a third
electrode at the tip of the lead for pacing. The second lead uses two electrodes in
a split ring arrangementfor current injection and a third tip electrode for pacing.
RATE ~ADAPTATION BY EGM AND IMPEDANCE 381
The separate use of current injection electrodes avoids the effect of electrode
polarization at the pacing electrode.
¸26
Whenthe current flow paths are considered, the split ring arrangement turns
out to have a lower sensitivity. Its sensitivity to impedancechanges drops off as
the ..cosine of the angle from the ,split between ~tlae electrodes. The electrode
position thenplays a large role ~in the quality of the impedancesignal, If,the split
line is , close to the heart,wall, then motionartifacts may.playan excessively large
role ~in the detected sigual.
, ;A monoP0!ararrangement uses the pacemaker case as a~remrrt point,,Geddes
et a~.~,:O991) found ~that bipolar !eads~d not produce a go~ current ~stribution.
~,. ~ ~urrent tended to,hug, to the ca~eter and not intercept the heart wall,
Increasing the spacing between the rings didnot improve, the-signal greatly. With
the bipolar!arrangement, if one of the,e!ec~odes .came, in contact with the heart
wall, ~e relationship of impedance change tostroke volume was altered. Geddes
et aLproposed ~that the monopolar arrangement ~produeed a radial flow. of
ct!~eut~ with a correspondingly large path through the blood ,of the ,heart and a
goodsignal. Figure 16.7 showsthe,situation.
In. general it would seem desirable to have the electrode far from the heart
wall~ ~sqm~what centered in the ventricIe chamber, however, due .to the irregular
~ati~ Of a damagedheart, the ideal electrode position-varies from-patie~it to
p~en~,¯ Geddes proposed to have multiple monopolar electrodes on the catheter
and use~.the remote programmer~andan internal.: multiple~Xer~.to ~choose the
electrode which best reflected the stroke volume, for the p~eular-patient. The
numberof wires needed wouldseem~tocause the lead to be: excessively stiff..
382 DESIGN OF CARDIAC PACEMAKERS
¯ "i il .
Figure16.7 Comparison ofthe current flow path for bipolar, eleetrode~:(l~ft) andmonopolar
electrode(right)eonfigurafions.Thecurrentin the bipolar arrangement hug~~e :.caOeter. The
monopolar lead causescurrenttoflowradially throughthe bloodof the:~art ~6~.rem0felylocated
sink (the paeemakdrcase). FromGeddes et al. (1991) ~~’
................
:’~
Even with the monopolar dectrode configuration, Geddeset al. (1991) noted
that the ’relationship between stroke volume and changes in itripedance is not
hnearover a very wide range of stroke volumes. However, this does not seem to
be a problem for rate adaptation since the goal of most schemes is to .hold the
stroke volume unchangedby changing the hedrt rate.
Attempts have been made tO come up with models for the observed changes in
impedance with the beating of the hem., The .heart Chamberhas been modeled as
a cylinder, a stackof different diameter disks and~ a crescent shaped volumeall
with little success in accurately predicting the time variations in impedance
observed: There are a number of reasoris why model attempts have been
unsuccessful. First,, the current"flow ~p~hs can- be. affected by the chambers
adjacent to the right ventricle. S~ond,title’heart tissue and blood~aredifferent in
bulk resistivity but-only by a~relativ~ly smallfact0r, with blood being
approximately three times more Conductive than heart wall~tissue. This means
that although we wish~;to detectonly blood Volumec~ges in the heart, changes
in the heart, wall position relative to the electrode can~havea significant affect on
the impedance. Third, and most troublesome’ for-modeling; is th/at the
conductivity of blood is highly anisotropic during flow due to the shape of blood
cells, It has been observed that the blood resistivity decreases with flowdueto the
alignment of blood cells (Ovsyshcher and Furman, 1993). ~This would appear
be asignificant complication in any.detailed modeling attempt.
The creation of an accurate geometric model for~the relation of impedance
to stroke volume-is not critical to,the uge ~f intracardiac impedance’ for rate~
adaptive pacemakers~, Numerousalgorithms have beeh~ foundthat: p,errnif the
available impedanceinformation to be successfully Osedfor rate adaptation:
RATE ADAPTATION BY EGM AND IMPEDANCE 383
and the paced heart rate in a resting individual over a moderate range of heart
rates. Fortunately, the effect provides negative feedback rather than positive
feedback as was the case for the Q-T interval.
Rate determination
The rate-adaptive algorithm has to determine, how to respond to the stroke
volumesignal. In response to exercise, an increase in SVindicates the heart rate
should increase so that the SVis returned .to its nominalvalue. The nominalvalue
can be a fixed value or just a long term average of the past measured values of
SV. The rate can be proportional to the error or it can integrate the error
between the SVsetpoint and the measured SVas shownin Saloet al. (1993). Salo
proposed that the rate be determined in response to the derivative of the SV. If
the SV decreased, the paced heart rate should be decreased. The proposed time
constant for the differentiator was 10 rain so that after 30 rain, the rate would
have settled back downto its resting rate.
Chirife (1992) proposed using the ejection fraction (EF) as the rate-controlling
parameter for the pacemaker. Ejection fraction is the SV divided by the End
Diastolic Volume.Figure 16.9 showsthat the ejection fraction tracks the SVquite
closely. The ejection fraction has an advantage in that preload effects are taken in
account in its calculation. Stroke volumeis somewhatlacking in specificity for
metabolic demandin part due to preload effe~tS~ Ventilation, posture~ and heart
rate can affect SV. At high heart rates, the EndDiastolic Volumefalls due to the
reducedtime available for filling and SVcorrespo~n"dmglyfalls.
Figure 16.10 shows the effect of reduced filling at high heart rates on the
stroke volume whena resting heart has its paced rate increased. The drop in SV
closely tracks the dropi n EDV.The ejection fraction howeverremainsrelatively
constant and thus is not coupled to the paced rate. This decoupling makes the
design of the control algorithm simpler, as feedback between the pacing rate and
the sensed parameter is minimal.
The stroke volume paradox has also been solved by using additional information
from the intracardiac impedance signal. Olive et al. (1988) used the peak of the
first derivative of the impedance(dZIdt)p as .a rate-correcting factor. The signal
(dZIdt)p was found to be proportional to contractility and~ independent of the
RATE-ADAPTATION BY EGM AND~IMPEDANCE 385
heart rate. The deviations in (dZ/dt)p are directly proportional to the level of
exercise. Therate response algorithm sets the pacing rate so that it is increased in
direct proportion to the difference between (dZIdt)p at rest and the current
measuredvalue of (dZtdt)p.
10
o. -10
independent of pacing rate. Lau (1992) reported that the PEPhas been shown
behave paradoxically upon assuming a standing posture. While the pre-ejection
interval has been shownto be. a good indication of heart contractility, Grubbet
al. (1993) proposed that pre-ejection period might also shorten due to reduced
filling arising from a drop in venous pressure.
WI PACING
.11 .... .
~4 , i I I 1 ~170
0 10 20 30 40 50 60 70
-"- TIME(sec)
The slope of the impedance over the region of in’terest is denoted as the
Regional effective slope Quantity (RQ). The RQis the regional slope of the
impedance over the ROI. The RQis different during exercise than it is during
rest. The RQchanges in response to mental stress the same way it changes with
physical stress. The change in RQis used to smoothlyadapt the pacing rate.
Impedance
(rel. counts)
150
. resting
t.=d .......... ,of interest,
120
110
~’ ~ Qexercis~’ "’
leO
104 120 136 152 168 184 216 232 248 264 280
Time frompace (ms) ....
:3chaldh~l{’, ft al. (1992) give a simple formula used to calculate the paced
heart ~ate using ;th~ changein slope of the intracardiac impedancein a particular
region of interest. Thestimulated heart rate is
¯
..... ¯ , -RQacmal - RQrest
HeartRate = BaseRate + R(~max - RQrest X~MaxRate L BaseRate) (16.,6)
1 benefit of
myocardiuminterface.
RATE ADAPTATION BY EGM AND IMPEDANCE 389
ONSETOF I R.V.
SYSTOLE
(ENO-DI~..~
!IASTOLE
! CARRIER
DIASTOLIC
LEVEL(DL) ""~
IEFffi (OL-SL)
~~ SYSTOUC
LE~U(SL)
The design parameters of the.cu~nt ~ource are its output impedance and voltage
compliance range~ Volt,age ComplianCeis the range of vOltages over which the
cu~r~nt sourc~ is able’to "p~0Vi~e’ a’~onstant current. The ~,~lta~e compliance
range is Usual!y !ess than ~ supply Voltage and Can be .muchless than the supply
voltage dep~t~ding on the~cifcuit"topo~ogy agd C0mpohent .valUes used to
im~l~nerlt ~e current source. ~n &e voi~age’ comliahce range.is exceeded,
the 6~rreilt ini~o the ibad become~dncontrolled and U~own.
The otltput impedance’ refleet~ the q~Jality of~’~he current s~urce. If the
current so~ce’s output current doffhot Change wi~i the vol,tage ~cross its
terminals, then the current sourc~ is said" to have infinit~ output impedance. A
finite output imp,ed ,~,cf provides an undesired coupling between ~the electrode
voltage and the cu~bnt ~lelivered :to the heart. Since ~e current is no .longer
preciseI3 ..... re~ent accuracy is reduced. ’
..... currdn(is06rce s accuracy can be
of the
affects
ofthe
output
RATE ADAPTATION BY EGM~ AND IMPEDANCE 391
Amplifier
source Zo Vo ~
Output Lumped
impedance impedance of
of current lead, electrode,
source and heart
? , IL = Is - ~o (16.7)
Io Zo
-~ (16.8)
~ =ZL + Zo
To show more ctearly the effect of the relativesize of the output impedance to
the load impedance, we use the ratio Of Zo to ZL~Toaehievea;eertain number Of
bitsof aceuracy, the ratio of Zo to ZLmust be
Z°
i 2Nbit~ ~ .~ (16.9)
ZL I=-1 ÷
negative feedback. The current that flows through the resistor Rg also flows
through the load impedance(the heart).
Cac Rg Cdcblock
I
Rbias
Vin
A relatively large capacitor, Cdc. bloek,~is put in series with the heart to
prevent de currents from flowing through the heart and electrode interface. A
high valued resistance, Rbias, is used to provide a path for the small de input bias
current of the amplifier and helps retain negative feedback at de.
The current source load floats relative to ground, so this circuit will not
work for a unipolar lead arrangement. In this application, the current source
electrodes are distinct fromthe pa,cing electrodes,
The output impedance of this current source is quite highand the voltage
compliance is also good. However, the full load current must flow through the
resistor Rg and thus there are losses in this resistor. This resistor also sets the
voltage-to-current gain ratio of the current source, so we can not make it
arbitrarily small as the voltage commandsignal would also have to be scaled
down.and resolution would becomea proble~m.
Another current source cire~t which can be~JSed for groundedioads uses
two instrumentationampliflers a n~ d a current sens~g resistor as shownin Figure
16.16., The arnpi~fier with gain A1 provides the ioad’~ .~. e~nt; all of whichp~ses
through resistorR. The amplifier with gain A2 senses and’amplifies the.voltage
drop across R and thus measures the load current that is being produced. By
having a large gain, A2, on the feedback amplifier, .,the voltage drop across R is
magnified.
The transfer function for the voltage-tO-current converter is
Io = (16.10)
RATE ADAPTATION BY EGM AND IMPEDANCE 393
E2
E1
The advantage of the circuit is that R can be made very small so that the
losses in it are minimal~The voltage-to-current conversion ratiois controlled by
the an~lifier gains instead of just the resistance value.
The output impedanceof this circuit is very~ high and the compliance range
can be controlled by~be second amplifier. The Circuit requires that the common
moderejection ratio of the amplifier A2 be good and that the sense and return
sense input on amplifier A 1 are of high input impedance and have good common
moderejection, since the electrode voltage is seen as a commonmodesignal to
these inputs.
Rate response ,curves for impedance-based~(PEP and VIP) and Q-T interval-
based pacemak~ are compared. Figure 16.17shows the rate.~s~nse of the Q-
T interval-based pacethaker has a long iag (120 s) in its ra~decay after the
cessation of exercise, The ideal sin~ xate is not shownin Figure 16A7, so. Be
~0fiallty of rate response can t~ be compared to .that of the PEP-based
pacemaker~ In ~FigurelO.18, the rate response of a PEP-based-pacemaker is
comparedtO sinus-~rate~For the PEP-based.SYstem,rat e :i~rease s rapidly ~ith
moderate onset delay, of approximately 10 s. The" rate doceleration is less,
occurring approximately~45 s after the cessation of exercise. .
Figure 16.19 Showsthat a pacemakerwhich detects tlae symp~tth6tic tone
respond to emotional demandsfor an increase in heart rate. The patient wasgiven
a~psyeho!0gLcMly~stressful color word test. During the test, differem cards
containing ~e name~ofa color (e.g. green) written in the letters of another color
(e,g.~ blue) are shown.to the patient whomust respond quickly with the color:of
the letters.
394 DESIGN OF CARDIAC PACEMAKERS~
PEP
HR-
min-1
120
Recovery Recovery Recovery---’~
110
16.8 REFERENCES
Dritsas, A., Joshi, J., Webb,S. B., Athatassopoulos, G.,. Oakley, C. M, and Nihoyannopoulos,
P., 1993. Beat to beat variability in stroke Volumeduring VV!/pacingas predictor of
bemodynamicbenefit from DDDpacing. PACE,16: 1713-1718.
Grubb,B. P.~, Wolfe,D. A., Samoil,D. A. Hahn,H.~, and Elliott, L. 1993. Adaptiverate pacing
controlled by right ventricular pre-ejection interval for severe refractory orthostatic
hypotension. PACE,16: 801-805.
Geddes, L. A. , Fearnot, N. E., and Wessale, J. L.. 1991. Multiple monopolar methodof
measuringstroke volumeof the heart. USpatent 5,058,583.
Hanck,J. A. 1991. Rate adaptivecardiac~pacerincorporatingswitchedcapacitor filter-with cutoff
frequencydeterminedby heart rate. USpatent 5,074,303. -~
Heijer, P. D., Nagelkerk~;D., Pert’ins, E. J., Horstman,E., VanWoersem,R. J., Neiderlag,
W., Jordaens, L., Wilde, P. D., Hameleers, W. B., and Lie,= H,I 1989. Improvediat~
responsive algorithm in Q-Tdriven Pacemakers--evaluati0nof initial;response to exercise.
PACE., 12: 805-811. "’~
Horstmann,E, and Koenn,B, 1990. Temporalrelationship betweenexercise and Q-Tshortening
in patients with Q-T pacemakers. PACE,12: 1080-1084.
Jordaens, L., Bakcers, J., Moerman,E., and Clement, D. ~L. 1990. Catecholarnine levels and
pacing behavior of Q-T~venpacemakers during exe~i~.PACE,13: 603--607.
KatritsiS, D. and Camm,A. J., 1992.~ Adaptive-rate pacemakers. Cardiologyclinics: Cardiac
Pacing, 10:671-688. .
Lau, C. 1992..:’I’he range of sensors ~andalgorithmsused in rate-adaptive cardiac pacing PACE,
15:1177-1211.
Maloney,J. D., Helguera, M. E., and Woscoboinik,L R. 1992. Physiology of rate responsive
pacing Cardiologyclinics: CardiacPacing,10: 619-629,
Klesch, L J. 1992. Precision voltage controlled current source with variable ~ compliance. US
patent 5,153,499.
Olive, A. L., Pederson,B. D., and Salo, R. W.1988. Closedloop control of stimulator utilizing
rate of changeof impedance.USpatent 4,733,667.
Ovsyshcher,I. and Furman,S. 1993. Impedancecardiography for cardiac output estimation in
pacemaker patients: reviewof the literature. PACE,16i i~t13--1422.
Ohte, N., Hashimoto,T.,Narita, H. Takase, R.,Kobayashi,K., Haynao,J., ~and~Fi~jinami,T.
1990. Noninvasive evaluation of left ventricular performancewith a newsystolic time interval:
.the QVpeak, and comparisonwith established systolic time intervals Am.J. Cardiol., 66:
1018-1020.
Pickett~ B. R. and Buell, J. C. 1993. Usefulness of,the impedancecardiogramtO’reflect left
ventricular diastolic function. Am..J.Cardiol., 71: 1099.-~1103
Salo, R: and Pederson, B. D. 1993. Biomedicalmethodfor controlling the ~administration of
therapyto a patient in responseto physio!6gicaldemand.US1~ t, ent 5,190,035
Schaldach,M. M. 1992:Electrotherapyof the heart. Berlin: Spfifi~r-Verlag.
Schaldach,M. and Hutten, H. 1992. i~tr:acardiac :impedanceto determinesympatheticactivity in
rate responsive pacing, PACE,.lS: 1~78-1786 .... ~ .... .~
Schaldach, M. M. 1990~(~ardiac ~e,~r’withphysiologidal c0ti~l, USpatent 4,919,137.
Steinhaus, B. M., Napph61z,T. A:,’ Nol/m,J. A. and Morris; R. A~:, 1993. Minutevolumerate
responsive pacemakeremployingimpedancesensing on a unipolar lead. USpatent 5,201,808
electrodes wherethe current density is high, rather than in the desired larger area of
the lung between the two electrodes where the current density is low. Twolarge
electrodes can be used to minimizethis problem of sensitivity to changes near the
electrodes because the current density is low everywhere.
Pacemakerscompromiseby using the three (tripolar) electrodes as shown
Figure 17.1. The pacemaker can serves ~as alarge electrode for both injecting
current and measuring voltage. Injecting current through the ring electrode and
measuring voltage at the tip electrode avoids the sensitivity to changes in
impedancethat wouldoccur by using a single small electrode.
Lead
can
Heaff
Inject
current
Measure
~ectrode
voltage
electrode
Figure17.1 Thepacemaker can is large and serves to both inject current andmeasurevoltage
withoutemphasizinglocal sensitivity: The:ringelectrode~itrjdcts currentandthe tip electrode
measuresvoltage, whichalso avoidsemphasizing local sensitivity. .
For Teletronics pacemakers, after determining the impedancefrom the samples, the
signal is filtered for information,~ro~ 0,!-1 ~z*, This~ miniroizgs undesirable
in~pedance changes caused by ~strtke~;olume’ch~ges. A zer0]6r6~S~g detector
measuresthe ventilatory rate. If 7 of 10 samples are of one polarity, the signal is
consideredof that pol, ~arity, Eachtime th~ po!a4"itYch ~anges,,it is rec~rde~!and used
to compute the ~’~ntilatory rate VR, assh(~wninFigureAT~3~ ~ i ’~, .
Tile impedlahee~ signal is also rectified a~d filtered to ~yie!d"a Signal
propoMOnal io tidal ~olu~ne: ~s i~ multipliedbY Ve~til!ttory~iiate ~i ~eld m~nute
Ventilation (MV). The resting MVdtfin~ ~e lo~er ~ pacing rate. B~cause the
impedance may dd’ft with~, the rtS~gMVis dttem~ned b~/a long-’term (1 h~
averager. The MVis determined by a short-term. (36’ s) averager. The long-term
RATE A;~DAPTATION BY MINUTE VENTILATION 399
resting MVis subtracted from the short-term measured MVto yield thechange in
MV,AMV.This algorithm works.correctly if the patient has been resting for 1 h,
then begins exercise. However,if the patient continues exercising for 1 h, the long-
term average creeps up to the short-term average and AMV goes to zero, whereas it
should stay high. To prevent this problem, any time that impedance change
exceeds 50%of its maximalvalue within 35 s, it locks the long-term average so it
doesn’t creep up.
TO SWITCHES
Figure17.2 When the lowerpair of switchesis closed, the pacemaker can sense electrograms
andstimulate the heart. Whenthe upperpair of switchesis closed, the pacemakercan measure
impedance
betweenthe tip andthe case (Nappholz,1990).
,~rI I ..
l i ’: -l’ ~ Target
~ ~,. : hei~t
r rate
¯ --
l-termI ~
Head
rate
For Medtronic pacemakers, so that heart rate changes are not sudden, the
physician also programs in acceleration and deceleration time constants of 0.25,
0.5, 1, 2.5, 5 or-lOrain.
17.4 INTERFERENCE
the DDDR atrial pace region~ the ~pacer stimulates the atrium at the metabolic
indicator rate (region IV) (Nappholzet al.,. 1992).
Figure 17.5 shows timing diagram of acardiac cycle and its associated time
intervals indicating howthe pacemakerresponds to a cardiac event whensensed in
different time intervals (Nappholzet al., 1992).
900 ms
Ird~’val
600rn~
180 ms
Hgure17.4 Whenthe atrial rhythmis at long intervals, the AMS pacemakeroperates in the
DDDRmode(regions tIT and IV). When
the interval is shorter than PVARP,
it operates in the
VVIRmode(regions I and ID. ThePVARP limit and AVdelay shorten with increasing AMV.
t I I Ti~
VBITRiCLEi/llx"
vvt ~ AU[RT
~I
TARP j I
ATRIkL "VVIR W,: r i |
SENSING
Ki~VlOIt I AD.,~X,UT~
NONITORI INHIBIT I , .....
ltEFIIACTO~Y
I Tr t ~Tr I
I
Figure 17.5 If a P wave is sensed, after the,AV delay the, ventricle is paced, During file atrial
alert time, a sensed P wavewill trigger a synchronized pacing of the ventricle (Nappholz et al.,
]992).
Figure 17.6 shows that the Legend Plus pacemaker from Medtronic, Inc. extracts
information from two sensors. A piezoelectric ceramic bondedto the inside of the
flexible pacemaker can provides a signal caused by the mechanical muscle
movementunder the can. Thus body movementprovides information that yields a
rapid increase in pacing rate at the onset of activity. An impedance-based
measurement of minute ventilation provides a more accurate pacing rate for
sustained exercise.
IO
I00 8~ 83
26
65 24 I
Figure17.6Dualsensorrate-responsivepacemaker.
Piezoelectricsensor20 yields fast response
to motions"Impedancecircuit 82 ,measuresminuteventilation to yield long-termaccuracy
(Wahlstrand
et al., 1993).
Cooper (1994) has designed an algorithm that combines inputs from activity
and minute ventilation inputs. The algorithm isbased on a hierarchical fuzzy logic
expert system. A group of expert etectrophysiologists recommended that: ~
(a) (b)
Figure 17.7 (a) Dual-sensor response to a simulated step function of exercise. (b) Dual-sensor
response to a simulated stepped ramp function of exercise. From Cooper, D. 1994. A dual-sensor
rate-responsive pacemaker algorithm incorporating a fuzzy logic expert system. Computers in
cardiology 1994. Piscataway, NJ: IEEE.
17.7 REFERENCES
17.1 Explain the advantages and disadvantages of the use of minute ventilation as compared
with the use of bodymotionfor rate-adaptive pacemakers.
17.2 Whenmeasuringelectrical impedancein the body, explain the problemwith using small
electrodes~ instead oftarge electrodes.
17.3 Whenmeasuring electrical impedancein the body, explain the advantage of using four
electrodes instead of twoelectrodes.
17.4 Whenmeas~ngelectrical impedance in the body, explain how to avoid undesirable
tissue stimulation.
17~5 Whenmeasuring electrical impedance in the body~ explain how to avoid pickup by
pacemaker~ sensing amplifiers.
17.6 Whenmeasuring electrical impedance in the body, explain how to avoid pickup by
surface ECGs.
17.7 Explain howto calculate minuteventilation from impedance.
17.8 Whenmeasuring minute ventilation, explain howto avoid errors due to the long-term
drift of impedance.
17.9 Whenmeasufiagminuteventilation by impedance, explain howto avoid errors due to
long-termsustained e~ercise.
¯ 17.10 Whenmeasuring’initiate ventilation by impedance,explain howto avoid sudden changes
of packingrate.
17.11 Explain the situations where external interference maycause malfunction in minute
ventilation r~te-adaptivep~acemakers.
17.12 Explain-theresu~ of sensing P wavesduringthe atrial monitorinterval.
17.13 E~plainthe resdl~ of sensin~P wavesduringthe atfialinhibit~interval.
17.14 Explainthe result of the atrial alert timertimingout.
17.15 Explain the- advantagesof using dual minuteventilation andbodymotionsensors in rate-
adaptive pacemakers.
18
Antitachycardia Pacing
Rex S. Piper
His
RBB
Ablation
LBB
Figure 18.2 Showsa reentrant circuit of His bundle branches. The right.bundle branch with its
slowconduction(shaded)is the critical part Of the self propagatingcircuit. A cut u~ing~urgery
~blafion"]h’ the slowconductionpath Canterminatethe tach~,c~rdia. His = His bundl&RBB = right
bundlebranch; LBB= left bundle"branch..
Long-term
efficacy /20 ~ 1’5~ :20 ~ ,~ 20 ’~15 .~~’~, ~ 1,5
Comfort/5 :
Sideeffects/5
Convenience/5 1 4 5 3 2
Cost/lO 6 6 10 . 5 2
Pmscfibin~Eas~e/5 5 I I, ’ " I’ " ’ I
CUre/15 ~ ~
Rescue/t5 , = .... 0 - 0, ,’=-
0 11: - 15
Risk/IS . , 13 4 ..... 12:: ,’::.7.’5" 12
Compliance/1O= ,3 ..... ,10 :10 i ~ ’8 " ¯ .... 8
Success/l~0 ,. 7 .8 6 , z,.~5 -10
electrical wave front and the tachycardia generating circuit. Here we present
these higher level models to enable study of pacemaker effectiveness and
interactions.
TachyCardia acceleration
Just as the dynamicnature of some reentrant tachycardia circuits is utilized to
terminate their encircling waves, this same principle can result in the formation
410 DESIGN OF CARDIAC PACEMAKERS
18.,4,1 Manualactivation
Manually activated antitachycardia pacemakers have been used since the first
antitachycardia pacemaker was installed in 1968. They are ~i~eally, suited for
patients whoremain conscious during their periods of tachyc~dia. The patieht,
by sensing, characteristic symptomso~ ia rapid pulse rate, can be as accurate asthe
piaysician s reading of an electrocardiogram. ~s tends to limit false triggfring,
although some patients can become overconfident in their ’abilities and may
initiate pacemaker therapy inappropriately. Older externally activated
antitachycardia pacemakers employ a sihaple n~agnet while, ,some newer models
are activated i b3~ radiofrequency transmitters. Because mantmi intervention is
requtred, their overall usefulne.ss and convemenceIs hmlted. Since patlen
cooperation is mandatory, disabling symptomsC~ be fatal. The~advantages of
manualdetection include simplicity and the ability to restrict patient activation to
a hospital setting wherea physician and defibrillation can~ be p~esent.
412 DESIGN OF CA~IAC PACEMAKERS
Rate-related parameters
More sophisticated detection algorithms now use additional rate-related
parameters to further classify the baseline heart rate. The rate of onset, the
stability of the heart rate and the duration of the rapid rate are commonly used to
enhance baseline heart rate data. Rapid onset of a high rate is associated with
tachycardia, whereas gradual onset usually characterizes exercise. A study
examining rate of onset in patient subjected to 30 s of maximal bicycling
(Mercandoet al., 1988) demonstrated excellent tachycardia detection results using
this scheme. A .number of’commercial devices already incorporate, rate of onset
parameters. Measurementof a regular R--R interval is .also used to distinguish
exercise related high rates, with its respiratory related variation, from more
stable tachycardia intervals. However,such stability parameters must be chosen
very carefully, since some tachycardias will include a limited amount of R-R
interval variation. Finally, the duration of the tachycardia in terms of the number
of consecutive cycles measured can be used to prevent responses to temporal
arrhythmias and to allow spontaneous termination of transient tachycardias.
Rate BC TC fibrillation
Activity low high high q0w
Electrogram
Some experimental third generation antitachyeardia pacemakers with
cardioversion/defibrillation capabilities perform advanced analysis of the
tachycardia wave form to differentiate between simple tachycardia and
fibrillation. Fast Fourier analysis, template matching, gradient pattern detection
and probability density function are someof these sophisticated calculations. As
these algorithms becomemore available and more efficient, they could be used to
distinguish tachycardia itself from normal rapid heart rate.
Burst Underdrive
Fta~gUre.18.8
hyca/dia Tachycardia:terrnination
wave algorithms-are
fdrm.PacedpU!s~s~are representedby~’ demonstrated
arr~vheads. PES~against a background
paciiaginclude~pulse
number"andtimingvariability: Burst; ramp;and ~indetdriveprotocols include pulse timing,
number~andfrequ~ncyvariation. Since’ a!tachycardia’s response to a pulse train is very
unpredietable,.for
clarity, the waveformis not~shownafter pulsetrains start.
ANTITACHYCARDIA PACING 415
Since rapid tachycardias rarely respond to the single capture techniques used by
PESstimulation, pulse train delivery is used. Within the two classes of pulse train
algorithms, burst and underdrive pacing, there are a large numberof protocols.
As can be seen in Figure 18.8, a number of delivery parameters are adjusted by
these different algorithms, including burst delivery initiation and termination
timing, pulse to pulse interval, and the numberof pulses.
Burst pacing utilizes a .short train of pulses,~ ,usually between 5 to 15,
delivered at a rate faster than the tachycardia to effect termination. Since the risk
of acceleration is greater for multiple pulses, even for slow well tolerated
tachycardias (Fisher, 1990),~a key operational objective is to keep the numberand
rate of pulses as low as possible. Earlier burst delivery algorithms used a short
burst of traditionally fixed cycle length pulses. Moresophisticated algorithms can
adaptively adjust the cycle timing as a percentage of the tachycardia cycle timing,
and/or shift the burst train forward or backward. Another type of burst pacing
uses ramping to .modify the pulse train. Ramping-the pulse frequency down is
designed to help minimize the risk of accderation. Somedevices will ramp the
frequency up and then down,which has less Side effects than a very fast burst but
still reaches a peak target frequency.
~-Underdrive is a technique used to treat some patients with slow, well
tolerated, hemodynamicallystable tachycardias (Echt et al., 1990). Unlike burst
pacing whichdelivers short trains of pulses, underdriving can continue for a long
period of time. It delivers a slow constant train of pulses, which randomly
interact with the tachycardia until an appropriately timed pulse terminates it:
Care must be taken to ensure that the frequency of the pulse train is not a
harmonic of the tachycardia wave form or the .tachycardia maybe reinforced or
accelerated. Burst pacing and PES are substantially more effective for the
treatment of frequent, reentrant tachycardias.
Overdrive suppression
The occurrence of triggers, for example early or late afterpotentials, .have a
strong, relationship toheart rate. There appears to be an optimalintermediate rate
in the range of 10 to15 beats .per minute above :the normaLheart rate where
trigger frequency is suppressed, Manystudies confirm thatov.er.driving the heart
rate can significantly reduce the number of tachycardias (Mehra, 1990).
416 DESIGN OF CARDIAC PACEMAKERS
Inhibition
Inhibition is used to prevent tachycardias by prolonging the refractory period of
the causative substrate. Preexcitation is administered to the site by a pulse during
the refractory period, thereby extending the refractory period of the tissue. If
applied to strategic points, the resultant prolonged refractoriness can choke the
substrate and prevent tachycardia from occurring. However, due to electrode
implantation restrictions, inhibition remains primarily a laboratory phenomenon.
ventricular signals must be sensed by the atrial sensor. In patients using DDD
mode, 41%have shownretrograde conduction from the ventricle (Nitzsche et al.,
1992). The incidence of ventricular atrial conduction varies from 66%to 100%
(note that VAconduction can appear and disappear) when AVconduction
normal to 0% to 25%with third degree AVblock (Bertholet et al., 1985). The
order of events starts whenthe pacemaker senses a premature atrial event. This
trigger can comefrom numerous sources: a premature atrial contraction (PAC),
a retrograde premature ventricular contraction (PVC), electromagnetic
interference with the pacemaker, and myopotentials from skeletal muscle. In
response, the pacemaker then paces the ventricle at a time whenthe retrograde
conduction pathway is not refractory, allowing the ventricular pulse to back
propagate. Whenthis is sensed as an atrial event, a racing endless loop
tachycardia cycle begins.
Pacemaker
~mode ! ( [7~r~go~tion
%
~anterograde_~ ~ VA c~nduction}
AV conductio n ~
VACT prevention
As modern pacemakers have sought to track higher rates by lowering the
PVARP,ELT has become much more prevalent. Some have :added an anti-ELT
algorithm which utilizes the ventricular atrial conduction time (VACT)
differentiate true atrial events from retrograde P waves (Fisher et al., 1986;
Limousin et aL, 1990). This procedure starts with monitoring ~the VACTfor
normal timing, If the VACTdrops below 450 ms, ELT is suspected and tested
for by shortening the ~AVIa~preprogrammedvalue (see Figure 18.10)~ If the
VACTremains constant within, a range limit,, back propagating P waves are
confirmed. A normal atrial pulse signal would be unaffected by an early
418 DESIGN OF CARDIAC PACEMAKERS
ventricular pace. To terminate the ELT, the PVARP is then lengthened to 450 ms
for one cycle, rendering the next P wave unsensable. Additional proposed
sophistication includes adjusting the length of the PVARP to a value longer than
the longest VACT as measured during the last ELT. This would prevent future P
wavesensing but also tends to limit high rate tracking.
(b)
I lAW- VACT
+ I
WARAD prevention
Since endless loop tachycardia trigger events are well known, it would be
advantageousto prevent the initiation of this tachycardia. To do this, Nitzsche et
al. (1992) introduced a new parameter called the window of atrial rhythm
acceleration detection (WARAD), which is equal to 75%of the preceding P-P
interval. As shown in Figure 18All if an atrial event is sensed during the
WARAD, it is considered to be a trigger (a ~ premature ventricular or atrial
contraction). Instead of resetting the AVI delay on this event, only the AEI
(Atrial escape interval, also knownas ventriculoatrial, interval, VAI)is reset and
atrial activity is monitored.If no atrial event occurs within the AEI, the trigger is
considered to be isolated and after the atrium is paced, the AVI.is temporarily set
to 31 ms to maximize the next atrial sensing: Endless loop tachycardias are
prevented if the trigger event falls within the W~ARAD. If a newatrial event is
detected during the AEI, the pacemaker suspects atrial tachycardia and
temporarily limits the pacing rate to 120 bpm. As a final fallback, the pacemaker
defaults to VDImode.
to exclude retrograde P waves from atrial sensing and during exercise the
PVARPcan be ,shortened to allow high rate tracking in DDDRmode. The main
disadvantage of using activity sensors for ELTprevention is if the sensor does
not have a fast response time, high heart rates from quick physical exertions
might be mistaken for tachycardia. However, the future prospect s of using
biosensors for pacemaker mediated tachycardia discrimination remain promising.
A very different approach to ELTprevention using atrial sensor sensitivity
adjustment has been successfully demonstrated by Rognoniet al, (1991).-Atrial
sensing in advanced DDDpacemakers with programmable sensing parameters
were optimized. ~The LEM/CCS Twinal 20/30 pacemaker allows a wide rangeof
atrial sensing options, and its telemetry options include transmissions of sensed
electrograms used ~for optimization. After recording and analyzing the sensed
characteristics of both atrial and retrograde P wave signals, optimal threshold
levels were used to filter P waves from true atrial beats. Reprogrammingthe
atrial sensing parameters resulted in immediate ELTinterruption in 14 out of 15
patients. The obvious risk in atrial sensing optimization is undersensing due to
small safety margins.
It is important ;to note that allthe patients selected for: antitachycardia pacing
therapy have been very carefully screened. This tends to bias the results. ~ The
420 DESIGN OF CARDIAC PACEMAKERS
18.7.2 Surveys
IL
Fahraeus, 1984
uderitz; 1982
] Kahn, 1976
8
9
12
PASAR4151
Intermedics
C-~bettach
Medtronic 5998
Auto, scan
Auto, manual
underdrive, burst
Manualburst-
4
5
10
I (multieenter)
The ideal antitachycardia device would include treatment options for bradycardia,
tachycardia; and fibrillation. The features of such a device are listed in Figure
1,8.!5. Third generatign antitachycardia devices including cardioversion and
defibrillation capability are the subject of Chapter 19..
Sensing’
Distinguishp~thologic from physiologietachycardias
Differentiatetwo(ormore) pathologic tachycardias
Recognize nonsustaincd tachycardias
.,Autorf!aticsensitivitygainfor differentrhythms
Pacing ~.. .
¯ Bradycardia pacing(single anddualchamber,physiologicsensor)
Antitaehye,ardiapacing(extmstimulus, burstor adaptive)
Differentantitachycardiatherapiesfor,differenttaehycardias
Noninvasiveelectrophysiologic studycapabilities
Taehyeardia prevention
Cardloversion/defibrillation
Lowenergysynchronizedcardioversion
Highenergycardioversion of defibrillation
Bidirectional,biphasic,or sequentialshocks
18.10 REFERENCES
Arzbaecher, R., Bump,T., Ripley, K. L., Yurkonis, C. Jenkins, J. and Noh, K. 1989.
Implantable microprocessor based devices for the management of arrhythmia. Computersin
Cardiology, 9: 29-34.
Barold, S., Ryan, G. F., and Goldstein, S. I989. The first implanted tachycardia-terminating
pacemaker. PACE,12: 870-874~
Bertholet, M.~Materne,P., Dubois, C., Mareelle, P., Beckers, L, Demo~ulin,J. C., Fourny,
and Kulbertus,~H. E. 1985. Artificial circus movement tachyeardias: incidence, mechanisms,
and prevention. PACE,8: 415"423.
Bertholet, M., Demoulin, J. C., Waleffe, A., and Kulbertus, H. 1985. Programmable
extrastimulus pacing for long-term managementof~supraventricular and:~ ventricular
tachycardias:clinical experiencein 16 patients. Am.HeartJ., 110: 582-589.
Block, M., Borggrefe, M., and Hammel,D., et al. 1991. Pacer-cardioverter-defibfillator (PDC),
utilization, efficacy and complicationsof antitachycardia pacing (abstract). J. Am.Coll.
Cardiol., 17: 54A "- ~ ¯
Bonnet, C. A., Fogoros,R. N~; Elson; J. J., Fiedler, S: B., and Burkholder,J. ’A. 1991. Long-
term efficacy of an antitachycardia pacemakerand implantable defibrillator combination.
PACE,14: 814-822.
de Bakker, J. M.T., Vancapelle, F. J. L., and Janse, M~.L 1987~..L0calizationof the site of
origin of Ve~tficulartachycardiain the ChrOnicptiase’of my0eardi~l:infarction. In G: B~ithardt,
M. Borggrefe and D. Zipe$ (eds)Nonpharmocological therapy of tachyarrhythmias. Mt.
Kisco, NY: ¯ Futura Publishing.
den Dulk, K., Bertholet, M.,. Bi’ugada, P., et~fl. 1984. Clinical experience with implantable
devices for control of tachyarrhythmia.PACE,7: 548-556. ¯
Echt, D. S., Lee, J. T. and Hammon, J.: W:1990. !mplantableand intra0perative .assessment of
antitaehycardia devices. In S. SaksenaandN:Goldschlager(eds) Electrical therapy Of cardiac
arrhythmias.Philadelphia: Saunders.
EI-Sherif, N. 1990. Electrophysiologicmechanisms in electrical therapyof ventricular tachycardia.
In S. Saksena and N. Goldschlager (eds) El, ectri,cal therapy of cardiac arrhythmias.
- Philadelphia: Saunders.
Ellenbogen,K., Welch,W., and Luceri, R., et al. 1991. Clinical evaluation of the GuardianATP
4210ir/aplaatable pacemaker/defibrillator:worldwideexperience(abstract). PACE,14: 623.
Fahraeus, T.,-.Lassvik, C., and Sonnhag, C. 1984. Tachycardias initiated by automatic
~ antitachycardia pacemakers. PACE,7: 1049.
Falkoff, M. D., Barold, S. S., Goodfriend, M. A~Otig, L. S., andHeinle, R. A. 1986. Long-
term managementof ventricular tachycard~a by implantable automatic burst tachycardia-
terminating pacemakers.PACE,9: 885--895.
Fisher, J. D. 1990. Clinical results with antitachycardia pacemakers. In S. Saksenaand N.
Goldschlager(eds) Electrical therapyof cardiacarrhythmias.Philadelphia:Saunders.
Fisher, J. D. 1990. Antitachycardia pacing in the acute care setting. In S. Saksena~ndN.
Goldschlager(eds) Electrical therapyof cardiacarrhythmias.Philadelphia:Sannders.
Fisher, J. D., Kim,S. G., and Mercando,A. D. 1987. Argumentsfor antitachyeardia therapies
using a graded point score model. In G. Breithardt~:M. Borggrefe ~a~a~dD. Zipes (eds)
Nonpharmocologicaltherapy oftachyarrhythmias. Mt: Kisco, N¥:Futura Publishing~
Fisher~ J. D., Johnston, S. K., Furman,S., and Mercando,A. M. 1986. Implantable phcers for
tachycardia termination: Stimulation techniquesand long:termefficacy. PACE,9:!325-~1333.
Fromer, M., Shensa, M., Kus, T.~ and Page, P. 198~ Managemen~t,of ~ a~patient with recurrent
sustained ventricular tachycardia with a newsoftware-baSedantitachycardia pacemaker.J.
ElectrophysioL, 1:133-139
Fromer, M., Schlapfer, J., Fischer, A., and Kappenberger, L. 1991. Experience witha new
implantablepacer-cardioverter-defibrillator for the therapy of recurrent sustained~ventricular
tachyarrhythmias:a step towarda universal ventficular tachyarrhythmiacontrol device. PACE,
14: 1288-1298.
Greve, H., Koch, T., Gulker, H., and Heuer, H. 1988. Termination of malignant ventricular
tachycardiaby use of an automaticdefibrillator (AICD)in combinationwith an antitachycardia
pacemaker. PACE,11: 2040--2044.
Griffin, J. C., and Sweeney,M. 1984. The management of paroxysmaltachycardias using the
Cybertach-60. PACE,7: ~ 1291-1295 ~-
Hiles, M. C., Bourland, J.. L,; Wessale, J. L., Geddes, L,A., and’ Voorhees, W.~D:~1993~
Detectionof ventricular tachycardiaand fibfillati0n using coronarysinus bloodtemperature:a
feasibility study. PACE,16: 2266-2278.
ANTITACHYCARDIA PACING 425
Kahn,A., Morals, J. J., and Citron, P. 1976. Patient,initiated rapid atrial pacing to manage
supraventriculartachycardia. Am..J. Cardiol., 38: 200..~
Klein, L. S., HackeR,F. K., Miles, W. M., Mohamed,Y:, and Zipes, D. P. 1993. Clinical
experiencewith newimplantableantitachycardiacardioverter-defibfillators: In G. V. Naccarelli
and E. P. Veltri (eds) Implantablecardioverters,defibrillators~
Boston:BlackwellScientific.
Lau, C. P., Tai, Y~T. Fong, P. C., Li, J. P., Chung, F. L., and Song, S. 1992. The use of
implantable sensors for ~the: control of pacemakermediated tachycardias: a comparative
evaluation betweenminute ventilation sensing and acceleration sensing dual chamberrate
adaptive pacemakers. PACE,15: 34-44.
Lau, C. P. 1991. Sensors and pacemakermediated tachycardias. PACE,14: 495-498.
Leiteh, J. W., Gillis, A. M., Wyse,D. ~G.et al. 199.1. Reductionin defibrillator, shockswith an
implantabledevice combiningantitachycardia pacing and shocktherapy. J. Am.Coll. Cardiol.,
18: 145-151.
Limousin,M., Bonnet, J. L., et al. 1990. A newalgorithm to solve endless loop tachyeardia in
DDD Pacing: a multi center study of 91 patients. PACE,13: 867-874.
Luderitz, B., d’Alnoncourt, C. N., Steinbeck, G., and Beyer, J. 1982..Tberapeutic pacing in
tachyarrhythmias by implanted pacemakers. PACE,5:366-371.
Ludedtz,B. 1991. The impact of antitachycardia pacing with defibrillation. PACE,14:312-316.
Mehra,R. 1990. Electrical stimulation techniquesfor preventionof ventricular tachyarrhythmi~s.
In S. Saksena and N. Goldsehlager (edS) Eledtribal therapy" of cardiac arrhythmiiiS.
Philadelphia: Saunders.
Mercando,A. D., Fisher, J. D., and Furman, S. 1988. Automateddetection of tachycardias by
antitachycardia devices. J. Am.ColLCardiol., 11: 308"316.
Moiler, M. Simonsen,E.Ing;P. A., and.Oxhj, H. 1989. Long,termfollow-up of patiet~tS treated
with automatic scanning antitachycardia pacemaker:PACE,12: 425-430.
Nathan,A., Hellestrand K., Bexton,R., et al. 1983. Problemswith patient activated pacemakers
for tachyeardiatermination(abstract).~PACE,6:.A-137
Newman, D.M., Lee,~M.A., Hen’e, J, M., Langberg, J. J., Scheinman,M: M., and Griffin, J~
C. 1989. Permanentantitachycardia pacemakertherapy for ventdcular tachycardia. PACE,12:
1387-1395.
Nitzsche, R., Girodo,S., Limousin,M,Cazean,S., et al. 1992. Useof a newfallback function to
prevent endless-looptachycardias: first clinical results. PACE,15:1851~-1857.
Nitzsche, R., Gueunoun,M., Lamaison,D., Lascault, G., PiogercG., Richard, M., Malherbe,
O., and Limousin,M. 1990. Endless-looptachycardia protection. PACE,13:’1712-1716.
Occhetta, E., Bolognese, L, Magnani,A., Francalacci, G., Rognoni, G., and ,Rossi, P. 1989
Clinical experience with Orthoeor II antitachycardia pacing system for recurrent
tachyarrhythmiatermination. J. Electrophysiol., 3: 289~300.
Palakurthy, P. R., and Slater, D. 1988’~Automaticimplantable scanning burst pacemakersfor
recurrent tachyarrhythmias. PACE,11: 185-192.
Peters, R. W., Shafton E., Frank S., et al 1978. Radiofrequency-triggeredpacemakers:uses and
limitations. Ann. Intern. Med., 88:17
Portillo, B., Medina-Ravell,V., Portillo-Leon, N., et al. 1982. Treatmentof drug resistant A-V
reciprocating tachycardias with multiprogrammabledual demandA-Vsequential (DVI, MN)
pacemakers. PACE,5: 814.
Rognoni,G., Occhetta, E., Perueca, A., Magnani,A., Francalacci, G., Audoglio,R., and Rossi,
P. 1991. A new approach to the prevention of endless loop tachycardia in DDDand VVD
pacing. PACE,14:1828-1834
Rosen, M. R. 1990. Mechanisms of cardiac impulse initiation and propagation. In S. Saksenaand
N. Goldschlager(eds) Electrical therapy of cardiac arrhythmias.Philadelphia:Saunders.
Rosenthal, M. E., Marchlinski, F. E., and Josephson, M. E. 1990. Complicationsof implantable
antitachycardia devices: diagnosis and management. In S. Saksenaand N. Goldschlager(eds)
Electrical therapyof cardiacarrhythmias.Philadelphia:Saunders.
Roth, J. A. and Fisher, J. D. 1993. Antitachycardiapacing-ICDinteraction. In G. V. Naccarelli
andE. P. Veltri (eds) lmplantablecardioverters-defibrillators.Boston:BlackwellScientific.
Rothman,M. T., and Keefe, J. M. 1984. Clinical results with Omni-Orthocor,an implantable
antitachvcardia pacing system. PACE,7: 1306-1312.
Ruskin, J. ~., Garan, H., Poulin, F., and Harthorne, J. W. 1980. Permanentradiofrequency
ventricular pacing for management of drug-resistant ventricular tachycardia. Am.J. Cardiol.,
46: 317-321.
Saksena, S., and An, H. 1990. Electrophysiologic mechanisms underlying managementof
supraventricular tachycardia by electrical stimulation. Mechanismsof supraventricular
tachyarrhythmias. In S. Saksenaand N. Goldschlager (eds) Electrical therapy of cardiac
arrhythmias.Philadelphia: Sannders.
426 DESIGN OF CARDIAC PACEMAKERS
Sudden cardiac death (SCD), a major problem in Europe and in North America,
accounts for 1200 deaths in the U.S. every day. In the Framinghamstudy, for
deaths between the age of 35 to 64, nearly one out of three is caused by heart
disease. Suddendeath itself is defined as death occurring within 24 h of onset .of
symptoms. :-
Depending on the underlying mechanism of SCD, the duration .of illness
beforethe-death vades~ For k example, most SCD caused by ventdcular
tachycardia/ventricular fibrillation occurs in a few minutes. ~Onthe.other hand,
SCD caused by pump failure could take considerably longer ........
SCDis a multifactodal problem that is usually associated with various forms
of structural cardiac abnormalities that interact with acute triggers. Someof the
more apparent triggers include myocardial ischemia, neurohumoral changes,
electrolyte abnormalities, pharmacological agents, and electrophysiologic events
(Naccarelli, et al., 1993).~ ~
sudden death
total
expected
TIMEFRONIMPLANTATION
IN Y~EARS
HV DISCRETES
CONTROL& DATABUS
POWERSUPPLY & PULSE DEL~ERY
Figure19.2 Microprocessor-based
third generationICDFromCarroll andPless (1991).
Each pacing circuit communicateswith the atrium (for pacing the atrium)
the ventricle (for pacing the ventricl~e) thn ~ugh two lines, Oneof the lines is
switchableground,and the other is ff~’p,a,~it i8 electr0de~ whichis also the input to
the sense amphfier in IC2. Both the atn~ an, t v~n~ial~ pacing lines pass through
IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS 431
19.4 BATTERY
A
~ _ _ ./~- ~,.,..~ .~.v,.¢o vet,.
19.3(b) shows the typical open circuit voltage for the’ lithium silver vanadium
pentoxide Cells.
Figure 19.3 shows that the vanadium~iiver pento~de pro~,ides approximately
3.2 V per cell ~nd’typicaliy two cellsare cbnne~tedins~ri~s in thedefibrillator so
that the battery voltage at ’the beginningof life is i~n ~the rang6of 6.4 V. Becausethe
digital circuits take 3.3 V, there is a ’n~ed for an ~ffi~ient~/oitage downconverter
and regulation. SomeICDdevices use two batteries~ ~~e lithium Silver vanadium
pentoxide for the high~v~ttage ~harging circuit and li~ium m~ganesedioxide or
lithium iodine for the 10w-voltagecircfiit. This arrangementeliminates the need of
a voltage downConverter.
and it is a function of the voltage drop in the regulator. Thus, an efficient voltage
downconverter to an intermediate voltage in between the battery voltage and the
regulated voltage would improve the efficiency of the systems. Efficiency without
a voltage down converter could be derived by dividing the power used in the
digital circuit by the powersupplied by the battery, and it could be Simplified as
Withregulated voltage at 3.3 Vand battery voltage at 6.4 V,~ the efficiency is
52%, and it would improve as the battery voltage goes downto 4.5 V, which gives
efficiency only up to 73%.
Using the downconverter, the efficiency is defined as
VBATTERY
i ~ SM2
aC2 7 I
S 10~,
!
!
I I I
P! P2 P3 ~
(a)
VBATTERY VBATTERY
CI~"
C2 ~"
C3 ~"
X7
~
.p1
to 5.5 V, 4/5 of~e ba~e~ vol~ge would~be ehoseq ~give ~ ~~ate vol~ge
feedback voltage and a reference voltage, as shown in Figure 19.6. OP1 and
IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS 435
COMP1 work as a backup voltage regulator in case the intermediate voltage drops
below the regulated voltage which will be detected by COMP1 and ~ mrri disable
OP2. As the current is not being supplied by OP2, the voltage at the inverting
terminal of OP1will drop to slightly belowreference voltage, and OP1will turn on
and regulate the voltage, taking the supply directly from the battery terminals.
PI P2 P3
MASTER ~
CLOCK
(a)
MASTER
CLOCK
V BATTERY
~ R BATTERY
BAbY
DETECTOR
....
CS~’b.EDI ~cos~v~ I
VOLTAGE
DOWN
CONVERTER
RGAP~
1
MASTER
CLOCK
Figure 19.6 Regulated power supply system for the digital circuit using a voltage down
converter, voltage regulator, and baeleap regulator. From Pless and Ryan, (1989).
used for external defibrillators. Thus, the exponential decay (til0 is morerapid.
solid-sate switch closes at 4-8 msto dumpthe.. charge. This truncates the waveform
to yield a higher success rate than is the exponential is allowed to continue. The
leads maybe interchanged half way through to yield the single pathway biphasic
truncated exponential waveform, which results in lower energy-requirements. The
leads maybe switched to a different pair of electrodes half waythrough to yield the
sequential biphasic truncated exponential waveform.
Monophasic ’~ Biphasic
truncated truncated
exponential exponential
19.6.2 Capacitorcircuits
While manyICDs use a single capacitor, some use two ~apacitors. An. efficient
capacitor-charging technique was developed based on flyback pritleiple~, where
energy was stored in an inductor and transferred to the capacitors .duriog a short
flyback pulse. Figure 19.8 showsthe circuit to charge the capacitors and generate
the shock pulse. The inductor was .formed by the primary of Tl.and the Sefondary
is split into two coils having advantage of charging each of the series-cormefted
energy storage capacitors to the same voltage, while providing a vol~ge doubler
action requiring fewer turns on each secondary. Reducing the~turns on:~each
secondary also has the advantage of reducing the energy lost ~ repeated~l~ycharging
and discharging the stray capacitance of each secondary.
Battery
Voltage
DI Q4:
R2 D2
T1
Patient
Terminals
The capacitors are discharged into the patient or the test load by silicon-
controlled rectifiers (SCRs) Q2,:Q3, and Q4 while D2 was added to~proteetthe
circuit from external defibrillation pulses. Oneinteresting feature about theeireuit
is the use of R1 and D1, whichconstantly supply battery voltage to the electrolytic
capacitors, as opposed to no voltage in the idle state, to reduce the effect of
deforming overlong periods of time when the c~pacitorsare not~i~ar~ed. Most
ICDsautomatically fully charge and discharge the~ap~cit0rs everyfew n~onths, i~
438 DESIGN OF CARDIAC PA( ~EMAKERS
Capacitors used in the implantable defibrillators have to be chosen for high energy
density. Generally, a capacitor consists of two electrically conductive layers with
dielectric or insdator in between. Manydifferent constructions and materials are
used for makingcapacitors, but, for an energy storage capacitor, there were two
technologies to choose from: the electrostatic type, which Uses film technology,
and the electrolytic style. Becausethe capacitance values are proportional to the
surface area, film capacitors optimized for high capacitance per volumeare often
the metalized type wherea very thin f’dm of metal is directly applied to a very thin
film of dielectric. The layers are then wrappedaround into a cylindrical structure
containing as muchsurface interface as possible. These film capacitors can be built
to accurate tolerance and are very reliable. Unfortunately, it is-difficult to achieve
high capacitance values at around 300 V as they have high-energy density only at
high voltage above 2500 V (Kolenik 1993).
In an~aluminumelectrolytic capacitor, a very thin layer of aluminumoxide.is
formed on the surface of an aluminumanode by an electrolyticprocess, where it
acts as a dielectric. This dielectric can be madevery ’thin and has high dielectric
strength. By etching or roughing the surface of the anode, the contact surface area
of the anode:can be multiplied’ manytimes: The film is formed by oxidation and
the thickness of the film is proportional to the oxidation voltage.
The forming process is important as it determines the maximal operating
voltage that can be applied to the capacitors. If a higher operating voltage than the
forming voltage is applied, high leakage current will flow. Deforming is one
problem associated with this oxide layer. With the absence of voltage across the
terminals of the capacitor, the thickness of the aluminumoxide layer is slowly
reduced due to normal chemical reactions within the capacitors. If the capacitors
were formed using a lower forming voltage, wheh higher voltage is applied,
leakage current increases and performancedifferences can he noticeable as it takes
longer to charge up the capacitors to ~he same voltage level. To solve these
deformation problems, the e~ly AICDfr~om~CPIused battery voltage to constantly
charge the capacitors and stop the deformation process. Another technique, as in
the Cadence VT-100 from:Ventritex uses automatic capacitor maintenance in
whichthe microprocessorwill charge the capacitors to a factory preset voltage and
discharge it through an internal 19ad. Figure 19~9 displays someICDdevices from
CPI and Medtronic with their d~ensions and volume, which are primarily due to
the size of the capacitors and battery, capacitance value, and energy output.
Figure 19.10 shows that, after a thoracotomy, patch electrodes maybe sutured to
the parietal pericardium or directly to the epicardial surface. These are madeof
titanium screen and maybe backed with silicone rubber or polyurethane. Patch-to-
patch defibrillation requires about:~800V. Patch:to-patch pacing could be achieved
with about 5 V, but the large area~elec~odes require larger pacing currents.
Therefore, separate small-area electrodes are Used for pacing for reliable sensing
and to reduce power consumption.~
~ A~oraebtomy
Figure"lg.~0 throughthe upper~ dashedline permitsi~iaeemeh~t
’ of twi5 pa{~h
el~ctrbde;sfor pulses~andtw0screw-inelectrodes for sensinga~d~p~in~elowerdashe~lline
incision pen~its 91aeementof
Wepulse generator, in ~ abd~en.
in the
440 DESIGN OF CARDIAC PACEMAKERS
19.8 ~~ DETECTION
tuency
,gramming programming wand can be
442 DESIGN OF CARDIA~C PACEMAKERS
ethylene oxide gas sterilized for use in the operating room, The programmeris
menu driven with parameter values selected from overlays that display the
available range of selections.
tqoclel V-100
Pulm ~eneratof
~
1~9’.9.2 Bradyc~dia pacing o~y ~ ~ ~, ~..,,
’niode
be
IMPLANTABLE CARDIOVERTER-DEFIBRH~LATORS 443
19.9,6Fibrillationidetection ¯ ~ ~
; cardioverter-
of
during
detection ~the
If a rate less
immediately
and no therapy is only when the device has
chargedto deliver a shock, not whenantitachycardia pacing therapy is initiated.
19.11 REFERENCES
Frame,R., Brodman,R., Furman,S., Gross, J., Kim,S. G., Ferdck, K., Roth, J., Hollinger, I.,
and Fisher, L D. 1993. Long-termstability of defibrillation thresholds with intrapericardial
defibrillator patches. PACE,16: 208-212.
Frumin, H., Goodman,G. R., and Pleatman, M. 1993. ICD implantation via thoracoscopy
without the need for stemotomyor thoracotomy. PACE,16: 257-260.
Gordon, T. and Kannel, W. B. 1971. Premature mortality from coronary heart disease: The
Framingham study. J. Am.Med. Assoc. 215: 1617-1625.
Haaser, R. G., Kurschinski, D. T., McVeiglLK., Thomas~A., anffMower,M. M. 1993. Clinical
result with nonthoracotomyICD. PACE,16:141-148.
Kelly, P. A., Mann,D. E., Dande,R. S., and Reiter, M. J. 1994. Oversensingduring ventrieular
pacingin patients with a third-generation implantablecardioverter-defibrillator. J. Am.Coll.
Cardiol., 23: 1531-1534.
Kolenik, S. A., Langer, A. A., Heilman, M. S., and Staewen, W. S. 1993. Engineering
considerations in the developmentof the automatic implantable cardioverter defibrillator.
Prog. Cardim, asc. Dis., XXXVI: 115-136.
Marchlinski, F. E., Gottlieb, C. D., Sarter, B., Finkle, J., Hook, B., Callans, D., and
Schwartzman,D. 1993. ICDdata storage: Value in arrhythmia management,PACE,16: 527-
53~4.
- Nacearelli, G.V., and VeltH~E. P. (eds) 1993. implantabIecardioverter-defibrillators. Boston:
Blackwell ScientifiC. ¯
Neuzner, J:, Pitschner, H.F., Huth, C:-; and SChlepper,M, 1994. Effect of biphasic waveform
pulse on endoeardial defibrillation effic~y ~in humans,PACE,~ 17: 207-212.
Pless, B. and Ryan, J. G. 198~. powersupply down-conv,~rsion, regulation and low battery
" dete~fion S~teii~.US patentS,g68,908.
Pless, B. arid Ryan, J. G. 1989. Power supply:down-conversion,regulati6n ~and low battery
~.
-itetecfion systerr~ USpatent 4,952,864
Hess, B, Sweeney,M., and Winkle,R.: 1989. Apparatusfor protecting the heart with protected
pacer. USpatent 4,827,936.
Ple~s~ B., Swe.~ey~M.~W~e, R. i989. Apparatfig for protecting the heart with protected pacer.
us patent 5;115,807.:
Saksena, S, .~t ~al. 1993,~¯D~fibrillation thresholds and pefiop~rative meatyassociated with
endoeardialan~¢pieardialdefibrillation lea~ ,syste ~ms
¯ ,, Singe#,i. ~ed~)1994~ImplahtableCardioverter-defibrillator.Arm~onk, NY:Futura Publishing.
Troup,P. J. 1989. Implantablecardioverters and defibrillators. Curr. Probl. Cardiol., XlV:679"
815.
Ventritex. 1993a. Cadence;Tiered therapy defibrillator system; V-IO0Series pulse generator
and programmer,Sunnyvale, CA.
Ventritex. 1993b. Using the Cadenceprogrammer,Sunnyv/de, CA.
Walleott, G. P,, Waleott, K. T., Knisley, S. B., Zhou, X., and Ideker, R. E. 1994. Mechanisms of
defibrillation for monophasicand biphasie waveforms.PACE,17: 478-498.
Winkle, R. A., Mead,R. H., Rudder,M. A., Gaudiani, V., Pless, B., Sweeney,M., and Schmidt,
P. 1989. Improvedlow energy defibrillation efficacy in .man with th~ use of a bipbasic
truncated exponential waveform.Am.Heart J.-, 117: 122-127.
Winter,J., Vester, E. G., Kuhls,S.,, Kantartzis, M., Perings, C., Pausehinger,.M., Straner, B. E.,
and Bir~ks, W. 1993. Defibrillation e~nergyr~quii~ments’withsingle endocardial (End0tak)
lead.PACE,16: 540-546.
19.12-INSTRUCTIONAL OBJECTIVES
UWPacemaker Tester
Mark D. Werkheiser
heartbeat.
A pace~?~aker
tester is a device that !s used t6teSt the operating characteristics of a
pacemaker. Someof these characteristics arepuls~ Width, pulse ampli~de, rate,
and demandsensitivity. Pacemaker testers are used in hospitals to verify the
performance of external pacemakers and to test implantable pacemakers before
they are implanted. Twocommercially available pacemaker :testers aredescfibed
below.
Netech®PMT 100 ¯
The Netech PMT100 is a portable, miO~oprocessor.basedpacemaker tester. It is
capable of testing invasive and transthoraoc pacemakers. The PMT100 measures
and displays amplitude, rate, energy, and pulse width of ~the pacing pulse as well
as demandsensitivity tests, refractory period, ~and susceptibifity to 50/60-Hz
intefferen6e~ " ¯ ¯
The UWpacemaker tester is different from the two pacemaker testers described
above in that the user can develop test signals and observe howthe pacemaker
behaves while sensing the signal. There are two aspects of the pacemaker tester
that allow for the observation-of a pacemaker’s behavior. The first is the
production of the pacemakertest signal, and the second is the detection of the
pacemaker pulses. Figure 20.1 shows a block diagram of the UWpacemaker
tester.
Pacemaker
DAC
Ventricular
Electrogram ECG
The sensing circuitry in a pacemaker does not look for an actual QRS
complex, it si~mply senses whenthe signal crosses a voltage threshold that can be
specified by external programming. The QRScomplex is the only part of the
ele6trophysiological signal that eXceedsthis threshold, so the other parts of the
signal, such as the P or T wave, do not have to be incorporated into the test
signal. The UWpacemaker tester uses triangle pulse waveformsto simulate the
QRSsignal and thus the ventricular electrogram. Figure 20.3 showsthe triangles
used in the model. The parameters of the triangles (width of 100 ms and
amplitude of 5 mV)are taken from the ventricular electrogram in Figure 20.2.
20.2.1 Software
The software aspect of the UWpacemaker tester is written in C code. VAXLAB
software is also used to communicate with the hardware (ADCand DAC), and
MOTIFgraphics environment is used to display the test signal and observe the
pacemaker’s behavior.
The user develops a 30-8 signal by specifying the location of all the triangles
and the ,time beiween each of the triangles. This is done by typing a "Q" for
triangle and typing the time interval (in seconds) between each of the triangles.
This is repeated until a 30-s signal has been developed. Figure 20.4 shows an
example of howthe signal is developed. The triangle data points were developed
using Matlab and are stored on the VAX’shard drive. The 100-ms width of the
triangles has to be accounted for by the user whendeveloping the 30-s test signal.
After the signal is developed, the user has the option of saving the signal. The
user also has the option of loading a predeveloped signal that.is saved, on disk
instead of designing a new signal. Future improvements made to the UW
pacemakertester will incorporate a graphical user interface to develop, load, and
save test signals.
5 2s
Figure20.4 Asampletest signal producedwhenthe user types Q, 1.0, Q,, 1.5, Q, 2, Q.
20.2.1 Hardware
After the signal is:put i~n ~e array, i~ is loaded into, four 16-kilobyte data buffers.
The,signal is,sent~ou t through a Digital AAVli-D DAC~,a~ a rate of 1092
samples/s.ThiSrate corresponds,.to, 30 s 0~, Signal for the 32,,7687elementbuffer,
The output signal has a magnitudeof aboilt ~ 5 V,, peak,~ a~ndth~ pacemakerneeds a
Signal of about 5 mVto sense, so the:signal cqming ~t ~oftile DACi s oltage
divided~-down from 5 V to 5 inV, The 5~mVsignal is then s~nsed by the
pacemakerelectrodes, Figure 20.5 is a .block diagram of the Si~iai deveiopment
aspect of the UWpacemakertester.
Whilethe DACis sending out the test signal, the pacemaker responds:by putsing
dependingon the signal it receives. These pulses must be detected so the user can
452 DESIGN OF CARDIAC PACEMAKERS
observe whenthe pulses occur with respect to the test signal. This is done by
sampling the pacemaker electrode voltage using a DECADQ-32ADC.
5V 5 mV
VAX
20.3.1 Software
The sol, are sa~_ples the pacemaker electrode voltage and uses the samples to
determine whenthe pacing pulses occur with respect to the test signal. First, the
pacing electrode voltage is sampledat the samerate as the test signal is being sent
out by the DAC.This is accomplished by using the ADQ-32internal clock as a
source for both the ADQ-32and the AAVll~D.Since the ’clock rates are the
same for the ADCand the DAC,the DACbegins sending out the test signal as
soon as the ADCbegins to sample the pacing electrode voltage, and each sample
from the ADCcorresponds to ~the same’ point in time a~ the .test signal being’sent
out by the DAC.So after the 39-s test signal is sent out, there is a 32,768-element
array coqtaining ~e test signal as well as a 32,7687element array containing the
sampled pulse dat~.Next, e~ch element of ~the sainpled pulse .data array is
compare~, to a threshold th~it iS detgrmined by the pacing puls~ amplit6de.
Finally; if the eleraent has’an amplitude thai’is greater than’ the threshold, the
corresponding element inthe test,signal arrgy~ is replaced by a flag denoting a
pulse.
20.3.2 Hardware
0.2-mspulsewidth
_ ~,/’5-V amplitude
ADC VAX
This section showsthe results of several tests that were developed and run on the
Medtronic Legend pacemaker. These results come directly from the test signal
array with pacing flags as described in section 20.3A, The pacemakerwas put in
ventricular sensing modeand programmedfor a rate of 70 bpm and a refractory
period of 425 ms.
The UWpacemaker tester sends a test signal to the pacemaker and records
whenthe pacemaker pulses. In an actual heart, a ventricular contraction would
result from the pacemaker pulses and a QRScomplex would result. The 30-s test
signal cannot be changed while being sensed by the pacemaker, so any changes in
the test signal, such as a QRSstimulated by a pacing pulse, Cannot be accounted
for. This open-loop characteristic of the UWpacemaker tester brings up three
timing cases that must be observed:
1) pulse-triangle interval > pacemaker refractory, heart
refractory: This case is accounted for by the UWpacemaker tester. If the time
interval between the pacing pulse and the next triangle producedby the tester is
greater than the pacemakerand heart refractory period, no alteration of the test
signal is needed.Figure 20.9 illustrates this timing case.
2) heart refractory < pulse-triangle interval < pacemaker
refractory: This case is also accounted for by the UWpacemaker tester. The
pacemaker pulse would produce a natural QRS, andthe next triangle wave
producedby the tester can occur because it is not in the QRS’srefractory period.
If the time interval between .the pacing pulse and the next triangle produced by
the tester is less than the pacemakerrefractory period, the triangle will not be
sensed by the pacemaker. Figure 20.10 illustrates this ~ng ,case.
¯ 3) pacemaker refractory < pulse-QRS interval ~ heart refractory:
This cage is .not ~/ccounted fOr by the ~UW pacemakertester. A pacing pulse could
occur that~ would produce .a QRS~and the next ffi~ngle .prod-u~ed by the tester
could Occur within the QR~s~efractory period. None~f ~e tests de’~eloped and
run on ’th~ pacemaker allow ~the third timing Case to occur. Figure 20.il
illustrates this timingcase.
454 DESIGN OF CARDIAC PACEMAKERS
I Start ADCclock I
Sendout test
signal data point
and sample
pacemakerlead
"~YES
I i=0
I i=i÷ I
~YES
[i] =flag
Saveoutput to disk
andplot to screen
Figure 20.7 A block diagramof the software used in the UWpacemakertester. First, the user
makesor loads a test signal. Next, the signal is loaded into the DAC data buffers. Next, the ADC
clock is started, whichstarts the electrode voltage Samplingand output of the test signal fromthe
DAC.After the ADC buffer is full, every elementof the buffer is checkedagainst a threshold. If
the elementis abovethe threshold, the correspondingelement~inthe.test signal array is flagged.
Finally, the test signal array containingthe pulse flags is savedto disk and plotted onthe screen,
The following diagrams (Figures 20.12 through 20.15) show the results
the tests run on the pacemaker. The top curve in all of the diagrams shows the
triangles produced by the’tester. The middle curve in all of the diagrams shows
the pacemaker pulses sensed by the tester. The bottom curves show the what the
final :results would look like: the triangles produced by the tester (solid lines) and
the QRSs that would be pr6duced by the pacing pulses (dashed lines)~ Remember
that ~theQRSs that would be stimulated by the pacing pulses are not produced by
the UWpacemaker tester.
UW PACEMAKERTESTER ~ :~:~ 455
This code makes the test signal. The user has already entered
the specifications (Qs and time interval s Between the Qs) into
an array called signal_spec. The triangle data has been loaded
into a ll4-element array called triangle. The test signal is
put into a 32,768-element array called test_signal. */
count = 0;
for(i=0; i<300; i++)
{
if (signal_spec[i] == ’done’) break;
if (signal_spec[i] == ’Q’)
/* put triangle data into next 114 elements of test_signal */
current_count = count;
for (j=0; j<l14; j++)
{
count++;
test_signal[current_count + j] = triangle[j];
else
for (j=0; j<((int) (signal_spec[i]*1092));
{
test_signal[coUnt] = 0.0;
count++;
}
count = 0;
for (i=0; i<4; i++)
for (j=0; j<8192; j++)
{
/* convert test_signal to short integers and scale for DACo */
DA_buffer[i][j] = (short) (204.8 * test_signal[count]);
count++;
}
/* This section of code checks each pacing pulse sample against
threshold (pulse_threshdld) ~,. -I~flthe ~ples is greater than the
threshold, the corresponding element in the test_signal array
is set equal to a flag (pulse_flag)~
Figure20.8C code~rmakingthetestsignMandsensing
~epacingpulses.
pulse-triangle
interval
Figure 20,9 Timingcase 1. The time interval betweenthe pacing pulse and the next triangle
producedby the tester is greater than the pacemaker
(lighter shadedrectangle) and heart refractory
period (darker shadedrectangle).
Pulse-~r~angle
in~ewal
Figure20.10 Timingcase 2. The time interval betweenthe pacing pulse and the next triangle
producedby the tester is less the pacemaker
refractory period (lighter shadedrectangle).
Pulse-triangle
interval
Figure 20.11 Timingcase 3. The time interval betweenthe pacing pulse and the next triangle
producedby the tester is less than the heart’s natural refractory period (darkershadedrectangle).
Figure 20.12 shows the results of the test while in inhibit mode. Two triangles
are produced by the tester 0.85 s apart (70 bpm). The pacemaker senses these
triangles and is inhibited from pulsing. After a third triangle is not sensed for
0~85 S, the pacemaker pulses. The bottomcurve shows that the pacemaker would
cause a steady heart rate of 70 bpm.
UW PACEMAKER TESTER 457
Patient
simulator
waveform
I
I
I
I
I
Sensed I
pacing I
I
pulses
0.85 s ~
Results
Figure 20.13 shows the results of the test while in trigger mode. A
pacemakerpulse is triggered by the first two triangles in the test signal. After a
third triangle is not sensed for 0.85 s (70 bpm), the pacemaker pulses. The
bottom eurve shows that the pacemakerwill cause a steady heart rate of 70 bpm.
Patient
A
simulator
waveform I
I
I
I
I
Sensed l~,
-pacing
pulses
0.85s 0.85s I
I
I
I
Results
Missed beat
The next t~st on the pacemakerwas to observe its behavior whenthe input signal
missed a beat. Figure 20.14 shows the resilltS of the miss6d beat test whi!e in
triggered mode. The pacemaker is triggered off the first two triangles. Afier a
third is. not sensed for 0.85 s, the pacemakerpulses. Next, the third triangle is
sensed.witl!in 0.85 s dter the pulse and the pacemaker is triggered. The bottom
curve :shows t.hat the pacemakerproduceda steady heart rate of 70 bpm.
458 DESIGN OF CARDIAC PACEMAKERS
Patient
simulator
waveform
Sensed
pacing
pulses
"--’-
0.85 ’~-
s
It
!
Results t
Figure 20.14 Representation of a missed beat. The top curve shows that two triangles are
produced by the simulator 0.85 s apart. The next triangle produced by the simulator does not occur
until 1.7 s later, representing amissed heat, The pacemakertriggers off the first two triangles and
produces a pulse after a third triangle is not sensed for 0.85 s. The pacemakertriggers off the third
triangle producedby the simulator.
Refractory period
The last test performed on the pacemaker was to observe its refractory period.
Figure 20.15 shows the results of the refractory period test while the pacemaker
is in trigger mode. The first triangle produced by the simulator is triggered on
by the pacemaker. The next triangle occurs 0.85 s later and is also triggered on
by the pacemaker. This pacing pulse starts the pacemaker 425-ms refractory
period. The next triangle produced by the simulator occurs within the refractory
period, so it is not sensed. The next pacing pulse occurs 0.85 s after the second
pacing pulse because no triangle wavehad been sensed.
Patient
simulator
waveform
425-mspacemaker
refractorypedod
Sensed
pacing
pulses ~\\\\",,\N1
0.85 s 0.85 s
I
!
lit
Results
Figure 20.15 Testing the refractory period of the pacemaker. The third triangle is not sensed by
the pacemakerbecause it occurs within the pacemaker425-msrefractory period. So the next pacing
pulse occurs 0.85 S after the Secondtriangle producedby the simulator.
The four tests shownabove are just a few that can be performed using the
patient simulator. The flexibility of the simulator allows manydifferent heart
signals to be developed and sensed by a pacemaker. Further study of different
UW PACEMAKER TESTER 459
20.5 REFERENCES
EnvironicsAdaptiveTechnology PaceAlyzer~
Malik,M., andCamm, A. J. 1988.Computermodelingof cardiac rhythmdisturbancesand heart-
pacemaker interaction. Pace,11: 2101-2108.
Malik, M., Cochrane,T.; Davies,D. W., andCamm, A. J. 1987.Clinically relevant computer
modelof cardiac rhythmandpacemaker/hem interaction. Med.Biol. Eng. Comput.,25: 504-
512.
Malik,M., Nathan,A., and Carom,A. J. 1985Computer simulationof dual chamberpacemaker
algorithms usinga realistic heart model:PacingClin. Electrophysiol.,
8: 579-588.
® PMT
Netech 100, 60 BethpageDrive, Hicksville, NY11801.
20.1 Explain the difference betweena the two commerciallyavailable pacemaker testers
describedin section20.1.1 andthe UW pacemaker tester.
20.2 Explainwhytriangle pulsewaveformsare adequatein modeling the electrogram
signal.
20.3 Calculatethe minimal pulsewidthof the one-shotoutputthat allowsat least twopointsin
the pulseto be sampled at a rate of 1092samples/s.
20.4 Explain howthe UWpacemakertester determineswhenthe pacing pulses occur with
respectto thetest signal.
20.5 Sketchthe test signalthat wouldbe developed if the usertypedQ,1.0, Q,2.0, Q,1.5, Q:
20~6 Explainthe "open-loop" characteristicof the UW pacemaker tester.
20~7 Explainwhythe UW pacemaker tester can accountfor the situationin Figure20.9.
20.8 Explainwhythe UW pacemakertester canaccountfor the situation in Figure20.10.
20.9 Explainwhythe UW pacemakertester cannotaccountfor the situation in Figure20.11.
20.10 Describeinhibit modeusingFigure20.12.
20.11 Describe trigger mode ~,s.ingFigure20.13.
20.12 Describethe pacemaker s behaviorwhenthe test signal missesa beat usingFigure20.14.
20:11 Explainwhythe pacemaker (whilein trigger mode)doesnot trigger onthe third triangle
Figure20.15.
Appendix
Microprocessor-based pacemakerdesign
Surekha Palreddy
This pacemakerdetects the onset of atrial depolarizations by detecting the P waveand the
subsequentventricular depolarization by identifying the Q wave.The microprocessorprovides the
pacemakercomputationalabilities and control. Otherforms of analog or digital circuitry can be
used in place of the microprocessor.Amicroprocessoris preferred for itsminiature size and its
flexibility, both of whichare of critical importancein the implantablesystem(Figure A. 1).
microprocessoris designed with input--output ports connecte~iin a conventional manner~;ia a
bidirectional bus to memory,an AVtimer, and a pacing timer. The timers are external to the
microprocessorand are conventionalup or downcountersof the type that are initially loadedwith a
count value and count up or downfrom the value and output a roll-over bit uponcompletingthe
programmed count. The initial count values are loaded into the timers and the roll-over bits are
output to the microprocessor.
i I generator
AV Atria[
[ interval I~ Microprocessor ~ .L~I "J
i~DatI
sense
I timer I I
I / Ventricular Part
sense
I Pacing. a bus
J interva~ ~ Ventricular
I timer / stimulus
|
.~ generator
Figure A.2 and Figure A.3 showthe flowcharts of the AVrate-responsive algorithm (adapted from
Baker, 1988). Whenthe microprocessor is alerted from a sleep state to execute the control
algorithm, it tests to determineif a P wavehas beendetected. If a P wavehas beendetected, the
AVinterval timer is loaded with a maxAV intervall (programmed)value and enabled to start
counting. The pacing interval timer is also loaded with the current pacing interval value (from
RAM) and is enabled to start counting. A paced/sensedflag is cleared (’0’) to indicate that
spontaneousatrial activity wassensed. The algorithmthen puts the microprocessorto sleep until
another wakeupevent is encountered.
Wakeup
No
/
pacing pulse Initiate ventricular
interval time
Loadatrial pacing
pacing pulse
Loadand start AV
interval timer
and start pacingi~] Set paced/
interval timer I~l sensed flag
I
|
Figure A.2 The AVrate-responsive algorithm. The detection of P-wave, Q-waveand time=out
si.gnals from the AVinterval timer and pacer interval timer serve as wakeupevents to the
rmcroprocessor(Baker, 1988).
~ I ..
I Currentinterval = ~1oi~_urr_em ~
lCurrentinterval - Increment
b
= Current
pacingintervalI
I ÷oo, I
Figure A.3 The pace routine to calculate the new pacing interval, based on the recorded AV
interval and the programmedparameters. TAVrefers to the measured AVinterval (Baker, 1988).
If no P wave or Q wave has been detected and the AVinterval has not timed out, the
microprocessor tests to determine if the pacing interval timer has expired. If the pacing interval
timer has timed out, without a P wavebeing detected, the control initiates pacing of the atrium by
loading atrial pulse,parameters into the atrial pulse generator. The controt algorithm then loads the
AVinterval,timer with max AVinterval2 and the pacing interval timerwith the current pacing
interval. The paced/sensed flag is set (,1") to indicate that the atrium has been paced before
entedngthe sleep state.
Figure A.5 shows the various programmableparameters used inthe algorithm for calculating
a newpacing interval based on the, AVinterval.
The microprocessor entersthe wakeup state only when one of the above four conditions are
detected to perform the appropriate functions before going to the sleep state again.
464 DESIGN OF CA~IAC PACEMAKERS
MaxAVintervall MaxAVinterval2
Maxinterval
(1 s)
Pacing
interval
Mininterval
(0.5s)
MeasuredAVinterval
Figure A.4 The rate response function of the AVrate-responsive pacemaker.Pacing rate is
calculated as a function of the patient’s measuredAVinterval. Twodifferent linear functions axe
indicated for rate calculation basedon whetherthe atriumis pacedor sensedin the previouscardiac
cycle. This compensates for the differences in conductiondelay whenthe atrium is pacedand When
it is sensed.Linearfunctionsare preferredfor ease of calculation. Mininterval (0.5 s) corresponds
to 120bpmand Maxinterval (1 s) correspondsto 60 bpm(Baker, 1988).
Fig.ure A.5 The parameters used in the AVrate-responsive algorithm for calculating a new
pacing interval based on AVinterval and programmed
parameters.
The difference betweenthe sensed and paced linear functions (slopes and offsets), which
mapthe measuredAVinterval to a new pacing interval is to compensatefor the effects of
spontaneousand inducedcardiac activity on the newpacinginterval. Whenthe atrium is paced, the
exact onset of the P waveis known,whereasthere is a delay associated with P-wavedetection
whenthe atrium is sensed. To avoid these dine delays whichinfluence the calculation of the new
pacinginterval, the offset andslope of the functionsare adjustedaccordingly.
If the newcalculatedpacinginterval is less than the current pacinginterval, the current pacing
interval is incrementedby a prefuted amount"increment".If the incrementedcurrent pacinginterval
is greater than the newpacinginterval, the current pacing,interval is set to the newpacinginterval.
If the current pacinginterval is greater than the newpac,nginterval, the current pacinginterval is
decremented by a prefixed amount.If the decremented value is less than the newpacing interval, it
is set equal to the newpacing interval. Thishelps to avoidthe rapid increasein pacingrate due to
increased catecholaminewith the onset of exercise. This algorithm provides a smoothgradual
responseas is obtainedin a natural heart.
The algorithm also tests to determineif the paced count exceededa predeterminedmaximal
count. If the pacedcount is greater than or equal to the maximalcount (i.e. the atrium has been
pacedfor the past maximalcount cardiac cycles), the paced count is reset and the current pacing
interval is incrementedby a prefixed amount"delta". This adjustmentenables the emergenceof
spontaneous activity in the atrium, if present.
Finally, before control is returned to the mainprogram,the sensed/pacedflag is set to "2" to
avoid the sensing of a prematureventricular contraction from incorrectly influencing the pacing
rate.
if( Asense == i)
Load MaxAVI -> R7
Loadtimer R7 -> AVtimer
Load Paceint -> R6
Loadtimer R6 -> Pacetimer
LoadImm R1 #0 # (Paceflag = 0)
SLEEP
if( Vsense == i)
Storetimer R7 <- AVtimer
Load MaxAvl -> R6
Loadtimer R6 -> AVtimer
Load Paceaddr -> R5
Jump&Link R5(to) R4(return)
Store R3(Acc) -> newpaceint
SLEEP
if( AVtimerout == i)
Load Vepulwid -> R7
Load Vepulamp -> R6
Enable # (R7 & R6)
Load MaxAVI -> R5
Loadtimer R5 -> AVtimer
SLEEP
if(Pacetimerout == i)
Load Atpulwid -> R7
Load Atpulamp -> R6
Enable # (R7 & R61
466 DESIGN OF CARDIAC PACEMAKERS
Figure A.6 The flowchart of the AVrate responsive algorithmsdiscussed by Baker, 1988 is
converted
into a Pseudocodeto decideuponthe instructionset architecture.
APPENDIX 467
Generalregister operations:
Opcode
field (7 - 3) I Re~ister 9r offset field (2- 0)
Wechosean accumulator-based architecture in whichall the arithmetic and logic operations
are performedbetweenthe register specified and the accumulator.This architecture is preferred by
the designersas the accumulatoris implicitly specified obviatingthe needfor moreinstructional
bits. This wouldallow more"bits" to be used by the opcodeand henceroomfor moreinstructions
in the instructionset.
Theinstruction set is shownin Figure A.7
A.4 3 Datapath
After the instruction set is designed,the hardwarecomponents required to execute the instructions
are identified whichconstitute the datapath. Basedon the complexityof the algorithms.that needto
be implemented,the sizeof storage elements suchas the register file, ROM and RAM is decided.
A register file is a collectionof registers in whichanyregister can be read fromor written to
by specifyingthenumberof the registe~ in the file. Thep~oposeddesign has eight registers, with
three special purposeregisters(0-2) and five ,general pu~s,e registers (3-7), as shown~inFigure
A.8. Register "0" alwaysholds the value "zero. Register ’1 is dedicated to the sensed/pacedflag.
Register "3" is an accumulatoron whichall the arithmetic and logic operations are performed.The
read/write addressport providesa 3-bit addressto identify the register beingread or written into.
The write data port provides 8-bit data to be written into the registers either fromRAM or timers.
Readenablecontrol, whenasserted ~nablesthe register f’de to provide’dataat the read data port.
Writeenable control enables writing of data being providedat the write data port into a register
specified by the read/write address.
The instructions and programmable parametersare stored in 256 bytes of ROM (Readonly
memory).The load/storeinstructions can read data from and write data into the locations of this
memory.The parametersUsedin the AVrate-responsive pacemakerare listed in Figure A.5. The
intermediate parameters are stored in RAM (Randomaccess memory).
Twotimers are required to implementan AVrate-responsive pacemaker.One is used to
measurethe AVinterval (~1"1) and the other for the pacing interval (T2). The timers are read’and
written into just as any other memory location. These timers are provided with read and write
enablecontrols.
468 DESIGN OF CARDIAC,PACEMAKERS
Instruction Function
LDI ### LoadAccumulatorwith 3-bit immediatevalue
LD (rx) LoadAccumulatorwith ROM location (rx)
LDR3I ##### LoadR3 with 5-bit immediatevalue
LDFLG Loadwakeupcondition inputs into R2
LDACC(rx) LoadAccumulatorwith RAM location (rx)
LDT1 rx LoadTimer#1 with value in rx
LDT2 rx LoadTimer#2 with value in rx:
ST (~) Store Accumulatorwith ROMlocation specified in rx
STT1 rx Store Timer#1 in rx
STT2 rx Store Timer#2 in rx
ALU
CP rx CopyAccumulatorto rx
Addrx to accumulator
Subtract (rx) from accumulator
Incrementrx
DEC rx Decrementrx
SHR rx Shift accumulatorrisht by amountin rx
SHL Shift accumulatorleft by amountin rx
NOP Nooperation
Control ’
BGE rx Branchif (ace) > (rx)
BLT rx Branchif (ace) < (rx)
BEQ rx Branchif (acc) ==(rx)
JMP rx Jumpto memory location in rx
Special
SLEEP Put the microprocessorto sleep
Read/write= Register
file
RAM
Intermediate
register
parameters
ROM
Instructions
Curpaceinterval Acc
Newpaceinterval
MaxAvl
MaxAv2
Paceinterval =~Write ,
PaceAddress
VPulWidth Read
VpulAmp
Apul Amp
Apul Width
Delta
Increment
Minlnt AVtimer
SlopeP
OffsetP Pacetimer
Slopes
Offsets Read Wnte
PACount Clock enable enable
Figure A.8. A schematicdiagramof the register file; timers and ROM/RAM. Three registers of
the register file’are dedicatedfor special purp0se s and the rest are general purposeregisters. An
instruction register is usedto load the instruction fetchedfrommemory.
APPENDIX 469
Control tells the datapath and Input/Output devices what to do according to the wishes of the
instructions of the program.The methodused to specify the multicycle control is afinite state
machine.A finite state machineconsists of a set of states and directions on howto changestates.
Eachstate specifies a set of controls that needto be asserted whenthe machineis in that state. An
exampleof a state machinefor RISCinstructions is shownin Chapter10. The control is designed
to be implementedusing a programmable logic array (PLA);whichenables the forcing of any set
of output conditions in responseto any set of defined input conditions. The PLAis programmed to
set the required controls of the datapath elementsshownin Figure A.9basedon the opcodeof the
instruction being executed. Since each instruction is executed in several clock cycles, a state
register is used as part of the control to indicate the next state to be executed.Thestate machine
changesstates basedon the inputs from the opcodeof the instruction and the state register. For
more information on the design of control, the book Computerorganization & design: The
hardware/software interface serves as anexcellent reference.
Oncethe datapath and control are designed, the required algorithm that needs tobe executed is
loaded into the memoryof the microprocessor.Eachinstruction fromthe memory is fetched one at
a tame and executedby setting the appropriate controls. Theloaded programscould be modifiedor
updated with newerversions by the use of telemetry, The simple instruction architecture chosen
makesit easy to implementcomplexalgorithmswith relatively simple hardware.
A.6 REFERENCES