Eschweiler Combi - Service Manual
Eschweiler Combi - Service Manual
Eschweiler Combi - Service Manual
Blood-Gas-Electrolyte-Analyser
Service Manual
Made in Germany
Service Manual
Manual Version
The latest Manual version is: 4.01
Order No.: 60 8 90 10
Copyright © 2010 by
Print: SMe COMBILINE 401.PDF
!
Eschweiler GmbH & Co. KG, Holzkoppelweg 35, 24118 Kiel
Manual History
Manual ANALYSER-Software Revision
Version Version Date
Software Copyrights
The software for the combiline Analyser is the User and Instrument Con-
trol Software for the combiline. The combiline Software is the intellectual
property of Eschweiler GmbH & Co. KG hereafter Eschweiler.
Intellectual property rights shall remain with Eschweiler.
You are entitled to use the combiline Software and the printed accompa-
nying material at your place of work only.
Manufacturer
Eschweiler GmbH & Co. KG
Holzkoppelweg 35
24118 Kiel, Germany
Internet: http://www.eschweiler-kiel.de
E-Mail: info@eschweiler-kiel.de
Service Manual
Consumable solutions and spare parts from other manufactur-
ers used in ESCHWEILER Analysers.
The name ESCHWEILER with its decades of experience in the field of blood
gas analysis also stands for the quality of the spare parts and consumable
solutions. ESCHWEILER sensors and solutions form a functional unit
whose components are optimally adapted to each other. Measuring accuracy
and long life of the sensors are the goal of this system.
For example, special wetting solutions are used which have a decisive
influence on the unconstrained contact between sensor and solution in the
measuring capillary. At the same time suitable preservatives prevent the
growth of micro-organisms in the consumable solutions and keep them from
being introduced into the measuring capillaries of the ESCHWEILER
analysers.
Service Manual
Table of Contents
Safety Issues ................................................................................................................................................ 2-5
Technical Data. ................................................................................................................................................ 6
Blockdiagram. ................................................................................................................................................. 17
Power Supply ..................................................................................................................................................18
Adjustment of the operating voltage for Rollerpumps RP1 and RP2. ................................................................ 21
Adjustment of Thermostate Temperature ......................................................................................................... 23
Keyboard. ....................................................................................................................................................... 51
Please read the Instruction Manual in its entirety prior to operating the
combiline
line. In order to ensure a high level of performance, all warnings and
line
references to technical safety in this Instruction Manual have to be followed.
CAUTION Warns of a risk of biological material like blood or other human or animal
specimen. Risk of infection! Potential infectious area/material!
Biohazard
Introduces rules to be observed.
Electrical safety
- Check that the operating voltage is set correctly before you connect the
device to the main power supply.
- To connect the analyser to the main power supply, use only sockets which
are grounded to avoid the risk of an electrical shock.
- Use only grounded extension power cords. The used power cord has to
correspond to the country specific regulations.
DANGER
- Never intentionally disconnect the grounding contacts.
There is the risk of electrical shock if the protective conductor is inter-
rupted within or outside the analyser, and/or the grounded contacts have
been disconnected from the line.
- Even after an analyser has been switched off, components (e.g. capaci-
tors) can be under voltage as the result of an electrical charge.
- Surfaces (floors, work tables) have not to be moist when you are working
with any electrical device.
- Carry out only the maintenance work and/or the replacement of parts
described in these Instruction Manual.
- Use only replacement fuses of the stated type and with the stated nomi-
nal current. Never use fuses, which have been “repaired”.
- Avoid any direct contact with samples that are potentially infectious or
which may generate other risks to the human body (Aids, Hepatitis etc).
Biohazard In case of direct body contact go and wash the contaminated area
immediately. Use a suitable skin disinfection solution. Ask your doctor for
aid.
- If sample material (blood) or reagent is spilled onto the analyser, wipe it off
immediately and refer to the chapter ! 2 Maintenance and Hygiene.
Operator Qualification
- The analyser should only be operated by trained personnel. Improper
operation of the analyser might cause inaccurate measuring results!
- Never spray or pour cleaning solution directly onto the analyser, that may
negatively impact the analyser’s functions significantly.
CAUTION - Keep the analyser clean and do not spill liquids onto it.
- In case liquids were spilled onto the analyser, immediately absorb liquid
with a suitable cloth.
Recommended Disinfectant
Solutions.
A suitable disinfectant is available by BODE Chemie, Hamburg:
Bacillol®, is recommended for areas at risk of infection, where rapid
contact times and drying times are necessary.
See the website for your nearest representation. Ask for an EC Material
Safety Data Sheet according to EC 91/155.
!3 First-Aid Measures
If a person swallowed any chemical solution may be while servicing,
wash out mouth with water.
If provided person is unconscious, than call a physician.
If inhaled, remove to fresh air. If breathing becomes difficult, than call a phy-
sician.
In case of contact, immediately wash skin with soap and copious amount of
water.
In case of contact with eyes, flush with copious amounts of water for at least
15 minutes.
Assure adequate flushing by separating the eyelids with fingers, than call a
physician.
!4 Repairs
- Repairs to the analyser may only be carried out by trained personnel, and
replacement parts have to comply with the analyser specifications.
Useful hint
In this Manual a bullet "•" draws your attention to an instruction.
Example:
POWER UP COMBILINE
Rev. 2.500
AUTO-CALIBRATION
OF SENSORS OPERATION
1 = RINSE
2 = CALIBRATION *
STAND BY 3 = CALIBRATION BGA * **
ECONOMY
DIALOGUE 4 = CALIBRATION ISE * **
PROGRAM
5 = ACTIVATION HB *
6 = SENSOR PARAMETER *
7= ECONOMY MODE *
8 = DISPLAY MEASURED
Opened Front
Order no.
2 - Thermoprinter GPT-4352-60 60 8 20 90
4 - Dampness Absorber 40 7 10 50
Order no.
12 - Intermediate Vessel 40 7 10 50
13 - Thermoprinter GPT-4352-60 60 8 20 90
14 - Analogboard ESW-ANALOG 60 8 10 xx / 60 8 11 xx
ESW_MVTC 2.0 60 8 10 80
1 - Potentiometer Rollerpump 2
2 - Potentiometer Rollerpump 1
3 - Buzzer
15 -RAM
S204-06
VE2
SR20.17
F200.25
combiline BGA
normally open
stromlos offen
SR20.24 suction-pump
Entlüftungsventil
Absaugpumpe
ventilation valve
F10.19
SR20.16
SP SR20.18
S104-08
R20.28
Rollerpumpe
F200.25
Feuchtefilter
Rollerpump
Reagent-Transport
Zwischengefäß
(peristaltic-pump) SV20.7
Intermediate-Vessel
Dampness Absorber
SR20.19
F200.25
S204-06 SR20.13
RP SR20.23 SR20.20
stromlos offen VE1
SR20.3
Light barrier for
normally open level control
(CL Füllstand)
Entlüftungsventil
R20.8 ventilation valve
R20.8
SR20.12
SR20.12
Ventilblock
Valve-block SR30.14
stromlos offen Kalibrierlösung
35-8.101 (set) normally open 40-8.101 (set) F200.26
SR20.19 Calibration-Solution
ABS
SR20.22
S204-06
F200.25
R20.8
stromlos offen PV
EV stromlos offen normally open
normally open
F200.25
F200.25 S204-06
S204-06
LS
Ref pH HB
Rm100.21
pCO2 pO2
waste bottle
Abfallflasche WASH
VERSCHLAUCHUNG / TUBING REV 1.1 ESCHWEILER GmbH & CO. KG KIEL GERMANY
COMBILINE BGA DATUM: 28.07.2008 BEZ.: SP-COMBILINE BGA+HB.TCD VERSION: 1.1
combiline
SV20.6
combiline
S204-06
VE2
SR20.17
combiline ISE
F200.25
normally open
stromlos offen
SR20.24 suction-pump
Entlüftungsventil
Absaugpumpe
ventilation valve
F10.19
SR20.16
SP SR20.18
S104-08
R20.28
Rollerpumpe
F200.25
Feuchtefilter
Rollerpump
Zwischengefäß
Reagent-Transport
(peristaltic-pump) SV20.7
Intermediate-Vessel
Dampness Absorber
SR20.19
F200.25
S204-06 SR20.13
RP SR20.23 SR20.20
stromlos offen VE1
SR20.3
Light barrier for
normally open level control
(CL Füllstand)
Entlüftungsventil
R20.8 ventilation valve
R20.8
SR20.12
SR20.12
Ventilblock
Valve-block SR30.14
stromlos offen
35-8.101 (set) normally open 35-8.101 (set) F200.26
SR20.19
ABS LV
SR20.22
S204-06 CAL 3
SR20.9
F200.25 CAL 4
sample-port
Probenport
R20.8
stromlos offen PV
EV stromlos offen normally open
normally open
F200.25
F200.25 S204-06
S204-06
LS Ref Cl pH Na Ca Li K E
Rm100.21
Rm100.21
Rm100.21
waste bottle
Abfallflasche WASH 2 CAL 3 CAL 4
Kalibrierlösung
Calibration-Solution
page 13
page 14
SV20.6
S204-06
VE2
SR20.17
F200.25
normally open
stromlos offen
SR20.24 suction-pump
Entlüftungsventil
Absaugpumpe
ventilation valve
F10.19
combiline BGA+E
SR20.16
SP SR20.18
S104-08
R20.28
Rollerpumpe
F200.25
Feuchtefilter
Rollerpump
Zwischengefäß
Reagent-Transport
(peristaltic-pump) SV20.7
Intermediate-Vessel
Dampness Absorber
SR20.19
F200.25
S204-06 SR20.13
SR20.23 SR20.20
RP VE1
stromlos offen
SR20.3
Light barrier for
normally open level control
(CL Füllstand)
Entlüftungsventil
R20.8 ventilation valve
R20.8
SR20.12
SR20.12
Ventilblock
Valve-block SR30.14
stromlos offen
35-8.101 (set) normally open 40-8.101 (set) F200.26
SR20.19
ABS LV Calibration-Solution
SR20.22
S204-06 CAL 3 Kalibrierlösung
SR20.9
R20.8
stromlos offen PV
EV stromlos offen normally open
normally open
F200.25
F200.25 S204-06
S204-06
LS
Ref pH Ca Na K pCO2 pO2 HB
Rm100.30
Rm100.31
Rm100.31
waste bottle
Abfallflasche WASH 2 CAL 3 CAL 4
calibration solution ISE
VERSCHLAUCHUNG / TUBING REV 1.1 ESCHWEILER GmbH & CO. KG KIEL GERMANY
COMBILINE BGA+E DATUM: 28.07.08 BEZ.: SP-COMBILINE BGA+E+HB.TCD VERSION: 1.1
combiline
SV20.6
combiline
S204-06
VE2
SR20.17
F200.25
normally open
stromlos offen
SR20.24 suction-pump
Entlüftungsventil
Absaugpumpe
combiline meta
ventilation valve
F10.19
SR20.16
SP SR20.18
S104-08
R20.28
Rollerpumpe Rollerpumpe
F200.25
Feuchtefilter
Rollerpump Rollerpump
Zwischengefäß
(peristaltic-pump) (peristaltic-pump) SV20.7
Intermediate-Vessel
Dampness Absorber
Reagent-Transport
SR20.19
F200.25
S204-06 SR20.13
SR20.23 SR20.20
RP 2 RP 1 VE1
stromlos offen
SR20.3
Light barrier for
normally open level control
(CL Füllstand)
Entlüftungsventil
R20.8 ventilation valve
R20.8
40-8.102 (set)
SR20.12
SR20.12
Ventilblock
Valve-block SR30.14
F200.26
40-8.102 (set) SR20.19
ABS stromlos offen LV Calibration-Solution
normally open sample-port SR20.22
Probenport CAL 3 Kalibrierlösung
S204-06 1-8.106 SR20.9
F200.25 CAL 4+M BGA 3 BGA 4
BGA 3
BGA 4
EV S204-06
LAC stromlos offen
normally open
F200.25
S204-06
LS
Ref pH Ca Na K pCO2 pO2 HB
Rm100.30
Rm100.31
Rm100.31
META
waste bottle
stromlos offen WASH
normally open Abfallflasche CAL 3 CAL 4+M
2+M
F200.25
S204-06 calibration solution ISE
VERSCHLAUCHUNG / TUBING REV 1.1 ESCHWEILER GmbH & CO. KG KIEL GERMANY
COMBILINE meta BGA+E DATUM: 01.09.2008 BEZ.: SP-COMBILINE II BGA+E HBII.TCD VERSION: 1.1
page 15
Valves
Determination of Valves
VE1 Ventilation-valve, intermediate vessel (one stage pinch valve, norm. open)
VE2 Ventilation-valve, suction pump (one stage pinch valve, normally open)
1
not available in combiline with electrolytes or metabolites
2
not available in combiline BGA
3
available in combiline meta only
combiline
CONVERTER Kontrast
Absaugpumpe Ventile Rollerpumpe 1 Roller-Pump 2
Contrast
Suction-Pump Valves Roller-Pump 1 for metabolites
Beleuchtung
Drucker Füllstands- TASTATUR illumination
Printer kontrolle
+24V
GPT-4352 Level-Control KEYBOARD DISPLAY
100 - 250 V~ CL FÜLLSTAND
50 - 60HZ LMBHT014GC
L1
2 x 250V / 2.5A T CL Tastatur DISP INT SD-CARD
Blockdiagram
M1 M1 IC25
BUZZER
L2 SER-TTL
9600Baud 8N1
M2 M2 IC26
Lüfter 24V DC SD CARD INT
drivers
Treiber
Fan 1.0W
SD Card Reader
V1 M3 M3 IC27 CPU Z80 ADC RS232
+24V +24V
+5V LCD-Converter +5V
LCD
V1
+12V Data-Output 9600Baud 8N1
+5V +5V +24V Printer MUX ADC
PRINT
-12V
+5V
+12V +12V +12V CPUZ80ADC L4
Z80
-12V
NTC
Schaltnetzteil
switching power-supply
PM110-45-2A
-12V -12V
COOL
Peltier-Element
PELTIER
ESW-MVCT LAC GLU
-12V Sensor Sensor
+12V
ANALOG
HB-Enable
LS HB TEMP K Na Ca Cl Li pH pCO pO
2 2
Sensoren / Sensors
page 17
Switching Power-Supply
Type: PM110-45-2A
+5V Supply for Logic (with Protection Diode1N5630A): Z80 ADC, Level Control board CL Füllstand,
LCD Converter
Type: ESW-MVTC
Rollerpump 1
Preparations:
4. The board ESW-MVTC is located at the inner side of the rear wall
4. Connect the voltmeter to the test-points MOTOR (+) and GND1 (ground).
5. Press key „3“ (calibration solution) to run the roller pump 1 with slow speed.
7. Press key „2“ (WASH solution) to run the roller pump 1 with high speed.
9. Connect the voltmeter to the test-points MOTOR_META (+) and GND1 (ground).
10. Press key „8“ (Meta) to run the roller pump 2 with slow speed.
11. Set the voltage at test-point MOTOR_META to 6.2V ±0.3V using potentiometer TRM_SLOW.
12. Press key „7“ (Gas) to run the roller pump 2 with high speed.
13. Set the voltage at test-point MOTOR_META to 8.0V ±0.3V using potentiometer TRM_FAST.
Remark:
The operating voltage for the roller pump has to be checked and adjusted only when an intact, good-as-new
roller pump tube is used.
Temperature-Regulation
Description:
Via the plug connection V1 the low voltage 24 V~/4.5A is supplied. Furthermore the external elements for
heavy-duty heating resistors as well as the temperature sensor (both in thermostats) are connected.
After exchange of one of these components, the temperature, indicated on the display, is adjusted to 37°C
with potentiometer T-CAL. The PIC IC must not be exchanged. Otherwise a re-adjustment of the circuit is
required. For this adjustment special equipment is required.
There should be made no adjustments without the required auxiliaries. Adjustments should not be done
without contacting Eschweiler Service department!
A yellow LED for Temperature indication is located at the analyser’s front, below the reagent-wall. This LED is
an indicator for temperature regulation status.
The more the actual thermostat temperature reaches to the nominal value (37°C), the longer the LED is
“ON”.
The LED is permanently “ON” when the temperature is in a range of ±0.5°C from 37°C.
The LED is blinking when the temperature deviation from 37°C is more than ±0.5°C.
The higher the blinking frequency, the higher is the temperature difference (e.g. Warm-up Phase).
In case of defective NTC-sensor or released temperature safety switch, the LED just blinks very short.
Layout
C
A
Preparations:
3. The board ESW ANALOG is located at the rear-side of the reagent wall
3. Adjust the indicated voltage to 0.700V with the potentiometer POL at pos. A
5. Close the analogue box, the reagent-wall and the frot cover again
Remark:
The polarization voltage is connected to the platinum cathode of the pO2 sensor. It is set to 0.700V at the
factory
1. Activate Testprogram 5, GAS, in the SERVICE menue and remove the plug from the pO2-sensor
2. Adjust the voltage, indicated on the screen, to -1.700V with potentimeter ZERO at pos. C
3. Connect sensor cable to pO2 sensor, stop the Testprogram and start a calibration
I
D
K
F
G
D
4. The atmospheric pressure measured by the internal air-pressure sensor is indicated as follows:
5. Adjust the indicated value to the actual Barometric Pressure with the potentiometer T PS at pos. D
Remark:
A testpoint for measurement of the output voltage of the Barometric Pressure sensor is located at pos. E
3. The board ESW ANALOG is located at the rear side of the reagentwall
The adjustment can be made with a suitable PT100-Transmitter or alternatively with 2 calibrated resistors. The
adjustment is made for the value +20°C with a resistance of 107.80 Ohm and for +40ºC with a resistance of
115.54 Ohm. At the transmitter and the resistors the same lengths of cable must be used as on the built-in
temperature sensor.
Adjustment procedure:
5. Set +20ºC at the PT100-Transmitter and connect it at pin1 and pin3 of plug M2. Alternatively, connect a
precision resistance* of 107.8 Ohm.
6. Set the temperature value in the display to 20.0°C with potentiometer T_ZERO at pos. G
8. Set the temperature value in the display to 40.0ºC with Potentiometer T_GAIN at pos F
It may be necessary to repeat this procedure several times until both values are reached.
The Sample Light Barrier can be adjusted at potentiometer TR_LB at pos. K on the analog board ESW
ANALOG
For Adjustment:
4. Adjust the voltage, indicated for air, with potentiometer TR_LB at pos. K to a value between 2,000 V -
2.300V
4. Press key TEST (WASH solution is filled into the measuring cell). The indicated voltage must be between
0.300V and 0.500V!
Should it not be posssible to reach the a.m. voltage values, an exchange of the Sample Light Barrier or its
capillary might be required. For instructions for LB-exchange please note next page.
Capillary exchange
in Sample Light-Barrier LB
figure 4
Measure method
The Hb-sensor is based on the combination of reflexion (scattered) and Transmission measurement of infrared
light.
The data processing of Hb-values in the range of 6 - 30g/dl is done by the Reflexion-measurement method.
The data processing of Hb-values in the range of 3 - 6g/dl is done by the Transmission-measurement method.
The Hb-measure unit is located on the analog amplifier board ESW-ANALOG 1_01.
- The use of Hb-calibration is not required, only the input of two parameters forrefexion and Transmission
(HB-R and HB-T) is necessary.
- After setting the Combiline into operation, sensor related parameters have to be entered by keyboard. After
that the Hb-sensor is active for measurement. These parameters (HB-R and HB-T) are printed on a white
flag at the Hb-sensor cable.
- The sample volume required for a fully equipped Combiline BGA+E+HB is min. 210 - 240µl. The
sample volume required for a combiline BGA+HB is approx. 160µl. So it is of particular
relevance to use capillaries with adequate volume.
- Please note that this measure method is suitable for measurement of whole-blood only! Quality controls
are not usable!
- Hb will be measured if the sample light-barrier detects a whole-blood sample. Otherwise the programmed
standard Hb-value is displayed (e.g. 15g/dl) and used for the calculation of the acid-base parameters.
- In case of too less sample volume, a “Transmission-voltage” of -4.000V is detected. The displayed Hb-
value will be 0.0g/dl then. This can be observed on a sensor-parameter printout which can be made after a
sample measurement (see attached example of printout)
In order to maintain the measuring accuracy for determination of Hb and to activate the Hb-sensor, an input of
two factors HB-R and HB-T has to be carried out in STAND-BY mode
In case the COMBILINE was switched off for less than 24 hours the Hb-factors are retained in the memory
and the sensor remains ACTIVE.
In case the COMBILINE was switched off for more than 24 hours the Hb-factors have to be entered as
described below, the Hb-sensor is marked with OFF.
The mentioned Hb-factors are printed on a white flag at the Hb-sensor cable (see attached photo)
• Press the 5-key in OPERATION MENU to perform the Hb-activation process, the following dialogue
appears:
HB-SENSOR
HB R : 0.000
HB T :
• Enter the HB-R value given on the white ensign at the Hb-sensor by keypad e.g. 0.970 and press the
Enter-key to confirm.
HB-SENSOR
HB R : 0.970
HB T : 0.000
• Enter the HB-T value given on the white ensign at the Hb-sensor by keypad e.g. 0.517 and press the
Enter-key to confirm.
For quality control of the Hb-Sensor, whole blood specimens whose Hb-concentration have already been
determined with the help of suitable measuring methods might be used (e.g. oximeter).
HB - SENSOR
Hb-sensor type II
Layout
+5V
+5V +5V
41 90 16 64
1 Vcc Vss
81 89 D0 1 2 11 10 11 10
combiline
2 CLK/TRG 0 D0 D0 A0 D0 A0
80 88 D1 /M1 3 4 /WR 12 9 12 9
3 CLK/TRG 1 D1 D1 A1 D1 A1
79 87 D2 A7 5 6 /WAIT 13 8 13 8
4 CLK/TRG 2 D2 D2 A2 D2 A2
78 86 D3 A6 7 8 CLK 15 7 15 7
5 CLK/TRG 3 D3 D3 A3 D3 A3
L1 77 85 D4 A5 9 10 /C0 16 6 16 6
6 ZC/TO 3 D4 D4 A4 D4 A4
76 84 D5 A4 11 12 /RD 17 5 17 5
7 ZC/TO 2 D5 D5 A5 D5 A5
HB INTERFACE
75 83 D6 A3 13 14 /C1 18 4 18 4
8 ZC/TO 1 D6 D6 A6 D6 A6
74 82 D7 A2 15 16 /IORQ 19 3 19 3
9 ZC/TO 0 D7 D7 A7 D7 A7
+5V A1 17 18 D7 25 25
10 +5V A8 +5V A8
33 6 A0 A0 19 20 D6 24 24
Circuitdiagram 1/3
RXC A A0 A9 A9
34 5 A1 D0 21 22 D5 J1 21 J2 21
1 TXC A A1 A10 A10
30 4 A2 D1 23 24 D4 23 23
2 W/RDY A A2 A11 A11
31 3 A3 D2 25 26 D3 32K 1 2 8K 26 2
3 SYNC A A3 Vpp/A15 A12 CE2/A13 A12
32 2 A4 /NMI 27 28 /INT 64K 26 32K 1
4 RXD A A4 A13 A14
35 1 A5 /RES 29 30 IEOUT 27
5 TXD A A5 A14 22 /RD
L2 36 100 A6 31 32 IC4 IC5 OE
6 DTR A A6 22 /RD 27 /WR
37 70 A7 A15 EPROM OE A13 RAM WE
7 RTS A A7 RF L5 20 20
38 98 A8 CE CE1
serielle Schnittstelle
serial interface
8 CTS A A8 EXPANSION 28 14 28 14
39 97 A9 +5V +5V Vcc GND +UB Vcc GND
9 DCD A A9
+5V 96 A10
10 A10
49 95 A11 +5V +UB +5V
RXC B A11
48 94 A12
1 TXC B A12 +5V
52 93 A13
2 W/RDY B A13 20 R3
51 92 A14 IC3
3 SYNC B A14 A4 1 Vcc 12 /C0 100K /RES 12
100nF
100nF
100nF
100nF
10uF Ta
100nF
100nF
100nF
10uF Ta
100nF
100nF
50 91 A15 I0 F0 11
4 RXD B A15 A5 2 13 /C1 13 D
CPU Board CL Z80 ADC12
47 +5V + + I1 F1
5 TXD B A6 3 14
L3 46 99 R5 I2 F2 74HC00
6 DTR B A7 A7 4 15 /M1
45 100K I3 F3
7 RTS B A12 5 16 /C90
C3
C4
C5
C6
C7
C8
C9
C11
44 72 IEI I4 F4 14
C10
C13
C14
8 CTS B IEI A13 6 17 +UB 1
Drucker
printer
43 71 IEO DGnd I5 F5 3
9 DCD B IC1 IEO A14 7 18 2 A
RN1 5*100K I6 IC6 F6 IC3
10 Z84C15 A15 8 19
19 /INT 2 I7 GAL F7 7 74HC00
INT /MREQ 9
63 3 /NMI I8 16V8
NMI /IORQ 11
11 /WAIT 4 I9 ESCHWEILER CombiLine
WAIT GND
DA0 29 10 5 CL Z80ADC Rev. 1.2
PA 0 BUSREQ 10 CL Z80ADC 12a.tcd / 01.02.2006
DA1 28 9 6 /RES
PA 1 RESET Ident-No. 2061
DA2 27
PA 2 +5V
DA3 26 8 /M1 +5V +UB
PA 3 M1
CPU-PLATINE / CPU BOARD CL Z80ADC REV. 1.2 ESCHWEILER GmbH & CO. KG KIEL GERMANY IDENT Nr.: 2061
SCHALTPLAN / CIRCUIT DIAGRAMS DATUM / DATE : 01.02.2006 NAME : CL Z80ADC 12a.TCD BLATT / PAGE : 1/4
page 37
+5V
+5V
+5V
page 38
22 53 61 66 78 20 10 L11 L2
1 Vcc D0 3 Vcc GND 2 D0
72 62 D0 9 11
2 CLK/TRG 0 D0 D1 4 5
73 63 D1 10 12
3 CLK/TRG 1 D1 D2 7 IC10 6
74 64 D2 7 13
4 CLK/TRG 2 D2 D3 8 74HC273 9
75 65 D3 8 14
5 CLK/TRG 3 D3 D4 13 12
L6 47 68 D4 D Q 5 15
6 ZC/TO 3 D4 D5 14 15
48 69 D5 6 16
7 ZC/TO 2 D5 D6 17 16
49 70 D6 3 17
8 ZC/TO 1 D6 D7 18 19 D7 GRAPHIC-DISPLAY
50 71 D7 4 18
9 ZC/TO 0 D7 CL
15 2
10 SEL0 11 1
1 A0 11 19
Circuitdiagram 2/3
A0 /RES /RES
84 A1 10
A1
+5V 83 A2 +5V +5V 16 3
A2
17 82 A3
RXC A A3
16 16 8
1 TXC A
32 80 /M1 3 Vcc GND 2 /WR
2 W/RDY A M1 12 5
31 79 /RD 4 5 /RS
3 SYNC A RD 14 6
18 77 /IORQ 6 IC11 7 C/D
4 RXD A IORQ 13 8
15 76 /RES 11 74HC174 10 /CE
5 TXD A RESET 1 7
L7 30 57 CLK 13 12
6 DTR A CLOCK D Q
29 54 /INT 14 15
7 RTS A INT DISPLAY ADAPTER DISPINT
9 52 IEOUT
8 CTS A IEO
8 51 IEO
9 DCD A IEI
81 /C90 CL
10 CS
SEL1 9 1
60
XTALI
CPU Board CL Z80 ADC12
59
XTALO +5V
56 +5V
CLKOUT +5V
55
10 OSC 20 10
PC7 1
9 26 18 Vcc GND 2 2
PC6 2
8 IC9 25 5 15 5
PC5 19 3
7 Z84C90 PA0 17 14 IC12 6 6 ESCHWEILER CombiLine
PC4 20 4
6 KIO PA1 15 9 74HC244 11 8 CL Z80ADC Rev. 1.2
L8 PC3 21 5 L13 FREI / FREE CL Z80ADC 12b.tcd / 01.02.2006
5 PA2 13 12 8 9
PC2 23 6 Ident-No. 2061
4 PA3 11 7 13 7
+5V PC1 25 7
3 PA4 9 16 4 4
PC0 26 8
2 PA5 7 3 17 3
pull up resistors on
ECO-TAST (4 x 100K)
CPU-PLATINE / CPU BOARD CL Z80ADC REV. 1.2 ESCHWEILER GmbH & CO. KG KIEL GERMANY IDENT Nr.: 2061
SCHALTPLAN / CIRCUIT DIAGRAM DATUM / DATE : 01.02.2006 NAME : CL Z80ADC12b.TCD BLATT / PAGE : 2/4
combiline
+5V
RN7 8*30K
ANALOG-INPUT L4 ANALOG-GND 10 20 RN4 1 8*330 1
D0 3 GND Vcc 9 8 9 RN10 4*4.7K M1
1
3
5
7
9
2
4
6
8
D
11
13
15
17
19
21
23
25
10
12
14
16
18
20
22
24
26
D1 4 2 7 10 2 8 11
1 ISE 1 / 3
D2 7 5 6 11 5 7 12
IC19 2 ISE 2 / 4
combiline
ESCHWEILER CombiLine D3 8 8 5 12 C 13
D 3 ( HB-LAMP )
CL Z80ADC Rev. 1.2 D4 13 7 4 13 14
CL Z80ADC 12c.tcd / 10.01.2006 ILQ1 6 C 4 LAMP - THERMOSTATE
D5 14 IC16 6 3 14 6 15
Ident-No. XXXX 5 PIEP - SAMPLE SOUND
19
20
21
22
23
24
25
26
11
10
9
8
7
6
5
4
D6 17 9 2 15 9 5 16
6 MZ - SAMPLE COUNTER
D7 18 74HC273 6 1 16 D IC25 17
7 VE2
RN11 18
IN1
IN2
IN3
IN4
IN5
IN6
IN7
IN8
IN9
ULN 8 AP - SUCTION PUMP
IN10
IN11
IN12
IN13
IN14
IN15
IN16
Q
5 1 16 RN10 2804
DB0 17 12 SEL4 11 B
A0 IC15 GND 12 2 15 3 4
DB1 16 10
A1 15 3 14 4 3 V+ 9
DB2 15 HI506 1 IC20
Circuitdiagram 3/3
A2 V++ +15V 4 4 13 A
DB3 14 /RES 1
A3 CL 3 5 12
27 ILQ1 B
V-- -15V 16 6 11 7 2
9
EN ANA IN 19 7 10 8 1 GND 10
2 8 9 A
18 28
RN11 4*4.7K
+15V +5V
+5V RN8 8*30K
+15V 10 20 RN5 1 8*330 1
2 VR1 1 -5V + C15 3 GND Vcc 9 8 9 RN12 4*4.7K M2
-15V D
LM79L05 10µF Ta 4 2 7 10 2 8 11
1 EV - INPUT VALVE
+ C12 7 5 6 11 5 7 12
2 ABS - SUCTION VALVE
3 10µF Ta + C16 8 8 5 IC21 12 13
D C 3 SP - WASH
10µF Ta 13 7 4 13 14
4 PV - SAMPLE
+5V -15V 14 6 3 ILQ1 14 6 C 6 15
IC17 5 BGA 1 / 3
-15V 17 9 2 15 9 5 16
6 BGA 2 / 4
18 6 1 16 17
CPU Board CL Z80 ADC12
74HC273 D IC26 7 LV -
RN13 18
8 VE1 -
1 40 28 Q ULN
5 1 16 RN12 2804
GND V+ V- SEL5 11 B
0.15µF CI1 12 2 15 3 4
DB4 18 32 10
LBEN INT 15 3 14 4 3 V+ 9
DB5 19 IC22
HBEN 0.33µF CAZ1 4 4 13 A
DB6 2 31 1
STATUS AZ CL 3 5 12
DB7 26 ILQ1 B
RUN/HOLD 200K RI1 16 6 11 7 2
30 9
BUF 19 7 10 8 1 GND 10
DA0 16
B1 2 8 9 A
DA1 15 38
B2 REFC- RN13 4*4.7K
DA2 14
B3 1uF CR1
DA3 13 37 +5V
B4 REFC+ RN9 8*30K
DA4 12 10 20 RN6 1 8*330 1
CPU-PLATINE / CPU BOARD CL Z80ADC REV. 1.2 ESCHWEILER GmbH & CO. KG KIEL GERMANY IDENT Nr.: xxxx
SCHALTPLAN / CIRCUIT DIAGRAM DATUM / DATE : 10.01.2006 NAME : CL Z80ADC 12c.TCD BLATT / PAGE : 3/4
page 39
CPU Board CL Z80 ADC12
Specification
AD-conversion: (2 x 8 Bit parallel) 12 Bit AD-converter, ICL7109 int., range -4.096V to +4.096V, resolution
1mV. 16 channel multiplexer, HI506
Driver Output (24 pieces) 3 x 8 Bit outputs, with D-Latch (74HC273), Optocoupler (galvanic insulation)
and driver circuits for loads up to 1A (ULN2803)
Level control: (1 x 8 Bit) 8 Input-Ports for checking the filling levels of reagents via L12 CL FÜLLSTAND
2. Thermo printer at L3
Reset-component: battery operation, automatic reset generation, switch over to battery operation
4 Timer and counter channels: Baudrate generation (2 channels) Timer (multitasking software modul)
Remark:
An automatic reset is generated if the operation voltage of the CPU board becomes lower than 4.75V or
higher than 5.25V.
Technical Data:
Databit: 8
Parity: no
no handshake (unidirectional)
Description:
All data which is sent to the printer after completition of a sample measurement, will also be transmitted to
the RS232 interface at the same time together with CR (carriage return) and LF (line feed) at the end of the
line. The transmission is unidirectional, there is no dialog between COMBILINE and PC. The data has to be
received in a polling mode after starting a measurement. The data is given out in standard ASCII code.
Data recalled from the memory (SD-Card) will not be sent via RS232 !
Remark:
The data given by the interface can be detected by any terminal program for example TERMINAL.
Printout:
2. Value - ADCMSK = sensors which exceeded the permissible time for measurement (measure-value with ?)
All values are available in ASCII-Format as hexadecimal value. The value range is from 0 to 32768.
Explanation:
00 1 Syringe PO2
01 2 Capillary PCO2
02 4 BGA only PH
03 8 ISE only K
04 16 QC NA
05 32 Respiratory Gas CA
06 64 Serum, Plasma CL
07 128 Blood LI
08 256 - HB
09 512 - -
10 1024 - -
11 2048 - -
12 4096 - -
13 8192 - -
14 16384 - -
15 32768 - -
3 2+1 pO2 + pCO2 did not comply with permissible measure times
255 128+64+32+16+8+4+2+1 pO2, pCO2, pH, K+, Na+, Ca++, Cl-, Li+ were ACTIVE during measurement
Layout
Circuit Diagram
combiline
1 1
6 RX RX 6
2 2
RX = Receive Data / Daten empfangen 7 7
TX = Transmit Data / Daten senden 3 3
Interface Cable
8 TX TX 8
4 4
9 9
5 5
GND
Computer Analysator / Analyser
1
Serial Interface RS-232 SD INT 1.4
1
14 RX RX 6
2 2
Computer
RS232 INTERFACEKABEL / RS232 INTERFACE-CABLE ESCHWEILER GmbH & CO. KG KIEL GERMANY
ECOSYS II, ECOLYTE, COMBISYS II, COMBILINE DATUM: 01.02.2006 BEZ.: RS232 Kabel.TCD BLATT: 1/1
page 47
Fill Level Control CL FÜLLSTAND
Description:
Via L2 the waste bottle sensor is connected. The control of the Calibrants is made by the Sample Light
Barrier in the Sensorholder (Thermostat).
Via L1 the voltage supply (+5V / GND digital) and the outputs of the comparators are connected. The switch
thresholds of the light barrier LS1 for WASH is adjusted with potentiometer TR1.
empty full
rinsing solution T8 0V +5 V
waste bottle T5 +5 0V
Transport of calibration solutions CAL3, CAL4, BGA1, BGA2, BGA3 and BGA4 is controled by the Sample
Light Barrier during calibration procedure!
Layout
page 50
R16 R17
200 4,7K
R1 R10 TS1 R8
S1.4 S1.1
100K 100K 9.1K
L1 VERBUNDEN MIT Z80 ADC L12
OPB 875 T51 L1 CONNECTED TO Z80 ADC L12
E S
R3 LS2
10 3
- T4
S1.3 S1.2 13
30K TLC272CD IC2_D L1.2 BIT 0
11
Circuit Diagram
ABFALLFLASCHE 4
WASTE BOTTLE R12 R13 - T8
1,1K 3K 2
IC2_B L1.5 BIT 3
5
TLS + 12 AUSGANG SPÜLLÖSUNG
Testpunkte / testpoints : OUTPUT WASH SOLUTION
T9 - ca. 0,5V Referenzspannung für Lichtschranken / reference voltage for light barriers
ESCHWEILER CombiLine2
CL FÜLLSTAND REV 2.0
CL FÜLLSTAND 20.tcd / 01.02.2010
Ident-No. 3011
Blatt 1 / 1
FÜLLSTANDSKONTROLLE / LEVEL CONTROL CL FÜLLSTAND REV 2.0 ESCHWEILER GmbH & CO. KG KIEL GERMANY IDENT Nr. : 3011
FÜR ABFALLFLASCHE UND REAGENZIEN / FOR WASTE BOTTLE AND REAGENTS DATUM / DATE : 01.02.2010 NAME : CL FÜLLSTAND 20.TCD BLATT / PAGE : 1/1
combiline
Keyboard
Specification
The keyboard is operated directly at a parallel 8-bit port. Debouncing of the contacts takes place by by a
progranmable Controller (PIC)
The PIC scans the keyboard-matrix permanently. If a key is pressed, the corresponding value is sent to the
port. By an interrupt the information is interpreted by the analyser software. The corresponding action will be
performed then.
Cleaning
Use a suitable cleaning solution as described in chapter !2 Maintenance and Hygiene and pay attention to
the notes.
+5V
AR1 C1
TASTATUR 4*10K 100nF
1 2 28
TOCKI +U MCLR
F1 F2 F4 F3
6
9 RA0 IC1
1 4 C 7 PIC 16C57C
7 25
8 RA1 26
ENTER 2 5 0 8
8 18
7 RA2 RC0 17
3 6 . 9
9 19
6 RA3 RC1 15
20
10 RC2 13
5 RB0
21 L10
11 RC3 11 Z80ADC
4 RB1
22
12 RC4 9
3 RB2
23
13 RC5 7
2 RB3
24
14 RC6 5
1 RB4
25
RC7 3
L2 27
OSC1 1,2
26 L1
OSC2
Q1
4.0 MHz GND
4
ESCHWEILER CombiLine
Tastaturadapter CL TASTATUR Rev. 1.2
CL TASTATUR 12.tcd / 01.02.2010
Ident-No. 3031
TASTATURADAPTER CL TASTATUR Rev. 1.2 ESCHWEILER GmbH & CO. KG KIEL GERMANY IDENT Nr.: 3031
SCHALTBILD UND BESTÜCKUNGSPLAN / CIRCUIT DIAGRAM AND LAYOUT DATUM / DATE : 01.02.2010 BEZ.: CL TASTATUR 12.TCD BLATT / PAGE : 1 / 1
Specification
Type: NLC240x128BTGLE
Contrast Adjustment
The trim potentiometer for contrast adjustment is located above the roller pump, as shown in figure 42a.
• Turn the potentiometer knob until the contrast of the display is pleasant for the operator.
Paper specifications:
Roll diameter: up to 60mm
Roll width: 57mm
Paper thickness: 0.04mm
Key-Functions
FEED / ENTER key (1)
Printer self test:
Press key 1 then switch analyser ON.
Keep key 1 pressed for at least 3 seconds.
The printer software version and the character
set will be printed
1 LED
air
CAL3
CAL4
BGA3/1
BGA4/2
outlet
The Suction Pump is mounted with an aluminium angle beside the trans-
former and is controlled by the Main Board M1.
• Wear gloves!
• Uninstall the Suction Pump as described in the chapter above.
• Remove lid (B) from pump head, see figure 49 on next page.Liquid-Valve
• Remove the two screws (A) and the front panel (don't loose centre ring for
motor axis).
• Clean the corresponding internal parts with a suitable solution, see chapter
!2 Hygiene.
G
D
E F
AS
F
A B
Ref. Haltbarkeit
(Monate)
pO2 Haltbarkeit
(Monate)
pCO2 Haltbarkeit
(Monate)
pH Haltbarkeit
(Monate)
durability durability durability durability
1 (month) (month) (month) (month)
12 - 36
1 1
2 1
2 12 - 48 2 12 - 24
3 3 3
12 - 36
2 12 - 24
4 3-6 4
4 3
3-6
5 12 - 48 5 12 - 48 5 12 - 48 3 12 - 24
6 6 6 4
Referenzsensor kompl. pO2 Sensor kompl. pCO2 Sensor kompl. pH Sensor kompl.
reference sensor compl. pO2 sensor compl. pCO2 sensor compl. pH sensor compl.
50 5 10 00 50 1 10 00 50 1 20 00 50 1 30 00
4 Hülse mit Membran 4 Hülse mit Membran 4 Hülse mit Membran 4 O-Ring 1.8 x 1.0
membrane shell membrane shell membrane shell o-ring 1.8 x 1.0
50 5 10 30 50 1 10 30 50 1 20 30 50 1 10 25
BGA_combiline/15012009HEU
Haltbarkeit Haltbarkeit
Na (Monate) ISE (Monate)
durability durability
(month) (month)
12 - 36
K+ 3 - 6
1 12 - 36 2 Ca++ 3 - 6
2 Li+ 3 - 6
Cl- 3 - 6
Na+12-24
3 3 12 - 24
4 4
Ersatzteile ISE Sensoren / spare parts ISE sensors K+ / Ca++ / Cl- / Li+
1 Ableitsystem 3 Sensorgehäuse
conducting system sensor casing
K+ 50 3 10 10 K+ 50 3 10 20
Ca++ 50 3 30 10 Ca++ 50 3 30 20
Cl- 50 3 40 10 Cl- 50 3 40 20
Li+ 50 3 50 10 Li+ 50 3 50 20
12 - 36 Monate 6 - 9 Monate
Ref. Referenzsensor
reference sensor 12 - 36 months 6 - 9 months
50 5 10 00
Sensor-Einheit Membran
sensor unit membrane
12 - 24 Monate
pH pH Sensor
pH sensor 12 - 24 months
50 1 30 00
Ableitsystem 1 Membran
conducting system 1 membrane
for combiline
line BGA+E
40 6 10 10 WASH 2 (Rinsing solution for BGA plus E) 6x250 ml
40 6 10 40 BGA 3 (Calibration solution BGA plus E/low) 12x130 ml
40 6 10 50 BGA 4 (Calibration solution BGA plus E/high) 12x130 ml
40 6 10 45 CAL 3 (Calibration solution for BGA plus E/low) 12x150 ml
40 6 10 55 CAL 4 (Calibration solution for BGA plus E/high) 12x150 ml
for combiline
line ISE
3-4.110 WASH 2 (Rinsing solution for ISE) 6x330 ml
3-4.400 CAL 3 (Calibration solution for ISE/low) 6x330 ml
3-4.500 CAL 4 (Calibration solution for ISE/high) 6x330 ml
Special Solutions
50 6 10 80 Protein remover 1 x 100 ml
50 6 10 84 Electrode cleaner 50 ml
50 6 10 88 Activating solution for pH-sensor 20 ml
Quality Controls
Control Solutions for BGA
50 6 20 10 Level I BGA (acidosis) 30 ampoules
50 6 20 20 Level II BGA (normal) 30 ampoules
50 6 20 30 Level III BGA (alkalosis) 30 ampoules
Sampling Systems
Capillaries
50 6 30 08 hep. capillaries 8 cm/lg 250 pcs.
50 6 30 10 hep. capillaries 10 cm/lg 250 pcs.
50 6 30 12 hep. capillaries 12 cm/lg 250 pcs.
50 6 30 14 hep. capillaries 14 cm/lg for instruments with Hb sensor 250 pcs.
50 6 30 16 hep. capillaries 16 cm/lg for instr. with meta and Hb sensor 250 pcs.
50 6 30 30 metal stirrers for capillaries 250 pcs.
50 6 30 35 magnet for stirrer
50 6 30 40 end-caps for capillaries 500 pcs.
Printer Paper
50 6 50 00 printer paper pack.=2 rolls
Cleaning Paste
50 1 10 41 cleaning paste (silicon carbide)
Sensors
pO2-Sensor
50 1 10 00 pO2-Sensor complete
50 1 10 10 pO2-Sensor-unit with membrane shell
50 1 10 15 gasket for Sensor-unit (8.0 x 1.0)
50 1 10 20 pO2-Sensor-casing with screw-cap
50 1 10 21 screw cap for Sensor-casing
50 1 10 25 gasket for Sensor-casing (1.8 x 1.0)
50 1 10 30 pO2-membrane shells pack.= 5 pcs.
50 1 10 35 pO2-fill solution bot.=20 ml
50 1 10 50 polishing kit for pO2 sensors
pCO2-Sensor
50 1 20 00 pCO2-Sensor complete
50 1 20 10 pCO2-Sensor-unit with membrane shell
50 1 10 15 gasket for Sensor-unit (8.0 x 1.0)
50 1 20 20 pCO2-Sensor-casing
50 1 10 21 screw cap for Sensor-casing
50 1 10 25 gasket for Sensor-casing (1.8 x 1.0)
50 1 20 30 pCO2-membrane shells pack.=5 pcs.
50 1 20 35 pCO2-fillsolution bot.=20 ml
pH-Sensor
50 1 30 00 pH-Sensor complete
50 1 30 10 pH-Sensor-unit
50 1 30 20 pH-Sensor-casing
50 1 10 21 screw cap for Sensor-casing
50 1 10 25 gasket for Sensor-casing (1.8 x 1.0)
K+-Sensor
50 3 10 00 K+-Sensor complete
50 3 10 10 K+-conducting system
50 3 10 20 K+-Sensor casing
50 1 10 25 gasket for Sensor-casing 1.8 x 1.0)
50 3 10 30 K+-membrane shell with membrane
50 3 10 35 K+-fillsolution (organic tenside, buffer substances) bot.=20 ml
Na+-Sensor
50 3 20 00 Na+-Sensor complete btl.= bottle
50 3 20 10 Na+-conducting system
50 3 20 15 gasket for conducting system (3.2 x 1.8)
50 3 20 20 Na+-Sensor-casing
50 1 10 25 gasket for Sensor-casing 1.8 x 1.0)
50 3 20 35 Na+-fillsolution bot.= 20 ml
Ca++-Sensor
50 3 30 00 Ca++-Sensor complete
50 3 30 10 Ca++-conducting system
50 3 30 20 Ca++-Sensor casing
50 1 10 25 gasket for Sensor-casing 1.8 x 1.0)
50 3 30 30 Ca++-membrane shell with membrane
50 3 30 35 Ca++- fillsolution bot.=20 ml
Cl--Sensor
50 3 40 00 Cl--Sensor complete
50 3 40 10 Cl-- conducting system
50 3 40 20 Cl--Sensor casing
50 1 10 25 gasket for Sensor-casing 1.8 x 1.0)
50 3 40 30 Cl--membrane shell with membrane
50 3 40 35 Cl--fillsolution bot.=20 ml
Li+-Sensor
50 3 50 00 Li+-Sensor complete
50 3 50 10 Li+-conducting system
50 3 50 20 Li+-Sensor casing
50 1 10 25 gasket for Sensor-casing 1.8 x 1.0)
50 3 50 30 Li+-membrane shell with membrane
50 3 50 35 Li+-fillsolution bot.=20 ml
Reference-Sensor
40 5 10 00 reference-sensor complete
50 5 10 10 reference-conducting system
50 3 20 15 gasket for conducting system (3.2 x 1.8)
50 5 10 20 reference-Sensor casing
50 1 10 25 gasket for Sensor-casing 1.8 x 1.0)
40 5 10 50 reservoir for reference-Sensor casing
50 5 10 30 ref.-membrane shell with membrane
50 5 10 35 reference-fillsolution bot.=50 ml
Calibration-bag adapter
40 7 10 35 bag-adapter COMBILINE, complete
consisting of:
G 40.1 metal-casing
40 7 10 40 slide-bar holder COMBILINE
G 41.3 slide-bar COMBI 1 set
G 40.4 spring for slide-bar 1 set
G 40.5 cannula 1set
Peristaltic pump
50 7 40 00 peristaltic pump, complete
50 7 40 10 micromotor with gear 97.3:1
50 7 40 20 bush for bearing
50 7 40 30 bearing
50 7 40 40 roller device for peristaltic pump
50 7 40 50 clutch
Reagent-wall
50 7 40 60 tube connector, black
50 7 40 61 tube connector, white, reagent-wall
50 7 40 62 tube connector, white, bottle compartment
50 7 40 70 holder for roller pump tubing, reagents (low)
40 7 40 70 holder for roller pump tubing WASH (high)
40 7 10 50 intermediate vessel
40 7 10 55 holder for intermediate vessel
Sample port
Rm 20.80 sample port device, complete with insert
consisting of:
R 20.81 (1-8.105) sample port insert with integr. Steel canula (white)
R 20.83 (1-8.106) sample port insert (red)
Wiring
SEm 100.21 sensor-cable pO2
SEm 200.21 sensor-cable pCO2
SEm 300.21 sensor-cable pH
SEm 400.21 sensor-cable Na+
SEm 500.21 sensor-cable K+
SEm 600.21 sensor-cable Ca++
SEm 700.21 sensor-cable Li+
SEm 800.21 sensor-cable Cl-
SEm 900.22 sensor-cable reference
Valves
F 200.25 tube valve (normally open)
F 200.28 sample port valve PV (normally open)
Declaration of EC-Conformity
as defined by EC In-Vitro Diagnostica (IVD) directive 98/79/EG
EN 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use:
General requirements
EN 61010-2-08/A1 Safety requirements for electrical equipment for measurement, control, and laboratory use:
Particular requirements for laboratory equipment for analysis
EN 61010-2-101 Safety requirements for electrical equipment for measurement, control, and laboratory use:
Particular requirements for in vitro diagnostic (IVD) medical equipment
Emission:
EN 55011 A2 Industrial, scientific and medical radio-frequency equipment – Emission: Class B
EN 61000-3-2 Harmonic current emissions
EN 61000-3-3 Voltage fluctuation and flicker
Immunity:
EN 61326-1 Electrical equipment for measurement, control and laboratory use – EMC requirements:
Industrial requirement
EN 61000-4-2 Electrostatic discharge immunity test
EN 61000-4-3 +A1 Radiated, radio-frequency electromagnetic field – immunity test
EN 61000-4-4 Electrical fast transient/burst immunity test
EN 61000-4-5 Surge immunity tests
EN 61000-4-6 Immunity to conducted disturbances, induced by radio frequency fields
EN 61000-4-11 Voltage dips, short interruptions and voltage variations immunity test
Kiel, 01.04.2010
.....................................
i.A Dr. H.-J. Heupke
(QCM)
EC declaration combilineV2 IVD engl/01012010
ESCHWEILER Telefon: 0431 / 546580 Deutsche Bank AG, Kiel Commerzbank, Kiel Postbank
GmbH & Co. KG Telefax: 0431 / 549423 Kto.-Nr. 0324 426 Kto.-Nr. 7224488 Hamburg 24676 - 204