Table 1: Reference ambient temperature read at thermometer = 30˚C

No Thermometer reading (˚C) Average (˚C)


1 30.0
30.0
2 30.0

TE1 at Tl(1), K TE2 at Tl(2), J TE3 at Tl(3), RTD


RUN 1
UNIT
(HEATER OFF) Set 1 Set 2 Avg Set 1 Set 2 Avg Set 1 Set 2 Avg

A Temperature ˚C 30.30 30.40 30.35 29.90 30.0 29.95 30.50 30.40 30.45

B1 mV 1.2153 1.2194 1.2174 1.5318 1.5370 1.5344


MV from TC table
B2 Ω 111.8650 111.8260 111.8455

C1 mV 0.0150 0.0130 0.0140 0.0220 0.0210 0.0215


Measure signal
C2 Ω 112.02 112.00 112.01

D1 = B1 + C1 mV 1.2303 1.2324 1.2314 1.5538 1.5580 1.5559


Actual signal
D2 = C2 Ω 112.02 112.00 112.01

E Calculated temperature ˚C 30.67 30.72 30.69 30.32 30.40 30.36 30.90 30.85 30.87

F= | E – A | Temperature deviation ˚C 0.37 0.32 0.35 0.42 0.40 0.41 0.40 0.45 0.43
Table 2: Reference ambient temperature read at thermometer = 40˚C

No Thermometer reading (˚C) Average (˚C)


1 30.0
30.0
2 30.0

TE1 at Tl(1), K TE2 at Tl(2), J TE3 at Tl(3), RTD


RUN 1
UNIT
(HEATER OFF) Set 1 Set 2 Avg Set 1 Set 2 Avg Set 1 Set 2 Avg

A Temperature ˚C 39.80 39.90 39.85 40.0 40.50 40.25 39.50 39.90 39.70

B1 mV 1.6038 1.6079 1.6059 2.0590 2.0850 2.0720


MV from TC table
B2 Ω 115.345 115.501 115.423

C1 mV 0.2980 0.3030 0.3005 0.3930 0.3890 0.3910


Measure signal
C2 Ω 129.260 129.230 129.245

D1 = B1 + C1 mV 1.9018 1.9109 1.9064 2.4520 2.4740 2.4630


Actual signal
D2 = C2 Ω 129.260 129.230 129.245

E Calculated temperature ˚C 47.07 47.28 47.18 47.47 47.89 47.68 75.71 75.63 75.67

F= | E – A | Temperature deviation ˚C 7.27 7.38 7.33 7.47 7.39 7.43 35.81 35.73 35.77
Table 3: Reference ambient temperature read at thermometer = 50˚C

No Thermometer reading (˚C) Average (˚C)


1 31.3
31.15
2 31.0

TE1 at Tl(1), K TE2 at Tl(2), J TE3 at Tl(3), RTD


RUN 1
UNIT
(HEATER OFF) Set 1 Set 2 Avg Set 1 Set 2 Avg Set 1 Set 2 Avg

A Temperature ˚C 51 51 51 51 51 51 50 50 50

B1 mV 2.0640 2.0640 2.0640 2.638 2.638 2.638


MV from TC table
B2 Ω 119.40 119.40 119.40

C1 mV 0.6030 0.6060 0.6045 0.765 0.761 0.763


Measure signal
C2 Ω 119.82 120.04 119.93

D1 = B1 + C1 mV 2.6670 2.6700 2.6685 3.4056 3.3990 3.4023


Actual signal
D2 = C2 Ω 119.82 120.04 119.93

E Calculated temperature ˚C 65.56 65.63 65.60 65.44 65.31 65.38 51.10 51.67 51.38

F= | E – A | Temperature deviation ˚C 14.56 14.63 14.60 14.44 14.31 14.38 1.10 1.67 1.39
Table 4: Reference ambient temperature read at thermometer = 60˚C

No Thermometer reading (˚C) Average (˚C)


1 33.0
33.0
2 33.0

RUN 1 TE1 at Tl(1), K TE2 at Tl(2), J TE3 at Tl(3), RTD


UNIT
(HEATER OFF) Set 1 Set 2 Avg Set 1 Set 2 Avg Set 1 Set 2 Avg
A Temperature ˚C 61.1 61.1 61.1 61.4 61.2 61.3 59.9 60.3 60.1

B1 mV 2.4822 2.4822 2.4822 3.1902 3.1796 3.1849


MV from TC table
B2 Ω 123.202 123.357 123.280

C1 mV 0.9100 0.9110 0.9105 1.1490 1.1470 1.1480


Measure signal
C2 Ω 123.86 138.2 131.03

D1 = B1 + C1 mV 3.3922 3.3932 3.3927 4.3392 4.3266 4.3329


Actual signal
D2 = C2 Ω 123.86 138.2 131.03

E Calculated temperature ˚C 83.03 83.95 83.49 82.84 82.60 82.72 61.56 99.18 80.37

F= | E – A | Temperature deviation ˚C 21.93 22.85 22.39 21.44 21.4 21.42 1.66 38.88 20.27
A B C
Element Type Deviation Range (°) Difference (°C) in the range

Type K (TE1) 0.35 - 22.39 22.04

Type J (TE2) 0.41 - 21.42 21.01

RTD (PT 100)(TE3) 0.43 - 35.77 35.34