Professional Documents
Culture Documents
Validation - DataPro 3 Full
Validation - DataPro 3 Full
CLiMET DataPro 3
Supplemental Validation
of Version 1.20
Notices
Copyright © 2003-2011 by CLiMET Instruments Company. All rights reserved.
Complying with applicable copyright laws, no part of this document may be reproduced, stored in or introduced to any retrieval system, or
retransmitted in any form or by any means (electronic, mechanical, photocopying, recording, or otherwise), or for any purpose, without the prior
written permission of CLiMET Instruments Company. Contact CLiMET Instruments to request written permission.
CLiMET Instruments Company may make improvements and/or changes in the product(s) and/or the program(s) described in this document at
any time without notice; these improvements and/or changes will be incorporated in new versions of the documentation. Unless otherwise noted,
the example companies, organizations, products, people and events depicted herein are fictitious and no association with any real company,
organization, product, person or event is intended or should be inferred.
Although every precaution has been taken in the preparation of this documentation, the author assumes no responsibility for technical
inaccuracies, typographical errors, or damages resulting from the use of the information contained herein.
The following paragraph does not apply to a country where such provisions are inconsistent with local law:
CLIMET INSTRUMENTS COMPANY PROVIDES THIS DOCUMENTATION “AS IS” WITHOUT WARRANTY OF ANY KIND,
EITHER EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF NON-INFRINGEMENT,
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Some states do not allow a disclaimer of express or implied
warranties in certain transactions, therefore, this statement may not apply to you.
Trademarks
CLiMET is a trademark of CLiMET Instruments Company in the United States and/or other countries. Microsoft, Windows, MSDE, SQL
Server 2005 Express, Visual Basic .NET, Visual Studios .NET and Excel are trademarks of Microsoft Corporation in the United States and/or
other countries. VeriSign is a trademark of VeriSign in the United States and/or other countries. RSA and MD5 are trademarks of RSA Security
Inc. in the United States and/or other countries. The names of actual companies and products mentioned herein may be the trademarks of their
respective owners.
Supplementary Features
Note indicates notes related to the nearby text.
Tip indicates tips related to the nearby text.
Typographical Conventions
Bold indicates important words, or names of buttons, menus, and windows.
Italic indicates new terms, or terms in glossary.
indicates a procedure in the table of contents.
§ indicates reference to a section number.
Table of Contents
1 Introduction ........................................................................................................................... 5
2 System Requirements & Equipment ................................................................................... 5
3 IQ: Installation Qualification ............................................................................................... 5
3.1 Scope ..................................................................................................................... 5
3.2 Peripherals Test ..................................................................................................... 5
1 Introduction
CLiMET Instruments Company designed, constructed and validated
CLiMET DataPro 3 Version 1.20 at the CLiMET facility in Redlands,
California. The validation method described herein is consistent with
CLiMET’s “SOFTWARE DEVELOPMENT, DOCUMENTATION AND
REVISION STANDARDS” policy (Revision 3.0 – 03/04/1999).
The purpose of this validation document is to provide documented
evidence that DataPro will correctly install and consistently operate in
accordance with the procedures outlined in the DataPro manual. This
protocol describes all of the conditions and tests performed for this
validation.
The manual describes DataPro. The validation system has inputs and
outputs. The inputs to the validation system are particle samples and
the DataPro user interface. The outputs of the system are printouts,
databases and the data recorded in this document.
Setup A personal computer with one of the supported operating systems and,
One DataPro installation CD.
Success Yes No
You will be told that you are not a member of the Administrators or Domain
Admins group.
5. Start the DataPro program as Administrators or Domain Admins member.
You will be told that you have fourteen-days left in your evaluation.
6. Change the date in your PC in increments until the evaluation period expires.
Make sure that reinstalling the DataPro client on a later day resets the
evaluation period anew.
7. Start the DataPro program as Users member.
Make sure you can now run the program after the Administrators or Domain
Admins member ran the program once.
Success Yes No
5. Click OK.
Make sure the product activation works under all user levels.
Success Yes No
Purpose Test that the system works when installing the system in one computer.
Success Yes No
Purpose Test that Overview and View Alarms tabs works correctly.
Setup Use CLiMET’s Validation Utility (VU).
Procedure Generate the following events.
Normal (a.k.a., OK) Coarse Alert
Off (a.k.a., Standby) Coarse Alert Recovery
Bad Flow Fine Action
Bad Flow Recovery Fine Action Recovery
Bad Laser Fine Alert
Bad Laser Recovery Fine Alert Recovery
Bad Sensor No Response
Bad Sensor Recovery No Response Recovery
Coarse Action Forced Counts
Coarse Action Recovery Forced Counts Recovery
Service Not Responding Sample Time Expired
Not Used (blank) Sample Time Expired Recovery
Make sure Overview and View Alarms tabs show the events.
The Overview tab does not display Recoveries, Forced Counts,
Sample Time Expired, or Alarm Tower No Response.
The View Alarms tab does not display Service Not Responding, Off,
Not Used or Normal.
Success Yes No
Purpose Test that the sub controls of the Reports tab work as outlined in
Functional Specification (FS).
Setup None
Success Yes No
Purpose Test that the Reports tab creates the documents and verifies the data
integrity.
Procedure For all locations and allowed sizes in Ampoule Filling Machine 1, make
sure the following documents are correct and attach them to the back of
this test:
1. Reports
1.1. EU GMP Annex 1
1.1.1. Alternate
1.1.1.1. In Operation
1.1.1.1.1. A
1.1.1.1.2. B
1.1.1.1.3. C
1.1.1.2. At Rest
1.1.1.2.1. A
1.1.1.2.2. B
1.1.1.2.3. C
DataPro 3 V1.20, Validation
August 1, 2011 Page 15 of 156 P/N 03036901
Validation: DataPro 3 software
1.1.2. Normal
1.1.2.1. In Operation
1.1.2.1.1. A
1.1.2.1.2. B
1.1.2.1.3. C
1.1.2.2. At Rest
1.1.2.2.1. A
1.1.2.2.2. B
1.1.2.2.3. C
1.2. ISO 14644-1 In Operation
1.2.1. C2
1.2.2. C3
1.2.3. C4
1.2.4. C5
1.2.5. C6
1.2.6. C7
1.2.7. C8
1.2.8. C9
1.3. Raw Data
1.3.1. Total Count
1.3.2. Rolling Count
1.4. Summary
1.4.1. CF
1.4.2. CM
1.4.3. Total Count
2. Graphs
2.1. Raw Data
2.1.1. CF Report
2.1.1.1. CM Samples
2.1.1.1.1. “.5”
2.1.1.1.2. “5”
2.1.1.2. CF Samples
2.1.1.2.1. “.5”
2.1.1.2.2. “5”
2.1.2. CM Report
2.1.2.1. CM Samples
2.1.2.1.1. “.5”
2.1.2.1.2. “5”
2.1.2.2. CF Samples
2.1.2.2.1. “.5”
2.1.2.2.2. “5”
2.1.3. Total Count Report
2.1.3.1. CM Samples
2.1.3.1.1. “.5”
2.1.3.1.2. “5”
2.1.3.2. CF Samples
2.1.3.2.1. “.5”
2.1.3.2.2. “5”
2.1.4. Alarm Test
2.1.4.1. CM Report
2.1.4.1.1. CF + CM + Logarithmic + Alarms
2.1.4.1.1.1. “.5”
2.1.4.1.1.2. “5”
2.1.4.2. CF Report
2.1.4.2.1. CF + CM + Logarithmic + Alarms
2.1.4.2.1.1. “.5”
2.1.4.2.1.2. “5”
2.1.4.3. Total Count
2.1.4.3.1. CF + CM + Logarithmic + Alarms
2.1.4.3.1.1. “.5”
2.1.4.3.1.2. “5”
2.1.4.3.2.
2.2. Time Scale
2.2.1. CF Report
2.2.1.1. CM Samples
2.2.1.1.1. “.5”
2.2.1.1.2. “5”
2.2.1.2. CF Samples
2.2.1.2.1. “.5”
2.2.1.2.2. “5”
2.2.2. CM Report
2.2.2.1. CM Samples
2.2.2.1.1. “.5”
2.2.2.1.2. “5”
2.2.2.2. CF Samples
2.2.2.2.1. “.5”
2.2.2.2.2. “5”
2.2.3. Total Count Report
2.2.3.1. CM Samples
2.2.3.1.1. “.5”
2.2.3.1.2. “5”
2.2.3.2. CF Samples
2.2.3.2.1. “.5”
2.2.3.2.2. “5”
2.2.4. Alarm Test
2.2.4.1. CM Report
2.2.4.1.1. CF + CM + Logarithmic + Alarms
2.2.4.1.1.1. “.5”
2.2.4.1.1.2. “5”
2.2.4.2. CF Report
2.2.4.2.1. CF + CM + Logarithmic + Alarms
2.2.4.2.1.1. “.5”
2.2.4.2.1.2. “5”
2.2.4.3. Total Count
2.2.4.3.1. CF + CM + Logarithmic + Alarms
2.2.4.3.1.1. “.5”
2.2.4.3.1.2. “5”
NOTE: The Total Count vs CM/CF will not match up for CM sample data as rolling count validation data is created as a static number with the VU
with variable range and not a rolling count of the actual intervals being generated. The data is verifiable as it comes from a counter, but not against
itself.
Success Yes No
Purpose Verifies that Excel can handle more than 35000 rows. 21 billion hard cap.
Procedure Create a raw data report and make sure it has more than 35000 rows.
Create a raw graph, timescale graph and make sure the resulting
Excel file contains more than 35000 rows of data.
Create 1 of each report and more sure resulting files report the proper
range.
Success Yes No
Purpose Verifies that large amounts of data can be handled by user and allows
the user to limit data according to their hardware spec’s.
Procedure Create a raw data report and make sure that the program asks for
user input before verifying the data.
Create a raw data graph and check to make sure that 3 options
appear in a dialog window.
Check each option to make sure that data is plotted with the
designated number of rows. If an error occurs limit time range and
attempt test again.
Success Yes No
Check to make sure each line has a different line designation, proper
name, value and color association assigned. Do this for both graphs.
Success Yes No
12/1/2011
Done by Date
Purpose Test that the alternate EU GMP Annex 1 concentration limits work
correctly. Note: Alternate does not mean that the limits are different just
that the sampling size is now full cubic meters only.
Procedure 1. Create cubic meter concentrations that both pass and fail each of the
2. EU GMP Annex 1 concentration limits of the manual.
Success Yes No
Setup None
Success Yes No
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 0.5 µm Cubic Meter 204439 204900 2012-01-24 10:39:31 3302 6 CM 0 0 0 Fail
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 5.0 µm Cubic Meter 484 606 2012-01-24 12:59:32 1 6 CM 0 0 0 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00 ≥ 0.5 µm Cubic Meter 204857 205488 2012-01-24 11:14:31 3517 6 CM 0 0 1 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00 ≥ 5.0 µm Cubic Meter 456 551 2012-01-24 11:49:33 13 6 CM 0 0 1 Fail
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 0.5 µm Cubic Meter 2999 3100 2012-01-24 11:53:14 99 6 CM 0 0 1 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6 CM 0 0 1 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00 ≥ 0.5 µm Cubic Meter 2993 3102 2012-01-24 11:53:14 79 6 CM 0 0 0 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6 CM 0 0 0 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 0.5 µm Cubic Meter 204439 204900 2012-01-24 10:39:31 3302 6 CM 0 0 0 Pass
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 5.0 µm Cubic Meter 484 606 2012-01-24 12:59:32 1 6 CM 0 0 0 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00 ≥ 0.5 µm Cubic Meter 204857 205488 2012-01-24 11:14:31 3517 6 CM 0 0 1 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00 ≥ 5.0 µm Cubic Meter 456 551 2012-01-24 11:49:33 13 6 CM 0 0 1 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 0.5 µm Cubic Meter 2999 3100 2012-01-24 11:53:14 99 6 CM 0 0 1 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6 CM 0 0 1 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00 ≥ 0.5 µm Cubic Meter 2993 3102 2012-01-24 11:53:14 79 6 CM 0 0 0 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6 CM 0 0 0 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 0.5 µm Cubic Meter 204439 204900 2012-01-24 10:39:31 3302 6 CM 0 0 0 Pass
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 5.0 µm Cubic Meter 484 606 2012-01-24 12:59:32 1 6 CM 0 0 0 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00 ≥ 0.5 µm Cubic Meter 204857 205488 2012-01-24 11:14:31 3517 6 CM 0 0 1 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00 ≥ 5.0 µm Cubic Meter 456 551 2012-01-24 11:49:33 13 6 CM 0 0 1 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 0.5 µm Cubic Meter 2999 3100 2012-01-24 11:53:14 99 6 CM 0 0 1 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6 CM 0 0 1 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00 ≥ 0.5 µm Cubic Meter 2993 3102 2012-01-24 11:53:14 79 6 CM 0 0 0 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6 CM 0 0 0 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 0.5 µm Cubic Meter 204439 204900 2012-01-24 10:39:31 3302 6 CM 0 0 0 Fail
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 5.0 µm Cubic Meter 484 606 2012-01-24 12:59:32 1 6 CM 0 0 0 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00 ≥ 0.5 µm Cubic Meter 204857 205488 2012-01-24 11:14:31 3517 6 CM 0 0 1 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00 ≥ 5.0 µm Cubic Meter 456 551 2012-01-24 11:49:33 13 6 CM 0 0 1 Fail
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 0.5 µm Cubic Meter 2999 3100 2012-01-24 11:53:14 99 6 CM 0 0 1 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6 CM 0 0 1 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00 ≥ 0.5 µm Cubic Meter 2993 3102 2012-01-24 11:53:14 79 6 CM 0 0 0 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6 CM 0 0 0 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 0.5 µm Cubic Meter 204439 204900 2012-01-24 10:39:31 3302 6 CM 0 0 0 Fail
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 5.0 µm Cubic Meter 484 606 2012-01-24 12:59:32 1 6 CM 0 0 0 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00 ≥ 0.5 µm Cubic Meter 204857 205488 2012-01-24 11:14:31 3517 6 CM 0 0 1 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00 ≥ 5.0 µm Cubic Meter 456 551 2012-01-24 11:49:33 13 6 CM 0 0 1 Fail
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 0.5 µm Cubic Meter 2999 3100 2012-01-24 11:53:14 99 6 CM 0 0 1 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6 CM 0 0 1 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00 ≥ 0.5 µm Cubic Meter 2993 3102 2012-01-24 11:53:14 79 6 CM 0 0 0 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6 CM 0 0 0 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 0.5 µm Cubic Meter 204439 204900 2012-01-24 10:39:31 3302 6 CM 0 0 0 Pass
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 5.0 µm Cubic Meter 484 606 2012-01-24 12:59:32 1 6 CM 0 0 0 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00 ≥ 0.5 µm Cubic Meter 204857 205488 2012-01-24 11:14:31 3517 6 CM 0 0 1 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00 ≥ 5.0 µm Cubic Meter 456 551 2012-01-24 11:49:33 13 6 CM 0 0 1 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 0.5 µm Cubic Meter 2999 3100 2012-01-24 11:53:14 99 6 CM 0 0 1 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6 CM 0 0 1 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00 ≥ 0.5 µm Cubic Meter 2993 3102 2012-01-24 11:53:14 79 6 CM 0 0 0 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6 CM 0 0 0 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Fail
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Meter 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412 Fail
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 1 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 0 240 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Pass
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Meter 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 1 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 0 240 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Pass
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Meter 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 1 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 0 240 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Fail
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Meter 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412 Fail
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 1 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 0 240 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Fail
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Meter 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412 Fail
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 1 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 0 240 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Pass
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Meter 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 1 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 0 240 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Fail
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Fail
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Fail
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Fail
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Fail
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Fail
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Fail
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Fail
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Fail
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Fail
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Meter 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412 Fail
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 1 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 0 240 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Fail
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Fail
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Meter 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412 Fail
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 1 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 0 240 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Pass
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Meter 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 1 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 0 240 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Pass
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Meter 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 1 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 0 240 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples Result
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412 Pass
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412 Pass
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Meter 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240 Pass
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 1 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240 Pass
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 0 240 Pass
Date: 1/26/12 .
Validation: DataPro 3 software
≥ 0.5 ≥ 5.0
Timestamp µm µm Flow Volume
2012-01-24 09:30:06 3302 1 1.00CFM 1/60 CM
2012-01-24 09:30:41 3458 0 1.00CFM 1/60 CM
2012-01-24 09:31:12 3374 0 1.00CFM 1/60 CM
2012-01-24 09:31:49 3497 0 1.00CFM 1/60 CM
2012-01-24 09:32:25 3352 23 1.00CFM 1/60 CM
2012-01-24 09:32:56 3356 0 1.00CFM 1/60 CM
2012-01-24 09:33:33 3326 0 1.00CFM 1/60 CM
2012-01-24 09:34:09 3338 0 1.00CFM 1/60 CM
2012-01-24 09:34:46 3491 0 1.00CFM 1/60 CM
2012-01-24 09:35:17 3437 24 1.00CFM 1/60 CM
2012-01-24 09:35:54 3462 27 1.00CFM 1/60 CM
2012-01-24 09:36:30 3360 0 1.00CFM 1/60 CM
2012-01-24 09:37:02 3420 0 1.00CFM 1/60 CM
2012-01-24 09:37:40 3327 0 1.00CFM 1/60 CM
2012-01-24 09:38:16 3486 20 1.00CFM 1/60 CM
2012-01-24 09:38:48 3502 15 1.00CFM 1/60 CM
2012-01-24 09:39:24 3394 0 1.00CFM 1/60 CM
2012-01-24 09:40:01 3462 19 1.00CFM 1/60 CM
2012-01-24 09:40:32 3504 7 1.00CFM 1/60 CM
2012-01-24 09:41:09 3460 12 1.00CFM 1/60 CM
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
≥ 0.5 ≥ 5.0
Timestamp µm µm Flow Volume
2012-01-24 09:30:06 3517 13 1.00CFM 1/60 CM
2012-01-24 09:30:41 3499 0 1.00CFM 1/60 CM
2012-01-24 09:31:17 3382 14 1.00CFM 1/60 CM
2012-01-24 09:31:49 3462 5 1.00CFM 1/60 CM
2012-01-24 09:32:25 3459 0 1.00CFM 1/60 CM
2012-01-24 09:33:02 3394 0 1.00CFM 1/60 CM
2012-01-24 09:33:33 3366 0 1.00CFM 1/60 CM
2012-01-24 09:34:09 3454 26 1.00CFM 1/60 CM
2012-01-24 09:34:46 3370 18 1.00CFM 1/60 CM
2012-01-24 09:35:22 3408 9 1.00CFM 1/60 CM
2012-01-24 09:35:54 3503 0 1.00CFM 1/60 CM
2012-01-24 09:36:30 3331 0 1.00CFM 1/60 CM
2012-01-24 09:37:08 3358 0 1.00CFM 1/60 CM
2012-01-24 09:37:40 3432 0 1.00CFM 1/60 CM
2012-01-24 09:38:16 3448 0 1.00CFM 1/60 CM
2012-01-24 09:38:53 3452 15 1.00CFM 1/60 CM
2012-01-24 09:39:24 3391 18 1.00CFM 1/60 CM
2012-01-24 09:40:01 3464 16 1.00CFM 1/60 CM
2012-01-24 09:40:37 3407 22 1.00CFM 1/60 CM
2012-01-24 09:41:09 3348 11 1.00CFM 1/60 CM
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
≥ 0.5 ≥ 5.0
Timestamp µm µm Flow Volume
2012-01-24 09:30:11 99 0 1.00CFM 1CF
2012-01-24 09:31:12 91 0 1.00CFM 1CF
2012-01-24 09:32:09 84 0 1.00CFM 1CF
2012-01-24 09:33:12 85 0 1.00CFM 1CF
2012-01-24 09:34:09 74 0 1.00CFM 1CF
2012-01-24 09:35:12 86 0 1.00CFM 1CF
2012-01-24 09:36:15 86 0 1.00CFM 1CF
2012-01-24 09:37:13 95 0 1.00CFM 1CF
2012-01-24 09:38:11 85 0 1.00CFM 1CF
2012-01-24 09:39:14 78 0 1.00CFM 1CF
2012-01-24 09:40:11 74 0 1.00CFM 1CF
2012-01-24 09:41:14 87 0 1.00CFM 1CF
2012-01-24 09:42:11 97 0 1.00CFM 1CF
2012-01-24 09:43:14 74 0 1.00CFM 1CF
2012-01-24 09:44:11 93 0 1.00CFM 1CF
2012-01-24 09:45:09 81 0 1.00CFM 1CF
2012-01-24 09:46:12 84 0 1.00CFM 1CF
2012-01-24 09:47:11 86 0 1.00CFM 1CF
2012-01-24 09:48:13 92 0 1.00CFM 1CF
2012-01-24 09:49:11 90 0 1.00CFM 1CF
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
≥ 0.5 ≥ 5.0
Timestamp µm µm Flow Volume
2012-01-24 09:30:11 79 0 1.00CFM 1CF
2012-01-24 09:31:12 71 0 1.00CFM 1CF
2012-01-24 09:32:15 90 0 1.00CFM 1CF
2012-01-24 09:33:12 79 0 1.00CFM 1CF
2012-01-24 09:34:15 71 0 1.00CFM 1CF
2012-01-24 09:35:12 74 0 1.00CFM 1CF
2012-01-24 09:36:15 98 0 1.00CFM 1CF
2012-01-24 09:37:13 71 0 1.00CFM 1CF
2012-01-24 09:38:11 80 0 1.00CFM 1CF
2012-01-24 09:39:14 84 0 1.00CFM 1CF
2012-01-24 09:40:11 79 0 1.00CFM 1CF
2012-01-24 09:41:14 94 0 1.00CFM 1CF
2012-01-24 09:42:11 98 0 1.00CFM 1CF
2012-01-24 09:43:14 75 0 1.00CFM 1CF
2012-01-24 09:44:11 86 0 1.00CFM 1CF
2012-01-24 09:45:15 89 0 1.00CFM 1CF
2012-01-24 09:46:12 79 0 1.00CFM 1CF
2012-01-24 09:47:16 74 0 1.00CFM 1CF
2012-01-24 09:48:13 81 0 1.00CFM 1CF
2012-01-24 09:49:12 90 0 1.00CFM 1CF
2012-01-24 09:50:15 88 0 1.00CFM 1CF
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
≥ 0.5 ≥ 5.0
Timestamp µm µm Flow Volume
2012-01-24 09:30:06 204678 898 1.00CFM 1/60 CM
2012-01-24 09:30:41 204608 893 1.00CFM 1/60 CM
2012-01-24 09:31:12 204642 886 1.00CFM 1/60 CM
2012-01-24 09:31:49 204660 914 1.00CFM 1/60 CM
2012-01-24 09:32:25 204655 889 1.00CFM 1/60 CM
2012-01-24 09:32:56 204614 887 1.00CFM 1/60 CM
2012-01-24 09:33:33 204638 890 1.00CFM 1/60 CM
2012-01-24 09:34:09 204676 903 1.00CFM 1/60 CM
2012-01-24 09:34:46 204499 902 1.00CFM 1/60 CM
2012-01-24 09:35:17 204689 886 1.00CFM 1/60 CM
2012-01-24 09:35:54 204692 901 1.00CFM 1/60 CM
2012-01-24 09:36:30 204584 903 1.00CFM 1/60 CM
2012-01-24 09:37:02 204630 904 1.00CFM 1/60 CM
2012-01-24 09:37:40 204529 887 1.00CFM 1/60 CM
2012-01-24 09:38:16 204602 905 1.00CFM 1/60 CM
2012-01-24 09:38:48 204519 911 1.00CFM 1/60 CM
2012-01-24 09:39:24 204681 900 1.00CFM 1/60 CM
2012-01-24 09:40:01 204521 901 1.00CFM 1/60 CM
2012-01-24 09:40:32 204672 902 1.00CFM 1/60 CM
2012-01-24 09:41:09 204672 908 1.00CFM 1/60 CM
2012-01-24 09:41:45 204663 907 1.00CFM 1/60 CM
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
≥ 0.5 ≥ 5.0
Timestamp µm µm Flow Volume
2012-01-24 09:30:06 204573 897 1.00CFM 1/60 CM
2012-01-24 09:30:41 204527 914 1.00CFM 1/60 CM
2012-01-24 09:31:17 204657 898 1.00CFM 1/60 CM
2012-01-24 09:31:49 204564 912 1.00CFM 1/60 CM
2012-01-24 09:32:25 204604 912 1.00CFM 1/60 CM
2012-01-24 09:33:02 204490 912 1.00CFM 1/60 CM
2012-01-24 09:33:33 204562 913 1.00CFM 1/60 CM
2012-01-24 09:34:09 204508 889 1.00CFM 1/60 CM
2012-01-24 09:34:46 204655 885 1.00CFM 1/60 CM
2012-01-24 09:35:22 204525 900 1.00CFM 1/60 CM
2012-01-24 09:35:54 204700 903 1.00CFM 1/60 CM
2012-01-24 09:36:30 204497 899 1.00CFM 1/60 CM
2012-01-24 09:37:08 204694 905 1.00CFM 1/60 CM
2012-01-24 09:37:40 204605 901 1.00CFM 1/60 CM
2012-01-24 09:38:16 204542 907 1.00CFM 1/60 CM
2012-01-24 09:38:53 204678 910 1.00CFM 1/60 CM
2012-01-24 09:39:24 204613 896 1.00CFM 1/60 CM
2012-01-24 09:40:01 204652 893 1.00CFM 1/60 CM
2012-01-24 09:40:37 204564 909 1.00CFM 1/60 CM
2012-01-24 09:41:09 204681 885 1.00CFM 1/60 CM
2012-01-24 09:41:45 204708 889 1.00CFM 1/60 CM
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
≥ 0.5 ≥ 5.0
Timestamp µm µm Flow Volume
2012-01-24 09:30:11 99 0 1.00CFM 1CF
2012-01-24 09:31:12 91 0 1.00CFM 1CF
2012-01-24 09:32:09 84 0 1.00CFM 1CF
2012-01-24 09:33:12 85 0 1.00CFM 1CF
2012-01-24 09:34:09 74 0 1.00CFM 1CF
2012-01-24 09:35:12 86 0 1.00CFM 1CF
2012-01-24 09:36:15 86 0 1.00CFM 1CF
2012-01-24 09:37:13 95 0 1.00CFM 1CF
2012-01-24 09:38:11 85 0 1.00CFM 1CF
2012-01-24 09:39:14 78 0 1.00CFM 1CF
2012-01-24 09:40:11 74 0 1.00CFM 1CF
2012-01-24 09:41:14 87 0 1.00CFM 1CF
2012-01-24 09:42:11 97 0 1.00CFM 1CF
2012-01-24 09:43:14 74 0 1.00CFM 1CF
2012-01-24 09:44:11 93 0 1.00CFM 1CF
2012-01-24 09:45:09 81 0 1.00CFM 1CF
2012-01-24 09:46:12 84 0 1.00CFM 1CF
2012-01-24 09:47:11 86 0 1.00CFM 1CF
2012-01-24 09:48:13 92 0 1.00CFM 1CF
2012-01-24 09:49:11 90 0 1.00CFM 1CF
2012-01-24 09:50:10 83 0 1.00CFM 1CF
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
≥ 0.5 ≥ 5.0
Timestamp µm µm Flow Volume
2012-01-24 09:30:11 79 0 1.00CFM 1CF
2012-01-24 09:31:12 71 0 1.00CFM 1CF
2012-01-24 09:32:15 90 0 1.00CFM 1CF
2012-01-24 09:33:12 79 0 1.00CFM 1CF
2012-01-24 09:34:15 71 0 1.00CFM 1CF
2012-01-24 09:35:12 74 0 1.00CFM 1CF
2012-01-24 09:36:15 98 0 1.00CFM 1CF
2012-01-24 09:37:13 71 0 1.00CFM 1CF
2012-01-24 09:38:11 80 0 1.00CFM 1CF
2012-01-24 09:39:14 84 0 1.00CFM 1CF
2012-01-24 09:40:11 79 0 1.00CFM 1CF
2012-01-24 09:41:14 94 0 1.00CFM 1CF
2012-01-24 09:42:11 98 0 1.00CFM 1CF
2012-01-24 09:43:14 75 0 1.00CFM 1CF
2012-01-24 09:44:11 86 0 1.00CFM 1CF
2012-01-24 09:45:15 89 0 1.00CFM 1CF
2012-01-24 09:46:12 79 0 1.00CFM 1CF
2012-01-24 09:47:16 74 0 1.00CFM 1CF
2012-01-24 09:48:13 81 0 1.00CFM 1CF
2012-01-24 09:49:12 90 0 1.00CFM 1CF
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CF Maximum /CF Time of Maximum Minimum /CF Volume Alert Action Events # of samples
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Foot 5791.1 5796.7 2012-01-24 10:33:40 5790.5 242.5CF 0 0 0 412
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Foot 13.7 25.9 2012-01-24 09:55:08 25.1 242.5CF 0 0 0 412
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Foot 5801.3 5796.7 2012-01-24 09:48:13 5790.5 242.5CF 0 0 1 412
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Foot 12.6 25.9 2012-01-24 10:17:58 25.1 242.5CF 0 0 1 412
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Foot 84.9 99 2012-01-24 09:30:11 71 240CF 0 0 1 240
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Foot 0 0 2012-01-24 09:30:11 0 240CF 0 0 1 240
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Foot 84.8 99 2012-01-24 10:09:15 71 240CF 0 0 0 240
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Foot 0 0 2012-01-24 09:30:11 0 240CF 0 0 0 240
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events # of samples
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 204509.4 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Cubic Meter 204870.4 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Cubic Meter 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Cubic Meter 2998.1 3496.2 2012-01-24 09:30:11 2507.3 6.8CM 0 0 1 240
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 1 240
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Cubic Meter 2994.2 3496.2 2012-01-24 10:09:15 2507.3 6.8CM 0 0 0 240
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Cubic Meter 0 0 2012-01-24 09:30:11 0 6.8CM 0 0 0 240
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average Maximum Time of Maximum Minimum Volume Alert Action Events # of samples
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Total Count 204509 204707 2012-01-24 10:33:40 204490 6.9CM 0 0 0 412
Point 1 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Total Count 482.2 915 2012-01-24 09:55:08 885 6.9CM 0 0 0 412
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 0.5 µm Total Count 204870 204710 2012-01-24 09:48:13 204490 6.9CM 0 0 1 412
Point 2 CI-3100-21 CI-3100-21-1-4-6 Mystic Pharma 1.00CFM ≥ 5.0 µm Total Count 445.2 915 2012-01-24 10:17:58 885 6.9CM 0 0 1 412
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 0.5 µm Total Count 84.9 99 2012-01-24 09:30:11 71 240CF 0 0 1 240
Point 3 CI-3100-21 CI-3100-21-1-4-6 1.00CFM ≥ 5.0 µm Total Count 0 0 2012-01-24 09:30:11 0 240CF 0 0 1 240
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 0.5 µm Total Count 84.8 99 2012-01-24 10:09:15 71 240CF 0 0 0 240
Point 4 CI-3100-21 CI-3100 -21 S/N 079090 1.00CFM ≥ 5.0 µm Total Count 0 0 2012-01-24 09:30:11 0 240CF 0 0 0 240
Date: 1/26/12 .
Validation: DataPro 3 software
All graphs have the graph type and date created at the top, as well as the start and end times at the bottom of the printed graphs. These are headers and footers respectively but will
not appear on the following graphs due to the way the data is converted for this document. The graph types can be referenced in the validation documents “Reports and Graphs”
section using the Validation number on each graph. The start and end times for all graphs are the same as this example. EXAMPLE:
Date: 1/26/12 .
Cubic Foot
Date:
5787
5788
5789
5790
5791
5792
5793
5794
5795
5796
5797
5798
2012-01-24 09:30:41
2012-01-24 09:33:33
2012-01-24 09:37:02
2012-01-24 09:40:01
Completed By:
2012-01-24 09:42:58
2012-01-24 09:46:22
2012-01-24 09:48:44
1/26/12
2012-01-24 09:52:51
2012-01-24 09:56:17
.
2012-01-24 09:59:12
2012-01-24 10:02:10
2012-01-24 10:04:33
2012-01-24 10:07:25
2012-01-24 10:09:48
2012-01-24 10:12:41
2012-01-24 10:15:37
2012-01-24 10:19:05
2012-01-24 10:22:00
2012-01-24 10:26:03
2012-01-24 10:29:37
2012-01-24 10:33:07
2012-01-24 10:36:01
2012-01-24 10:39:31
2012-01-24 10:42:22
2012-01-24 10:45:17
2012-01-24 10:47:43
2012-01-24 10:51:09
2012-01-24 10:54:36
2012-01-24 10:57:03
2012-01-24 11:00:33
2012-01-24 11:58:13
2012-01-24 12:01:43
2012-01-24 12:04:41
2012-01-24 12:08:07
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:11:05
2012-01-24 12:13:58
2012-01-24 12:17:31
2012-01-24 12:20:24
2012-01-24 12:23:57
2012-01-24 12:27:28
2012-01-24 12:30:22
2012-01-24 12:33:16
2012-01-24 12:36:08
2012-01-24 12:39:08
2012-01-24 12:42:35
2012-01-24 12:46:05
2012-01-24 12:49:03
2012-01-24 12:51:55
2012-01-24 12:54:53
2012-01-24 12:58:18
2012-01-24 13:01:17
2012-01-24 13:04:11
2012-01-24 13:07:04
2012-01-24 13:09:59
2012-01-24 13:12:57
2012-01-24 13:15:52
2012-01-24 13:18:42
2012-01-24 13:22:16
2012-01-24 13:25:10
2012-01-24 13:28:08
2.1.1.1.1
Validation
Point 2 ≥ 0.5 µm
Point 1 ≥ 0.5 µm
Cubic Foot
Date:
24.6
24.8
25.2
25.4
25.6
25.8
25
26
2012-01-24 09:30:41
2012-01-24 09:34:09
2012-01-24 09:37:40
2012-01-24 09:41:09
Completed By:
2012-01-24 09:44:42
2012-01-24 09:48:08
1/26/12
2012-01-24 09:52:15
2012-01-24 09:56:17
.
2012-01-24 09:59:17
2012-01-24 10:02:48
2012-01-24 10:06:12
2012-01-24 10:08:38
2012-01-24 10:11:33
2012-01-24 10:15:31
2012-01-24 10:19:05
2012-01-24 10:22:36
2012-01-24 10:27:16
2012-01-24 10:31:22
2012-01-24 10:34:53
2012-01-24 10:38:23
2012-01-24 10:41:50
2012-01-24 10:45:17
2012-01-24 10:48:15
2012-01-24 10:52:19
2012-01-24 10:55:18
2012-01-24 10:59:20
2012-01-24 12:01:43
2012-01-24 12:05:14
2012-01-24 12:08:44
2012-01-24 12:12:13
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:16:18
2012-01-24 12:19:47
2012-01-24 12:23:20
2012-01-24 12:27:28
2012-01-24 12:30:53
2012-01-24 12:33:52
2012-01-24 12:37:22
2012-01-24 12:41:22
2012-01-24 12:45:27
2012-01-24 12:48:58
2012-01-24 12:51:55
2012-01-24 12:55:24
2012-01-24 12:58:54
2012-01-24 13:03:01
2012-01-24 13:05:57
2012-01-24 13:09:22
2012-01-24 13:12:52
2012-01-24 13:15:52
2012-01-24 13:18:47
2012-01-24 13:22:49
2012-01-24 13:26:18
2012-01-24 13:29:49
2.1.1.1.2
Validation
Point 2 ≥ 5.0 µm
Point 1 ≥ 5.0 µm
Cubic Foot
Date:
0
20
40
60
80
100
120
2012-01-24 09:31:12
2012-01-24 09:34:15
2012-01-24 09:39:14
2012-01-24 09:44:11
Completed By:
2012-01-24 09:47:16
2012-01-24 09:50:15
1/26/12
2012-01-24 09:54:16
2012-01-24 09:57:16
.
2012-01-24 10:02:10
2012-01-24 10:05:15
2012-01-24 10:09:09
2012-01-24 10:12:15
2012-01-24 10:16:14
2012-01-24 10:19:11
2012-01-24 10:22:10
2012-01-24 10:26:13
2012-01-24 10:29:16
2012-01-24 10:32:15
2012-01-24 10:36:11
2012-01-24 10:41:13
2012-01-24 10:45:12
2012-01-24 10:48:15
2012-01-24 10:52:14
2012-01-24 10:56:10
2012-01-24 10:59:15
2012-01-24 11:59:10
2012-01-24 12:02:15
2012-01-24 12:07:10
2012-01-24 12:10:13
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:13:16
2012-01-24 12:18:13
2012-01-24 12:22:13
2012-01-24 12:26:09
2012-01-24 12:29:14
2012-01-24 12:33:11
2012-01-24 12:35:16
2012-01-24 12:39:13
2012-01-24 12:44:14
2012-01-24 12:48:11
2012-01-24 12:52:16
2012-01-24 12:57:16
2012-01-24 13:01:11
2012-01-24 13:05:14
2012-01-24 13:10:14
2012-01-24 13:14:10
2012-01-24 13:17:12
2012-01-24 13:20:11
2012-01-24 13:23:09
2012-01-24 13:26:13
2012-01-24 13:29:12
2.1.1.2.1
Validation
Point 4 ≥ 0.5 µm
Point 3 ≥ 0.5 µm
Cubic Foot
Date:
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
0
1
2012-01-24 09:31:12
2012-01-24 09:34:09
2012-01-24 09:37:13
2012-01-24 09:41:14
2012-01-24 09:45:09
Completed By:
2012-01-24 09:47:16
2012-01-24 09:50:10
2012-01-24 09:53:13
1/26/12
2012-01-24 09:56:11
2012-01-24 09:58:14
.
2012-01-24 10:02:10
2012-01-24 10:05:10
2012-01-24 10:07:15
2012-01-24 10:10:14
2012-01-24 10:13:14
2012-01-24 10:16:14
2012-01-24 10:18:13
2012-01-24 10:21:11
2012-01-24 10:23:14
2012-01-24 10:27:11
2012-01-24 10:29:16
2012-01-24 10:32:10
2012-01-24 10:34:11
2012-01-24 10:38:12
2012-01-24 10:42:11
2012-01-24 10:45:12
2012-01-24 10:48:10
2012-01-24 10:51:10
2012-01-24 10:54:10
2012-01-24 10:56:15
2012-01-24 10:59:15
2012-01-24 11:56:12
2012-01-24 11:59:10
2012-01-24 12:02:09
2012-01-24 12:05:14
2012-01-24 12:08:13
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:11:10
2012-01-24 12:13:16
2012-01-24 12:17:15
2012-01-24 12:21:15
2012-01-24 12:24:14
2012-01-24 12:26:15
2012-01-24 12:29:14
2012-01-24 12:32:11
2012-01-24 12:34:14
2012-01-24 12:36:14
2012-01-24 12:40:11
2012-01-24 12:44:14
2012-01-24 12:47:13
2012-01-24 12:51:13
2012-01-24 12:54:15
2012-01-24 12:58:13
2012-01-24 13:01:11
2012-01-24 13:04:11
2012-01-24 13:08:12
2012-01-24 13:12:10
2012-01-24 13:14:15
2012-01-24 13:17:12
2012-01-24 13:19:14
2012-01-24 13:22:11
2012-01-24 13:24:12
2012-01-24 13:27:11
2012-01-24 13:29:12
2.1.1.2.2
Validation
Point 4 ≥ 5.0 µm
Point 3 ≥ 5.0 µm
Cubic Meter
Date:
204350
204400
204450
204500
204550
204600
204650
204700
204750
2012-01-24 09:30:41
2012-01-24 09:34:09
2012-01-24 09:37:40
Completed By:
2012-01-24 09:41:09
2012-01-24 09:44:42
2012-01-24 09:48:08
1/26/12
2012-01-24 09:52:15
.
2012-01-24 09:56:17
2012-01-24 09:59:17
2012-01-24 10:02:48
2012-01-24 10:06:12
2012-01-24 10:08:38
2012-01-24 10:11:33
2012-01-24 10:15:31
2012-01-24 10:19:05
2012-01-24 10:22:36
2012-01-24 10:27:16
2012-01-24 10:31:22
2012-01-24 10:34:53
2012-01-24 10:38:23
2012-01-24 10:41:50
2012-01-24 10:45:17
2012-01-24 10:48:15
2012-01-24 10:52:19
2012-01-24 10:55:18
2012-01-24 10:59:20
2012-01-24 12:01:43
2012-01-24 12:05:14
2012-01-24 12:08:44
2012-01-24 12:12:13
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:16:18
2012-01-24 12:19:47
2012-01-24 12:23:20
2012-01-24 12:27:28
2012-01-24 12:30:53
2012-01-24 12:33:52
2012-01-24 12:37:22
2012-01-24 12:41:22
2012-01-24 12:45:27
2012-01-24 12:48:58
2012-01-24 12:51:55
2012-01-24 12:55:24
2012-01-24 12:58:54
2012-01-24 13:03:01
2012-01-24 13:05:57
2012-01-24 13:09:22
2012-01-24 13:12:52
2012-01-24 13:15:52
2012-01-24 13:18:47
2012-01-24 13:22:49
2012-01-24 13:26:18
2012-01-24 13:29:49
2.1.2.1.1
Validation
Point 2 ≥ 0.5 µm
Point 1 ≥ 0.5 µm
Cubic Meter
Date:
870
875
880
885
890
895
900
905
910
915
920
2012-01-24 09:30:41
2012-01-24 09:34:09
2012-01-24 09:37:40
2012-01-24 09:41:09
Completed By:
2012-01-24 09:44:42
2012-01-24 09:48:08
2012-01-24 09:52:15
1/26/12
2012-01-24 09:56:17
.
2012-01-24 09:59:17
2012-01-24 10:02:48
2012-01-24 10:06:12
2012-01-24 10:08:38
2012-01-24 10:11:33
2012-01-24 10:15:31
2012-01-24 10:19:05
2012-01-24 10:22:36
2012-01-24 10:27:16
2012-01-24 10:31:22
2012-01-24 10:34:53
2012-01-24 10:38:23
2012-01-24 10:41:50
2012-01-24 10:45:17
2012-01-24 10:48:15
2012-01-24 10:52:19
2012-01-24 10:55:18
2012-01-24 10:59:20
2012-01-24 11:02:49
2012-01-24 12:01:43
2012-01-24 12:05:14
2012-01-24 12:08:44
2012-01-24 12:12:13
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:16:18
2012-01-24 12:19:47
2012-01-24 12:23:20
2012-01-24 12:27:28
2012-01-24 12:30:53
2012-01-24 12:33:52
2012-01-24 12:37:22
2012-01-24 12:41:22
2012-01-24 12:45:27
2012-01-24 12:48:58
2012-01-24 12:51:55
2012-01-24 12:55:24
2012-01-24 12:58:54
2012-01-24 13:03:01
2012-01-24 13:05:57
2012-01-24 13:09:22
2012-01-24 13:12:52
2012-01-24 13:15:52
2012-01-24 13:18:47
2012-01-24 13:22:49
2012-01-24 13:26:18
2012-01-24 13:29:49
2.1.2.1.2
Validation
Point 2 ≥ 5.0 µm
Point 1 ≥ 5.0 µm
Cubic Meter
Date:
1000
1500
2000
2500
3000
3500
4000
0
500
2012-01-24 09:31:12
2012-01-24 09:34:15
2012-01-24 09:39:14
Completed By:
2012-01-24 09:44:11
2012-01-24 09:47:16
2012-01-24 09:50:15
1/26/12
2012-01-24 09:54:16
.
2012-01-24 09:57:16
2012-01-24 10:02:10
2012-01-24 10:05:15
2012-01-24 10:09:09
2012-01-24 10:12:15
2012-01-24 10:16:14
2012-01-24 10:19:11
2012-01-24 10:22:10
2012-01-24 10:26:13
2012-01-24 10:29:16
2012-01-24 10:32:15
2012-01-24 10:36:11
2012-01-24 10:41:13
2012-01-24 10:45:12
2012-01-24 10:48:15
2012-01-24 10:52:14
2012-01-24 10:56:10
2012-01-24 10:59:15
2012-01-24 11:59:10
2012-01-24 12:02:15
2012-01-24 12:07:10
2012-01-24 12:10:13
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:13:16
2012-01-24 12:18:13
2012-01-24 12:22:13
2012-01-24 12:26:09
2012-01-24 12:29:14
2012-01-24 12:33:11
2012-01-24 12:35:16
2012-01-24 12:39:13
2012-01-24 12:44:14
2012-01-24 12:48:11
2012-01-24 12:52:16
2012-01-24 12:57:16
2012-01-24 13:01:11
2012-01-24 13:05:14
2012-01-24 13:10:14
2012-01-24 13:14:10
2012-01-24 13:17:12
2012-01-24 13:20:11
2012-01-24 13:23:09
2012-01-24 13:26:13
2012-01-24 13:29:12
2.1.2.2.1
Validation
Point 4 ≥ 0.5 µm
Point 3 ≥ 0.5 µm
Cubic Meter
Date:
0.1
0.2
0.3
0.4
0.5
0.6
0.8
0.9
0.7
0
1
2012-01-24 09:31:12
2012-01-24 09:34:15
2012-01-24 09:39:14
2012-01-24 09:44:11
Completed By:
2012-01-24 09:47:16
2012-01-24 09:50:15
1/26/12
2012-01-24 09:54:16
2012-01-24 09:57:16
.
2012-01-24 10:02:10
2012-01-24 10:05:15
2012-01-24 10:09:09
2012-01-24 10:12:15
2012-01-24 10:16:14
2012-01-24 10:19:11
2012-01-24 10:22:10
2012-01-24 10:26:13
2012-01-24 10:29:16
2012-01-24 10:32:15
2012-01-24 10:36:11
2012-01-24 10:41:13
2012-01-24 10:45:12
2012-01-24 10:48:15
2012-01-24 10:52:14
2012-01-24 10:56:10
2012-01-24 10:59:15
2012-01-24 11:03:10
2012-01-24 11:59:10
2012-01-24 12:02:15
2012-01-24 12:07:10
2012-01-24 12:10:13
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:13:16
2012-01-24 12:18:13
2012-01-24 12:22:13
2012-01-24 12:26:09
2012-01-24 12:29:14
2012-01-24 12:33:11
2012-01-24 12:35:16
2012-01-24 12:39:13
2012-01-24 12:44:14
2012-01-24 12:48:11
2012-01-24 12:52:16
2012-01-24 12:57:16
2012-01-24 13:01:11
2012-01-24 13:05:14
2012-01-24 13:10:14
2012-01-24 13:14:10
2012-01-24 13:17:12
2012-01-24 13:20:11
2012-01-24 13:23:09
2012-01-24 13:26:13
2012-01-24 13:29:12
2.1.2.2.2
Validation
Point 4 ≥ 5.0 µm
Point 3 ≥ 5.0 µm
Total Count
Date:
3150
3200
3250
3300
3350
3400
3450
3500
3550
2012-01-24 09:30:41
2012-01-24 09:34:09
2012-01-24 09:37:40
Completed By:
2012-01-24 09:41:09
2012-01-24 09:44:42
2012-01-24 09:48:08
1/26/12
2012-01-24 09:52:15
2012-01-24 09:56:17
.
2012-01-24 09:59:17
2012-01-24 10:02:48
2012-01-24 10:06:12
2012-01-24 10:08:38
2012-01-24 10:11:33
2012-01-24 10:15:31
2012-01-24 10:19:05
2012-01-24 10:22:36
2012-01-24 10:27:16
2012-01-24 10:31:22
2012-01-24 10:34:53
2012-01-24 10:38:23
2012-01-24 10:41:50
2012-01-24 10:45:17
2012-01-24 10:48:15
2012-01-24 10:52:19
2012-01-24 10:55:18
2012-01-24 10:59:20
2012-01-24 12:01:43
2012-01-24 12:05:14
2012-01-24 12:08:44
2012-01-24 12:12:13
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:16:18
2012-01-24 12:19:47
2012-01-24 12:23:20
2012-01-24 12:27:28
2012-01-24 12:30:53
2012-01-24 12:33:52
2012-01-24 12:37:22
2012-01-24 12:41:22
2012-01-24 12:45:27
2012-01-24 12:48:58
2012-01-24 12:51:55
2012-01-24 12:55:24
2012-01-24 12:58:54
2012-01-24 13:03:01
2012-01-24 13:05:57
2012-01-24 13:09:22
2012-01-24 13:12:52
2012-01-24 13:15:52
2012-01-24 13:18:47
2012-01-24 13:22:49
2012-01-24 13:26:18
2012-01-24 13:29:49
2.1.3.1.1
Validation
Point 2 ≥ 0.5 µm
Point 1 ≥ 0.5 µm
Total Count
Date:
0
5
10
15
20
25
30
35
2012-01-24 09:30:41
2012-01-24 09:34:09
2012-01-24 09:37:40
2012-01-24 09:41:09
Completed By:
2012-01-24 09:44:42
2012-01-24 09:48:08
2012-01-24 09:52:15
1/26/12
2012-01-24 09:56:17
.
2012-01-24 09:59:17
2012-01-24 10:02:48
2012-01-24 10:06:12
2012-01-24 10:08:38
2012-01-24 10:11:33
2012-01-24 10:15:31
2012-01-24 10:19:05
2012-01-24 10:22:36
2012-01-24 10:27:16
2012-01-24 10:31:22
2012-01-24 10:34:53
2012-01-24 10:38:23
2012-01-24 10:41:50
2012-01-24 10:45:17
2012-01-24 10:48:15
2012-01-24 10:52:19
2012-01-24 10:55:18
2012-01-24 10:59:20
2012-01-24 11:02:49
2012-01-24 12:01:43
2012-01-24 12:05:14
2012-01-24 12:08:44
2012-01-24 12:12:13
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:16:18
2012-01-24 12:19:47
2012-01-24 12:23:20
2012-01-24 12:27:28
2012-01-24 12:30:53
2012-01-24 12:33:52
2012-01-24 12:37:22
2012-01-24 12:41:22
2012-01-24 12:45:27
2012-01-24 12:48:58
2012-01-24 12:51:55
2012-01-24 12:55:24
2012-01-24 12:58:54
2012-01-24 13:03:01
2012-01-24 13:05:57
2012-01-24 13:09:22
2012-01-24 13:12:52
2012-01-24 13:15:52
2012-01-24 13:18:47
2012-01-24 13:22:49
2012-01-24 13:26:18
2012-01-24 13:29:49
2.1.3.1.2
Validation
Point 2 ≥ 5.0 µm
Point 1 ≥ 5.0 µm
Total Count
Date:
0
20
40
60
80
100
120
2012-01-24 09:31:12
2012-01-24 09:34:15
2012-01-24 09:39:14
2012-01-24 09:44:11
Completed By:
2012-01-24 09:47:16
2012-01-24 09:50:15
1/26/12
2012-01-24 09:54:16
2012-01-24 09:57:16
.
2012-01-24 10:02:10
2012-01-24 10:05:15
2012-01-24 10:09:09
2012-01-24 10:12:15
2012-01-24 10:16:14
2012-01-24 10:19:11
2012-01-24 10:22:10
2012-01-24 10:26:13
2012-01-24 10:29:16
2012-01-24 10:32:15
2012-01-24 10:36:11
2012-01-24 10:41:13
2012-01-24 10:45:12
2012-01-24 10:48:15
2012-01-24 10:52:14
2012-01-24 10:56:10
2012-01-24 10:59:15
2012-01-24 11:59:10
2012-01-24 12:02:15
2012-01-24 12:07:10
2012-01-24 12:10:13
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:13:16
2012-01-24 12:18:13
2012-01-24 12:22:13
2012-01-24 12:26:09
2012-01-24 12:29:14
2012-01-24 12:33:11
2012-01-24 12:35:16
2012-01-24 12:39:13
2012-01-24 12:44:14
2012-01-24 12:48:11
2012-01-24 12:52:16
2012-01-24 12:57:16
2012-01-24 13:01:11
2012-01-24 13:05:14
2012-01-24 13:10:14
2012-01-24 13:14:10
2012-01-24 13:17:12
2012-01-24 13:20:11
2012-01-24 13:23:09
2012-01-24 13:26:13
2012-01-24 13:29:12
2.1.3.2.1
Validation
Point 4 ≥ 0.5 µm
Point 3 ≥ 0.5 µm
Total Count
Date:
0.1
0.2
0.3
0.4
0.5
0.6
0.8
0.9
0.7
0
1
2012-01-24 09:31:12
2012-01-24 09:34:15
2012-01-24 09:39:14
2012-01-24 09:44:11
Completed By:
2012-01-24 09:47:16
2012-01-24 09:50:15
1/26/12
2012-01-24 09:54:16
2012-01-24 09:57:16
.
2012-01-24 10:02:10
2012-01-24 10:05:15
2012-01-24 10:09:09
2012-01-24 10:12:15
2012-01-24 10:16:14
2012-01-24 10:19:11
2012-01-24 10:22:10
2012-01-24 10:26:13
2012-01-24 10:29:16
2012-01-24 10:32:15
2012-01-24 10:36:11
2012-01-24 10:41:13
2012-01-24 10:45:12
2012-01-24 10:48:15
2012-01-24 10:52:14
2012-01-24 10:56:10
2012-01-24 10:59:15
2012-01-24 11:03:10
2012-01-24 11:59:10
2012-01-24 12:02:15
2012-01-24 12:07:10
2012-01-24 12:10:13
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:13:16
2012-01-24 12:18:13
2012-01-24 12:22:13
2012-01-24 12:26:09
2012-01-24 12:29:14
2012-01-24 12:33:11
2012-01-24 12:35:16
2012-01-24 12:39:13
2012-01-24 12:44:14
2012-01-24 12:48:11
2012-01-24 12:52:16
2012-01-24 12:57:16
2012-01-24 13:01:11
2012-01-24 13:05:14
2012-01-24 13:10:14
2012-01-24 13:14:10
2012-01-24 13:17:12
2012-01-24 13:20:11
2012-01-24 13:23:09
2012-01-24 13:26:13
2012-01-24 13:29:12
2.1.3.2.2
Validation
Point 4 ≥ 5.0 µm
Point 3 ≥ 5.0 µm
Cubic Meter
Date:
1000
10000
1
100000
10
1000000
100
2012-01-24 09:30:11
2012-01-24 09:34:09
Completed By:
2012-01-24 09:37:40
2012-01-24 09:41:14
2012-01-24 09:45:09
1/26/12
2012-01-24 09:48:13
2012-01-24 09:52:15
.
2012-01-24 09:56:12
2012-01-24 09:59:17
2012-01-24 10:03:14
2012-01-24 10:06:17
2012-01-24 10:09:15
2012-01-24 10:12:15
2012-01-24 10:16:09
2012-01-24 10:19:06
2012-01-24 10:22:15
2012-01-24 10:26:39
2012-01-24 10:30:13
2012-01-24 10:33:14
2012-01-24 10:37:09
2012-01-24 10:40:40
2012-01-24 10:44:08
2012-01-24 10:47:12
2012-01-24 10:51:09
2012-01-24 10:54:15
2012-01-24 10:57:34
2012-01-24 12:03:33
2012-01-24 12:07:36
2012-01-24 12:11:10
2012-01-24 12:14:38
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:18:13
2012-01-24 12:22:13
2012-01-24 12:26:14
2012-01-24 12:29:14
2012-01-24 12:32:45
2012-01-24 12:35:37
2012-01-24 12:39:08
2012-01-24 12:43:12
2012-01-24 12:46:42
2012-01-24 12:51:13
2012-01-24 12:54:15
2012-01-24 12:58:18
2012-01-24 13:01:49
2012-01-24 13:05:19
2012-01-24 13:09:22
2012-01-24 13:12:57
2012-01-24 13:16:23
2012-01-24 13:19:14
2012-01-24 13:22:49
2012-01-24 13:26:13
2.1.4.1.1.1
Validation
2012-01-24 13:29:49
*Alert
*Alarm
Point 4 ≥ 0.5 µm
Point 3 ≥ 0.5 µm
Point 2 ≥ 0.5 µm
Point 1 ≥ 0.5 µm
Cubic Meter
Date:
1000
10000
1
10
100
2012-01-24 09:30:11
2012-01-24 09:34:09
2012-01-24 09:37:40
Completed By:
2012-01-24 09:41:14
2012-01-24 09:45:09
2012-01-24 09:48:13
1/26/12
2012-01-24 09:52:15
2012-01-24 09:56:12
.
2012-01-24 09:59:17
2012-01-24 10:03:14
2012-01-24 10:06:17
2012-01-24 10:09:15
2012-01-24 10:12:15
2012-01-24 10:16:09
2012-01-24 10:19:06
2012-01-24 10:22:15
2012-01-24 10:26:39
2012-01-24 10:30:13
2012-01-24 10:33:14
2012-01-24 10:37:09
2012-01-24 10:40:40
2012-01-24 10:44:08
2012-01-24 10:47:12
2012-01-24 10:51:09
2012-01-24 10:54:15
2012-01-24 10:57:34
2012-01-24 11:01:15
2012-01-24 12:03:33
2012-01-24 12:07:36
2012-01-24 12:11:10
2012-01-24 12:14:38
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:18:13
2012-01-24 12:22:13
2012-01-24 12:26:14
2012-01-24 12:29:14
2012-01-24 12:32:45
2012-01-24 12:35:37
2012-01-24 12:39:08
2012-01-24 12:43:12
2012-01-24 12:46:42
2012-01-24 12:51:13
2012-01-24 12:54:15
2012-01-24 12:58:18
2012-01-24 13:01:49
2012-01-24 13:05:19
2012-01-24 13:09:22
2012-01-24 13:12:57
2012-01-24 13:16:23
2012-01-24 13:19:14
2012-01-24 13:22:49
2012-01-24 13:26:13
2.1.4.1.1.2
Validation
2012-01-24 13:29:49
*Alert
*Alarm
Point 4 ≥ 5.0 µm
Point 3 ≥ 5.0 µm
Point 2 ≥ 5.0 µm
Point 1 ≥ 5.0 µm
Total Count
Date:
1000
10000
1
10
100
2012-01-24 09:30:11
2012-01-24 09:33:33
2012-01-24 09:37:08
2012-01-24 09:40:32
Completed By:
2012-01-24 09:44:06
2012-01-24 09:47:11
2012-01-24 09:50:10
1/26/12
2012-01-24 09:54:10
.
2012-01-24 09:57:11
2012-01-24 10:00:25
2012-01-24 10:03:51
2012-01-24 10:06:48
2012-01-24 10:09:15
2012-01-24 10:12:10
2012-01-24 10:15:31
2012-01-24 10:18:13
2012-01-24 10:21:16
2012-01-24 10:24:53
2012-01-24 10:28:24
2012-01-24 10:31:54
2012-01-24 10:34:47
2012-01-24 10:38:12
2012-01-24 10:41:49
2012-01-24 10:44:14
2012-01-24 10:47:12
2012-01-24 10:50:38
2012-01-24 10:54:05
2012-01-24 10:56:57
2012-01-24 11:00:27
2012-01-24 12:00:34
2012-01-24 12:04:04
2012-01-24 12:07:36
2012-01-24 12:11:05
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:13:58
2012-01-24 12:17:31
2012-01-24 12:21:15
2012-01-24 12:25:06
2012-01-24 12:28:00
2012-01-24 12:30:53
2012-01-24 12:33:52
2012-01-24 12:36:14
2012-01-24 12:39:39
2012-01-24 12:43:43
2012-01-24 12:46:42
2012-01-24 12:50:47
2012-01-24 12:53:38
2012-01-24 12:57:16
2012-01-24 13:00:14
2012-01-24 13:03:38
2012-01-24 13:07:15
2012-01-24 13:11:07
2012-01-24 13:14:10
2012-01-24 13:17:11
2012-01-24 13:19:55
2012-01-24 13:23:09
2012-01-24 13:26:13
2.1.4.3.1.1
Validation
2012-01-24 13:29:18
*Alert
*Alarm
Point 4 ≥ 0.5 µm
Point 3 ≥ 0.5 µm
Point 2 ≥ 0.5 µm
Point 1 ≥ 0.5 µm
Total Count
Date:
1000
10000
1
10
100
2012-01-24 09:30:11
2012-01-24 09:33:33
2012-01-24 09:37:08
2012-01-24 09:40:32
Completed By:
2012-01-24 09:44:06
2012-01-24 09:47:11
2012-01-24 09:50:10
1/26/12
2012-01-24 09:54:10
.
2012-01-24 09:57:11
2012-01-24 10:00:25
2012-01-24 10:03:51
2012-01-24 10:06:48
2012-01-24 10:09:15
2012-01-24 10:12:10
2012-01-24 10:15:31
2012-01-24 10:18:13
2012-01-24 10:21:16
2012-01-24 10:24:53
2012-01-24 10:28:24
2012-01-24 10:31:54
2012-01-24 10:34:47
2012-01-24 10:38:12
2012-01-24 10:41:49
2012-01-24 10:44:14
2012-01-24 10:47:12
2012-01-24 10:50:38
2012-01-24 10:54:05
2012-01-24 10:56:57
2012-01-24 11:00:27
2012-01-24 12:00:34
2012-01-24 12:04:04
2012-01-24 12:07:36
2012-01-24 12:11:05
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:13:58
2012-01-24 12:17:31
2012-01-24 12:21:15
2012-01-24 12:25:06
2012-01-24 12:28:00
2012-01-24 12:30:53
2012-01-24 12:33:52
2012-01-24 12:36:14
2012-01-24 12:39:39
2012-01-24 12:43:43
2012-01-24 12:46:42
2012-01-24 12:50:47
2012-01-24 12:53:38
2012-01-24 12:57:16
2012-01-24 13:00:14
2012-01-24 13:03:38
2012-01-24 13:07:15
2012-01-24 13:11:07
2012-01-24 13:14:10
2012-01-24 13:17:11
2012-01-24 13:19:55
2012-01-24 13:23:09
2012-01-24 13:26:13
2.1.4.3.1.2
Validation
2012-01-24 13:29:18
*Alert
*Alarm
Point 4 ≥ 5.0 µm
Point 3 ≥ 5.0 µm
Point 2 ≥ 5.0 µm
Point 1 ≥ 5.0 µm
Cubic Foot
Date:
5792.5
5793.5
5794.5
5795.5
5796.5
5793
5795
5796
5797
5794
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:15:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:15:00
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.1.1.1
Validation
Point 2 ≥ 0.5 µm
Point 1 ≥ 0.5 µm
Cubic Foot
Date:
25.3
25.4
25.5
25.6
25.7
25.8
25.9
26
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
P/N 03036901
DataPro 3 V1.20, Validation
2012-01-24 12:15:00
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.1.1.2
Validation
Point 2 ≥ 5.0 µm
Point 1 ≥ 5.0 µm
Cubic Foot
Date:
0
20
40
60
80
100
120
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.1.2.1
Validation
Point 4 ≥ 0.5 µm
Point 3 ≥ 0.5 µm
Cubic Foot
Date:
0.1
0.2
0.3
0.4
0.5
0.6
0.8
0.9
0.7
0
1
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
2012-01-24 09:55:00
1/26/12
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.1.2.2
Validation
Point 4 ≥ 5.0 µm
Point 3 ≥ 5.0 µm
Cubic Meter
Date:
204560
204580
204600
204620
204640
204660
204680
204700
204720
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:15:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:15:00
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.2.1.1
Validation
Point 2 ≥ 0.5 µm
Point 1 ≥ 0.5 µm
Cubic Meter
Date:
895
900
905
910
915
920
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.2.1.2
Validation
Point 2 ≥ 5.0 µm
Point 1 ≥ 5.0 µm
Cubic Meter
Date:
1000
1500
2000
2500
3000
3500
4000
0
500
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:15:00
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.2.2.1
Validation
Point 4 ≥ 0.5 µm
Point 3 ≥ 0.5 µm
Cubic Meter
Date:
0.1
0.2
0.3
0.4
0.5
0.6
0.8
0.9
0.7
0
1
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
2012-01-24 09:55:00
1/26/12
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.2.2.2
Validation
Point 4 ≥ 5.0 µm
Point 3 ≥ 5.0 µm
Total Count
Date:
3380
3400
3420
3440
3460
3480
3500
3520
3540
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:15:00
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.3.1.1
Validation
Point 2 ≥ 0.5 µm
Point 1 ≥ 0.5 µm
Total Count
Date:
0
5
10
15
20
25
30
35
2012-01-24 09:35:00
2012-01-24 09:45:00
Completed By:
2012-01-24 09:55:00
1/26/12
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.3.1.2
Validation
Point 2 ≥ 5.0 µm
Point 1 ≥ 5.0 µm
Total Count
Date:
0
20
40
60
80
100
120
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.3.2.1
Validation
Point 4 ≥ 0.5 µm
Point 3 ≥ 0.5 µm
Total Count
Date:
0.1
0.2
0.3
0.4
0.5
0.6
0.8
0.9
0.7
0
1
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
2012-01-24 09:55:00
1/26/12
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.3.2.2
Validation
Point 4 ≥ 5.0 µm
Point 3 ≥ 5.0 µm
Cubic Meter
Date:
1000
10000
1
100000
10
1000000
100
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:15:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.4.1.1.1
Validation
*Alert
*Alarm
Point 4 ≥ 0.5 µm
Point 3 ≥ 0.5 µm
Point 2 ≥ 0.5 µm
Point 1 ≥ 0.5 µm
Cubic Meter
Date:
1000
10000
1
10
100
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.4.1.1.2
Validation
*Alert
*Alarm
Point 4 ≥ 5.0 µm
Point 3 ≥ 5.0 µm
Point 2 ≥ 5.0 µm
Point 1 ≥ 5.0 µm
Cubic Foot
Date:
1000
10000
1
10
100
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.4.2.1.1
Validation
*Alert
*Alarm
Point 4 ≥ 0.5 µm
Point 3 ≥ 0.5 µm
Point 2 ≥ 0.5 µm
Point 1 ≥ 0.5 µm
Cubic Foot
Date:
1000
10000
1
10
100
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.4.2.1.2
Validation
*Alert
*Alarm
Point 4 ≥ 5.0 µm
Point 3 ≥ 5.0 µm
Point 2 ≥ 5.0 µm
Point 1 ≥ 5.0 µm
Total Count
Date:
1000
10000
1
10
100
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.4.3.1.1
Validation
*Alert
*Alarm
Point 4 ≥ 0.5 µm
Point 3 ≥ 0.5 µm
Point 2 ≥ 0.5 µm
Point 1 ≥ 0.5 µm
Total Count
Date:
1000
10000
1
10
100
2012-01-24 09:35:00
Completed By:
2012-01-24 09:45:00
1/26/12
2012-01-24 09:55:00
.
2012-01-24 10:05:00
2012-01-24 10:15:00
2012-01-24 10:25:00
2012-01-24 10:35:00
2012-01-24 10:45:00
2012-01-24 10:55:00
2012-01-24 11:05:00
2012-01-24 11:25:00
2012-01-24 11:35:00
2012-01-24 11:45:00
2012-01-24 11:55:00
Ampule Filling Machine 1
2012-01-24 12:05:00
2012-01-24 12:15:00
P/N 03036901
DataPro 3 V1.20, Validation
Validation: DataPro 3 software
2012-01-24 12:25:00
2012-01-24 12:35:00
2012-01-24 12:45:00
2012-01-24 12:55:00
2012-01-24 13:05:00
2012-01-24 13:15:00
2012-01-24 13:25:00
2.2.4.3.1.2
Validation
*Alert
*Alarm
Point 4 ≥ 5.0 µm
Point 3 ≥ 5.0 µm
Point 2 ≥ 5.0 µm
Point 1 ≥ 5.0 µm
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events Result
Point 5 CI-3100-21 CI-3100-21-2-4-6 S/N 000186 1.00 ≥ 0.5 µm Cubic Meter 60 60 2012-01-27 09:50:52 1 1 CM 0 0 2 Pass
Point 5 CI-3100-21 CI-3100-21-2-4-6 S/N 000186 1.00 ≥ 5.0 µm Cubic Meter 0 0 2012-01-27 09:16:47 0 1 CM 0 0 2 Pass
Point 6 CI-3100-21 CI-3100-21-2-4-6 S/N 000196 1.00 ≥ 0.5 µm Cubic Meter 3540 3540 2012-01-27 09:51:36 59 1 CM 0 0 0 Fail
Point 6 CI-3100-21 CI-3100-21-2-4-6 S/N 000196 1.00 ≥ 5.0 µm Cubic Meter 60 60 2012-01-27 09:51:36 1 1 CM 0 0 0 Fail
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events Result
Point 5 CI-3100-21 CI-3100-21-2-4-6 S/N 000186 1.00 ≥ 0.5 µm Cubic Meter 342000 342000 2012-01-27 12:14:48 5700 1 CM 0 0 0 Pass
Point 5 CI-3100-21 CI-3100-21-2-4-6 S/N 000186 1.00 ≥ 5.0 µm Cubic Meter 2820 2820 2012-01-27 12:14:48 47 1 CM 0 0 0 Pass
Point 6 CI-3100-21 CI-3100-21-2-4-6 S/N 000196 1.00 ≥ 0.5 µm Cubic Meter 354000 354000 2012-01-27 12:14:48 5900 1 CM 0 0 0 Fail
Point 6 CI-3100-21 CI-3100-21-2-4-6 S/N 000196 1.00 ≥ 5.0 µm Cubic Meter 2940 2940 2012-01-27 12:14:48 49 1 CM 0 0 0 Fail
Date: 1/26/12 .
Validation: DataPro 3 software
Location Sensor Unit Id Flow Average Size Count Mode Average /CM Maximum /CM Time of Maximum Minimum /CM Volume Alert Action Events Result
Point 5 CI-3100-21 CI-3100-21-2-4-6 S/N 000186 1.00 ≥ 0.5 µm Cubic Meter 342000 342000 2012-01-27 13:13:42 5700 1 CM 0 0 0 Pass
Point 5 CI-3100-21 CI-3100-21-2-4-6 S/N 000186 1.00 ≥ 5.0 µm Cubic Meter 28200 28200 2012-01-27 13:13:42 470 1 CM 0 0 0 Pass
Point 6 CI-3100-21 CI-3100-21-2-4-6 S/N 000196 1.00 ≥ 0.5 µm Cubic Meter 3540000 3540000 2012-01-27 13:13:42 59000 1 CM 0 0 0 Fail
Point 6 CI-3100-21 CI-3100-21-2-4-6 S/N 000196 1.00 ≥ 5.0 µm Cubic Meter 29400 29400 2012-01-27 13:13:42 490 1 CM 0 0 0 Fail
Date: 1/26/12 .
Validation: DataPro 3 software
CLiMET DataPro 3
Supplemental Validation
of Version 1.21
Notices
Copyright © 2003-2012 by CLiMET Instruments Company. All rights reserved.
Complying with applicable copyright laws, no part of this document may be reproduced, stored in or introduced to any retrieval system, or
retransmitted in any form or by any means (electronic, mechanical, photocopying, recording, or otherwise), or for any purpose, without the prior
written permission of CLiMET Instruments Company. Contact CLiMET Instruments to request written permission.
CLiMET Instruments Company may make improvements and/or changes in the product(s) and/or the program(s) described in this document at
any time without notice; these improvements and/or changes will be incorporated in new versions of the documentation. Unless otherwise noted,
the example companies, organizations, products, people and events depicted herein are fictitious and no association with any real company,
organization, product, person or event is intended or should be inferred.
Although every precaution has been taken in the preparation of this documentation, the author assumes no responsibility for technical
inaccuracies, typographical errors, or damages resulting from the use of the information contained herein.
The following paragraph does not apply to a country where such provisions are inconsistent with local law:
CLIMET INSTRUMENTS COMPANY PROVIDES THIS DOCUMENTATION “AS IS” WITHOUT WARRANTY OF ANY KIND,
EITHER EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF NON -INFRINGEMENT,
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Some states do not allow a disclaimer of express or implied
warranties in certain transactions, therefore, this statement may not apply to you.
Trademarks
CLiMET is a trademark of CLiMET Instruments Company in the United States and/or other countries. Microsoft, Windows, MSDE, SQL
Server 2005 Express, Visual Basic .NET, Visual Studios .NET and Excel are trademarks of Microsoft Corporation in the United States and/or
other countries. VeriSign is a trademark of VeriSign in the United States and/or other countries. RSA and MD5 are trademarks of RSA Security
Inc. in the United States and/or other countries. The names of actual companies and products mentioned herein may be the tra demarks of their
respective owners.
Supplementary Features
Note indicates notes related to the nearby text.
Tip indicates tips related to the nearby text.
Typographical Conventions
Bold indicates important words, or names of buttons, menus, and windows.
Italic indicates new terms, or terms in glossary.
indicates a procedure in the table of contents.
§ indicates reference to a section number.
Table of Contents
1 Introduction ....................................................................................................................... 5
2 System Requirements & Equipment ................................................................................. 5
3 IQ: Installation Qualification ............................................................................................ 5
3.1 Scope ................................................................................................................. 5
3.2 Peripherals Test .................................................................................................. 5
1 Introduction
CLiMET Instruments Company designed, constructed and validated
CLiMET DataPro 3 Version 1.21 at the CLiMET facility in Redlands,
California. The validation method described herein is consistent with
CLiMET’s “SOFTWARE DEVELOPMENT, DOCUMENTATION AND
REVISION STANDARDS” policy (Revision 3.0 – 03/04/1999).
The purpose of this validation document is to provide documented
evidence that DataPro will correctly install and consistently operate in
accordance with the procedures outlined in the DataPro manual. This
protocol describes all of the conditions and tests performed for this
validation.
The manual describes DataPro. The validation system has inputs and
outputs. The inputs to the validation system are particle samples and
the DataPro user interface. The outputs of the system are printouts,
databases and the data recorded in this document.
F1.1.0, F1.1.1,
U1.1.0 Software Package N N N 4.5
F1.2.0
Dropped XP from
U1.2.0 minimum N N N F1.4.0 4.5
requirements
Purpose Observe that versions have been updated and that XP is removed from
minimum requirements.
Setup Use the manual to determine the correct version numbering for the
software package, the client and DAS.
Success Yes No
Purpose Test that Overview correctly shows “Bad Sensor” when blocked at start
up.
Procedure Generate the following events by blocking the air flow and starting the
unit. Leave the air flow blocked for 2 minutes.
Bad Sensor
Make sure Overview and View Alarms tabs show the Bad Sensor
event.
Test on XP
Text on Vista
Test on Windows 7
Once a Bad Sensor alarm is thrown, put the counter into standby,
remove the restricted airflow, and start the counter again.
The Overview and View Alarms tab does not display Bad Sensor
recovery. This type of error must be manually resolved by the
operator at the physical location to prevent unit damage. Cycling the
unit into a Standby state and then Active again forces this interaction.
The unit will resume counting if powered off and on after a flow
obstruction is cleared however it will take a few minutes to show on
the over view. To expedite this process put the counter into Standby
state and only setting it to Active after the obstruction is cleared.
Success Yes No