OPERATION MANUAL
Thank you for choosing the EndoPAL (Powerful And Long-lasting) cordless battery-
powered handpiece. The EndoPAL provides mechanical assistance to complete root
canal therapy quicker and with less fatigue. It is designed to work in conjunction
with rotary nickel-titanium root canal instruments. Be certain that the specifications
of the EndoPAL (Instructions for Use, Page 3) are compatible with the instrument
manufacturer's requirements before using.
Existence of a hazard that could result in death or serious injury, if the safety instructions
Danger are not followed.
Existence of a hazard that could result in bodily injury or damage of the device, if the
Warning safety instructions are not followed.
Possibility of a hazard that could result in light or medium degree of bodily injury or
damage of device, if the safety instructions are not followed.
Caution
Danger
Never disassemble or alter the handpiece or charging unit.
Never use this handpiece or charger in an area with combustible liquids or gases.
Never plug anything, other than the supplied charger, into the charging port on the rear of the handpiece.
Doing so will void the warranty.
Warning
Do not attempt to replace batteries. Opening this device will void warranty. Return to service center for
all repairs and service.
Never allow water or other liquids to spill or seep into the handpiece as it could cause damage to the
handpiece and void warranty.
Do not use handpiece if any liquid leaks from inside the handpiece and report leak immediately to service
center. Immediately wash clothes or part of the body in water if you come in contact with battery fluid.
Caution
This cordless handpiece is only to be used by licensed dental practitioners or hygienists for rotary
root canal treatments or polishing.
Charging should be done in a dry place at 0° to 40° C (32° to 104° F) ambient temperature.
Battery life may be shortened when left uncharged for an extended period of time.
Never use a bent, damaged, deformed or non-ISO conforming instrument in this handpiece.
Do not use or leave this device in a high heat environment such as direct sun, near fire or heater.
Warming of the handpiece casing is not normal. If you feel it become warm, stop using the device
and contact us immediately.
Looseness, vibrations or rattling noises are not normal. If you feel or hear such, stop using the device
and contact us immediately.
Never attempt to sterilize the handpiece body itself. Only the heads may be sterilized. Handpiece body
may be cleaned only with alcohol.
General Instructions for Use 3
On/off Switch
Head
Main Body
Charging Port
Figure 1. Handpiece
Instructions
Warnings
• DO NOT attempt to open handpiece body for any reason at any time.
Return to service center if service is required.
• DO NOT autoclave the handpiece body.
• DO NOT ultrasonic the handpiece body.
4 Contra Angle Head Maintenance and Installation
Latch Push
Button
Collar
Collar
Figure 3. Push Button Latch Head
Figure 2. Latch Head
Head Maintenance
• Only use the contra angle head provided with the cordless handpiece or one that is
sold by LightSpeed Technology Inc. specifically for the LightSpeed EndoPAL
handpiece. Failure to do so will void the warranty.
• Please follow the contra angle maintenance and sterilization instructions between
each patient use.
• Steam autoclaving is the only acceptable sterilizing method. Remove from
handpiece and autoclave between patient uses. Sterilize between 121°C and
132°C for 15 to 20 minutes at 15-PSI minimum pressure. Follow manufactures
directions of the unit you are using for autoclaving. Never autoclave with an
instrument in head. Allow head to cool to room temperature before use.
• You may use rubbing alcohol to wipe off the head but this is not a sufficient
sterilization process.
• Lubricate the head with handpiece oil between each use and prior to autoclaving.
Clean off any excess oil that could seep into handpiece body, damage its
components or get into the patient’s mouth.
Figure 6. Battery Figure 6a. Model 1: 120 Volt Figure 6b. Model 2: 230 Volt
Charging Unit 60 Hz., with USA type plug 50 Hz, with Euro type plug
• The battery charger has a lighted charge indicator as shown in Figure 6. A red light
indicates that your battery is being charged. A green light indicates charging is complete,
is not needed, or that the charger is not connected to handpiece.
• DO NOT attempt to replace the batteries yourself. Return the handpiece to LightSpeed
Technology, Inc. to replace the battery pack. Please call for pricing and a Return
Authorization number.
6 Troubleshooting
If trouble is found, please refer to the following chart before consulting your dealer. If
none of the following scenarios are applicable or the solution does not remedy the
problem, a more serious problem is suspected. Contact your dealer.
The handpiece torque is noticeably weaker. The battery has become weak. Recharge the battery.
The speed of the handpiece is low. The battery has become weak. Recharge the battery.
Handpiece unit has come apart. Joint connections have failed. Call for RA#.
Battery Charger
Input rating . . . . . . . . . . . . . . . . . . . .Nominal 90-264VAC/47-63Hz
Reliability (MIL-GDBK-217F) . . . . . . .MTBF > 250,000 hours at 30ºC and full load
Battery Pack
Cell Type . . . . . . . . . . . . . . . . . . . . . .Rechargeable Lithium-Ion Cylindrical Cell
Weight . . . . . . . . . . . . . . . . . . . . . . . .34.5g
8 Warranty and Repairs
One-year parts and labor warranty for handpiece body and internal components
(excluding the battery), heads and charging unit. Warranty is void if units are
opened for any reason. Warranty may be void if handpiece is used without
protective moisture barrier or dropped. Before units are returned for repair or
service a return authorization number must be obtained. Shipping to our repair
facility must be prepaid if unit is out of warranty period.
Only use the contra angle provided with the cordless handpiece or one that is
sold by LightSpeed Technology Inc. specifically for the LightSpeed EndoPAL
handpiece. Failure to do so will void the warranty. Instructions, warranty and
repairs are subject to change without notice.