Professional Documents
Culture Documents
Wlzard 230: Instructions For Use
Wlzard 230: Instructions For Use
Wlzard 230: Instructions For Use
Mr.
WlZARD 230
Nasal Pillows Mask
Instructions For Use
Reorder Information
Description
(English/Espaiiol/Portugues/Fran1rais/Deutsch/
Nederlands/ltaliano)
[]]
A SM03001
Mr. WiZARD 230 Nasal Pillows Mask I Mascarilla nasal Mr.
WiZARD 230 para hombre I Mascara de Almofadas Nasais Mr.
WiZARD 230 I M. WiZARD 230 Masque Narinaire I Mr. WiZARD
230 Nasenpolster-Maske I Mr. WiZARD 230 lntranasaal
neusmasker I Maschera con cuscinetti nasali da uomo WiZARD
230
.
Please read the instruction
manual before use.
SM00019
Cuscinetto nasale M
www.apexbrand.com
Print-2012/All rights reserved
686004-0000 V1 .2
CE
0197
nasale S
Intended Use
The WiZARD 230 Nasal Pillows Mask is intended for use in the home
environment, hospitals and care facilities by adult patients (more than 30 kg) for
whom positive airway pressure (CPAP or bi-level system) has been prescribed .
Warnings
•The vent holes must be kept clear.
•Follow all precautions when using supplemental oxygen. Oxygen flow must be
turned off when the flow generator is not operating.
•Use your mask as prescribed by your physician. It should be used only together
with the appropriate flow generator equipment (CPAP or bi-level device) to
ensure optimum therapy.
•Contact a health care professional if you experience ANY adverse reaction to
the use of the mask.
Symbols
.&. Caution [I!J Consult instructions for use
ffim Reference number ~ Lotnumber
CJ Date of manufacture Use by
,, .., Temperature limitation Si)" Humidity limitation
I@ Does not contain latex ~- Medical prescriptions
.. Manufacturer ~ Authorized representative in the European
Community
Before Use
• Read the instructions completely.
• Remove plastic cover from the silicon pillows.
• Hand-wash the mask.
• Inspect the mask for damage or wear. Do not use if damaged.
f.... Do not use bleach, chlorine, alcohol, scented oils or aromatic based
ill solutions. These solutions may damage the mask, cause cracks and reduce
the life of the product. Do not iron the headgear.
Technical Specifications
Note: The mask system does not contain latex, PVC or DEHP materials.
Exhaust Flow
The mask has vent holes to flush out the exhaled air and ensure that fresh air is
inhaled.
Pressure (cmH20) 4 8 12 16 20
WiZARD 230 Exhaust Flow Flow (UMin) 17 26 34 40 47
Refer the Pressure-Flow Curve to Fig 4 on the front page.
50
Mask Dead S ace
45
100 ml
40
Thera Pressure Ran e
35
~ 4-20 cmH20
c
.E 30
Resistance pressure drop
2: 25 at 50 I/min: 0.5 cm/H20 at 100 I/min: 1. 7 cm/H20
~ 20
0 Operating Range
iI
15 +5°C to +35°C (+41°F to +95°F)
10 15% to 95% relative humidity non condensing
5 Storage and transport
0 -15°C to +60°C (+5°F to +140°F)
-
4 8 12 16 20 10% to 90% relative humidity non condensing
Pressure (cmH20) Store the mask in a dry place out of direct sunlight.
Disposal
Your mask does not contain any hazardous substance and may be disposed of
with your normal household waste.