Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

ARTICLE BACK TO BASICS

Centrifugation:
a guide to equipment
use and maintenance
With increased emphasis now placed on work-based learning and the
20 50 20000
need for trainees and support workers to receive a solid grounding in 18 45 15000
16 40
laboratory techniques, The Biomedical Scientist once again goes back to 14 10000
35
12 30000
basics. This first in a series of articles sees Peter Riddle consider the use 30
20000
6000
10 25 10000
and maintenance of laboratory centrifuges.
9 5000 4000
8 20 3000
The centrifuge is ubiquitous in biomedical Earth’s gravitational force and is known as 7 18 2000
laboratories and a basic knowledge of the the relative centrifugal field (RCF) or, more 6 15 1000 2000
theory of centrifugation is more than useful. commonly, the ‘g’ force. 14 500
Centrifuge performance can be classified as 5
low-speed, high-speed and ultra-speed. Usual RELATIVE CENTRIFUGAL FIELD 12
200 1000
applications include the separation of serum The RCF generated by a rotor depends on 4 10 100
or plasma from red blood cells, the separation the speed of the rotor in revolutions per 9
of precipitated solids from the liquid phase minute (rpm) and the radius of rotation 50
3 8 500
of a mixture, or the separation of liquids of (ie the distance from the axis of rotation).
7 20
varying density. The equations that permit calculation of the
RCF from a known rpm and radius of 6 10
PRINCIPLES OF CENTRIFUGATION rotation, and calculation of the rpm from a
Particles suspended in a fluid move, under known RCF and radius are shown in Table 1. 2 5 3 200
the influence of gravity, towards the bottom The RCF value can also be obtained using
of a vessel at a rate that depends, in general, a nomogram (Fig 1). Using a straight-edged Inches cm Relative Revolutions
on their size and density. Centrifugation is ruler, line up the known rotating radius centrifugal per minute
a technique designed to utilise centrifugal (distance from the centre of the rotor to the field (xg)
forces, which are greater than the force of bottom of the centrifuge bucket) on the left
gravity, to speed up the sedimentation rate with the known rpm on the far right and read Fig 1. Nomogram used to convert
of particles. This is achieved by spinning the RCF value where the line crosses the revolutions per minute to relative
the vessel containing the fluid and particles graph in the centre. Most manufacturers centrifugal field (xg), and vice versa.
about an axis of rotation so that the particles include a nomogram in the instruction
experience a force acting away from the axis. manual; however, most modern centrifuges there are instruments with a temperature
The force is measured in multiples of the now have the facility to swap the figure control facility. Most instruments now include
displayed on the control panel between rpm a sensor that will detect any imbalance when
Table 1. Calculations used to convert and RCF, making manual calculation the rotor is running and cut off power to the
rpm to RCF, and vice versa. unnecessary. drive mechanism if imbalance is present.
Low-speed instruments are used to separate
RCF = 11.18 x r (rpm/1000)2
LOW-SPEED INSTRUMENTS serum or plasma from red blood cells, and to
rpm = 299.07 √RCF/r Low-speed centrifuges have maximum rotor harvest and purify chemical precipitates,
r: radius (cm) speeds of less than 10,000 rpm, which do not intact cells, nuclei, large mitochondria and
require the rotors to be run in a vacuum, and large plasma-membrane fragments.

76 THE BIOMEDICAL SCIENTIST FEBRUARY 2013


ARTICLE

HIGH-SPEED INSTRUMENTS door-locking mechanism manually to


In general, high-speed centrifuges are retrieve the samples. Manufacturers’
capable of rotor speeds up to 21,000 instructions should be checked for
rpm, although the new generation of the exact procedure required.
super-speed instruments are capable
of rotor speeds of 30,000 rpm, Refrigerator
in which RCFs of 120,000 xg are A centrifuge generates heat as
possible. These instruments require it rotates and if samples are
refrigeration systems to overcome the temperature labile then a refrigerated
heat generated by the friction of the centrifuge should be used. Some
spinning rotor, and the higher-speed centrifuges enable the rotor and
machines must incorporate vacuum chamber to be precooled before a
systems. High-speed centrifuges are run.
used in the separation of a number of
cell constituents and in the isolation Braking system
and purification of viruses. Braking devices are incorporated
to provide rapid rotor deceleration.
ULTRACENTRIFUGES Modern instruments have an
Ultracentrifuges are capable of electrical braking system that
speeds in excess of 30,000 rpm functions by reversing the polarity
and RCFs of over 600,000 xg. of the electrical current to the
RSP

They can be used in the isolation motor. Other machines may have
and purification of membrane Owing to their aerodynamic construction, fixed-angle rotors a mechanical brake.
components such as the endoplasmic can attain much higher speeds.
reticulum and Golgi membrane, CENTRIFUGE TUBES
endosomes, ribosomes, DNA and RNA. Motor It is advisable to use a conical-bottomed
Once again, refrigeration and vacuum In general, centrifuge motors are high-torque, tube in a swing-out bucket rotor for the
systems are necessary. series-wound DC motors, the rotation of which sedimentation of cells. This tube type will
increases as the voltage is increased. The rotor retain the pellet of cells more effectively as the
INSTRUMENT COMPONENTS shaft is driven directly or through a gyro, supernatant is removed. All tubes for use with
Rotor although occasionally a pulley system is used. high-speed rotors are round-bottomed. Pyrex
The design of most centrifuges allows the Electrical contact to the commutator is provided glass tubes can withstand forces of around
drive system to accept rotors of different sizes by graphite brushes, which gradually wear down 2000 xg, while Corex tubes can be used up to
and capacities, although most instrument as they press against the commutator turning 12,000 xg. Polycarbonate or polyallomer are
rotors are now capable of accepting a large at high speed, and thus should be replaced at the most common plastic tubes in use but
range of tube sizes through the use of specified intervals. Modern centrifuges have great care must be taken when using organic
adaptors. Rotors have three basic designs: induction drive motors that have no brushes to solvents. Manufacturers usually provide
horizontal, in which the tubes are carried change. The shaft of the motor turns through extensive information about solvent, salt
in buckets that can swing outwards to a sleeve bearings located at the top and bottom and pH resistance, as well as sterilisation
horizontal position and can operate at speeds of the motor. Most instruments contain sealed procedures.
to about 3000 rpm; fixed angle, in which the bearings that are permanently lubricated, while
sample tubes are held at a fixed angle to the others require periodic application of oil or PREVENTIVE MAINTENANCE
vertical position and can attain much higher grease. r If the bearings on the upper and lower
speeds (approximately 7000 rpm) because The speed of the centrifuge is controlled ends of the motor shaft are not of the
of the aerodynamic construction of the rotor; by a potentiometer that raises and lowers the sealed type then they should be lubricated
and vertical, in which the tubes are fixed voltage supplied to the motor. The calibrations as per the manufacturer’s instructions.
in the vertical position. In general, the on the speed control are often only relative r Brushes should be removed regularly and
horizontal rotor offers advantages to the voltage increments and should never be taken checked for wear; they should be replaced
clinical laboratory because sedimentation of as accurate indicators of speed. Therefore, if they are worn to more than one-half of
large particles (eg red blood cells) is efficient periodic recalibration is required. their original length. When reinserting
at low force and because a flat sediment is used brushes, replace them in the same
produced. Imbalance detector orientation. New brushes should be broken
The load on the rotor should always be Some instruments have an internal imbalance in by slowly accelerating the unloaded unit
balanced before operating the centrifuge, detector that monitors the rotor during to mid-speed and then allowing it to run
particularly when using high-speed operation, causing automatic shutdown if for a period of time.
instruments in which the buckets and caps rotor loads are severely out of balance. r The rotor, buckets and shields or carriers
are often numbered so that they can be should be examined for signs of
matched on opposite sides of the rotor. Tachometer mechanical stress (eg cracks, corrosion).
The load must be balanced both by equal A tachometer indicates the speed in rpm. Most r Some manufacturers etch the expiry date
mass and by centres of gravity across the modern centrifuges use electronic tachometers, on the rotor and this should be checked
centre of rotation. Thus, it is important not in which a magnet rotates around a coil to periodically.
to run the centrifuge with buckets, carriers produce a current that can be measured. r Regularly lubricate the contact areas
or shields missing from the unit, and not to between the centrifuge buckets and the
exceed the maximum rated speed of the Safety lid pins.
rotor in use. Most modern rotors have Modern centrifuges must have a door-locking r Regularly check the condition of the
microprocessor-controlled automatic rotor mechanism to prevent the lid from being O-ring on the tie-down nut on top of the
identification so that it is impossible to set opened while the instrument is running. If rotor, and replace it if worn or damaged.
the speed beyond the safety limit for that there is a power failure or the safety latch fails r Always follow a manufacturer’s specific
rotor. for some reason it may be necessary to trip the instructions. r

FEBRUARY 2013 THE BIOMEDICAL SCIENTIST 77

You might also like