SD BIOLINE HBsAg WB Quick Reference Guide Caution:

For more detailed instructions please refer to Instructions for use. DO NOT USE THE TEST MORE THAN ONCE.

Before you • Please read the • Cover your work space • All components must be brought 30˚C Preparation: • Test pouch (Test Device) • Timer (Not Provided)
begin: Instructions for use in its with a clean, disposable to room temperature (between • Micropipette & Disposable Tip • Disposable Gloves (Not Provided)
entirety, prior to use. absorbent workspace cover. 15-30˚C) prior to testing. 15˚C (Not Provided) • Permanent pen (Not Provided)

1 2 3
Check the expiry date. Put on the gloves. Use new gloves for each patient. Open the test pouch and write the unique
If expiry date has passed, use another kit. patient identifier on the test device.

Check the desiccant.


If the color is green, use another kit. HBsAg WB
HBsAg WB

HBsAg WB
XXXXXXXX
YYYY.MM.DD
YYYY.MM.DD

4 5 6
Use the micropipette to collect serum, Add drawn serum, plasma or whole blood Interpret test results at 20 minutes.
plasma or whole blood specimen (100μl). specimen into the specimen well (marked “S”). The control line should
Discard the disposable tip in the appear for all results. If it
Do not read test results after 20 does not appear, the result
sharps box after adding specimen. minutes. Reading too late can give is considered invalid and the
false results. specimen should be retested
using a new test kit.

Positive Negative Invalid


C T C T C T

C T
HBsAg WB
HBsAg WB

www.standardia.com © 2016 Alere. All rights reserved. The Alere Logo and Alere are trademarks of the Alere group of companies.
All other trademarks referenced are trademarks of their respective owners. 120003168E-01, 11/16