HDI 5000 SonoCT System Specifications

® ®

premium
Bringing
performance
ultrasound to the whole picture
System Overview
Applications
• Abdominal
• Obstetrical
• Cerebrovascular
• Peripheral vascular
54.6 in
• Gynecological and fertility 138 cm
• Advanced breast imaging
• Small parts and superficial
• Musculoskeletal
• Transcranial Doppler
43.4 in
• Pediatric general imaging 110.3
28.3 in
• Prostate 72 cm
• Adult cardiology
• Pediatric cardiology Physical Dimensions
• Endoscopic ultrasound* Width: 28.3 in/72 cm; Height: 62 in/157.5 cm with VCR,
54.6 in/138 cm without VCR; Depth: 43.4 in/110.3 cm;
Imaging Modes Weight: approximately 380 - 440 lbs/172-200 kg, depending
• MicroFine TM Grayscale Imaging on peripherals

• SonoCT Real-time Compound Imaging


• XRES Visualization Technology • Reconstructed zoom with pan
• Tissue Harmonic Imaging with pulse • Chroma ® Imaging
inversion technology • Contrast Specific Imaging (2D Harmonic, Power
• Harmonic SonoCT Imaging Contrast Harmonic Imaging, Pulse Inversion
• Simultaneous M-mode Harmonic Imaging, Grayscale Pulse Inversion
• Duplex and simultaneous pulsed Doppler Contrast Imaging, Power Pulse Inversion
• Continuous wave Doppler Imaging, MicroVascular Imaging* and Agent
• Simultaneous 2D, continuous wave Doppler Detection Imaging*)
• Broadband Flow® Imaging • Color Power Angio ® Imaging
• Simultaneous 2D, color Doppler, pulsed Doppler • Panoramic SonoCT Imaging
• Dual composite image • Power Motion Imaging
• High DefinitionTM Zoom • Tissue Doppler Imaging
• 3D SonoCT Imaging • Supercomputing Image Processing Unit (IPU)
• 3D Color Power Angio Imaging • Dynamic Focal Tuning
• 3D Grayscale Imaging • 170 dB full time input dynamic range
• 1,024 digitally processed channels
Control Panel and User Interface
• Up to nine frames compounded in real time with
• Easy-to-learn graphical user interface
SonoCT imaging
• Primary controls readily accessible and
• Advanced Extended Signal Processing (ESP) with
logically grouped
parallel digital signal processing and frequency
• Commonly used secondary controls located on
compounding:
“superkeys” for quick access
- Improves tissue contrast resolution and
• Other secondary controls accessible through
textural perception
on-screen menus
- Sharpens lateral beam profile for finer dot size
• Slide pot control of TGC curve, overall gain
- Reduces speckle artifacts for increased
control knobs for 2D, color and Doppler
image clarity
• Smart TGC: pre-defined TGC curves optimized
• Intelligent Tissue Specific TM Imaging software
for consistently excellent imaging with minimal
provides total system optimization by exam type
TGC adjustment
of more than 5,000 parameters
Gray Shades • Patient Optimization key for rapid system
• 256 (8 bits) in 2D, Doppler spectral analysis optimization in 2D and color Doppler
• 64 (6 bits) in M-mode • iSCAN Intelligent Optimization*
- In 2D mode, one-button automatic
Acquisition Frame Rate
adjustment of:
• Greater than 500 frames/sec (dependent on
- TGC and receiver gain to achieve optimal
field-of-view, depth and angle)
uniformity and brightness of tissues
User- centered Design Focus - Compression curve based on range of
• High mobility cart detectable tissue signals
- Designed for easy maneuverability and mobility - In Doppler mode, one-button automatic
- Four-wheel swivel cart adjustment of:
- Lightweight aluminum monocoque frame - Doppler gain and PRF based on signal
- Single, digitally controlled, 15-inch non- strength and detected velocity
interlaced high resolution color monitor with - Doppler baseline based on detected flow
tilt and swivel direction
- User adjustable control panel lighting
Display Annotation
- Universal peripheral bay provides easy access
• On-screen annotation of all pertinent imaging
to hardcopy/documentation devices
parameters for complete documentation,
- Built-in A/C line conditioner provides
including transducer type and frequency, active
isolation from voltage fluctuations and
clinical options and optimized presets, display
electrical noise interference
depth, TGC curve, grayscale, color map, frame
- Four high-capacity fans with automatic speed
rate, dynamic range, compression and contrast
adjustment to optimize cooling efficiency with
enhancement, color gain, color image mode, and
minimal audible noise
hospital and patient demographic data
System Architecture • Annotation data can be turned off for generating
• All-digital broadband beamformer images used in publication and presentation
• MicroFine 2D Imaging • Sector Steering icon for endocavitary transducers
• Digital Broadband Flow Imaging • Real-time display of Mechanical Index (MI)
• Real-time display of Thermal Index
(TIb, TIc, TIs)
• Multiple trackball-driven annotation arrows • L7-4 Broadband Linear Array
• Quick Text keys that will memorize up to 30 - 4 to 7 MHz extended operating frequency range
phrases or names per key - Steerable pulsed Doppler, color Doppler and
• Predefined body markers (application specific Color Power Angio, SonoCT, XRES, harmonic
anatomic annotation) and MicroVascular imaging*
- General imaging and superficial imaging
Image Presentation
applications
• Up/down
• C7-4 Broadband Curved Array
• Left/right
- 4 to 7 MHz extended operating frequency range
• Multiple duplex image formats
- Steerable pulsed Doppler, color Doppler and
(1/3-2/3, 1/2-1/2, 2/3-1/3)
Color Power Angio, SonoCT, XRES and
• Depth to 24 cm
harmonic imaging
- High resolution abdominal, pediatric,
Transducers
obstetrical and gynecological applications
Transducer Selection
• C8-5 Broadband Curved Array
• Electronic switching of transducers using three
- 5 to 8 MHz extended operating frequency range
universal connectors
- Steerable pulsed Doppler, color Doppler
• Dedicated (Pedoff) continuous wave
and Color Power Angio, SonoCT and
Doppler connector
XRES imaging
• Automatic optimization of each transducer for
- Pediatric abdominal and cephalic imaging,
exam type by Tissue Specific imaging
neurosurgery applications
Transducers • C5-2 Broadband Curved Array
• Multiple user-selectable transmit focal zones on - 2 to 5 MHz extended operating frequency range
all transducers - Steerable pulsed Doppler, high PRF Doppler,
• Continuous dynamic receive focusing on color Doppler and Color Power Angio, SonoCT,
all transducers XRES and harmonic imaging
• L12-5 38 mm Broadband Linear Array - General purpose abdominal, obstetrical and
- Fine pitch, 192 element, high resolution gynecological applications
linear array • P12-5 Broadband Phased Array
- 5 to 12 MHz extended operating - 5 to 12 MHz extended operating
frequency range frequency range
- Steerable pulsed Doppler, color Doppler - Steerable pulsed and continuous wave Doppler,
and Color Power Angio, SonoCT and color Doppler, and Color Power Angio and
XRES imaging XRES imaging
- High resolution superficial applications - Pediatric cardiology, neonatal cephalic and
including vascular, small parts and neonatal abdominal applications
musculoskeletal imaging • P7-4 Broadband Phased Array
• L12-5 50 mm Broadband Linear Array - 4 to 7 MHz extended operating frequency range
- Fine pitch, 256 element, high resolution - Steerable pulsed and continuous wave Doppler,
linear array color Doppler, and Color Power Angio and
- 5 to 12 MHz extended operating XRES imaging
frequency range - Pediatric cardiology, neonatal cephalic and
- Steerable pulsed Doppler, color Doppler neonatal abdominal applications
and Color Power Angio, SonoCT and
XRES imaging
- High resolution superficial applications
including small parts, breast and
musculoskeletal imaging
• P6-3 Broadband Phased Array - Uniquely designed for intraoperative use in
- 3 to 6 MHz extended operating frequency range vascular and cardiac surgery
- Steerable pulsed Doppler, color Doppler, - Steerable pulsed Doppler, color Doppler,
and Color Power Angio, XRES and and Color Power Angio, SonoCT, XRES
harmonic imaging and harmonic imaging
- Abdominal (includes pediatrics and small - Superficial imaging and intraoperative
adults 30-140 lbs/14-64 kg) and obstetrical applications
applications • CL10-5 Broadband Compact Linear Array
• P5-3 Broadband Phased Array - 5 to 10 MHz extended operating
- 3 to 5 MHz extended operating frequency range frequency range
- Steerable pulsed and continuous wave Doppler, - Uniquely designed for intraoperative use in
and Color Power Angio and XRES imaging vascular and cardiac surgery
- Abdominal and cardiology (includes pediatrics - Steerable pulsed Doppler, color Doppler and
and small adults 30-140 lbs/14-64 kg) and Color Power Angio imaging
obstetrical applications - Superficial imaging and intraoperative
• P4-2 Broadband Phased Array applications
- 2 to 4 MHz extended operating frequency range • LI9-5 Broadband Intraoperative
- Steerable pulsed and continuous wave Doppler, - 5 to 9 MHz extended operating frequency range
color Doppler, and Color Power Angio and - Uniquely designed for intraoperative use in
harmonic imaging abdominal surgery
- Abdominal, cardiac and transcranial - Pulsed Dopper, color Doppler and Color Power
applications Angio imaging
• P4-1 Broadband Phased Array - Abdominal surgery applications
- 1 to 4 MHz extended operating frequency range • CT8-4 Broadband Intraoperative
- Steerable pulsed wave Doppler, color Doppler, - 4 to 8 MHz extended operating frequency range
and Color Power Angio and XRES imaging - Uniquely designed for intraoperative use in deep
- Deep abdominal, obstetrical and abdominal surgery
gynecological applications - Steerable pulsed Doppler, color Doppler and
• C9-5 ICT Broadband Curved Array Color Power Angio imaging
- 5 to 9 MHz extended operating frequency range - Deep abdominal surgery applications
- End-fire sector, 8 mm radius of curvature, • MPT7-4 Broadband Phased Array
150° field-of-view Multiplane TEE
- Steerable pulsed Doppler, color Doppler, and - 4 to 7 MHz extended operating frequency range
Color Power Angio, SonoCT and - Rotation through 180°, stepwise or continuous
XRES imaging (dual speed control)
- Endovaginal and endorectal applications - On-screen position indicator
• C8-4v Broadband Curved Array - Steerable pulsed and continuous wave
- 4 to 8 MHz extended operating frequency range Doppler, color Doppler, and Power Motion
- End-fire sector, 11 mm radius of curvature, and harmonic imaging
135° field-of-view - Cardiac transesophageal applications
- Steerable pulsed Doppler, color Doppler, and • BPT9-5 Broadband Phased Array
Color Power Angio, SonoCT, XRES and Biplane TEE
harmonic imaging - 5 to 9 MHz extended operating frequency range
- Endovaginal applications - 90° field-of-view
• CL15-7 Broadband Compact Linear Array - 10 mm tip diameter
- 7 to 15 MHz extended operating - Steerable pulsed and continuous wave Doppler,
frequency range color Doppler and Color Power Angio imaging
- Pediatric cardiac transesophageal applications
• BPTRT9-5 Broadband Dual Array • Operates in conjunction with SonoCT real-time
- 5 to 9 MHz extended operating frequency range compound imaging on the L12-5 38 mm, L12-5
- 150° field-of-view (curved), 55 mm 50 mm, L7-4, C7-4, C5-2, BPTRT9-5*, C8-4v*
field-of-view (linear) and C8-5* transducers
- 19 mm tip diameter
M -mode
- Steerable pulsed continuous wave Doppler,
• Available with all imaging transducers
color Doppler, and Color Power Angio (curved),
• Selectable sweeping rates
SonoCT and XRES imaging
• Time markers: 0.1 and 0.2 seconds
- X,Y, Z volume measurements
• Acquisition zoom
- Prostate transrectal applications
• Chroma colorization with multiple color maps
• M-mode review for retrospective analysis of
Imaging Modes
M-mode data
MicroFine Grayscale Imaging
• Color M-mode on all phased array transducers,
• Frame rate selection
and C5-2 and C7-4 curved array transducers
• 8-level digital reconstructed zoom with pan
• High Definition zoom concentrates all image Doppler
processing power into a user-defined area • Display annotations including Doppler
of interest mode, scale (cm/sec or kHz), Nyquist limit,
• Image orientation marker pulse repetition frequency, wall filter setting,
• Cineloop ® Image Review (up to 800 B/W frames) gain, acoustic output status, sample volume size,
• Persistence, adjustable in real time and normal/inverted, angle correction, grayscale curve
Cineloop Review • Angle correction with automatic velocity scale
• Selectable compression curves adjustment
• Sector size and steering control • Adjustable frequency/velocity display ranges
• Selectable line density • 8-position zero baseline shift
• Up to five transmit focal zones plus • Normal/invert display around horizontal
separation control zero line
• Dual imaging • Selectable sweep speeds
• Chroma imaging with multiple color maps • 1 millisecond spectral FFT rate
• Selectable low-frequency signal filtering with
SonoCT Real-time Compound Imaging
adjustable wall filter settings
• Available on the L12-5 38 mm and L12-5 50 mm,
• Selectable grayscale curve for optimal display
L7-4, C9-5, C8-5*, C8-4v*, C5-2, C7-4,
• Selectable display format
CL15-7 and BPTRT9-5 transducers
(1/3-2/3, 1/2-1/2, 2/3-1/3)
• Two modes of operation: Survey and Target
• Doppler Review for retrospective analysis of
• Operates in conjunction with Tissue Harmonic
Doppler data
imaging, 3D mode and panoramic imaging
• Operates in conjunction with XRES imaging Pulsed Wave (PW) Doppler
on the L12-5 38 mm, L12-5 50 mm, L7-4, • Available on all imaging transducers
CL15-7, C9-5, C8-5*, C8-4v*, C7-4, C5-2, and • Adjustable sample volume size: 1.0-25 mm
BPTRT9-5 transducers • Simultaneous or duplex mode of operation
• Simultaneous 2D, color Doppler, pulsed Doppler
XRES Visualization Technology
• High PRF capability
• Available on the L12-5 38 mm, L12-5 50 mm,
L7-4, CL15-7*, CL10-5*, C9-5, C8-5*, C5-2,
C7-4, C8-4v, P12-5*, P6-3*, P4-2, P4-1 and
BPTRT9-5* transducers
Continuous Wave (CW) Doppler • Available on the P4-2 and P5-3 transducers
• Cardiac phased array transducers only for cardiology applications and on the C5-2,
• Steerable through 90° L7-4, L12-5 and P4-2 transducers for general
• Maximum velocity range: 19 m/sec imaging applications
• Pulse Inversion Harmonic imaging on
Broadband Flow Imaging
P4-2, C5-2, L12-5 38mm, L12-5 50mm and
• Available on all imaging transducers
L7-4 transducers
• Cineloop review
• Power Pulse Inversion Harmonic imaging on
• Dynamic Motion Differentiation with intelligent
P4-2 and C5-2 transducers
proprietary algorithms; adapts to various
• Grayscale low MI image blended with
application types to selectively eliminate virtually
fundamental in general imaging applications on
all color motion artifact
the C5-2 and P4-2 transducers
• Chroma colorization with multiple color maps
• Surgical contrast options available on the
• 256 color bins
CT8-4 transducer*
• Color steering on L12-5 linear array transducers:
0° and ±20° Tissue Harmonic Imaging (THI)
• Trackball-controlled color region of interest: size • Patented pulse inversion phase cancellation
and position technology
• Maps, filters, color sensitivity, line density, • System processing of second harmonic
smoothing, echo write priority, color persistence, frequencies (nonlinear energy) in tissue
gain and baseline optimized automatically by • Extends high performance imaging capabilities
exam type or is user selectable to all patient body types
• Velocity and variance displays • Available on the P4-2, P5-3, C5-2, C7-4, C8-4v,
• Color invert P6-3 and L7-4 transducers
• Color capture • Image display virtually free of artifacts
• Color/2D line density control • Supports SonoCT Real-time Compound
• Selection of color bar display units imaging (Harmonic SonoCT imaging)
• Color tag capability
Color Power Angio Imaging (CPA)
Contrast Specific Imaging (CSI) • Highly sensitive mode for small vessel visualization
• System optimized for detecting harmonic agent • Available on all transducers
signatures—optimized for all available agents • Fully user-configurable
• Available on P4-2, P4-1, C5-2, L7-4, • Cineloop review
L12-5 38 mm and L12-5 50 mm transducers • User-definable presets
• MicroVascular Imaging* for detection of contrast • Multiple maps
agents in superficial scanning of areas with very • Individual controls for gain, filters, sensitivity,
low flow velocity and signal strength; available on echo write priority and color invert
the L7-4 transducer • Dynamic Motion Differentiation
• Agent Detection Imaging* for quick verification • Adjustable CPA region of interest: size
of presence or absence of contrast agents and position
• Cardiac applications in 2D Harmonic, Power • Background on/off
Contrast Harmonic imaging, Pulse Inversion • User-selectable persistence
Harmonic imaging, Grayscale Pulse Inversion • Topographic and high-resolution displays
Contrast imaging and Power Pulse • User-selectable blend levels
Inversion imaging • TGC control
Tissue Doppler Imaging (TDI) • User can display or remove skin-line ruler
• Velocity mapping of cardiac tissue and vessel • Scaling information will be included for
wall motion connectivity prints allowing for measurements
• Available on the P4-2, L12-5, P5-3, MPT7-4 on a workstation
and BPT9-5 transducers • Available on linear and curved array transducers
• TGC control
Power Motion Imaging (PMI)
• Supports harmonic TDI in conjunction with
• Highly sensitive mode for visualization of cardiac
grayscale THI
endocardium and wall motion
3D SonoCT Imaging • Available on all phased array transducers in
• Qualitative freehand acquisition supported on cardiac applications only
all imaging transducers • Multiple maps
• 2D grayscale and CPA imaging • Individual controls for gain, filters, sensitivity,
• Volume display with surface rendering echo write priority and color invert
• Multiplanar view display • Adjustable PMI region of interest: size
• Full trim tools on both volume and and position
multiplanar views • User-selectable persistence
• Supports SonoCT Real-time Compound • TGC control
imaging (3D SonoCT imaging)
Measurements and Analysis
3D Color Power Angio Imaging (3D CPA)
Measurement Tools and General Description
• Provides comprehensive visualization of
• 2D distance
vascular anatomy
• 2D circumference/area by ellipse, continuous
• Volume rendering of CPA images
trace, trace by points
• Available on all imaging transducers
• 2D curved-linear distance
• Individual controls for manipulating the
• M-mode distance (depth, time, slope)
on-screen 3D rendering and display options
• Manual Doppler distance
3D Grayscale Imaging • Manual Doppler trace
• Provides a qualitative volume display of • Time/slope measurements in Doppler and
2D data set M-mode
• Available on all imaging transducers • High Q ® Automatic Doppler Analysis
• Specialized algorithms and maps maximize • Doppler values including PI, RI, S/D indices
three-dimensional display • 2D volume
• Individual controls for manipulating the • Heart rate
on-screen 3D rendering and display options • Trackball-controlled electronic measurement
calipers: 6 sets
Panoramic SonoCT Imaging
• User-defined protocols, measurements
• Real-time extended field-of-view composite
and equations
imaging, acquired in SonoCT mode
• On-the-fly measurement labels
• Ability to back up and realign the image
• Fully editable results data sheet
during acquisition
• Integrated patient exam report
• Full zoom, pan, Cineloop review and image
• ABD (Assisted Border Detection)
rotation capabilities
- Available in Acquisition and Review
• Auto fit of composite image
- Provides simultaneous left ventricle (LV)
• User can measure distance, curved-linear
endocardial traces for end diastole (ED) and
distance and area in review mode with distance
end systole (ES) frames
marker displayed via skin-line ruler
- Calculates LV volumes and ejection fraction • Gynecology/Fertility
(EF) using Simpson’s Rule - Uterus
- Available for apical 4- and 2-chamber views - Right and left ovary
- Supports biplane and single plane results - Right and left follicles
- Tools available for quick trace adjustment, • General imaging
manual trace or reselection of ED or ES frame - General
- Requires DVSTM technology • Advanced breast imaging
- Breast
Clinical Option Analysis Packages
• Prostate
• Cardiac analysis
- Prostate gland
- Volume by area/length method
• Pediatric
- Ejection fraction (via Teichholz or cubed
- General
method)
- Simpson’s biplane and single plane High Q Automatic Doppler Analysis
- LV mass • Automatic real-time and retrospective tracing of:
- M-mode analysis - Instantaneous peak velocity (or frequency)
- Doppler - Instantaneous intensity weighted mean velocity
- Peak and mean gradients (or frequency)
- Pressure half time • Vascular
- E/A ratio - Automatic real-time display of:
- Continuity equation - Volume flow
- Diastolic function - Time-averaged peak velocity (or frequency)
- Cardiac output - Time-averaged mean velocity (or frequency)
- Qp: Qs ratio - Resistive index
- Proximal Isovelocity Surface Area (PISA) - Pulsatility index
• Vascular analysis - Systolic/diastolic ratio
- Right and left carotid artery protocols - Acceleration/deceleration times
- ICA/CCA ratio • Cardiology
- Right and left vertebral flow direction - Automatic real-time display of:
descriptive list - Peak velocity
- Transcranial vasculature protocols (Doppler - Peak gradient
angle and SV-depth reported) - Display of:
- Right and left lower extremity protocols - Cardiac output
- Optional tools: percent diameter area reduction - VTI
- User comments - Mean velocity
• OB analysis - Mean gradient
- Fetal biometry
Cineloop Review
- Biophysical profile
• Acquisition, storage in memory, and display in
- Amniotic fluid index
real-time and duplex modes of up to 800 frames
- Early gestation
of 2D and color images or up to 100 seconds
- Fetal long bones
of Doppler data and M-mode for retrospective
- Fetal cranium
review and image selection
- Other OB measurements
• Extended loop capability up to 1000-frame
- 2D echo
loops with variable capture rates; allows
- Fetal heart M-mode
continuous capture for extended periods in
- Fetal Doppler
Contrast Specific Imaging*
- Echo Doppler
• Trackball control of image selection
• Variable playback speed • NetLink
• Choice of loop or sweep playback - DICOM 3.0 print and store service class user
• Trim capability - Fully integrated user interface
• Functions in: - 10-100 BaseT Ethernet output
- 2D and Tissue Harmonic imaging - Transceiver included with NetLink
- M-mode - Site configurable IP address
- PW Doppler - Requires 50% available network capacity and
- CW Doppler RJ45 connectors
- Color Doppler - Performed Procedure Step
- Color Power Angio imaging - Storage Commitment
- Power Motion imaging - Modality Worklist
- Contrast Harmonic Imaging - Works in conjunction with RIS systems
- Automatic entry of patient demographics
ECG/Physio
- Provides support for HIPAA requirements to
• One 3-lead ECG input
safeguard patient information*
• One external ECG input
• WebLink TM Server
• Two physio input channels (1V p-p)
- HTTP protocol, HTML format provides
• External pulse-phono module (option)
World Wide Web compatibility
• External respiration module (option)
- Stores a maximum of 4,500 NTSC/3,500
• Selectable ECG triggered skipping between
PAL images
1 and 20
- Provides access to saved images from a computer
using Web browser, e.g. Netscape ® 3.0 or higher
Miscellaneous System Data
or Microsoft ® Internet Explorer 3.0 or higher
Footswitch
- Uses lossy JPEG image compression, ratio = 10:
• Three pedal
1- 20:1 (typically 16:1)
• Allows Freeze, start/stop Doppler scroll
- Site configurable IP address
• User definable between Print and VCR Record
- Integrated server capability for remote internet/
Connectivity intranet image access
• DiskLink - Password security
- Direct digital storage of single frame color
Exam Documentation
and B/W images to internal hard disk and
• Super VHS VCR
magneto-optical disk
• Color or B/W page printer
- Direct digital storage of B/W and color loops
to internal hard disk DVS Digital Video Streaming and
- Stores a maximum of 4,500 NTSC/3,500 PAL Digital Stress Echocardiography
images to internal hard drive • Free-form capture of Full Screen, Dual or Quad
- 540 MB magneto-optical disk stores a image clips
minimum of 1,680 NTSC/1180 PAL • Optional fully-configurable protocol engine
B/W images with four standard stress echo protocols and one
- Password protected access to images stored on routine digital echo protocol
the hard drive • Qualitative Wall Motion Scoring standard with
- Fully-integrated interface Protocol Acquisition
- Supports DEFF compatible printing via • User can modify existing protocols or create
optical disk entirely new protocols
- On-screen recall, measurement and text editing • User can store up to 44 custom protocols;
- Exam directory each protocol can have up to 10 stages with 75
- Delete and replace recalled image capability views each
• User-configurable capture parameters, Safety Requirements
including: Capture Format, Capture Length • Electromechanical standards met:
and Compression - C22.2 No. 601.1, Canadian Standards
• Image capture triggered to the ECG or over time Association, Standard for Medical Electrical
up to one minute Equipment
• Capture available in all modes including sweeping - JIS 60601-1, Japanese Standard for Medical
Doppler and M-mode Electrical Equipment
• Protocol suspend at any time to add additional - EN 60601-1, European Norm, Safety of
image clips Medical Electrical Equipment
• Trim, Mix and Match, and Label tools - EN 60601-1-2 European Norm, Collateral
• Three playback options in Review Standard: Electromagnetic compatibility
• DICOM 3.0 Store and DICOM Scaling - UL 2601-1 Underwriters Laboratories Standard
• DICOM Modality Worklist capability for Medical Electrical Equipment
• Standard 3.5" or available 5.25" magneto-optical • Agency approvals:
drive for image transfer and storage - Canadian Standards Association (CSA)
• Three selectable levels of JPEG lossy compression; - CE Mark in accordance with the European
typical image rotation less than 30:1 Medical Device Directive issued by British
• Optional hand controller (standard with stress Standards Institute (BSI)
echo) allows remote access to major system - Japanese Ministry of Health, Labor and Welfare
controls, including: Capture, Freeze and Capture
Region of Interest (ROI) box position
• NTSC and PAL formats

Electrical Power/Video Parameters


• 100V-120V/50 Hz and 60 Hz – NTSC
• 220V-240V/50 Hz and 60 Hz – NTSC and PAL
• Integrated A/C line conditioner
• Power consumption: 775 to 1,250 watts,
depending on peripherals
* All Performance 2003 Upgrade package features/capabilities are available soon.

“MicroFine,” “XRES,” “High Definition,” “Tissue Specific,” “WebLink” and “DVS” are
trademarks and “HDI,” “SonoCT,” “Broadband Flow,” “Chroma,” “Color Power Angio,”
“High Q” and “Cineloop” are registered trademarks of Philips Ultrasound.

Other trademarks appearing herein are trademarks of their respective owners.


Philips Medical Systems is part North America
Tel: 800 982 2011
of Royal Philips Electronics
Asia Pacific
At Philips our responsibility to the environment Tel: 852 2821 5888
influences our actions globally every day. From
Europe
our manufacturing and business processes to our
Tel: 49 89 321 75 0
products and their use, we constantly strive to
discover new and innovative means of caring for Latin America
our planet and her inhabitants. Tel: 954 835 2600

www.ultrasound.philips.com
© Koninklijke Philips Electronics N.V. 2002
All rights are reserved. Reproduction in
whole or in part is prohibited without prior
written consent of the copyright holder.

Philips Medical Systems Nederland B.V.


reserves the right to make changes in
specifications or to discontinue any
product, at any time without prior notice
or obligation, and will not be liable for any
consequences resulting from the use of
this publication.

Printed in USA 8 - 02
G59981

You might also like