Professional Documents
Culture Documents
Analyst2004129607 612
Analyst2004129607 612
discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/8494068
CITATIONS READS
38 413
2 authors, including:
Ross Aylott
Aylott Scientific
21 PUBLICATIONS 194 CITATIONS
SEE PROFILE
All content following this page was uploaded by Ross Aylott on 17 June 2014.
Whilst the authentication of Scotch whisky brands using laboratory based gas chromatography is well established, there
is a need for authentication methods for use under field test conditions. The UV/visible absorbance spectra of specific
brands of Scotch whisky were found to produce consistent absorbance ranges which allowed the development of a faster,
cheaper and more mobile test method. Test samples with spectra outside the acceptable ranges for genuine reference
samples were therefore deemed suspect, enabling their authenticity to be later confirmed in the laboratory by GC. This
novel application allowed brand authenticity analyses to be undertaken at remote sites where gas chromatography was
unavailable and has also allowed the introduction of field testing using a small portable hand held spectrophotometer.
Introduction Experimental
Absorbance spectra of test samples were collected using two
As part of its worldwide popularity, blended Scotch whiskies can be
different spectrophotometers, one laboratory based and one
subject to illegal counterfeiting and substitution. Consumer protec-
portable. Absorbance data at set wavelengths from both instru-
tion agencies and producers require analytical laboratories to
ments were transferred electronically via RS232 interfaces into
support their consumer and brand protection programmes with
Microsoft® Excel spreadsheets.
effective brand authenticity analyses.
Initial laboratory development was conducted on a WPA
Current authenticity analysis is based upon the gas chromato-
graphic determination of the major volatile congeners present in Lightwave® S2000 UV/visible diode-array spectrophotometer
(Walden Precision Apparatus Ltd., The Old Station, Linton,
Scotch whisky. Normal concentration ranges are first set for
Cambridge, UK CB1 6NW) fitted with deuterium and tungsten
genuine product, against which the results for suspect samples may
lamps. Samples were introduced by drawing liquid up into a 1 mm-
be compared. Suspect samples are then analysed for alcoholic
path length flow cell using a 5 ml plastic syringe. Absorbance data
strength and major volatile congeners. If the suspect sample’s
was collected at 1 nm intervals between 200 and 500 nm and
alcoholic strength and congener results fall within the normal
transferred into the spreadsheet using WPA Lightwave® software.
ranges for the genuine brand, then the sample is considered
authentic. If the results are outside the normal ranges, the sample is At the start of operation, the spectrophotometer was referenced
against 40% vol. ethanol in water.
not authentic. These analyses may then form part of the evidence
The development of mobile authentication was conducted on a
used in the prosecution of offenders.1,2
However, investigations are enhanced when supported by fast SAD Brand Authenticator® (Spectroscopic and Analytical De-
vices, West Grinstead, West Sussex, UK RH13 8LH. http://
and reliable forensic information, ideally based on field tests.
www.spectroscopic.co.uk). This small novel instrument was de-
Whilst the current gas chromatographic approach provides accurate
signed and built specifically by SAD for the purpose of enabling
and precise results which enable reliable conclusions, it can take
authenticity analysis to be conducted under field conditions. The
significant time and cost to locate and utilise local laboratories with
instrument measures 27 cm long 3 9 cm wide 3 10 cm deep,
appropriate equipment and expertise. Spectroscopic techniques
weighs 1.75 kg and is shown in Fig. 1. It is powered by a 20 V
were therefore considered as an alternative means of authenticating
whiskies. rechargeable battery that enables approximately 2 h continuous use.
UV and visible absorbency in Scotch whisky is primarily derived
from three sources. First, the Scotch whisky maturation process
results in the addition of a range of congeners, which are extracted
from the oak cask into liquid through the interaction of ethanol with
wood lignin. Secondly, volatile phenolic compounds are found in
whiskies where peat is used in the kiln drying of the malted barley
used in the fermentation. The spectroscopic properties of the
maturation and phenolic congeners are employed in their high
performance liquid chromatographic determination with UV and
fluorescence detection. Thirdly, batch to batch colour consistency
is often maintained through the addition of trace spirit caramel
(E150a). Indeed, the colour consistency of Scotch whisky is
DOI: 10.1039/b403068k
Fig. 1 Photograph of a test sample being injected into the SAD Brand
† For Part 1 see ref. 1. Authenticator®.
This journal is © The Royal Society of Chemistry 2004 Analyst, 2004, 129, 607–612 607
The Brand Authenticator® was fitted with a deuterium lamp and a represent blend compositions containing increasing proportions of
diode array detector. Samples were introduced by either injecting or malt whiskies. Compositions may be shown from their 2- and
drawing liquids into the 1 mm-path flow cell using a plastic syringe. 3-methylbutanol concentrations, congeners primarily associated
Absorbance data was collected at defined intervals at approx- with the malt whisky contents of blended whiskies.3
imately 1 nm resolution between 200 and 460 nm. Four patents The spectra for single samples taken from 10 different batches
have been applied for covering various features of this instrument, representing five different brands (labelled A to E) produced over
which is designed to be simple to operate in stand-alone mode a two-year period are shown in Fig. 2. There were clear separations
controlled by internal software. There are five control keys, a two- between the spectra of certain brands, particularly brands A, B, C
line text screen and three result lights. Reference data for genuine and E. There was partial overlap between others (such as brands C
brands is uploaded to the authenticator via its electronic interface. and D) particularly at higher wavelengths. Whiskies of known
Similarly, result data from field measurements may be downloaded authenticity were then analysed in a similar way. Fig. 3 shows that
to the result spreadsheet for onward transmission and critical a sample of genuine brand B (sample T1) fitted, as expected, within
examination at a central laboratory. the minimum and maximum ranges for that brand, whereas four
A process for drawing conclusions on sample authenticity was non-authentic samples (samples T2–T5), which were known
devised. An algorithm was constructed based on the sum of squared counterfeits, each had spectra outside these ranges.
scores (S z2) for absorbance results at 10 pre-selected wavelengths The principles of this spectroscopic method were then trans-
ranging from 230 to 500 nm for the WPA Lightwave® S2000 and ferred from the laboratory WPA Lightwave® spectrophotometer to
between 220 and 360 nm for the SAD Brand Authenticator®. The the portable SAD Brand Authenticator®. Brand range data were
absorbance of the test sample and the average and standard generated on the Brand Authenticator® in the same way as that
deviation of at least 10 reference samples were entered into the used for the laboratory instrument. Fig. 4 shows equivalent spectra
calculation. The z-score5 was calculated from z = (x 2 X)/s for for 10 samples of each of the five Scotch whisky brands A to E.
absorbance data at 10 wavelengths chosen to reflect changes in the Minor differences in the absorbencies at higher wavelengths
slope of the spectral curve, where x is the determined value, X is the between the two instruments (Figs. 2 and 4) were thought to arise
average of the reference values and s is the standard deviation of from secondary transmissions caused by the higher excitation
the reference values. For example, the Brand Authenticator® used energy used in the Brand Authenticator®. This highlighted the
wavelengths at 220, 230, 240, 250, 260, 280, 300, 320, 340 and necessity to establish brand ranges on each model of instrument
360nm. The formula for the z score at each given wavelength (l used at this stage of method development.
nm) was therefore Analytical precision was investigated on the Brand Authentica-
tor® over the range 220 to 360nm. The repeatability of 10
measurements on one sample of brand C (expressed as ±2s/mean 3
100%) ranged from ±0.32% at 260nm to ±1.43% at 360nm (Table
1). The equivalent confidence limits for brand authenticity analysis
by gas chromatography were approximated double the spectral
Finally, S z2 = Sum of squared scores (z(l)) at each monitored
limits and ranged from ±1.6% (for methanol analysis) to ±2.7% (for
wavelength.
2-methyl butanol analysis).1 The spectral confidence limits for 10
The same test samples were subjected to brand authenticity
production batches of brand C analysed once ranged from ±9.8 to
analysis by gas chromatographic methods previously described.1
±10.9%, indicating that variation in instrument repeatability was
Methanol, n-propanol, isobutanol and the combined value for 2-
very small compared to batch to batch variation.
and 3-methyl butanol were determined. For the gas chromato-
The authenticities of a series of samples were then assessed both
graphic process a similar algorithm was applied to each of the 4
by the spectroscopic method on the Brand Authenticator® and by
congeners.
gas chromatography. Reference samples of the same five brands
were analysed for their major volatile congeners. Minimum and
Results maximum range data, shown in Table 2, indicated that similar
Initial stages of this spectroscopic method development determined analytical overlap of certain congeners used in authenticity
the consistency and specificity of the spectral characteristics of a assessment occurs and therefore specific congeners, such as 2- and
range of Scotch whisky brands. The brands were selected to 3-methyl butanols, are particularly important in reaching au-
Fig. 2 Absorbance spectra of 10 samples from different production batches of five different brands of Scotch whisky collected using the WPA Lightwave®
S2000 spectrophotometer.
Fig. 4 Absorbance spectra of 10 samples from different production batches of five different brands of Scotch whisky collected using the SAD Brand
Authenticator®.
Table 1 Precision of spectral analyses for brand C using the Brand Authenticator®. The repeatability of ten measurements on one sample and the results
obtained from single analyses on ten production batches of brand C
Ten measurements on one sample of brand C Single analyses on ten production batches of brand C.
Fig. 5 Absorbance spectra of 20 suspect samples purporting to be brand B collected using the SAD Brand Authenticator®.
2- and
Sample Isobutyl 3-Methyl Authenticity S z2 by Authenticity
number Methanol n-Propanol alcohol butanol S z2 by GC by GC UV/VIS by UV/VIS
Fig. 6 The sum of squared scores (S z2) for 20 suspect samples and 15 other brands of Scotch whisky determined both spectrally, using the SAD Brand
Authenticator®, and by gas chromatography.