Smart Density Gauge
Smart Density Gauge
Smart Density Gauge
Density Gauge
Emitter
Detector
JM_202 aSMART DENSITY GAUGE BRO UK 11/8/04 3:47 pm Page 4
PC
9 A +ve 1
10 B -ve 2
11 A Scm 3
12 B 4
45 3 1 2 13 5 HART
14 6 Modem
-ve 8
PRI 160 A 15
PRI 116C Power supply
+ve 7 DENSITY B 16
10-34v DC
Internals 2.5 VA GAUGE
Level/Density/Concentration Measurement HART™ compatible 4-20mA Output. External HART™ input for Pressure/Temperature correction.
External 4-20mA correction signal can be achieved when combined with a PRI 159.
Electrical Power Requirements 10-34V DC 2.5VA Power supply required for PRI 160. The PRI 160 supplies power
to the PRI 116c field mounted instrument.
Electrical Safety Flameproof enclosure certified EEx d IIC T5 (Tamb= -20°C to +60°C)
Detector cable 5 core screened cable is required for connection between the PRI 160 and the
field mounted PRI 116c unit
The PRI 116c is supplied with a pre-assembled flameproof gland and 4m flying lead as standard.
A flameproof junction box with appropriate glands is required to connect the PRI 116c-ATX to the PRI 160.
Indicators LED indicators provide instant recognition of power and connection to the PRI 116c-ATX.
TRACERCO and TRACERCO Diagnostics are trademarks of the Johnson Matthey Group of companies.
PO Box 1 Unit 4, The Technology Centre 1100 Hercules Avenue Tracerco Asia
Belasis Avenue Offshore Technology Park Suite 200 Tracerco/Dialog Alliance
Billingham Bridge of Don Houston 109, Block G,
Cleveland Aberdeen Texas Phileo Damansara 1
TS23 1LB AB23 8GD 77058-2759 46350 Petaling Jaya
UK UK USA Selangor
Tel +44 (0)1642 375500 Tel +44 (0)1224 826628 Tel +1 281 488 0039 Malaysia
Fax +44 (0)1642 370704 Fax +44 (0)1224 827095 Fax +1 281 488 1646 Tel + 603 79551199
Fax + 603 79546953
© 2003 Johnson Matthey Group
www.tracerco.com tracerco@matthey.com XM0117/0/0/JM202