Philips iU22 Annual QC Test Procedure 

Equipment
·  Gammex RMI grayscale ultrasound phantom
·  Ultrasound gel 

19  20 

23 21 

22 

Philips iU22 console 


1  Res/Spd Select between “Resolution” and “Speed”. Changes line density 
2  Gray Map Select the gray map to be used for image display 
3  Sector Width Change the number of sectors used in the transducer 
4  Compress Change the amount of image compression 
5  Protocol Select a pre­set protocol to determine scanning parameters 
6  Review Review images that have been obtained during the current exam 
7  Annotate Annotate the current image 
8  End Exam End the current exam 
9  Transducer Select a transducer to activate 
10  Patient Data Switch to a screen to input patient data 
11  Zoom 
12  Caliper Use the caliper tool to make measurements on the image 
13  Focus* Change the number of foci / change the focal depth 
14  Depth Change the depth 
15  Pointer Show a cursor 
16  Freeze Freeze on the current image / unfreeze the current image 
17  Print Save the current image 
18  iScan Return to the image acquisition screen 
19 XRes 
20 SonoCT 
21 2D Opt Transducer frequency setting (Pen, Gen or Res) 
22 2D Adjusts the image gain 
23 Dual display mode 
* This button has dual functions.  Toggle between the two functions by pressing the button.


Test Procedure:  Philips iU22 
(1) Power up machine. Verify accuracy of displayed time and date. 
(2) Perform mechanical inspection (wheel locks, etc). Visually inspect each transducer face 
(cracks, cuts, dents, other damage), connector pins (dust, damage), transducer and power 
cables (cuts, defects, other damage). Check the air filter for cleanliness*. 
(3) Start new patient using the ZPAF convention specified in the 
Ultrasound_annual_test_schedule.xls spreadsheet. 
(4) Select transducer and preset (usually Gen/Gen). (Button 9)  The active transducer is the one 
shown in green. 
(5) Remember: NEVER CONNECT/DISCONNECT AN ACTIVE TRANSDUCER 
(To de­activate the current transducer, attach another transducer and select that second 
transducer.  You will hear a clicking sound when the second transducer has been activated.) 
(6) Set TGC to be centered. 
(7) Add the transducer serial number to image annotation.  (Button 7) 
(8) Add the phantom serial number to image annotation. 
(9) For each transducer, turn off SONO­CT, X­RES and harmonics if activated.  (These buttons 
are located on the touch­screen.  These options are OFF when buttons are blue and ON if 
buttons are green). 
(10) All tests are performed at the highest frequency available for each transducer. This is done 
by setting the “2D Opt” to “Res.” (Button 21) 
(11) Verify all settings each time you select a different transducer. 
(12) To make sure you are not capturing cine images, freeze each image before saving.  Make 
sure the image is annotated properly before it is saved. (Buttons 16 (Freeze) and 17 (Print)) 

*Air Filter 
The air filter is located on the lower left­hand side of the ultrasound unit. 
Turn off power before you remove the air filter. Do not turn power on without an air filter installed.


(1) Uniformity 
a.  Position the probe in most homogeneous area of phantom (parallel to short axis of the 
phantom, if possible). 
b.  Set depth at maximum achievable, but no deeper than just below reflection from the 
bottom of phantom. (Button 14) 
c.  Use only one focus (Button 13), set at the deepest depth or just above the base of 
phantom. 
d.  2D Opt: Res (Button 21) 
e.  Set gain to demonstrate uniformity. (Button 22) 
f.  Adjust TGC if necessary to eliminate horizontal banding 
g.  Evaluate for artifacts, dropouts, or excessive banding that cannot be corrected with 
TGC adjustments. If artifacts are present, store additional images to show these 
artifacts. 

For curvilinear transducers, display the full FOV.  Roll the probe left and right to evaluate for 


dropout across transducer face. Store 2­3 images to demonstrate image quality across the entire 
transducer face.


For curvilinear transducers,  the FOV (sector width) can be reduced (Button 3) for the 
remaining tests. 

(2) Depth of penetration 
a.  Set depth to the maximum achievable depth, but no deeper than just below the 
reflection from the bottom of the phantom. 
b.  Set the focus at the deepest depth or just above the base of phantom. 
c.  Make sure Gain is set to 100% for this test.  (If setting the gain to 100% over­ 
saturates the image, a lower gain setting can be used). 
d.  If the beam does not penetrate to the bottom of phantom, display images in dual mode 
by selecting Dual mode.  (Button 23)  The second image should be obtained “in air.” 
Measure the depth where there are no more usable echoes, or where the electronic 
noise is equivalent in both images. Store the dual image. 
e.  If the beam penetrates to bottom of phantom, a single image can be stored. 
Reminder: Measure the depth of penetration on the system before the image is stored. 

The depth of penetration is less than the depth of the phantom (i.e., a reflection off the 
bottom of the phantom is not visible).


Example where the beam penetrates to the bottom of the 
phantom, shown by the reflection.


(3) Caliper accuracy(Axial and Lateral) 
a.  Set the depth at  5­6 cm 
b.  Set the focus between the sets of axial and lateral caliper accuracy pins 
c.  2D Opt: Res 
d.  Set the gain so that the pins are highly visible 
e.  Adjust the angle of the transducer on the phantom until the pins are as circular as 
possible.  (This will be easier with the linear arrays than with the curvilinear arrays). 
f.  Make 2 measurements: For lateral caliper accuracy, position the cursor above the 
pins. For vertical caliper accuracy, position the cursor beside the pins (i.e. the cursors 
should not obscure the pins). 
Reminder: Make the caliper measurements on the system before the image is stored.


(4) Resolution (Axial and Lateral) 
a.  Set the depth to ~ 4cm (the depth of the resolution pattern). 
b.  Focus at the level of the resolution pattern. 
c.  2D Opt: Res 
d.  Set the gain to minimize background and visualize pins. 
e.  Adjust the angle of the transducer on the phantom until the pins are as circular as 
possible.  (This will be easier with the linear arrays than with the curvilinear arrays). 
f.  Freeze the image. 
g.  Zoom in on the pins. (Button 11) 
h.  Make a measurement of FWHM.  Decrease the gain so that the background is not 
visible.  Areas of reverb beneath the pins should not be included in the FWHM 
measurement. 
i.  Freeze the image displaying the FWHM measurements. 
j.  When filling out the QC report, note the minimum distance between pins that is fully 
resolvable. 
Reminders: The FWHM measurement should agree with the resolution limits shown by 
the resolution pattern.


(5) Dead zone 
a.  Set the depth at approximately 2.0 cm 
b.  Focus at the shallowest level. 
c.  2D Opt: Res 
d.  Set the gain to visualize pins and the presence or absence of reverb near surface. 
e.  Record the number of pins that are visible.


(6) Anechoic cyst 
a.  Reduce the depth to ~5.0 cm 
b.  Focus at the depth that is just below the largest anechoic object. 
c.  2D Opt: Res 
d.  Turn up the gain to look for reverberations inside the cysts. (Storing this image is 
optional unless there are reverbs/artifacts to demonstrate) 
e.  Turn the gain down to demonstrate a lack of internal reflections/noise. All three cysts 
should be round (non­distorted) with distal enhancement. (Store this image.) 

A zoomed image of the anechoic objects in the phantom. 

A diagram of the anechoic objects, adapted from the Gammex manual.

10 
(8) Testing an Additional Transducer 

• Test remaining transducers within the same exam (if possible). 
• Connect the next transducer to be tested and select the new transducer. (Button 9) Before 
disconnecting a transducer, be sure to de­activate that transducer. 
• Repeat the set­up described on p.3 to ensure that all of the proper settings are selected. 
• Modify the image annotation with the serial number of the next transducer. Verify X­Res and 
SonoCT are off for each transducer and that the resolution is set at the highest available. 

(9) After all Acceptance Testing is Finished 

• Review the images (Button 6) to ensure that all of the necessary images were captured. 
• End exam. (Button 8) 
• Send the images to Stentor. 
• Shut down the machine 
• Clean the gel from the transducers and the phantom with a soft, clean towel. 

(10) Additional Reminders / Some Common Mistakes 

• Make sure that for the depth of penetration measurements, the focal depth is set appropriately. 
(It should be set to the maximum focal depth, or, if the bottom of the phantom is visible in the 
image, to just above the bottom of the phantom. 
• Record distance measurements (depth of penetration, caliper accuracy, FWHM) on the 
ultrasound image.  Make sure to make measurements before capturing images, so that these 
measurements are recorded on the image sent to Stentor. 
• If you use the zoom option, make sure you zoom the frozen image, NOT the live image. 
• When testing multiple transducers on a single unit, be sure to update the transducer serial 
number on the images before testing a new transducer. 

Deleting Individual Images 
• To delete individual images, review the images (Button 6) and select the image number next to 
the images you want to delete.  The selected images will have image numbers highlighted in 
orange.  Then select the “delete selected images” option.  Warning: This permanently deletes 
the selected images from the ultrasound hard drive.