UV Nr. 4
UV Nr. 4
Abstract: Solid-phase extraction method followed by direct UV spectrophotometry at 264 nm was developed
and applied for the selective ibuprofen determination in two-component formulation of ibuprofen and pseu-
doephedrine-HCl, combined powder which contains ibuprofen in the form of salt with L-arginine and 10%
ibuprofen cream. Procedures for ibuprofen determination in complex pharmaceutical preparations by direct UV
spectrophotometry lack selectivity because of interferences of other active substances and fat components. A
limited number of spectrophotometric methods applicable to these samples are based on derivative (first and
second-order) UV spectroscopy. Common HPLC procedures are more selective but more expensive and for
creams also require some type of extraction because the large amount of oily excipients would clog up the col-
umn. The proposed solid-phase extraction method proved to be suitable for analysis of ibuprofen in combined
tablets, powders and creams by direct UV spectrophotometry. Also the method provides an effective clean-up
of the cream and allows ibuprofen determination by HPLC analysis. For the extraction three different com-
mercial sorbents were tested: anion exchange OasisÆ MAX, hydrophilic-lipophilic balanced OasisÆ HLB and
reverse-phase ChromabondÆ C18ec. The optimization of the SPE method was first done on standard ibuprofen
solutions and then the suitability of the method was checked on solutions of commercial pharmaceutical sam-
ples. The method yields good results for all three types of commercial preparations on the anion-exchange
OasisÆ MAX cartridges, with recoveries of 90-100.2%. The interferences in UV analysis were not registered
and good precision (RSD < 6%) was obtained. The present method has been verified as accurate as the refer-
ence HPLC with the great advantage of less expensive instrumentation. For this reason, the method would be
suitable for a routine and rapid drug quality control.
Ibuprofen (non-steroidal anti-inflammatory fen arginate) is a highly soluble salt formed by com-
drug) is present on the market in a number of prod- bining racemic ibuprofen with the amino acid L-
ucts in which it is either the only active substance or arginine. L-arginine, a naturally occurring essential
in combination with other active ingredients or amino acid, renders ibuprofen more soluble in water
excipients. In addition to classical anti-inflammato- and facilitates its rapid absorption across gastric and
ry drugs which contain only ibuprofen and excipi- enteric mucosa, which delivers the well-known
ents, two-component formulations in which ibupro- analgesic properties of conventional ibuprofen in a
fen is combined with pseudoephedrine or where faster, more effective way (1). On the market there
ibuprofen is in the form of salt, most commonly L- is also topical semi-solid product of ibuprofen in the
arginine salt, have often been prescribed in the past form of cream, which minimizes the gastrointestinal
few years. Two-component formulation of ibupro- side-effects of oral ibuprofen and provides relative-
fen and pseudoephedrine-HCl is effective in cold ly consistent drug levels at the application site for
and flu symptoms and is used as an anti-inflamma- prolonged periods (2).
tory and nasal decongestant for upper respiratory In all of these complex formulations other
infections. The L-arginine salt of ibuprofen (ibupro- active ingredients or excipients may interfere with
403
404 SLAVICA SUNARIC et al.
ibuprofen determination. Tablets and powders con- liquid extraction which is limited to equilibrium dis-
tain significant amount of various excipients and tribution of the substance between two liquid phas-
fillers. Many of them are polar compounds and es. In contrast to the large number of papers relating
show absorption in the UV range. Further, pseu- to the extraction and determination of ibuprofen
doephedrine and ibuprofen have overlapping UV from biological samples, there are a limited number
spectra in aqueous and alcoholic solutions and can- of literature references concerning its extraction
not be determined by direct UV spectrophotometry. from combined pharmaceutical preparations or
In cream, the active substance is dispersed in com- creams.
plex emulsion containing hydrophobic and The aim of this work was to develop a method
hydrophilic compounds, emulsifying agents and for determination of ibuprofen without interferences
preservatives. Sodium and potassium salts of fatty in complex pharmaceutical matrices such as creams
acids, cationic surfactants and non-ionic surfactants and combined dosage forms. The method involves
are commonly used as excipients in creams (3). sample pretreatment by solid-phase extraction
These excipients do not have strong chromophores (SPE) and analytical determination by UV spec-
but they have to be removed, because they can sig- trophotometry at 264 nm. The optimal conditions
nificantly contaminate chromatographic column in for the extraction as well as the recovery have been
HPLC analysis, whereas in UV spectrophotometry tested for three different types of sorbents/car-
they disable ibuprofen determination because the tridges: OasisÆ HLB (Waters), ChromabondÆ C18ec
obtained solution is an emulsion. (Macherey-Nagel) and OasisÆ MAX (Waters). The
Different methods for simultaneous determina- described method has been applied to the determi-
tion of ibuprofen and pseudoephedrine in combined nation of ibuprofen in commercial pharmaceutical
dosage forms are developed. High-performance liq- preparations.
uid chromatography is one of the most popular and
sensitive method which can separate ibuprofen and EXPERIMENTAL
other active substances in tablets (4), granules (5),
soft capsules (6), creams (7) and syrup preparation Apparatus
(8). For the HPLC determination in creams protec- The spectrophotometric analyses were per-
tive precolumn and washing after 10 injections were formed on Evolution 60 spectrophotometer
used. Capillary zone electrophoresis was also pro- (Thermo-Fisher Scientific USA). Spectra were
posed for the active constituents determination in measured with a 10 mm quartz cell. Solid-phase
ibuprofen tablets (9). extraction was done on Visiprep SPE Vakuum
For the analysis of multicomponent ibuprofen Manifold 12-port model, Supelco 57030-U, Sigma-
preparations many UV spectropfotometric methods Aldrich Chemie. pH was measured by using Hanna
have been proposed. Because of near absorption Instruments pH meter. HPLC analysis was per-
maximums of ibuprofen (264 nm and 272 nm) and formed on Agilent 1200 Series apparatus (Agilent
pseudoephedrine (240-260 nm) these methods are Technologies Inc., Santa Clara, CA, USA) equipped
based on first or second derivative spectrophotomet- with the diode array and multiple wavelength
ric assay (10, 11), ratio spectra derivative spec- detector.
trophotometry and chemometric techniques (12, 13)
or formation of color copper complexes with differ- Materials and reagents
ent absorption wavelengths (14). Only one paper Pharmaceutical certified product of ibuprofen
deals with the method of direct and derivative UV was of analytical grade (99.6%) and provided by the
spectrophotometric analysis of ibuprofen in pharma- Pharmaceutical Laboratory Galenika, a.d., Belgrade,
ceutical cream after solid-phase extraction (15). Serbia. Stock solution of ibuprofen, concentration of
Solid phase extraction is widely used tools for 0.5 mg/mL was prepared in methanol and stored at
sample cleanup, but in the analysis of drugs from 4OC.
pharmaceuticals is rarely used. In comparison with SPE was performed on OasisÆ MAX (1 cc,
classical liquid-liquid extraction it has many advan- Waters Corporation, Milford, Mass., USA), OasisÆ
tages, the most important is a much higher selectiv- HLB (1 cc, Waters Corporation, Milford, Mass.,
ity, the use of smaller quantities of toxic organic sol- USA) and ChromabondÆ C18ec (1 mL/100 mg,
vents and the shorter time of performance. High Macherey-Nagel, GmbH, D¸ren, Germany).
selectivity and specificity of SPE methods is Methanol 99.9% Reag. Ph. Eur.-Reag. USP for
achieved by a wide range of the sorbents for polar, analysis was from Carlo Erba Italy. All other
hydrophobic and ionic interactions, unlike liquid- reagents used were of analytical grade.
Determination of ibuprofen in combined dosage forms and cream.. 405
Commercial pharmaceutical preparations of tion should be 0.5 mg/mL. The obtained solutions
ibuprofen were purchased as follows: film tablets were filtered through MN 619 de. ∅ 125 mm
under the trade name ìDefrinol forteî by Galenika- (Macherey-Nagel, Germany) filter paper.
Serbia, powder under the trade name ìSpedifen
Zambonî by Bonifar-Serbia and cream under the Solutions for SPE
trade name ìBrufen 10%î by Galenika-Serbia. Working solutions of ibuprofen standard were
prepared from the stock ibuprofen solution (0.5
Procedure for dosage forms preparation mg/mL). Working solutions of pharmaceutical sam-
The developed SPE method was applied to the ples were prepared from the solutions obtained as
analysis of ibuprofen in commercially available described in Procedure for dosage forms prepara-
pharmaceuticals in three dosage forms: tablets, pow- tion.
der and cream. Each formulation was differently Different methods of preparation of standard
prepared. The obtained sample solutions were and sample solution for HLB, MAX and C18ec sor-
applied to the selected SPE cartridges. bents are based on acid-base properties and polarity
of active substance and provide more effective sep-
Tablets aration of ibuprofen from interfering compounds.
According to the label, ìDefrinol forteî con- For HLB and C18 cartridges 1 mL of ibuprofen
tains 200 mg of ibuprofen and 30 mg of pseu- solution was mixed with 1.5 mL of 2% CH3COOH
doephedrine HCl per tablet. A total of ten tablets to obtain final solution which has pH 3.0. For MAX
were weighed and then the protective film was cartridge, 1 mL of ibuprofen solution was mixed
removed with deionized water. After drying, the with 1 mL 2% NH4OH and 0.5 mL of methanol or
tablets were pulverized and homogenized. After fine deionized water. The obtained solution had pH 10.0.
powdering, accurately weighed three portions which
contained 25 mg of ibuprofen each, were dissolved UV spectrophotometric analysis
in 30 mL of methanol and stirred on magnetic stirrer After the SPE step obtained eluates were ana-
for 30 min. After extraction with methanol, the lyzed by direct UV spectrophotometry. The amount
obtained solutions were quantitatively transferred to of ibuprofen was determined by using calibration
a 50 mL volumetric flasks and diluted to volume curve at 264 nm. For this purpose calibration graph
with methanol. The concentration of these solutions was constructed. From the stock standard ibuprofen
of samples should be 0.5 mg/mL. Further, the solution (0.5 mg/mL) working standard solutions in
obtained solutions were filtered through MN 619 de. the range of 0.1-0.5 mg/mL were prepared by appro-
∅ 125 mm (MachereyñNagel, Germany) filter priate dilution with methanol.
paper. The analysis of all samples was repeated 5 times.
wavelengths. For this purpose, separate calibration reverse-phase SPE systems. This cartridge can
curves at 264 and 272 nm were constructed by a retain both polar and non-polar molecules due to the
least squares fit. Linear regression analysis was per- presence of lipophilic divinylbenzene and
formed for calibration curves at two hydrophilic N-vinylpyrrolidone groups. C18ec car-
wavelengths.The limit of detection and limit of tridge contains endcapped octadecyl phase and
quantitation were calculated as kSD/b, where k = 3 shows interaction with very nonpolar, hydrophobic
for LOD and 10 for LOQ. SD is the standard devia- organic compounds.
tion of the intercept, and b is the slope of the cali- Critical parameters in SPE methodology were
bration line. Limit of quantitation was 0.028 mg/mL active substances solubility and pH selection.
for both wavelengths but the slope of the line is bet- Ibuprofen is a weak acid which is water insoluble
ter for calibation graph obtained at 264 nm, and but shows good solubility in methanol (17).
therefore, this wavelength was selected for further Pseudoephedrine, on the other hand, is a base and is
ibuprofen determination. both water and alcohol soluble. Chemical structures
of these drugs are shown in Figure 1. Ionization of
Optimization of SPE method ibuprofen depends on the pH values of the solution.
SPE method optimization has been done with It is negatively charged above pH 6.4 and in neutral
the standard methanol ibuprofen solution concentra- form at pH less than 2.4. The equation (1) was
tion of 0.5 mg/mL. A number of variables were test- applied in order to calculate the percentage of both
ed during the optimization procedure: the nature of ionized and non-ionized forms in different pH:
the sorbent, the influence of a washing step, the pH 10pH ñ pKa
% ionization = ñññññññññññññññ (1)
values of the sample solutions, the pH values of sol- 1 + 10pH ñ pKa
vents mixtures for washing and eluting, as well as Pseudoephedrine is a weak base with pKa 9.8
the volume of the solvents and mixture of solvents and it is cationic in acidic and neutral solutions. L-
for conditioning the cartridges. Optimization was arginine has guanidine group and it is a strong
achieved through several steps: first, the recovery of organic base with pKa 12.48. Guanidine functional
ibuprofen from the standard solution was tested on group is positively charged even at alkaline condi-
three types of cartridges (HLB, C18ec and MAX) tions. In order to investigate the retention of ibupro-
with the pH values being changed in the interval fen on hydrophilic-lipophilic sorbent (HLB),
from pH 3.0 to pH 10.0. These sorbents have differ- octadecyl-modified silica (C18) and strong anion
ent retention mechanism. MAX cartridge has been exchanger (MAX), pH of the ibuprofen solutions
optimized to achieve high selectivity and sensitivity was varied in the range from 3.0 to 10.0.
for extracting acidic compounds with anion- The first step in SPE methodology is charac-
exchange groups. This is mixed-mode polymeric terization of sample and selection of appropriate
sorbent and powerful anion-exchanger which con- preparation procedure. This step depends on physi-
tains ionized quaternary amino groups located on cal form of dosage forms and physico-chemical
the surface of the divinylbenzene-N-vinylpyrroli- properties of active substance. Characterization of
done copolymer. HLB cartridge (hydrophilic- the samples was done in terms of molecular weight,
lipophilic balance) finds its application in the functional groups, polarity, acid-base properties,
pKa values and solubility of analyte and other sub-
stances in the sample matrix. Dissolving a sample of
ìDefrinolî tablets in methanol results in extraction
of both ibuprofen and pseudoephedrine, therefore,
methanol extraction is not enough for separating
these two active ingredients. UV spectrophotometric
determination of ibuprofen in the obtained extract is
inaccurate because these two substances absorb at
very similar wavelengths (264 nm for ibuprofen and
257 nm for pseudoephedrine) (18). The additional
extraction and separation of ibuprofen from pseu-
doephedrine from the tablet solution is necessary.
According to the product declaration, íSpedifení
contains 200 mg of ibuprofen in salt form with L-
Figure 1. Chemical structures of ibuprofen and pseudoephedrine- arginine. Since ibuprofen is in the form of salt, good
HCl water solubility is to be expected. L-arginine, on the
Determination of ibuprofen in combined dosage forms and cream.. 407
other hand, is water soluble but not alcohol soluble, tion of acetic acid was used, while MAX sorbent was
so adjusting the pH values of the solution and choos- washed up by 2% solution of ammonium hydroxide.
ing appropriate washing and eluting solvents mix- We assumed that the washing of sorbents was neces-
tures, can result in good extraction of ibuprofen and sary for accurate recovery values because of our
its separation from L-arginine. According to the results showed that recovery was slightly higher than
product declaration, 1 gram of the íBrufení cream expected when the washing step wasnít done.
contains 100 mg of ibuprofen in molecular form and
a number of hydrophobic and hydrophilic excipi- SPE protocols
ents. Dissolving the cream samples in methanol Before use, the SPE columns were conditioned
results in formation of emulsion which cannot be and equilibrated with 2 mL of methanol and 3 mL of
accurately analyzed by UV spectrophotometry. For deionized water. Solutions of ibuprofen, which were
the quantitatively extraction of ibuprofen from the passed through HLB and C18 columns are adjusted
methanol emulsion, appropriate sorbent and solvent by 2% acetic acid to pH 3.0. For extraction on MAX
selection is very important. cartridges pH of the ibuprofen solution is set to 10.0
On the basis of the physico-chemical properties with 2% ammonium hydroxide. The sample solu-
of sorbents, as well as the polarity and solubility of tions were passed through the conditioned cartridges
ibuprofen, pseudoephedrine and L-arginine, it was at the flow rate no greater than 1 mL/min. After
assumed that ibuprofen will be retained on the HLB sample loading, the sorbents were washed with 1
and C18ec cartridges if the solution has pH 3, while mL of 2% CH3COOH (HLB and C18ec) and 1 mL
good extraction of ibuprofen on the MAX cartridge is of 2% NH4OH (MAX). Ibuprofen was eluted from
achieved in basic solution with the pH 10. Two per- HLB and C18 cartridges with 3 ◊ 1 mL of methanol,
cent solutions of weak acid and base were used for pH while elution from MAX sorbent was performed
regulation according to Marlatt et al. (19). As a wash- with 3 ◊ 1 mL of a mixture methanol : 2%
ing solution for HLB and C18ec sorbents a 2% solu- CH3COOH (2 : 1,v/v).
Found IB from
Declared
Dosage the dosage form Recovery RSD
Cartridge content
form (mg/mL) (%) (%)
(mg/mL)
xñ ± SD
HLB 0.444 ± 0.017 88.8 3.83
Tablets
C18ec 0.438 ± 0.005 0.500 87.6 1.14
IB + PE
MAX 0.465 ± 0.007 93.0 1.51
HLB 0.816 ± 0.005 163.2 0.62
Powder
C18ec 0.414 ± 0.020 0.500 82.8 4.83
IB-arginate
MAX 0.454 ± 0.025 90.8 5.51
Figure 2. Absorbance spectra of ibuprofen standard solutions concentration of 0.5 mg/mL after extraction: 1) HLB, 2) MAX, 3) C18
on the C18ec and HLB which indicates incomplete ples and those of standard after SPE (Figs. 3 and
retention or elution from these sorbents. A good 2), it was proved that ibuprofen was well retained
extraction of ibuprofen and its separation from on the MAX sorbent and was not eluted in the
pseudoephedrine with acceptable RSD value was washing step. The absorption maximums corre-
achieved on the MAX cartridge. By comparing UV spond to those for the standard passed through the
spectra and absorbance of ibuprofen from the sam- cartridges.
Figure 3. Absorbance spectra of ibuprofen from tablet dosage form (0.5 mg/mL) after extraction: 1) C18, 2) HLB, 3) MAX
Figure 4. Absorbance spectra of ibuprofen from powder dosage form (0.5 mg/mL) after extraction: 1) C18, 2) MAX, 3) HLB
Figure 5. Absorbance spectra of ibuprofen from cream solution (0.5 mg/mL) after extraction: 1) MAX, 2) C18, 3) HLB
410 SLAVICA SUNARIC et al.
13. Wahbi A.A, Hassan E., Hamdy D., Khamis E., 17. Delgado J.N., Remers W.A.: in Wilson and
Barary M.: Pak. J. Pharm. Sci. 18 (4), 1 (2005). Gisvoldís Textbook of Organic Medicinal and
14. Singhvi I., Chaturvedi S.C.: Asian J. Chem.10, Pharmaceutical Chemistry, 10th edn.; Lippincott
879 (1998). Williams & Wilkins, Philadelphia 1998.
15. Bonazzi D., Andrisano V., Gatti R., Cavrini V.: 18. Clarkeís Analysis of Drugs and Poisons;
J. Pharm. Biomed. Anal. 13, 1321 (1995). Pharmaceutical Press, London 2006.
16. British Pharmacopoeia, electronic version 19. Marlatt L., Trinh A., Bell D.S.: http://www.sig-
11.0.; The Stationery Office on behalf of the maaldrich.com/etc/medialib/docs/Supelco/Post
Medicines and Healthcare products. Regulatory ers/t404049h.Par.0001.File.tmp/t404049h.pdf
Agency (MHRA), Crown Copyright, 2006. (accessed June 25, 2010).