MANUAL
1
SK3002 CHEMISTRY MANUAL
TEL: 0086-755-86238653
FAX: 0086-755-33229823
CATALOGUE
First chapter: Outline
First section:Outline ---------------------------------------------------------- 2
Second section:Assembly attention ------------------------------------------ 2
Third section:Device introduce ------------------------------------------------- 4
Second chapter: Sample test---------------------------------------------------------- 6
Third chapter: Main menu introduce------------------------------------------------- 8
First section: TEST arrange------------------------------------------------------- 8
Second section: Report------------------------------------------------------------ 12
Third section: Set parameter ---------------------------------------------------- 15
Fourth section: QC report--------------------------------------------------------- 21
Fifth section:System maintenance ------------------------------------------- 22
Fourth chapter:Daily operation and maintenance----------------------------- 26
First section:Daily operation -------------------------------------------------- 26
Second section: Routine maintenance ------------------------------------- 27
Fifth chapter Icon explanation ----------------------------------------------------- 29
App1: English character input method -------------------------------------------- 30
App2: paper installation method ----------------------------------------------- 30
NOTATION:
As this instrument uses the quartz flowing cell , every day before the instrument
be power off, please enters into the ‘ MAINTAIN '—‘3-ADC', and press the
numeric key `8' to aspirate air in order to avoid the cell ice up to breakage.
1.Preparement
⑴ Grounding
As the SK3002 semiautomatic biochemistry analyzer is the precise electronic
instrument ,It must be earthed in working, the earth resistance is smaller than 0.5
ohm. The professional electrician to install the electric power grounding in the
room.
⑵ Power source manostat
If the power voltage is unstable, which will influence the instrument. before the
installment please purchase the high grade power source manostat which power
supply is large than 500W, and its voltage turning on the earth.
⑶ Working conditions
This instrument cannot uses a power source plug or be put into one room with
high efficiencies, the big interferometric technique and so on centrifuge, electric
refrigerator, thermostat; Avoids the sunlight perpendicular incidence.
2.Instrument installment
⑴ Open paper box, checks in the packing detailed list and the packing case
instrument, if there is a lacks or the damage, please relate with the supplier;
⑵ If there was unmistakably, takes out the instrument carefully, puts on the stable
smooth plate, prevents the intense vibration;
⑶ Please insert one end of the power line into the grounding power source plug,
another end into the SK3002 back panel;
⑷ assemble aspirate tube as the follows :
3
SK3002 CHEMISTRY MANUAL
Look for a waste liquid bottle, puts the SK3002 waste liquid tube into the waste
liquid bottle;
⑸ Prepare one cup of distilled water to be used;
⑹ After turns on the power switch, the instrument will start the system self-check,
after self-checking completed the system demonstration to be as follows:
SK3002 CHEMISTRY ANALYZER
SYSTEM CHECK
PLEASE WAITING
CHECK ABS
‘ASPIRATE DISTILLED WATER’ to flash, let aspirating tube into the distilled water,
presses down ‘ENTER' key to aspirate the distilled water, then the instrument will
be able to detect the extinction ADC of 0-5 wave length for 10 seconds, various wave
4
SK3002 CHEMISTRY MANUAL
CHECK ABS
WAVE 340 -0.078 0
WAVE 380 -0.082 5
The extinction ADC examination completed, then the system will turn to the main menu, the
demonstration is as follows:
1, ‘0' – ‘9' these 10 keys already may be used to choose the menu, also we may use
the button to input the numeral and the letter (refer to appendix - input method).
2, 4 direction keys (‘↑' ‘↓' ‘ ← ’‘ → ’), uses in moving the cursor; In the input method,
uses to choose and delete the character.
3, ‘BLANK', ‘STANDARD', ‘CLEAN' these three special-purpose keys, BLANK' used
for to aspirate the blank fluid, ‘STANDARD' for standard fluid and ‘CLEAN' for
cleaning liquid separately;
4,‘FEED' used in replacing the printing paper. During the printing process ‘IN-LINE'
indicating lamp must be at the bright condition, pressed key may cause this
indicating lamp bright according to ‘IN-LINE'.
5, ‘RETURN' used to withdraw from some operation sequence or the sub-menu,
‘ENTER' used to enter some operation sequence or the menu.
5
SK3002 CHEMISTRY MANUAL
1. pre-arrangement pattern
Before samples test, input the TEST to test, the instrument will arranges the test order
automatically, then test will carry on according to the instrument promptly. The brief step
of operation is as following:
(1)、TEST arrangement
A In the main menu press ‘ ↑ ’‘ ↓ ’‘ ← ’‘ → ’Key choose ‘ARRANGE’, enters the TEST
arrangement contact surface.
B Presses numeric keyboard ‘1' to choose ‘SINGLE INPUT '.
C, Input a patient number (≤59999) , press ‘ENTER’, then press ‘ ← ’‘ → ’ key to
choose patient rank ‘COMMON' or ‘EMERGENCY ' ,the patient number cannot be
‘0'.
D, Input the TEST NO. which the patient wanted to test, press ‘ENTER’ key to confirm
every time
E, The input completed, Press key ‘RETURN’ returns to the level menu.
F, Repeating step C-E can input the TEST of the other patient. When finished, presses
`RERURN' key until returning the main menu continuously.
(2)Sample test
A, pressing numeric key‘2’ in the main menu or press the ‘ ↑ ’‘ ↓ ’‘ ← ’‘ → ’Key to
choose ‘TEST SAM ', enter into the test contact surface.
B, Operation according to the display in the monitor fourth line of prompt information.
C, After the TEST test, the system demonstrate automatically the measurement
result and absorb cleaning liquid promptly, when the absorption finished the system
will be able to enter the next TEST test, until test finished, the system can the self
return to the main menu.
(3)Print report
A, In the main menu, press‘3’or press the ‘ ↑ ’‘ ↓ ’‘ ← ’‘ → ’Key to choose the
‘REPORT’, the system enters the result report.
B, Presses numeric keyboard ‘4', enters the ‘PRINT/DELETE REPORT’.
C, Presses numeric keyboard ‘1' to choose the ` PRINT SELECTED’
D, Input the patient no, press ‘ENTER’ key.
E, After the operation finished, please continuously press ‘RETURN' key until returns
the main menu.
2, stochastic pattern:
Choose ‘TEST SAM ’in the main menu, the system demonstration is as follows:
TEST NO:
Please input the TEST NO. which want to test, after presses ‘ENTER’ key, the
instrument demonstration is as follows:
6
SK3002 CHEMISTRY MANUAL
TEST NO 2 ALP
PAT NO:
TEST SUM:
ASPIRATE WATER
‘ASPIRATE WATER ' flashes, press the ’ENTER' key aspirate distilled water and
clean pipeline .System may jump over the clean procedure according to pressing ‘↓
'.If the first measurement system can aspirate the blank liquid promptly, or pressing
‘BLANK' key, the system prompt is as follows:
TEST NO:2 ALP
PAT NO:0
TEST SUM:0
ASPIRATE BLANK
‘ASPIRATE BLANK ' flash, press the ‘ENTER' key to aspirate the blank liquid, after
the aspiration completed, ‘ASPIRATE BLANK ' turn to `TEST BLANK'. When the
testing completed, it will show the blank extinction value. If the first measurement
system can promptly test the standard fluid, or key designated according to
‘STANDARD' key, the system prompt is as follows:
TEST NO: 1 ALT
PAT NO:0
TEST SUM:0
ASPIRATE STANDARD 1
ABS:-0.080
ABS:0.08
FACTOR:1300
‘ASPIRATE SAMPLE ' flashes, this time keys may change the test No.. according to
‘6', after changes the test No., according to ‘ENTER’, presses `7' keys to be possible
again to change the patient No., After confirms the test No. and the patient no correct,
7
SK3002 CHEMISTRY MANUAL
3. Report
This instrument uses two kind of pattern printing result to report ,as follows:
① Menu operation pattern
(1) In the main menu, key‘3’or press the ‘ ↑ ’‘ ↓ ’‘ ← ’‘ → ’Key chooses the
‘REPORT' ,enters the result report.
(2) Presses numeric keyboard ‘4' to enter the ‘PRINT/DELETE REPORT’.
(3) Presses numeric keyboard ‘1' to choose the ‘PRINT SELECTED'.
(4) Input the patient NO which want to print (insufficient 5, please press ‘ENTER’
key).
(5) Operation finished please continuously do press ‘RETURN' key until returns main
menu.
②. quick ways
In the main menu press numeric keyboard ‘8’ may print the result in the main menu
directly, demonstration as following. Press ‘RETURN’ to return the main menu.
PRINT REPORT
PATIENT NO:
1-SINGLE INPUT
2-BATCH INPUT
3-MODIFY
8
SK3002 CHEMISTRY MANUAL
PAT SUM: XX
PAT NO:_
LEVEL: ROUTINE
Here may input ‘1---59999' patient no which has used i.e., it must be already existed,
might be the new number, if it was new patients, the number do not be the same.
After inputs the patient no, the patient no and the TEST total quantity will display in
the monitor first line, simultaneously the cursor will be located the last at the end of line,
reminds you input new TEST, you are able to input the numeral, but not can use the
search function, if did not remember the test number, please read list of TEST no and
the name.
Each patient may input 16 TEST mostly, when is more than 16, may input 16 first,
then test, then inputs again, if the insufficient 16 projects, may enter the next patient
after ‘RETURN'.
One time, you may input 40 project mostly
2 -BATCH INPUT
If many patients have many same test TEST, arrangement work can use this function
After chooses this function the system to demonstrate:
PAT SUM: XX
START NO: _
PAT SUM: XX
START NO: XXXXX
END NO: _
Please input the number will be equal or more than the start patient no. But the end
no subtract the start no adds the quantity that input before cannot be more than
40.Then, the demonstration on the screen first line as follows:
The cursor is located at the end of line, you may input TEST, similarly the TEST
quantity is more than 16, the system can prompt:
‘RETURN’TO EXIT
3-MODIFY
If want to delete the TEST or the patient, please choose enter this menu, the screen
demonstrated:
Now you may input exist patient no(1 - 59999),also might use the search function
according to the‘↓'key. You will use‘↑' ‘↓' to search patient no which input before, after
found the no which will need, according to `ENTER' might enter next step, this time
the system demonstrated:
Simultaneously the cursor is located first project no place, this time may carry on
operates:
a. ‘↑' ‘↓ ' key moves the cursor.
b. The numeric keyboard revises the TEST no.
c. ‘ENTER' key cancels some TEST.
d. ‘RETURN’ key to return.
(2) Choose ‘2-delete test patient’ into next contact surface as follows:
PAT NO:_
10
SK3002 CHEMISTRY MANUAL
Input the patient no that want to delete, after press `ENTER' key the system will
delete the selected patient no and the project which this patient will be test, then will
return to demonstrate the ‘PAT NO: '(patient no:), until the patient no completely
were deleted (screen to demonstrate ‘NO INFORMATION DISPLAY') or key
‘RETURN’ to return.
(3) Choose ‘3-delete test TEST’ into next contact surface as follows:
TEST NO:_
You might input 1-2 test code directly (which must be already exist), also may press
‘↓’ key to start the search function, press ‘↑'‘↓' key search the need TEST no, then
press ‘ENTER' key to delete the TEST no that all patient will test. After completes
the system to return to the ‘TEST NO:_' place, until all waiting test project deletes
or ‘RETURN' key is pressed down.
11
SK3002 CHEMISTRY MANUAL
You might input patient code directly, also use the search function through ‘↓’ key,
after searches the need patient code , please press ‘ENTER' key. when the patient
code does not exit, the system will prompt:
PAT SUM: XXX
PAT NO:XXXXX
PATIENT NOT EXIST
RETURNTO EXIT ENTER TO RETRY!
If replied ‘Y' or according to ‘ENTER' key you may found this patient record in the
system, then enters next step.If replies ‘N' then the system will return, you might input
the new patient code. After found the patient code or create new patient code the
system enters next step, demonstrates in the second line:
PAT SUM:XXX
PAT NO:XXXXX NAME:XXXXXXXXXXXX
SEX : MAN AGE:21
BLOOD TYPE: AB
Name: May input most 12 letters in this (letter input method please to refer appendix
1), insufficient 12 please press `ENTER' key;
Sex: Cursor key chooses with ‘↑' ‘↓' : MAN, WOMAN, CHILD, NULL;
Age: May input 0-199, the unit is: ‘Y' (year old);
Blood-group: Cursor key chooses A, B, O, AB , NULL with ‘ ↑ ’‘ ↓ ’
If which you input patient code has existed, then the system can preserve had the
information to demonstrate directly, you may revise willfully, the revision finished
according to ‘RETURN' key returns.
2 DELETE RESULT
You may delete the incorrect result by this function. In ‘REPORT’ menu chooses ‘2 —
DELETE RESULT’, the system demonstration is as follows:
Inputs the patient code that must be already exists, or use the search function choice
correct patient code then presses ‘ENTER’ key the system demonstrated:
You may use ‘↑' ‘↓’ key to examine the result of the same patient , use ‘ ← ’‘ → ’ key to
check the different patient result. If the result which does not need, pressing ‘ENTER’
the key to delete it. Any time presses ‘RETURN’ to return:
The patient code which input here must be exist also be the new code or which might
use the search function. Afterwards the system demonstration is as follows:
PAT NO:XXXXX NAME: XXXXXXXX
TEST NO: _
This time also may use search function ‘↓' or input the code directly, but the code
which already defined. next the system to demonstrate :
PAT NO:XXXXX NAME: XXXXXXX
TEST NO: XXXX
RESULT: _
Please input the result which obtains through the other means, press ‘MENU’key to
return.
4,PRINT/DELETE REPORT
1 — PRINT SELECTED
2 — PRINT ALL PATIENT
3 —DELETE SELECTED
4 — DELETE ALL PATIENT
13
SK3002 CHEMISTRY MANUAL
PAT NO: _
You may input the patient code to print directly, also may through pressing `↓ ' key to
use the search function, after found the patient code which want to report, press
‘ENTER’. Press ‘RETRUN’ to return.
4.1.2 Print all patients
a. If want to print all patient’s report you may choose ‘2 — PRINT ALL PATIENT’,
then the system demonstrated as follows:,
START DATE(MMDD): _
This time you may input the patient's copyright date which want to print, after like on
August 5 should input ‘0805', the system demonstrated:
START DATE(MMDD):0805
ENDDATE(MMDD): _
After you input the end date that equal to or more than the start date, the system
search all patient code which fit in the condition in memory , then starts to print, found
one, prints one, until prints finished, if wants to stop printing in the middle, only close
the power.
b. When system prompt input start date, if you want to print all results, then press
the ‘ENTER' key, the system demonstrated:
You may directly input with the numeric keyboard, also may use the search function,
after found press ‘ENTER' to delete.
14
SK3002 CHEMISTRY MANUAL
PAT NO:_
PAT NO:_
PATIENT NOT EXIST!
RETURN TO EXIT ENTER TO RETRY
1 — DEFINE TEST
2 — DISPLAY TEST SUM
3 — DELETE TEST
1 . DIFINITION TEST
Each test project hands-on operation parameter must input into instrument, in order to
the instrument can defer to the correct instruction execution test .
(1)、INPUT PROJECT PARAMETER
Choose ‘ 1 — DIFINE TEST TEST’, the system prompt as follows:
ITEM NO :__
NAME :
In the cursor place input 1-3 code, the code cannot be 0, please press ‘ENTER' key.
If the code inputs which already existed in the system, then the system assigns out
it, various parameters will display on the screen, then the cursor stops behind the name.
1 May use‘ ← ’‘ → ’ key to choose the different project name:
ITEM NO: 1
NAME: ALT
FACTOR: 1746 UNIT: U/L
METHOD: K-KINETIC
3 With‘←’‘→’key change unit, according to ‘↓' or ‘ENTER' key may enter the next
TEST, altogether has 11 kind of units to choose
:null 、u/L 、u/mL 、umol/L 、mmol/L 、g/L 、mg/L、mg/dL、g/ml、g/dL、X1012/L。
4 With ‘←’‘→’ key may choose the analysis method. The instrument provides “K-
KINETIC”” E- ENDPOINT” method, when purchase the reagent you must attention to the
method whether suits for the this analyzer, According to ‘↓' or `ENTER' key may enter
the next TEST.
ALT
TEMP: 37℃ DECIMAL:0
WAVE1:340 nm
REAGENT BLANK: Y SAMPLE BLANK: N
ALT
GEN MAN WOMAN CHILD
HIGH: 40 40 40 40
LOW: 0 0 0 0
The value of linear range can find in the reagent introduce , please input in
‘LINEAR:’place,generally the ‘DELAY’ and ‘INTERVAL’ were equal to 5s, Sample
points(at ‘S-POINT’place) was equal to 1 for the E-END POINT method or 2 for the K-
KINETIC method
the aspirate time(at ‘asp time: ’place)to be only effective to this test, does not have
the influence to other tests; generally cannot to modify .Regarding the general test, the
16
SK3002 CHEMISTRY MANUAL
ALT
LINEAR:〈 406 DELAY:60 S
INTERVAL: 60S S-POINT: 2
ASP TIME: 250 STANUM: 0
Inputs the multi-spot standard curve the contact surface as follows, its
establishment is this project standard or the standard curve Press ‘↓' key enters the
next TEST.
The system prompts ‘SAVE MODIFIED?Y' If save, please press ‘Y' or ‘ENTER'
key, otherwise according to ‘N' key; Inputs after completely completes the system to
return to this start position, may input or the revision next test project. The operator
must input the related reagent parameter after the first use this instrument or the system
replacement, by all means must careful reading reagent instruction for use,
corresponding input. Explains with examples as follows:
End-point method parameter establishment example one (Shanghai Long March
reagent
TEST NO: 16
NAME: TG
FACTOR: 0 UNIT:mmol/L
METHOD: E-ENDPOINT
TEMP:37℃ DECIMAL: 2
WAVE1:510nm
REAGENT BLANK: Y SAMPLE BLANK: N
GEN MAN WOMAN CHILD
HIGH: 1.81 1.81 1.53 1.53
LOW: 0.45 0.45 0.40 0.40
LINEAR:< 10 DELAY: 5 S
INTERAL: 5 S S-POINTS: 1
ASP TIME: 250 STDNUM: 1
STD1: 2.5
TEST NO: 1
NAME: ALT
FACTOR: 0 UNIT: u/L
METHOD: E-ENDPOINT
TEMP:37℃ DECIMAL: 0
WAVE1:510nm
REAGENT BLANK:Y SAMPLE BLANK:N
17
SK3002 CHEMISTRY MANUAL
TEST NO: 3
NAME: TBIL
FACTOR: 215 UNIT: umol/l
METHOD: E-ENDPOINT
TEMP: 37℃ DECIMAL: 1
WAVE1: 578 NM
REAGENT BLANK: N SAMPLE BLANK: Y
GEN MAN WOMAN CHILD
HIGH: 17.1 17.1 17.1 17.1
LOW: 5.1 5.1 5.1 5.1
LINEAR: < 342 DELAY: 5S
INTERAL: 5 S S-POINT: 1
ASP TIME: 250 STDNUM: 0
TEST NO: 1
NAME: ALT
FACTOR: 1746 UNIT:U/L
METHOD: K-KINETIC
TEMP:37℃ DECIMAL: 1
WAVE1:340nm
REAGENT BLANK: Y SAMPLE BLANK: N
GEN MAN WOMAN CHILD
HIGH: 40 40 40 40
LOW: 0 0 0 0
LINEAR:< 600 DELAY: 60 S
INTERAL: 60 S S-POINTS: 2
ASP TIME: 250 STDNUM: 0
18
SK3002 CHEMISTRY MANUAL
TEST NO: 11
NAME: CRE
FACTOR: 133 UNIT:umol/L
METHOD: K-KINETIC
TEMP: 37℃ DECIMAL: 0
WAVE1:510nm
REAGENT BLANK: Y SAMPLE BLANK: N
GEN MAN WOMAN CHILD
HIGH: 97 97 80 80
LOW: 53 53 44 44
LINEAR:< 884 DELAY: 20 S
INTERAL: 40 S S-POINTS: 2
ASP TIME: 250 STDNUM: 0
May see test project total, according to ‘1' then the system will restore leaves the
plant establishment, system preservation any modification will all lose, after ‘1', the
system could remind once more, if wanted to continue, including pressed 2 time ‘Y ' key,
pressed other key to return to the level contact surface. The instrument leaves the plant
when solidified has 48 test projects, the project number and the name is in turn as
follows:
19
SK3002 CHEMISTRY MANUAL
3. Delete TEST
The project which does not need might be deleted using this function, if the patient
result has this project, please delete it first, otherwise can appear the wrong prompt.
20
SK3002 CHEMISTRY MANUAL
1 – CHOOSE QC TEST
2 – CHOOSE QC NO
3 – PRINT QC CUVE
4 – PRINT QC RESULT
Operates this contact surface in turn 4 TESTs, then prints the quality control report:
1, After choose ‘1- CHOOSES QC PROJECT’, the system prompt is as follows:
TEST NO:
Please input the QC project code, this code must be the test project which
already existed,after input the QC project code, the system prompt was as follows:
Please input the target value and the deviation scope of the QC TEST;
If the TEST no is not exist,display as follows:
TEST NO:
CODE NOT EXIST
RETURN TO EXIT ENTER TO RETRY
The QC code is the same as the patient code, must be already existed, the qc
control result and the patient result treats impartially, stored up together, managed
together, when arrange the project the qc control code and the patient code not be
the same.
3, completes the above 1,2 operations, may print the curves of the qc control thing
using ‘3'key, using ‘4' key may print the qc control statistical result: The average
value, the standard deviation .
21
SK3002 CHEMISTRY MANUAL
Use the key “ ↑ ”、“ ↓ ”can set the temperature by 0.1℃step ,use the key‘←’‘→’can
set the temperature by 1℃step,current temp is the device temperature.
You can press “9” aspirate the volume of the sample,press”8” aspirate the volume of the
cleaner, though aspirating the quantity of liquid volume can find the appropriate amount.
The SK3002 biochemistry analyzer recommend the quantity of the liquid aspiration
might be bigger than 450ul, the clean time can be clean well the pipeline
6.3 ADC:
The operator will use this function in the routine maintenance. Chooses ‘2-ADC’ in
22
SK3002 CHEMISTRY MANUAL
WAVE (0-7): 0
With ‘ ← ’‘ → ’key may change the different wave length, this wave length
extinction value demonstrates immediately. Press “6” can open the light, Press ‘8’ key
aspirate water, after five or more minutes can observe the absorbance of each
wavelength.
WAVE:340 RESULT
WAVE:340 RESULT 2.357
-0.063
WAVE (0-7): 0
WAVE (0-7): 0
Press ‘8’ key aspirate liquid according to the clean quantity, press ‘9’ key aspirate
liquid according to the sample liquid aspiration. Using this function may inspect the
extinction of each wave length, also available in daily pipeline clean work.
1— PRINT CONTENT
2— PRINT MODE
3— PRINT FONT
a. Printing content:
After chooses ‘1 - print all result', the system will print all results of the same patient
NO.
Choose the ‘2 - print today result', the system prints the patient's all results today.
The instrument power off one time, the system regards as one day. For example test
result has not printed in the morning , power off the instrument ,in the afternoon want to
print again , the instrument cannot regard as the morning result the same day result,
this time have to print please choose ‘1- print all result', it can print.
b. Print mode:
1—CLOSE AUTO PRINT
2—PRINT EACH PATIENT
3—PRINT EACH SAMPLE
PRINT CONTENT: XXX
RETURN TO EXIT
XXX is OFF,2, 3
① If will not want to print the report in the test process, please establish the print mode
‘OFF';
23
SK3002 CHEMISTRY MANUAL
② If want to print after test the all project of the patient automatically, please establish
the print mode ‘2';
③ If want to print after one project has been test every time, please establish the print
mode ‘3'.
C. Print capital or lower case:
Choose prints the capital or the lowercase according to needs, the system
demonstration is as follows.
1—CAPITAL
2—LOWERCASE
PRINT FONT : 1
RETURN TO EXIT
CHECK ABS
ASPIRATE DISTILLED WATER
Please defer to the screen prompt operation, the absorbance examination is in order
to cause the system determination different time the extinction change, then adjusts,
causes the result to be more reliable, also may avoid each test will have to test the
blank and the standard. The absorbance of each wave length must be between
±0.15,otherwise need to adjust or should to remove the air bubble.
A
A:The system preservation patient counts, scope: 1 - 255; B: The system examines
actually the patient database patient counts.
A and B should be equal. If is not equal, system when starting can appear the above
prompt dialog box, chooses ‘1- modify patient sum' cursor to appear in A place,
afterwards inputs and the B place same numeral. Next time starting will not be able
again to appear the above dialog box.
Through pressing ‘1' or ‘2' completes the establishment, the system does not have
any prompt, if wants to know the establishment is correct, may enter the establishment
time&date, the viewing time demonstration.
1 SYSTEM DEFAULT
2 RESET PATIENT
RETURN TO EXIT
Please act according to the special details choice, if wants to reposition the system,
please press ‘1', then according to ‘Y' key 2 times, if only must eliminate the patient
material, please press ‘2', then according to ‘Y' 2 times then; If does not want to
reposition, please press ‘RETURN ' key, after the replacement all patient material and
the result loses, therefore please use this function carefully.
25
SK3002 CHEMISTRY MANUAL
1.Starting
① Please insert the power line into the grounding power source plug, Turns on the
power switch which is behind the instrument.
② After let aspirate tube into the distilled water, presses down `ENTER' key to
aspirate the distilled water ,then the instrument will be able to detect the extinction of
0-5 wave length, after the examination complete, the instrument will be able to enter
the main menu automatically.
③ Waiting for 5 minutes after starting ,the instrument preheat
2. Operation
According to the prompt of the instrument you can test the sample.
3. Clean pipeline
The operation finished ,you can closes down after use the distilled water clean
pipeline
26
SK3002 CHEMISTRY MANUAL
This instrumentation use liquid crystal display monitor, as weather change, the light
and shade contrast can change, when instrument each operation is all normal, the
display monitor was not clearly, may adjust ‘contrast gradient adjusting knob’ until it
display clear.
: Dangerous pressure, in
instrument internal power unit
labeling
: ‘ brittle goods’, in instrument
outside wrapping box marking, when
transport should be careful puts lightly.
28
SK3002 CHEMISTRY MANUAL
: ‘ Upward’ symbol, in instrument outside wrapping box marking, when transport the
correct position is the graphical representation direction vertical upward
: ‘ Rainproof’ symbol, in instrument outside wrapping box marking, this package of
supports feared the rain drenches
:‘Piles the code layer limit’, in instrument outside wrapping box marking, the same
package support biggest pile of code limit is 4
29
SK3002 CHEMISTRY MANUAL
30