Professional Documents
Culture Documents
10
10
10
TABLE I
C OMPARISON W ITH O THER LC S ENSORS
shown in Fig. 7(b). The gap between the blue squares (without
LHM) and red dots (with LHM) is the signal strength differ-
ence received at the readout terminal. The enhancing effect
stays almost consistent during all the frequency spectrum of
the sensor. Frequency shift, which is due to frequency splitting
as analyzed in [10] and [11], does not happen here. Hence,
the measured sensitivity of the system with an LHM slab is
kept around −34.67 kHz/%RH.
According to Fig. 5, the real part of effective permeability
of the LHM slab has a value of −1 at 35.4 MHz, and the
point is also marked in Fig. 7(b). It is worth to notice that the
signal strength difference becomes great near Re(μe f f ) = −1.
This means the metamaterial with Re(μe f f ) = −1 has a better
Fig. 7. Humidity sensor interrogation with an LHM slab. (a) The frequency enhancement performance due to less dielectric loss and mag-
response of the LC humidity sensor with an LHM slab. The inset is the netic loss. Hence, setting the frequency with Re(μe f f ) = −1
measurement setup. (b) The resonance points with and without LHM. of the LHM at the center of the frequency spectrum of LC
sensor maintains a harmonious enhancement in the whole
working range of the sensor.
is maintained unchanged, the readout distance of the sensors
with/without a metamaterial slab is compared to show the IV. D ISCUSSIONS AND C ONCLUSION
improvement. There have been several methods by using a relay/repeater
The humidity sensor is first monitored as shown in Fig. 6(a). to enhance the readout. A comparison between the LC sensors
The humidity level is controlled by a humidity test chamber in this paper with and without a metamaterial slab and other
OMEGA 205. The relative humidity is varied from 15 %RH LC humidity sensors in [11] and [12] are listed in Table I when
to 90 %RH (25 °C). The distance between the sensor and the humidity is 15%RH. To show the exact influences brought
the readout coil is 7mm. Fig. 6(b) gives the frequency by the relay/repeater, the comparison has only adopted several
response of the LC humidity sensor with the sensitivity as results that are measured in the same circumstances, including
−34.67 kHz/%RH. the inductor size, testing environment, the Q factor of LC
An LHM slab is then added between the LC sensor and sensor, and the signal received. From the series of research
the readout coil. The measurement setup is shown as the inset work in Table I, it is obvious that the LHM slab is not the one
in Fig. 7(a). The distance between the LHM and the readout with optimal performance in readout enhancement, but it is the
coil is 5 mm, and the distance between the LHM and the sensor energy-efficient one. No extra power is required in the system
is 7.5 mm. The distance is designed to keep the S11 value when LHM is applied. While other techniques demand a power
closet to the former data that are without LHM slab. Hence, supply or a readout equipment with transmitting power. The
the readout distance is extended to 12.5 mm after adding an application of LHM slab can no doubt simplify the operating
LHM slab. The large peak at 30 MHz belongs to the LHM system by omitting the energy supply.
slab. Due to the relative location, the LHM slab is closer to Another merit of the LHM slab is that there is no frequency
the readout coil than the LC sensor, hence the signal strength splitting to concern when utilized in LC sensors. Fig. 7(b)
(S11 value) is much stronger than the sensor’s. shows that the resonant frequency of LC sensor does not
Now the total distance between the readout coil and the LC change after the LHM slab is inserted. This is the basic
sensor is increased, while the signal strength (magnitude of reason that the sensitivity can be remained the same. However,
S11) is kept unchanged. To have a detailed view of the S11 other kinds of repeaters [11], [12] have to deal with the
values in the two situations mentioned above, the resonance frequency splitting problem by setting the critical distance
point of each curve is then extracted to have a comparison between the repeater and the sensor. The intrinsic nature of
between the received signal with and without an LHM as the frequency splitting is the mutual coupling between the
DONG et al.: APPLYING METAMATERIAL-BASED REPEATER IN LC PASSIVE WIRELESS SENSORS TO ENHANCE READOUT 1759
repeater and the sensor coil. This is avoided in the LHM the inductance is reduced. However, the capacitance has an
system because the enhancement of LHM is realized by its upper limit (the capacitance is so much larger than the induc-
negative permeability not the mutual coupling with the LC tance that the whole element appears capacitive, then the LC
sensor. Hence, the position of LHM slab has an amount of tank cannot work as a resonance), which means the working
flexibility. frequency is unable to be lowered down immoderately by just
The readout distance talked about in this paper indicates increasing the capacitance. This means the size of unit cell
the distance between the readout coil and the LC sensor, cannot be diminished with ease. Hence, the design of the LHM
which is widely used in researches that study a repeater/relay. slab should be appropriate not only in the aspect of size, but
In addition to this readout distance, the separation between also in the aspect of working frequency.
LHM slab and the LC sensor is another focus of attention, From the experiments results in Fig. 7, it can be deduced
because it has the practical significance when the LC sensor that the point where Re(μe f f ) = −1 is critical in this design
is inserted in some sealed environments. In this situation, when the LHM is used in the LC sensor system. When both
the distance between the repeater and the sensor is more permittivity ε = −1 and permeability μ = −1 are satisfied,
essential when the repeater is stamped onto the coverage. From then the fraction index of a medium can be expressed as [16]
Fig. 7(a), it seems that the distance between the LHM and the √
n = εμ = −1, (8)
LC sensor is about half of the total readout distance. Actually,
the position of the LHM does not have be the middle point, and where a negative square root should be chosen when both ε
the accessible range of the location is varied from the 1/3 to and μ are negative. The medium is a perfect match to free
2/3 of the whole readout distance. For example, the whole space and the interfaces show no reflection. The medium can
readout distance is 12.5 mm in Fig. 7(a), then the distance refocus and cancel the decay of evanescent waves (both far
between the sensor and the LHM is varied around from field and near field). The evanescent waves emerge from the far
4.2 mm to 8.3 mm. In this range, the enhanced performance side of the medium enhanced in amplitude by the transmission
is maintained. This extreme distance (8.3 mm in this paper) process. Hence, an LHM slab giving μ = −1 at an appropriate
helps extend the separation to some extent. frequency can focus sources of magnetic fields into sharp
Adding more unit cells in the LHM slab is conducive to images.
the further enhancement since it will cover a larger space. The result of the experiments has an improvement differ-
But a large LHM slab is not compatible in the LC sensor ent from other points, though it is not very dramatic. It is
system since the sensor size has been decreased. Hence it is because at this point the medium is a perfect match to free
unadvisable to enlarge the whole size of LHM slab at random. space and the interface shows no reflection. The metamaterial
For an LHM slab with the outer diameter fixed, the enhancing with a larger absolute value of μe f f (such as Re(u e f f ) =
performance depends on the fraction volume. The fraction −160) can also achieve evanescent wave amplification. How-
volume of the cell occupied by the interior two-layer inductor ever, this causes higher dielectric loss and magnetic loss in
can be described as metamaterials.
r2 The metamaterial slab is used as a separated element in the
F= , (5) system. The best position is along the axis of the sensor and
(r + s)2
the reader to achieve the best performance. Hence, it will not
where r is the diameter of the inductor, and r is the separation be integrated with the reader design. In practical applications,
between adjacent inductors. The effective permeability and the this slab is used to enhance the readout without any wired-
magnetic plasma frequency ωmp can be expressed as [14] connections. For example, when this approach is applied to the
Fω2 LC pressure sensor, the LHM slab can be integrated with the
μe f f = 1 − (6) glasses worn by the patient when a readout device interrogates
ω2 − ω02 + i γ ω the LC sensor implanted in eyes for intraocular pressure.
ω0
ωmp = √ , (7) In addition, in the paint spray booth of some factories, wired-
1− F connections and power supply (which are the components of
where ω0 is the original resonant frequency of the double-layer the readout device) are not allowed to appear to avoid electric
inductor and γ is the damping factor. The magnetic plasma sparks. In this situation, a wireless passive repeater is essential
frequency is a point where the effective permeability μe f f ends to extend the readout distance.
in a non-negative value, which determines the negative band. In conclusion, a left-handed metamaterial is utilized to
If the fraction volume F can be maintained unchanged when enhance the readout of the LC passive wireless sensor. The
adding more unit cells in the LHM slab, then the effective working frequency of the left-handed metamaterial is designed
permeability μe f f can keep the same by using a larger external to be adjustable to adapt to the frequency spectrum of LC
capacitor to adjust the original resonant frequency (since the sensors by connecting a discrete capacitor. An LC humidity
inductance is decreased). It is worth to mention that a large sensor is measured both with and without the LHM slab. The
F can cause a big absolute value of μe f f and a broad band interrogating distance is extended over 1.8 times when the
of negative part. Hence, when scaling down the unit cells, signal strength is kept the same with the same sensitivity.
the fraction volume should be maintained as large as possible. The measurement results indicate that the enhancement per-
It is workable to downsize the unit cell, then a larger formance becomes better when the real relative permeability
capacitance is needed to adjust the working frequency since is around −1.
1760 IEEE SENSORS JOURNAL, VOL. 18, NO. 4, FEBRUARY 15, 2018